Search Filters

Search Results

Found 57 results

510(k) Data Aggregation

    K Number
    K221634

    Validate with FDA (Live)

    Date Cleared
    2023-05-24

    (352 days)

    Product Code
    Regulation Number
    868.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Oxymag is a controlled volume, pressure and time cycled emergency and transport ventilator. It is intended for use with infant, child, and adult patients with a tidal volume from 50 ml upwards who are in respiratory and/or cardiac arrest or respiratory distress and who require the ventilatory support.

    It is intended for pre-hospital and hospital use including intra-hospital, inter-hospital and transport settings.

    Device Description

    Oxymag provides a mixture of ambient air and oxygen at concentrations adjusted by the operator using the accurate oxygen concentration system using the venturi principle. O2 concentration is obtained through a galvanic cell by passing gas through the sensor. In addition, it performs the control of flows and pressures in the respiratory circuit to provide the ventilation modalities appropriate to the patient's condition.

    The associated accessories include:

    • Power outlet 12V/3,34A
    • AC cable
    • Disposable respiratory circuit
    • 02 extension
    • Environment filters
    AI/ML Overview

    The provided document is a 510(k) summary for the Oxymag - Transport and Emergency Ventilator. It focuses on demonstrating substantial equivalence to a predicate device (O-two e700, K141595) rather than establishing novel performance acceptance criteria or conducting a comprehensive study to prove these criteria.

    Therefore, the document does not contain the specific information requested in most of your numbered points, as it's not a study designed to establish and prove acceptance criteria for a new device's performance in a clinical setting with human subjects. Instead, it describes bench testing, software verification, electrical safety, and biocompatibility to confirm the device operates as intended and is comparable to an already cleared device.

    Here's what can be extracted and what is missing:


    Acceptance Criteria and Device Performance (Based on Comparison to Predicate):

    The submission demonstrates substantial equivalence by comparing the Oxymag's parameters and features to those of the predicate device, e700. The "acceptance criteria" are implicitly that the Oxymag's performance is equivalent or does not raise new safety or effectiveness concerns compared to the predicate.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Since this is a 510(k) summary, formal "acceptance criteria" for a study are not explicitly stated in the way one might find in a clinical trial protocol. Instead, the performance is demonstrated by comparison with a predicate device and through various forms of engineering and non-clinical testing.

    The provided "Table 1. Comparison table between the subject and predicate devices" serves as the primary evidence for this. Instead of a 'reported device performance' against a strict new acceptance criterion, it shows 'equivalence' or 'similarity' to the predicate.

    CharacteristicAcceptance Criteria (Implied: Equivalent/Similar to Predicate)Reported Device Performance (Oxymag)Discussion/Proof of Equivalence
    Indications for UseEquivalent to e700Controlled volume, pressure and time cycled emergency and transport ventilator for infant, child, and adult patients (tidal volume 50ml upwards) in respiratory/cardiac arrest or distress requiring ventilatory support. For pre-hospital and hospital use (intra-hospital, inter-hospital, transport settings).The indications for use are equivalent to e700.
    Patient PopulationEquivalent to e700Infant, child, and adult patientsThe patient population is equivalent to e700.
    Environment of UseEquivalent to e700Pre-hospital and hospital use including intra-hospital, inter-hospital and transport settings.The environment of use is equivalent to e700.
    Product CodeEquivalent to e700BTLThe product code is equivalent to e700.
    Ventilation ModesSimilar to e700VCV, PCV, V-SIMV, P-SIMV, CPAP/PSSimilar to e700. Except for CPR (not in Oxymag), other modes have different nomenclature but are equivalent (e.g., ACV = VCV/PCV, SIMV = V-SIMV, BiLVL = P-SIMV, CPAP is same).
    Breathing CircuitPerformance similar to e700Dual limb, unique for all patient typesDifferent from e700 (which is single limb). "Despite the difference, the performance of both devices is similar as demonstrated in bench test."
    Exhalation ValvePerformance similar to e700Connected to the equipment in the exhalation connector. Electronically controlled.Different from e700 (which has pneumatic control, valve in circuit). "Despite this difference, the exhalation for both devices is similar."
    WaveformsEquivalent to e700Volume-time, pressure-time and flow-timeWaveforms are equivalent to e700 as demonstrated in bench test.
    Flow SensorEquivalent to e700PneumotachographFlow sensor is equivalent to e700.
    Flow ControlPerformance similar to e700Proportional valves controlled by microprocessorSimilar to e700 (solenoid valves). "Despite the difference, the performance of both devices is similar as demonstrated in bench test."
    Trigger SensitivityParameter range equivalent to e700OFF; 1 to 15 L/minParameter range is equivalent to e700.
    Input PressurePerformance not affected by difference from e70039 to 87 psiDifferent to e700 (45-87 PSI). "The lower limit of Oxymag's input pressure is lower than e700, but the difference does not affect the performance of Oxymag. Oxymag delivers all ventilation parameters between 39 and 87 psi."
    PSV (Pressure Support Ventilation)Parameter range equivalent to e700OFF; 4 to 35 cmH2O (± 10% or ± 2 cmH2O)Parameter range is equivalent to e700.
    Ventilation FrequencyParameter range equivalent to e7005 to 60 breath/min (± 10% or ± 1 bpm)Parameter range is equivalent to e700.
    Tidal Volume (L)Parameter range equivalent to e70050 to 2000 mL (±20ml or ±15%)Parameter range is equivalent to e700.
    Manual Ventilation/Inspiration HoldEquivalent to e700YesParameter range is equivalent to e700.
    Inspiration time to expiration time ratioParameter range equivalent to e7001:4 to 3:1 (± 20%)Parameter range is equivalent to e700.
    Inspiration time Ti (sec.)Parameter range equivalent to e7000.2 to 9 s (± 20%)Parameter range is equivalent to e700.
    PEEP/CPAP (cm H2O)Parameter range equivalent to e700OFF; 4 to 20 (± 10% or ± 2 cmH2O)Parameter range is equivalent to e700.
    FiO2 (%)Parameter range equivalent to e70060 or 100 (± 15%)Parameter range is equivalent to e700.
    PmaxDoes not raise safety questions10 to 60 (± 10% or ± 2 cmH2O)Different to e700 (10-80 cmH2O). Oxymag has a more restricted upper limit, which generally implies enhanced safety.
    Safety relief valveEquivalent to e700YesSafety valve is equivalent to e700.
    Inhalation pressure (cmH2O)Parameter range equivalent to e700OFF; 4 to 50 cmH2O (± 10% or ± 2 cmH2O)Parameter range is equivalent to e700.
    Apnea Backup timeParameter range equivalent to e70010 to 60 secParameter range is equivalent to e700.
    MonitoringSimilar to e700, does not affect performanceMinute Volume, Volume Measured, Instant pressure measured, maximum inspiratory pressure, Respiratory Rate, Plateau pressure, PEEP, Flow, inspiratory time, expiratory time, Ratio I:E, Airway resistance, Dynamic compliance, Static compliance, FiO2, O2 consumptionOxymag has monitored parameters that e700 has and additional others, but the difference does not affect the performance of Oxymag.
    Waveform DisplayedEquivalent to e700Pressure and flowWaves are equivalent to e700.
    Alarms Audible/Visual & IndicationsSignificant alarms similar to e700; differences do not raise safety questionsDisconnection, Low airway pressure, High airway pressure, Low minute volume, High minute volume, Obstruction, Low supply pressure, Apnea, Low battery, High PEEP, Low PEEP, AC input fail, High volume and Low volume, High r. rate, Low r. rate, Low internal temperature, High internal temperature, Flow sensor off, HW:High O2 int.Significant alarms are similar to e700. Differences like additional ISO-required alarms or different alarm trigger mechanisms are stated not to raise safety questions.
    AccessoriesSignificant accessories equivalent to e700AC/DC power supply, Patient ventilation circuit, Oxygen supply hose, Environment filterSignificant accessories are equivalent to e700.
    Reprocessing of Patient CircuitEquivalent to e700Single usePatient circuit reprocessing is equivalent to e700.

