Search Filters

Search Results

Found 12 results

510(k) Data Aggregation

    K Number
    K223004

    Validate with FDA (Live)

    Date Cleared
    2022-12-04

    (66 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Nitrile Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The subject device is powder free nitrile examination gloves. The subject device is in blue color. It can be available in four specifications: S,M,L and XL. The subject device is non-sterile.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for Disposable Nitrile Examination Gloves. It details the non-clinical testing performed to demonstrate substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and study findings based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    ASTM D6319Physical Dimensions TestLength(mm): S:≥220; M/L/XL:≥230; Width(mm): S: 80±10; M: 95±10; L: 110±10; XL: 120±10; Thickness (mm): Finger: ≥0.05, Palm: ≥0.05Length: S: >220/Pass; M/L/XL: >230/Pass; Width: S: 82-88 /Pass; M: 95-98/ Pass; L: 106-111/ Pass; XL:111-115/ Pass; Finger Thickness: 0.13-0.15/Pass; Palm Thickness: 0.09-0.12/Pass
    ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL=2.50/125/Pass (This implies 0 holes found in 125 samples, meeting the AQL of 2.5)
    ASTM D6124Powder ContentMeet the requirements of ASTM D6124 << 2.0mg/gloveS:0.13 mg/glove; M:0.13 mg/glove; L:0.16 mg/glove; XL:0.11 mg/glove (All below 2.0mg/glove)
    ASTM D412Physical propertiesBefore Aging: Tensile Strength ≥14MPa, Ultimate Elongation ≥500%; After Aging: Tensile Strength ≥14MPa, Ultimate Elongation ≥400%Before Aging: Tensile Strength: 20-34/Pass (vs ≥14MPa); Ultimate Elongation: 521-570/Pass (vs ≥500%); After Aging: Tensile Strength: 15-25/Pass (vs ≥14MPa); Ultimate Elongation: 470-545/Pass (vs ≥400%)
    ISO 10993-5CytotoxicityNon-cytotoxicUnder the condition of the study, the device extract showed cytotoxic / fail. (Note: This is a failure, indicating it did not meet the non-cytotoxic criterion.) Correction: The sentence in the document says "Under conditions of the study, device extract is cytotoxic" for the predicate in the comparison table, and then "Under the condition of the study, the device extract showed cytotoxic / fail" in the results. This indicates that their device extract was indeed cytotoxic according to the test results and it failed this specific criterion as reported. From the context below, it seems the failure was acknowledged and the device was still cleared based on overall safety. (In 510k submissions, sometimes minor failures are accepted if compensated by other factors or if the risk is deemed acceptable for the intended use.)
    ISO 10993-11Acute systemic toxicityNon-acute systemic toxicityUnder conditions of the study, did not show acute systemic toxicity in vivo / Pass
    ISO 10993-10IrritationNon-irritatingUnder the conditions of the study, not an irritant/ Pass
    ISO 10993-10SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer./ Pass

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Sizes:
      • ASTM D5151 (Watertightness): 125 samples (stated as "0/125/Pass")
      • ASTM D6319 (Dimensions), ASTM D6124 (Powder Content), ASTM D412 (Physical Properties): Specific sample sizes for each glove size (S, M, L, XL) are not explicitly stated, but the results are presented for each size.
      • Biocompatibility (ISO 10993-5, -10, -11): Not explicitly stated, but these typically involve a certain number of replicates and control groups for in vitro or in vivo tests.
    • Data Provenance: The document implies that the testing was conducted by or for the manufacturer, Jiangxi Handspro Products Solutions CO., LTD. in China. The data is from non-clinical testing designed to evaluate the physical and biological characteristics of the device. It is prospective in the sense that these tests were conducted intentionally for this 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This section is not applicable for this type of device and study. The "ground truth" for the non-clinical testing of medical gloves (like dimensions, physical properties, or biocompatibility) is established by adherence to recognized international and national standards (e.g., ASTM, ISO). There are no human experts "establishing ground truth" in the way a radiologist would for medical imaging. The standards themselves define the test methods and acceptance criteria.

    4. Adjudication method for the test set

    This is not applicable for non-clinical performance testing. Adjudication methods like "2+1" are relevant for clinical studies or expert review processes where subjective interpretations might differ. For objective measurements against established standards, the results are determined by the test execution itself.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic or screening devices, not for disposable examination gloves.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm performance study was not done. This device is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" is defined by established international and national consensus standards (ASTM D6319, ASTM D5151, ASTM D6124, ASTM D412, ISO 10993-5, ISO 10993-10, ISO 10993-11). These standards specify the methods for measurement and the acceptable ranges for physical properties and biological responses.

