Search Filters

Search Results

Found 17 results

510(k) Data Aggregation

    K Number
    K240538

    Validate with FDA (Live)

    Date Cleared
    2024-07-19

    (144 days)

    Product Code
    Regulation Number
    872.3770
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Premium Teeth Resin, when utilized to print 3D printed dental appliances such as denture teeth for complete and partial removable dentures, try-in dentures, provisional full arch implant-supported restoration and provisional restorations such as temporary crowns and bridges, inlays, onlays and veneers, is indicated to replace or restore missing tooth structures or missing teeth.

    Device Description

    Formlabs Premium Teeth Resin is a light-curable polymer-based resin of 3D printed dental appliances, such as denture teeth for complete and partial res, try-in dentures, provisional full arch implant-supported restoration, and provisional restorations including temporary crowns and bridges, inlays, onlays and veneers. Formlabs Premium Teeth Resin is used to fabricate patient-specific dental appliances in a stereolithographic (SLA) 3D printer using layer additive manufacturing.

    AI/ML Overview

    This document is a 510(k) summary for a medical device (Premium Teeth Resin) and does not describe acceptance criteria or a study that proves the device meets those criteria in the context of an AI/ML-based medical device.

    The provided text focuses on demonstrating substantial equivalence of a new 3D printing resin (Premium Teeth Resin) to a legally marketed predicate device (VarseoSmile Temp) for dental applications. It details the product, its intended use, and compares its technological characteristics and performance to the predicate through non-clinical testing.

    Therefore, I cannot extract the information required for your request from this document. The concepts of "acceptance criteria" and "study proving device meets acceptance criteria" as they relate to AI/ML device performance (e.g., sensitivity, specificity, AUC, human reader improvement with AI assistance, ground truth establishment, sample sizes for AI model training/testing) are not present in this regulatory submission for a material.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200602

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-06-03

    (86 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Premium Cooling Personal Lubricant is a personal lubricant for penile, anal and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex, polyisoprene and polyurethane condoms.

    Premium Anal Cooling Lubricant is a personal lubricant for penile, anal and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex, polyisoprene and polyurethane condoms.

    Device Description

    Premium Cooling Personal Lubricant and Premium Anal Cooling Lubricant are clear, semi-viscous personal lubricants that are compatible with natural rubber latex, polyisoprene and polyurethane condoms. These devices are non-sterile personal lubricants for penile, anal and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity supplement the body's natural lubrication.

    Premium Cooling Personal Lubricant is sold as an over-the-counter (OTC) product in 1 fl. oz./30 mL and 2 fl. oz./60 mL sizes. This product is provided in a clear polyethylene terephthalate (PET) cylinder bottle. The 1 fl. oz./30 mL size bottles are capped with natural disc tops. The 2 fl. oz./60 mL size bottles are capped with silver disc tops. The individual bottles are hermetically sealed during the production process.

    Premium Anal Cooling Lubricant is sold as an over-the-counter (OTC) product in 2 fl. oz./60 mL and 4 fl. oz./120 mL sizes. This product is provided in a clear polyethylene terephthalate (PET) cylinder bottle. The 2 fl. oz./60 mL and 4 fl. oz./120 mL size bottles are capped with silver disc tops. The individual bottles are hermetically sealed during the production process.

    These devices are composed of Dimethicone, Dimethiconol and Menthol.

    AI/ML Overview

    The provided FDA 510(k) summary document for "Premium Cooling Personal Lubricant" and "Premium Anal Cooling Lubricant" is for a personal lubricant, not an AI/ML device. Therefore, the questions related to AI/ML device acceptance criteria and studies (such as sample sizes for test sets, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, training set sample size, and ground truth establishment for training) are not applicable to this document.

    However, I can extract the acceptance criteria and study information for this non-AI/ML medical device from the document.

