Search Results
Found 31 results
510(k) Data Aggregation
(252 days)
The VELO Respiratory Test is an automated rapid multiplex real-time, reverse transcriptase polymerase chain reaction (RT-PCR) test performed on the VELO Instrument and is intended for the simultaneous qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A, and influenza B virus nucleic acid in anterior nasal swab (ANS) specimens from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2 and influenza can be similar.
The VELO Respiratory Test is intended for use as an aid in the differential diagnosis of SARS-CoV-2, influenza A, and/or influenza B infection if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A and influenza B viral nucleic acid are generally detectable in ANS specimens during the acute phase of infection.
Positive results do not rule out co-infection with other organisms. The agent(s) detected by the VELO Respiratory Test may not be the definite cause of disease.
Negative results do not preclude SARS-CoV-2, influenza A, and/or influenza B infection. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.
The VELO System is comprised of a single-use VELO Respiratory Test, and a reusable VELO Instrument:
- The Test contains individually packaged consumables: a "Test Cartridge" and an anterior nasal swab.
- The Instrument is a small, benchtop device with preloaded software for running tests and viewing the results. For a full description of the VELO Instrument, please refer to the VELO Instrument Manual.
Each Test Cartridge contains all necessary reagents for the detection of Flu A, Flu B and SARS-CoV-2 viral RNA from anterior nasal swab specimens. Primers and probes in the VELO Respiratory Test are designed to amplify and detect unique sequences within the following regions of each target pathogen: influenza A genome (matrix protein gene), influenza B genome (non-structural protein gene), and SARS-CoV-2 genome (ORF 1a/b non-structural region, and membrane protein gene). Each Test Cartridge also contains an endogenous Sample and Process Control (SPC). The SPC acts as an Internal Control (IC), to control for adequate sample collection and processing, the detection of failures in the reaction resulting from PCR inhibition, or a failure of the reagents. The SPC also ensures that the RT-PCR reaction conditions (temperature and time) are appropriate for the amplification reaction and that the resulting signal can be detected and processed by the system. Test outcomes are reported to the operator in real-time via the Instrument viewscreen with 'Not Detected' results available in under 10 minutes, when all cycles have completed. When the Test ends, all results can be viewed via the Instrument viewscreen, and the Test Cartridge may be removed for disposal.
N/A
Ask a specific question about this device
(179 days)
The LIAISON NES FLU A/B, RSV & COVID-19 assay is a real-time RT-PCR assay intended for use on the LIAISON NES instrument for the simultaneous in vitro qualitative detection and differentiation of nucleic acid from severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A (Flu A) virus, influenza B (Flu B) virus, and respiratory syncytial virus (RSV) in anterior nasal swab specimens from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, influenza B, and RSV can be similar.
The LIAISON NES FLU A/B, RSV & COVID-19 assay is intended for use as an aid in the differential diagnosis of SARS-CoV-2, influenza A, influenza B, and RSV infections when used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A, influenza B, and RSV viral RNA are generally detectable in anterior nasal swab specimens during the acute phase of infection. This test is not intended to detect influenza C virus infections.
Positive results are indicative of the presence of the identified virus, but do not rule out bacterial infection or co-infection with other pathogens not detected by the test. The agent(s) detected by the LIAISON NES FLU A/B, RSV & COVID-19 assay may not be the definite cause of the disease. Negative results do not preclude SARS-CoV-2, influenza A, influenza B or RSV infection and should not be used as the sole basis for patient management decisions.
The LIAISON NES Instrument (NES1001) is capable of analysis of a single cartridge containing a single specimen. A set of parameters specific to the assay is included in the instrument software to name target molecules, assign dyes to probes, specify cycling conditions, and to analyze data from runs. Fluorescence intensity is monitored at each PCR cycle by detection modules within the instrument. The instrument software controls the thermocycling and, upon completion of the run, automatically interprets and displays results for the specimen.
The LIAISON NES Instrument is comprised of the following:
- Touchscreen User Interface
- Status LED Indicator
- Audio Speaker
- Barcode Scanner
The LIAISON NES software is a graphical user interface (GUI) application that is the end-user interface to the LIAISON NES Instrument. The software is installed in an embedded computer. The LIAISON NES software is responsible for providing the environment in which a user runs assays and obtains results.
The LIAISON NES instrument is intended to accept a cartridge, containing either a quality control (QC) or patient sample, to process and detect for the target nucleic acid.
The LIAISON NES FLU A/B, RSV & COVID-19 assay used on the LIAISON NES instrument is a real-time RT-PCR system that enables the direct amplification, detection, and differentiation of influenza A viral RNA, influenza B viral RNA, RSV RNA, and SARS-CoV-2 RNA directly from nasal swabs. Nasal swabs can be professionally collected by a healthcare provider or self-collected under the healthcare provider's supervision.
The LIAISON NES FLU A/B, RSV & COVID-19 assay consists of the LIAISON NES instrument, the LIAISON NES FLU A/B, RSV & COVID-19 Cartridge containing all the required PCR reagents, the NES Sample Vial containing sample release buffer, and the NES Swab for sample collection.
In the LIAISON NES FLU A/B, RSV & COVID-19 assay, fluorescent probes are used with corresponding forward and reverse primers to amplify influenza A, influenza B, RSV, and SARS-CoV-2, and internal control (IC) RNA. Conserved regions of influenza A viral RNA, influenza B viral RNA, RSV RNA, and SARS-CoV-2 RNA are targeted to identify the viruses in the specimen, while the internal control (IC) RNA is used to detect any PCR failures and/or inhibition.
N/A
Ask a specific question about this device
(91 days)
The Simplexa COVID-19 / Flu A/B & RSV Direct is a real-time RT-PCR assay intended for use on the LIAISON MDX instrument for the simultaneous in vitro qualitative detection and differentiation of nucleic acid from severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A (Flu A) virus, influenza B (Flu B) virus and respiratory syncytial virus (RSV) in nasopharyngeal swab and anterior nasal swab specimens from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, influenza B, and RSV can be similar.
The Simplexa COVID-19 / Flu A/B & RSV Direct assay is intended for use as an aid in the differential diagnosis of SARS-CoV-2, influenza A, influenza B and RSV infections if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A, influenza B, and RSV viral RNA are generally detectable in nasopharyngeal swab and anterior nasal swab specimens during the acute phase of infection. This test is not intended to detect influenza C virus infections.
Positive results are indicative of the presence of the identified virus, but do not rule out bacterial infection or co-infection with other pathogens not detected by the test. The agent(s) detected by the Simplexa COVID-19 / Flu A/B & RSV Direct real-time RT-PCR assay may not be the definite cause of the disease. Negative results do not preclude SARS-CoV-2, influenza A, influenza B, or RSV infection and should not be used as the sole basis for patient management decisions.
The Simplexa™ COVID-19 & Flu A/B & RSV Direct assay is a qualitative, multiplex real-time reverse transcription polymerase chain reaction (RT-PCR) test intended for the simultaneous detection and differentiation of RNA from SARS-CoV-2, Influenza A, Influenza B, and Respiratory Syncytial Virus (RSV) in nasopharyngeal swabs (NPS) and anterior nasal swabs (NS) in UTM/UVT and M4RT specimen transport media. The assay is performed on the LIAISON® MDX Instrument using a Direct Amplification Disc (DAD) format, enabling sample-to-answer processing without separate nucleic acid extraction.
The LIAISON® MDX Instrument is a benchtop real-time PCR thermocycler that utilizes a self-contained, single-use direct amplification disc (DAD) to process samples. It performs thermal cycling and real-time fluorescence detection using optical detection modules, each with specific excitation and emission wavelengths. The instrument includes a laser enclosed in a laser product housing, with integrated hardware and software interlocks to ensure user safety. It is operated via a USB connection to a dedicated computer running the LIAISON® MDX Studio software.
The LIAISON® MDX Studio software controls the instrument and provides a user interface for assay setup, execution, and result analysis. The software automatically interprets results for in vitro diagnostic (IVD) assays using pre-defined assay definitions encoded in barcode inserts included with the assay kits. It performs spectral compensation, verifies internal control amplification, and checks for sufficient sample volume prior to amplification. The software also includes user authentication, audit logging, laboratory information system (LIS) connectivity, and cybersecurity features.
The assay kit includes single-use reaction mix vials, a positive control pack with inactivated viral particles in transport media, and the Direct Amplification Disc consumable, which supports up to eight simultaneous reactions.
The assay format is designed for direct amplification, with 24 single-use reaction mix vials per kit. The required sample volume input is 50 µL. The reaction mix is provided in single-use vials and includes DNA polymerase, reverse transcriptase, RNase inhibitor, primers, probes, and encapsulated RNA templates. The buffer component in the reaction mix maintains optimal pH and ionic strength to support enzyme activity and amplification efficiency throughout the RT-PCR process.
The assay includes an encapsulated RNA internal control (RNA IC) in each reaction to monitor for potential RT-PCR inhibition or process failure. The RNA IC is derived from bacteriophage MS2. This non-target RNA is co-amplified with the assay's viral targets and detected independently using post-amplification melting curve analysis. The presence of the RNA IC in a negative specimen confirms that the amplification process functioned as expected, while its absence—along with no target detection—results in an invalid outcome. Detection of the RNA IC is not required in the Positive Control but is expected in the No Template Control (NTC) to verify assay validity.
N/A
Ask a specific question about this device
(226 days)
QIAstat-Dx Respiratory Panel Plus:
The QIAstat-Dx Respiratory Panel Plus is a multiplexed nucleic acid test intended for use with the QIAstat-Dx system for the simultaneous in vitro qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals with clinical signs and symptoms of respiratory tract infections, including Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2).
The following organism types and subtypes are identified using the QIAstat-Dx Respiratory Panel Plus: Adenovirus, Human Coronavirus 229E, Human Coronavirus HKU1, Human Coronavirus NL63, Human Coronavirus OC43, Human Metapneumovirus, Influenza A, Influenza A H1, Influenza A H1N1 pdm09, Influenza A H3, Influenza B, Parainfluenza Virus 1, Parainfluenza Virus 2, Parainfluenza Virus 3, Parainfluenza Virus 4, Respiratory Syncytial Virus, Human Rhinovirus/Enterovirus (not differentiated), SARS-CoV-2, Bordetella pertussis, Chlamydophila pneumoniae, and Mycoplasma pneumoniae.
