510(k) Data Aggregation

    K Number
    K241485

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-10-09

    (138 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TruAbutment DS is a patient-specific CAD/CAM abutment, which is directly connected to endosseous dental implants and is intended to be used as an aid in prosthetic rehabilitation. It is compatible with the following systems: Astra OsseoSpeed EV (K130999, K120414), Biomet 3i Full OSSEOTITE Tapered Certain (K130949), DIO UF (II) Internal Submerged (K161987, K170608, K173975), Neoss ProActive® (K083561), Osstem TS (K161604), Camlog Screw-Line (K083496), Conelog Screw-Line (K113779), Implant Direct Legacy2 (K192221), BioHorizons Internal Implant System (K093321, K143022, K071638), MegaGen AnyRidge Internal Implant (K140091). All digitally designed abutments and/or copings for use with the TruAbutments are intended to be sent to a TruAbutment-validated milling center for manufacture.

    TruBase is a titanium component that is directly connected to endosseous dental implants to provide support for patient-specific prosthetic restorations, such as copings or crowns. It is indicated for a screw-retained single tooth or cement-retained single tooth and bridge restorations. It is compatible with the following systems: Astra OsseoSpeed EV (K130999), Biomet 3i Full OSSEOTITE Tapered Certain (K130949), DIO UF(II) Internal Submerged (K161987, K170608, K173975), Neoss ProActive® (K083561), Camlog Screw-Line (K083496), Conelog Screw-Line (K113779), Implant Direct Legacy2 (K192221). All digitally designed abutments and/or copings for use with the TruAbutment are intended to be sent to a TruAbutment-validated milling center for manufacture.

    Device Description

    TruAbutment DS, TruBase and abutment screw are made of Titanium grade Ti-6A1-4V ELI (meets ASTM Standard F136). TruAbutment DS, TruBase are supplied with two identical screws which are used: (1) For fixing the abutment into the endosseous implant. (2) For dental laboratory use during construction of related restoration. TruAbutment DS, TruBase are provided non-sterile. Therefore, it must be sterilized before use. TruAbutment DS, TruBase are devices that can only be sold, distributed, or used upon the order of an authorized healthcare provider, generally referred to as prescription (Rx) devices.

    TruAbutment DS system includes patient-specific abutments that are placed into the dental implant to provide support for the prosthetic restoration. The subject abutments are indicated for serew-retained restorations. The design and manufacturing of the patient-specific abutments take into consideration the shape of the final prosthesis based on the patient's intra-oral indications using CAD/CAM system during the manufacturing. All manufacturing processes of TruAbutment DS are conducted at the TruAbutment milling center.

    TruBase is a two-piece abutment. The base component is premanufactured and is used to support a cemented CAD/CAM zirconia superstructure. The base and the zirconia superstructure together form the final abutment. CAD/CAM customized superstructure that composes the final abutment is intended to be sent to a TruAbutment-validated milling center to be designed and milled, according to the prosthetic planning and patient clinical situation. The superstructure is cemented to the TruBase in the lab. Use "RelyX Unicem 2Automix" as an adhesive extra orally to connect.

    AI/ML Overview

    The provided text is a 510(k) summary for the TruAbutment DS and TruBase devices. It primarily focuses on demonstrating substantial equivalence to a predicate device (TruAbutment DS, K203649) and does not detail an acceptance criteria table with reported device performance in the manner of a clinical study. The text describes non-clinical testing performed, but not a study designed to prove the device meets acceptance criteria related to a specific clinical outcome or diagnostic accuracy.

    Therefore, many of the requested items (acceptance criteria table, sample size for test/training sets, data provenance, expert ground truth, adjudication, MRMC studies, standalone performance, type of ground truth) are not applicable based on the content of this 510(k) summary, which is a premarket notification for a medical device primarily based on demonstrating substantial equivalence through engineering and mechanical testing, not clinical performance or AI algorithm validation studies.

    However, I can extract the information provided regarding non-clinical testing for the devices.


    Acceptance Criteria and Study for TruAbutment DS & TruBase

    Based on the provided 510(k) summary, the "acceptance criteria" and "study" described are focused on non-clinical mechanical performance testing and demonstration of substantial equivalence to a predicate device, rather than a clinical study proving performance against specific clinical or diagnostic accuracy metrics with human or AI components.

    Here's the relevant information extracted and presented based on the document:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table with specific quantitative acceptance criteria alongside actual reported numerical performance results for the new devices in the context of a comparative study proving their performance against such criteria. Instead, it states that "The results of the above tests have met the criteria of the standard and demonstrated substantial equivalence with the reference devices." This implies a qualitative "met standard" outcome rather than specific numerical performance data.

