K Number
K163447

Validate with FDA (Live)

Date Cleared
2017-06-20

(194 days)

Product Code
Regulation Number
878.4370
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Premium Plus Disposable Barrier Sleeve is intended to be used as a disposable barrier for dental instruments and equipment. This device is non-sterile and intended for single patient use only.

Device Description

Premium Plus Disposable Barrier Sleeves are made of polyethylene film and are used as accessories to dental instruments and equipment used during dental procedures. These barrier sleeves consist of various sizes and shapes which are positioned on various small hand-held dental instruments such as handpieces, curing lights, air/water syringes and similar hand instruments. In other forms, they are used to cover various devices such as dental chairs, dental instrument travs, x-ray heads, and others. The products are sold nonsterile, prepackaged, and are disposable, single use only.

AI/ML Overview

The provided text is a 510(k) premarket notification from the FDA for a device called "Premium Plus Disposable Barrier Sleeve." This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving a medical device's performance against specific acceptance criteria through a study involving diagnostic accuracy, human readers, or AI.

Therefore, the requested information about acceptance criteria for diagnostic performance, sample sizes for test/training sets, expert ground truth establishment, MRMC studies, or standalone algorithm performance, as would be relevant for an AI/ML-driven diagnostic device, cannot be found in this document.

The document describes the physical and material characteristics of the barrier sleeves and confirms that they meet certain safety and performance standards for barrier efficacy, not diagnostic accuracy.

Here's what can be extracted from the document regarding the device's "performance" in its specific context:


Device: Premium Plus Disposable Barrier Sleeve
Indications for Use: Intended to be used as a disposable barrier for dental instruments and equipment. Non-sterile and intended for single patient use only.

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't present acceptance criteria in a quantitative diagnostic performance metric (e.g., sensitivity, specificity, AUC). Instead, it compares the new device to a predicate device ("Pac-Dent Barrier Sleeve") based on various physical and biological properties relevant to its function as a barrier. The "acceptance criteria" are implied by the "Comparison of Technological Characteristics and Performance" table, where the new device is stated to be "comparable" or "equivalent" to the predicate.

Characteristic/TestAcceptance Criteria (Implied by Predicate)Reported Device Performance (Premium Plus Disposable Barrier Sleeve)
Intended UseTo be used as a barrier to cover dental instrumentsTo be used as a barrier for dental instruments and equipment
ClassificationPEM (Product Code)PEM (Product Code)
MaterialPolyethylene filmPolyethylene film
Material CompositionLow density polyethylene and linear low density polyethylene filmLow density polyethylene and linear low density polyethylene film
BiocompatibilityNon-cytotoxic, Non-sensitizing, Non-irritatingNon-cytotoxic, Non-sensitizing, Non-irritating (Biocompatibility test result indicates the disposable barrier sleeves do not have the potential cytotoxicity, hypersensitivity and irritation.)
Film Thickness0.02mm - 0.06mm0.03mm
Mechanical PropertiesTensile Strength – tested in compliance with ASTM D882; Puncture Resistance - tested in compliance with ASTM F1342; Tear Resistance - tested in compliance with ASTM D1424Mechanical properties are comparable to the predicate device. Tensile Strength, puncture resistance and tear resistance are equivalent to the predicate device.
SterilityNon-sterileNon-sterile
Single UseSingle use deviceSingle use device
Performance PropertiesSynthetic Blood Penetration - Pass; Viral Penetration - PassSynthetic Blood Penetration - Pass; Viral Penetration - Pass
DimensionsDetermined by the size and shape of dental instruments and equipmentDifference in dimensions is due to the size of instruments and equipment they cover (device comes in many models for various instruments)
FDA-Recognized StandardsASTM F1670, ASTM F1671, ASTM D882, ASTM F1342, ASTM D1004, ISO 10993-5, ISO 10993-10ASTM F1670, ASTM F1671, ASTM D882, ASTM F1342, ASTM D1004, ISO 10993-5, ISO 10993-10
X-Ray Compatibility(Implied: not negatively impacting X-ray device function)Effectiveness of x-ray devices covered with Premium Plus Disposable Barrier Sleeves was performed to demonstrate the subject device does not impact the function of dental x-ray devices.

2. Sample Size Used for the Test Set and Data Provenance

This document does not describe studies that would have "test sets" in the context of diagnostic performance (e.g., patient data for an AI model). The "tests" mentioned are for material properties and barrier efficacy.

