510(k) Data Aggregation

    K Number
    K222282

    Validate with FDA (Live)

    Date Cleared
    2022-10-27

    (90 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 2.4mm Secure Locking Variable Angle Distal Radius Plate and Ø2.4 mm Secure Locking Screws, Self-Tapping are intended for fixation of complex intra- and extra-articular fractures arid osteotomies of the distal radius.

    The 2.7mm/3.5mm Secure Locking Distal Humerus Medial Plate, 2.7mm/3.5mm Secure Locking Distal Humerus, Dorsolateral Plate, and 2.7mm/3.5mm Secure Locking Distal Humerus Plate, Dorsolateral With Lateral Support are indicated for intra-articular fractures of the distal humerus, comminuted supracondylar fractures, and nonunions of the distal humerus.

    The 3.5mm Secure Locking Small Plate, 3.5mm Secure Locking T-Plate, Small with 3 Head Holes, and Ø3.5 mm Secure Locking Screws, Self-Tapping, are indicated for fixation of fractures, and non-unions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia, fibula.

    The 3.5 mm Secure Locking Superior-Anterior Clavicle Plate and 3.5 mm Secure Locking Superior-Anterior Clavice Plate With Lateral Extension is indicated for fixation of fractures, malunions, and ostectornies of the clavice in adults, and in both adolescents (12-18 years) and transitional adolescents (18-21 years), in which the clavicular growth plates have fused or in which the growth plates will not be crossed by the plate system.

    The 3.5mm Secure Locking Olecranon Plate is indicated for fractures, osteotomies, malunions and non-unions of the olecranon.

    The 3.5 mm Secure Locking Philos Proximal Humeral Internal Plate is indicated for fractures and fracture dislocations, osteotomies, and non-unions of the proximal humerus.

    The 4.5/5.0mm Secure Locking Narrow LC Dynamic Compression Plate is indicated for fixation of various long bones, such as the humerus, femur and tibia and for use in fixation of peri-prosthetic fractures, and fixation of nonunions or malunions in adult patients.

    The 4.5mm/5.0mm Secure Locking Distal Femoral Plate is intended for buttressing multifragmentary distal femur fractures including: supracondylar, intra-articular condylar, periprosthetic fractures, nonunions and malunions, and osteotomies of the femur.

    The 4.5mm/5.0mm Secure Locking Medial Proximal Tibia Plate is intended to buttress metaphyseal fractures of the medial tibia plateau, split-type fractures of the medial split fractures with associated depressions and split or depression fractures of the medial tibia plateau. Also, for use in the fixation of nonunions of the medial proximal tibia and tibia shaft.

    The 4.5mm/5.0mm Secure Locking L Buttress Plate and 4.5mm/5.0mm Secure Locking T Buttress Plate are intended to buttress metaphyseal fractures of the proximal humerus, medial tibial tibia. Also, for use in fixation of non-unions and malunions.

    The 3.5mm/4.5mm/5.0mm Secure Locking Distal Tibia Plates are intended treatment of non-unions, and fractures of the distal tibia, including simple, comminuted, lateral wedge, depression, medial wedge, bicondylar, combinations of lateral wedge and depression, and fractures with associated shaft fractures.

    The Ø3.5mm CORTICAL SCREW and Ø4.5mm CORTICAL SCREW are intended for fixation of fractures, osteotomies and non-unions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, tibia, calcaneous, femur and fibula in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation.

    The Ø2.7 mm Secure Locking Screws, Self Tapping, and Ø 2.7mm Cortical Screws are intended for fractures and osteotomies of small bone fragments, including the foot, ankle, and hand in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by screw fixation.

    The Ø5.0mm Secure Locking Screws, Self-Tapping are intended for fixation of various long bones, such as the humerus, femur and tibia. It is also for use in fixation of non-unions or malunions.

    The 6.5mm Cancellous Screw, 16mm Thread, 32 Thread, Ø4.0mm CANCELLOUS SCREW, Partial Thread, Full Thread, and Ø4.0mm SMALL CANCELLOUS CANNULATED SCREW, Partial Thread, Full Thread are indicated for use in hindfoot and midfoot fusions, subtalar fusions, calcaneal osteotomies, midfoot reconstruction, and ankle arthrodeses.

    The 6.5mm Cancellous Cannulated Screw, 16mm Thread, 32 Thread, Full Thread, and 7.0mm Cancellous Cannulated Screw, 16mm Thread, 32 Thread, are indicated for fracture fixation of large bone fragments, such as tibial plateau fractures, ankle arthrodeses, intercondylar femur fractures; and subtalar arthrodeses.

    Device Description

    HEMC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System consists of various shapes and sizes of plates featuring compression and locking holes, full-threaded-cortical, locking selftapping screws, compression and dynamic screws. The subject device system also consists of a variety of general use instruments (Class I), which include drill bits, forceps, plate benders, and drill guides.

    The plates and screws are manufactured from Stainless Steel and Titanium alloy.

    The system contains several models based on the size of the device and application site such as fixation/reconstruction of small fragment bones, forefoot, mid-foot, rear-foot, ankle, or other bones appropriate for the size of the device. The plate implants are in many models available, such as:, Reconstruction Plates, T-Plates, Anatomical Plates, Clavicle Plates.

    These all are mainly divided into:

    • . Large Fragment Plates
    • Small Fragment Plates ●
    • Mini Fragment Plates .

    The locking screw implants are offered in corresponding diameter ranges from 2.4mm, 2.7mm, 3.5mm, 5.0mm diameters with lengths varying from a minimum length of 6 mm to maximum length of 90mm.

    The non-locking screw implants are offered in 2.7mm, 3.5mm and 4.5mm diameters, with lengths ranqing from 10mm to 80mm.

    The cancellous screw implants are 4.0mm and 6.5mm in diameter, with lengths ranging from 10 to 120 mm. The cancellous cannulated screw implants are offered in 4.0mm. 6.5mm and 7.0mm diameters and lengths ranging from 16mm to 130 mm.

    HEMC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System are provided non-sterile, the products must be sterilized prior to use. All implants are for single use only.

    AI/ML Overview

    This document describes the 510(k) summary for the HEMC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System. However, it does not contain information about acceptance criteria or a study proving the device meets those criteria in the context of an AI/ML device.

    The information provided pertains to the substantial equivalence of a medical implant (bone plates and screws) to predicate devices, focusing on material, design, and mechanical performance rather than AI/ML algorithm performance.

    Therefore, I cannot extract the requested information regarding acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth for an AI/ML device from the given text.

    The document discusses the following:

    • Device Name: HEMC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System
    • Intended Use: Fixation of various bone fractures and osteotomies.
    • Predicate Devices: A list of Synthes Locking Compression Plate Systems and Screws.
    • Non-Clinical Testing:
      • Material Standards: ASTM F136, ASTM F138, ASTM F139 (for Stainless Steel and Titanium alloy).
      • Performance Standards: ASTM F382 (Metallic Bone Plates), ASTM F384 (Metallic Angled Orthopedic Fracture Fixation Devices), ASTM F543 (Metallic Medical Bone Screws), and FDA guidance documents for orthopedic screws and fracture fixation plates.
      • Performance Results: Conforms to Static Four Point Bend Test, Dynamic Four Point Bend Test (for plates), Torsional Properties, Driving Torque, and Pull-out Test (for screws).
    • Clinical Evaluation: Not necessary to demonstrate substantial equivalence, as the device is similar in design, pattern, and intended use to predicate devices.

    In summary, the provided text does not describe an AI/ML device or its performance evaluation. It details the regulatory clearance process for a traditional medical implant based on substantial equivalence.