Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K231895
    Manufacturer
    Date Cleared
    2023-12-04

    (159 days)

    Product Code
    Regulation Number
    868.5330
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K883038, K925982

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Maxtec MaxBlend 2+p is intended to provide a continuous air/oxygen gas mixture and to continuously monitor the concentration of oxygen and pressure being delivered to infant, pediatric, and adult patients. It is a restricted medical device intended for use by qualified, trained personnel, under the direction of a physician, in professional healthcare settings, i.e., hospital, subacute, and nursing-care facilities where the delivery and monitoring of air/oxygen mixtures is required. This is not intended as a life-supporting device or life sustaining device.

    Device Description

    The Maxtec MaxBlend 2+p is an oxygen delivery device which incorporates an air/oxygen blender, battery powered oxygen and pressure monitor, and an adjustable flowmeter, all in a single assembly. The integral air/oxygen blender provides precise mixing of medical grade air and oxygen. The flowmeter provides control of the flow rate delivered. The oxygen monitor measures the oxygen concentration from the blender's gas flow, displays these measured concentrations, and provides user selectable high and low oxygen alarms. It also allows the user to monitor pressure simultaneously using adjustable high and low alarm limits.

    AI/ML Overview

    The document describes the Maxtec MaxBlend 2+p, a medical device combining an air/oxygen blender, oxygen monitor, pressure monitor, and flowmeter. The 510(k) submission seeks substantial equivalence to existing predicate devices (Maxtec MaxBlend 2 - K161718 and MaxO2ME+p - K221734), particularly highlighting the addition of a pressure monitoring feature.

    Here's an analysis of the acceptance criteria and the study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for several performance aspects are implicitly derived from the comparative tables (Table 1 and 2) where the subject device's specifications are listed and compared to the predicate devices. The "Reported Device Performance" for the subject device is simply its stated characteristics, which are claimed to meet or be identical to the predicate device specifications, thereby meeting the acceptance criteria based on substantial equivalence.

    Feature / Performance AspectAcceptance Criteria (from predicate devices)Reported Device Performance (Maxtec MaxBlend 2+p)
    Air/Oxygen Mixer Features
    Gas Supply TypeAir / OxygenAir / Oxygen
    Pressure30 to 75 psi30 to 75 psi
    Mixed gas stability± 1% oxygen± 1% oxygen
    Flow range of BlendersLow flow model – 0-30 Lpm; High flow model – 0-100 LpmLow flow model – 0-30 Lpm; High flow model – 0-70 Lpm (Primary Predicate) / 0-100 Lpm (Secondary Predicate)
    Pressure supply differential alarmAir / oxygen pressure
    Ask a Question

    Ask a specific question about this device

    K Number
    K182956
    Date Cleared
    2019-01-18

    (86 days)

    Product Code
    Regulation Number
    868.5925
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K883038, K925982, K093913

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Puffin Lite Infant Resuscitator is intended to provide the basic equipment required for pulmonary resuscitation of neonatal infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the neonatal infant.

    Device Description

    The resuscitation system provides the basic equipment required for pulmonary resuscitation of neonatal infants. The Puffin Lite Infant Resuscitator is a gas powered emergency resuscitation system. It is intended to be used inside the hospital by trained medical professionals to provide precise FIO2 delivery, and manual ventilation as established by resuscitation guidelines to neonatal infants weighing less than 10 kg (22 lb). The resuscitation system includes two medical gas flowmeters, an integrated oxygen blender, airway pressure manometer, peak inspiratory pressure (PIP) control, positive end expiratory pressure (PEEP) control, a gas supply pressure alarm and T-piece resuscitator.

    AI/ML Overview

    The Puffin Lite Infant Resuscitation System underwent bench testing to demonstrate conformance to performance specifications.

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance CriteriaReported Device Performance
    Valve Function after VomitusThe proper function of the circuit shall be verified within 20 seconds of becoming disabled by vomitus. Function is verified by verifying flow valve accuracy.Passed
    Inspiratory ResistancePressure generated at the patient connection port during expiration should not exceed -5 cmH2O with inspiratory airflow set to 6 L/min.Passed
    Expiratory ResistancePressure generated at the patient connection port during expiration should not exceed 5 cmH2O with expiratory airflow set to 6 L/min.Passed
    Dead SpaceThe deadspace volume of the T-Piece circuit should be less than 7 mL.Passed
    FIO2 accuracyThe proper function of the FIO2 adjustment knob shall be verified by comparing the FIO2 setting value with the output oxygen concentration. Values shall be within 5%.Passed
    Primary and Secondary Flow Valve - Peak FlowThe proper function of the primary and secondary flow valves shall be verified by comparing the flow setting with the actual measured output flow.Passed
    Airway Manometer AccuracyThe proper function of the airway manometer shall be verified by comparing the pressure readings with the actual measured output pressure.Passed
    VOC TestingThe device should not add volatile organic compounds (VOCs) to the output gas delivered to the patient.Passed
    Particulate AnalysisThe output of particulate matter sizes 2.5 microns or less are no more than 12 micrograms/cubic meter of air at one atmospheric pressure.Passed

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the specific sample sizes used for each individual bench test. The testing was described as "bench testing," which typically involves laboratory-controlled environments and simulated conditions rather than real patient data. The data provenance is from this bench testing, not from human subjects or real-world clinical data. Therefore, there is no information on country of origin or whether the data was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable. The study involved bench testing of a medical device, where acceptance criteria were based on engineering and performance specifications (e.g., pressure, flow, FIO2 accuracy) rather than expert interpretation of medical images or data. Ground truth was established by measuring the physical performance parameters of the device against predefined technical requirements.

