510(k) Data Aggregation

    K Number
    K243455

    Validate with FDA (Live)

    Date Cleared
    2025-07-31

    (266 days)

    Product Code
    Regulation Number
    866.3981
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    DEN200031

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The cobas® Respiratory 4-flex for use on the cobas® 5800/6800/8800 Systems is an automated, multiplex, nucleic acid test that utilizes real-time polymerase chain reaction (PCR) technology intended for simultaneous in vitro qualitative detection and differentiation of severe acute respiratory syndrome coronavirus (SARS-CoV-2), influenza A virus, influenza B virus, and respiratory syncytial virus (RSV) in nasopharyngeal swab specimens obtained from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2, influenza A, influenza B and RSV can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza A, influenza B, and RSV infections in humans and is not intended to detect influenza C virus infections.

    Nucleic acids from the viral organisms identified by this test are generally detectable in nasopharyngeal swab specimens during the acute phase of infection. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative of the presence of the identified virus, and aid in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings.

    The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Negative results do not preclude SARS-CoV-2, influenza A virus, influenza B virus, or RSV infections. Conversely, positive results do not rule out coinfection with other organisms, and the agent(s) detected by the cobas® Respiratory 4-flex for use on the cobas® 5800/6800/8800 Systems may not be the definite cause of disease.

    Device Description

    cobas® Respiratory 4-flex for use on the cobas® 5800/6800/8800 Systems (cobas® Respiratory 4-flex) is based on fully automated sample preparation (nucleic acid extraction and purification) followed by PCR amplification and detection. The cobas® 5800 System is designed as one integrated instrument. The cobas® 6800/8800 Systems consist of the sample supply module, the transfer module, the processing module, and the analytic module. Automated data management is performed by the cobas® 5800 System or cobas® 6800/8800 Systems software(s), which assigns results for all tests. Results can be reviewed directly on the system screen and printed as a report.

    Nucleic acid from patient samples and added Internal Control RNA (RNA IC) molecules are simultaneously extracted. Nucleic acid is released by addition of proteinase and lysis reagent to the sample. The released nucleic acid binds to the silica surface of the added magnetic glass particles. Unbound substances and impurities, such as denatured protein, cellular debris and potential PCR inhibitors, are removed with subsequent wash steps and purified nucleic acid is eluted from the magnetic glass particles with elution buffer at elevated temperature. External controls (positive and negative) are processed in the same way.

    Selective amplification of target nucleic acid from the sample is achieved by the use of target-specific forward and reverse primers detecting conserved viral genome regions as shown in Table 1.

    Selective amplification of RNA IC is achieved by the use of non-competitive, sequence specific forward and reverse primers, which have no homology with the viral-target specific genomes. Amplified target is detected by the cleavage of fluorescently labeled oligonucleotide probes. Roche's temperature assisted generation of signal (TAGS) technology, short TAGS technology, is introduced to differentiate up to three targets per fluorescence channel, enabling the detection of up to14 targets, and the Internal Control, per well. A thermostable DNA polymerase enzyme is used for amplification.

    Multiplicity of target detection is enabled with temperature-dependent quenching of cleaved fluorescent target-specific probes. This is achieved by separating signals from probes into introduced thermal channels, where fluorescence is acquired at two additional fixed temperatures for each amplification cycle.

    During the PCR amplification step, hybridization of the probes to the specific single-stranded DNA template results in cleavage of the probe by the 5' to 3' exonuclease activity of the DNA polymerase, resulting in separation of the reporter and quencher dyes, and the generation of a fluorescent signal. Conventional probes release fluorescence signal immediately upon separation of reporter from quencher. TAGS probes rely on temperature dependent fluorescence activation, requiring both nuclease cleavage during the extension phase, as well as an increase in reaction temperature, to activate the otherwise dormant fluorophore. For this reason, during each PCR cycle the TAGS technology captures fluorescence in five available fluorescence channels in combination with three thermal channels (detection of fluorescence at three defined temperatures T1, T2 and T3).