    Regarding the Study (Performance Data provided for Substantial Equivalence):

    The document explicitly states: "There were no animal or clinical studies done for the subject device." This means that most of the specific questions about sample sizes, human experts, adjudication, and MRMC studies are not applicable to the data presented for this 510(k) submission.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not applicable. No human test subjects were used. Performance data was derived from bench testing.
    • Data Provenance: Not applicable for human data. Bench testing was performed on the device itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. No human test set requiring expert ground truth was used.

    4. Adjudication method for the test set:

    • Not applicable. No human test set was used for adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. No MRMC study was done, as this is a ventilator and not an AI-assisted diagnostic device, and no human studies were conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable in the context of an algorithm's diagnostic performance. However, "Performance testing was conducted on the Oxymag to determine its specifications regarding ventilatory parameters" and "verification of technical data, comparison between Oxymag and the predicate e700, human factors evaluation, validation of ventilatory modes, alarm system, monitored parameters, auto test, hardware and mechanical specifications" serve as standalone device performance evaluations in a non-clinical setting.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For Biocompatibility: International standard ISO 10993-1 and FDA Guidance as recognized by FDA.
    • For Electrical Safety and EMC: IEC 60601-1, IEC 60601-1-8, IEC 60601-1-12, IEC 80601-2-12, IEC 80601-2-55 for safety; IEC 60601-1-2 for EMC; AIM 7351732 for RFID.
    • For Software: FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
    • For Mechanical Testing: Device specifications based on engineering design and comparison to the predicate device's established performance. The "ground truth" here is the established and expected performance curves and parameter ranges for ventilators, often derived from industry standards and the predicate device's known capabilities.

    8. The sample size for the training set:

    • Not applicable. There is no mention of a training set as this is not an AI/machine learning device requiring such.

    9. How the ground truth for the training set was established:

    • Not applicable. No training set was used.
    Ask a Question

    Ask a specific question about this device

    K Number
    K202219

    Validate with FDA (Live)

    Date Cleared
    2021-02-11

    (189 days)

    Product Code
    Regulation Number
    868.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VORTRAN GO2VENT with PEEP Valve is to be used by properly trained personnel to deliver emergency, shortterm, constant flow, pressure-cycled ventilatory support on patients weighing 10 kg and above.

    Device Description

    The device consists of a modulator (an exhalation valve that opens at PIP and closes at PEEP), a mechanical manometer, a resistor (which is referred to as the PEEP Valve) attached within the exhalation path for elevating the PEEP further than can be set with the modulator, a patient connector tee to supply gas flow, entrain additional air, and connect to a face mask or a tube inserted into a patient's airway, and two associated accessories. It is intended to be used by properly trained personnel in any environment in which emergency ventilation is required.

    The VORTRAN GO₂VENT with PEEP Valve provides short-term, constant flow, pressure-cycled ventilatory support in either pressure control or pressure support mode. In pressure support mode, the rate dial of the VORTRAN GO₂VENT with PEEP Valve is set so that the baseline pressure is set above the set PEEP. This allows the patient to initiate inhalation by drawing the baseline pressure down to the set PEEP. In pressure control mode, the device will automatically cycle between PIP and PEEP when connected to a patient's airway.

    The working mechanism of the VORTRAN GO₂VENT with PEEP Valve consists of a moving diaphragm that opens the exhalation path when the pressure reaches PIP and closes when the pressure reaches PEEP. Without the PEEP Valve, PEEP will be approximately 20% of the set PIP. With the PEEP Valve attached, the PEEP will be increased from this value up to 24 cm-H₂O depending on the patient's compliance and the VORTRAN GO₂VENT with PEEP Valve's settings. The PEEP Valve is attached within the exhalation path by connecting it between the modulator and the patient connector tee. The pressure at which the diaphragm opens and closes against the valve is controlled by the amount of spring force acting against the diaphragm's movement. The spring force is adjusted by manually turning a threaded knob that varies the amount of spring compression.

    The device is constructed of a variety of plastics such as K-Resin, HDPE, polycarbonate, and silicone, as well as copper beryllium springs. The VORTRAN GO₂VENT with PEEP Valve is an external communicating device with limited and prolonged tissue contact duration (up to 30 days) via the breathing gas pathway.

    The associated accessories were included with the cleared predicate device (K162968) and include:

    • . Elbow flex hose for connecting the patient connector tee to a face mask or a tube inserted into a patient's airway
    • Oxygen tubing for connecting the patient connector tee to a gas supply
    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the VORTRAN GO2VENT with PEEP Valve. It includes information on performance testing which addresses acceptance criteria.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of acceptance criteria and the reported device performance

    Acceptance Criteria (Standard)Reported Device Performance (Compliance)
    ISO 10993-1:2018 "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" and ISO 18562-1:2017 "Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 1: Evaluation and testing within a risk management process."Biocompatibility testing was not needed as a result of the evaluation to these standards. The device is an external communicating device with limited and prolonged tissue contact duration (up to 30 days) via the breathing gas pathway.
    ISO 10651-5:2006 "Lung ventilators for medical use – Particular requirements for basic safety and essential performance – Part 5: Gas-powered emergency resuscitators."The subject device complies with this standard. Performance testing was conducted to determine its specifications regarding ventilatory parameters.
    ISO 5356-1:2015 "Anaesthetic and respiratory equipment – Conical connectors – Part 1: Cones and sockets."Connection ports on the VORTRAN GO2VENT with PEEP Valve were designed to this standard.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample size used for the mechanical performance testing. It also does not provide details on the data provenance, such as country of origin or whether the studies were retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and not provided in the document. The device is a medical device, and its performance is evaluated against engineering and biological standards, not through expert-labeled data like in AI/imaging studies.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and not provided in the document for the reasons stated above.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The device is a powered emergency ventilator, not an AI or imaging diagnostic tool that would involve human readers or AI assistance in interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable for the reasons stated above.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For biocompatibility, the "ground truth" was regulatory guidance and recognized ISO standards. For mechanical testing, the "ground truth" was compliance with internationally recognized ISO standards for lung ventilators and connectors.

    8. The sample size for the training set

    This information is not applicable as this is a medical device approval, not an AI/machine learning model.

