K Number
K243400

Validate with FDA (Live)

Date Cleared
2025-04-25

(176 days)

Product Code
Regulation Number
866.3981
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The cobas liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test is an automated rapid multiplex real-time reverse transcription polymerase chain reaction (RT-PCR) test intended for the simultaneous qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A virus and influenza B virus nucleic acids in anterior nasal (nasal) and nasopharyngeal swab specimens from individuals exhibiting signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2 and influenza can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza A and influenza B infections in humans and is not intended to detect influenza C virus infections.

Nucleic acids from the viral organisms identified by this test are generally detectable in nasopharyngeal and nasal swab specimens during the acute phase of infection. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative of the presence of the identified virus, and aid in diagnosis if used in conjunction with other clinical and epidemiological information and laboratory findings.

The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organisms. The organism(s) detected by the cobas liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus or influenza B virus infections.

Device Description

The cobas liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test is performed on the cobas liat analyzer which automates and integrates sample purification, nucleic acid amplification, and detection of the target sequence in biological samples using real-time PCR assays. The assay targets both the ORF1 a/b non-structural region and membrane protein gene that are unique to SARS-CoV-2, a well-conserved region of the matrix gene of influenza A (Flu A target), and the nonstructural protein 1 (NS1) gene of influenza B (Flu B target). An Internal Control (IC) is included to control for adequate processing of the target virus through all steps of the assay process and to monitor the presence of inhibitors in the RT-PCR processes.

AI/ML Overview

This document describes the validation study for the cobas liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test.

Here's an analysis of the acceptance criteria and the study proving the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

Since this is a diagnostic test, the primary acceptance criteria revolve around analytical and clinical performance metrics like Limit of Detection, Inclusivity, Cross-Reactivity, Reproducibility, and Clinical Agreement (Positive Percent Agreement and Negative Percent Agreement). The document doesn't explicitly state "acceptance criteria" values in a separate table, but these are implied by the performance metrics reported and the general standards for diagnostic device clearance. I will extract the reported device performance from the provided text.

Performance MetricTarget AnalyteSpecimen TypeReported Performance (Value)Implied Acceptance Criteria (Typically high for diagnostic tests)
Analytical Sensitivity (LoD)SARS-CoV-2Co-spiked panels0.0350 TCID50/mLLowest detectable concentration for 95% positivity
Influenza ACo-spiked panels0.00325 TCID50/mLLowest detectable concentration for 95% positivity
Influenza BCo-spiked panels0.183 TCID50/mLLowest detectable concentration for 95% positivity
Reactivity/InclusivitySARS-CoV-2Respective variants100% detection at 3x LoDDetection of various strains/variants
Influenza ARespective variants100% detection at varying LoD (up to 12x)Detection of various strains/variants
Influenza BRespective variants100% detection at 3x LoDDetection of various strains/variants
Cross-Reactivity/Microbial InterferenceAll targetsVarious microorganismsNo cross-reactivity/interferenceNo false positives or interference from other common pathogens
Competitive InhibitionAll targetsCo-spiked samplesNo interferenceAccurate detection of all targets even in co-infection
Endogenous/Exogenous InterferenceAll targetsVarious substancesNo interferenceRobust performance in presence of common respiratory interferents
Reproducibility (Negative)N/ANegative samples100.0% AgreementHigh agreement for negative samples across sites, lots, days
Reproducibility (1x-2x LoD)SARS-CoV-2Low Positive samples100.0% AgreementHigh agreement for low positive samples
Influenza ALow Positive samples99.6% AgreementHigh agreement for low positive samples
Influenza BLow Positive samples99.6% AgreementHigh agreement for low positive samples
Reproducibility (3x-5x LoD)SARS-CoV-2Moderate Positive100.0% AgreementHigh agreement for moderate positive samples
Influenza AModerate Positive100.0% AgreementHigh agreement for moderate positive samples
Influenza BModerate Positive100.0% AgreementHigh agreement for moderate positive samples
*Clinical Performance (PPA)ProspectiveSARS-CoV-2NPS94.5% (90.7-96.8 CI)High sensitivity (ability to detect true positives)
SARS-CoV-2ANS96.7% (93.4-98.4 CI)High sensitivity (ability to detect true positives)
Influenza ANPS100.0% (93.4-100.0 CI)High sensitivity (ability to detect true positives)
Influenza AANS100.0% (93.2-100.0 CI)High sensitivity (ability to detect true positives)
Influenza BNPS100.0% (85.1-100.0 CI)High sensitivity (ability to detect true positives)
Influenza BANS100.0% (86.2-100.0 CI)High sensitivity (ability to detect true positives)
*Clinical Performance (NPA)ProspectiveSARS-CoV-2NPS97.6% (96.7-98.3 CI)High specificity (ability to correctly identify true negatives)
SARS-CoV-2ANS97.2% (96.2-97.9 CI)High specificity (ability to correctly identify true negatives)
Influenza ANPS99.3% (98.8-99.6 CI)High specificity (ability to correctly identify true negatives)
Influenza AANS99.3% (98.8-99.6 CI)High specificity (ability to correctly identify true negatives)
Influenza BNPS99.3% (98.8-99.6 CI)High specificity (ability to correctly identify true negatives)
Influenza BANS99.5% (99.0-99.7 CI)High specificity (ability to correctly identify true negatives)
*Clinical Performance (PPA)RetrospectiveInfluenza BNPS100.0% (89.8-100.0 CI)High sensitivity (ability to detect true positives)
Influenza BANS100.0% (89.8-100.0 CI)High sensitivity (ability to detect true positives)
*Clinical Performance (NPA)RetrospectiveInfluenza BNPS97.9% (94.7-99.2 CI)High specificity (ability to correctly identify true negatives)
Influenza BANS98.3% (95.0-99.4 CI)High specificity (ability to correctly identify true negatives)

2. Sample Sizes Used for the Test Set and Data Provenance

  • Prospective Clinical Study:

    • Sample Size: 1729 symptomatic subjects enrolled.
      • 1705 evaluable NPS specimens for analysis (19 non-evaluable due to missing/invalid results, 5 due to handling).
      • 1706 evaluable ANS specimens for SARS-CoV-2 and Influenza B analysis (22 non-evaluable due to missing/invalid results, 1 due to handling).
      • 1704 evaluable ANS specimens for Influenza A analysis (2 additional found inconclusive for comparator).
    • Data Provenance: Prospective, collected between September 2023 and March 2024 at 14 point-of-care testing sites in the United States (US).
  • Retrospective Clinical Study (Influenza B Supplement):

    • Sample Size: 223 archived NPS specimens and 206 archived ANS specimens (total 429).
      • One NPS sample pre-characterized as positive for influenza B was non-evaluable.
    • Data Provenance: Retrospective, frozen archived (Category III) specimens collected between 2019 and 2023. Distributed to 6 sites for testing.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

The document does not mention the use of experts to establish ground truth for the clinical test sets. For molecular diagnostic tests like this, the ground truth is typically established by comparing the investigational device's results against a highly accurate, accepted comparator method (another FDA-cleared Nucleic Acid Amplification Test specific for the target analytes). The expertise lies in the development and validation of these comparator methods, not in individual expert review of each sample for ground truth in this context.

