510(k) Data Aggregation

    K Number
    K211078

    Validate with FDA (Live)

    Device Name
    Progreat Lambda
    Date Cleared
    2021-12-21

    (253 days)

    Product Code
    Regulation Number
    870.1200
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Progreat Lambda is intended for the infusion of contrast media, or embolic materials for hemostasis, into the peripheral vasculature, excluding the blood vessels belonging to the central circulatory system. Progreat Lambda is also indicated for drug infusion in intra-arterial therapy in the peripheral vasculature. Progreat Lambda should not be used in cerebral or coronary vessels.

    Device Description

    Progreat Lambda is a single use, ethylene oxide sterilized device that is intended for the infusion of contrast media, or embolic materials for hemostasis, into the peripheral vasculature, excluding the blood vessels belonging to the central circulatory system. Progreat Lambda is also indicated for drug infusion in intra-arterial therapy in the peripheral vasculature. Progreat Lambda should not be used in cerebral or coronary vessels. The catheter consists of metal wire mesh reinforced multi-layer polymer tubing. The mesh is embedded in the catheter wall the entire length of the catheter with the exception of the distal tip section. This increases the flexibility, kink resistance, and pressure resistance of the catheter. The inner layer of the catheter is made of PTFE (polytetrafluoroethylene) to ensure smooth movement of devices such as the guide wire. The outer surface of the catheter is coated with a hydrophilic polymer with the exception of the proximal end that is 60cm from the catheter hub. The coating becomes lubricious when wet with saline solution or blood. The device is offered in effective lengths of 110, 130, 150, 165 and 175 cm.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and supporting studies for the "Progreat Lambda" device based on the provided FDA 510(k) summary:

    The document does not contain information relevant to AI/ML or a multi-reader multi-case (MRMC) study. The device described is a microcatheter, a physical medical device, not a diagnostic software or AI/ML algorithm. Therefore, many of the requested categories like effect size of AI assistance, standalone algorithm performance, number of experts for ground truth, and training set information are not applicable to this submission.

    The acceptance criteria and studies focus on the physical and biological performance of the catheter itself.


    Acceptance Criteria and Reported Device Performance

    1. A table of acceptance criteria and the reported device performance

    The document states that "Performance testing met the predetermined acceptance criteria and is acceptable for clinical use throughout its shelf life." However, it does not explicitly list the quantitative acceptance criteria for each test. It only lists the tests performed. Without the specific numerical thresholds for "acceptance criteria," a direct comparative table with reported performance cannot be fully constructed from this document.

    Below is a table summarizing the types of performance tests conducted, implying that the device met the unstated acceptance criteria for each.