    Acceptance Criteria and Reported Device Performance

    PropertySpecification (Acceptance Criteria)Reported Device Performance
    AppearanceSemi-viscous liquidNot explicitly stated as "met", but implied by description
    ColorColorlessNot explicitly stated as "met", but implied by description
    OdorOdorlessNot explicitly stated as "met", but implied by description
    Viscosity (cps) per USP <911>800 to 1,075Met (implied by "The subject device met the device specifications at all time points" for shelf-life)
    Specific Gravity per USP <841>0.875 to 1.020Met (implied by "The subject device met the device specifications at all time points" for shelf-life)
    Antimicrobial effectiveness per USP <51>Meets USP <51> acceptance criteria for Category 2 products.Met (explicitly stated "Antimicrobial effectiveness per USP <51> meets USP <51> acceptance criteria for Category 2 products")
    Total aerobic microbial count (TAMC) per USP <61> and <1111>Less than 100 cfu/gMet (explicitly stated "Less than 100 cfu/g")
    Total yeast and mold count (TYMC) per USP <61> and <1111>Less than 10 cfu/gMet (explicitly stated "Less than 10 cfu/g")
    Presence of Pathogens per USP <62>Pseudomonas aeruginosa: AbsentStaphylococcus aureus: AbsentSalmonella/Shigella: AbsentEscherichia coli: AbsentCandida albicans: AbsentMet (explicitly stated "Absent" for all listed pathogens)
    Biocompatibility - CytotoxicityPassing reactivity score (non-cytotoxic)"only slightly cytotoxic with a passing reactivity score"
    Biocompatibility - SensitizationNon-sensitizing"non-sensitizing"
    Biocompatibility - Vaginal IrritationNon-irritating"non-irritating"
    Biocompatibility - Acute Systemic ToxicityNon-systemically toxic"non-systemically toxic"
    Shelf-LifeMaintain all device specifications for 3 years"met the device specifications at all time points" (0, 1, 2, and 3 years)
    Condom CompatibilityCompatible with natural rubber latex, polyisoprene, and polyurethane condoms"are compatible with natural rubber latex, polyisoprene and polyurethane condoms"

    Study Information (for non-AI/ML device):

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Biocompatibility Studies: Not specified in terms of sample size for living organisms. The tests performed are standard in vitro and in vivo toxicology tests (Cytotoxicity, Sensitization, Vaginal Irritation, Acute Systemic Toxicity). Data provenance is not mentioned.
      • Shelf-Life Study: Samples were tested at 0, 1, 2, and 3 years. The number of individual product units tested at each time point is not specified. Data provenance is not mentioned.
      • Condom Compatibility Study: Evaluated in accordance with ASTM D7661-18. The number of condoms or lubricant samples tested is not specified. Data provenance is not mentioned.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable as this is not an AI/ML diagnostic device requiring expert interpretation for ground truth. Performance is assessed against scientific/regulatory standards and specifications.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable as this is not an AI/ML diagnostic device.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable as this is not an AI/ML diagnostic device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable as this is not an AI/ML device. The device itself is the "standalone" product being evaluated.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The "ground truth" here refers to established scientific and regulatory standards/specifications (e.g., USP monographs for viscosity, microbial limits, and antimicrobial effectiveness; ISO standards for biocompatibility; ASTM standards for condom compatibility).
    7. The sample size for the training set:

      • Not applicable as this is not an AI/ML device that requires training data.
    8. How the ground truth for the training set was established:

      • Not applicable as this is not an AI/ML device that requires training data.
    Ask a Question

    Ask a specific question about this device

    K Number
    K200457

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-05-22

    (87 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    0 - 999
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Premium Warming Personal Lubricant, Premium Anal Warming Lubricant and All-In-One Warming are personal lubricants for penile, anal and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. These products are compatible with natural rubber latex, polyisoprene and polyurethane condoms.

    Device Description

    Premium Warming Personal Lubricant is sold as an over-the-counter (OTC) product in 1 fl. oz./30 mL and 2 fl. oz./60 mL sizes. This product is provided in a clear polyethylene terephthalate (PET) cylinder bottle. The 1 fl. oz./30 mL size bottles are capped with natural disc tops. The 2 fl. oz./60 mL size bottles are capped with silver disc tops. The individual bottles are hermetically sealed during the production process.

    Premium Anal Warming Lubricant is sold as an over-the-counter (OTC) product in 2 fl. oz./60 mL size. This product is provided in a clear polyethylene terephthalate (PET) cylinder bottle. The 2 fl. oz./60 mL size bottles are capped with silver disc tops. The individual bottles are hermetically sealed during the production process.

    All-In-One Warming is sold as an over-the-counter (OTC) product in 1 fl. oz./30 mL and 4 fl. oz./120 mL sizes. This product is provided in a clear polyethylene terephthalate (PET) cylinder bottle. The 1 fl. oz./30 mL size bottles are capped with natural disc tops. The 4 fl. oz./120 mL size bottles are capped with silver disc tops. The individual bottles are hermetically sealed during the production process.

    This device is composed of Dimethicone, Dimethiconol and Capsicum Frutescens Fruit Extract.