4. Adjudication Method for the Test Set

The document describes discrepant result analysis for both prospective and retrospective clinical studies.

  • For the prospective study, "discrepant NAAT results" are detailed for SARS-CoV-2 (NPS and ANS), Influenza A (NPS and ANS), and Influenza B (NPS and ANS).
  • For the retrospective study, discrepant NAAT results are detailed for Influenza B (NPS and ANS).

The method appears to be:

  • The cobas liat test result is compared to the FDA-cleared comparator NAAT result.
  • When there's a discrepancy (e.g., cobas liat positive, comparator negative), it explicitly states how many were "positive" and "negative" upon further investigation or re-evaluation (e.g., with "discrepant NAAT results").
    • For example: "Of 12 specimens negative on cobas® liat and positive on the comparator, 8 were positive and 4 were negative." This implies some form of re-testing or deeper analysis (not specified as "adjudication by experts" but rather "discrepant NAAT results"). It's more of a re-confirmation of the comparator or a third method, rather than a human expert consensus process. Such re-evaluation often involves re-testing using the comparator or a reference method.

Therefore, while there's no "2+1" or "3+1" expert adjudication method described as would be seen in imaging studies, there is a discrepant resolution process based on further NAAT results. It's not "none" in the sense that discrepancies are just reported without follow-up; rather, they are further investigated using additional NAAT results to re-confirm the original comparator status if possible.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

No. This is a standalone diagnostic test (RT-PCR), not an AI-assisted imaging device or a test that involves human "readers" interpreting results. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, implicitly. This is a fully automated RT-PCR test run on the cobas liat analyzer. The performance metrics (LoD, inclusivity, cross-reactivity, reproducibility, and clinical agreement) are measures of the device's performance on its own against established ground truth (comparator NAAT). While humans load samples and interpret the final digital result (positive/negative), the core detection and differentiation is algorithm-driven within the instrument, making its performance essentially "standalone" in the context of diagnostic accuracy.

7. The Type of Ground Truth Used

  • Clinical Performance (Prospective and Retrospective): The ground truth for clinical sample testing was established by comparing the cobas liat results against an FDA-cleared Nucleic Acid Amplification Test (NAAT), which serves as the reference or "ground truth" method for molecular diagnostic assays. The document explicitly states: "PPA and NPA were determined by comparing the results of cobas® liat SARS-CoV-2 & Influenza A/B v2 to the results of an FDA-cleared Nucleic Acid Amplification Test (NAAT)." and "The comparator method was an acceptable FDA-cleared molecular assay."
  • Analytical Studies (LoD, Inclusivity, Cross-Reactivity, Interference, Reproducibility): Ground truth was established by preparing precisely known concentrations of viral material (cultured or inactivated viruses) or specific microorganisms in controlled laboratory settings. For these studies, the "ground truth" is meticulously prepared and verified laboratory standards.

8. The Sample Size for the Training Set

The document does not specify a separate "training set" sample size. For an RT-PCR diagnostic platform, the "training" involves the fundamental biochemical and optical engineering, and the optimization of assay (reagent) design to achieve sensitivity and specificity. This is distinct from machine learning models that often require large, labeled datasets for "training." The analytical and clinical validation studies described here are verification and validation (V&V) studies, akin to a "test set" to prove the device's performance against its design specifications and clinical utility.

9. How the Ground Truth for the Training Set Was Established

Since no explicit "training set" for a machine learning algorithm is mentioned (as this is a molecular diagnostic test), this question is not directly applicable. However, the ground truth for assay development and optimization (which can be considered analogous to "training" in a broader sense of device development) would have been established through extensive laboratory work using:

  • Highly characterized viral cultures or purified nucleic acids: Used to define target sequences, optimize primer/probe design, and determine initial analytical sensitivity.
  • Spiked samples: Adding known quantities of targets or interferents to negative clinical matrices to mimic real-world conditions during early development.
  • Early clinical samples: Used to refine assay performance and resolve initial issues prior to formal validation studies.

These processes ensure the assay correctly identifies the target nucleic acids.

U.S. Food & Drug Administration 510(k) Clearance Letter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

April 25, 2025

Roche Molecular Systems, Inc.
Sumedha Sinha
Clinical Development Lead
4300 Hacienda Drive
Pleasanton, California 94588

Re: K243400
Trade/Device Name: cobas liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test
Regulation Number: 21 CFR 866.3981
Regulation Name: Device To Detect And Identify Nucleic Acid Targets In Respiratory Specimens From Microbial Agents That Cause The Sars-Cov-2 Respiratory Infection And Other Microbial Agents When In A Multi-Target Test
Regulatory Class: Class II
Product Code: QOF
Dated: October 30, 2024
Received: October 31, 2024

Dear Sumedha Sinha:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K243400 - Sumedha Sinha Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K243400 - Sumedha Sinha Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joseph Briggs -S

Joseph Briggs, Ph.D.
Deputy Division Director
Division of Microbiology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K243400

Device Name: cobas liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test

Indications for Use (Describe)

The cobas liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test is an automated rapid multiplex real-time reverse transcription polymerase chain reaction (RT-PCR) test intended for the simultaneous qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A virus and influenza B virus nucleic acids in anterior nasal (nasal) and nasopharyngeal swab specimens from individuals exhibiting signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2 and influenza can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza A and influenza B infections in humans and is not intended to detect influenza C virus infections.

Nucleic acids from the viral organisms identified by this test are generally detectable in nasopharyngeal and nasal swab specimens during the acute phase of infection. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative of the presence of the identified virus, and aid in diagnosis if used in conjunction with other clinical and epidemiological information and laboratory findings.

