510(k) Data Aggregation

    K Number
    K212676

    Validate with FDA (Live)

    Date Cleared
    2022-03-03

    (191 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Osteon Precision Milled Suprastructure is indicated for attachments in the treatment of partially orfully edentulous for the purpose of restoring chewing function. The Osteon Precision Milled Suprastructures are intended for attachment to a minimum of two (2) abutments.

    The Osteon Milled Suprastructure is indicated for compatibility with the following abutment systems:

    • · Nobel Biocare Multi Unit Abutment Plus, 4.8mm, max 30°
    • · Nobel Biocare Xeal Abutments, 4.8mm, max 30°
    • · Nobel Biocare Multi Unit Abutment, 4.8mm, max 30°
    • MIS Multi-unit Abutments, 4.8mm
      • · C1 Conical Connection Implant System, max 30°
      • · V3 Conical Connection Implant System, max 30°
      • · Internal Hex Implant System, max 30°
      • · Conical Connection, max 30°
    • Southern Compact Conical Abutments, 4.8mm
      • · MAX Implant System, 0°
      • · Provata Implant System, max 30°
      • · Deep Conical (DC) Implants, 0°
      • · Piccolo Implants, 0°
      • · External Hex Implants, max 30°
    • · Astra Tech Implant System® Multi Base Abutment EV, 4.8mm, max 30°
    • Keystone Multi Unit Abutment, 4.8mm, 0°
    • · Neodent GM Mini Conical Abutment, 4.8mm, max 30°
    • · Implant Direct GPS® Angled Abutment, 5.0mm, max 30°
    • · Dentium SuperLine® Abutments, 4.5-5.5mm, max 30°
    • · Zimmer Angled Tapered Abutments, 4.5mm, max 30°
    • · Paltop Multi Unit Abutment, 5.0mm, max 17°
    Device Description

    The Osteon Precision Milled Suprastructures (also referred as superstructures) are metallic dental restorative device that is intended for attaching by screw retention to dental abutments to aid in the treatment of partial and totally edentulous patients for the purpose of restoring their chewing function. These suprastructures attach to previously-cleared original equipment manufacturers (OEM) dental abutments using the (OEM) prosthetic screws. The abutmentborne subject devices are indicated for placement only on OEM implant/abutment constructs placed according to the labeling of the previously-cleared systems, and not to exceed the maximum angulation allowed for each OEM implant/abutment construct as identified in the Indications for Use Statement of the subject system.

    The Osteon Precision Milled Suprastructure is designed for an individual patient from scans of the patient's impression. The suprastructure is manufactured in biocompatible Titanium alloy with the aid of Computer Aided Design (CAD) and Computer Aided Manufacturing (CAM) technology. All CAD/CAM fabrication is performed by Osteon Medical, within our premises.

    Osteon Precision Milled Suprastructures facilitate the attachment of both removable and fixed dental prosthesis and hence categorized as type A and type B.

    AI/ML Overview

    This document is a 510(k) summary for a dental device, the Osteon Precision Milled Suprastructure. It outlines the regulatory process and demonstrates substantial equivalence to predicate devices, rather than presenting a study that proves the device meets specific performance acceptance criteria. Therefore, most of the requested information about acceptance criteria, study details, sample sizes, expert involvement, and ground truth establishment is not available in this document.

    Here's why and what information can be extracted:

    • This is a 510(k) submission: The primary goal of a 510(k) is to demonstrate that a new device is "substantially equivalent" to a legally marketed predicate device, not necessarily to prove its performance against a set of novel acceptance criteria through a clinical trial.
    • Focus on Substantial Equivalence: The document compares features like indications for use, materials, design, and manufacturing processes to existing devices.
    • Non-Clinical Testing Mentioned: The document refers to non-clinical tests performed, but these are primarily for demonstrating compatibility, material safety, and sterility, not performance against specific clinical efficacy or accuracy metrics.

    Information that can be extracted or inferred:

    1. A table of acceptance criteria and the reported device performance:

    This document does not present explicit "acceptance criteria" in the traditional sense of performance metrics for an AI/algorithm-driven device. Instead, it demonstrates similarity to predicate devices based on design specifications and material properties. The "performance" is implicitly deemed acceptable if it's shown to be substantially equivalent to the established predicate devices.