Nucleic acids from viral and bacterial organisms identified by this test are generally detectable in NPS specimens during the acute phase of infection. Detecting and identifying specific viral and bacterial nucleic acids from individuals presenting with signs and symptoms of a respiratory infection aids in the diagnosis of respiratory infection, if used in conjunction with other clinical, epidemiological and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment or other patient management decisions.
Negative results in the presence of a respiratory illness may be due to infection with pathogens that are not detected by the test, or due to lower respiratory tract infection that is not detected by a NPS specimen.
Conversely, positive results are indicative of the presence of the identified microorganism, but do not rule out co-infection with other pathogens not detected by the QIAstat-Dx Respiratory Panel Plus. The agent(s) detected by the QIAstat-Dx Respiratory Panel Plus may not be the definite cause of disease.
The use of additional laboratory testing (e.g., bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection.
QIAstat-Dx Respiratory Panel Mini:
The QIAstat-Dx Respiratory Panel Mini is a multiplexed nucleic acid test intended for use with the QIAstat-Dx system for the simultaneous in vitro qualitative detection and identification of multiple respiratory viral nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals with clinical signs and symptoms of respiratory tract infections, including Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2).
The following viruses are identified using the QIAstat-Dx Respiratory Panel Mini: Influenza A, Influenza B, Respiratory Syncytial Virus, Human Rhinovirus, and SARS-CoV-2.
Nucleic acids from viral organisms identified by this test are generally detectable in NPS specimens during the acute phase of infection. Detecting and identifying specific viral nucleic acids from individuals presenting with signs and symptoms of a respiratory infection aids in the diagnosis of respiratory infection, if used in conjunction with other clinical, epidemiological and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment or other patient management decisions.
Negative results in the presence of a respiratory illness may be due to infection with pathogens that are not detected by the test or due to lower respiratory tract infection that is not detected by a NPS specimen.
Conversely, positive results are indicative of the presence of the identified microorganism, but do not rule out co-infection with other pathogens not detected by the QIAstat-Dx Respiratory Panel Mini. The agent(s) detected by the QIAstat-Dx Respiratory Panel Mini may not be the definite cause of disease.
The use of additional laboratory testing (e.g., bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection.
The QIAstat-Dx Respiratory Panel Plus and the QIAstat-Dx Respiratory Panel Mini are multiplexed nucleic acid tests which are designed for use with the QIAstat-Dx system (currently QIAstat-Dx Analyzer 1.0 and QIAstat-Dx Analyzer 2.0). The device modification is to add the QIAstat-Dx Rise as an additional instrument for use with the QIAstat-Dx Respiratory Panel Plus and the QIAstat-Dx Respiratory Panel Mini ("QIAstat-Dx Respiratory Panels"). The QIAstat-Dx Rise is a higher throughput platform, incorporating up to eight QIAstat-Dx Analytical Modules (AM) on a small footprint. The instrument allows queuing up to 18 cartridges, which are scheduled for processing and delivered to the appropriate AM by an integrated robotic handler. The AM used with the QIAstat-Dx Rise is the same AM that can be used with the QIAstat-Dx Analyzer 1.0 or 2.0.
The modified QIAstat-Dx Respiratory Panel Plus and QIAstat-Dx Respiratory Panel Mini are identical to the QIAstat-Dx Respiratory Panel Plus (K233100) and the QIAstat-Dx Respiratory Panel Mini (K242353), respectively, with the exception of the Instructions for Use which were updated to include the assay-specific procedure for the QIAstat-Dx Rise.
The QIAstat-Dx Respiratory Panels are intended to be used with one nasopharyngeal swab (NPS) eluted in Universal Transport Media (UTM), which is not provided with the QIAstat-Dx Respiratory Panels.
All the reagents required for the complete execution of the test are pre-loaded and self-contained in a QIAstat-Dx Respiratory Panel cartridge. The user does not need to manipulate any reagents. During the test, reagents are handled by pneumatically-operated microfluidics without any direct contact with the user or the analyzer actuators.
Within the cartridge, multiple steps are automatically performed in sequence by using pneumatic pressure and a multiport valve to transfer the sample and fluids via the Transfer Chamber (TC) to their intended destinations. Following the introduction of the sample from a disposable transfer pipette, the following assay steps occur automatically and sequentially:
- Resuspension of Internal Control
- Cell lysis using mechanical and/or chemical means
- Membrane-based nucleic acid purification
- Mixing of the purified nucleic acid with lyophilized master mix reagents
- Transfer of defined aliquots of eluate/master mix to different reaction chambers
- Performance of multiplex real-time RT-PCR testing within each reaction chamber
The QIAstat-Dx Respiratory Panel Assay Definition File (ADF) automatically interprets test results and displays a summary on the instrument display screen. The detected analytes are displayed in red. All other tested but not detected analytes are listed in green. The instrument will report if an error occurs during processing, in which case the test will fail and no results will be provided (screen will show "FAIL").
The provided text describes a 510(k) premarket notification for the QIAstat-Dx Respiratory Panel Plus and QIAstat-Dx Respiratory Panel Mini, with a modification to include the QIAstat-Dx Rise instrument. The key takeaway from this document is that the FDA determined the device is substantially equivalent to previously cleared devices. Therefore, the "acceptance criteria" discussed here refer to the demonstration of equivalence to a predicate device, rather than specific performance metrics against a clinical ground truth for a new device.
Here's an analysis based on your questions:
1. A table of acceptance criteria and the reported device performance
Since this is a submission for a modification to an already cleared device, the acceptance criteria are not explicitly stated in terms of clinical performance numbers (e.g., sensitivity, specificity). Instead, the acceptance criteria are focused on demonstrating that adding the new instrument (QIAstat-Dx Rise) does not negatively impact the performance, and that the new system is "substantially equivalent" to the predicate devices.
The "reported device performance" is essentially that the studies "successfully demonstrated the equivalent performance."
| Acceptance Criteria (Implied for Substantial Equivalence) | Reported Device Performance |
|---|---|
| Equivalence at Low Analyte Concentration | Successfully demonstrated equivalent performance |
| Carryover | Successfully demonstrated equivalent performance |
| Reproducibility | Successfully demonstrated equivalent performance |
| Maintenance of original Intended Use/Indications for Use | Maintained the same Intended Use/Indications for Use as predicate devices |
| Maintenance of device technology (specimen type, amplification/detection, controls, extraction, assay targets, operational aspects) | Maintained all technological characteristics as predicate devices |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify the sample size for the test set used in the "Equivalence at Low Analyte Concentration," "Carryover," or "Reproducibility" studies. It also does not mention the data provenance (country of origin, retrospective or prospective). These details would typically be found in the actual study reports, which are not included in this FDA clearance letter.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable or provided in this document. For an in vitro diagnostic device like this, ground truth for clinical performance would typically be established by comparing against FDA-cleared or gold standard laboratory methods (e.g., culture, sequencing, or other highly sensitive PCR assays) rather than expert consensus on imaging or clinical findings. Since this submission focuses on establishing equivalence and not initial clinical performance, such details are not expected.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not provided and would not typically be part of a 510(k) clearance letter for an IVD device unless specific clinical adjudication was required for complex diagnostic outcomes.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The QIAstat-Dx Respiratory Panels are automated in vitro diagnostic devices for detecting nucleic acids. They do not involve human readers' interpretation of images or other data in a way that would necessitate an MRMC study or AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is an in vitro diagnostic test. It is inherently a "standalone" system in its operation, as the instrument performs the test and provides a result. There is no mention of a human-in-the-loop component beyond loading the sample and reading the final result from the display screen. The device's "algorithm" (i.e., the assay's detection mechanism and interpretation software) operates without human intervention once the run starts.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The document does not explicitly state the "ground truth" used for the analytical studies (Equivalence at Low Analyte Concentration, Carryover, Reproducibility). For such studies, the ground truth would be established by controlled laboratory experiments, where samples with known concentrations of analytes are used, and the assay's results are compared against these known inputs. For initial clinical performance, the ground truth would typically be a highly sensitive and specific reference method, but those studies are for the predicate devices, not this modification.
8. The sample size for the training set
This information is not provided. Training sets are typically associated with machine learning or AI models. While instruments like the QIAstat-Dx have underlying algorithms, they are based on established PCR principles and assay design, not a machine learning training paradigm in the way AI image analysis would be.
9. How the ground truth for the training set was established
This information is not provided and is not applicable for this type of IVD device in the context of a 510(k) modification for instrument compatibility.
Ask a specific question about this device
(272 days)
The BIOFIRE SPOTFIRE Respiratory/Sore Throat (R/ST) Panel Mini is an automated multiplexed polymerase chain reaction (PCR) test intended for use with the BIOFIRE SPOTFIRE System for the simultaneous, qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swab (NPS) or anterior nasal swab (ANS) specimens obtained from individuals with signs and symptoms of respiratory tract infection, including COVID-19; (Respiratory menu) or in throat swab (TS) specimens from individuals with signs and symptoms of pharyngitis; (Sore Throat menu).
The following analytes are identified and differentiated using the BIOFIRE SPOTFIRE R/ST Panel Mini:
Respiratory Menu
Viruses
- Coronavirus SARS-CoV-2
- Human rhinovirus
- Influenza A virus
- Influenza B virus
- Respiratory syncytial virus
Sore Throat Menu
Viruses
- Human rhinovirus
- Influenza A virus
- Influenza B virus
- Respiratory syncytial virus
Bacteria
- Streptococcus pyogenes (group A Strep)
Nucleic acids from the viral and bacterial organisms identified by this test are generally detectable in NPS/ANS/TS specimens during the acute phase of infection. The detection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and symptoms of respiratory infection and/or pharyngitis is indicative of the presence of the identified microorganism and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.