The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organisms. The organism(s) detected by the cobas liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus or influenza B virus infections.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

cobas® liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

Submitter NameRoche Molecular Systems, Inc.
Address4300 Hacienda Drive, Pleasanton, CA 94588-2722
ContactSumedha SinhaPhone: 925-366-0146Email: sumedha.sinha@roche.com
Date Prepared14 April 2025
Proprietary Namecobas® liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test
Common Namecobas® liat SARS-CoV-2 & Influenza A/B v2
Classification21 CFR 866.3981Device To Detect And Identify Nucleic Acid Targets In Respiratory Specimens From Microbial Agents That Cause The SARS-Cov-2 Respiratory Infection And Other Microbial Agents When In A Multi-Target Test
Product CodesQOF
Predicate DevicesHOLOGIC Panther Fusion® SARS-CoV-2/Flu A/B/RSV Assay (K241240)
Establishment RegistrationRoche Molecular Systems, Inc. (2243471)

Page 6

1. DEVICE DESCRIPTION

The cobas® liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test is performed on the cobas® liat analyzer which automates and integrates sample purification, nucleic acid amplification, and detection of the target sequence in biological samples using real-time PCR assays. The assay targets both the ORF1 a/b non-structural region and membrane protein gene that are unique to SARS-CoV-2, a well-conserved region of the matrix gene of influenza A (Flu A target), and the nonstructural protein 1 (NS1) gene of influenza B (Flu B target). An Internal Control (IC) is included to control for adequate processing of the target virus through all steps of the assay process and to monitor the presence of inhibitors in the RT-PCR processes.

2. INDICATIONS FOR USE

The cobas® liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test is an automated rapid multiplex real-time reverse transcription polymerase chain reaction (RT-PCR) test intended for the simultaneous qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A virus and influenza B virus nucleic acids in anterior nasal (nasal) and nasopharyngeal swab specimens from individuals exhibiting signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2 and influenza can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza A and influenza B infections in humans and is not intended to detect influenza C virus infections.

Nucleic acids from the viral organisms identified by this test are generally detectable in nasopharyngeal and nasal swab specimens during the acute phase of infection. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative of the presence of the identified virus, and aid in diagnosis if used in conjunction with other clinical and epidemiological information and laboratory findings.

The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.

Positive results do not rule out coinfection with other organisms. The organism(s) detected by the cobas® liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus or influenza B virus infections.

Page 7

3. TECHNOLOGICAL CHARACTERISTICS

The primary technological characteristics and intended use of the Roche cobas® liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test are substantially equivalent to other legally marketed nucleic acid amplification tests intended for the qualitative detection of SARS-CoV-2 & Influenza A/B.

As indicated in Table 1, the Roche cobas® liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test is substantially equivalent to the identified predicate device, the cleared HOLOGIC Panther Fusion® SARS-CoV-2/Flu A/B/RSV Assay (K241240).

Table 1: Comparison of the cobas® liat SARS-CoV-2 & Influenza A/B v2 test and the Predicate Device

Submitted Device: cobas® liat SARS-CoV-2 & Influenza A/B v2 nucleic acid testPredicate Device: Panther Fusion SARS-CoV-2/Flu A/B/RSV Assay (K241240)
Regulation Name21 CFR 866.398121 CFR 866.3981
Product CodeQOFQOF
Intended UseThe cobas® liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test is an automated rapid multiplex real-time reverse transcription polymerase chain reaction (RT-PCR) test intended for the simultaneous qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A virus and influenza B virus nucleic acids in anterior nasal (nasal) and nasopharyngeal swab specimens from individuals exhibiting signs and symptoms of respiratory tract infection. Clinical signs and symptoms of respiratory tract infection due to SARS-CoV-2 and influenza can be similar. This test is intended to aid in the differential diagnosis of SARS-CoV-2, influenza A and influenza B infections in humans and is not intended to detect influenza C virus infections.Nucleic acids from the viral organisms identified by this test are generally detectable in nasopharyngeal and nasal swab specimens during the acute phase of infection. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative ofThe Panther Fusion SARS-CoV-2/Flu A/B/RSV assay is a fully automated multiplexed real-time polymerase chain reaction (RT-PCR) in vitro diagnostic test intended for the qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A virus (Flu A), influenza B virus (Flu B), and respiratory syncytial virus (RSV). Nucleic acids are isolated and purified from nasopharyngeal (NP) swab specimens and anterior nasal (AN) swab specimens obtained from individuals exhibiting signs and symptoms of a respiratory tract infection.Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2, influenza, and RSV can be similar. This assay is intended to aid in the differential diagnosis of SARS-CoV-2, Flu A, Flu B, and RSV infections in humans and is not intended to detect influenza C virus infections.Nucleic acids from the viral organisms identified by this test are generally detectable in NP and AN swab specimens during the acute phase of infection. The

Page 8

Submitted Device: cobas® liat SARS-CoV-2 & Influenza A/B v2 nucleic acid testPredicate Device: Panther Fusion SARS-CoV-2/Flu A/B/RSV Assay (K241240)
the presence of the identified virus, and aid in diagnosis if used in conjunction with other clinical and epidemiological information and laboratory findings.The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.Positive results do not rule out coinfection with other organisms. The organism(s) detected by the cobas® liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus or influenza B virus infections.detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infection are indicative of the presence of the identified virus and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.Positive results do not rule out coinfection with other organisms. The organism(s) detected by the Panther Fusion SARS-CoV-2/Flu A/B/RSV assay may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus, influenza B virus, or RSV infections. This assay is designed for use on the Panther Fusion system.The Hologic RespDirect Collection Kit is cleared for NP swab specimens only for testing with the Panther Fusion SARS-CoV2/Flu A/B/RSV assay.
Sample TypeNasopharyngeal and anterior nasal swabsNasopharyngeal and anterior nasal swabs
AnalyteSARS-CoV-2, Flu A, Flu B RNASARS-CoV-2, Flu A, Flu B, RSV RNA
Ancillary Collection Kits• Copan FLOQSwabs™ with UTM™• BD UVT with flocked swabs• Sterile flocked swabs with a synthetic tip with other viral transport media (VTM) – Remel™M4RT, M4, M5 and M6• 0.9% Saline• RespDirect Swab, intended for collection of NPS specimens with UTM™/ VTM or eSTM• Enhanced Direct Load Tube (eDLT), containing enhanced specimen transport media (eSTM).
Time to resultAbout 15 minutes~ 2.5 hours
Amplification TechnologyReal-time PCRReal-time PCR
Detection ChemistryAssay using different reporter dyes for target and controlAssay using different reporter dyes for target and control
Controls UsedSample processing control (IC), Positive and negative controlSample processing control (IC), Positive and negative control
Instrumentationcobas® liat SystemPanther Fusion System

Page 9

4. SPECIAL CONTROLS/STANDARDS/GUIDANCE REFERENCED

Class II Special Controls as per 21 CFR 866.3981.