    The cobas® Respiratory 4-flex master mix contains detection probes which are specific for influenza A virus, influenza B virus, RSV, SARS-CoV-2 and the RNA Internal Control (RNA IC) nucleic acid, which enables simultaneous detection and differentiation of influenza A virus, influenza B virus, RSV, and SARS-CoV-2 viral targets and the RNA IC.

    The master mix includes deoxyuridine triphosphate (dUTP), instead of deoxythymidine triphosphate (dTTP), which is incorporated into the newly synthesized DNA (amplicon). Any contaminating amplicons from previous PCR runs are destroyed by the AmpErase enzyme [uracil-N-glycosylase], which is included in the PCR mix, when heated in the first thermal cycling step. However, newly formed amplicons are not destroyed since the AmpErase enzyme is inactivated once exposed to temperatures above 55°C.

    The RESP-4FLEX ASAP enables the system to differentiate and report the qualitative results of the four targets influenza A virus, influenza B virus, RSV and SARS-CoV-2. For each specimen the customer can test for any combination of the four enabled virus targets. Also, additional target calculation (digital reflex) can be ordered for the four enabled virus targets (influenza A virus, influenza B virus, RSV and SARS-CoV-2) on the cobas® 5800 System.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study detailed in the provided FDA clearance letter for the cobas Respiratory 4-flex, structured according to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA clearance letter does not explicitly state pre-defined acceptance criteria for the clinical performance. Instead, it reports the observed performance metrics (PPA and NPA) from the clinical studies. For the purpose of this table, I will present the reported clinical performance as the "met acceptance criteria," assuming these values were deemed acceptable by the FDA for clearance.

    AnalyteAcceptance Criteria (Reported PPA CI)Device Performance (PPA %)Acceptance Criteria (Reported NPA CI)Device Performance (NPA %)
    Influenza A
    Fresh Pros.(90.5%, 98.5%)96.2(99.3%, 99.9%)99.7
    Frozen Pros.(89.0%, 99.1%)96.8(99.4%, 99.8%)99.7
    Combined Pros.(92.3%, 98.3%)96.4(99.5%, 99.8%)99.7
    Retrospective(94.1%, 100.0%)100.0(97.6%, 99.4%)98.8
    Influenza B
    Fresh Pros.(94.5%, 100.0%)100.0(99.6%, 100.0%)99.9
    Frozen Pros.(79.8%, 99.3%)95.8(99.8%, 100.0%)100.0
    Combined Pros.(94.0%, 99.8%)98.9(99.8%, 100.0%)99.9
    Retrospective(87.1%, 99.6%)97.5(98.3%, 99.7%)99.3
    RSV
    Fresh Pros.(77.4%, 94.7%)88.7(99.8%, 100.0%)100.0
    Frozen Pros.(81.5%, 95.3%)90.4(99.8%, 100.0%)100.0
    Combined Pros.(83.1%, 93.9%)89.7(99.9%, 100.0%)100.0
    Retrospective(96.4%, 100.0%)100.0(98.2%, 99.7%)99.3
    SARS-CoV-2
    Fresh Pros.(92.4%, 98.6%)96.7(98.0%, 99.1%)98.6
    Frozen Pros.(95.3%, 98.9%)97.7(97.8%, 98.8%)98.4
    Combined Pros.(95.4%, 98.5%)97.3(98.0%, 98.8%)98.5

    Study Proving Device Meets Criteria: The detailed clinical performance evaluation described in "5. CLINICAL PERFORMANCE EVALUATION" (pages 23-26) demonstrates that the cobas Respiratory 4-flex achieves the reported Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) values against a U.S. FDA-cleared molecular comparator assay. These reported percentage agreements, along with their 95% confidence intervals, serve as the evidence that the device meets the performance expectations for clinical use.

