Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K231455

    Validate with FDA (Live)

    Date Cleared
    2023-08-15

    (89 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Implant-One™ System Abutments are intended for use as an aid in prosthetic rehabilitation in the mandible or maxilla for support of single-unit or multi-unit restorations.

    The Implant-One Ti-Base Abutment consists of the titanium base and a mesostructure component, making up a two-piece abutment, and will be attached to the implant using an abutment screw. The mesostructure for use with the Implant-One Ti-Base is intended only to be designed and manufactured according to digital dentistry workflow that integrates scan files from lab scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories.

    Device Description

    The subject devices comprise abutments designed for the 300, 400, and 500 Series of the Implant-One" system. All subject device abutments incorporate a Morse taper at the implant/abutment interface, have a hexagonal male end for alignment purposes and are screw retained. The series is grouped according to the implant/abutment interface size and each series is color coded for ease of identification. The subject devices are Titanium Base Abutments.

    The Implant-One™ Titanium Base Abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation. The Implant-One" Titanium Base Abutments consist of two major components. Specifically, the titanium base and the mesostructured component make up a two-piece abutment. The system integrates multiple components of the digital dentistry workflow: scan files from scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling.

    The standard Ti Base Abutments are offered in cuff heights of 0.5mm, 1.5mm and have a Ø4.0mm profile for the 300 series and a Ø4.5mm profile for the 400 & 500 series.

    The following components are intended to be used for scanning, designing and manufacturing of ceramic mesostructures for the Implant-One™ Titanium Base Abutment:

    • Lab scanner: Medit/Identica T500 3D Scanner (Medit Corp) ●
    • CAD software (PNP): 3Shape Abutment Designer & Abutment Design (3Shape A/S -● K151455 & K200100) and exocad AbutmentCAD (exocad GmbH - K193352)
    • Ceramic material: Katana™ Zirconia (Kuraray Noritake Dental, Inc. K143439) ●
    • . Milling machine: Ceramill Motion 2 (Amann-Girrbach AG)
    • Milling software: Ceramill Match 2 CAM Software (Amann Girrbach AG)
    • Cement: RelyX™ Luting Plus Automix Cement (3M ESPE Dental Products K11185) ●
    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for a dental implant abutment system (Implant-One™ System). It focuses on demonstrating substantial equivalence to a predicate device rather than presenting an AI/ML-based device performance study.

    Therefore, the document does not contain the information requested regarding acceptance criteria and performance data for an AI/ML device, as it describes a traditional medical device submission.

    Specifically, the following points from your request cannot be addressed based on the provided text:

    • A table of acceptance criteria and the reported device performance: This document does not describe acceptance criteria for an AI/ML system or statistical performance metrics. Instead, it describes mechanical testing (ISO 14801) and sterilization validation (ISO 17665-1 and ISO 17665-2) for a physical medical device.
    • Sample size used for the test set and the data provenance: There is no "test set" in the context of AI/ML evaluation. The performance data refers to mechanical and sterilization tests, not data for an algorithm.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Ground truth establishment is irrelevant in this context, as there is no AI/ML algorithm being evaluated.
    • Adjudication method: Not applicable due to the absence of an AI/ML performance study.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done: The document explicitly states "No clinical data were used in support of this submission," meaning no MRMC study was performed.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is not an AI/ML device.
    • The type of ground truth used: Not applicable.
    • The sample size for the training set: Not applicable, as this is not an AI/ML device.
    • How the ground truth for the training set was established: Not applicable, as this is not an AI/ML device.

    The document primarily focuses on demonstrating that the Implant-One™ System is substantially equivalent to a legally marketed predicate device (NobelActive™ Universal Base Abutments) by comparing their indications for use, technological characteristics, materials, dimensions, and sterilization methods. It mentions "Software verification and validation testing was conducted to demonstrate that the restrictions prevent design of the mesostructure component outside of design limitations," which refers to CAD/CAM software used in the manufacturing process, not an AI/ML diagnostic or prognostic algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173701

    Validate with FDA (Live)

    Date Cleared
    2018-11-29

    (360 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Implant-One™ System is indicated for surgical placement in partially or completely edentulous upper or lower jaws to provide a means for prosthetic attachment to restore a patient's chewing function. The Implant-One™ system is indicated for immediate loading only when primary stability is achieved and with the appropriate occlusal loading.