    9. How the ground truth for the training set was established

    This information is not applicable as this is a medical device approval, not an AI/machine learning model. The relevant "ground truth" for compliance was established through recognized international standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193191

    Validate with FDA (Live)

    Date Cleared
    2020-11-20

    (367 days)

    Product Code
    Regulation Number
    868.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MEDUMAT Easy CPR is an electrical, pneumatically operated emergency and transport ventilation used for ventilation and oxygen inhalation with either a mask or tube.

    Patient groups
    Adults and children with a body weight of over 22 lbs (10 kg) where spontaneous respiration has failed or is inadequate.

    Users Qualified medical personnel only

    Intended environments of use
    · Mobile use for emergency medicine and primary care during emergency deployments
    · During land or air transport or transfer between hospital rooms and departments

    Device Description

    MEDUMAT Easy CPR is an electrical, pneumatically operated emergency and transport ventilator. Highly compressed medical oxygen is used as the ventilation gas; this is reduced to the necessary operating pressure via an external pressure reducer. The oxygen is supplied at the compressed gas connection. The ventilation parameters – frequency and tidal volume - are linked together and can be set using the adjusting knob on the device. The ventilation gas is transported to the patient through the ventilation hose via the patient valve and ventilation mask or via the tube. The lip membrane in the patient valve allows the patient exhalation of expiration gas. In order to monitor the patient, the device features continuous measurement of the airway pressure as well as a visual and audible alarm system.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study that proves the device meets those criteria for the MEDUMAT Easy CPR device.

    Here's a breakdown of the requested information:

    1. Table of acceptance criteria and the reported device performance:

    The document doesn't explicitly list "acceptance criteria" for each performance metric in a table format with specific numerical targets. Instead, it states that various tests were conducted to "verify the safety and effectiveness" and that "the performance of MEDUMAT Easy CPR is at least as good as the performance of the predicate device." The "Test results" section for performance testing only indicates "Passed."

    However, we can infer some criteria and reported performance from the comparison table and performance testing section.

    FeatureAcceptance Criteria (Inferred)Reported Device Performance
    BiocompatibilityMeet ISO 10993 and ISO 18562 series standardsPassed (Non-cytotoxic, non-sensitizer, non-irritant, no systemic toxicity, exposures to particulates and VOCs unlikely to result in toxicological effects, no organic compounds identified as leachable)
    EMC and Electrical SafetyMeet IEC 60601-1-2, -6, -8, -12 and AIM 7351731 standardsPassed
    Ventilation Mode Characteristics (IPPV, Manual Mode)Similar to predicate device (Pneupac VR1 Standard)Passed
    Demand Flow Mode (Inspiration Trigger, Expiration Trigger, Flow)Similar to predicate device (Pneupac VR1 Standard)Passed
    Pressure LimitsSimilar to predicate device (Pneupac VR1 Standard)Passed (Differences noted but determined not to raise additional concerns)
    Endurance and Functional Test of Pressure Relief Valve / Safety ValveFunctional operation and endurancePassed
    Inspiratory and Expiratory Respiratory ResistancesComparable to predicate device (Pneupac VR1 Standard) across different operating conditionsPassed
    Tidal Volume (Vt) Accuracy± 40 ml or ± 20% (the larger tolerance applies)Stated as the tolerance; assumed to be met. (Predicate device has similar tolerances: ±15%)
    Operating Conditions (Temperature, Air Pressure, Humidity)Within specified ranges and similar to predicate deviceTested for the subject device; operating conditions are similar and tested.
    Alerts (High Airway Pressure, Low Airway Pressure, Low gas supply, Low battery)Functionality as specifiedPresent and stated to increase safety. (Predicate only has high airway pressure alarm).

    2. Sample size used for the test set and the data provenance:

    The document does not specify the sample size for the test set in terms of number of devices or data points used for each performance test.
    The data provenance is not explicitly mentioned (e.g., country of origin, retrospective/prospective). The tests were conducted internally by the manufacturer (Weinmann Emergency Medical Technology GmbH + Co. KG, Germany) or by testing labs on their behalf.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. The performance tests appear to be engineering verification and validation against technical specifications and predicate device performance, rather than clinical studies requiring expert ground truth establishment.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not applicable/provided. The performance testing described is objective, technical verification against standards and predicate device characteristics, not subjective assessment requiring adjudication by multiple experts.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The MEDUMAT Easy CPR is a hardware device (ventilator), not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study is irrelevant to this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not applicable. The device is a ventilator, not an algorithm. Performance testing was done on the device itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the performance testing can be considered:

    • International Standards: For biocompatibility (ISO 10993, ISO 18562) and electrical safety/EMC (IEC 60601 series, AIM 7351731).
    • Predicate Device Performance: For ventilation characteristics, pressure limits, demand flow mode, and respiratory resistances, the predicate device (Pneupac VR1 Standard) served as a benchmark for comparison.
    • Internal Specifications/Design Requirements: For various technical data and functional tests.

    8. The sample size for the training set:

    This is not applicable. The MEDUMAT Easy CPR is a hardware medical device and does not involve machine learning or AI algorithms that require a "training set."

    9. How the ground truth for the training set was established:

    This is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182956

    Validate with FDA (Live)

    Date Cleared
    2019-01-18

    (86 days)

    Product Code
    Regulation Number
    868.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Puffin Lite Infant Resuscitator is intended to provide the basic equipment required for pulmonary resuscitation of neonatal infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the neonatal infant.

    Device Description

    The resuscitation system provides the basic equipment required for pulmonary resuscitation of neonatal infants. The Puffin Lite Infant Resuscitator is a gas powered emergency resuscitation system. It is intended to be used inside the hospital by trained medical professionals to provide precise FIO2 delivery, and manual ventilation as established by resuscitation guidelines to neonatal infants weighing less than 10 kg (22 lb). The resuscitation system includes two medical gas flowmeters, an integrated oxygen blender, airway pressure manometer, peak inspiratory pressure (PIP) control, positive end expiratory pressure (PEEP) control, a gas supply pressure alarm and T-piece resuscitator.

    AI/ML Overview

    The Puffin Lite Infant Resuscitation System underwent bench testing to demonstrate conformance to performance specifications.

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance CriteriaReported Device Performance
    Valve Function after VomitusThe proper function of the circuit shall be verified within 20 seconds of becoming disabled by vomitus. Function is verified by verifying flow valve accuracy.Passed
    Inspiratory ResistancePressure generated at the patient connection port during expiration should not exceed -5 cmH2O with inspiratory airflow set to 6 L/min.Passed
    Expiratory ResistancePressure generated at the patient connection port during expiration should not exceed 5 cmH2O with expiratory airflow set to 6 L/min.Passed
    Dead SpaceThe deadspace volume of the T-Piece circuit should be less than 7 mL.Passed
    FIO2 accuracyThe proper function of the FIO2 adjustment knob shall be verified by comparing the FIO2 setting value with the output oxygen concentration. Values shall be within 5%.Passed
    Primary and Secondary Flow Valve - Peak FlowThe proper function of the primary and secondary flow valves shall be verified by comparing the flow setting with the actual measured output flow.Passed
    Airway Manometer AccuracyThe proper function of the airway manometer shall be verified by comparing the pressure readings with the actual measured output pressure.Passed
    VOC TestingThe device should not add volatile organic compounds (VOCs) to the output gas delivered to the patient.Passed
    Particulate AnalysisThe output of particulate matter sizes 2.5 microns or less are no more than 12 micrograms/cubic meter of air at one atmospheric pressure.Passed

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the specific sample sizes used for each individual bench test. The testing was described as "bench testing," which typically involves laboratory-controlled environments and simulated conditions rather than real patient data. The data provenance is from this bench testing, not from human subjects or real-world clinical data. Therefore, there is no information on country of origin or whether the data was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable. The study involved bench testing of a medical device, where acceptance criteria were based on engineering and performance specifications (e.g., pressure, flow, FIO2 accuracy) rather than expert interpretation of medical images or data. Ground truth was established by measuring the physical performance parameters of the device against predefined technical requirements.