    2. Sample Size Used for the Test Set and Data Provenance

    • Prospective Clinical Study:

      • Total NPS specimens enrolled: 4,475
      • Total NPS specimens tested: 4,378 (1,832 fresh, 2,546 frozen)
      • Total NPS specimens evaluable: 4,341 (1,827 fresh, 2,514 frozen)
      • Data Provenance: Fresh prospective specimens collected at eleven collection sites during the 2023-2024 respiratory viral season. Frozen prospective specimens collected during parts of the 2022-2023 respiratory viral season at seven sites and the 2023-2024 respiratory viral season from 14 collection sites. The specific country of origin is not explicitly stated but implies U.S. based on the "U.S. testing sites" mention for retrospective samples, and the FDA clearance context. The data is prospective.
    • Retrospective Clinical Study:

      • Total NPS specimens enrolled: 770
      • Total NPS specimens evaluable:
        • Influenza A: 657
        • Influenza B: 647
        • RSV: 659
      • Data Provenance: Archived NPS specimens collected between 2014 and 2022 from individuals with signs and symptoms of respiratory viral infection. Tested at three (3) U.S. testing sites. The data is retrospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not mention the use of human experts to establish ground truth for the test set. The ground truth was established by a "U.S. FDA-cleared molecular assay" as the comparator method.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method involving experts for discrepant results. Instead, it states that an "FDA 510(k) cleared comparator" method was used to establish the "ground truth." For the influenza A target in the prospective study, "three (3) additional specimens (two (2) fresh and one (1) frozen) were excluded from analysis due to inconclusive results obtained from the comparator test," implying a reliance on the comparator's definitive result rather than external adjudication. The same pattern is noted for retrospective samples, with exclusions for failed or invalid comparator test results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Human Readers' Improvement with AI vs. Without AI Assistance

    This section is Not Applicable to the provided document. The cobas Respiratory 4-flex is an automated, multiplex nucleic acid detection test, not an AI-powered diagnostic tool requiring human reader interpretation or assistance. Therefore, an MRMC study or analysis of human reader improvement with AI is not relevant.

    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, this was a standalone performance evaluation. The device is an automated in vitro diagnostic (IVD) test. The clinical performance evaluation directly compares the cobas Respiratory 4-flex's results (algorithm only, as it's an automated system) against a predicate FDA-cleared molecular assay. There is no human interpretation or intervention in the generation of the primary test result from the cobas system itself.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The ground truth for the clinical performance evaluation was established by a "U.S. FDA-cleared molecular assay." This means another legally marketed and validated molecular diagnostic test was used as the reference standard.

    8. The Sample Size for the Training Set

    The document does not specify a separate training set or its sample size. This is typical for in vitro diagnostic (IVD) devices that use established laboratory techniques (like PCR) rather than machine learning algorithms which require explicit training data. The development of such assays involves extensive analytical validation (LoD, precision, specificity, inclusivity, etc.) and then clinical validation with independent samples.

    9. How the Ground Truth for the Training Set Was Established

    Since no explicit training set for a machine learning algorithm is mentioned, the concept of "ground truth for the training set" is not applicable in the context of this device's validation as described. The analytical studies (LoD, inclusivity, specificity, etc.) characterize the inherent performance of the assay's chemical and hardware components.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241806

    Validate with FDA (Live)

    Date Cleared
    2025-01-08

    (201 days)

    Product Code
    Regulation Number
    866.3981
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    BioFire® Respiratory Panel 2.1 (DEN200031)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel is a multiplex, real-time reverse transcription polymerase chain reaction (RT-PCR) in vitro diagnostic test for the qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza B virus, and respiratory syncytial virus (RSV). Nucleic acids are isolated and purified from nasopharyngeal (NP) swab and anterior nasal (AN) swab specimens obtained from individuals exhibiting signs and symptoms of a respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2, influenza, and RSV can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza B, and RSV A/B (undifferentiated) infections in humans and is not intended to detect influenza C virus infections.

    Nucleic acids from the viral organisms identified by this test are generally detectable in NP and AN swab specimens during the acute phase of infection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infective of the presence of the identified virus and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings.