    AI/ML Overview

    The provided text describes the 510(k) summary for the Premium Warming Personal Lubricant, Premium Anal Warming Lubricant, and All-In-One Warming. This document is a regulatory submission to the FDA, and as such, it presents performance data to demonstrate substantial equivalence to a predicate device, rather than a "study that proves the device meets acceptance criteria" in the context of an AI/algorithm-driven device.

    Given that this document is for a personal lubricant (a physical product, not an AI/algorithm), most of the requested information (like sample size for test/training sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, ground truth types) is not applicable to this type of device and submission.

    However, I can extract the acceptance criteria and reported device performance from the provided "Device Specifications" table and the "Summary of Performance Data."

    Here's the information that can be extracted, and where applicable, a note on why certain requested fields are not relevant for this device type:


    1. Table of Acceptance Criteria and Reported Device Performance

    PropertyAcceptance CriteriaReported Device Performance
    AppearanceSemi-viscous liquidSemi-viscous liquid (implied, as it meets specification)
    ColorColorlessColorless (implied, as it meets specification)
    OdorOdorlessOdorless (implied, as it meets specification)
    Viscosity (cps) per USP <911>800 to 1,075Meets specification (actual value not provided, but within range)
    Antimicrobial effectiveness per USP <51>Meets USP <51> acceptance criteria for Category 2 products.Meets USP <51> acceptance criteria for Category 2 products.
    Total aerobic microbial count (TAMC) per USP <61> and <1111>Less than 100 cfu/gLess than 100 cfu/g (implied, as it meets specification)
    Total yeast and mold count (TYMC) per USP <61> and <1111>Less than 10 cfu/gLess than 10 cfu/g (implied, as it meets specification)
    Presence of Pathogens per USP <62> (Pseudomonas aeruginosa)AbsentAbsent (implied, as it meets specification)
    Presence of Pathogens per USP <62> (Staphylococcus aureus)AbsentAbsent (implied, as it meets specification)
    Presence of Pathogens per USP <62> (Salmonella/Shigella)AbsentAbsent (implied, as it meets specification)
    Presence of Pathogens per USP <62> (Escherichia coli)AbsentAbsent (implied, as it meets specification)
    Presence of Pathogens per USP <62> (Candida albicans)AbsentAbsent (implied, as it meets specification)
    BiocompatibilityPass (based on ISO standards)The subject lubricant is biocompatible.
    Shelf-Life3 years (maintaining all device specifications)3 years (met device specifications at 0, 1, 2, and 3 years)
    Condom CompatibilityCompatible with natural rubber latex, polyisoprene and polyurethane condoms (per ASTM D7661-18)Compatible with natural rubber latex, polyisoprene, and polyurethane condoms.

    Regarding the studies conducted, the following information is available:

    • Biocompatibility Studies:

      • Tests performed: Acute Systemic Toxicity, Vaginal Irritation Testing, Penile Irritation Testing, Cytotoxicity, and Sensitization Testing.
      • Standards: ISO 10993-1:2016, ISO 10993-1:20019, ISO 10993-5:2009, ISO 10993-10:2010, ISO 10993-11:2006.
      • Outcome: "The results of this testing demonstrated that the subject lubricant is biocompatible."
    • Shelf-Life Study:

      • Design: Real-time aging study.
      • Duration: 3 years.
      • Testing points: 0, 1, 2, and 3 years.
      • Parameters tested: All device specifications listed in Table 1 (Appearance, Color, Odor, Viscosity, Antimicrobial effectiveness, Microbial counts, Pathogen presence).
      • Outcome: "The subject device met the device specifications at all time points."
    • Condom Compatibility Study:

      • Standard: ASTM D7661-18 Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms.
      • Results: "Premium Warming Personal Lubricant, Premium Anal Warming Lubricant and All-In-One Warming are compatible with natural rubber latex, polyisoprene, and polyurethane condoms."

    The following requested information is NOT APPLICABLE or NOT PROVIDED for this type of medical device (personal lubricant) in this regulatory submission:

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable for a physical product; refers to software/AI. The "test set" here would refer to the physical samples of the lubricant subjected to the various chemical, microbiological, and physical tests. The document does not specify the number of units tested for each specific criterion but implies standard laboratory practices were followed.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable to a physical product; refers to AI/algorithm performance evaluation. Ground truth for these tests comes from established scientific methods, not expert consensus in the diagnostic sense.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable; refers to AI/algorithm. Laboratory tests have defined endpoints and pass/fail criteria.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable; refers to AI/algorithm.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable; refers to AI/algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): For this physical product, "ground truth" is established through standardized laboratory testing methods (e.g., USP monographs for microbial testing, ISO standards for biocompatibility, ASTM for condom compatibility).
    7. The sample size for the training set: Not applicable; refers to AI/algorithm.
    8. How the ground truth for the training set was established: Not applicable; refers to AI/algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K161989