  • Sample Size: Not specified in terms of "test set" for diagnostic performance. For mechanical and barrier tests, samples of the material/product would have been tested according to relevant ASTM/ISO standards, but specific sample numbers are not provided.
  • Data Provenance: Not applicable in the context of patient data. The tests are laboratory-based material property tests.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

Not applicable. This is not a diagnostic device or an AI/ML device requiring expert ground truth for interpretation of medical images or patient data. The "ground truth" for this device relates to established material performance standards (e.g., a material either passes or fails a tear strength test as per an ASTM standard).

4. Adjudication Method for the Test Set

Not applicable. No expert adjudication process is described as it's not a diagnostic study.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No. An MRMC study is relevant for evaluating the impact of a diagnostic tool (often AI-assisted) on human readers' performance. This device is a physical barrier, not a diagnostic tool, so an MRMC study is not applicable.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

No. This device is not an algorithm or an AI system. Its performance relates to its physical properties as a barrier.

7. The Type of Ground Truth Used

The "ground truth" is based on:

  • Physical and Chemical Properties: Material composition (polyethylene film), thickness.
  • Biocompatibility Standards: ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization). These involve laboratory tests on cell cultures and animal models (though the document doesn't specify if human or animal data were used beyond stating "non-cytotoxic," etc.).
  • Mechanical and Barrier Performance Standards: ASTM D882 (tensile strength), ASTM F1342 (puncture resistance), ASTM D1424 (tear resistance), ASTM F1670 (synthetic blood penetration), ASTM F1671 (viral penetration). These are well-defined laboratory test methods.
  • Functional Testing: "Effectiveness of x-ray devices covered with Premium Plus Disposable Barrier Sleeves was performed to demonstrate the subject device does not impact the function of dental x-ray devices." This implies functional testing against a benchmark of an X-ray device operating without the barrier, but the specifics of this "ground truth" are not detailed.

8. The Sample Size for the Training Set

Not applicable. This is not an AI/ML device that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable. No training set is involved.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it, overlaid on three human profiles facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 20, 2017

Premium Plus International Limited Jessica Mao OA Engineer Flat 1001 Yuen Long Trading Centre, 33 Wang Yip Street West Yuen Long N.t. Hong Kong, CN

Re: K163447

Trade/Device Name: Premium Plus Disposable Barrier Sleeve Regulation Number: 21 CFR 878.4370 Regulation Name: Surgical Drape And Drape Accessories Regulatory Class: Class II Product Code: PEM Dated: November 30, 2016 Received: December 8, 2016

Dear Jessica Mao:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Lori A. Wiggins -S6

Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K163447

Device Name

Premium Plus Disposable Barrier Sleeve

Indications for Use (Describe)

Premium Plus Disposable Barrier Sleeve is intended to be used as a disposable barrier for dental instruments and equipment. This device is non-sterile and intended for single patient use only.

Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Flat 1001 Yven Long Trading Centre, No. 33 Wang Yip Street West, Yven Long, HONG KONG Phone: (852) 2761 3211, Facsimile: (852) 2761 3210, Email: jessicamao@premiumplushk.com

Section 5 510(k) Summary [As Required by 21 CFR 807.92]

510(k) Number: K163447

Premium Plus International Limited Submitter: Flat 1001 Yuen Long Trading Centre, No. 33 Wang Yip Street West, Yuen Long, N.T. Hong Kong, China (Establishment registration number: 3006847937)

Contact Person: Jessica Mao QA Engineer Tel: 852-27613211 Fax: 852-27613210 Email: jessicamao@premiumplushk.com

Trade Name: Premium Plus Disposable Barrier Sleeve Common Name: Dental Barriers and Sleeves Device Panel: General and Plastic Surgery Basis for Submission: New Device Regulation Name: Surgical Drape and Drape Accessories Device Classification: Class II Regulation Number: 21 CFR 878.4370 Product Code: PEM

Predicate Device to Premium Plus Disposable Barrier Sleeves:

Trade Name: Pac-Dent Barrier Sleeve 510(k) Number: K151123 Manufactured by: Pac-Dent International, Inc.