    The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organism(s) detected by the Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus, influenza B virus, or RSV infections.

    The Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel is intended for use by qualified and trained clinical laboratory personnel specifically instructed and trained in the techniques of real-time PCR and in vitro diagnostic procedures.

    Device Description

    The Applied Biosystems™ TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel is a multiplex, real-time reverse transcription polymerase chain reaction (RT-PCR) test. The severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A, influenza B, respiratory syncytial virus (RSV) A/B and RNase P primer and probe sets are designed to detect viral RNA in nasopharyngeal (NP) and anterior nasal (AN) swab specimens from individuals exhibiting signs and symptoms of a respiratory tract infection.

    Each TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel includes the following components:

    • TaqPath™ COVID-19, Flu A, Flu B, RSV Select Assay-Multiplex assays that contain . primer and probe sets specific to the following targets:
      • Three SARS-CoV-2 targets (Orfla, Orf1b, and N genes) .
      • . Two influenza A virus targets (PB1 & M genes)
        • I Two influenza B virus targets (M & NS genes)
        • . Three RSV targets (NP, M, and L protein genes)
        • l RNase P (internal human sample collection control)
    • TaqPath™ COVID-19, Flu A, Flu B, RSV Select Positive Control—Inactivated viral . control that contains SARS-CoV-2, influenza A, influenza B, and RSV.
    • TaqPath™ COVID-19, Flu A, Flu B, RSV Select Negative Control—MS2 packaged RNA . control that contains targets specific to RNase P genomic regions targeted by the assay.
    • TaqPath™ 1-Step Select Master Mix (No ROX)—Ready-to-use PCR mix, including . reverse transcriptase, polymerase, dNTPs, salts, and buffer.
    • . Package Insert -- Provides the instructions and the link to download the instructions for use and other assets (including the ADF)
    • An Assay Definition File (ADF) applicable to the instrument used in the workflow . (available via download).

    In addition to the SARS-CoV-2, influenza A, influenza B and RSV viral assay targets, the assay portion of the panel includes RNase P, which serves as an endogenous internal process control to monitor extraction and amplification of each clinical sample. The TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel also contains external process positive and negative controls. The positive control (PC) component included is an inactivated viral control that contains SARS-CoV-2, influenza A, influenza B, and RSV viruses. The PC monitors extraction and real-time RT-PCR by demonstrating that each of the four viruses can be detected when present and that RNase P is not detected when absent. The negative control (NC) component included is an MS2 packaged RNA control that contains targets specific to RNase P genomic regions targeted by the assay. The NC also monitors extraction and real-time RT-PCR by demonstrating RNase P can be detected when present and that the four viruses are not detected when absent. The TaqPath™ 1-Step Select Master Mix (No ROX) included as a component of the kit is a ready-to-use PCR mix which contains a deoxyribonucleotide triphosphate mix (dNTPs), enzymes, and other components to permit reverse transcription and amplification of the assay targets. The TagPath™ 1-Step Select Master Mix (No ROX) also contains ribonuclease (RNase) inhibitors as well as deoxyuridine triphosphate (dUTP) and uracil N-glycosylase (known as UNG or UDG).

    The TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel is provided in two overall kit configurations: either 200 reactions or 1.000 reactions.

    AI/ML Overview

    The provided text describes the analytical and clinical performance studies of the "Applied Biosystems TaqPath COVID-19, Flu A, Flu B, RSV Select Panel" (hereafter referred to as "the Device"). This device is an in vitro diagnostic test for the qualitative detection and differentiation of SARS-CoV-2, influenza A, influenza B, and respiratory syncytial virus (RSV).

    Below is a breakdown of the acceptance criteria and study details based on the provided information, presented in the requested format.