5. NON-CLINICAL PERFORMANCE EVALUATION

5.1. Analytical sensitivity (Limit of Detection)

Limit of detection (LoD) studies determine the lowest detectable concentration of SARS-CoV-2, influenza A, and influenza B at which equal to or greater than 95% of all replicates test positive. Two strains for each target (SARS-CoV-2, influenza A, influenza B) were evaluated. To determine the LoD for each target, co-spiked panels were formulated using cultured or inactivated viral material diluted in pooled negative nasopharyngeal swab matrix. Twenty-one replicates per lot of assay tubes per dilution were tested for five or six 2-fold dilutions using three lots of assay tubes. Each strain evaluated as well as the corresponding LoD are shown in Table 2.

Table 2: LoD determination for SARS-CoV-2, influenza A & influenza B strains

VirusStrainConcentration at LoDHit rate (Mean Ct)
SARS-CoV-2*USA-WA1/20200.0350 TCID50/mL20/21 (35.5)
SARS-CoV-2*WHO International Standard 20/146, v3, 11/202165.1 IU/mL21/21 (34.9)
Influenza ADarwin/6/20210.295 TCID50/mL21/21 (35.0)
Influenza ABrisbane/02/20180.00325 TCID50/mL21/21 (36.4)
Influenza BB/Austria/1359417/20210.979 TCID50/mL20/21 (34.6)
Influenza BPhuket/3073/20130.183 TCID50/mL21/21 (34.1)

*Inactivated virus

A separate study was performed to demonstrate that the LoD of each strain individually was equivalent to the co-spiked LoD.

5.2. Reactivity/Inclusivity

The inclusivity study evaluates the ability of the test to detect SARS-CoV-2, influenza A and influenza B isolates/variants. The reactivity/inclusivity was evaluated with 10 SARS-CoV-2, 28 influenza A (15 H1N1 & 10 H3N2, 1 H5N1, 1 H5N2 and 1 H7N9), and ten (10) influenza B (5 Victoria and 5 Yamagata) variants. All strains were individually tested at 3x LoD in 3 replicates

Page 10

to evaluate inclusivity. If < 100% hit rate was observed, the concentration was doubled until 3/3 replicates were detected.

SARS-CoV-2

The SARS-CoV-2 isolates/variants were tested as inactivated viruses in the study and the lowest concentrations detected are listed in Table 3.

In silico analysis on January 15, 2025 indicates 99.9% detection of all available SARS-CoV-2 sequences in the GISAID (>7.94M sequences) and NCBI (>15.04M sequences) databases.

Table 3: Results of Testing SARS-CoV-2 Isolate/Variants

Lineage/Subtype*Isolate/VariantTest Concentration (TCID50/mL)Relative to LoD
AlphaHong Kong/VM20001061/20200.1053x
Beta, B.1.595_2020 (was B.1.2)NY-Wadsworth-33126-01/20200.1053x
Delta, B.1.617.2USA/MD-HP05285/20210.1053x
Epsilon, B.1.427USA/CA/VRLC009/20210.1053x
Gamma, P.1Japan/TY7-503/20210.1053x
Iota, B.1.526_2021USA/NY-Wadsworth-21025952-01/20210.1053x
Kappa, B.1.617.1USA/CA-Stanford-15_S02/20210.1053x
Omicron, B.1.1.529, CH.1.1USA/MD-HP41275/20220.1053x
Omicron, B.1.1.529, XBB.1.5USA/MD-HP40900/20220.1053x
Zeta, P2_2021USA/NY-Wadsworth-21006055-01/20210.1053x

*These strains are in addition to the SARS-CoV-2 USA-WA1/2020 and WHO Standard 20/146, v3, 11/2021 used in the analytical sensitivity study.

Influenza A

The influenza A isolates/variants tested in the study and the lowest concentrations detected are listed in Table 4.

Table 4: Results of Testing Influenza A Isolate/Variants

Lineage/SubtypeIsolate/Variant*Test Concentration (TCID50/mL) äRelative to LoD
H1N1A/Brisbane/59/070.8853x

Page 11

Lineage/SubtypeIsolate/Variant*Test Concentration (TCID50/mL) äRelative to LoD
H1N1A/Christ Church/16/20201.77 EID50/mL6x
H1N1A/Denver/01/570.8853x
H1N1England/221740513/2022**75 copies/mL ‡1x
H1N1A England/224020815/2022**91.5 copies/mL ‡1x
H1N1A/Fort Monmouth/01/470.8853x
H1N1A/Malaya/302/540.8853x
H1N1A/New Caledonia/20/990.8853x
H1N1A/Swine/Iowa/15/301.77 CEID50/mL6x
H1N1A/Sydney/5/2021β1.776x
H1N1A/Victoria/2570/20193.54 EID50/mL12x
H1N1A/WS/330.8853x
H1N1A/Brisbane/14/202375 copies/mL ‡1x
H1N1A/Townsville/1A/202375 copies/mL ‡1x
H1N1A/ Townsville/2A/202375 copies/mL ‡1x
H3N2A/Aichi/2/680.885 CEID50/mL3x
H3N2A/Brisbane/10/070.8853x
H3N2A/Cambodia/E0826360/20200.8853x
H3N2A/Darwin/9/20211.776x
H3N2A/Hong Kong/8/680.8853x
H3N2A/H3/Perth/16/090.8853x
H3N2A/Tasmania/503/20200.885 EID50/mL3x
H3N2A/Victoria/3/750.885 CEID50/mL3x
H3N2A/Wisconsin/67/20050.8853x
H3N2A/Singapore/INFIMH-16-0019/20160.8853x
H5N1A/mallard/Wisconsin/2576/20090.885 CEID50/mL3x
H5N2A/ruddy turnstone/New Jersey/828212/20010.885 CEID50/mL3x
H7N9A/northern shoveler/ Mississippi/ 11OS145/20113.54 CEID50/mL12x

*These strains are in addition to the influenza A Darwin/6/2021 and Brisbane/02/2018 strains used in the analytical sensitivity study.
**A/England/221740513/2022 GISAID ID is EPI_ISL_14387941 and for A/England/224020815/2022 GISAID ID is EPI_ISL_15803829.
‡ The concentrations tested are near the 1x LOD value (69 copies/mL) determined by digital PCR for the influenza A/Darwin/6/2021 strain in Table 8.
ä Concentrations are shown in TCID50/mL unless otherwise indicated.
β Inactivated virus

Page 12

Influenza B

The influenza B isolates/variants tested in the study and the lowest concentrations detected are listed in Table 5.