Device Description:

Premium Plus Disposable Barrier Sleeves are made of polyethylene film and are used as accessories to dental instruments and equipment used during dental procedures. These barrier sleeves consist of various sizes and shapes which are positioned on various small hand-held dental instruments such as handpieces, curing lights, air/water syringes and similar hand instruments. In other forms, they are used to cover various devices such as dental chairs, dental instrument travs, x-ray heads, and others. The products are sold nonsterile, prepackaged, and are disposable, single use only.

Indications for Use:

Premium Plus Disposable Barrier Sleeves are intended to be used as a disposable barrier for dental instruments and equipment. This device is non-sterile and intended for single patient use only.

Page 1 of 7

{4}------------------------------------------------

Flat 1001 Yuen Long Trading Centre, No. 33 Wang Yip Street West, Yuen Long, HONG KONG
Phone: (852) 2761 3211, Facsimile: (852) 2761 3210, Email: jessicamao@premiumplushk.com

Device Models:

Model#DescriptionDesigned for
100ASyringe Sleeves with AdhesiveTape, 500 pcs/box3-way syringes, saliva ejectors and HVE valve
100BSyringe Sleeves Blue, 500pcs/box3-way syringes, saliva ejectors and HVE valve
100CSyringe Sleeves Clear, 500pcs/box3-way syringes, saliva ejectors and HVE valve
101Tray Sleeves Standard 'B', 500pcs/boxInstrument trays
102Tray Sleeves Mini 'F', 500pcs/boxInstrument trays
103High Speed / Pen Sleeves, 500pcs/boxMost handpieces, jet polishers and ultrasonicscalers
104Low Speed / Universal Sleeves,500 pcs/boxMost handpieces, jet polishers andultrasonicscalers
105Curing Light Handle Sleeves, 500pcs/boxCuring light, pistol type
106T Light Handle Sleeves, U-shape,500 pcs/boxMost T-style dental chair light handles
107Impression Gun Sleeves, 500pcs/boxImpression guns
108X-Ray Head / Keyboard Sleeves,250 pcs/boxX-ray head/keyboard
109X-Ray Head Sleeves Large, 250pcs/boxX-ray head
111Full Chair Sleeves, 125 pcs/boxDental chairs
112Half Chair / Stool Sleeves, 125pcs/boxDental chairs/stools
113Headrest Sleeves Small, 500pcs/boxDental chair headrest
114Headrest Sleeves Large, 500pcs/boxDental chair headrest
118Straight Tubing Sleeves, 366metersDental chair tubing
119Coiled Tubing Sleeves, 366metersDental chair tubing
123LCuring Light Sleeves Pen TypeLarge, 500 pcs/boxCuring lights, pen type
123SCuring Light Sleeves Pen TypeSmall, 500 pcs/boxCuring lights, pen type
123SSCuring Light Sleeves Pen TypeSmall, 500 pcs/boxCuring lights, pen type

{5}------------------------------------------------

Flat 1001 Yuen Long Trading Centre, No. 33 Wang Yip Street West, Yuen Long, HONG KONG

Phone: (852) 2761 3211, Facsimile: (852) 2761 3210, Email: jessicamao@premiumplushk.com
Phone: (852) 2761 3211, Facsimile: (852) 2761 3210, Email: jessicamao@premiumplushk.com
125Intraoral Camera Sleeves, 500 pcs/boxIntraoral Cameras
125SSIntraoral Camera/Curing Light Sleeves, 500 pcs/boxIntraoral Cameras/curing lights
128Light Guide Sleeves, Small, 200 pcs/boxCuring light guides
129Light Guide Sleeves, Large, 200 pcs/boxCuring light guides
133PC Mouse Sleeves, 500 pcs/boxPC mouses
135HVE Suction Tube Sleeves, 500 pcs/boxHVE suction tubes
136T Light Handle Sleeves, Rectangular, 500 pcs/boxT-style dental chair light handles
138Flowable Composite Syringe Sleeves, 200 pcs/boxComposite syringes
140LX-Ray Sensor Sleeves, Large, 500 pcs/boxX-ray sensors
140MX-Ray Sensor Sleeves, Medium, 500 pcs/boxX-ray sensors
140SX-Ray Sensor Sleeves, Small, 500 pcs/boxX-ray sensors
140XSX-Ray Sensor/Mouth Mirror Sleeves, 500 pcs/boxX-ray sensors/mouth mirrors
145Bite Block Sleeves, 1000 pcs/boxBite Block
155Tablet Sleeves, 500pcs/boxTablet/Pad
158Keyboard Sleeves, 500 pcs/boxKeyboards
183-1X-Ray Sensor Sleeves Two Step, Size 1, 500X-ray sensors
183-2X-Ray Sensor Sleeves Two Step, Size 2, 500X-ray sensors
190ISO Tray Sleeves, 500 pcs/boxInstrument Trays
105TCuring Light Sleeves Complete, 500 pcs/boxCuring lights
121Lab Transport Bags, 100 pcs/boxTransport of dental materials
188-0PSP X-Ray Barrier Envelopes, No. 0, 100 pcs/boxPhosphor Plates
188-1PSP X-Ray Barrier Envelopes, No. 1, 100 pcs/boxPhosphor Plates
188-2PSP X-Ray Barrier Envelopes, No. 2, 100 pcs/boxPhosphor Plates
188-2BPSP X-Ray Barrier Envelopes, No. 2, 500 pcs/boxPhosphor Plates
188-3PSP X-Ray Barrier Envelopes, No. 3, 50 pcs/boxPhosphor Plates

