510(k) Data Aggregation

    K Number
    K090778

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-04-28

    (36 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MTA Protective Sheet is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue. MTA Protective Sheet minimizes tissue attachment to the device in case of direct contact with the tissues. The device is indicated for open and endoscopic procedures. MTA Protective Sheet is intended for one-time use.

    Device Description

    MTA Protective Sheet is composed of microbial-derived cellulose. The non-resorbable surgical mesh is used for the management and protection of tendon injuries. MTA Protective Sheet minimizes tissue attachment to the device in case of direct contact with the tissues. The implantable device is presented in a sterile double-pouched package for appropriate removal in preparation for surgery.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    BiocompatibilityBiologically and chemically identical to Securian Tissue Reinforcement Matrix (K083823) and very similarly manufactured.
    Biomechanical EquivalenceMet test requirements in material property and suture pull-out testing, demonstrating biomechanical equivalence to MAST Biosurgery Ortho-Wrap Bioresorbable Sheet.
    Intended UseProvides reasonable assurance of device performance for its intended use.
    Minimizes Tissue AttachmentMinimizes tissue attachment to the device in case of direct contact with the tissues (stated in device description and indications).
    Management & Protection of Tendon InjuriesIndicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.
    Open and Endoscopic ProceduresIndicated for open and endoscopic procedures.
    Single UseIntended for one-time use.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document describes "various performance tests typical for its product code such as material property and suture pull out testing." However, it does not specify the sample size used for these tests.

    The data provenance is from bench testing, meaning it was conducted in a laboratory setting, not on human or animal subjects, and is neither retrospective nor prospective in the typical clinical sense.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    This information is not provided in the document. The studies conducted were bench tests on material properties, not clinical studies requiring expert consensus for ground truth.

    4. Adjudication Method for the Test Set:

    An adjudication method is not applicable as the studies were bench tests focused on material properties and biomechanical equivalence, not on interpreting images or clinical outcomes that typically require adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No MRMC comparative effectiveness study was mentioned. The document primarily focuses on demonstrating substantial equivalence through material and biomechanical testing against predicate devices.

    6. Standalone (Algorithm Only) Performance:

    Not applicable. This device is a physical surgical mesh, not a software algorithm, so the concept of standalone performance does not apply in this context.

    7. Type of Ground Truth Used:

    The ground truth for the performance testing was based on objective measurements from material property and suture pull-out tests, and comparison against the known characteristics and performance of the predicate devices (Securian Tissue Reinforcement Matrix for biocompatibility, and MAST Biosurgery Ortho-Wrap Bioresorbable Sheet for biomechanical properties).

    8. Sample Size for the Training Set:

    The document does not mention a "training set" as this is not a machine learning or AI-based device. The testing described is for performance validation, not for training an algorithm.

    9. How the Ground Truth for the Training Set Was Established:

    As there is no mention of a "training set," this information is not applicable.