    Acceptance Criteria and Device Performance

    The acceptance criteria for this diagnostic device are primarily demonstrated through various analytical performance evaluations (Limit of Detection, Inclusivity/Reactivity, Precision, Interfering Substances, Cross-Reactivity, Stability, Carryover) and clinical performance studies (PPA and NPA against a comparator device). The studies aim to show that the device performs as expected and is "substantially equivalent" to a legally marketed predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Study Aspect / Performance MetricAcceptance Criteria (Implied/Directly Stated where available)Reported Device Performance (Summary)
    Limit of Detection (LoD)≥ 95% positive detection at the determined LoD concentrations.All viruses confirmed at ≥ 95% positive detection at the determined LoD concentrations (e.g., SARS-CoV-2 at 107 GCE/mL; Flu A H1N1 at 340 GCE/mL). WHO Standard LoD for SARS-CoV-2 confirmed at 150.36 IU/mL.
    Inclusivity (in silico)Less than 1% (SARS-CoV-2) and less than 5% (Flu A, Flu B, and RSV) of publicly available sequences have mutations that reduce melting temperature below annealing temperature for all gene targets.> 99.99% of SARS-CoV-2, 99.9% of Flu B, and 99.49% of RSV strains met the criteria. For Flu A, 95.27% of strains were considered reactive after wet testing.
    Reactivity (wet-lab)100% detection at approximately 3x LoD.Most tested strains showed 100% detection at approximately 3x LoD. Strains not at 100% initially were further tested until 100% detection was achieved at higher concentrations.
    Precision (Within-Laboratory)N/A (Statistical measures like %CV and %PPA are reported).Positivity Rate (Negative Sample): 0%.
    Overall PPA (1x LoD, 3x LoD): 100.00% for all targets.
    Observed Precision and Repeatability %CV (1x and 3x LoD): < 2%CV for all targets.
    Site-to-Site ReproducibilityQualitative PPA performance: Generally high (e.g., above 90-95% at 1x LoD, 100% at 3x LoD). Quantitative %CV for Cq variability: Generally low (e.g., usually < 5%).NPA (Negative Sample): 99.63%.
    Overall Site-to-Site Reproducibility PPA (1x LoD): 92.59% (SARS-CoV-2) to 98.89% (Flu A).
    Overall Reproducibility PPA (3x LoD): 100.00% for all targets.
    Observed Reproducibility, Within-lab precision, Repeatability %CV (1x & 3x LoD): < 5% CV for all targets.
    Interfering SubstancesAll positive samples remain positive; negative samples remain negative. No false positives (FP) or false negatives (FN).All conditions produced expected results. No interference observed from 26 substances, 2 solvents, 5 VTMs, or extraction reagent carryover, except for expected positive results from FluMist® due to live influenza components.
    Competitive Interference100% positivity for both low-concentration and high-concentration viruses in multi-target samples. No off-target false positives.All conditions satisfied acceptance criteria, showing 100% positivity for both viruses in pairs and no off-target false positives.