{6}------------------------------------------------

Flat 1001 Yuen Long Trading Centre, No. 33 Wang Yip Street West, Yuen Long, HONG KONG

Phone: (852) 2761 3211, Facsimile: (852) 2761 3210, Email: jessicamao@premiumplushk.com
189-0No. 0, 100 pcs/boxPSP X-Ray Barrier Envelopes,Phosphor Plates
189-1No. 1, 100 pcs/boxPSP X-Ray Barrier Envelopes,Phosphor Plates
189-2No. 2, 100 pcs/boxPSP X-Ray Barrier Envelopes,Phosphor Plates
198-0No. 0, 100 pcs/boxPSP X-Ray Barrier Envelopes,Phosphor Plates
198-2No. 2, 100 pcs/boxPSP X-Ray Barrier Envelopes,Phosphor Plates
198-2BNo. 2, 500 pcs/boxPSP X-Ray Barrier Envelopes,Phosphor Plates
199-0No. 0, 100 pcs/boxPSP X-Ray Barrier Envelopes,Phosphor Plates
199-1No. 1, 100 pcs/boxPSP X-Ray Barrier Envelopes,Phosphor Plates
199-2No. 2, 100 pcs/boxPSP X-Ray Barrier Envelopes,Phosphor Plates
8008-Bsheets/rollBarrier Film 4" x 6", Clear, 1200Surface area that might be touchedduring procedure
8008-Csheets/rollBarrier Film 4" x 6", Blue, 1200Surface area that might be touchedduring procedure

Overall Dimensions (by Inches):

Model#Device DescriptionSize
100ASyringe Sleeves with Adhesive Tape, 500 pcs/box10 15/64"x2 9/16"
100BSyringe Sleeves Blue, 500 pcs/box10 15/64"x2 9/16"
100CSyringe Sleeves Clear, 500 pcs/box10 15/64"x2 9/16"
101Tray Sleeves Standard 'B', 500 pcs/box14 11/64"x10 5/8"
102Tray Sleeves Mini 'F', 500 pcs/box11 1/32"x7 31/64"
103High Speed / Pen Sleeves, 500 pcs/box8 21/32"x15/16"
104Low Speed / Universal Sleeves, 5008 21/32"x1 1/2"
105Curing Light Handle Sleeves, 500 pcs/box10 19/32"x5 19/32"
106T Light Handle Sleeves, U-shape, 500 pcs/box5 29/32"x3 11/32"
107Impression Gun Sleeves, 500 pcs/box8 17/64"x9 27/32"
108X-Ray Head / Keyboard Sleeves, 250 pcs/box23 5/8"x7 7/8"
109X-Ray Head Sleeves Large, 250 pcs/box29 17/32"x13 25/32"
111Full Chair Sleeves, 125 pcs/box70 55/64"x28 11/32"
112Half Chair / Stool Sleeves, 125 pcs/box39 3/8"x25 19/32"
113Headrest Sleeves Small, 500 pcs/box10 15/64"x10 53/64"
114Headrest Sleeves Large, 500 pcs/box11 1/32"x13 25/32"
118Straight Tubing Sleeves, 366 meters11 13/16"x1 31/32"
119Coiled Tubing Sleeves, 366 meters11 13/16"x3 15/16"