Table 5: Results of Testing Influenza B Isolate/Variants

Lineage/SubtypeIsolate/Variant*Test Concentration (TCID50/mL) äRelative to LoD
VictoriaB/Brisbane/60/20082.9373x
VictoriaB/Colorado/06/20172.9373x
VictoriaB/Malaysia/2506/042.9373x
VictoriaB/Michigan/09/20112.937 EID50/mL3x
VictoriaB/Washington/02/20192.9373x
YamagataB/Florida/04/062.9373x
YamagataB/Massachusetts/2/20122.9373x
YamagataB/Texas/6/20112.9373x
YamagataB/Texas/81/20162.937 EID50/mL3x
YamagataB/Wisconsin/1/20102.9373x

*These strains are in addition to the influenza B Austria/1359417/2021 and Phuket/3073/2013 strains used in the analytical sensitivity study.
ä Concentrations are shown in TCID50/mL unless otherwise indicated.

5.3. Cross reactivity and microbial interference

Cross-reactivity and microbial interference were evaluated by testing a panel of microorganisms (Table 6). High titer stocks of the potentially cross-reacting microorganisms were tested for cross-reactivity, and also in the presence of SARS-CoV-2, influenza A and influenza B at 3x LoD concentrations for microbial interference. Three (3) replicates in target negative background and three (3) replicates in target positive background were tested for each non-target microorganism. The testing concentrations for potentially interfering viruses are ≥1.0E+05 units/mL except for three viruses (SARS Coronavirus, Urbani, Human Rhinovirus Type 1A, and Human Parainfluenza Virus Type 4A) which were tested at a concentration less than 1.0e+5, but higher than 1.0e+4 units/mL due to their low stock concentration. Other microorganisms (non virus) were tested at ≥1.0E+06 units/mL. Clinical specimens containing Human Coronavirus HKU1 and Pneumocystis jirovecii were also tested (concentration was unknown). None of the organisms tested cross reacted or interfered with cobas® liat SARS-CoV-2 & Influenza A/B v2 performance at the concentrations tested.

Page 13

Table 6: Microorganisms tested for cross-reactivity and microbial interference

Adenovirus Type 1 βMERS-Coronavirus βMycobacterium tuberculosis
Adenovirus Type 7MeaslesMycoplasma genitalium β
CytomegalovirusMumpsMycoplasma pneumoniae
Epstein-Barr virus βRSV (Long/Subtype A)Neisseria elongata
Human Coronavirus OC43RSV (9320/Subtype B)Neisseria flava
Human Coronavirus 229ESARS Coronavirus, Urbani* βNeisseria meningitidis
Human Coronavirus HKU ƗBordetella parapertussisPneumocystis jirovecii Ɨ
Human Coronavirus NL63 βBordetella pertussisPseudomonas aeruginosa
Human Enterovirus 68Chlamydophila pneumoniaeStaphylococcus aureus
Human Metapneumovirus 27Corynebacterium flavescensStaphylococcus epidermidis
Human Parainfluenza Virus Type 1 βEscherichia coliStreptococcus pneumoniae
Human Parainfluenza Virus Type 2Fusobacterium necrophorum subsp. necrophorumStreptococcus pyogenes
Human Parainfluenza Virus Type 3Haemophilus influenzaeStreptococcus salivarius
Human Parainfluenza Virus Type 4A* βLactobacillus crispatusAspergillus flavus var. flavus
Human Rhinovirus Type 1A* βLegionella pneumophilaCandida albicans
Human Rhinovirus BMoraxella catarrhalis-
  • Tested at highest concentration available
    β Inactivated virus Ɨ Clinical specimens at unknown concentrations were tested.

5.4. Competitive inhibition

To assess competitive inhibition between SARS-CoV-2, influenza A and influenza B, each target prepared at high concentration with the other targets at low concentrations (approximately three times the respective LoD) was tested to evaluate any potential impact by the high concentration target on the detection of the other targets (Table 7). Five (5) replicates were tested for each concentration combination.

Testing results indicated that when one viral target was present at high concentration, no interference was observed for the other viral targets that were present at low concentrations (~3x LoD).

Page 14

Table 7: Competitive inhibition strains and concentrations tested with the cobas® liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test

Assay TargetStrain3x LoD Concentration (TCID50/mL)High Concentration Used in Combination with other Low Concentration Targets (TCID50/mL)
SARS-CoV-2*SARS WA 1/20200.1051.14E+05
Influenza AA/Darwin/6/20210.8851.48E+06
Influenza BB/Austria/1359417/20212.9379.79E+06
  • Inactivated virus

5.5. Endogenous and exogenous interference

Potentially interfering substances that may be commonly encountered in respiratory specimens were evaluated. Each substance was tested, by introducing potential interferents. Five (5) replicates were tested with and five (5) replicates were tested without 3x LoD SARS-CoV-2, influenza A and influenza B targets. The substances listed in Table 8 at the concentrations tested did not interfere with the detection of SARS-CoV-2, influenza A or influenza B nor did they produce invalid results in negative samples.

Table 8: Endogenous and Exogenous Interference

Potential InterferentConcentration Tested
Peripheral blood mononuclear cell (PBMC)1.00E+06 cell/mL
Mucin: bovine submaxillary gland, type I-S5 mg/mL
Human Whole Blood5% v/v
Nasal spray - Afrin / Anefrin15% v/v
Nasal corticosteroids - Flonase5% (v/v)
Nasal gel - Zicam5% (v/v)
Throat lozenges, oral anesthetic and analgesic – Cepacol**5 mg/mL
Antibiotic, nasal ointment - Bactroban mupirocin ointment5 mg/mL
Antiviral drug – Relenza5 mg/mL
Antiviral drug - Tamiflu7.5 mg/mL
Antimicrobial, systemic- Tobramycin4 µg/mL
Intranasal Vaccine – FluMist*6.25% (v/v)

*FluMist® contains influenza A and B virus and will test positive if present in the sample. See Procedural limitations.
**One invalid result was obtained in the presence of target analytes. Repeat testing was performed and all targets were detected. The one invalid result was most likely caused by other factors such as general tube processing error, lot variation, etc. that are not related to the interference test condition.