    Validate with FDA (Live)

    Date Cleared
    2018-02-01

    (562 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PREMIUM-SHELTA Abutments are intended to be used in conjunction with PREMUM-SHELTA Implants Systems in fully edentulous or partially edentulous maxillary and/or mandibular arches. PREMIUM-SHELTA Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prostheses, in the mandible or maxilla. The prosthesis can be cemented, screw retained or friction fit to the abutment screw is intended to secure the abutment to the endosseous implant. PREMIUM-SHELTA Abutments are compatible with PREMIUM-SHELTA Implants Systems

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text is a cover letter from the FDA regarding a 510(k) premarket notification for "PREMIUM-SHELTA Prosthetic Components". It also includes the "Indications for Use" for these components.

    This document does not contain any information about acceptance criteria, device performance studies, sample sizes, expert qualifications, or ground truth establishment. The text focuses on regulatory approval and general controls for a medical device.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets them based on the provided input.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172560

    Validate with FDA (Live)

    Date Cleared
    2017-12-21

    (118 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PREMIUM ONE Implant Systems are intended for both one- and two-stage surgical procedures. PREMIUM ONE Implant Systems are intended for immediate placement and function on single tooth and/or multiple tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore chewing function. Multiple tooth applications may be splinted with a bar.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text from the FDA 510(k) letter for the "PREMIUM ONE Implant Systems" does not contain information about acceptance criteria or the study that proves the device meets those criteria. This document is a clearance letter, indicating that the device has been found substantially equivalent to a predicate device, and outlines regulatory requirements and contact information. It does not elaborate on performance studies or acceptance criteria.

    Therefore, I cannot provide the requested information based on the given input.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170261

    Validate with FDA (Live)

    Date Cleared
    2017-10-19

    (265 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Premium Plus Prophy Air Motor is intended to be used with a disposable prophy angle and polishing paste by a trained dental clinician to perform dental prophylaxis.

    Device Description

    Premium Plus Prophy Air Motor is a dental low speed prophy air motor with pneumatic rotor, scroll bearing, stainless steel inner parts and anodized aluminum body surface. With the impulse of air current supplied. it can drive a disposable prophy angle with polishing paste to polish teeth.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Premium Plus Prophy Air Motor. Based on the document, this is a Class I medical device (dental handpiece and accessories). For Class I devices, the level of regulatory scrutiny and required evidence of "performance" to establish substantial equivalence is generally much lower than for Class II or Class III devices, especially those involving AI/ML.

    The provided document describes a 510(k) premarket notification, which aims to demonstrate "substantial equivalence" to a legally marketed predicate device, not necessarily to prove efficacy or performance in the same way an AI/ML device would be scrutinized. Therefore, the "acceptance criteria" here relate to meeting basic safety and performance standards for a simple mechanical device and demonstrating similarity to an existing device. The "study" refers to the engineering and biocompatibility tests conducted.

    Most of the specific information requested in your prompt (e.g., sample size for test set regarding AI algorithms, number of experts for ground truth, MRMC studies, standalone AI performance) is not applicable or not present in this type of submission for a simple mechanical dental device. This FDA submission is not for an AI/ML device.

    Here's the closest breakdown based on the provided text:


    Acceptance Criteria and Device Performance for Premium Plus Prophy Air Motor (K170261)