    Device Description

    Endosseous implants are self-tapping and threaded, and offered having root-form or wide thread forms. Root-form implant diameters range from 3.25mm to 5.5mm having lengths from 8mm to 14mm. Wide thread implant diameters are available in 4.1 and 4.5mm (8mm to 14mm lengths), 5.5mm (8mm to 12mm lengths) and 6.5mm (8mm to 10mm lengths). Cover screws and healing caps provide protection to the threads of the abutment connection during endosseous and gingival healing. Cover screws are pre- packaged with each implant. Healing caps are provided as an alternative to the cover screw and are packaged separately. The Implant-One™ dental implants and cover screws are provided sterile. Not all abutments can be used for single-unit restorations. The conical, angled conical, ball, locator and glueless abutments are intended only for multi-unit loaded restorations. The ball, locator and glueless abutments are to be used in fully removable dentures. The conical and angled conical abutments are to be used in screw retained dentures and with the titanium sleeve for screw retention. The final design parameters for the custom blank abutment are as follows: maximum total height, 12.5mm; minimum/maximum gingival height, 0.5mm/6mm; minimum post height, 4mm; maximum angulation, 30°; minimum wall thickness, 0.78 (at 1.5mm above the proximal end); minimum diameter, 3.75 mm for the 300 Series, 4.25 mm for the 300 Series and 4.75 mm for the 500 Series.

    AI/ML Overview

    The provided text is a 510(k) summary for the Implant-One™ System, a dental implant device. It demonstrates the device's substantial equivalence to existing legally marketed predicate devices through non-clinical performance data.

    However, the document does not describe a study involving an algorithm, AI assistance, human readers, or any form of "acceptance criteria" related to diagnostic performance or accuracy as one would expect for an AI/ML medical device.

    Instead, the "acceptance criteria" and "study" described in this document pertain to engineering and biocompatibility performance of a physical medical device (dental implants and abutments), not a software or AI-driven system.

    Therefore, for each of the requested points, the answer is that the information is not applicable (N/A) in the context of this document, as it describes a physical dental implant system and not an AI/ML device that would have such performance criteria.

    Here's how to address each point based on the provided text:


    Acceptance Criteria and Device Performance (Based on Device Type - Dental Implants)

    Since this is for a physical dental implant system, the "acceptance criteria" are related to mechanical integrity, biocompatibility, sterilization, and packaging/shelf-life, rather than diagnostic accuracy. The "device performance" indicates that the device met these criteria by demonstrating substantial equivalence to predicates.

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific Acceptance Criteria (Demonstrated Equivalence To)Reported Device Performance
    Mechanical PerformanceISO 14801 (for worst-case construct performance)"Non-clinical mechanical testing of the worst case Implant-One™ System construct was performed according to ISO 14801 and demonstrated that the Implant-One™ system performs as well as or better than the predicate devices."
    BiocompatibilityISO 10993-5"Biocompatibility testing was performed according to ISO 10993-5 and demonstrated substantial equivalence."
    SterilizationISO 11137 and ISO 17665"Sterilization validations were performed according to ISO 11137 and 17665 and demonstrated substantial equivalence."
    Packaging & Shelf-lifeASTM D4169 (including ASTM F1886, ASTM F88, and ASTM F1929)"Packaging and shelf-life validations were performed according to ASTM D4169 including ASTM F1886, ASTM F88 and ASTM F1929 and demonstrated substantial equivalence."
    Endotoxin Limit<20 EU/Device (per AAMI ST72)"The Limulus amebocyte lysate (LAL) test was performed on sterile devices per AAMI ST72 and met the <20EU/Device endotoxin limit."
    Material/Surface AnalysisIdentification of blast media/particles, removal treatments/agents, chemical analysis, SEM photomicrographs (per FDA Guidance)"For the modified implant surfaces, the following information was submitted per FDA Guidance...: identification of the blast media, composition of the particles, identification of the treatments and agents used to remove the particles, a chemical analysis of the surface and SEM photomicrographs of the surface."