    The tables provided describe the design limits of the devices and compare them to the predicate device, not performance data from a test:

    TruAbutment DS Design Parameters (Acceptance Criteria are implied by meeting these limits)

    Design ParameterSubject Device (TruAbutment DS) Design LimitPrimary Predicate Device (K203649) Design LimitReported Device Performance (Implied)
    Minimum and Maximum abutment angle (°)0 ~ 250 ~ 25Met specified range
    Minimum and Maximum cuff height (mm)0.5 ~ 6.00.5 ~ 6.0Met specified range
    Minimum and Maximum diameter at abutment/implant interface (Ø, mm)3.3 ~ 8.03.3 ~ 8.0Met specified range
    Minimum and Maximum length of the abutment (mm)6 ~ 116 ~ 11Met specified range
    Minimum wall thickness at abutment/implant interface (mm)0.40.4 ~ 0.9Met specified range
    Minimum and Maximum length of abutment post (length above the abutment collar / gingival height) (mm)4.0 ~ 7.04.0 ~ 7.0Met specified range

    TruBase Design Parameters (Acceptance Criteria are implied by meeting these limits)

    Design ParameterSubject Device (TruBase) Design LimitPrimary Predicate Device (K203649) Design LimitReported Device Performance (Implied)
    Minimum and Maximum angulation (°)0 ~ 150 ~ 15Met specified range
    Minimum and Maximum gingival (cuff) height (mm)0.5 ~ 5.00.5 ~ 5.0Met specified range
    Minimum and Maximum diameter at abutment/implant interface (Ø, mm)5.0 ~ 8.05.0 ~ 8.0Met specified range
    Minimum thickness (mm)0.40.4Met specified value
    Minimum and Maximum length of abutment post (length above the abutment collar / gingival height) (mm)4.0 ~ 6.04.0 ~ 6.0Met specified range

    For mechanical performance, the document states:
    "Mechanical performance testing was performed according to ISO 14801. For compatible OEM implant line, worst-case constructs were subjected to static compression and compression fatigue testing. The fatigue limit data for all other implant lines demonstrated the construct strengths to be sufficient for their intended use."

    This confirms that the acceptance criteria for mechanical performance were "sufficient for their intended use" as defined by ISO 14801 and worst-case testing, but quantitative results are not provided.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: The document mentions "worst-case constructs" were tested for mechanical performance, and "the entire system including all variations (all compatible implant bodies, dental abutments, and fixation screws)" was evaluated for MRI environment conditions. However, specific numerical sample sizes for these tests are not provided.
    • Data Provenance: The data comes from non-clinical laboratory testing following international standards (ISO 14801, ISO 17665-1/2, ISO 10993 series). The country of origin and retrospective/prospective nature are not applicable as it's not a clinical data study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This is not applicable. The "ground truth" for this type of device (dental abutments) is established through adherence to engineering design specifications, material standards (ASTM F136), and performance under mechanical stress tests (ISO 14801), as well as compliance with sterilization and biocompatibility standards. It does not involve expert interpretation of images or clinical outcomes in the same way an AI diagnostic device would.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This is not applicable as there is no human interpretation or subjective assessment of data requiring adjudication. Testing is based on objective measurements against engineering standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This is not applicable. The device is an endosseous dental implant abutment, not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This is not applicable. The device is a physical dental component, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for validating these devices is adherence to engineering specifications, material properties, and performance standards (e.g., passing specific load-bearing and fatigue tests per ISO 14801, meeting biocompatibility requirements, maintaining dimensional accuracy). "Dimensional analysis and reverse engineering" were used to confirm compatibility.

    8. The sample size for the training set

    • This is not applicable. There is no "training set" as this is a physical medical device, not an AI/machine learning algorithm.

    9. How the ground truth for the training set was established

    • This is not applicable for the reason above.

    Summary of the Study Proving Device Meets Criteria (Based on Provided Text):

    The "study" conducted was primarily non-clinical laboratory testing to demonstrate the mechanical performance, sterilization efficacy, and biocompatibility of the TruAbutment DS and TruBase devices. The objective was to show substantial equivalence to an existing legally marketed device (TruAbutment DS, K203649) by proving that the new devices meet established design limits and performance standards relevant to dental implant abutments.

    • Mechanical Testing: Performed on "worst-case constructs" according to ISO 14801 for static compression and compression fatigue. The outcome was that "construct strengths [were] sufficient for their intended use."
    • Sterilization Testing: Performed per ISO 17665-1:2006, 17665-2:2009 and ANSI/AAMI ST79:2010.
    • Biocompatibility Testing: Performed per ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010.
    • MRI Environment Evaluation: A non-clinical worst-case MRI review was done using scientific rationale and published literature to assess magnetically induced displacement force and torque.
    • Dimensional Analysis and Reverse Engineering: Conducted on the implant-to-abutment connection platform to assess critical design aspects and tolerances, confirming compatibility.

    The overall conclusion was that the devices "met the criteria of the standard and demonstrated substantial equivalence with the reference devices," thus indicating they met their implied acceptance criteria for safety and performance as medical devices. Clinical testing was explicitly stated as "not necessary."

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1