    The Premium Plus Prophy Air Motor is a Class I mechanical dental device. The "acceptance criteria" in this context are defined by compliance with recognized international standards for dental handpieces and by demonstrating substantial equivalence to a predicate device. The "performance" refers to the device meeting the specifications outlined in these standards and its functional characteristics.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/StandardReported Device Performance
    Indications for Use"The Premium Plus Prophy Air Motor is intended to be used with a disposable prophy angle and polishing paste by a trained dental clinician to perform dental prophylaxis." (Same as predicate)Met (Stated as same as predicate device: Prophy Star 3 Hygiene Handpiece K070869)
    Functional Performance & SafetyCompliance with ISO 14457:2012 (Dentistry Handpieces and Motors Test)"The test results confirm that Premium Plus Prophy Air Motor conforms to the requirements in ISO 14457:2012..." This standard likely covers aspects like mechanical safety, operating characteristics, and durability. The device also claims "Robust construction to withstand forces generated during cleaning cycle." It passed a performance test in accordance with ISO 14457.
    Compliance with ISO 9168:2009 (Dentistry Hose Connectors for Air Driven Dental Handpieces Test)"The test results confirm that Premium Plus Prophy Air Motor conforms to the requirements... ISO 9168:2009..." This standard ensures proper connection and compatibility with dental unit air hoses.
    Biocompatibility- Cytotoxicity per ISO 10993-5: 2009 (Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity)"Biocompatibility tests were conducted in accordance with the requirements of ISO 10993-5..." and stated as "Made of non-toxic anodized aluminum." The document implies the tests were successfully passed, leading to the "Same" conclusion in the comparison table.
    - Oral Irritation per ISO 10993-10: (Biological evaluation of medical device - Part 10: Tests for irritation and delayed -type hypersensitivity)Biocompatibility tests were conducted. Implied to have met criteria.
    - Sensitization per ISO 10993-10: (Biological evaluation of medical device - Part 10: Tests for irritation and delayed -type hypersensitivity)Biocompatibility tests were conducted. Implied to have met criteria.
    Reprocessing (Cleaning & Sterilization) ValidationRecommendations of FDA Guidance on "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" followed. Specific tests:"Validation Testings" were performed, including: Cleaning Validation Protein Analysis, Cleaning Validation - Hemoglobin Analysis, Sterilization Validation - Gravity cycle 132 ℃ 15min, Sterilization Validation - Gravity cycle 121 ℃ 30min, Sterilization Validation - Prevacuum cycle 132 ℃ 4min. The successful completion of these tests suggests validation criteria were met.
    Device Characteristics (Comparison to Predicate)- Operating Air Pressure: Close to predicate (Predicate: 35-40 psi; Subject: 200kPa ≈ 29 psi)Met (Difference noted but considered substantially equivalent for a Class I device).
    - Operating Speed: Within reasonable range of predicate (Predicate: 0 to 5,000 rpm; Subject: 0 to 2,480 rpm)Met (Difference noted, but "The subject device's operational speed is within the extent of the predicate device's speed." and deemed substantially equivalent).
    - Power Output: Close to predicate (Predicate: 5.8 Watts; Subject: 6.0 Watts)Met (Very similar).
    - Design, Materials (e.g., anodized aluminum), Compatibility with standard connectors (Mid-West style air hose, Doriot style prophy angles), Sterility (Non-sterile)All stated as "Same" or very similar to the predicate device, contributing to the claim of substantial equivalence. The slight differences in length/diameter/weight were noted but not considered to affect substantial equivalence for this device type.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated for each test (e.g., "Performance Test," "Biocompatibility Test," "Validation Testings"). These typically involve testing a representative sample of devices, but specific numbers are not provided in this summary. For Class I mechanical devices, the number of units tested is generally small as standardized tests are conducted on representative samples to ensure manufacturing quality and design integrity.
    • Data Provenance: The device manufacturer, Premium Plus International Limited, is located in Hong Kong, China. The testing itself is conducted to international standards (ISO) and FDA guidance, but the exact location of the testing laboratories (e.g., in Hong Kong, or a third-party lab elsewhere) is not specified. The tests are prospective in the sense that they are performed on the device being submitted for marketing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This concept is not applicable here. "Ground truth" established by experts is relevant for diagnostic or AI/ML devices where there is an interpretation or classification task. For a mechanical dental handpiece, the "ground truth" is determined by objective physical measurements and adherence to engineering standards.

    4. Adjudication Method for the Test Set

    This concept is not applicable. Adjudication methods are used in studies involving human interpretation or clinical outcomes, often for AI/ML devices or diagnostic tests. Here, the "tests" are compliance with engineering standards and validation protocols.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, Effect Size

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices, particularly AI/ML-driven ones, to assess how AI assistance impacts human reader performance. This device is a simple mechanical prophy air motor, not a diagnostic tool or an AI/ML product.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    No, a standalone performance study in the context of an "algorithm" was not done. This device does not contain an algorithm or AI component. Its performance is measured directly through mechanical, airflow, and material safety tests.

    7. The Type of Ground Truth Used

    As this is a mechanical device, the "ground truth" is established by:

    • Compliance with objective engineering standards (e.g., ISO 14457, ISO 9168).
    • Results from biocompatibility assays (e.g., cytotoxicity, irritation, sensitization tests per ISO 10993 standards) where "ground truth" is defined by the test's positive/negative control and specific criteria for material response.
    • Results from reprocessing validation tests (e.g., protein/hemoglobin analysis, sterilization cycles) designed to demonstrate effective cleaning and sterilization.
    • Direct physical and operational measurements (e.g., dimensions, weight, air pressure, operating speed, power output) compared against specifications and the predicate device.