    4. Adjudication method for the test set:

    Not applicable. As the study was bench testing against predefined technical specifications, there was no need for an adjudication method by human experts for the "ground truth." The "passing" or "failing" of a test was determined by whether the device's measured performance fell within the specified limits.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a 510(k) submission for a physical medical device (infant resuscitator) and not an AI or imaging diagnostic device. Therefore, an MRMC study comparing human reader performance with and without AI assistance was not conducted.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable. This device is not an algorithm or an AI system. It is a physical medical device (infant resuscitator) that is intended for use by trained medical professionals.

    7. The type of ground truth used:

    The ground truth used for this study was the engineering and performance specifications outlined in the acceptance criteria. The device's performance was measured against these objective, predefined technical requirements. There was no reliance on expert consensus, pathology, or outcomes data in the context of this bench testing.

    8. The sample size for the training set:

    Not applicable. This device is not an AI or machine learning system that requires a "training set" of data.

    9. How the ground truth for the training set was established:

    Not applicable, as there was no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173373

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-11-16

    (385 days)

    Product Code
    Regulation Number
    868.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SafeT T-Piece Resuscitator is a gas-powered emergency resuscitator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway. It is intended for use with neonates and infants weighing less than 10 kg (22 lb).

    Device Description

    The SafeT T-Piece Resuscitator is a single-use, non-sterile, manually operated, gas-powered resuscitator for use with patients less than 10 kg (22 lb). It is a simple T-Piece resuscitator with a manometer and the ability to adjust Peak Inspiratory Pressure (PIP) and Positive End-Expiratory Pressure (PEEP). It incorporates a pressure relief valve to protect against excessive pressure. The T-Piece resuscitator can be connected to the patient via a face mask or tube. The subject device consists of several components: T-Piece patient valve with variable PEEP dial and integrated manometer, Adjustable inspiratory pressure controller, 40 cm H2O pressure relief valve, 7' Oxygen tubing with a red universal (Fits-all) connector, 20" x 10 mm circuit tubing, Face mask.

    AI/ML Overview

    The provided text is a 510(k) summary for the Ventlab, LLC SafeT T-Piece Resuscitator, a medical device for emergency respiratory support. It details the device's characteristics and its comparison to predicate devices to establish substantial equivalence for FDA clearance.

    Crucially, this document does not describe a study involving an AI/Machine Learning device or a Multi-Reader Multi-Case (MRMC) study. It focuses on the substantial equivalence of a physical medical device (a resuscitator) based on its technological characteristics and performance testing against established standards, not on an algorithm's performance in interpreting medical images or data.

    Therefore, many of the requested points, such as acceptance criteria for AI performance, sample sizes for test/training sets, expert adjudication, MRMC studies, standalone algorithm performance, and ground truth establishment for AI, are not applicable to the context of this document.

    However, I can extract the closest analogous information regarding "acceptance criteria" and "study proving the device meets the acceptance criteria" in the context of this physical medical device:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't present a specific "acceptance criteria" table in the way one might for an AI device (e.g., sensitivity, specificity thresholds). Instead, "acceptance criteria" are implied by the compliance with applicable performance standards and the comparative data to predicate devices. The "reported device performance" is demonstrated by stating that the subject device "met all quantitative and qualitative requirements" of these standards and by the head-to-head comparison data presented in the "Substantial Equivalence Comparison Table" and subsequent discussion.

    The most direct representation of "acceptance criteria" and "reported device performance" is found within the "Substantial Equivalence Comparison Table" itself, where critical performance metrics of the SafeT T-Piece Resuscitator are compared against those of the predicate device (Mercury Medical Neo-Tee) and a reference device (Fisher & Paykel NeoPuff).

    Here is a partial table, extracted and reformatted from the "Substantial Equivalence Comparison Table" that acts as the primary evidence of meeting "acceptance criteria" (i.e., being substantially equivalent to the predicate device):

    Characteristic / Performance MetricSafeT T-Piece Resuscitator (Reported Performance)Mercury Medical Neo-Tee (Predicate Device)Acceptance Criteria/Comparison Outcome (from document)
    Indications of UseEmergency respiratory support for neonates/infants < 10kg.Emergency respiratory support for neonates/infants < 10kg.Same (Acceptable for Substantial Equivalence)
    Patient PopulationInfants and neonates < 10 kg (22 lb)Infants and neonates < 10 kg (22 lb)Same (Acceptable for Substantial Equivalence)
    Cycling Pressure Range0 to 40 cm H2O0 to 40 cm H2OSame (Acceptable for Substantial Equivalence)
    Ventilatory FrequencyUp to 60 breaths/min (ISO 10651-5 Table 1)Up to 60 breaths/min (ISO 10651-5 Table 1)Same (Acceptable for Substantial Equivalence)
    Manometer RangeUp to 60 cm H2OUp to 60 cm H2OSame (Acceptable for Substantial Equivalence)
    Manometer AccuracyUp to 15 cm H2O - ± 3 cm H2O, Greater than 15 cm H2O - ± 5 cm H2OUp to 15 cm H2O - ± 3 cm H2O, Greater than 15 cm H2O - ± 5 cm H2OSame (Acceptable for Substantial Equivalence)
    Peak Inspiratory Pressure (PIP)0 to 40 H2O @ 15 LPM0 to 40 H2O @ 15 LPMSame (Acceptable for Substantial Equivalence)
    Maximum Pressure Relief (high pressure pop-off)40 cm H2O40 cm H2OSame (Acceptable for Substantial Equivalence)
    Visual or Audible Indication of High PressureAudibleAudibleSame (Acceptable for Substantial Equivalence)
    PEEP Flow Range (Example at 15 LPM)9 to 23 cm H2OUp to approx. 15 cm H2OSubject device range and predicate device's upper limit is different, however, substantially equivalent. Subject device range falls within those of the reference device and do not raise questions of safety and effectiveness. (Deemed Acceptable)
    Delivered Oxygen ConcentrationUp to 100% O2Up to 100% O2Same (Acceptable for Substantial Equivalence)
    Resuscitator Dead Space (no accessory attached)3.07 ml2.67 mlThe subject and predicate device have similar dead space. In ISO 10651-5 for manual resuscitators, allowable dead space is 30% of the minimum delivered tidal volume of 18 mL for a total of 5.4 mL. (Deemed Acceptable)
    Resuscitator Dead Space (with infant face mask - worst case)29.7 ml29.4 mlThe subject and predicate device have similar dead space with similar infant face mask and do not raise questions of safety and effectiveness. (Deemed Acceptable)
    Operating time (E cylinder)660 L @ 15 LPM ~ 47 min.660 L @ 15 LPM ~ 44 min.Similar; +3 min for subject device does not impact substantial equivalence as purpose is backup oxygen. (Deemed Acceptable)