    4. Adjudication method for the test set:

    Not applicable. As the study was bench testing against predefined technical specifications, there was no need for an adjudication method by human experts for the "ground truth." The "passing" or "failing" of a test was determined by whether the device's measured performance fell within the specified limits.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a 510(k) submission for a physical medical device (infant resuscitator) and not an AI or imaging diagnostic device. Therefore, an MRMC study comparing human reader performance with and without AI assistance was not conducted.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable. This device is not an algorithm or an AI system. It is a physical medical device (infant resuscitator) that is intended for use by trained medical professionals.

    7. The type of ground truth used:

    The ground truth used for this study was the engineering and performance specifications outlined in the acceptance criteria. The device's performance was measured against these objective, predefined technical requirements. There was no reliance on expert consensus, pathology, or outcomes data in the context of this bench testing.

    8. The sample size for the training set:

    Not applicable. This device is not an AI or machine learning system that requires a "training set" of data.

    9. How the ground truth for the training set was established:

    Not applicable, as there was no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161718
    Manufacturer
    Date Cleared
    2016-10-27

    (127 days)

    Product Code
    Regulation Number
    868.5330
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K153659, K883038, K925982

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MaxBlend 2 and MaxBlend Lite are designed to provide a continuous air/oxygen gas mixture and to continuously monitor the concentration of oxygen being delivered to infant, pediatric, and adult patients. It is a restricted medical device intended for use by qualified, trained personnel, under the direction of a physician, in professional healthcare settings, i.e.. hospital, sub-acute, and nursing-care facilities where the delivery and monitoring of air/oxygen mixtures is required. This is not intended as a life supporting device.

    Device Description

    The MaxBlend family of blenders, MaxBlend 2 and MaxBlend Lite, are oxygen delivery devices which incorporate an air/oxygen blender, battery powered oxygen monitor, and an adjustable flowmeter, all in a single assembly. The air/oxygen blender provides precise mixing of medical grade air and oxygen. The flowmeter provides control of the flow rate delivered. The oxygen monitor measures the oxygen concentration from the blender's gas flow, displays these measured concentrations, and provides user selectable high and low oxygen alarms.

    The MaxBlend 2 incorporates the air/oxygen blender within its enclosure. The MaxBlend Lite may be provided with or without the blender component pre-assembled, allowing the user to install the oxygen monitor/flowmeter module, the MaxBlend Lite component, on an existing compatible blender. The addition of the MaxBlend Lite module is intended to improve the safety of an existing air/oxygen blender that is in the user's possession. Both devices use the exact same oxygen monitor sensor and electronics, MaxO2ME (K153659). Both devices use the exact same components/assemblies in the flowmeter and sensor bleed manifold which form the gas pathway to the patient. The only substantive difference is the form of the enclosure for the monitor electronics.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from Maxtec, LLC, regarding their MaxBlend 2 and MaxBlend Lite devices. It aims to demonstrate substantial equivalence to a predicate device, the Bird Sentry Blender (K973646), and references other cleared devices (K153659 – Maxtec MaxO2ME, K883038 - CareFusion / Bird - Blender, K925982 - BioMed Devices - Blender).

    Here's an analysis based on your request, highlighting that this document outlines non-clinical testing for a medical device that mixes and monitors breathing gases, not a typically AI-powered diagnostic imaging device. Therefore, many of the requested categories (like MRMC studies, roles of experts for ground truth, sample sizes for training/test sets for AI models, etc.) are not applicable or are addressed differently in the context of hardware device performance testing.