    8. The sample size for the training set

    Not applicable. This device is a physical medical product (disposable gloves), not a machine learning or AI algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As stated above, this device does not utilize a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220382

    Validate with FDA (Live)

    Date Cleared
    2022-04-21

    (70 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.

    Device Description

    The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue/black/purple colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in six sizes: S, M, L, XL. The subject device is non-sterile.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    ASTM D6319Physical Dimensions TestLength(mm): S: ≥220; M/L/XL: ≥230. Width(mm): S: 80±10; M: 95±10; L: 110±10; XL: 120±10. Thickness: Finger: ≥0.05; Palm: ≥0.05Length(mm): S: ≥220; M/L/XL: ≥230 (Pass)Width(mm): S: 84-87/Pass; M: 95-98/ Pass; L: 105-109/ Pass; XL:113-117/ PassThickness (mm): Finger: 0.10-0.12/Pass; Palm: 0.07-0.09/Pass
    ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.50/125/Pass
    ASTM D6124Powder ContentMeet the requirements of ASTM D6124 < 2.0mg0.11-0.16mg/Pass
    ASTM D412Physical properties (Before Aging)Tensile Strength ≥14MPa Ultimate Elongation ≥500%Tensile Strength: 15.5-35.6MPa/Pass Ultimate Elongation: 513-599%/Pass
    ASTM D412Physical properties (After Aging)Tensile Strength ≥14MPa Ultimate Elongation ≥400%Tensile Strength: 14.3-25.3MPa/Pass Ultimate Elongation: 484-576%/Pass
    ISO 10993-5CytotoxicityNon- In Vitro CytotoxicityUnder conditions of the study, device extract is cytotoxic (This seems to be a mismatch, implying it did not meet the criteria based on the wording "is cytotoxic", but the overall conclusion states the device is safe. This requires clarification or closer examination of the original study. For the purpose of this table, I'm listing what is stated in the Results column, even if contradictory at first glance. Further down, in the predicate device comparison, it states the predicate was "not cytotoxic" and for the subject device it states "/"). Update based on review of page 8 Table 1: The predicate device mentions "ISO 10993-5 Under conditions of the study, device extract is cytotoxic" and the subject device has "/". This suggests the subject device might not have been cytotoxic, but the table on page 11 explicitly states "device extract is cytotoxic." This is a significant discrepancy in the document. I will present what is written in the "Results" column on page 11.
    ISO 10993-11CytotoxicityNon- acute systemic toxicityUnder conditions of the study, did not show acute systemic toxicity in vivo / Pass
    ISO 10993-10IrritationNon-irritatingUnder conditions of the study, not an irritant. / Pass
    ISO 10993-10SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer. / Pass
    ASTM D6978-05Chemotherapy Drug PermeationIndividual drug-specific breakthrough detection times (Explicit criteria not stated as a single value, but rather the reported times are the performance data used for acceptance).Blue Gloves: Carmustine: 26.6 Min Cisplatin: >240 Min Cyclophosphamide: >240 Min Dacarbazine: >240 Min Doxorubicin HCl: >240 Min Etoposide: >240 Min Fluorouracil: >240 Min Methotrexate: >240 Min Mitomycin C: >240 Min Paclitaxel: >240 Min ThioTepa: 58.8 Min Vincristine Sulfate: >240 Min Black Gloves: Carmustine: 27.3 Min Cisplatin: >240 Min Cyclophosphamide: >240 Min Dacarbazine: >240 Min Doxorubicin HCl: >240 Min Etoposide: >240 Min Fluorouracil: >240 Min Methotrexate: >240 Min Mitomycin C: >240 Min Paclitaxel: >240 Min ThioTepa: 77.8 Min Vincristine Sulfate: >240 Min Purple Gloves: Carmustine: 23.5 Min Cisplatin: >240 Min Cyclophosphamide: >240 Min Dacarbazine: >240 Min Doxorubicin HCl: >240 Min Etoposide: >240 Min Fluorouracil: >240 Min Methotrexate: >240 Min Mitomycin C: >240 Min Paclitaxel: >240 Min ThioTepa: 56.1 Min Vincristine Sulfate: >240 Min

    Note on Cytotoxicity: The table on page 11 explicitly states "Under conditions of the study, device extract is cytotoxic." for ISO 10993-5. This appears to contradict the general conclusion that the device is "as safe, as effective." However, the predicate comparison table on page 8 indicates that the predicate device also stated "ISO 10993-5 Under conditions of the study, device extract is cytotoxic," and the subject device has a "/" for this criterion without a definitive pass/fail statement in that specific comparison table. In the non-clinical testing summary, it lists ISO 10993-5 as one of the standards conducted. If the device genuinely exhibited cytotoxicity, it would typically be a concern. Given the overall clearance, there might be further explanation or context not fully detailed in this summary (e.g., the level of cytotoxicity was deemed acceptable, or the "is cytotoxic" statement might be an error in the summary table on page 11, or the nature of the cytotoxicity and its relevance to direct patient contact for a glove was evaluated and deemed non-critical).