    Feature / Acceptance Criteria (Implied)Reported Device Performance (Osteon Precision Milled Suprastructure)
    Regulation DescriptionEndosseous dental implant abutment (Similar to Predicate)
    Indications for UseFor attachment to dental abutments in the treatment of partially or fully edentulous jaws for restoring chewing function. Intended for attachment to a minimum of two (2) abutments and compatible with various listed OEM abutment systems. (Similar to Predicate)
    Device MaterialTitanium alloy Ti-6Al-4V (Same as Reference Device K173466)
    Design/TechnologySingle milling blocks (Similar to Predicate)
    Fixation MethodCAD/CAM milling from single milling blanks (Similar to Predicate)
    Design/ConstructionAbutment-borne (Similar to Predicate)
    SterilitySupplied Nonsterile (Similar to Predicate)
    Target PopulationAdult patients (Similar to Predicate)
    Prescription/OTCPrescription only (Similar to Predicate)
    Recommended Cleaning/MaintenanceProper oral hygiene (Similar to Predicate)
    Design Specifications (Type A - Removable Prosthesis)
    Total CylindersMin 2, Max 10 (Similar to Reference Device K173466)
    Suprastructure Span Between CylindersMin 1mm, Max 30mm (Similar to Reference Device K173466)
    Suprastructure HeightMin 3mm, Max 12mm (Allows wider than Reference Device K173466)
    Suprastructure WidthMin 3.4mm, Max 12mm (Allows wider than Reference Device K173466)
    Distal Cantilever SectionMin 0mm, Max 15mm (Similar to Reference Device K173466)
    Cylinder HeightMin 0mm, Max 4.6mm (Lower max than Reference Device K173466)
    Cylinder DiameterMin 4.5mm, Max 8mm (Higher min than Reference Device K173466)
    Design Specifications (Type B - Fixed Prosthesis)
    Total CylindersMin 2, Max 10 (Similar to Reference Device K173466)
    Suprastructure Span Between CylindersMin 1mm, Max 30mm (Similar to Reference Device K173466)
    Suprastructure HeightMin 3mm, Max 22mm (Allows wider than Reference Device K173466)
    Suprastructure WidthMin 3.4mm, Max 12mm (Allows wider than Reference Device K173466)
    Distal Cantilever SectionMin 0mm, Max 15mm (Similar to Reference Device K173466)

    (Note: "Similar" indicates that the values or characteristics fall within or are comparable to those of the predicate/reference devices, with stated differences not impacting safety/efficacy from the manufacturer's perspective.)

    2. Sample sizes used for the test set and the data provenance:

    • Test Set Sample Size: Not applicable. This submission relies on "non-clinical tests" (dimensional analysis, biocompatibility, sterilization validation) and comparison to existing predicate marketing clearances. It does not describe a "test set" in the context of an AI/ML algorithm evaluation with human interpretation or image analysis. The "test" consists of demonstrating manufacturing conformance and material properties for the physical device.
    • Data Provenance: The device is manufactured by Implant Solutions PTY LTD (trading as Osteon Medical) in Mulgrave, Victoria, 3170 Australia. The non-clinical tests (e.g., biocompatibility and sterilization validation) would have been performed there or at certified labs. These are bench tests, not clinical data sets.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. There's no "ground truth" to establish in the context of an AI/ML algorithm or diagnostic accuracy study. The assessment is based on physical device characteristics and established standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This is not a study requiring adjudication of expert interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a physical dental device (suprastructure), not an AI algorithm.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical dental device (suprastructure), not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Not applicable. The "ground truth" for this medical device submission is compliance with engineering specifications, material standards (e.g., ASTM F136), and validated manufacturing processes, all demonstrated through non-clinical bench testing.

    8. The sample size for the training set:

    • Not applicable. This is a physical dental device. While it is manufactured using CAD/CAM technology, it does not involve an AI/ML model that requires a training set of data.

    9. How the ground truth for the training set was established:

    • Not applicable. See point 8.