    Acceptance Criteria and Device Performance (Non-Clinical/Hardware)

    The acceptance criteria are primarily based on meeting the performance specifications of the predicate device and established standards for medical devices of this type. The "reported device performance" is generally stated as "passed" or "meets the acceptance criteria" for various tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    AttributeAcceptance Criteria (Predicate / Standards Equivalent)Reported Device Performance (MaxBlend 2 and MaxBlend Lite)
    Indications for UseContinuous air/oxygen gas mixture, monitor O2 concentration for infant, pediatric, adult patients in institutional settings; not life supporting.Continuous air/oxygen gas mixture, continuously monitor O2 concentration for infant, pediatric, adult patients in professional healthcare settings (hospital, sub-acute, nursing-care facilities); not life supporting.
    Environments of UseInstitutional (healthcare settings), Not for MRI.Professional healthcare settings (hospital, sub-acute, nursing-care facilities), Not for MRI.
    Patient PopulationInfant, pediatric, and adult patients.Infant, pediatric, and adult patients.
    Weight2 kg2.4 kg
    Power source of oxygen monitor2 x AA Alkaline batteries4 x AA Alkaline batteries
    Gas Supply TypeAir / Oxygen, 30 to 75 psiAir / Oxygen, 30 to 75 psi
    % Oxygen Control21 - 100%, Accuracy ± 3%21 - 100%, Accuracy ± 3%
    Mixed Gas Stability± 1% oxygen± 1% oxygen
    Flow Range of BlendersLow flow: 0-30 Lpm; High flow: 0-100 Lpm.Low flow: 0-30 Lpm; High flow: 0-100 Lpm.
    Pressure Supply Differential AlarmAir / oxygen pressure
    Ask a Question

    Ask a specific question about this device

    K Number
    K140707
    Date Cleared
    2014-05-28

    (69 days)

    Product Code
    Regulation Number
    868.5925
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K883038, K925982, K093913

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LifeBorne Infant Resuscitator is intended to provide the basic equipment required for pulmonary resuscitation of neonatal infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the neonatal infant.

    Device Description

    The resuscitation system provides the basic equipment required for pulmonary resuscitation of neonatal infants. The LifeBorne Infant Resuscitator is a gas powered emergency resuscitation system. It is intended to be used inside the hospital by trained medical professionals to provide precise FIO2 delivery, manual ventilation, and emergency airway clearance as established by resuscitation guidelines to neonates and infants weighing less than 10 kg (22 lb).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the LifeBorne Infant Resuscitator, based on the provided text:

    Acceptance Criteria and Device Performance

    TESTTEST REQUIREMENTS (Acceptance Criteria)REPORTED DEVICE PERFORMANCE
    Valve Function after VomitusThe proper function of the circuit shall be verified within 20 seconds of becoming disabled by vomitus. Function is verified by verifying flow valve accuracy.Passed.
    Inspiratory ResistancePressure generated at the patient connection port during expiration should not exceed -5 cmH2O with inspiratory airflow set to 6 L/min.Passed.
    Expiratory ResistancePressure generated at the patient connection port during expiration should not exceed 5 cmH2O with expiratory airflow set to 6 L/min.Passed.
    Dead SpaceThe deadspace volume of the T-Piece circuit should be less than 7 mL.Passed.
    FIO2 accuracyThe proper function of the FIO2 adjustment knob shall be verified by comparing the FIO2 setting value with the output oxygen concentration. Values shall be within 5%.Passed.
    Primary and Secondary Flow Valve - Peak FlowThe proper function of the primary and secondary flow valves shall be verified by comparing the flow setting with the actual measured output flow.Passed.
    Airway Manometer AccuracyThe proper function of the airway manometer shall be verified by comparing the pressure readings with the actual measured output pressure.Passed.
    VOC TestingThe device should not add volatile organic compounds (VOCs) to the output gas delivered to the patient.Passed.
    Particulate AnalysisThe output of particulate matter sizes 2.5 microns or less are no more than 12 micrograms/cubic meter of air at one atmospheric pressure.Passed.

    Study Details

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated as a separate "test set" sample size. The document refers to "bench testing" as the method for performance evaluation. The tests themselves are functional evaluations of the device's components and overall system.
    • Data Provenance: The testing was "bench testing," meaning it was conducted in a laboratory or simulated environment. The country of origin of the data is not specified, but the submission is from International Biomedical, located in Austin, TX, U.S.A. It is retrospective in the sense that the testing was completed before the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This device's performance was evaluated through bench testing against predefined technical specifications, not by expert interpretation of data. Therefore, no experts were used to establish ground truth in this context.

    4. Adjudication method for the test set:

    • Not applicable. As described above, the evaluation was based on meeting objective technical requirements through bench testing, not on human interpretation requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a medical device (resuscitator) that does not involve "human readers" interpreting medical cases or utilizing AI. Its effectiveness is determined by its mechanical and functional performance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, in a way. The device itself operates "standalone" in performing its functions (e.g., delivering precise FIO2, controlling PIP/PEEP). The "bench testing" represents the performance of the algorithm/device without human intervention during the test itself (though humans operate the test setup). There is no "algorithm" in the sense of AI or image analysis involved.

    7. The type of ground truth used:

    • The ground truth for the device's performance was based on predefined technical specifications and engineering standards. Each test had a specific, measurable requirement (e.g., FIO2 within 5%, dead space less than 7 mL) which served as the objective ground truth against which the device's performance was measured.

    8. The sample size for the training set:

    • Not applicable. This device is a mechanical/electronic system, not one that uses machine learning or AI that would require a "training set" of data.

    9. How the ground truth for the training set was established:

    • Not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1