    8. The Sample Size for the Training Set

    This concept is not applicable. This device is a mechanical product, not an AI/ML algorithm. There is no "training set" for an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    This concept is not applicable as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163447

    Validate with FDA (Live)

    Date Cleared
    2017-06-20

    (194 days)

    Product Code
    Regulation Number
    878.4370
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Premium Plus Disposable Barrier Sleeve is intended to be used as a disposable barrier for dental instruments and equipment. This device is non-sterile and intended for single patient use only.

    Device Description

    Premium Plus Disposable Barrier Sleeves are made of polyethylene film and are used as accessories to dental instruments and equipment used during dental procedures. These barrier sleeves consist of various sizes and shapes which are positioned on various small hand-held dental instruments such as handpieces, curing lights, air/water syringes and similar hand instruments. In other forms, they are used to cover various devices such as dental chairs, dental instrument travs, x-ray heads, and others. The products are sold nonsterile, prepackaged, and are disposable, single use only.

    AI/ML Overview

    The provided text is a 510(k) premarket notification from the FDA for a device called "Premium Plus Disposable Barrier Sleeve." This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving a medical device's performance against specific acceptance criteria through a study involving diagnostic accuracy, human readers, or AI.

    Therefore, the requested information about acceptance criteria for diagnostic performance, sample sizes for test/training sets, expert ground truth establishment, MRMC studies, or standalone algorithm performance, as would be relevant for an AI/ML-driven diagnostic device, cannot be found in this document.

    The document describes the physical and material characteristics of the barrier sleeves and confirms that they meet certain safety and performance standards for barrier efficacy, not diagnostic accuracy.

    Here's what can be extracted from the document regarding the device's "performance" in its specific context:


    Device: Premium Plus Disposable Barrier Sleeve
    Indications for Use: Intended to be used as a disposable barrier for dental instruments and equipment. Non-sterile and intended for single patient use only.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't present acceptance criteria in a quantitative diagnostic performance metric (e.g., sensitivity, specificity, AUC). Instead, it compares the new device to a predicate device ("Pac-Dent Barrier Sleeve") based on various physical and biological properties relevant to its function as a barrier. The "acceptance criteria" are implied by the "Comparison of Technological Characteristics and Performance" table, where the new device is stated to be "comparable" or "equivalent" to the predicate.

    Characteristic/TestAcceptance Criteria (Implied by Predicate)Reported Device Performance (Premium Plus Disposable Barrier Sleeve)
    Intended UseTo be used as a barrier to cover dental instrumentsTo be used as a barrier for dental instruments and equipment
    ClassificationPEM (Product Code)PEM (Product Code)
    MaterialPolyethylene filmPolyethylene film
    Material CompositionLow density polyethylene and linear low density polyethylene filmLow density polyethylene and linear low density polyethylene film
    BiocompatibilityNon-cytotoxic, Non-sensitizing, Non-irritatingNon-cytotoxic, Non-sensitizing, Non-irritating (Biocompatibility test result indicates the disposable barrier sleeves do not have the potential cytotoxicity, hypersensitivity and irritation.)
    Film Thickness0.02mm - 0.06mm0.03mm
    Mechanical PropertiesTensile Strength – tested in compliance with ASTM D882; Puncture Resistance - tested in compliance with ASTM F1342; Tear Resistance - tested in compliance with ASTM D1424Mechanical properties are comparable to the predicate device. Tensile Strength, puncture resistance and tear resistance are equivalent to the predicate device.
    SterilityNon-sterileNon-sterile
    Single UseSingle use deviceSingle use device
    Performance PropertiesSynthetic Blood Penetration - Pass; Viral Penetration - PassSynthetic Blood Penetration - Pass; Viral Penetration - Pass
    DimensionsDetermined by the size and shape of dental instruments and equipmentDifference in dimensions is due to the size of instruments and equipment they cover (device comes in many models for various instruments)
    FDA-Recognized StandardsASTM F1670, ASTM F1671, ASTM D882, ASTM F1342, ASTM D1004, ISO 10993-5, ISO 10993-10ASTM F1670, ASTM F1671, ASTM D882, ASTM F1342, ASTM D1004, ISO 10993-5, ISO 10993-10
    X-Ray Compatibility(Implied: not negatively impacting X-ray device function)Effectiveness of x-ray devices covered with Premium Plus Disposable Barrier Sleeves was performed to demonstrate the subject device does not impact the function of dental x-ray devices.