Negative results in the setting of a respiratory illness and/or pharyngitis may be due to infection with pathogens that are not detected by this test, or a respiratory tract infection that may not be detected by an NPS, ANS, or TS specimen. Positive results do not rule out co-infection with other organisms. The agent(s) detected by the BIOFIRE SPOTFIRE R/ST Panel Mini may not be the definite cause of disease.
Additional laboratory testing (e.g., bacterial and viral culture, immunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection and/or pharyngitis.
The BIOFIRE SPOTFIRE R/ST Panel Mini (SPOTFIRE R/ST Panel Mini) simultaneously identifies 5 different respiratory viral pathogens in nasopharyngeal swabs (NPS) or anterior nasal swabs (ANS), or 5 different viral and bacterial pharyngitis pathogens in throat swabs (TS) from individuals with signs and symptoms of respiratory tract infections or pharyngitis, respectively, (see Table 1). The SPOTFIRE R/ST Panel Mini is compatible with the BIOFIRE® SPOTFIRE® System, a polymerase chain reaction (PCR)-based in vitro diagnostic system for infectious disease testing. The BIOFIRE SPOTFIRE System Software executes the SPOTFIRE R/ST Panel Mini test and interprets and reports the test results. The SPOTFIRE R/ST Panel Mini was designed to be used in CLIA-waived environments.
A test is initiated by loading Hydration Solution into the hydration solution injection port of the SPOTFIRE R/ST Panel Mini pouch and NPS, ANS, or TS specimen, mixed with the provided Sample Buffer, into the other sample injection port of the SPOTFIRE R/ST Panel Mini pouch and placing it in the SPOTFIRE System. The pouch contains all the reagents required for specimen testing and analysis in a freeze-dried format; the addition of Hydration Solution and Sample/Buffer Mix rehydrates the reagents. After the pouch is prepared, the SPOTFIRE System Software guides the user through the steps of placing the pouch into the instrument, scanning the pouch barcode, entering the sample identification, and initiating the run.
Nucleic acid extraction occurs within the SPOTFIRE R/ST Panel Mini pouch using mechanical and chemical lysis followed by purification using standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, the SPOTFIRE System performs a nested multiplex PCR that is executed in two stages. During the first stage, the SPOTFIRE System performs a single, large volume, highly multiplexed reverse transcription PCR (rt-PCR) reaction. The products from first stage PCR are then diluted and combined with a fresh, primer-free master mix and a fluorescent double-stranded DNA binding dye (LC Green® Plus, BioFire Diagnostics). The solution is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The 2nd stage PCR, or nested PCR, is performed in singleplex fashion in each well of the array. At the conclusion of the 2nd stage PCR, the array is interrogated by melt curve analysis for the detection of signature amplicons denoting the presence of specific targets. A digital camera placed in front of the 2nd stage PCR captures fluorescent images of the PCR reactions and software interprets the data.
The SPOTFIRE System Software automatically interprets the results of each DNA melt curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the SPOTFIRE R/ST Panel Mini.
The FDA 510(k) clearance letter details the acceptance criteria and study that proves the BIOFIRE SPOTFIRE Respiratory/Sore Throat Panel Mini meets these criteria, specifically for the addition of Anterior Nasal Swabs (ANS) as a sample type for the Respiratory Menu.
Here's the breakdown:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the reported performance metrics (Positive Percent Agreement - PPA and Negative Percent Agreement - NPA) in the clinical study. The device is deemed to meet these criteria if the lower bound of the 95% Confidence Interval (95% CI) for PPA and NPA is above acceptable thresholds (though specific numerical thresholds for "acceptable" are not explicitly stated as separate criteria, the observed high performance and clearance imply they were met).
| SPOTFIRE R/ST Panel Mini R Menu Analyte | Performance Metric | Reported Performance (Prospective Study) |
|---|---|---|
| Coronavirus SARS-CoV-2 | Positive Percent Agreement (PPA) | 96.2% (95% CI: 87.0-98.9%) |
| Negative Percent Agreement (NPA) | 99.6% (95% CI: 98.8-99.9%) | |
| Human rhinovirus | Positive Percent Agreement (PPA) | 95.7% (95% CI: 92.2-97.6%) |
| Negative Percent Agreement (NPA) | 95.0% (95% CI: 92.9-96.5%) | |
| Influenza A virus | Positive Percent Agreement (PPA) | 94.3% (95% CI: 84.6-98.1%) |
| Negative Percent Agreement (NPA) | 100% (95% CI: 99.5-100%) | |
| Influenza B virus | Positive Percent Agreement (PPA) | 100% (95% CI: 77.2-100%) |
| Negative Percent Agreement (NPA) | 100% (95% CI: 99.5-100%) | |
| Respiratory syncytial virus | Positive Percent Agreement (PPA) | 95.0% (95% CI: 83.5-98.6%) |
| Negative Percent Agreement (NPA) | 99.9% (95% CI: 99.3-100%) |
Archived Specimen Performance for Influenza B virus:
| Analyte | Performance Metric | Reported Performance (Archived Study) |
|---|---|---|
| Influenza B virus | Positive Percent Agreement (PPA) | 100% (95% CI: 90.1-100%) |
| Negative Percent Agreement (NPA) | 100% (95% CI: 98.2-100%) |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size:
- Clinical Performance (Prospective Study): 797 specimens (out of 820 initially enrolled, 23 excluded).
- Archived Specimen Testing: 241 specimens for Influenza B virus (35 positive, 206 negative).
- Data Provenance:
- Country of Origin: US (prospective multi-center study at five geographically distinct urgent care or emergency department study sites).
- Retrospective/Prospective: The study was primarily prospective, conducted from March 2024 to February 2025. This was supplemented with archived specimens for Influenza B due to low prevalence in the prospective study.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not explicitly state the number or qualifications of experts used to establish the ground truth. It states that the performance was evaluated by comparing the test results with those from a "commercially available FDA-cleared multiplexed respiratory pathogen panel." This suggests that the ground truth was established by the results of this comparator method, which themselves would have been validated. No human expert interpretation of the comparator method is described.
4. Adjudication Method for the Test Set
The document mentions "Investigations of discrepant results are summarized in the footnotes." These footnotes indicate that for discrepant cases (e.g., false positives, false negatives), "additional molecular methods" were used to re-test the specimens. This implies a form of post-hoc adjudication using a more definitive or orthogonal molecular method to resolve discrepancies between the SPOTFIRE R/ST Panel Mini and the initial comparator.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. This study is for a diagnostic PCR test, not an AI-assisted imaging or interpretation device. Therefore, an MRMC study and analysis of human reader improvement with AI assistance are not applicable.
6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done
Yes. The study evaluates the performance of the BIOFIRE SPOTFIRE R/ST Panel Mini as a standalone diagnostic device. The results are automatically interpreted and reported by the system software, with no human interpretation step in the primary analysis flow. The study compares the device's output directly to the comparator method.
7. The Type of Ground Truth Used
The primary ground truth was established by a commercially available FDA-cleared multiplexed respiratory pathogen panel. For discrepant results, "additional molecular methods" were used for confirmatory testing, indicating a molecular gold standard approach.
8. The Sample Size for the Training Set
The document does not provide details about a training set size. This notice is a 510(k) clearance for a PCR-based in vitro diagnostic test, not a machine learning or AI algorithm in the traditional sense that requires distinct training and test sets in the same manner. The "test set" described is the clinical validation cohort for demonstrating performance. PCR assays are generally developed and optimized through laboratory analytical studies, not typically "trained" on large datasets in the way an AI model would be.
9. How the Ground Truth for the Training Set Was Established
As noted above, the concept of a "training set ground truth" is not applicable in this context, as the device is a PCR assay. The development and optimization of such assays involve different molecular and analytical validation processes to ensure specificity and sensitivity.
Ask a specific question about this device
(266 days)
The cobas® Respiratory 4-flex for use on the cobas® 5800/6800/8800 Systems is an automated, multiplex, nucleic acid test that utilizes real-time polymerase chain reaction (PCR) technology intended for simultaneous in vitro qualitative detection and differentiation of severe acute respiratory syndrome coronavirus (SARS-CoV-2), influenza A virus, influenza B virus, and respiratory syncytial virus (RSV) in nasopharyngeal swab specimens obtained from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2, influenza A, influenza B and RSV can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza A, influenza B, and RSV infections in humans and is not intended to detect influenza C virus infections.
Nucleic acids from the viral organisms identified by this test are generally detectable in nasopharyngeal swab specimens during the acute phase of infection. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative of the presence of the identified virus, and aid in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings.
The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Negative results do not preclude SARS-CoV-2, influenza A virus, influenza B virus, or RSV infections. Conversely, positive results do not rule out coinfection with other organisms, and the agent(s) detected by the cobas® Respiratory 4-flex for use on the cobas® 5800/6800/8800 Systems may not be the definite cause of disease.
cobas® Respiratory 4-flex for use on the cobas® 5800/6800/8800 Systems (cobas® Respiratory 4-flex) is based on fully automated sample preparation (nucleic acid extraction and purification) followed by PCR amplification and detection. The cobas® 5800 System is designed as one integrated instrument. The cobas® 6800/8800 Systems consist of the sample supply module, the transfer module, the processing module, and the analytic module. Automated data management is performed by the cobas® 5800 System or cobas® 6800/8800 Systems software(s), which assigns results for all tests. Results can be reviewed directly on the system screen and printed as a report.
Nucleic acid from patient samples and added Internal Control RNA (RNA IC) molecules are simultaneously extracted. Nucleic acid is released by addition of proteinase and lysis reagent to the sample. The released nucleic acid binds to the silica surface of the added magnetic glass particles. Unbound substances and impurities, such as denatured protein, cellular debris and potential PCR inhibitors, are removed with subsequent wash steps and purified nucleic acid is eluted from the magnetic glass particles with elution buffer at elevated temperature. External controls (positive and negative) are processed in the same way.
Selective amplification of target nucleic acid from the sample is achieved by the use of target-specific forward and reverse primers detecting conserved viral genome regions as shown in Table 1.