{7}------------------------------------------------

Flat 1001 Yuen Long Trading Centre, No. 33 Wang Yip Street West, Yuen Long, HONG KONG

rang rade reduced reduction and trainer respendent reduction comment coment coment coment ordered.
Phone: (852) 2761 3211, Facsimile: (852) 2761 3210, Email: jessicamao@premiumplushk.com
123LCuring Light Sleeves Pen Type Large, 500pcs/box13 25/32"x2 9/16"
123SCuring Light Sleeves Pen Type Small, 500pcs/box12 19/32"x1 31/32"
123SSCuring Light Sleeves Pen Type X-Small, 500pcs/box11 1/32"x1 31/32"
125Intraoral Camera Sleeves, 500 pcs/box12 19/32"x1 57/64"
125SSIntraoral Camera/Curing Light Sleeves, 500pcs/box11 1/32"x1 57/64"
128Light Guide Sleeves, Small, 200 pcs/box3 15/32"x3 11/32"
129Light Guide Sleeves, Large, 200 pcs/box3 15/16"x3 35/64"
133PC Mouse Sleeves, 500 pcs/box5 29/32"x3 11/32"
135HVE Suction Tube Sleeves, 500 pcs/box2 11/64"x13 25/32"
136T Light Handle Sleeves, Rectangular, 500 pcs/box5 29/32"x3 15/16"
138Flowable Composite Syringe Sleeves, 200 pcs/box4 59/64"x1 31/32"
140LX-Ray Sensor Sleeves, Large, 500 pcs/box9 27/32"x2 23/64"
140MX-Ray Sensor Sleeves, Medium, 500 pcs/box7 7/8"x1 49/64"
140SX-Ray Sensor Sleeves, Small, 500 pcs/box7 7/8"x1 21/32"
140XSX-Ray Sensor/Mouth Mirror Sleeves, 500 pcs/box7 7/8"x1 1/2"
145Bite Block Sleeves, 1000 pcs/box2 11/64"x1 1/32"
155Tablet Sleeves, 500pcs/box7 7/8"x11 39/64"
158Keyboard Sleeves, 500 pcs/box6 19/64"x21 21/32"
183-1X-Ray Sensor Sleeves Two Step, Size 1, 5008 21/32"x1 31/32"
183-2X-Ray Sensor Sleeves Two Step, Size 2, 5008 21/32"x2 11/64"
190ISO Tray Sleeves, 500 pcs/box12 3/16"x8 1/4"
105TCuring Light Sleeves Complete, 500 pcs/box10 19/32"x5 19/32"
121Lab Transport Bags, 100 pcs/box8 21/32"x7 3/32"
188-0PSP X-Ray Barrier Envelopes, No. 0, 100 pcs/box1 27/64"x1 49/64"
188-1PSP X-Ray Barrier Envelopes, No. 1, 100 pcs/box1 31/32"x1 29/64"
188-2PSP X-Ray Barrier Envelopes, No. 2, 100 pcs/box2 1/64"x1 11/16"
188-2BPSP X-Ray Barrier Envelopes, No. 2, 500 pcs/box2 1/64"x1 11/16"
188-3PSP X-Ray Barrier Envelopes, No. 3, 50 pcs/box2 33/64"x1 1/2"
188-4PSP X-Ray Barrier Envelopes, No. 4, 50 pcs/box3 15/32"x2 43/64"
189-0PSP X-Ray Barrier Envelopes, No. 0, 100 pcs/box1 1/2"x1 3/8"
189-1PSP X-Ray Barrier Envelopes, No. 1, 100 pcs/box1 3/16"x2 3/32"
189-2PSP X-Ray Barrier Envelopes, No. 2, 100 pcs/box2 3/32"x1 29/64"
198-0PSP X-Ray Barrier Envelopes, No. 0, 100 pcs/box1 57/64"x1 7/32"
198-2PSP X-Ray Barrier Envelopes, No. 2, 100 pcs/box2 1/8"x1 1/2"
198-2BPSP X-Ray Barrier Envelopes, No. 2, 500 pcs/box2 1/8"x1 1/2"
199-0PSP X-Ray Barrier Envelopes, No. 0, 100 pcs/box1 47/64"x1 3/8"
199-1PSP X-Ray Barrier Envelopes, No. 1, 100 pcs/box2 3/64"x1 29/64"
199-2PSP X-Ray Barrier Envelopes, No. 2, 100 pcs/box2 1/8"x1 47/64"