    The study that proves the device meets the acceptance criteria is primarily the "Bench Performance Testing" described on page 12-13, assessed against the "Applicable Standards" listed on page 11-12.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not applicable. This is a bench performance test for a physical device, not an AI model. The "tests" refer to evaluations of the manufacturing design and performance characteristics, often involving one or a few units under controlled conditions (e.g., measuring pressure, flow, resistance). The phrase "head-to-head testing" is mentioned for resuscitator dead space but does not specify a "sample size" in the statistical sense for a data set.
    • Data Provenance: Not applicable. The "data" comes from the physical testing of the device itself by the manufacturer (Ventlab, LLC/SunMed) to demonstrate compliance with standards. There's no country of origin of patient data, nor is it retrospective or prospective in the context of clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. This is not an AI/ML device that requires expert-established ground truth from medical images or clinical observations. The "ground truth" for this device's performance is objective measurement against established engineering and safety standards (e.g., ISO 10651-5 for lung ventilators). The experts involved would be engineers and quality control personnel performing the bench testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No human interpretation or adjudication of outputs is described, as this is a physical device subject to objective performance measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is explicitly stated as not being an AI device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a mechanical resuscitator.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for this device's performance is established by engineering standards and specifications (e.g., ISO 10651-5, ISO 5356-1) which define acceptable ranges for physical parameters (pressure, flow, dead space, etc.). Compliance is determined by direct physical measurement during bench testing.

    8. The sample size for the training set

    • Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established

    • Not applicable. This is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K161420

    Validate with FDA (Live)

    Device Name
    MOVES SLC
    Date Cleared
    2017-06-06

    (379 days)

    Product Code
    Regulation Number
    868.5925
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MOVES® SLC™ is a portable computer controlled, electrically powered emergency ventilator intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.

    a. Suction
    The MOVES® SLC™ suction pump is intended for aspiration and removal of fluids, tissue (including bone), gases, bodily fluids or infectious materials from wounds or from a patient's airway or respiratory support system.

    b. Supplementary Oxygen
    The MOVES® SLC™ is intended to provide supplemental oxygen enriched air to patients that require supplemental oxygen.

    c. Patient Monitoring
    The MOVES® SLC™ is intended to monitor physiological parameters of patients and provide these parameters to a health care provider for interpretation in the form of physiological data and system alarms. Physiological data and system alarms will be available to the care provider from the monitor.

    Device Description

    The MOVES® SLC™ (SLC) is an upgraded version of the cleared MOVES® SLC™ device (K140049), a portable multifunction patient support and monitoring system with the following capabilities:

    • Computer controlled, electrically powered circle ventilator intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.
    • Delivery of oxygen-enriched air that may be supplied from an external oxygen source or generated internal to the system with the on-board oxygen concentrator.
    • Patient monitoring functions including the following patient parameters: Pulse Rate, Noninvasive BP (NIBP), Invasive BP (IBP), SPO2, Temperature, Respiration Rate, CO2, and O2.
    • Suction/aspirator pump for medical suction procedures where secretions, blood and other body fluids must be removed through the application of continuous negative pressure.

    The MOVES® SLCTM is capable of operating under battery power or external AC supply. It includes a handle and mounting equipment that allows it to attach to a stretcher.

    AI/ML Overview

    The provided text is a 510(k) summary for the MOVES® SLC™ medical device, which is an emergency ventilator with additional functions. This document describes the device, its indications for use, and its comparison to predicate devices to demonstrate substantial equivalence. However, it does not include detailed acceptance criteria and a study proving the device meets those criteria in the format requested.

    Specifically, the document:

    • Does not provide a clear table of acceptance criteria and reported device performance. It offers a comparison table of features and characteristics between the MOVES® SLC™ and predicate devices, including some performance specifications like frequency range, tidal volume, and SPO2 accuracy, but these are comparative, not acceptance criteria.
    • Does not mention anything about sample sizes used for test sets, data provenance, number of experts, adjudication methods, multi-reader multi-case (MRMC) studies, or standalone algorithm performance. This is because the device is a piece of hardware (ventilator, monitoring, suction, oxygen concentrator) and not an AI/software device that would typically involve such studies for regulatory clearance.
    • Does not discuss a training set or how ground truth for a training set was established. This is irrelevant for a hardware medical device of this type.
    • States that "Testing was conducted in accordance with all referenced standards and regulations, and to validate all system requirements" and a "Summary of Performance Testing" section mentions that "The results of performance testing demonstrate that the characteristics the MOVES® SLC™ are substantially equivalent to the identified predicates in terms of ventilator characteristics, patient monitoring performance, ability to delivery supplemental oxygen, and provide airway suction." However, it does not explicitly detail the specific performance test results against a defined set of acceptance criteria.

    Therefore, I cannot extract the requested information as it is not present in the provided document. The document focuses on demonstrating substantial equivalence to existing predicate devices based on features, characteristics, and compliance with general medical device standards, rather than proving performance against specific acceptance criteria through clinical studies involving human readers or AI algorithms.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162968

    Validate with FDA (Live)

    Device Name
    VORTRAN GO2VENT
    Date Cleared
    2017-04-10

    (168 days)

    Product Code
    Regulation Number
    868.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This Device is to be used by properly trained personnel to deliver emergency, short term, constant flow, pressure cycled ventilator support on patients weighing 10kg and above.

    Device Description

    The VORTRAN® GO2VENT™ provides short term, constant flow, pressure cycled ventilatory support in either pressure control or pressure support modes on patients weighing 10kg and above. In the pressure support mode, the rate dial of the VORTRAN® GO2VENT™ is set so that the baseline pressure is above the set PEEP allowing the patient to initiate inhalation by drawing the baseline pressure down to the set PEEP. The device includes the pulmonary modulator (an exhalation valve that opens at PIP and closes at PEEP) and a patient connector tee to supply gas flow, entrain additional air, and provides a redundant pop-off valve for patient care. The working mechanism of the VORTRAN® GO2VENTTM consists of a moving diaphragm which adds or subtracts spring force when it is moved from a horizontal to a vertical position, the addition or subtraction of spring force will affect the PIP setting by 1~3 cm-H2O. The VORTRAN® GO2VENT™ will function in any position as long as the final adjustments are made in a secured position (strapped or taped to the patient). The VORTRAN® GO2VENT™ is not an ICU stand alone ventilator with multiple monitoring features. Set up and use of the VORTRAN® GO2VENT™ is simple. Set desired flow and adjust pressure dial to obtain desired I-time and/or tidal volume (see tidal volume chart in instructions), and adjust rate dial to obtain desired rate and I to E ratio.