Page 15

5.6. Reproducibility study

A reproducibility study assessed the total variability of the assay in detecting SARS-CoV-2, influenza A and influenza B across operators, study sites, testing days, analyzers, and assay tube lots. The reproducibility was evaluated at three (3) study sites representative of intended use settings. Two (2) operators at each of the three sites tested a 3-member reproducibility panel in triplicate on five different days for three assay tube lots. The reproducibility panel comprises a low positive (1-2x LoD) and a moderate positive (3-5x LoD) co-formulated for SARS-CoV-2, influenza A and influenza B in addition to negative samples. The expected result for the true negative panel member is "Not Detected," while the expected result for the low positive and moderate positive panel members is "Detected." Percent agreement with expected result is shown in Table 9.

Table 9: Reproducibility Results for the cobas® liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test

Target AnalyteExpected Panel Member ConcentrationValid Tests (N)Results in Agreement with Target Analyte (n)Percent Agreement n/N x 10095% Score CI
Negative0265265100.0(98.6, 100.0)
SARS-CoV-21x-2x LoD270270100.0(98.6, 100.0)
SARS-CoV-23x-5x LoD267267100.0(98.6, 100.0)
Influenza A1x-2x LoD26926899.6(97.9, 99.9)
Influenza A3x-5x LoD267267100.0(98.6, 100.0)
Influenza B1x-2x LoD26926899.6(97.9, 99.9)
Influenza B3x-5x LoD267267100.0(98.6, 100.0)

Note: Results were in agreement when a positive panel member had a valid result of "Detected" for the analyte or when the negative panel member had a valid result of "Not Detected" for the analyte.

The means, standard deviations, and coefficients of variation (%) for cycle threshold (Ct) values by target analyte and expected concentration (Positive Panel Members) are shown in Table 10.

Page 16

Table 10: Overall Mean Estimate, Standard Deviations, and Coefficients of Variation (%) for Cycle Threshold Values by Target Analyte and Expected Concentration (Positive Panel Members) for the cobas® liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test

Target AnalyteExpected Concentrationn/NaMean CtSite SDSite CV%Lot SDLot CV%Day SDDay CV%Run SDbRun CV%Within-Run (Residual) SDWithin-Run (Residu-al) CV%Total SDTotal CV%
SARS-CoV-21x-2x LoD270/27033.80.000.00.381.10.170.50.361.11.033.01.163.4
SARS-CoV-23x-5x LoD267/26732.40.130.40.541.70.270.80.000.01.003.11.183.6
Influenza A1x-2x LoD268/26934.80.110.30.080.20.130.40.000.00.621.80.651.9
Influenza A3x-5x LoD267/26733.80.000.00.050.20.000.00.110.30.391.20.411.2
Influenza B1x-2x LoD268/26933.80.040.10.290.90.000.00.130.40.662.00.732.2
Influenza B3x-5x LoD267/26732.80.050.20.471.40.120.40.000.00.521.60.712.2

Note: Ct = cycle threshold; LoD = Limit of Detection; SARS-CoV-2 = Severe acute respiratory syndrome coronavirus-2; SD = standard deviation; CV% = percent coefficient of variation.
a n is the number of positive tests, which contribute Ct values to the analysis. N is the total number of valid tests for the panel member by target analyte.
b In the reproducibility study, each panel member was tested in triplicate, defining one run.

6. CLINICAL PERFORMANCE EVALUATION

6.1. Prospective Study

The clinical performance of cobas® liat SARS-CoV-2 & Influenza A/B v2 for the detection of SARS-CoV-2, influenza A and influenza B was evaluated using paired prospective fresh nasopharyngeal swab (NPS) and anterior nasal swab (ANS) specimens collected in Universal Viral Transport medium (UVT) or Universal Transport Medium (UTM) from individuals with signs and symptoms of respiratory viral infection. For prospectively enrolled subjects an NPS specimen was collected from each subject along with either a self-collected or a healthcare-provider collected ANS specimen. Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) were determined by comparing the results of cobas® liat SARS-CoV-2 & Influenza A/B v2 to the results of an FDA-cleared Nucleic Acid Amplification Test (NAAT).

Page 17

Prospective clinical (Category I) specimens were collected and tested in a non-interventional study between September 2023 and March 2024 at 14 point of care testing sites in the United States (US). Of the 1729 prospective symptomatic subjects enrolled, 1705 NPS specimens were evaluable for analyses, 19 were non-evaluable due to missing or invalid cobas® liat test results, and 5 were non-evaluable due to specimen handling issues. Of the 1729 prospective symptomatic subjects enrolled, 1706 ANS specimens were evaluable in the SARS-CoV-2 and influenza B analyses, 22 were non-evaluable due to missing or invalid cobas® liat test results, and 1 was non-evaluable due to specimen handling issues. Two (2) additional ANS specimens obtained inconclusive comparator results for influenza A, leaving 1704 ANS specimens in the influenza A analysis. Available demographic data regarding the individuals from whom specimens were obtained are presented in Table 11.

Table 11: Demographics of Prospectively Enrolled Individuals

CharacteristicsOverall N(%) (N=1729)
Sex at Birth-
Male681 ( 39.4%)
Female1048 ( 60.6%)
Age Group (Years)-
<133 ( 1.9%)
1 - <18434 ( 25.1%)
18 - <30381 ( 22.0%)
30 - <40246 ( 14.2%)
40 - <50200 ( 11.6%)
50 - <60181 ( 10.5%)
>=60254 ( 14.7%)
Ethnicity-
Hispanic / Latino244 ( 14.1%)
Not Hispanic / Latino1478 ( 85.5%)
Not Reported4 ( 0.2%)
Unknown3 ( 0.2%)
Race-
American Indian / Alaska Native8 ( 0.5%)
Asian24 ( 1.4%)
Black / African American212 ( 12.3%)
Native Hawaiian / Other Pacific Islander3 ( 0.2%)
White1413 ( 81.7%)

Page 18

CharacteristicsOverall N(%) (N=1729)
Other Race52 ( 3.0%)
Not Reported17 ( 1.0%)

For the NPS specimens, cobas® liat SARS-CoV-2 & Influenza A/B v2 demonstrated a PPA and NPA of 94.5% and 97.6% for SARS-CoV-2, respectively; 100.0% and 99.3% for influenza A, respectively; 100.0% and 99.3% for influenza B, respectively (Table 12)). For the ANS specimens, cobas® liat SARS-CoV-2 & Influenza A/B v2 demonstrated a PPA and NPA of 96.7% and 97.2% for SARS-CoV-2, respectively; 100.0% and 99.3% for influenza A, respectively; 100.0% and 99.5% for influenza B, respectively (Table 12). The initial invalid rate of cobas® liat SARS-CoV-2 & Influenza A/B v2 on NPS and ANS specimens was 0.5% and 0.7% respectively. Upon repeat testing, the final assay invalid rate on NPS and ANS specimens was 0% and 0.1% respectively.