    Cross-Reactivity & Microbial InterferenceNo false positive results from tested microbes in negative samples. No inhibition of PCR or interference with extraction/RNase P detection in positive samples with microbes.None of the 49 tested microorganisms produced false positive results or interfered with detection. One of 39 CoV-HKU1 clinical samples produced a positive SARS-CoV-2 result, the rest were negative.
    Specimen Stability100% positivity for all targets at specified storage conditions and time points.100% positivity for NP and AN specimens at all time points tested (RT for 24h, refrigerated for 72h, ≤ -70°C for 30 days).
    Eluate Stability100% positivity for contrived positive sample eluates at specified conditions.100% of contrived positive sample eluates produced expected results for up to 24 hours at 2-8°C, 5 days at ≤ -70°C, and after one freeze/thaw cycle.
    Fresh vs Frozen EquivalencySupport the use of frozen samples in clinical studies and stated freeze/thaw cycles in IFU.All samples showed 100% concordance with expected results, supporting use of frozen samples and specific F/T cycles (up to 2 F/T for Copan/BD, 1 F/T for others).
    In-Use and Hold Time StabilityAll components meet specified stability criteria for storage conditions and freeze/thaw cycles.Assay Mix Tube and Master Mix stable for 72h at 2-8°C and up to 4 F/T cycles. PC and NC stable for 24h at 2-8°C.
    Carryover Cross-ContaminationCarryover/cross-contamination rate of 0%.0% carryover/cross-contamination rate reported among 492 replicates tested.
    RNase P Internal Control Cutoff ConfirmationOverall PPA of 98.07% (NPA 100%) for NP swabs and 92.30% (NPA 100%) for AN swabs. Improved AN swab collection technique reduces invalid rate and shows 100% RNase P detection.NP PPA: 98.07%, NPA: 100%. AN PPA: 92.30%, NPA: 100%. Subsequent study with improved AN collection showed 100% RNase P detection and 100% validity for AN samples.
    Matrix Equivalency100% concordance between NP and AN swab specimens.All samples produced 100% concordance with expected results.
    Clinical Performance (Prospective Study)Demonstrate substantial equivalence based on PPA and NPA compared to a comparator device. Target typically high PPA/NPA (e.g., >85-90%).SARS-CoV-2: NP Overall PPA 96.6%, NPA 96.47%; AN Overall PPA 95.3%, NPA 96.46%.
    Flu A: NP Overall PPA 94.9%, NPA 99.36%; AN Overall PPA 97.8%, NPA 99.06%.
    Flu B: NP Overall PPA 98.0%, NPA 99.87%; AN Overall PPA 100.0%, NPA 99.87%.
    RSV: NP Overall PPA 93.8%, NPA 99.75%; AN Overall PPA 100.0%, NPA 99.6%.
    Clinical Performance (Enrichment Study)Maintain high PPA/NPA for Flu A, Flu B, RSV (where prospective numbers might be low).Flu A: NP Overall PPA 97.9%, NPA 100.0%; AN Overall PPA 100.0%, NPA 95.5%.
    Flu B: NP Overall PPA 92.3%, NPA 98.2%; AN Overall PPA 92.3%, NPA 100.0%.
    RSV: NP Overall PPA 100.0%, NPA 94.9%; AN Overall PPA 100.0%, NPA 96.6%.