Selective amplification of RNA IC is achieved by the use of non-competitive, sequence specific forward and reverse primers, which have no homology with the viral-target specific genomes. Amplified target is detected by the cleavage of fluorescently labeled oligonucleotide probes. Roche's temperature assisted generation of signal (TAGS) technology, short TAGS technology, is introduced to differentiate up to three targets per fluorescence channel, enabling the detection of up to14 targets, and the Internal Control, per well. A thermostable DNA polymerase enzyme is used for amplification.
Multiplicity of target detection is enabled with temperature-dependent quenching of cleaved fluorescent target-specific probes. This is achieved by separating signals from probes into introduced thermal channels, where fluorescence is acquired at two additional fixed temperatures for each amplification cycle.
During the PCR amplification step, hybridization of the probes to the specific single-stranded DNA template results in cleavage of the probe by the 5' to 3' exonuclease activity of the DNA polymerase, resulting in separation of the reporter and quencher dyes, and the generation of a fluorescent signal. Conventional probes release fluorescence signal immediately upon separation of reporter from quencher. TAGS probes rely on temperature dependent fluorescence activation, requiring both nuclease cleavage during the extension phase, as well as an increase in reaction temperature, to activate the otherwise dormant fluorophore. For this reason, during each PCR cycle the TAGS technology captures fluorescence in five available fluorescence channels in combination with three thermal channels (detection of fluorescence at three defined temperatures T1, T2 and T3).
The cobas® Respiratory 4-flex master mix contains detection probes which are specific for influenza A virus, influenza B virus, RSV, SARS-CoV-2 and the RNA Internal Control (RNA IC) nucleic acid, which enables simultaneous detection and differentiation of influenza A virus, influenza B virus, RSV, and SARS-CoV-2 viral targets and the RNA IC.
The master mix includes deoxyuridine triphosphate (dUTP), instead of deoxythymidine triphosphate (dTTP), which is incorporated into the newly synthesized DNA (amplicon). Any contaminating amplicons from previous PCR runs are destroyed by the AmpErase enzyme [uracil-N-glycosylase], which is included in the PCR mix, when heated in the first thermal cycling step. However, newly formed amplicons are not destroyed since the AmpErase enzyme is inactivated once exposed to temperatures above 55°C.
The RESP-4FLEX ASAP enables the system to differentiate and report the qualitative results of the four targets influenza A virus, influenza B virus, RSV and SARS-CoV-2. For each specimen the customer can test for any combination of the four enabled virus targets. Also, additional target calculation (digital reflex) can be ordered for the four enabled virus targets (influenza A virus, influenza B virus, RSV and SARS-CoV-2) on the cobas® 5800 System.
Here's an analysis of the acceptance criteria and study detailed in the provided FDA clearance letter for the cobas Respiratory 4-flex, structured according to your request:
1. Table of Acceptance Criteria and Reported Device Performance
The FDA clearance letter does not explicitly state pre-defined acceptance criteria for the clinical performance. Instead, it reports the observed performance metrics (PPA and NPA) from the clinical studies. For the purpose of this table, I will present the reported clinical performance as the "met acceptance criteria," assuming these values were deemed acceptable by the FDA for clearance.
| Analyte | Acceptance Criteria (Reported PPA CI) | Device Performance (PPA %) | Acceptance Criteria (Reported NPA CI) | Device Performance (NPA %) |
|---|---|---|---|---|
| Influenza A | ||||
| Fresh Pros. | (90.5%, 98.5%) | 96.2 | (99.3%, 99.9%) | 99.7 |
| Frozen Pros. | (89.0%, 99.1%) | 96.8 | (99.4%, 99.8%) | 99.7 |
| Combined Pros. | (92.3%, 98.3%) | 96.4 | (99.5%, 99.8%) | 99.7 |
| Retrospective | (94.1%, 100.0%) | 100.0 | (97.6%, 99.4%) | 98.8 |
| Influenza B | ||||
| Fresh Pros. | (94.5%, 100.0%) | 100.0 | (99.6%, 100.0%) | 99.9 |
| Frozen Pros. | (79.8%, 99.3%) | 95.8 | (99.8%, 100.0%) | 100.0 |
| Combined Pros. | (94.0%, 99.8%) | 98.9 | (99.8%, 100.0%) | 99.9 |
| Retrospective | (87.1%, 99.6%) | 97.5 | (98.3%, 99.7%) | 99.3 |
| RSV | ||||
| Fresh Pros. | (77.4%, 94.7%) | 88.7 | (99.8%, 100.0%) | 100.0 |
| Frozen Pros. | (81.5%, 95.3%) | 90.4 | (99.8%, 100.0%) | 100.0 |
| Combined Pros. | (83.1%, 93.9%) | 89.7 | (99.9%, 100.0%) | 100.0 |
| Retrospective | (96.4%, 100.0%) | 100.0 | (98.2%, 99.7%) | 99.3 |
| SARS-CoV-2 | ||||
| Fresh Pros. | (92.4%, 98.6%) | 96.7 | (98.0%, 99.1%) | 98.6 |
| Frozen Pros. | (95.3%, 98.9%) | 97.7 | (97.8%, 98.8%) | 98.4 |
| Combined Pros. | (95.4%, 98.5%) | 97.3 | (98.0%, 98.8%) | 98.5 |
Study Proving Device Meets Criteria: The detailed clinical performance evaluation described in "5. CLINICAL PERFORMANCE EVALUATION" (pages 23-26) demonstrates that the cobas Respiratory 4-flex achieves the reported Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) values against a U.S. FDA-cleared molecular comparator assay. These reported percentage agreements, along with their 95% confidence intervals, serve as the evidence that the device meets the performance expectations for clinical use.
2. Sample Size Used for the Test Set and Data Provenance
-
Prospective Clinical Study:
- Total NPS specimens enrolled: 4,475
- Total NPS specimens tested: 4,378 (1,832 fresh, 2,546 frozen)
- Total NPS specimens evaluable: 4,341 (1,827 fresh, 2,514 frozen)
- Data Provenance: Fresh prospective specimens collected at eleven collection sites during the 2023-2024 respiratory viral season. Frozen prospective specimens collected during parts of the 2022-2023 respiratory viral season at seven sites and the 2023-2024 respiratory viral season from 14 collection sites. The specific country of origin is not explicitly stated but implies U.S. based on the "U.S. testing sites" mention for retrospective samples, and the FDA clearance context. The data is prospective.
-
Retrospective Clinical Study:
- Total NPS specimens enrolled: 770
- Total NPS specimens evaluable:
- Influenza A: 657
- Influenza B: 647
- RSV: 659
- Data Provenance: Archived NPS specimens collected between 2014 and 2022 from individuals with signs and symptoms of respiratory viral infection. Tested at three (3) U.S. testing sites. The data is retrospective.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not mention the use of human experts to establish ground truth for the test set. The ground truth was established by a "U.S. FDA-cleared molecular assay" as the comparator method.
4. Adjudication Method for the Test Set
The document does not describe an adjudication method involving experts for discrepant results. Instead, it states that an "FDA 510(k) cleared comparator" method was used to establish the "ground truth." For the influenza A target in the prospective study, "three (3) additional specimens (two (2) fresh and one (1) frozen) were excluded from analysis due to inconclusive results obtained from the comparator test," implying a reliance on the comparator's definitive result rather than external adjudication. The same pattern is noted for retrospective samples, with exclusions for failed or invalid comparator test results.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Human Readers' Improvement with AI vs. Without AI Assistance
This section is Not Applicable to the provided document. The cobas Respiratory 4-flex is an automated, multiplex nucleic acid detection test, not an AI-powered diagnostic tool requiring human reader interpretation or assistance. Therefore, an MRMC study or analysis of human reader improvement with AI is not relevant.
6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done
Yes, this was a standalone performance evaluation. The device is an automated in vitro diagnostic (IVD) test. The clinical performance evaluation directly compares the cobas Respiratory 4-flex's results (algorithm only, as it's an automated system) against a predicate FDA-cleared molecular assay. There is no human interpretation or intervention in the generation of the primary test result from the cobas system itself.
7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)
The ground truth for the clinical performance evaluation was established by a "U.S. FDA-cleared molecular assay." This means another legally marketed and validated molecular diagnostic test was used as the reference standard.
8. The Sample Size for the Training Set
The document does not specify a separate training set or its sample size. This is typical for in vitro diagnostic (IVD) devices that use established laboratory techniques (like PCR) rather than machine learning algorithms which require explicit training data. The development of such assays involves extensive analytical validation (LoD, precision, specificity, inclusivity, etc.) and then clinical validation with independent samples.
9. How the Ground Truth for the Training Set Was Established
Since no explicit training set for a machine learning algorithm is mentioned, the concept of "ground truth for the training set" is not applicable in the context of this device's validation as described. The analytical studies (LoD, inclusivity, specificity, etc.) characterize the inherent performance of the assay's chemical and hardware components.
Ask a specific question about this device
(30 days)
The Xpert Xpress CoV-2/Flu/RSV plus test, performed on the GeneXpert Xpress System, is an automated multiplexed real-time reverse transcriptase polymerase chain reaction (RT-PCR) test intended for use in the simultaneous in vitro qualitative detection and differentiation of severe acute respiratory syndrome coronavirus (SARS-CoV-2), influenza A, influenza B, and/or respiratory syncytial virus (RSV) viral RNA in nasopharyngeal swab and anterior nasal swab specimens collected from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, influenza B, and RSV can be similar.
The Xpert Xpress CoV-2/Flu/RSV plus test is intended for use in the differential detection of SARS-CoV-2, influenza A, influenza B and/or RSV RNA and aids in the diagnosis of COVID-19, influenza and/or RSV infections if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A, influenza B, and RSV viral RNA are generally detectable in nasopharyngeal swab and anterior nasal swab specimens during the acute phase of infection.
Positive results are indicative of the presence of the identified virus, but do not rule out bacterial infection or co-infection with other pathogens not detected by the test. The agent (s) detected by the Xpert Xpress CoV-2/Flu/RSV plus test may not be the definite cause of the disease.
