K Number
K162968

Validate with FDA (Live)

Device Name
VORTRAN GO2VENT
Date Cleared
2017-04-10

(168 days)

Product Code
Regulation Number
868.5925
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This Device is to be used by properly trained personnel to deliver emergency, short term, constant flow, pressure cycled ventilator support on patients weighing 10kg and above.

Device Description

The VORTRAN® GO2VENT™ provides short term, constant flow, pressure cycled ventilatory support in either pressure control or pressure support modes on patients weighing 10kg and above. In the pressure support mode, the rate dial of the VORTRAN® GO2VENT™ is set so that the baseline pressure is above the set PEEP allowing the patient to initiate inhalation by drawing the baseline pressure down to the set PEEP. The device includes the pulmonary modulator (an exhalation valve that opens at PIP and closes at PEEP) and a patient connector tee to supply gas flow, entrain additional air, and provides a redundant pop-off valve for patient care. The working mechanism of the VORTRAN® GO2VENTTM consists of a moving diaphragm which adds or subtracts spring force when it is moved from a horizontal to a vertical position, the addition or subtraction of spring force will affect the PIP setting by 1~3 cm-H2O. The VORTRAN® GO2VENT™ will function in any position as long as the final adjustments are made in a secured position (strapped or taped to the patient). The VORTRAN® GO2VENT™ is not an ICU stand alone ventilator with multiple monitoring features. Set up and use of the VORTRAN® GO2VENT™ is simple. Set desired flow and adjust pressure dial to obtain desired I-time and/or tidal volume (see tidal volume chart in instructions), and adjust rate dial to obtain desired rate and I to E ratio.

AI/ML Overview

The VORTRAN® GO2VENT™ is a powered emergency ventilator. The device was found substantially equivalent to its predicate device (VORTRAN® Automatic Resuscitator (VAR-Plus), K041473) based on non-clinical performance data and material biocompatibility.

1. Table of Acceptance Criteria and Reported Device Performance:

The device's acceptance criteria are based on compliance with the following standards:

  • ASTM Designation: F 920 - 93 (Reapproved 1999): "Standard Specification for Minimum Performance and Safety Requirements for Resuscitators Intended for Use with Humans."
  • ISO 10651-5: "Lung ventilators for medical use - Particular requirements for basic safety and essential performance - Part 5: Gas-powered emergency resuscitators."

The table below summarizes some key operational characteristics of the device and how they compare to the predicate, demonstrating that the new device meets the same specifications:

CharacteristicAcceptance Criteria (Predicate)Reported Device Performance (New Device)
Maximum Inspiratory Flow40 L/min40 L/min
Ventilatory FrequencyAuto-adjusting to lung capacityAuto-adjusting to lung capacity
Peak Pressure Range10 - 45 cm-H2O10 - 45 cm-H2O
PEEP2 - 9 cm-H2O2 - 9 cm-H2O
Required Source Pressure50 psig50 psig
Dead Space4 ± 3 mL4 ± 3 mL
Inspiratory Resistance3 ± 1 cm-H2O / L / sec3 ± 1 cm-H2O / L / sec
Expiratory Resistance3 ± 1 cm-H2O / L / sec3 ± 1 cm-H2O / L / sec
High Pressure Pop-offYes, 60 cm H2OYes, 60 cm H2O
Visual or Audible Indication of High PressureYesYes
FiO2 Delivery (50% setting)50% FiO2 by entraining room air50% FiO2 by entraining room air
FiO2 Delivery (>85% setting)FiO2 of >85% (100% O2 supply)FiO2 of >85% (100% O2 supply)

In addition, specific modifications were tested:

  • MR Conditional: The new beryllium-copper springs were tested according to ASTM F2052 and verified as MR Conditional for a static magnetic field of 3-Tesla or less, and a spatial gradient magnetic field of 10,000-gauss/cm or less.
  • New Entrainment Connectors: All testing data showed that the change in the knobs for FiO2 delivery did not degrade performance and the device delivers the required specifications.
  • New Beryllium-Copper Springs: Tested for performance and fatigue.

2. Sample size used for the test set and the data provenance:

The document concerns non-clinical performance testing only. There is no specific "test set" sample size in terms of clinical data or patient cases mentioned. The testing was conducted on the device itself and its components. The data provenance is from non-clinical laboratory testing performed by the manufacturer and a third-party (Shellock MR Testing for MR Conditional verification). The data is retrospective in the sense of comparing the new device against previously documented performance of the predicate device and established standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

Not applicable, as this was a non-clinical device performance and material compatibility study, not an expert-driven ground truth establishment for a diagnostic or AI algorithm.