    2. Sample Size Used for the Test Set and Data Provenance

    • Clinical Performance Study (Prospective Cohort):

      • Subjects: 1,909 subjects enrolled (1,840 prospectively, 69 enrichment).
      • Test Set (Evaluable Specimens): 1,620 Nasopharyngeal (NP) swabs and 1,541 Anterior Nasal (AN) swabs were included in the final data analysis for the prospective cohort. (Minor adjustments for inconclusive comparator results in Flu A, specifically 1,539 AN and 1,616 NP for Flu A analysis).
      • Enrichment Study: 69 subjects, yielding 69 NP swabs and 68 AN swabs for analysis.
      • Data Provenance:
        • Country of Origin: Not explicitly stated as a single country, but "14 different collection sites across the U.S." indicate United States data.
        • Retrospective or Prospective: Primarily prospective data, with an "enrichment phase" also involving prospective collection from subjects with confirmed positive PCR results for Flu A, Flu B, and/or RSV.
        • Specimen Handling: Specimens were tested either fresh (Category I) or frozen (Category II).
    • Analytical Studies (Test Sets):

      • LoD Confirmation: At least 24 replicates per virus strain/specimen type.
      • Precision (within-lab): 96 replicates for each panel member.
      • Reproducibility (site-to-site): 9 replicates per panel member (3 sites x 3 reagent lots x 1 replicate/lot). Total ~270 replicates per analyte level across 3 sites.
      • Interfering Substances: 3 replicates per substance/condition.
      • Competitive Interference: 3 replicates per condition.
      • Cross-Reactivity & Microbial Interference: 3 replicates per organism/condition.
      • Specimen Stability: Multiple replicates per time point/storage condition.
      • Eluate Stability: Multiple replicates per time point/storage condition.
      • Fresh vs Frozen Equivalency: Multiple replicates per F/T cycle and concentration.
      • In-Use and Hold Time Stability: Multiple replicates per time point/F/T cycle.
      • Carryover Cross-Contamination: 492 replicates (negative wells).
      • RNase P Internal Control Cutoff Confirmation: Individually collected NP and AN swabs (specific number not given for the initial confirmation, but the follow-up AN study used 138 AN samples).
      • Matrix Equivalency: Multiple replicates per matrix and concentration.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • For the clinical performance study, the ground truth appears to be established by a comparator FDA-cleared molecular test.
    • The document does not specify the number or qualifications of human experts (e.g., radiologists, pathologists) involved in establishing the ground truth for any of the studies. This is typical for molecular diagnostic assays where the "ground truth" is often defined by the results of a highly sensitive and specific reference method (e.g., another molecular test or a composite reference method incorporating multiple tests or clinical findings).

    4. Adjudication Method for the Test Set

    • For the clinical performance study, samples that produced a discordant call between the Device and the comparator test were further tested with "another FDA cleared molecular assay" (a resolver assay). This implies a reconciliation method where a third test acts as an adjudicator for discordant results between the device under evaluation and the primary comparator.
    • The specific rules for how the resolver assay's result determined the final ground truth (e.g., 2+1 majority rule, or if the resolver result was the final truth) are not explicitly detailed but are common in diagnostic test evaluation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not done. The evaluated device is a molecular diagnostic test (RT-PCR kit), not an imaging AI algorithm that typically involves human readers. The performance is assessed based on the device's ability to detect viral nucleic acids against a comparator molecular test, not on improvements in human reader performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Yes, the primary evaluation of this device is a standalone performance study. The "TaqPath™ COVID-19, Flu A, Flu B, RSV Select Panel" is a laboratory-based RT-PCR diagnostic kit. Its performance, as detailed in the analytical and clinical studies, is measured by its output (detection/differentiation of viral targets) directly from patient samples, independent of human interpretation or "human-in-the-loop" assistance for the test result itself. The Diomni™ Software provides automated interpretation, making it an algorithm-only result at the interpretation stage.

    7. The Type of Ground Truth Used

    • For analytical studies (e.g., LoD, Inclusivity, Precision, Interference, Cross-Reactivity, Stability), the ground truth was established by contrived samples with known concentrations of viral material or known presence/absence of interfering/cross-reactive substances. This is a common and appropriate method for analytical validation.
    • For the clinical performance study, the ground truth was established by an FDA-cleared molecular test (comparator device), supplemented by a resolver FDA-cleared molecular assay for discordant results. This constitutes a form of "composite reference standard" where the truth is determined by a highly reliable, already approved diagnostic method.

    8. The Sample Size for the Training Set

    • The document describes the validation of a diagnostic kit, not an AI model requiring a separate "training set" in the machine learning sense. The information provided heavily details the test set performance (analytical and clinical validation).
    • While the "Diomni™ Software" performs data analysis and interpretation based on "assay-specific parameters from the Assay Definition File (ADF)," the document does not specify a distinct training set size typically associated with machine learning model development. The "training" of such a system would typically involve internal development data used to establish parameters like Cq cutoffs, but this is not detailed as a separate "training set" with specific sample numbers. The development study that determined the RNase P IC cutoff of Cq = 33.0 could be considered part of the "training" or parameter establishment phase, but no sample size is given for it.

    9. How the Ground Truth for the Training Set Was Established

    • As noted above, a distinct "training set" for an AI/algorithm is not explicitly described in the context of this diagnostic kit's submission. The "ground truth" for establishing device parameters (like Cq cutoffs) would generally come from extensive analytical experiments and internal validation studies using characterized samples (e.g., known concentrations of viruses, negative controls).
    • For example, the preliminary LoD determined by probit analysis and the RNase P IC cutoff determined in a development study are examples of how internal parameters were established. These processes rely on highly controlled experiments with well-defined inputs (the "ground truth" being the known composition of the contrived samples).