    2. Sample Size Used for the Test Set and Data Provenance

    This document does not describe studies that would have "test sets" in the context of diagnostic performance (e.g., patient data for an AI model). The "tests" mentioned are for material properties and barrier efficacy.

    • Sample Size: Not specified in terms of "test set" for diagnostic performance. For mechanical and barrier tests, samples of the material/product would have been tested according to relevant ASTM/ISO standards, but specific sample numbers are not provided.
    • Data Provenance: Not applicable in the context of patient data. The tests are laboratory-based material property tests.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. This is not a diagnostic device or an AI/ML device requiring expert ground truth for interpretation of medical images or patient data. The "ground truth" for this device relates to established material performance standards (e.g., a material either passes or fails a tear strength test as per an ASTM standard).

    4. Adjudication Method for the Test Set

    Not applicable. No expert adjudication process is described as it's not a diagnostic study.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No. An MRMC study is relevant for evaluating the impact of a diagnostic tool (often AI-assisted) on human readers' performance. This device is a physical barrier, not a diagnostic tool, so an MRMC study is not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    No. This device is not an algorithm or an AI system. Its performance relates to its physical properties as a barrier.

    7. The Type of Ground Truth Used

    The "ground truth" is based on:

    • Physical and Chemical Properties: Material composition (polyethylene film), thickness.
    • Biocompatibility Standards: ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization). These involve laboratory tests on cell cultures and animal models (though the document doesn't specify if human or animal data were used beyond stating "non-cytotoxic," etc.).
    • Mechanical and Barrier Performance Standards: ASTM D882 (tensile strength), ASTM F1342 (puncture resistance), ASTM D1424 (tear resistance), ASTM F1670 (synthetic blood penetration), ASTM F1671 (viral penetration). These are well-defined laboratory test methods.
    • Functional Testing: "Effectiveness of x-ray devices covered with Premium Plus Disposable Barrier Sleeves was performed to demonstrate the subject device does not impact the function of dental x-ray devices." This implies functional testing against a benchmark of an X-ray device operating without the barrier, but the specifics of this "ground truth" are not detailed.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. No training set is involved.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162028

    Validate with FDA (Live)

    Date Cleared
    2017-03-24

    (245 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Abutments:

    PREMIUM-SHELTA Conico Abutments are intended to be used in conjunction with a PREMIUM-SHELTA Implants Systems in fully edentulous or partially edentulous maxillary and/or mandibular arches.

    The PREMIUM-SHELTA Conico Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prostheses, in the mandible or maxilla. The prosthesis can be cemented, screw retained or friction fit the abutment screw is intended to secure the abutment to the endosseous implant.

    PREMIUM-SHELTA Conico Abutments are compatible with PREMIUM-SHELTA Implants Systems. Anchorage of dentures retained by taper friction and supported by PREMIUM-SHELTA Implants Systems.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for a dental prosthetic component. It confirms that the device, "Premium Shelta Prosthetic Components," is substantially equivalent to legally marketed predicate devices.

    However, this document does not contain any information regarding acceptance criteria, device performance studies, sample sizes, ground truth establishment, or expert qualifications. The letter is a regulatory approval, not a technical report on performance validation.

    Therefore, I cannot provide the requested information based on the given text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162004

    Validate with FDA (Live)

    Date Cleared
    2016-12-05

    (138 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PREMIUM Implant Systems -SHELTA Implant Systems are intended for both one- and two-stage surgical procedures. PREMIUM Implant Systems-SHELTA Implant Systems are intended for immediate placement and function on single tooth and/or multiple tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore chewing function. Multiple tooth applications may be splinted with a bar. Abutments: PREMIUM-SHELTA Abutments are intended to be used in conjunction with a PREMIUM-SHELTA Implants Systems in fully edentulous or partially edentulous maxillary and/or mandibular arches. The PREMIUM-SHELTA Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single tooth prostheses, in the mandible or maxilla. The prosthesis can be cemented, screw retained or friction fit to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant. PREMIUM-SHELTA Abutments are compatible with PREMIUM-SHELTA Implants Systems.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) premarket notification letter from the FDA regarding dental implants and abutments. It does not contain information about acceptance criteria, device performance studies, sample sizes, ground truth establishment, expert qualifications, or MRMC studies.

    Therefore, I cannot extract the requested information from this document. This document is primarily an FDA clearance letter, confirming that the device is substantially equivalent to a legally marketed predicate device. It does not detail the technical performance studies of the device beyond its stated indications for use.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153514

    Validate with FDA (Live)

    Date Cleared
    2016-08-22

    (259 days)

    Product Code
    Regulation Number
    872.6070
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Premium Plus C01/C02 Dental Curing Light is intended for use by trained dental professionals for the purpose of curing dental composites by light.