{8}------------------------------------------------

Flat 1001 Yuen Long Trading Centre, No. 33 Wang Yip Street West, Yuen Long, HONG KONG
Phone: (852) 2761 3211, Facsimile: (852) 2761 3210, Email: jessicamao@premiumplushk.com

8008-BBarrier Film 4" x 6", Clear, 1200 sheets/roll4"x6"
8008-CBarrier Film 4" x 6", Blue, 1200 sheets/roll4"x6"

Comparison of Technological Characteristics and Performance:

Comparison of Technological Characteristics and Performance:
DevicePac-Dent Barrier Sleeve510(K) Number: K151123Premium Plus DisposableBarrier Sleeves
Intended UseTo be used as a barrier to coverdental instrumentsTo be used as a barrier fordental instruments andequipment
ClassificationProduct CodePEMPEM
MaterialPolyethylene filmPolyethylene film
Material CompositionLow density polyethylene andlinear low density polyethylenefilmLow density polyethylene andlinear low density polyethylenefilm
BiocompatibilityNon-cytotoxicNon-sensitizingNon-irritatingNon-cytotoxicNon-sensitizingNon-irritating
Film Thickness0.02mm- 0.06mm0.03mm
MechanicalProperties(Pac-Dent101 B Tray Sleevewas tested forcomparison with theproposed device)Tensile Strength – tested incompliance with ASTM D882Puncture Resistance - tested incompliance with ASTM F1342Tear Resistance - tested incompliance with ASTM D1424Mechanical properties of thesubject device are comparableto the predicate device.Tensile Strength, punctureresistance and tear resistanceare equivalent to the predicatedevice.
SterilityNon-sterileNon-sterile
Single UseSingle use deviceSingle use device
PerformancePropertiesSynthetic Blood Penetration-PassViral Penetration-PassSynthetic Blood Penetration-PassViral Penetration-Pass
DimensionsDetermined by the size andshape of dental instruments andequipmentDifference in dimensions is dueto the size of instruments andequipment they cover
FDA- RecognizedStandardsASTM F1670ASTM F1671ASTM D882ASTM F1342ASTM D1004ISO 10993-5ISO 10993-10ASTM F1670ASTM F1671ASTM D882ASTM F1342ASTM D1004ISO 10993-5ISO 10993-10

{9}------------------------------------------------

Flat 1001 Yuen Long Trading Centre, No. 33 Wang Yip Street West, Yuen Long, HONG KONG Phone: (852) 2761 3211, Facsimile: (852) 2761 3210, Email: jessicamao@premiumplushk.com

Tests:

Performance test including resistance to penetration, tear strength, tensile properties, resistance to puncture was performed. Biocompatibility test in accordance with ISO 10993-5, ISO 10993-10 has been performed. Effectiveness of x-ray devices covered with Premium Plus Disposable Barrier Sleeves was performed to demonstrate the subject device does not impact the function of dental x-ray devices.

The biocompatibility test result indicates that the disposable barrier sleeves do not have the potential cytotoxicity, hypersensitivity and irritation.

Clinical Performance Data:

No data from human clinical studies have been included to support the substantial equivalence of the proposed devices, Premium Plus International Limited, as clinical studies is not required for this medical device.

Conclusion

Based on the tests performed the subject device Premium Plus Disposable Barrier Sleeve is as safe and as effective, and performs as well as the legally marketed predicate device.

§ 878.4370 Surgical drape and drape accessories.

(a)
Identification. A surgical drape and drape accessories is a device made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. The device includes a plastic wound protector that may adhere to the skin around a surgical incision or be placed in a wound to cover its exposed edges, and a latex drape with a self-retaining finger cot that is intended to allow repeated insertion of the surgeon's finger into the rectum during performance of a transurethral prostatectomy.(b)
Classification. Class II (special controls). The device, when it is an ear, nose, and throat surgical drape, a latex sheet drape with self-retaining finger cot, a disposable urological drape, a Kelly pad, an ophthalmic patient drape, an ophthalmic microscope drape, an internal drape retention ring (wound protector), or a surgical drape that does not include an antimicrobial agent, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.