    Test ItemAcceptance Criteria (Implied)Reported Device Performance (Implied)
    Radio-detectabilityDevice is detectable as required.Met predetermined acceptance criteria.
    SurfaceSurface characteristics meet specified standards (e.g., smoothness, freedom from defects).Met predetermined acceptance criteria.
    Peak tensile forceWithstands specified tensile forces without failure.Met predetermined acceptance criteria.
    Freedom from leakageNo leakage occurs under specified conditions.Met predetermined acceptance criteria.
    Fluid leakage (Hub)Hub does not leak fluid under specified pressure.Met predetermined acceptance criteria.
    Sub-atmospheric pressure air leakage (Hub)Hub does not leak air under specified negative pressure.Met predetermined acceptance criteria.
    Stress cracking (Hub)Hub maintains integrity without cracking under stress.Met predetermined acceptance criteria.
    Resistance to separation from axial load (Hub)Hub remains securely attached under specified axial load.Met predetermined acceptance criteria.
    Resistance to separation from unscrewing (Hub)Hub remains securely attached despite unscrewing attempts.Met predetermined acceptance criteria.
    Resistance to overriding (Hub)Hub maintains proper connection without overriding.Met predetermined acceptance criteria.
    Power injectionCatheter withstands specified power injection pressures.Met predetermined acceptance criteria.
    Distal tipDistal tip performance meets specified criteria (e.g., flexibility, shape retention).Met predetermined acceptance criteria.
    Particulate evaluationDevice is free from unacceptable levels of particulates.Met predetermined acceptance criteria.
    Coating integrityHydrophilic coating remains intact and functional.Met predetermined acceptance criteria.
    Torque strengthCatheter withstands specified torque without damage or failure.Met predetermined acceptance criteria.
    Distal marker strengthDistal marker remains securely attached and detectable.Met predetermined acceptance criteria.
    Product dimensionDimensions conform to specifications.Met predetermined acceptance criteria.
    Embolic device compatibilityCompatible with specified embolic devices for intended use.Met predetermined acceptance criteria.
    Flexibility and kink testCatheter maintains flexibility and resists kinking under specified conditions.Met predetermined acceptance criteria.
    Wire compatibilityCompatible with specified guidewires.Met predetermined acceptance criteria.
    Simulated UsePerformance under simulated clinical conditions meets requirements.Met predetermined acceptance criteria.
    Biocompatibility (various tests listed below)Meets ISO 10993 standards for biological safety.Found to be biocompatible throughout the shelf life.
    Sterilization (Ethylene Oxide)Achieves Sterility Assurance Level (SAL) of 10-6.Confirmed SAL of 10-6 in accordance with ISO 11135:2014.

    2. Sample size used for the test set and the data provenance

    • Performance Testing: The document does not explicitly state the numerical sample size for each performance test. It mentions that tests were performed on "nonaged and accelerated aged samples," indicating multiple samples were used for each condition. The provenance of the data is internal to the manufacturer (Terumo Clinical Supply Co., Ltd., Japan). The tests are presumably prospective laboratory tests conducted on manufactured device units.
    • Biocompatibility Testing: The specific numerical sample size is not provided, but it notes tests were conducted on "non-aged, sterile, whole device" and "accelerated-aged (3 years), sterile, whole device." These tests were also prospective laboratory tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is a physical medical device, not an AI/ML diagnostic. Ground truth as typically defined for AI/ML (e.g., expert consensus on medical images) does not apply. The "ground truth" for this device's performance is established by objective engineering and biocompatibility standards.

    4. Adjudication method for the test set

    Not applicable. This is not a diagnostic device requiring human reader adjudication. Performance is assessed against predefined engineering and biological criteria.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This is a physical medical device (microcatheter), not an AI/ML diagnostic system. Therefore, MRMC studies and AI assistance effect sizes are not relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. This is a physical medical device. The concept of an "algorithm only" performance is not applicable.

    7. The type of ground truth used

    The "ground truth" for evaluating this device's performance is based on established engineering standards and specifications (e.g., tensile strength, leak rates, dimensions, coating integrity), and biocompatibility standards (ISO 10993-1, pyrogenicity, hemolysis, cytotoxicity, etc.). These are objective, measurable criteria. For manufacturing and quality, the "ground truth" is adherence to design specifications and validated processes.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not an AI/ML algorithm that requires a "training set."

    9. How the ground truth for the training set was established

    Not applicable. (See #8).

    Ask a Question

    Ask a specific question about this device

    K Number
    K170223

    Validate with FDA (Live)

    Device Name
    Progreat
    Date Cleared
    2017-07-21

    (177 days)

    Product Code
    Regulation Number
    870.1200
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Progreat is intended for the infusion of contrast media into all peripheral vessels up to and including the cervical vessels, all vessels in the lower and upper extremities and all coronary vessels. The Progreat is also intended for drug infusion in intra-arterial therapy and the infusion of embolic materials for hemostasis. The Progreat should not be used in cerebral vessels.