Table 12: Clinical Performance of cobas® liat SARS-CoV-2 & Influenza A/B v2 relative to the comparator by Specimen Type and Target

TargetSpecimen Typea/ (a+c)PPA (%)PPA 95% CId/ (b+d)NPA (%)NPA 95% CI
SARS-CoV-2*NPS207/21994.590.7-96.81451/148697.696.7-98.3
SARS-CoV-2**ANS208/21596.793.4-98.41449/149197.296.2-97.9
Influenza A#NPS54/54100.093.4-100.01640/165199.398.8-99.6
Influenza A##ANS53/53100.093.2-100.01640/165199.398.8-99.6
Influenza B&NPS22/22100.085.1-100.01671/168399.398.8-99.6
Influenza B&&ANS24/24100.086.2-100.01673/168299.599.0-99.7

Abbreviations: PPA = Positive Percent Agreement; CI = Confidence Interval; NPA = Negative Percent Agreement; NPS = Nasopharyngeal swab; ANS = Anterior Nasal Swab; SARS-CoV-2 = Severe acute respiratory syndrome coronavirus 2.

Note: N = Total number of specimens; a = number of specimens where both cobas® liat and the comparator are positive; b = number of specimens where cobas® liat is positive and the comparator is negative; c = number of specimens where cobas® liat is negative and the comparator is positive; d = number of specimens where both cobas® liat and the comparator are negative.

  • SARS-CoV-2 NPS discrepant NAAT results: Of 12 specimens negative on cobas® liat and positive on the comparator, 8 were positive and 4 were negative. Of 35 specimens positive on cobas® liat and negative on the comparator, 12 were positive and 23 were negative.

Influenza A NPS discrepant NAAT results: Of 11 specimens positive on cobas® liat and negative on the comparator, 2 were positive and 9 were negative.

& Influenza B NPS discrepant NAAT results: Of 12 specimens positive on cobas® liat and negative on the comparator, all 12 were negative.
** SARS-CoV-2 ANS discrepant NAAT results: Of 7 specimens negative on cobas® liat and positive on the comparator, 6 were positive and 1 was negative. Of 42 specimens positive on cobas® liat and negative on the comparator, 8 were positive and 34 were negative.

Page 19

Influenza A ANS discrepant NAAT results: Of 11 specimens positive on cobas® liat and negative on the comparator, 1 was positive and 10 were negative.

&& Influenza B ANS discrepant NAAT results: Of 9 specimens positive on cobas® liat and negative on the comparator, all 9 were negative.

6.2. Retrospective Study

Influenza B was of lower prevalence and was not encountered in sufficiently large numbers during the prospective clinical study to adequately demonstrate assay performance. To supplement the results of the prospective clinical study, retrospective frozen clinical NPS and ANS specimens collected in Universal Viral Transport medium (UVT) or Universal Transport Medium (UTM) from individuals with signs and symptoms of respiratory viral infection were tested. Frozen archived (Category III) influenza B positive and negative NPS (n=223) and ANS (n=206) specimens obtained between 2019 and 2023 were distributed to 6 sites and tested in the course of the daily workflow. One NPS sample pre-characterized as positive for influenza B was non-evaluable due to missing/invalid results on the comparator method. The comparator method was an acceptable FDA-cleared molecular assay. Available demographic data regarding the individuals from the retrospective study are shown in Table 13.

Table 13: Demographics of Subjects from Retrospective Population

CharacteristicsOverall N(%) (N=429)
Sex at Birth-
Male151 ( 35.20%)
Female226 ( 52.68%)
Not Reported52 ( 12.12%)
Age Group (Years)-
<10 ( 0.00%)
1 - <18129 ( 30.07%)
18 - <3077 ( 17.95%)
30 - <4079 ( 18.41%)
40 - <5028 ( 6.53%)
50 - <6014 ( 3.26%)
>=6023 ( 5.36%)
Not Reported79 ( 18.41%)
Ethnicity-
Hispanic / Latino98 ( 22.84%)

Page 20

CharacteristicsOverall N(%) (N=429)
Not Hispanic / Latino112 ( 26.11%)
Not Reported213 ( 49.65%)
Other5 ( 1.17%)
Unknown1 ( 0.23%)
Race-
American Indian / Alaska Native2 ( 0.47%)
Asian3 ( 0.70%)
Black / African American12 ( 2.80%)
White196 ( 45.69%)
Other Race3 ( 0.70%)
Unknown5 ( 1.17%)
Not Reported208 ( 48.48%)

For the retrospective NPS specimens, cobas® liat SARS-CoV-2 & Influenza A/B v2 demonstrated a PPA and NPA of 100.0% and 97.9% for influenza B, respectively (Table 14). For the retrospective ANS specimens, cobas® liat SARS-CoV-2 & Influenza A/B v2 demonstrated a PPA and NPA of 100.0% and 98.3% for influenza B, respectively (Table 14). No invalid results were observed for the cobas® liat SARS-CoV-2 & Influenza A/B v2 with retrospective NPS or ANS specimens for an invalid rate of 0.0%.

Table 14: Agreement of cobas® liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test and Comparator Test for influenza B in Retrospective Samples

TargetSpecimen Typea/(a+c)PPA (%)PPA 95% CId/(b+d)NPA (%)NPA 95% CI
Influenza B*NPS34/34100.089.8-100.0184/18897.994.7-99.2
Influenza B**ANS34/34100.089.8-100.0169/17298.395.0-99.4

Abbreviations: PPA = Positive Percent Agreement; CI = Confidence Interval; NPA = Negative Percent Agreement; NPS = Nasopharyngeal swab; ANS = Anterior Nasal Swab; SARS-CoV-2 = Severe acute respiratory syndrome coronavirus 2.

Note: N = Total number of specimens; a = number of specimens where both cobas® liat and the comparator are positive; b = number of specimens where cobas® liat is positive and the comparator is negative; c = number of specimens where cobas® liat is negative and the comparator is positive; d = number of specimens where both cobas® liat and the comparator are negative.

  • Influenza B NPS discrepant NAAT results: Of 4 specimens positive on cobas® liat and negative on the comparator, all 4 were negative.
    ** Influenza B ANS discrepant NAAT results: Of 3 specimens positive on cobas® liat and negative on the comparator, 1 was positive and 2 were negative.