4. Adjudication method for the test set:

Not applicable, as this was non-clinical device performance testing against established standards and predicate device specifications, not a clinical trial requiring adjudication of patient outcomes or interpretations.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No. This is not an AI-enabled device; it is a powered emergency ventilator. Therefore, no MRMC study or AI assistance evaluation was performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This is not an algorithm, but a medical device.

7. The type of ground truth used:

The ground truth for the device's performance is compliance with established engineering and medical device standards (ASTM F 920-93 and ISO 10651-5) and the performance specifications of the predicate device. Material compatibility was assessed against materials used in previously cleared devices.

8. The sample size for the training set:

Not applicable, as this is hardware medical device. There is no "training set" in the context of machine learning.

9. How the ground truth for the training set was established:

Not applicable, no training set or ground truth in the context of machine learning.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are connected and appear to be part of a single, flowing design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 10, 2017

Vortran Medical Technology 1, Inc. James Lee Executive Vice President & COO 21 Goldenland Court, #100 Sacramento, California 95834

Re: K162968

Trade/Device Name: VORTRAN® GO2VENT™ Regulation Number: 21 CFR 868.5925 Regulation Name: Powered Emergency Ventilator Regulatory Class: Class II Product Code: BTL Dated: March 2, 2017 Received: March 10, 2017

Dear James Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Tina
Kiang-S

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162968

Device Name

VORTRAN® GO2VENT™

Indications for Use (Describe)

This Device is to be used by properly trained personnel to deliver emergency, short term, constant flow, pressure cycled ventilator support on patients weighing 10kg and above.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary / K162968 - VORTRAN® GO2VENT™

I. SUBMITTER

VORTRAN® Medical Technology 1, Inc. 21 Goldenland Court, Suite 100 Sacramento, CA 95834 USA

Phone: (800) 434-4034 Fax: (916) 648-9751

Contact Person: James Lee Date Prepared: March 31, 2017

II. DEVICE

Name of Device: VORTRAN® GO2VENT™ Common or Usual Name: Ventilator, Emergency, Powered (Resuscitator) Classification Name: Powered Emergency Ventilator Regulation Number: 21 CFR 868.5925 Regulatory Class: II Product Code: BTL Classification Advisory Committee: Anesthesiology Review Advisory Committee: Anesthesiology

III. PREDICATE DEVICE

K041473VORTRAN® Automatic Resuscitator (VAR-Plus)(Primary Predicate Device)
K153733VORTRAN® Manometer(Reference Device – For Material Compatibility Only)

IV. DEVICE DESCRIPTION

The VORTRAN® GO2VENT™ provides short term, constant flow, pressure cycled ventilatory support in either pressure control or pressure support modes on patients weighing 10kg and above. In the pressure support mode, the rate dial of the VORTRAN® GO2VENT™ is set so that the baseline pressure is above the set PEEP allowing the patient to initiate inhalation by drawing the baseline pressure down to the set PEEP. The device includes the pulmonary modulator (an exhalation valve that opens at PIP and closes at PEEP) and a patient connector tee to supply gas flow, entrain additional air, and provides a redundant pop-off valve for patient care. The working mechanism of the VORTRAN® GO2VENTTM consists of a moving diaphragm which adds or subtracts spring force when it is moved from a horizontal to a vertical position, the addition or subtraction of spring force will affect the PIP setting by 1~3 cm-H2O. The VORTRAN® GO2VENT™ will function in any position as long as the final adjustments are made in a secured position (strapped or taped to the patient).

{4}------------------------------------------------

510(k) Summary / K162968 - VORTRAN® GO2VENTTM

The VORTRAN® GO2VENT™ is not an ICU stand alone ventilator with multiple monitoring features. Set up and use of the VORTRAN® GO2VENT™ is simple. Set desired flow and adjust pressure dial to obtain desired I-time and/or tidal volume (see tidal volume chart in instructions), and adjust rate dial to obtain desired rate and I to E ratio.

Device Model Number:6123-10
Device Accessories:Elbow Flex Hose, Oxygen Tubing, Pressure Manometer
Interaction with Patient:The Elbow Flex Hose has indirect contact with patient

INDICATIONS FOR USE V.

This device is intended to be use by properly trained personnel to deliver emergency, short term, constant flow, pressure cycled, ventilatory support on patients weighing 10kg and above.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

A. Technical Modifications:

MR conditional - The new model of the VORTRAN® Automatic Resuscitator has been modified with new springs made of Beryllium-Copper instead of the Stainless-Steel. This change was verified (in the Shellock MR Testing on Vortran Products Report) to make the device MR Conditional. The report concluded that the VORTRAN® GO2VENT™ can be used in an MRI environment with a static magnetic field of 3-Tesla or less, and a spatial gradient magnetic field of 10,000-gauss/cm or less. Shellock performs their tests based on ASTM F2052 - Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment. We have verified that the performance of the beryllium-copper springs is substantially equivalent to the stainless-steel springs.