    AI/ML Overview

    The VORTRAN® GO2VENT™ is a powered emergency ventilator. The device was found substantially equivalent to its predicate device (VORTRAN® Automatic Resuscitator (VAR-Plus), K041473) based on non-clinical performance data and material biocompatibility.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The device's acceptance criteria are based on compliance with the following standards:

    • ASTM Designation: F 920 - 93 (Reapproved 1999): "Standard Specification for Minimum Performance and Safety Requirements for Resuscitators Intended for Use with Humans."
    • ISO 10651-5: "Lung ventilators for medical use - Particular requirements for basic safety and essential performance - Part 5: Gas-powered emergency resuscitators."

    The table below summarizes some key operational characteristics of the device and how they compare to the predicate, demonstrating that the new device meets the same specifications:

    CharacteristicAcceptance Criteria (Predicate)Reported Device Performance (New Device)
    Maximum Inspiratory Flow40 L/min40 L/min
    Ventilatory FrequencyAuto-adjusting to lung capacityAuto-adjusting to lung capacity
    Peak Pressure Range10 - 45 cm-H2O10 - 45 cm-H2O
    PEEP2 - 9 cm-H2O2 - 9 cm-H2O
    Required Source Pressure50 psig50 psig
    Dead Space4 ± 3 mL4 ± 3 mL
    Inspiratory Resistance3 ± 1 cm-H2O / L / sec3 ± 1 cm-H2O / L / sec
    Expiratory Resistance3 ± 1 cm-H2O / L / sec3 ± 1 cm-H2O / L / sec
    High Pressure Pop-offYes, 60 cm H2OYes, 60 cm H2O
    Visual or Audible Indication of High PressureYesYes
    FiO2 Delivery (50% setting)50% FiO2 by entraining room air50% FiO2 by entraining room air
    FiO2 Delivery (>85% setting)FiO2 of >85% (100% O2 supply)FiO2 of >85% (100% O2 supply)

    In addition, specific modifications were tested:

    • MR Conditional: The new beryllium-copper springs were tested according to ASTM F2052 and verified as MR Conditional for a static magnetic field of 3-Tesla or less, and a spatial gradient magnetic field of 10,000-gauss/cm or less.
    • New Entrainment Connectors: All testing data showed that the change in the knobs for FiO2 delivery did not degrade performance and the device delivers the required specifications.
    • New Beryllium-Copper Springs: Tested for performance and fatigue.

    2. Sample size used for the test set and the data provenance:

    The document concerns non-clinical performance testing only. There is no specific "test set" sample size in terms of clinical data or patient cases mentioned. The testing was conducted on the device itself and its components. The data provenance is from non-clinical laboratory testing performed by the manufacturer and a third-party (Shellock MR Testing for MR Conditional verification). The data is retrospective in the sense of comparing the new device against previously documented performance of the predicate device and established standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable, as this was a non-clinical device performance and material compatibility study, not an expert-driven ground truth establishment for a diagnostic or AI algorithm.

    4. Adjudication method for the test set:

    Not applicable, as this was non-clinical device performance testing against established standards and predicate device specifications, not a clinical trial requiring adjudication of patient outcomes or interpretations.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No. This is not an AI-enabled device; it is a powered emergency ventilator. Therefore, no MRMC study or AI assistance evaluation was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an algorithm, but a medical device.

    7. The type of ground truth used:

    The ground truth for the device's performance is compliance with established engineering and medical device standards (ASTM F 920-93 and ISO 10651-5) and the performance specifications of the predicate device. Material compatibility was assessed against materials used in previously cleared devices.

    8. The sample size for the training set:

    Not applicable, as this is hardware medical device. There is no "training set" in the context of machine learning.

    9. How the ground truth for the training set was established:

    Not applicable, no training set or ground truth in the context of machine learning.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141595

    Validate with FDA (Live)

    Date Cleared
    2015-03-16

    (276 days)

    Product Code
    Regulation Number
    868.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    o two e700, e600 and e500 are a time-cycled, volume-constant and pressure-controlled (only e700) emergency and transport ventilator designed for use in the pre-hospital, inter-hospital and transport settings. It is intended for use with adult, child, infant patients with a tidal volume from 50 ml (100 ml for e500) upwards who are in respiratory and/or cardiac arrest or respiratory distress and who require ventilatory support.

    Device Description

    The proposed o_two e700, o_two e600 and o_two e500 are time-cycled, volume-constant and pressure controlled (o_two e700 only) emergency and transport ventilators. They are electronically controlled, pneumatically powered ventilators which can be run using AC/DC power adapter or an internal rechargeable Lithium Ion battery pack. The devices use a 4.3" TFT screen to display live ventilation parameters (Tidal and minute volumes, proximal airway pressure and breathing rate) as well as ventilation pressure & volume wave forms, ventilation modes, settings, alarm limits and battery status. The wide range of both visual and audible alarms provides the healthcare professional with warnings of any changes in patient or device parameters.

    AI/ML Overview

    The provided text describes the O-Two Medical Technologies' e700, e600, and e500 Electronic Transport Ventilators and their substantial equivalence to predicate devices, but it does not contain the specific information requested in the prompt regarding acceptance criteria, performance tables, sample sizes, expert qualifications, or ground truth establishment.

    The document is a 510(k) premarket notification for device clearance, which focuses on demonstrating substantial equivalence to existing legally marketed devices (predicates) rather than providing detailed acceptance criteria and performance study outcomes in the format you've requested.

    Here's what can be extracted based on the document, and what is explicitly not available:


    1. A table of acceptance criteria and the reported device performance

    • Not explicitly provided in the requested format with specific numerical acceptance criteria and performance metrics for an AI/algorithm-based device.

    • The document primarily presents a comparative table showing the characteristics and specifications of the proposed devices against the predicate devices (Oxylog 3000 and CAREvent PAR) to demonstrate "Substantial Equivalence." Instead of acceptance criteria, it lists features and parameters and asserts "Yes" for substantial equivalence or notes specific differences.

      • Example from the document (partial):
    CharacteristicProposed o_two e700, o_two e600 and o_two e500Predicate K062267 Oxylog 3000 or K081330 CAREvent PARSubstantial Equivalence
    Ventilation modesCMV, ACV, SIMV, SIMV / PS, BiLVL, BiLVL /PS, CPAP, CPAP /PSCMV, CMVassist (ACV), SIMV, SIMV /PS, BIPAP(BiLVL), BIPAP (BiLVL)/PS, CPAP, CPAP /PSEquivalent to Oxylog 3000
    Tidal Volume (L)50 ml to 2.0 L50 ml to 2.0 LEquivalent to Oxylog 3000
    Battery Operating time18 hrs4 hrsBetter than Oxylog 3000
    Safety relief valveOpens at 80 cmH2OOpens at 80 cmH2OEquivalent to Oxylog 3000
    Alarms Audible/Visual & indicationsPmax, Pmin, MVHigh, MVLow, Low Battery (20% increments), BCI, Supply pressure Low or No, APNEAPmax, Pmin, MVHigh, MVLow, Low Battery (25% increments), Leakage, Supply pressure, APNEAEquivalent to Oxylog 3000
    • The "Summary of Performance Testing" section lists qualitative findings from a comparative bench test:
      • "Both ventilators delivered equivalent volume-time, pressure-time and flow-time wave forms under the same ventilation mode;"
      • "Both ventilators were responsive to spontaneous breathing trigger;"
      • "Ventilation Vt, frequency and I:E ratio or Ti of both units were close to target or preset parameters;"
      • "Both units delivered similar pressure supports."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not available. The document mentions "comparative side-by-side bench testing" but does not specify sample sizes (e.g., number of test cases, simulated patient scenarios, or real patient data) or data provenance. It's bench testing, implying simulated rather than patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable / Not available. This is a ventilator device, and the testing described is bench testing against established performance metrics and comparison to predicate devices, not an AI/imaging diagnostic device requiring human expert ground truth for interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable / Not available. Adjudication methods are typically used for establishing ground truth in diagnostic studies, which is not the nature of this submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, this was not done. This is not relevant for an electronic transport ventilator. The study focused on proving substantial equivalence through bench testing and comparison of specifications to predicate devices, not on human reader performance with or without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. The device is a ventilator, a physical system with embedded software, not a standalone algorithm for diagnostic or interpretative tasks. The "Summary of Performance Testing" refers to bench testing of the device's operational performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For the bench testing, the "ground truth" would be the expected physical performance parameters and waveforms based on established engineering and medical device standards (e.g., IEC 60601 series, EN 794-3, ISO 10651-3, ISO 80601-2-12) and the known performance of the predicate devices. The device's output (volume-time, pressure-time, flow-time waveforms, Vt, frequency, I:E ratio, etc.) was compared against these established targets.