Negative results do not preclude SARS-CoV-2, influenza A, influenza B and/or RSV infection. The results of this test should not be used as the sole basis for diagnosis, treatment or other patient management decisions.
The Xpert Xpress CoV-2/Flu/RSV plus test is an automated in vitro diagnostic test for the simultaneous qualitative detection and differentiation of SARS-CoV-2, Flu A, Flu B, and RSV viral RNA in nasopharyngeal swab (NPS) and anterior nasal swab (NS) specimens collected from individuals showing signs and symptoms of respiratory viral infection.
The Xpert Xpress CoV-2/Flu/RSV plus test is performed on GeneXpert Xpress System, which consist of a GeneXpert IV instrument that executes sample preparation, nucleic acid amplification and real-time fluorescent signal detection for the tests, and a GeneXpert Hub with preloaded GeneXpert Xpress software for running the tests and viewing the test results. The GeneXpert Hub accessory integrates the computer, touchscreen monitor and barcode scanner. Each of the GeneXpert modules in the GeneXpert IV instrument can perform independent sample preparation and testing. The GeneXpert Xpress System requires the use of single-use disposable cartridges that hold the RT-PCR reagents and host sample purification, nucleic acid amplification, and detection of the target sequences. Because the cartridges are self-contained, cross-contamination between samples is minimized.
The Xpert Xpress CoV-2/Flu/RSV plus test includes reagents for the detection of SARS-CoV-2, Flu A, Flu B and RSV viral RNA from NPS and NS specimens. The primers and probes in the Xpert Xpress CoV-2/Flu/RSV plus test are designed to amplify and detect unique sequences in the genes that encode the following proteins: SARS-CoV-2 nucleocapsid (N), SARS-CoV-2 envelope (E), SARS-CoV-2 RNA-dependent RNA polymerase (RdRP), influenza A matrix (M), influenza A basic polymerase (PB2), influenza A acidic protein (PA), influenza B matrix (M), influenza B non-structural protein (NS), and the RSV A and RSV B nucleocapsid.
A Sample Processing Control (SPC) and a Probe Check Control (PCC) are also included in the cartridge utilized by the GeneXpert instrument. The SPC is present to control for adequate processing of the sample and to monitor for the presence of potential inhibitor(s) in the RT-PCR reaction. The SPC also ensures that the RT-PCR reaction conditions (temperature and time) are appropriate for the amplification reaction and that the RT-PCR reagents are functional. The PCC verifies reagent rehydration, PCR tube filling, and confirms that all reaction components are present in the cartridge including monitoring for probe integrity and dye stability.
The Xpert Xpress CoV-2/Flu/RSV plus test is designed for use with NPS or NS specimens collected with nylon flocked swabs and placed into viral transport medium (VTM), Universal Transport Medium (UTM), or eNAT®. The ancillary specimen collection kits, swabs and transport media validated for use with the Xpert Xpress CoV-2/Flu/RSV plus test included:
• Nasopharyngeal Sample Collection Kit for Viruses
- Copan UTM® 3C057N (Flexible Minitip Flocked Swab with UTM® Medium without Beads)
- Copan eNAT® Molecular Collection and Preservation Medium P/N 6U074S01 (Flexible Minitap Flocked Swab with eNAT® Medium)
• Nasal Sample Collection Kit for Viruses
- Copan UTM® 3C064N (Regular Flocked Swab with UTM® Medium without Beads)
- Copan eNAT® Molecular Collection and Preservation Medium P/N 6U073S01 (Regular Flocked Swab with eNAT® Medium)
• Alternatively, swabs and transport media can be obtained separately:
- Nylon flocked swab (Copan P/N 502CS01, 503CS01)
- Viral Transport Medium, 3 mL (Copan P/N 330C, 3C047N, BD Universal Transport Medium, Remel M4RT, or Remel M5)
The ancillary reagents allow NPS and NS specimens from patients to be collected, preserved and transported to laboratory prior to analysis with the Xpert Xpress CoV 2/Flu/RSV plus test.
Based on the provided FDA 510(k) clearance letter and summary, here's an analysis of the acceptance criteria and study that proves the device meets them:
Important Note: This document describes a "Special 510(k) submission." This type of submission is used when changes are made to a previously cleared device that do not affect its fundamental technology, intended use, or safety/effectiveness. In this specific case, the changes were to the Assay Definition File (ADF) – essentially software parameter settings. Therefore, the "study that proves the device meets the acceptance criteria" largely relies on re-analysis of existing studies from the original device clearance (K242071) rather than entirely new, large-scale clinical trials.
1. Table of Acceptance Criteria and Reported Device Performance
The provided document doesn't explicitly state quantitative acceptance criteria in a dedicated table. However, since this is a Special 510(k) for software parameter changes, the primary "acceptance criteria" appear to be demonstrating non-inferiority or equivalency to the previously cleared predicate device in terms of:
- Valid Test Runs: The number of tests that yield a valid result.
- Non-Determinate (ND) Test Results: The number of tests that do not yield a definitive positive or negative result (e.g., "NO RESULT").
- Performance Claims: That the modifications did not negatively impact the overall analytical and clinical performance claims established for the predicate device (sensitivity, specificity for detecting SARS-CoV-2, Flu A, Flu B, and RSV).
Inferred Acceptance Criteria and Reported Performance:
| Acceptance Criterion (Inferred) | Reported Device Performance (with Updated ADF) |
|---|---|
| Maintain number of valid test runs (vs. predicate) | "The comparison of 16264 test results demonstrated that the numbers of valid test runs... were the same between the original ADF and the updated ADF with Flu B SLD Off." |
| Maintain number of non-determinate (ND) test results (vs. predicate) | "The comparison of 16264 test results demonstrated that... non-determinate (ND) test results were the same between the original ADF and the updated ADF with Flu B SLD Off." "All non-determinate GeneXpert ERROR test results with error code 5011 (SLD) were changed to NO RESULT." (This indicates a change in reporting for specific errors to a more appropriate "NO RESULT" category, but the number of non-determinate results remained the same.) |
| Correctly implement revised algorithm for SARS-CoV-2 only mode | "The comparison of 25 test results demonstrated that the number of valid test runs and ND test results were the same between the original and the updated ADF." (for the SARS-CoV-2 only test mode) "The revised algorithm produced the expected NO RESULT – REPEAT TEST GeneXpert test results for the Xpress SARS-CoV-2 plus test mode." (for simulated conditions) |
| No negative impact on established performance claims | "The assessment of the re-analysis results determined that the performance claims of the Xpert Xpress CoV-2/Flu/RSV plus test were not impacted by the modifications made to the predicate device." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size:
- Main analysis: 16,264 test results. This number represents the re-analysis of "verification, validation and flex studies data from the original studies."
- SARS-CoV-2 only test mode analysis: 25 test results.
- Simulated conditions for revised algorithm: The document mentions "Test cartridges were simulated to generate SARS-CoV-2 INVALID and SPC FAIL conditions," but does not explicitly state the number of simulated tests.
- Data Provenance: The data used for re-analysis originated from the "original studies" (K242071 submission). The document does not specify the country of origin or whether these original studies were retrospective or prospective, though typical clinical validation studies for IVDs are often prospective. Given it references "flex study," those can sometimes include both retrospective and prospective elements.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not provide information on the number or qualifications of experts used to establish the ground truth. This is likely because the ground truth was established during the original predicate device clearance (K242071) and this Special 510(k) focused on demonstrating equivalence in specific software-related metrics through re-analysis. For IVD devices like this, ground truth is typically established by comparing against a highly sensitive and specific reference method (e.g., an FDA-cleared laboratory developed test or a combination of clinical diagnosis and other accepted testing methods).
4. Adjudication Method for the Test Set
The document does not specify any adjudication method (e.g., 2+1, 3+1, none) for the test set. Again, this specific submission involved re-analysis of existing data rather than new clinical trials where such adjudication might be more explicitly detailed if human reads were involved. For an automated RT-PCR test, adjudication as typically understood for image-based AI would not be directly applicable.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted or is not mentioned in this document. This is an in vitro diagnostic device (RT-PCR test), not an AI-assisted diagnostic imaging system that would typically involve human readers interpreting images with or without AI assistance. The device is fully automated.
6. Standalone Performance (Algorithm Only)
Yes, the performance evaluated here is inherently standalone (algorithm only). The Xpert Xpress CoV-2/Flu/RSV plus is an automated RT-PCR test system. Its output is directly generated by the instrument and its embedded software (ADF). The "performance data" detailed in section 18.5 describes how the device's algorithmic and software changes affected its ability to produce valid results and interpret them correctly, without human intervention in the result generation.
7. Type of Ground Truth Used
The document does not explicitly state the type of ground truth used for the original studies, beyond stating "analytical, clinical and flex studies data." For RT-PCR assays, the ground truth for clinical performance is typically established by:
- Clinical Reference Method: Comparison against a highly sensitive and specific laboratory reference method (e.g., another FDA-cleared or EUA-authorized RT-PCR test, or a composite reference standard using multiple methods).
- Clinical Diagnosis/Outcomes Data (less common for purely diagnostic tests): While the device aids in diagnosis ("aids in the diagnosis of COVID-19, influenza and/or RSV infections if used in conjunction with other clinical and epidemiological information, and laboratory findings"), the performance claims themselves are based on detection of viral RNA, rather than patient outcomes as primary ground truth.
Given the nature of a molecular diagnostic test for viral RNA, the ground truth for the performance claims (sensitivity, specificity) of the original device would have been established by comparing the device's results against a highly reliable reference molecular test, often considered the "gold standard" for pathogen detection.
8. Sample Size for the Training Set
The document does not provide information on the sample size for the training set. This is because:
- This is an RT-PCR test, not a deep learning AI model that undergoes "training" in the typical machine learning sense. The "algorithm" here refers to the pre-defined logical rules and parameters within the Assay Definition File (ADF) for interpreting RT-PCR signals.