    Device Description

    The Progreat catheter consists of metal coil reinforced multi-layer polymer tubing with a hydrophilic coating. The guidewire is comprised of an alloy core wire with radiopaque marker and hydrophilic coating. The subject 2.4 Fr Progreat catheter and 0.018" guidewire will be marketed as a combined unit and will be an extension of the existing Progreat product family. The design and technological characteristics of the subject 2.4 Fr Progreat with 0.018" guidewire are identical to the predicate 2.7 Fr Progreat with 0.021" guidewire.

    The Progreat catheter is available with or without accessories. The accessories to the catheter are supplied in different configurations depending on the product code:

    • The Guidewire has a super-elastic alloy core and is surface coated with a hydrophilic polymer.
    • I The Inserter is used to assist the physician in the placement of the guidewire within the catheter.
    • I The Catheter Mandrel (stylet) is used in the shaping of the catheter for procedures that require a catheter with a tip configuration other than straight.
    • I The Syringe is used in the priming of the catheter.
    • . The Wire Stopper can be clipped onto the guide wire to adjust the length of the guidewire that extends past the catheter tip.
    • I The Catheter Stopper S can be clipped onto the catheter to adjust the insertion length of the catheter.
    • . The Y-connector can be used to connect a power injector unit to the end of the catheter for infusion of contrast media.
    AI/ML Overview

    The provided text is a 510(k) summary for the Progreat catheter, which is a medical device. This document focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than providing a detailed study proving the device meets specific acceptance criteria in the context of an AI/ML-driven medical device.

    Therefore, many of the requested criteria for AI/ML device studies (such as types of ground truth, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance) are not applicable to this document.

    However, I can extract information related to the device's technical specifications and the non-clinical tests performed to assess its performance against predetermined criteria.


    Description of the Acceptance Criteria and the Study that Proves the Device Meets the Acceptance Criteria

    This 510(k) submission for the Progreat (2.4 Fr Catheter with 0.018" Guidewire) aims to demonstrate substantial equivalence to its predicate device (K033583, Terumo Progreat). The acceptance criteria for this submission are primarily focused on ensuring that the modified device maintains the safety and effectiveness of the predicate device and conforms to applicable external and internal standards. The study supporting this is a series of non-clinical performance tests.

    1. Table of Acceptance Criteria and the Reported Device Performance

    The document states that "Performance testing met the predetermined acceptance criteria and is acceptable for clinical use throughout its shelf life." While the specific numerical acceptance criteria for each test are not explicitly detailed in this summary, the types of performance tests conducted serve as the basis for these criteria.

    Acceptance Criteria Category/Test TypeReported Device Performance
    Material/Surface IntegrityMet predetermined criteria
    Radio-detectabilityMet predetermined criteria
    Mechanical Integrity:
    - Fracture testMet predetermined criteria
    - Flexing testMet predetermined criteria
    - Peak tensile force of guidewireMet predetermined criteria
    - Torque StrengthMet predetermined criteria
    - TorqueabilityMet predetermined criteria
    - Tip FlexibilityMet predetermined criteria
    - Bending strengthMet predetermined criteria
    Functional Performance:
    - Sliding resistanceMet predetermined criteria
    - Particulate evaluationMet predetermined criteria
    - Hermeticity testMet predetermined criteria
    - Shaping testMet predetermined criteria
    Dimensional ConformanceMet predetermined criteria
    BiocompatibilityDeemed equivalent to predicate (materials, formulation, processing, sterilization, geometry are identical or not raising new concerns)
    SterilizationDeemed equivalent to predicate (identical method, not raising new concerns)
    Shelf Life2 years (Identical to predicate)

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the specific sample sizes for each non-clinical test conducted. It mentions that tests were performed on "non-aged and accelerated aged samples." The provenance of the data is from Terumo Corporation's Ashitaka Factory in Japan, where the device is manufactured and where these non-clinical tests were presumably conducted. This is retrospective in the sense that the testing was performed on manufactured devices to support the 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable as the device is a physical medical instrument, not an AI/ML diagnostic tool requiring expert interpretation for ground truth. Performance was assessed against engineering specifications and industry standards.