    8. The sample size for the training set

    • Not applicable / Not available. This is not an AI/machine learning device that uses a "training set" in the conventional sense. The device's software logic is developed based on engineering principles and validated through testing.

    9. How the ground truth for the training set was established

    • Not applicable / Not available. As this is not an AI/machine learning device requiring a training set, the concept of establishing ground truth for a training set does not apply.

    In summary: The provided FDA document is a 510(k) clearance letter and summary for a medical device (electronic transport ventilator). It demonstrates "substantial equivalence" to predicate devices through comparative bench testing and specifications. It does not contain the specific information about AI/algorithm performance, detailed acceptance criteria tables, sample sizes for AI model development/testing, or expert-based ground truth evaluations as typically found in submissions for AI-enabled diagnostic or interpretive devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142096

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-10-17

    (77 days)

    Product Code
    Regulation Number
    868.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The T-Piece Resuscitator is a gas powered emergency resuscitator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway. It is intended for use with patients weighing greater than 10kg (>22lb).

    Device Description

    The Mercury T-Piece Resuscitator is manually operated, gas powered resuscitator for use with patients greater than 10 kg (>22 lb).

    It is a simple T-piece, with a manometer and the ability to adjust Peak Inspiratory Pressure (PIP) and Positive End-Expiratory Pressure (PEEP). It incorporates a pressure relief valve for excessive pressure.

    The T-Piece Resuscitator can be connected to the patient via a face mask, Supraglottic airway or endotracheal tube.

    AI/ML Overview

    The provided document describes the Mercury T-Piece Resuscitator and its substantial equivalence to predicate devices, focusing on non-clinical testing for performance. This is a 510(k) premarket notification summary, which means the device is being compared to an already legally marketed device (predicate device) and does not involve AI or complex analytical studies in the way modern AI/ML medical devices would.

    Therefore, many of the requested categories in the prompt are not applicable to this document. I will focus on the information that can be extracted directly from the provided text.

    Here's the breakdown based on the provided document:

    Acceptance Criteria and Device Performance

    The acceptance criteria are generally aligned with the performance requirements outlined in ISO 10651-5 - Particular Requirements for Basic Safety and Essential Performance for Gas Powered Resuscitators. The "Reported Device Performance" column reflects the specifications of the proposed Mercury Medical T-Piece Resuscitator as presented in comparison tables with its predicate devices.

    Acceptance Criteria (Derived from Predicate & ISO 10651-5)Reported Device Performance (Mercury Medical T-Piece Resuscitator)
    Indications for Use: Provide emergency respiratory support by face mask or tube; patients > 10kg (>22lbs).The T-Piece Resuscitator is a gas-powered emergency resuscitator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway. It is intended for use with patients weighing greater than 10kg (>22lb).
    Environment of Use: Hospital, sub-acute facilities, and pre-hospital (EMS).Hospital, sub-acute facilities, and pre-hospital (EMS).
    Patient Population: Patients greater than 10 Kg (>22 lbs.).Patients greater than 10 Kg (>22 lbs.)
    Ventilation Frequency: Manually delivered by the user up to 60 BPM.Manually delivered by the user up to 60 BPM.
    Maximum Pressure Relief: Factory set at 40 cm H2O (predicate range 5-80 cm H2O).60 cm H2O (Factory set at 40 cm H2O).
    Delivered Pressure: Up to 60 cm H2O (predicate range 5-80 cm H2O).Up to 60 cm H2O.
    Delivered Volume: (Predicate range 190 - 675 ml with flow rates between 4 - 36 lpm).60 - 700 ml with flow rates between 5 - 35 lpm.
    Inspiratory Resistance: (Predicate: Less than - 5 cm H2O).2 cm H2O at minimum PEEP setting @ 60 lpm.
    Expiratory Resistance: (Predicate: Less than + 5 cm H2O).2.4 cm H2O at minimum PEEP setting @ 60 lpm.
    Oxygen Concentration with optional blender: (Predicate range: 21-100%).21 - 98% based upon blender setting.
    Manometer Range: (Predicate range: -20 to 80 cm H2O).Up to 60 cmH2O (Cleared under K954486).
    Manometer Accuracy: (Predicate: +/- 2 cm H2O).+/- 3 cm H2O up to 15 cm H2O; +/- 5 cm H2O > 15 cm H2O.
    Peak Inspiratory Pressure (PIP): (Predicate range: 5-80 cm H2O).0-60 cm H2O.
    Positive End-Expiratory Pressure (PEEP): (Predicate uses CPAP instead; reference NeoTee has PEEP).0 to 60 cm H2O.
    Operational Gas Flow Rate: (Predicate range: 0-60 lpm).0 - 35 lpm.
    Operational time with 400 L cylinder: (Predicate @ 36 lpm - 10 minutes).@ 35 lpm - 11 minutes.
    Dead space of circuit: (Predicate: < 15 ml).< 7.5 ml.
    Operating Temperature Range: (Predicate range: 10 to 40°C).-18 to 60°C.
    Storage Temperature: (Predicate range: -20 to 60°C).-40 to 50°C.
    Immersion Resistance: Temporary submersion of circuit does not affect functionality.Temporary submersion of circuit does not affect functionality.
    Vomitus Resistance: Meets ISO 10651-5.Tested per ISO 10651-5.
    Water Immersion: Meets ISO 10651-5.Tested per ISO 10651-5.
    Oxygen Concentration: Meets ISO 10651-5.Tested per ISO 10651-5.
    Pressure Limitation: Meets ISO 10651-5.Tested per ISO 10651-5.
    Storage/Operating Conditions: Meets ISO 10651-5.Tested per ISO 10651-5.
    Drop: Meets ISO 10651-5.Tested per ISO 10651-5.