New Entrainment Connectors for the 50% and 100% FiO2 Delivery - The new device connectors include a combination knob that can be rotated from the 100% FiO2 (a fully closed position of entrainment ports) to the 50% FiO2 (a position of having entrainment ports open) versus the two connectors of the predicate device that the 100% FiO2 connector should be completely removed to access the 50% FiO2 connector. The modification facilitates the FiO2 change from 50% to 100% FiO2 and vice versa. All testing data have shown and verified that the change in the knobs has not degraded the performance and the device delivers the required specifications.

{5}------------------------------------------------

510(k) Summary / K162968 – VORTRAN® GO2VENT™

The tables below show the technological differences and similarities of our VORTRAN® GO2VENT™ compared to the predicate device (VAR-Plus).

Operational Characteristics:

NEW DEVICEPREDICATE DEVICE
Device NameVORTRAN® GO2VENT™VORTRAN® Automatic Resuscitator (VAR-Plus)
510(k) NumberN/AK041473
Maximum Inspiratory Flow40 L/minSame as predicate device40 L/min
Ventilatory FrequencyAuto-adjusting to lung capacity [c]Same as predicate deviceAuto-adjusting to lung capacity [c]
Peak Pressure Range10 - 45 cm-H2O [d]Same as predicate device10-45 cm-H2O [d]
PEEP2 - 9 cm-H2O [d]Same as predicate device2 - 9 cm-H2O [d]
Required Source Pressure50 psigSame as predicate device50 psig
Dead Space4 ± 3 mLSame as predicate device4 ± 3 mL
Inspiratory Resistance3 ± 1 cm-H2O / L / secSame as predicate device3 ± 1 cm-H2O / L / sec
Expiratory Resistance3 ± 1 cm-H2O / L / secSame as predicate device3 ± 1 cm-H2O / L / sec
High Pressure Pop-offYes, 60 cm H2OSame as predicate deviceYes, 60 cm H2O
Visual or Audible Indicationof High PressureYesSame as predicate deviceYesSame as predicate device

{6}------------------------------------------------

510(k) Summary / K162968 – VORTRAN® GO2VENT™

Functional Characteristics:

NEW DEVICEPREDICATE DEVICE
Device NameVORTRAN® GO2VENT™VORTRAN® Automatic Resuscitator (VAR-Plus)
510(k) NumberN/AK041473
FiO2 Devliery50% FiO2 by entraining room airFiO2 of >85% when supplied with 100% O250% FiO2 by entraining room airFiO2 of >85% when supplied with 100% O2
Supply 100% O2[A]Image: BSupply 100% O2[B]Supply 100% O2[C]Supply 100% O2[D]
Method ofChangingDelivery ofFiO2 whensupplied with100% O2[A] Delivers FiO2 at50% by entrainingroom air whensetting the selectordial to the 50%mark on the newpatient tee[B] Delivers FiO2 of>85% by NOTentraining room airwhen setting theselector dial to the100% mark on thenew patient tee[C] Delivers FiO2 at50% by entrainingroom air when the100% adaptor isremoved, exposingthe 50%entrainment nozzleon the patient tee[D] Delivers FiO2 of>85% by NOTentraining room airwhen the 100%adaptor isconnected to thepatient tee
OxygenConcentrationDelivers FiO2 >85% or 50% by entraining room airSame as predicate deviceDelivers FiO2 >85% or 50% by entraining room air[e]

Image /page/6/Figure/5 description: This image shows a comparison between a new device and a predicate device. The new device is labeled as "VORTRAN® GO2VENT™" with a 510(k) number of N/A. The predicate device is labeled as "VORTRAN® Automatic Resuscitator (VAR-Plus)" with a 510(k) number of K041473. The image shows the devices in two different scenarios: 50% FiO2 by entraining room air and FiO2 of >85% when supplied with 100% O2.

{7}------------------------------------------------

510(k) Summary / K162968 – VORTRAN® GO₂VENT™

Material of Construction:

NEW DEVICEPREDICATE DEVICE
Device NameVORTRAN® GO2VENT™VORTRAN® Automatic Resuscitator(VAR-Plus)VORTRAN® Manometer
510(k) NumberN/AK041473K153733
Date SE DecisionNAJuly 15, 2004September 16, 2016
HousingK-Resin®, PolycarbSame as predicate deviceK-Resin®Polycarb
Internal SpringsBeryllium-Copper [f]Same as predicate device302 Stainless SteelBeryllium-Copper [f]
Pressure and RateDialsHDPESame as predicate deviceDOW® HDPE 12450 color blueAcetal
One-Way ValveBodyHDPESame as predicate deviceDOW® HDPE 12450N/A
Flapper ValveSiliconeSame as predicate deviceSilicon PolymerN/A
DiaphragmSiliconeSame as predicate deviceSiliconeSILICONE/Natural
Disk CenterNylonSame as predicate deviceNylonN/A

VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility Testing ●

Device Nature of Body Contact Category:Duration of Contact for the Device:
External CommunicatingUp to 30 days
PartNo.Part DescriptionPatient Contact StatusSubstantial EquivalenceDevice (510K #)Material/Color
1.6007100% - 50% VAR NozzleIndirect patient contactK153733Polycarbonate/White
2.6008FiO2 Controller KnobIndirect patient contactK041473HDPE/Green
3.6009Entrainment BarrelIndirect patient contactK153733Polycarbonate/White
4.2015One-Way Valve BodyIndirect patient contactK041473HDPE/Natural
5.2016One-Way Valve FlapperIndirect patient contactK153733Silicone/Red
6.6005Pop-Off Valve CapNo patient contactPolycarbonate/Blue
7.6010Pop-Off Valve SpringNo patient contactBeryllium Copper
8.2012Pop-Off Valve PistonIndirect patient contactK041473HDPE/Red
9.6006Patient TeeIndirect patient contactK041473K-Resin®/Clear/Blue
10.2291Manometer AssemblyIndirect patient contactK153733N/A
11.6003Modulator Bottom, SinglePortIndirect patient contactK153733Polycarbonate/Clear/Blue
14.2182BDiaphragmNo patient contactSilicone/Natural
2181BHard CenterIndirect patient contactK041473Nylon/Natural

{8}------------------------------------------------

510(k) Summary / K162968 - VORTRAN® GO2VENT™

Biocompatibility Testing (Continued) ●

Device Nature of Body Contact Category:External CommunicatingDuration of Contact for the Device:Up to 30 days
Part No.Part DescriptionPatient Contact StatusSubstantial Equivalence Device (510K #)Material/Color
15.6004Pressure Dial SpringNo patient contactBeryllium Copper
17.6002Modulator TopNo patient contactPolycarbonate/Clear/Blue
18.6017Rate DialNo patient contactHDPE/Blue
19.6016Pressure DialNo patient contactHDPE/Blue
Discussion:1) Rationale for Not Needing Biocompatibility Testing: Considering the materialused in the components of the new device (VORTRAN® GO2VENT™) and the factthat the same material used as those in the predicated devices manufactured byVORTRAN Medical and produced with the same manufacturing process, thematerial of the new device is compatible and requires no additional biocompatibility

Animal Study ●

None

  • Non-Clinical Performance Testing/Performance Data/Compliance with Performance Standards
    The VORTRAN® GO2VENT™, just as the VAR-Plus, meets the "Standard Specification for Minimum Performance and Safety Requirements for Resuscitators Intended for Use with Humans" ASTM Designation: F 920 -93. A separate non-clinical test showed that all requirements were also met for ISO 10651-5, "Lung ventilators for medical use - Particular requirements for basic safety and essential performance - Part 5: Gas-powered emergency resuscitators". The determination for substantial equivalence was based on a comparison of performance data between the new device and predicate device (FDA 510(k) No. K041473). The predicate device performance tests were documented. To demonstrate substantial equivalence, the new device was also tested using "ASTM F920-93 (Reapproved 1999)". In addition, the features that are considered technological differences were tested. The function of the new FiO2 entrainment feature (at 50% and 100% FiO2 settings) and device shelf-life were tested with the results documented. The new Beryllium-Copper springs were tested for performance and fatigue.

{9}------------------------------------------------

510(k) Summary / K162968 - VORTRAN® GO2VENT™

  • . Clinical Studies None

VII. CONCLUSIONS

  • The VORTRAN® GO2VENT™ is substantially equivalent to a predicate ● device: The VAR-Plus. The results from the nonclinical tests show that the device achieves predefined acceptance criteria for all assessments that were previously performed on the predicate device. VORTRAN® GO2VENT™, just as the VAR-Plus, meets the "Standard Specification for Minimum Performance and Safety Requirements for Resuscitators Intended for Use with Humans" ASTM Designation: F 920 - 93. All requirements were also met for ISO 10651-5, "Lung ventilators for medical use - Particular requirements for basic safety and essential performance - Part 5: Gaspowered emergency resuscitators". The VORTRAN® GO2VENT™ has been shown to be substantially equivalent to the predicate device.

§ 868.5925 Powered emergency ventilator.

(a)
Identification. A powered emergency ventilator is a demand valve or inhalator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.(b)
Classification. Class II (performance standards).