    4. Adjudication Method for the Test Set

    This information is not applicable for a physical device's non-clinical performance testing. Adjudication methods like 2+1 or 3+1 typically apply to human interpretation of diagnostic images or data.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted, and such a study is not applicable for this type of medical device (intravascular catheter). This type of study is relevant for evaluating the impact of AI assistance on human reader performance in diagnostic tasks.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    A standalone algorithm-only performance assessment was not conducted, as this device is a physical catheter and not an algorithm. This criterion is not applicable.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance tests effectively refers to predetermined engineering specifications, design requirements, and relevant industry standards for intravascular catheters (e.g., ISO standards for medical devices, internal Terumo specifications). The tests confirm if the device meets these established benchmarks.

    8. The Sample Size for the Training Set

    This information is not applicable. The concept of a "training set" applies to machine learning models, not to the manufacturing and testing of a physical medical device.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the reasons stated above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033913

    Validate with FDA (Live)

    Device Name
    PROGREAT
    Date Cleared
    2004-03-11

    (85 days)

    Product Code
    Regulation Number
    870.1210
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Progreat™ is intended for the infusion of contrast media into all peripheral vessels up to and including the cervical vessels, all vessels in the lower and upper extremities, visceral vessels, and all coronary vessels. The Progreat is also intended for drug infusion in intra-arterial therapy and the infusion of embolic materials for hemostasis in procedures including but not limited to Uterine Fibroid Embolization. The Progreat should not be used in cerebral vessels.

    Device Description

    The Progreat catherer is available with or without the following accessories: guide wire, inserter, mandrel (stylet), syringe, wire stopper, and Y-connector. The catheter consists of metal coil reinforced multi-layer polymer tubing. The coil is embedded in the catheter wall the entire length of the catheter. This increases the flexibility, kink resistance, and pressure resistance of the catheter. The inner layer of the catheter is made of PTFE (polytetrafluoroethylene) to ensure smooth movement of devices such as the guide wire. The outer surface of the catheter is coated with a hydrophilic polymer which becomes lubricious when wet with saline solution or blood.

    AI/ML Overview

    This document outlines an application for clearance of a medical device, the Progreat™ catheter, based on substantial equivalence to a previously cleared device (K033583). As such, it does not contain a study report for the device meeting specific quantitative acceptance criteria in the way a medical AI product might.

    Therefore, many of the requested sections (2-9) regarding sample size, ground truth, expert adjudication, and comparative effectiveness studies are not applicable to this type of regulatory submission (a 510(k) for substantial equivalence of a physical device with minor changes/expanded indications where no new performance data was required).

    However, I can extract information related to the device description, intended use, and the basis for its perceived performance and safety as presented.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (from text)Reported Device Performance (from text)
    Intended UseInfusion of contrast media into specified peripheral, upper/lower extremity, visceral, and coronary vessels. Drug infusion in intra-arterial therapy and infusion of embolic materials for hemostasis (including UFE).The Progreat™ is intended for these uses. (This reflects its design purpose, rather than a quantifiable performance metric in this context).
    Safety (Biocompatibility)Blood contacting materials must be biocompatible per FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO-10993."Results of the testing demonstrate that the blood contacting materials are biocompatible."
    Safety (Sterilization)Sterilization conditions validated to a SAL of 10^-6 per AAMI/ANSI/ISO 11135 Medical Devices - Validation and Routine Control of Ethylene Oxide Sterilization. ETO residuals within maximum limits proposed for Part 821 of Title 21 in FR June 23, 1978."Sterilization conditions have been validated in accordance to AAMI/ANSI/ISO 11135... The device is sterilized to a SAL of 10^-6. ETO residuals for the Progreat will not exceed the maximum limits proposed..."
    Performance (General)Demonstrate performance substantially equivalent to the predicate device (K033583). No new issues of safety and effectiveness with expanded indications."This device is the identical device cleared under K033583... There were no new issues of safety and effectiveness with regards to the new indication. No additional testing was deemed necessary. Therefore the performance of the Progreat Catheter is substantially equivalent to the performance of the Progreat, cleared under K033583."
    Design/MaterialsIdentical design and materials to previously cleared device (K033583)."This device is the identical device cleared under K033583. There are no material or design changes therefore there are no new issues of safety and effectiveness."
    Risk ManagementRisk/hazard analysis conducted according to EN1441 and ISO 14971."A risk/hazard analysis was conducted according to EN1441 Medical devices- Risk analysis and ISO 14971 Medical Devices -Application of risk management to medical devices."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable. This submission is based on substantial equivalence to a prior device where no new performance testing data (i.e., a "test set") was required for the expanded indications. The performance is deemed identical to the predicate device due to identical design and materials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. As no new performance studies were conducted, there was no test set requiring expert-established ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. No test set was used for performance evaluation in this submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is a physical medical device (catheter), not an AI diagnostic tool. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not Applicable. This is a physical medical device (catheter), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable. Ground truth, in the context of performance studies for diagnostic or AI devices, is not relevant here as no such studies were performed for this substantial equivalence application. The "ground truth" for this device's safety and effectiveness relies on the previously cleared device's performance, safety, and the adherence to recognized standards for biocompatibility and sterilization.