- The re-analysis was performed on verification, validation, and flex studies data, which are typically considered test or validation sets in the context of device development, not training sets.
9. How the Ground Truth for the Training Set Was Established
As noted above, a "training set" and associated ground truth establishment in the context of deep learning AI are not applicable to this RT-PCR device. The "ground truth" for the device's performance relies on the known characteristics of the viral targets and the performance of the RT-PCR chemistry against reference methods.
Ask a specific question about this device
(30 days)
The Xpert Xpress CoV-2/Flu/RSV plus test, performed on the GeneXpert Dx and GeneXpert Infinity Systems, is an automated multiplexed real-time reverse transcriptase polymerase chain reaction (RT-PCR) test intended for use in the simultaneous in vitro qualitative detection and differentiation of severe acute respiratory syndrome coronavirus (SARS-CoV-2), influenza A, influenza B, and/or respiratory syncytial virus (RSV) viral RNA in nasopharyngeal swab and anterior nasal swab specimens collected from individuals with signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2, influenza A, influenza B, and RSV can be similar.
The Xpert Xpress CoV-2/Flu/RSV plus test is intended for use in the differential detection of SARS-CoV-2, influenza A, influenza B and/or RSV RNA and aids in the diagnosis of COVID-19, influenza and/or RSV infections if used in conjunction with other clinical and epidemiological information, and laboratory findings. SARS-CoV-2, influenza A, influenza B, and RSV viral RNA are generally detectable in nasopharyngeal swab and anterior nasal swab specimens during the acute phase of infection.
Positive results are indicative of the presence of the identified virus, but do not rule out bacterial infection or co-infection with other pathogens not detected by the test. The agent (s) detected by the Xpert Xpress CoV-2/Flu/RSV plus test may not be the definite cause of the disease.
Negative results do not preclude SARS-CoV-2, influenza A, influenza B and/or RSV infection. The results of this test should not be used as the sole basis for diagnosis, treatment or other patient management decisions.
The Xpert Xpress CoV-2/Flu/RSV plus test is an automated in vitro diagnostic test for the simultaneous qualitative detection and differentiation of SARS-CoV-2, Flu A, Flu B, and RSV viral RNA in nasopharyngeal swab (NPS) and anterior nasal swab (NS) specimens collected from individuals showing signs and symptoms of respiratory viral infection.
The Xpert Xpress CoV-2/Flu/RSV plus test is performed on GeneXpert Instrument Systems (GeneXpert Dx, GeneXpert Infinity-48s and GeneXpert Infinity-80 systems, GeneXpert System with Touchscreen), which consist of an instrument, computer or touchscreen, and preloaded software for running tests and viewing the results. The GeneXpert Instrument Systems automate and integrate sample preparation, nucleic acid extraction and amplification, and detection of the target sequences in simple or complex samples using real-time reverse transcription (RT)-polymerase chain reaction (PCR) and PCR technology. Depending on the instrument, the GeneXpert Instrument Systems can have from 1 and up to 80 randomly accessible modules, each capable of performing separate sample preparation and real-time RT-PCR and PCR tests. Each module contains a syringe drive for dispensing fluids (i.e., the syringe drive activates the plunger that works in concert with the rotary valve in the cartridge to move fluids between chambers), an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time RT-PCR and PCR as well as detection. The systems require the use of single-use disposable cartridges that hold the RT-PCR reagents and host sample purification, nucleic acid amplification, and detection of the target sequences. Because the cartridges are self-contained, cross-contamination between samples is minimized.
The Xpert Xpress CoV-2/Flu/RSV plus test includes reagents for the detection of SARS-CoV-2, Flu A, Flu B and RSV viral RNA from NPS and NS specimens. The primers and probes in the Xpert Xpress CoV-2/Flu/RSV plus test are designed to amplify and detect unique sequences in the genes that encode the following proteins: SARS-CoV-2 nucleocapsid (N), SARS-CoV-2 envelope (E), SARS-CoV-2 RNA-dependent RNA polymerase (RdRP), influenza A matrix (M), influenza A basic polymerase (PB2), influenza A acidic protein (PA), influenza B matrix (M), influenza B non-structural protein (NS), and the RSV A and RSV B nucleocapsid.
A Sample Processing Control (SPC) and a Probe Check Control (PCC) are also included in the cartridge utilized by the GeneXpert instrument. The SPC is present to control for adequate processing of the sample and to monitor for the presence of potential inhibitor(s) in the RT-PCR reaction. The SPC also ensures that the RT-PCR reaction conditions (temperature and time) are appropriate for the amplification reaction and that the RT-PCR reagents are functional. The PCC verifies reagent rehydration, PCR tube filling, and confirms that all reaction components are present in the cartridge including monitoring for probe integrity and dye stability.
The Xpert Xpress CoV-2/Flu/RSV plus test is designed for use with NPS or NS specimens collected with nylon flocked swabs and placed into viral transport medium (VTM), Universal Transport Medium (UTM), or eNAT®.
The FDA 510(k) Clearance Letter for the Xpert Xpress CoV-2/Flu/RSV plus device describes modifications to an existing PCR test. The information provided outlines the acceptance criteria implicitly through the modifications and performance data presented. However, it's important to note that this document is a 510(k) summary, which provides a high-level overview of the submission and does not contain the full details of all studies conducted. Therefore, some requested information may not be explicitly present in the provided text.
Based on the provided text, here's an analysis:
Summary of Acceptance Criteria and Device Performance
The core of this 510(k) submission is a "Special 510(k)" which means the device (Xpert Xpress CoV-2/Flu/RSV plus) is largely the same as a previously cleared predicate device (K231481), but with minor design changes. Therefore, the acceptance criteria are implicitly tied to demonstrating that these changes do not negatively impact the previously established performance claims of the predicate device.
Specifically, the modifications include:
- Turning off Signal Loss Detection (SLD) for the Flu B channel.
- Revising the result reporting algorithm for the SARS-CoV-2 only test mode.
- Previously: SARS-CoV-2 NEGATIVE if SARS-CoV-2 analyte result is INVALID and SPC is FAIL.
- Revised: INVALID GeneXpert test result if SARS-CoV-2 analyte result is INVALID and SPC is FAIL.
Table of Acceptance Criteria (Implicit) and Reported Device Performance:
| Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|
| No impact on validity of test runs with SLD off for Flu B channel. | "the numbers of valid test runs and non-determinate (ND) test results were the same between the original ADF and the updated ADF with Flu B SLD Off." |
| No impact on non-determinate results with SLD off for Flu B channel. | "the numbers of valid test runs and non-determinate (ND) test results were the same between the original ADF and the updated ADF with Flu B SLD Off." |
| Correctly change specific "ERROR" results to "NO RESULT" due to SLD being off. | "All non-determinate GeneXpert ERROR test results with error code 5011 (SLD) were changed to NO RESULT." |
| Revised algorithm produces expected "INVALID" results for SARS-CoV-2 only test mode under specific conditions (SARS-CoV-2 INVALID and SPC FAIL). | "The revised algorithm produced the expected INVALID GeneXpert test results for the Xpress SARS-CoV-2 plus test mode." |
| Overall performance claims of the device are not impacted by the modifications. | "The assessment of the re-analysis results determined that the performance claims of the Xpert Xpress CoV-2/Flu/RSV plus test were not impacted by the modifications made to the predicate device." |
Study Details Proving Acceptance:
-
Sample sizes used for the test set and the data provenance:
- Test Set (Re-analysis Data):
- Analytical Test Results: 15,645
- Clinical, Reproducibility-Precision, and Single-Site Precision Test Results: 9,525 (comprising 8,535 specimens and 990 controls).
- Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective for the re-analyzed verification and validation studies. However, since it refers to "original studies" (K231481), it implies data previously collected and used for the predicate device's clearance.
- Test Set (Re-analysis Data):
-
Number of experts used to establish the ground truth for the test set and their qualifications:
- This information is not provided in the given text. For an in vitro diagnostic (IVD) device like this, ground truth is typically established by comparative methods (e.g., another FDA-cleared PCR test, or a consensus of multiple clinical/laboratory results), not by human expert readers in the way an imaging AI device might use radiologists. The "ground truth" for this device's performance would be the presence or absence of viral RNA, determined by well-established laboratory methods.
-
Adjudication method for the test set:
- This information is not provided and is generally not applicable to the type of re-analysis done for an IVD device where ground truth is established by laboratory methods rather than human interpretation requiring adjudication.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done:
- No, an MRMC study was not done. This type of study is primarily relevant for imaging AI devices that assist human readers in interpretation. This device is an automated, standalone diagnostic test.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Yes, implicitly. The re-analysis of performance data, particularly the comparison between the original and updated Assay Definition Files (ADF) and the verification of the revised algorithm, represents the standalone performance of the device's software (algorithm) logic. The device itself is an automated system.
-
The type of ground truth used:
- Implicitly, the ground truth was established by laboratory reference methods (e.g., RT-PCR) for the original clinical specimens. For the re-analysis, the "ground truth" for evaluating the impact of the changes was the consistency of results and proper operation of the updated algorithm against the expected behavior, confirmed by the original analytical and clinical study data. For the SARS-CoV-2 algorithm change, "test cartridges were simulated to generate SARS-CoV-2 INVALID and SPC FAIL conditions," meaning the ground truth for this specific verification was a controlled simulation designed to trigger the specific algorithm conditions.
-
The sample size for the training set:
- Not explicitly stated in relation to this Special 510(k) submission. As this is a modification to an existing device, it's likely that the original training/development data for the predicate device were used, but the size of that dataset is not provided here. The 15,645 analytical test results and 9,525 clinical/precision test results mentioned are re-analyzed verification and validation data, not training data.
-
How the ground truth for the training set was established:
- Not explicitly stated in the provided text. For an RT-PCR diagnostic, the ground truth for training/development would typically involve characterized clinical samples or contrived samples with known viral concentrations and presence/absence of targets, verified by highly sensitive and specific reference methods (e.g., sequencing, confirmatory PCRs, or sometimes clinical outcomes correlated with virology).