Page 21

7. CONCLUSIONS

A comparison of the intended use, technological characteristics, and the results of non-clinical analytical and clinical performance studies demonstrate that cobas® liat SARS-CoV-2 & Influenza A/B v2 nucleic acid test is substantially equivalent to the predicate device.

§ 866.3981 Device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test.

(a)
Identification. A device to detect and identify nucleic acid targets in respiratory specimens from microbial agents that cause the SARS-CoV-2 respiratory infection and other microbial agents when in a multi-target test is an in vitro diagnostic device intended for the detection and identification of SARS-CoV-2 and other microbial agents when in a multi-target test in human clinical respiratory specimens from patients suspected of respiratory infection who are at risk for exposure or who may have been exposed to these agents. The device is intended to aid in the diagnosis of respiratory infection in conjunction with other clinical, epidemiologic, and laboratory data or other risk factors.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The intended use in the labeling required under § 809.10 of this chapter must include a description of the following: Analytes and targets the device detects and identifies, the specimen types tested, the results provided to the user, the clinical indications for which the test is to be used, the specific intended population(s), the intended use locations including testing location(s) where the device is to be used (if applicable), and other conditions of use as appropriate.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed descriptions of the performance characteristics of the device for each specimen type claimed in the intended use based on analytical studies including the following, as applicable: Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, precision, reproducibility, and clinical studies;
(iii) Detailed descriptions of the test procedure(s), the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) A warning statement that viral culture should not be attempted in cases of positive results for SARS-CoV-2 and/or any similar microbial agents unless a facility with an appropriate level of laboratory biosafety (
e.g., BSL 3 and BSL 3+, etc.) is available to receive and culture specimens; and(v) A prominent statement that device performance has not been established for specimens collected from individuals not identified in the intended use population (
e.g., when applicable, that device performance has not been established in individuals without signs or symptoms of respiratory infection).(vi) Limiting statements that indicate that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) There is a risk of incorrect results due to the presence of nucleic acid sequence variants in the targeted pathogens;
(D) That positive and negative predictive values are highly dependent on prevalence;
(E) Accurate results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(F) When applicable (
e.g., recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer-reviewed literature), that the clinical performance may be affected by testing a specific clinical subpopulation or for a specific claimed specimen type.(4) Design verification and validation must include:
(i) Detailed documentation, including performance results, from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, as appropriate, additional characterized clinical samples. The clinical study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained using a comparator that FDA has determined is appropriate. Detailed documentation must include the clinical study protocol (including a predefined statistical analysis plan), study report, testing results, and results of all statistical analyses.
(ii) Risk analysis and documentation demonstrating how risk control measures are implemented to address device system hazards, such as Failure Modes Effects Analysis and/or Hazard Analysis. This documentation must include a detailed description of a protocol (including all procedures and methods) for the continuous monitoring, identification, and handling of genetic mutations and/or novel respiratory pathogen isolates or strains (
e.g., regular review of published literature and periodic in silico analysis of target sequences to detect possible mismatches). All results of this protocol, including any findings, must be documented and must include any additional data analysis that is requested by FDA in response to any performance concerns identified under this section or identified by FDA during routine evaluation. Additionally, if requested by FDA, these evaluations must be submitted to FDA for FDA review within 48 hours of the request. Results that are reasonably interpreted to support the conclusion that novel respiratory pathogen strains or isolates impact the stated expected performance of the device must be sent to FDA immediately.(iii) A detailed description of the identity, phylogenetic relationship, and other recognized characterization of the respiratory pathogen(s) that the device is designed to detect. In addition, detailed documentation describing how to interpret the device results and other measures that might be needed for a laboratory diagnosis of respiratory infection.
(iv) A detailed device description, including device components, ancillary reagents required but not provided, and a detailed explanation of the methodology, including molecular target(s) for each analyte, design of target detection reagents, rationale for target selection, limiting factors of the device (
e.g., saturation level of hybridization and maximum amplification and detection cycle number, etc.), internal and external controls, and computational path from collected raw data to reported result (e.g., how collected raw signals are converted into a reported signal and result), as applicable.(v) A detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(vi) For devices intended for the detection and identification of microbial agents for which an FDA recommended reference panel is available, design verification and validation must include the performance results of an analytical study testing the FDA recommended reference panel of characterized samples. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(vii) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens, the design verification and validation must include a detailed description of the identity, phylogenetic relationship, or other recognized characterization of the Influenza A and B viruses that the device is designed to detect, a description of how the device results might be used in a diagnostic algorithm and other measures that might be needed for a laboratory identification of Influenza A or B virus and of specific Influenza A virus subtypes, and a description of the clinical and epidemiological parameters that are relevant to a patient case diagnosis of Influenza A or B and of specific Influenza A virus subtypes. An evaluation of the device compared to a currently appropriate and FDA accepted comparator method. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.
(5) When applicable, performance results of the analytical study testing the FDA recommended reference panel described in paragraph (b)(4)(vi) of this section must be included in the device's labeling under § 809.10(b) of this chapter.
(6) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiation between the Influenza A virus subtypes in human clinical specimens in addition to detection of SARS-CoV-2 and similar microbial agents, the required labeling under § 809.10(b) of this chapter must include the following:
(i) Where applicable, a limiting statement that performance characteristics for Influenza A were established when Influenza A/H3 and A/H1-2009 (or other pertinent Influenza A subtypes) were the predominant Influenza A viruses in circulation.
(ii) Where applicable, a warning statement that reads if infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to State or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
(iii) Where the device results interpretation involves combining the outputs of several targets to get the final results, such as a device that both detects Influenza A and differentiates all known Influenza A subtypes that are currently circulating, the device's labeling must include a clear interpretation instruction for all valid and invalid output combinations, and recommendations for any required followup actions or retesting in the case of an unusual or unexpected device result.
(iv) A limiting statement that if a specimen yields a positive result for Influenza A, but produces negative test results for all specific influenza A subtypes intended to be differentiated (
i.e., H1-2009 and H3), this result requires notification of appropriate local, State, or Federal public health authorities to determine necessary measures for verification and to further determine whether the specimen represents a novel strain of Influenza A.(7) If one of the actions listed at section 564(b)(1)(A) through (D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those influenza viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized influenza viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's labeling required under § 809.10(b) of this chapter that accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that accompanies the device, prominently providing a hyperlink to the manufacturer's public website where the analytical reactivity testing data can be found. The manufacturer's website, as well as the primary part of the manufacturer's website that discusses the device, must provide a prominently placed hyperlink to the website containing this information and must allow unrestricted viewing access.