    8. The sample size for the training set

    • Not Applicable. This is a physical medical device. There is no "training set" in the context of AI or machine learning.

    9. How the ground truth for the training set was established

    • Not Applicable. As there is no training set, this question is not relevant.

    Summary of the Study that Proves the Device Meets Acceptance Criteria:

    The "study" here is not an empirical performance trial but rather a demonstration of substantial equivalence to a previously cleared predicate device (K033583) and adherence to recognized standards.

    The principal arguments for meeting acceptance criteria are:

    • Identical Device: The Progreat™ catheter submitted in K033913 is explicitly stated as being the "identical device" cleared under K033583, with no material or design changes. This implies that its performance characteristics (e.g., flexibility, kink resistance, pressure resistance, smooth movement of guide wires) are the same as the predicate.
    • Expanded Intended Use with No New Safety/Effectiveness Issues: The new application expands the intended use to include additional vessels and specific procedures like Uterine Fibroid Embolization. However, the submitter asserts that "There were no new issues of safety and effectiveness with regards to the new indication. No additional testing was deemed necessary." This implies that the existing safety and performance profile of the identical device is considered sufficient for these expanded indications.
    • Adherence to Standards:
      • Risk Management: A risk/hazard analysis was performed according to EN1441 and ISO 14971, indicating a structured approach to identifying and mitigating risks.
      • Biocompatibility: Blood-contacting materials were tested in accordance with ISO-10993, and results demonstrated they are biocompatible.
      • Sterilization: Sterilization conditions were validated to a Sterility Assurance Level (SAL) of 10^-6 according to AAMI/ANSI/ISO 11135, and ETO residuals meet regulatory limits.
    • Accessories: The accessories supplied with the Progreat™ are also the "same as the ones used with the currently cleared device K033583."

    In essence, the "proof" that the device meets acceptance criteria is the argument of substantial equivalence based on:

    1. Being the exact same device as a previously cleared one.
    2. Demonstrating that the expanded indications for use do not introduce new safety or effectiveness concerns.
    3. Compliance with established regulatory standards for biocompatibility, sterilization, and risk management.
    Ask a Question

    Ask a specific question about this device

    K Number
    K033583

    Validate with FDA (Live)

    Device Name
    PROGREAT
    Date Cleared
    2003-12-12

    (29 days)

    Product Code
    Regulation Number
    870.1200
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PROGREAT Catheter is intended for the infusion of contrast media into all peripheral vessels up to and including the cervical vessels in the lower and upper extremities and all coronary vessels. The PROGREAT Catheter is also intended for drug infusion in intra-arterial therapy and the infusion of embolic materials for hemostasis. The Catheter should not be used in cerebral vessels.