Ask a specific question about this device
(175 days)
The cobas liat SARS-CoV-2, Influenza A/B & RSV nucleic acid test is an automated rapid multiplex real-time reverse transcription polymerase chain reaction (RT-PCR) test intended for the simultaneous qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A virus, influenza B virus and respiratory syncytial virus (RSV) nucleic acids in anterior nasal (nasal) and nasopharyngeal swab specimens from individuals exhibiting signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2, influenza and RSV can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza A, influenza B, and RSV infections in humans and is not intended to detect influenza C virus infections.
Nucleic acids from the viral organisms identified by this test are generally detectable in nasopharyngeal and nasal swab specimens during the acute phase of infection. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative of the presence of the identified virus, and aid in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings.
The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organisms. The organism(s) detected by the cobas liat SARS-CoV-2, Influenza A/B & RSV nucleic acid test may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus, influenza B virus, or RSV infections.
The cobas liat SARS-CoV-2, Influenza A/B & RSV nucleic acid test is performed on the cobas liat analyzer which automates and integrates sample purification, nucleic acid amplification, and detection of the target sequence in biological samples using real-time PCR assays. The assay targets both the ORF1 a/b non-structural region and membrane protein gene that are unique to SARS-CoV-2, a well-conserved region of the matrix gene of influenza A (Flu A target), the nonstructural protein 1 (NS1) gene of influenza B (Flu B target) and the matrix gene of RSV (RSV target). An Internal Control (IC) is included to control for adequate processing of the target virus through all steps of the assay process and to monitor the presence of inhibitors in the RT-PCR processes.
The provided text describes the analytical and clinical performance evaluation of the cobas® liat SARS-CoV-2, Influenza A/B & RSV nucleic acid test, which is a real-time RT-PCR assay. The information mainly focuses on the performance characteristics required for FDA clearance (510(k)).
Here's a breakdown of the requested information based on the provided document:
Acceptance Criteria and Device Performance
The document does not explicitly present a table of "acceptance criteria" in a pass/fail format for clinical performance. Instead, it demonstrates the device's performance through various analytical studies and clinical agreement percentages relative to a comparator method. The acceptance for a 510(k) submission is typically that the device is "substantially equivalent" to a legally marketed predicate device, which implies demonstrating comparable performance characteristics.
The key performance metrics are the Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) in clinical studies. While there aren't explicit numeric acceptance criteria stated, the achieved performance values are presented as evidence of substantial equivalence.
Here’s a table summarizing the reported clinical device performance based on the prospective study (Table 13) and the retrospective study (Table 15). These are the metrics by which the device's clinical performance would be "accepted" as substantially equivalent.
Table of Reported Device Performance (Clinical)
| Target | Specimen Type | PPA (%) (95% CI) | NPA (%) (95% CI) |
|---|---|---|---|
| SARS-CoV-2 | NPS | 94.5 (90.7-96.8) | 97.6 (96.7-98.3) |
| SARS-CoV-2 | ANS | 96.7 (93.4-98.4) | 97.2 (96.2-97.9) |
| Influenza A | NPS | 100.0 (93.4-100.0) | 99.3 (98.8-99.6) |
| Influenza A | ANS | 100.0 (93.2-100.0) | 99.3 (98.8-99.6) |
| Influenza B | NPS (Prospective) | 100.0 (85.1-100.0) | 99.3 (98.8-99.6) |
| Influenza B | ANS (Prospective) | 100.0 (86.2-100.0) | 99.5 (99.0-99.7) |
| Influenza B | NPS (Retrospective) | 100.0 (89.8-100.0) | 97.9 (94.7-99.2) |
| Influenza B | ANS (Retrospective) | 100.0 (89.8-100.0) | 98.3 (95.0-99.4) |
| RSV | NPS | 100.0 (94.8-100.0) | 99.0 (98.3-99.3) |
| RSV | ANS | 97.5 (91.4-99.3) | 98.8 (98.2-99.3) |
Note on "Acceptance Criteria" for Analytical Performance: The document describes detailed analytical studies (LoD, inclusivity, cross-reactivity, interference, reproducibility), and the reported hit rates and concentrations demonstrate that the device met the internal analytical performance specifications, which are implicitly the "acceptance criteria" for these aspects. For instance, for LoD, the acceptance criterion is implied to be ≥95% hit rate at the determined concentration. For inclusivity, it's detection at or near 3x LoD. For cross-reactivity and interference, the acceptance criterion is no cross-reactivity/interference observed. The document states that "none of the organisms tested cross reacted or interfered" and that "substances... did not interfere," indicating successful meeting of these criteria. For reproducibility, the agreement percentages for positive and negative samples are above 99% for most categories.
Study Details
-
Sample sizes used for the test set and the data provenance:
- Prospective Clinical Study (Category I):
- NPS specimens: 1729 enrolled subjects leading to 1704 evaluable specimens for SARS-CoV-2, Flu A, Flu B, and 1705 for RSV.
- ANS specimens: 1729 enrolled subjects leading to 1705 evaluable specimens for SARS-CoV-2, Flu B, 1703 for Flu A, and 1706 for RSV.
- Data Provenance: Fresh specimens, prospective, collected between September 2023 and March 2024 at 14 point-of-care testing sites in the United States (US).
- Retrospective Clinical Study (Category III):
- Specimens: Frozen archived clinical NPS (n=223) and ANS (n=206) specimens.
- Data Provenance: Retrospective, collected between 2019 and 2023. Distributed to 6 sites for testing. Country of origin not explicitly stated but implied to be US given the overall context of a US FDA clearance (though not definitively stated for the retrospective part).
- Prospective Clinical Study (Category I):
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
The ground truth for the clinical test set was established by comparing the results of the cobas® liat test to the "results of an FDA-cleared Nucleic Acid Amplification Test (NAAT)." The document does not specify the number of human experts, their qualifications, or their role in establishing this ground truth. The "ground truth" here is the result from the comparator NAAT, not human expert interpretation of images or clinical data. -
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
The document describes a discrepant analysis for cases where the cobas® liat test results differed from the comparator NAAT. For the prospective study, the discrepancy analysis showed how many of the discrepant results (e.g., cobas® liat negative, comparator positive) were ultimately confirmed as positive or negative by further investigation (implied to be by the comparator method or potentially a third method, though not explicitly detailed beyond "discrepant NAAT results"). The details provided are:- SARS-CoV-2 NPS: Of 12 negative cobas® liat/positive comparator, 8 were positive and 4 negative. Of 35 positive cobas® liat/negative comparator, 12 were positive and 23 negative.
- Similar analyses are provided for other targets and specimen types.
This implies an adjudication method where discrepant results were further investigated, likely with repeat testing or a confirmatory reference method, but the specific "2+1" or "3+1" reader/expert adjudication model (common in imaging studies) is not applicable or described for this in vitro diagnostic (IVD) device.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No, an MRMC comparative effectiveness study was not done. This type of study is primarily relevant for imaging devices that assist human readers (e.g., AI for radiology). The cobas® liat test is an automated molecular diagnostic test directly detecting nucleic acids, not an AI-assisted interpretation device for human "readers." -
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
Yes, the performance reported (PPA, NPA, and analytical studies) represents the standalone performance of the cobas® liat device. It is an automated system where the "algorithm" (the RT-PCR assay and its interpretation software) directly produces a qualitative result (Detected/Not Detected), without human "in-the-loop" interpretation for the primary result. Human operators load samples and review results, but the analytical detection and differentiation itself is automated. -
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The ground truth for both prospective and retrospective clinical studies was based on the results of an FDA-cleared Nucleic Acid Amplification Test (NAAT), which served as the comparator method. For discrepant samples, further re-testing against the comparator or a reference method was performed for adjudication. This falls under a "reference standard" or "comparator method" type of ground truth. -
The sample size for the training set:
The document does not specify the sample size for a "training set." This type of molecular diagnostic device typically relies on analytical validation (LoD, inclusivity, specificity) and clinical validation through comparison to a reference method, rather than a machine learning model that requires explicit training data. The development process would involve iterative optimization of primers, probes, and assay conditions, but this is not typically referred to as a "training set" in the context of IVD submissions, especially for traditional PCR assays. -
How the ground truth for the training set was established:
As no explicit "training set" for a machine learning model is described, the question of how its ground truth was established is not applicable based on the provided text. The "ground truth" in the context of this traditional IVD development refers to the reliable identification of the target analytes in samples for analytical and clinical validation, often through established reference methods or characterized materials.
Ask a specific question about this device
(176 days)
The cobas liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test is an automated rapid multiplex real-time reverse transcription polymerase chain reaction (RT-PCR) test intended for the simultaneous qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A virus and influenza B virus nucleic acids in anterior nasal (nasal) and nasopharyngeal swab specimens from individuals exhibiting signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2 and influenza can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza A and influenza B infections in humans and is not intended to detect influenza C virus infections.
Nucleic acids from the viral organisms identified by this test are generally detectable in nasopharyngeal and nasal swab specimens during the acute phase of infection. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative of the presence of the identified virus, and aid in diagnosis if used in conjunction with other clinical and epidemiological information and laboratory findings.
The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organisms. The organism(s) detected by the cobas liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus or influenza B virus infections.
The cobas liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test is performed on the cobas liat analyzer which automates and integrates sample purification, nucleic acid amplification, and detection of the target sequence in biological samples using real-time PCR assays. The assay targets both the ORF1 a/b non-structural region and membrane protein gene that are unique to SARS-CoV-2, a well-conserved region of the matrix gene of influenza A (Flu A target), and the nonstructural protein 1 (NS1) gene of influenza B (Flu B target). An Internal Control (IC) is included to control for adequate processing of the target virus through all steps of the assay process and to monitor the presence of inhibitors in the RT-PCR processes.
This document describes the validation study for the cobas liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test.