    Device Description

    Premium Plus C01/C02 LED Curing Light is classified as an Ultraviolet Activator for Polymerization (21 CFR 872.6070) because it is used for polymerization of dental light cured materials by dental professionals. The device is based on blue LED (light emitting diode) technology. It has three curing modes. Using different modes gives dental professionals the flexibility to polymerize virtually almost all types of composites, boding agents and sealants available in the market.

    Premium Plus C01/C02 LED Curing Light consists of a handpiece and a charging station, which is connected via a plug-in transformer to an AC outlet for charging. The handpiece contains high intensity dental blue LED light source and a fibre optic light guide(C01-1) that conduct light to the treatment area on the patients or direct light source head(C02-1). A protective light shield and transformer are accessories provided with Premium Plus C01/C02 LED Curing Light, which support the operation of the device. The protective light shield filters blue light to protect eyes of dental professionals and patients. An FDA cleared barrier sleeves must be used between each patient such as the Pac-Dent Barrier Sleeve.

    AI/ML Overview

    This document is a 510(k) premarket notification for a dental curing light, the Premium Plus C01/C02 LED Curing Light. It aims to establish substantial equivalence to existing predicate devices. The information provided focuses on technical specifications and safety standards rather than clinical performance or AI algorithm validation.

    Therefore, many of the requested details, such as acceptance criteria based on clinical metrics (e.g., sensitivity, specificity), sample size for test sets, ground truth establishment by experts, adjudication methods, or effects of AI assistance, are not applicable or not available in this document.

    The document primarily relies on comparisons of technical characteristics and adherence to recognized safety and performance standards to demonstrate substantial equivalence.

    Here's a breakdown of the available and unavailable information based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in terms of clinical performance metrics (like sensitivity, specificity, or human improvement with AI). Instead, it relies on demonstrating that the device meets recognized industry standards and performs comparably to predicate devices based on technical specifications and safety tests.

    Therefore, instead of a clinical performance table, below are the relevant technical and safety performance criteria that were met:

    Acceptance Criterion (Standard/Predicate Feature)Reported Device Performance (Premium Plus C01/C02)
    Electrical SafetyConforms to IEC 60601-1
    Electromagnetic CompatibilityConforms to IEC 60601-1-2
    LED Lamps PerformanceConforms to ISO 10650-2 / ANSI/ADA Specification No. 48-2 (LED Curing Lights)
    Depth of Cure"Depth of Cure" test conducted, results confirm conformance to requirements in ISO 10650-2 / ANSI/ADA Specification No. 48-2 (implied)
    Intended Use (vs. Predicate)Same as predicate devices (Ledex WL-070, Coltolux® LED Curing Light)
    Light Source (vs. Predicate)Same (LED light)
    Wavelength Range (C01 vs. Ledex WL-070)Same (440nm~480nm, peak: 460nm)
    Radiant Intensity (C01 vs. Ledex WL-070)Within comparable range (750 mw/cm²- 1200 mw/cm² vs. 1100 mw/cm²- 1200 mw/cm²)
    Radiant Intensity (C02 vs. Coltolux®)Meets or exceeds (750 mw/cm²- 1200 mw/cm² vs. 300 mw/cm² or higher)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified for any clinical or performance comparison study. The document refers to "Performance Tests" for Electrical Safety, EMI, LED lamps, and Depth of Cure, but the number of units or materials tested is not mentioned.
    • Data Provenance: Not applicable as there's no patient data or clinical imagery discussed. All testing appears to be laboratory-based and related to device specifications and safety.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable. The "ground truth" here is adherence to technical standards and specifications, not clinical diagnoses or interpretations. These standards are established by regulatory bodies and consensus organizations, not individual experts in a test set.

    4. Adjudication Method for the Test Set

    Not applicable. There's no interpretive task that would require adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a dental curing light, not an AI-powered diagnostic or assistive tool.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device does not involve an algorithm for standalone performance.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's approval is based on:

    • Compliance with recognized international and national standards: IEC 60601-1 (Electrical Safety), IEC 60601-1-2 (Electromagnetic Compatibility), ISO 10650-2 / ANSI/ADA Specification No. 48-2 (LED Curing Lights).
    • Demonstrated technical equivalence in key specifications (intended use, light source, wavelength, radiant intensity, operational modes, power, environmental factors, safety standards) to already approved predicate devices.

    8. The Sample Size for the Training Set

    Not applicable. There is no training set mentioned, as this is not an AI/machine learning device.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. There is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2