    Device Description

    The PROGREAT catheter is available with or without the following accessories: guide wire, inserter, mandrel (stylet), syringe, wire stopper, and Y-connector. The catheter consists of metal coil reinforced multi-layer polymer tubing. The coil is embedded in the catheter wall the entire length of the catheter. This increases the flexibility, kink resistance, and pressure resistance of the catheter. The inner layer of the catheter is made of PTFE (polytetrafluoroethylene) to ensure smooth movement of devices such as the guide wire. The outer surface of the catheter is coated with a hydrophilic polymer which becomes lubricious when wet with saline solution or blood.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the PROGREAT Catheter, asserting its substantial equivalence to a predicate device, the GT LEGGIERO (K981359). The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the performance tests conducted to demonstrate this substantial equivalence.

    Here's an analysis based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table labeled "acceptance criteria." Instead, it lists "verification tests" performed to demonstrate substantial equivalence to the predicate device. For each test, it implicitly states that the performance was satisfactory to deem the device substantially equivalent, meaning it met an internal standard set by the manufacturer to compare to the predicate.

    Verification Test (Acceptance Criteria)Reported Device Performance
    Ease of removal from holderMet (No new issues of safety and effectiveness raised)
    Frictional resistance of catheter and guide wireMet (No new issues of safety and effectiveness raised)
    Joint strength of hub on catheter and guide wireMet (No new issues of safety and effectiveness raised)
    Pressure resistance of catheterMet (No new issues of safety and effectiveness raised)
    Kink resistanceMet (No new issues of safety and effectiveness raised)

    In addition to these performance tests, the device also met biological and sterilization acceptance criteria:

    Acceptance CriteriaReported Performance
    Biocompatibility (for blood contacting materials)Blood contacting materials tested in accordance with FDA G95-1; results demonstrate biocompatibility.
    SterilizationValidated in accordance with AAMI/ANSI/ISO 11135 to a SAL of 10⁻⁶; ETO residuals within limits.

    2. Sample Size Used for the Test Set and Data Provenance
    The document does not explicitly state the sample sizes used for the specified verification tests (e.g., Ease of removal, Frictional resistance, etc.). These are likely internal engineering tests, and the exact number of units tested is not typically disclosed in the public summary. The data provenance is internal to the manufacturer, Terumo Medical Corporation, for the "modified device (PROGREAT catheter)." These would be prospective tests performed during the development and verification phase of the PROGREAT catheter.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
    This information is not applicable in the context of this 510(k) submission. The "ground truth" for the performance tests mentioned (e.g., frictional resistance, kink resistance) would be the physical properties and performance characteristics of the predicate device (GT LEGGIERO K981359) and established engineering standards. There is no mention of human expert consensus being used to establish the "ground truth" for these physical device performance characteristics.

    4. Adjudication Method for the Test Set
    Not applicable. The verification tests are objective physical and mechanical tests, not subjective evaluations requiring adjudication by experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
    Not applicable. This device is a physical medical device (angiographic catheter), not an AI imaging or diagnostic algorithm. Therefore, an MRMC study related to AI assistance for human readers is not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
    Not applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used
    The "ground truth" for demonstrating substantial equivalence for the PROGREAT Catheter primarily relied on:

    • Predicate Device Performance: The established performance characteristics of the GT LEGGIERO (K981359) served as the benchmark for comparison.
    • Engineering Standards: Standardized tests for mechanical properties (e.g., pressure resistance, kink resistance, joint strength) and material properties were used.
    • Regulatory Standards: Biocompatibility was assessed against FDA General Program Memorandum #G95-1 (ISO-10993), and sterilization against AAMI/ANSI/ISO 11135.

    8. The Sample Size for the Training Set
    Not applicable. This device does not involve a "training set" in the context of machine learning or AI.

    9. How the Ground Truth for the Training Set Was Established
    Not applicable. As noted above, there is no training set for this type of device submission.