    Regarding AI/ML Device Specifics (Not Applicable to this Document):

    The following points are typically relevant for AI/ML device submissions, and the provided document for a physical dental implant system explicitly states "No clinical data was used in support of this submission." Therefore, these points are not applicable.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • N/A. The submission is for a physical medical device (dental implant). No test set of clinical data was used. The document explicitly states: "No clinical data was used in support of this submission."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • N/A. No clinical test set or ground truth established by experts was used.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • N/A. No clinical test set requiring adjudication was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • N/A. This is a physical device, not an AI/ML system, so an MRMC study is not relevant and was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • N/A. This is a physical device, not an AI/ML system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • N/A. No clinical ground truth was established or used, as no clinical data was involved in this submission.

    8. The sample size for the training set

    • N/A. This is a physical device, not an AI/ML system requiring a training set.

    9. How the ground truth for the training set was established

    • N/A. This is a physical device, not an AI/ML system.
    Ask a Question

    Ask a specific question about this device

    K Number
    K102822

    Validate with FDA (Live)

    Date Cleared
    2011-01-10

    (104 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Implant-One™ system is indicated for surgical placement in partially or completely edentulous upper or lower jaws to provide a means for prosthetic attachment to restore a patient's chewing function. The Implant-One™ system is indicated for immediate loading only when primary stability is achieved and with the appropriate occlusal loading.

    Device Description

    The Implant-One™ Dental Implants include endosseous dental implants, cover screws, healing caps, dental implant abutments and abutment screws in a variety of sizes to accommodate differing patient anatomy. Endosseous implants are self-tapping, root-form and threaded. They range from 3.25mm to 5.5mm in diameter with lengths ranging from 8mm to 14mm and have an internal, threaded abutment connection having a Morse style taper. Cover screws and healing caps provide protection to the threads of the abutment connection during endosseous and gingival healing. Cover screws are pre-packaged with each implant. Healing caps are provided as an alternative to the cover screw and are packaged separately. Abutment options include Standard and Restorative in various heights. These are fastened to the implant using an abutment screw.

    AI/ML Overview

    The provided text describes the 510(k) summary for Custom Dental Implants Inc.'s Implant-One™ Dental Implants. This document focuses on demonstrating substantial equivalence to predicate devices through non-clinical testing rather than through clinical studies involving human patients or complex AI algorithms. Therefore, many of the requested elements for describing a study proving device acceptance (like sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, or specific ground truth types for AI models) are not applicable to this submission.

    The acceptance criteria are implicitly defined by the demonstration of equivalent performance to predicate devices through mechanical testing.

    Here's a breakdown based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The acceptance criteria for this device are not explicitly stated as quantitative thresholds in the document. Instead, the acceptance is based on demonstrating that the "worst case Implant-One™ system construct performed as well as or better than the predicate devices" in non-clinical dynamic testing according to ISO 14801.

    Acceptance Criteria (Implicit)Reported Device Performance (Non-Clinical Dynamic Testing according to ISO 14801)
    Mechanical performance for worst-case construct is equivalent to or better than predicate devices.The Implant-One™ system performed as well as or better than the predicate devices.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified for non-clinical testing, however it refers to "worst case Implant-One™ system construct" which usually implies a limited number of samples designed to represent the most challenging scenario.
    • Data Provenance: Not applicable as it's a non-clinical mechanical test. The test was performed by Custom Dental Implants Inc.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable as no human experts were involved in establishing ground truth for mechanical testing. The "ground truth" here is the objective measurement of mechanical properties.

    4. Adjudication method for the test set

    • Not applicable as no adjudication was needed for objective mechanical testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is a medical device (dental implant) 510(k) submission, not an AI/software as a medical device (SaMD) submission, and no MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a medical device (dental implant) 510(k) submission, not an AI/SaMD submission.

    7. The type of ground truth used

    • Ground Truth Type: Objective mechanical measurements (e.g., fatigue life, fracture strength) as defined by the ISO 14801 standard for dynamic testing of endosseous dental implants. This is a physical, measurable "ground truth" for the device's performance under simulated physiological loading conditions.

    8. The sample size for the training set

    • Not applicable. There was no training set as no machine learning or AI algorithm was developed or evaluated.

    9. How the ground truth for the training set was established

    • Not applicable. There was no training set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1