    The study that proves the device meets the acceptance criteria is a series of "Bench testing" performed according to ISO 10651-5 for gas powered resuscitators. The document explicitly states: "We have demonstrated that the T-Piece Resuscitator meets the performance and design specifications requirements as outlined in requirements and specifications as outlined in ISO 10651-5 - Particular Requirements for Basic Safety and Essential Performance for Gas Powered Resuscitators."


    Additional Information on the Study:

    1. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

      • The document describes "bench testing" which implies tests on physical devices rather than patient data. Therefore, details like "country of origin of the data" or "retrospective/prospective" are not applicable. The sample size for the bench tests is not specified in the provided summary.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • This is not applicable as the study involves bench testing of a physical medical device, not a diagnostic or AI algorithm relying on expert interpretation of data. The ground truth is established by the performance standards outlined in ISO 10651-5.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This is not applicable to a physical device performance test assessed against engineering standards.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • This is not applicable. This is a physical medical device (resuscitator), not an AI/ML diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • This is not applicable. This is a physical medical device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for the device's performance is based on engineering and safety standards, specifically ISO 10651-5 - Particular Requirements for Basic Safety and Essential Performance for Gas Powered Resuscitators.
    7. The sample size for the training set:

      • This is not applicable. The device is a physical resuscitator, not an AI/ML algorithm that requires a "training set."
    8. How the ground truth for the training set was established:

      • This is not applicable, as there is no training set for a physical device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K140049

    Validate with FDA (Live)

    Device Name
    MOVES SLC
    Date Cleared
    2014-05-29

    (140 days)

    Product Code
    Regulation Number
    868.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MOVES® SLC™ is a portable computer controlled electrically powered emergency transport ventilator intended to provide continuous or intermittent ventilatory support for the care of adults who require mechanical ventilation.

    MOVES® SLC™ is intended to deliver high inspired oxygen concentrations to spontaneously breathing patients who require elevated inspired oxygen.

    MOVES® SLC™ is intended to be used in a transport or emergency setting on adult patients who weigh between 40kg and 120kg.

    MOVES® SLC™ provides the following supplemental functions for patients that it is ventilating or supplying with supplemental oxygen:

    • Suction a.
      The MOVES® SLC™ suction pump is intended for aspiration and removal of fluids, tissue (including bone), gases, bodily fluids or infectious materials from wounds or from a patient's airway or respiratory support system.

    • Supplementary Oxygen b.
      The MOVES® SLCTM is intended to provide supplemental oxygen enriched air to patients that require supplemental oxygen.

    • c. Patient Monitoring
      The MOVES® SLC™ is intended to monitor physiological parameters of patients and provide these parameters to a health care provider for interpretation in the form of physiological data and system alarms. Physiological data and system alarms will be available to the care provider from the monitor.

    Device Description

    The MOVES® SLC™ (SLC) is an upgraded version of the cleared MOVES® device (K093261), a portable multifunction patient support and monitoring system with the following capabilities:

    • . Computer controlled, electrically powered circle ventilator intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.
    • Delivery of oxygen-enriched air that may be supplied from an external oxygen . source or generated internal to the system with the on-board oxygen concentrator.
    • Patient monitoring functions including the following patient parameters: Pulse . Rate, Noninvasive BP (NIBP), Invasive BP (IBP), SPO2, Temperature, Respiration Rate, CO2, and O2.
    • Suction/aspirator pump for medical suction procedures where secretions, blood . and other body fluids must be removed through the application of continuous negative pressure.
      The MOVES® SLCTM is capable of operating under battery power or external AC supply. It includes a handle and mounting equipment that allows it to attach to a stretcher.
    AI/ML Overview

    The provided text is a 510(k) summary for the Thornhill Research Inc. MOVES® SLC™ medical device. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than conducting extensive clinical studies to establish new safety and effectiveness.

    Therefore, the document does not contain the kind of detailed information about acceptance criteria and study designs (like sample sizes for test and training sets, expert qualifications, adjudication methods, or MRMC studies) that would be present in a submission establishing de novo safety and effectiveness for a novel device or AI algorithm.

    The document mainly describes the device, its intended use, and states that performance testing was conducted to demonstrate substantial equivalence to predicate devices and compliance with relevant standards.

    Here's a breakdown of the information available in the document regarding acceptance criteria and performance, as much as can be extracted:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a specific table of quantitative acceptance criteria with corresponding performance metrics. Instead, it relies on demonstrating substantial equivalence to existing predicate devices.

    The "Summary of Performance Testing" section states:

    "The results of performance testing demonstrate that the characteristics SLC are substantially equivalent to the identified predicates in terms of ventilator characteristics, patient monitoring performance, ability to delivery supplemental oxygen, and provide airway suction."

    And the "Determination of Substantial Equivalence" section reiterates:

    "Where differences in performance or technology exist, it has been demonstrated that they do not adversely impact safety or effectiveness. In addition, SLC has been tested to comply with relevant recognized consensus safety and performance standards as well as voluntary standards (detailed above)."

    The acceptance criteria, implicitly, are that the device's performance characteristics for ventilation, patient monitoring, oxygen delivery, and suction must be substantially equivalent to those of its predicate devices and comply with all referenced standards and regulations.


    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided in the 510(k) summary. The testing mentioned is "performance testing" and likely refers to engineering, bench, and potentially animal testing, rather than human clinical trials with specific patient sample sizes. The provenance of any data (country, retrospective/prospective) is also not mentioned.


    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This information is not applicable/provided. The document describes a medical device (ventilator, monitor, concentrator, suction pump), not an AI algorithm requiring expert-established ground truth on a test set (e.g., for image classification or diagnosis).


    4. Adjudication Method for the Test Set

    This information is not applicable/provided for the reasons stated above.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a MRMC comparative effectiveness study was not conducted or reported in this 510(k) summary. This type of study is typically performed for AI-powered diagnostic aids where the impact on human reader performance is being evaluated.


    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The MOVES® SLC™ is a hardware medical device with integrated software/control. It is not an "algorithm only" device in the sense of a standalone AI diagnostic tool. Therefore, the concept of "standalone performance" for an algorithm without human-in-the-loop is not applicable in this context. The performance testing refers to the overall device's functionality.


    7. The Type of Ground Truth Used

    Given that this is a hardware device primarily demonstrating substantial equivalence and compliance with standards, the "ground truth" would be established through a combination of:

    • Engineering specifications and measurements: Comparing output parameters (e.g., ventilator pressures, flow rates, oxygen concentration, monitoring accuracy) against known physical standards or the specifications of predicate devices.
    • Performance against predicate device specifications: Ensuring that the MOVES® SLC™ meets or exceeds the demonstrated performance of its predicate devices for the various functions.
    • Compliance with recognized consensus standards: Meeting the benchmarks set by standards like IEC 60601 series, EN-794-3, ISO 8359, and ASTM E1112-00.

    There is no mention of expert consensus, pathology, or outcomes data being used as "ground truth" in the context of this 510(k) summary.


    8. The Sample Size for the Training Set

    This information is not applicable/provided. The MOVES® SLC™ is not described as a device that uses machine learning or an AI algorithm that requires a "training set" of data. Its control system and functionalities are based on established engineering principles and algorithms, not data-driven machine learning models that are "trained."


    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable/provided as there is no mention of a training set for an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 6