Here's an analysis of the acceptance criteria and the study proving the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a diagnostic test, the primary acceptance criteria revolve around analytical and clinical performance metrics like Limit of Detection, Inclusivity, Cross-Reactivity, Reproducibility, and Clinical Agreement (Positive Percent Agreement and Negative Percent Agreement). The document doesn't explicitly state "acceptance criteria" values in a separate table, but these are implied by the performance metrics reported and the general standards for diagnostic device clearance. I will extract the reported device performance from the provided text.
| Performance Metric | Target Analyte | Specimen Type | Reported Performance (Value) | Implied Acceptance Criteria (Typically high for diagnostic tests) |
|---|---|---|---|---|
| Analytical Sensitivity (LoD) | SARS-CoV-2 | Co-spiked panels | 0.0350 TCID50/mL | Lowest detectable concentration for 95% positivity |
| Influenza A | Co-spiked panels | 0.00325 TCID50/mL | Lowest detectable concentration for 95% positivity | |
| Influenza B | Co-spiked panels | 0.183 TCID50/mL | Lowest detectable concentration for 95% positivity | |
| Reactivity/Inclusivity | SARS-CoV-2 | Respective variants | 100% detection at 3x LoD | Detection of various strains/variants |
| Influenza A | Respective variants | 100% detection at varying LoD (up to 12x) | Detection of various strains/variants | |
| Influenza B | Respective variants | 100% detection at 3x LoD | Detection of various strains/variants | |
| Cross-Reactivity/Microbial Interference | All targets | Various microorganisms | No cross-reactivity/interference | No false positives or interference from other common pathogens |
| Competitive Inhibition | All targets | Co-spiked samples | No interference | Accurate detection of all targets even in co-infection |
| Endogenous/Exogenous Interference | All targets | Various substances | No interference | Robust performance in presence of common respiratory interferents |
| Reproducibility (Negative) | N/A | Negative samples | 100.0% Agreement | High agreement for negative samples across sites, lots, days |
| Reproducibility (1x-2x LoD) | SARS-CoV-2 | Low Positive samples | 100.0% Agreement | High agreement for low positive samples |
| Influenza A | Low Positive samples | 99.6% Agreement | High agreement for low positive samples | |
| Influenza B | Low Positive samples | 99.6% Agreement | High agreement for low positive samples | |
| Reproducibility (3x-5x LoD) | SARS-CoV-2 | Moderate Positive | 100.0% Agreement | High agreement for moderate positive samples |
| Influenza A | Moderate Positive | 100.0% Agreement | High agreement for moderate positive samples | |
| Influenza B | Moderate Positive | 100.0% Agreement | High agreement for moderate positive samples | |
| *Clinical Performance (PPA)Prospective | SARS-CoV-2 | NPS | 94.5% (90.7-96.8 CI) | High sensitivity (ability to detect true positives) |
| SARS-CoV-2 | ANS | 96.7% (93.4-98.4 CI) | High sensitivity (ability to detect true positives) | |
| Influenza A | NPS | 100.0% (93.4-100.0 CI) | High sensitivity (ability to detect true positives) | |
| Influenza A | ANS | 100.0% (93.2-100.0 CI) | High sensitivity (ability to detect true positives) | |
| Influenza B | NPS | 100.0% (85.1-100.0 CI) | High sensitivity (ability to detect true positives) | |
| Influenza B | ANS | 100.0% (86.2-100.0 CI) | High sensitivity (ability to detect true positives) | |
| *Clinical Performance (NPA)Prospective | SARS-CoV-2 | NPS | 97.6% (96.7-98.3 CI) | High specificity (ability to correctly identify true negatives) |
| SARS-CoV-2 | ANS | 97.2% (96.2-97.9 CI) | High specificity (ability to correctly identify true negatives) | |
| Influenza A | NPS | 99.3% (98.8-99.6 CI) | High specificity (ability to correctly identify true negatives) | |
| Influenza A | ANS | 99.3% (98.8-99.6 CI) | High specificity (ability to correctly identify true negatives) | |
| Influenza B | NPS | 99.3% (98.8-99.6 CI) | High specificity (ability to correctly identify true negatives) | |
| Influenza B | ANS | 99.5% (99.0-99.7 CI) | High specificity (ability to correctly identify true negatives) | |
| *Clinical Performance (PPA)Retrospective | Influenza B | NPS | 100.0% (89.8-100.0 CI) | High sensitivity (ability to detect true positives) |
| Influenza B | ANS | 100.0% (89.8-100.0 CI) | High sensitivity (ability to detect true positives) | |
| *Clinical Performance (NPA)Retrospective | Influenza B | NPS | 97.9% (94.7-99.2 CI) | High specificity (ability to correctly identify true negatives) |
| Influenza B | ANS | 98.3% (95.0-99.4 CI) | High specificity (ability to correctly identify true negatives) |
2. Sample Sizes Used for the Test Set and Data Provenance
-
Prospective Clinical Study:
- Sample Size: 1729 symptomatic subjects enrolled.
- 1705 evaluable NPS specimens for analysis (19 non-evaluable due to missing/invalid results, 5 due to handling).
- 1706 evaluable ANS specimens for SARS-CoV-2 and Influenza B analysis (22 non-evaluable due to missing/invalid results, 1 due to handling).
- 1704 evaluable ANS specimens for Influenza A analysis (2 additional found inconclusive for comparator).
- Data Provenance: Prospective, collected between September 2023 and March 2024 at 14 point-of-care testing sites in the United States (US).
- Sample Size: 1729 symptomatic subjects enrolled.
-
Retrospective Clinical Study (Influenza B Supplement):
- Sample Size: 223 archived NPS specimens and 206 archived ANS specimens (total 429).
- One NPS sample pre-characterized as positive for influenza B was non-evaluable.
- Data Provenance: Retrospective, frozen archived (Category III) specimens collected between 2019 and 2023. Distributed to 6 sites for testing.
- Sample Size: 223 archived NPS specimens and 206 archived ANS specimens (total 429).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts
The document does not mention the use of experts to establish ground truth for the clinical test sets. For molecular diagnostic tests like this, the ground truth is typically established by comparing the investigational device's results against a highly accurate, accepted comparator method (another FDA-cleared Nucleic Acid Amplification Test specific for the target analytes). The expertise lies in the development and validation of these comparator methods, not in individual expert review of each sample for ground truth in this context.
4. Adjudication Method for the Test Set
The document describes discrepant result analysis for both prospective and retrospective clinical studies.
- For the prospective study, "discrepant NAAT results" are detailed for SARS-CoV-2 (NPS and ANS), Influenza A (NPS and ANS), and Influenza B (NPS and ANS).
- For the retrospective study, discrepant NAAT results are detailed for Influenza B (NPS and ANS).
The method appears to be:
- The cobas liat test result is compared to the FDA-cleared comparator NAAT result.
- When there's a discrepancy (e.g., cobas liat positive, comparator negative), it explicitly states how many were "positive" and "negative" upon further investigation or re-evaluation (e.g., with "discrepant NAAT results").
- For example: "Of 12 specimens negative on cobas® liat and positive on the comparator, 8 were positive and 4 were negative." This implies some form of re-testing or deeper analysis (not specified as "adjudication by experts" but rather "discrepant NAAT results"). It's more of a re-confirmation of the comparator or a third method, rather than a human expert consensus process. Such re-evaluation often involves re-testing using the comparator or a reference method.
Therefore, while there's no "2+1" or "3+1" expert adjudication method described as would be seen in imaging studies, there is a discrepant resolution process based on further NAAT results. It's not "none" in the sense that discrepancies are just reported without follow-up; rather, they are further investigated using additional NAAT results to re-confirm the original comparator status if possible.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
No. This is a standalone diagnostic test (RT-PCR), not an AI-assisted imaging device or a test that involves human "readers" interpreting results. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, implicitly. This is a fully automated RT-PCR test run on the cobas liat analyzer. The performance metrics (LoD, inclusivity, cross-reactivity, reproducibility, and clinical agreement) are measures of the device's performance on its own against established ground truth (comparator NAAT). While humans load samples and interpret the final digital result (positive/negative), the core detection and differentiation is algorithm-driven within the instrument, making its performance essentially "standalone" in the context of diagnostic accuracy.
7. The Type of Ground Truth Used
- Clinical Performance (Prospective and Retrospective): The ground truth for clinical sample testing was established by comparing the cobas liat results against an FDA-cleared Nucleic Acid Amplification Test (NAAT), which serves as the reference or "ground truth" method for molecular diagnostic assays. The document explicitly states: "PPA and NPA were determined by comparing the results of cobas® liat SARS-CoV-2 & Influenza A/B v2 to the results of an FDA-cleared Nucleic Acid Amplification Test (NAAT)." and "The comparator method was an acceptable FDA-cleared molecular assay."
- Analytical Studies (LoD, Inclusivity, Cross-Reactivity, Interference, Reproducibility): Ground truth was established by preparing precisely known concentrations of viral material (cultured or inactivated viruses) or specific microorganisms in controlled laboratory settings. For these studies, the "ground truth" is meticulously prepared and verified laboratory standards.
8. The Sample Size for the Training Set
The document does not specify a separate "training set" sample size. For an RT-PCR diagnostic platform, the "training" involves the fundamental biochemical and optical engineering, and the optimization of assay (reagent) design to achieve sensitivity and specificity. This is distinct from machine learning models that often require large, labeled datasets for "training." The analytical and clinical validation studies described here are verification and validation (V&V) studies, akin to a "test set" to prove the device's performance against its design specifications and clinical utility.
9. How the Ground Truth for the Training Set Was Established
Since no explicit "training set" for a machine learning algorithm is mentioned (as this is a molecular diagnostic test), this question is not directly applicable. However, the ground truth for assay development and optimization (which can be considered analogous to "training" in a broader sense of device development) would have been established through extensive laboratory work using:
- Highly characterized viral cultures or purified nucleic acids: Used to define target sequences, optimize primer/probe design, and determine initial analytical sensitivity.
- Spiked samples: Adding known quantities of targets or interferents to negative clinical matrices to mimic real-world conditions during early development.
- Early clinical samples: Used to refine assay performance and resolve initial issues prior to formal validation studies.
These processes ensure the assay correctly identifies the target nucleic acids.
Ask a specific question about this device
Page 1 of 4