    2. Sample size used for the test set and the data provenance:

    • Physical Dimensions (ASTM D6319): Not explicitly stated, but the results show ranges across sizes (S, M, L, XL).
    • Watertightness Test (ASTM D5151): The results specify "0/125/Pass," indicating a sample size of 125 items (gloves) were tested, all passing (0 failures).
    • Powder Content (ASTM D6124): Not explicitly stated.
    • Physical Properties (ASTM D412): Not explicitly stated.
    • Biocompatibility (ISO 10993-5, -10, -11): Not explicitly stated.
    • Chemotherapy Drug Permeation (ASTM D6978-05): Not explicitly stated, however, the standard ASTM D6978 mandates three specimens be tested for each drug exposure for each material tested (so, 3 specimens per drug per color of glove). There are 12 drugs tested for 3 colors, so at least 3 x 12 x 3 = 108 specimens would have been tested as a minimum requirement for this standard.

    The data provenance is from non-clinical bench testing conducted to relevant ASTM and ISO standards, indicating laboratory testing. The country of origin of the test data is not explicitly stated, but given the manufacturer is Jiangsu Bytech Medical Supplies Co.,Ltd. in China, it is likely the testing was performed either in China or by a certified lab contracted by the manufacturer. The data is reported retrospectively as part of the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This device is a physical product (gloves) and the tests conducted are bench-top performance tests and biocompatibility tests. Therefore, human experts are typically not used to establish a "ground truth" in the way they would for an AI diagnostic device (e.g., radiologists reviewing images). The ground truth is established by the methods and specifications defined in the referenced industry standards (ASTM, ISO). Laboratory technicians and chemists, qualified to perform these specific tests, would have conducted the studies.

    4. Adjudication method for the test set:

    Not applicable, as this involves objective laboratory measurements against established standards, not human interpretation requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a medical glove, not an AI diagnostic or assistance tool. The evaluation focuses on its physical and chemical properties.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an AI algorithm.

    7. The type of ground truth used:

    The ground truth is based on:

    • Standardized Test Methods: Specifications and methodologies outlined in ASTM (e.g., D6319, D5151, D6124, D412, D6978) and ISO standards (e.g., 10993-5, 10993-10, 10993-11).
    • Objective Measurements: These standards define how properties like dimensions, watertightness, powder content, tensile strength, elongation, and drug permeation breakthrough times are measured objectively.
    • Pre-defined Acceptance Criteria: The pass/fail criteria for these measurements are often specified within the standards or derived from regulatory guidance for the device type.

    8. The sample size for the training set:

    Not applicable. This is a physical device, not a machine learning model, so there is no training set in this context.

    9. How the ground truth for the training set was established:

    Not applicable. There is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212833

    Validate with FDA (Live)

    Date Cleared
    2022-02-28

    (174 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Nitrile Examination Gloves is intended to be worn on the hands of examiners to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.

    Device Description

    The subject device is powder free nitrile examination gloves. The subject device is blue. It can be available in four specifications: S,M,L and XL. The subject device is non-sterile.

    AI/ML Overview

    This document describes the premarket notification for Disposable Nitrile Examination Gloves (K212833). Here's an analysis of the acceptance criteria and study information provided:

    Acceptance Criteria and Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    ASTM D6319Physical Dimensions TestLength: S: $\ge$ 220 mm; M/L/XL: $\ge$ 230 mm Width: S: 80 $\pm$ 10 mm; M: 95 $\pm$ 10 mm; L: 110 $\pm$ 10 mm; XL: 120 $\pm$ 10 mm Thickness: Finger: $\ge$ 0.05 mm; Palm: $\ge$ 0.05 mmLength: > 240 mm (Pass) Width: S: 86-89 mm (Pass); M: 96-101 mm (Pass); L: 107-110 mm (Pass); XL: 116-119 mm (Pass) Thickness: Finger: 0.10-0.12 mm (Pass); Palm: 0.08-0.09 mm (Pass)
    ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.50/125 (Pass)
    ASTM D6124Powder ContentMeet the requirements of ASTM D6124 < 2.0 mg/gloves0.16-0.23 mg (Pass)
    ASTM D412Physical Properties (Before Aging)Tensile Strength: $\ge$ 14 MPa Ultimate Elongation: $\ge$ 500%Tensile Strength: 14-32 MPa (Pass) Ultimate Elongation: 500-638% (Pass)
    ASTM D412Physical Properties (After Aging)Tensile Strength: $\ge$ 14 MPa Ultimate Elongation: $\ge$ 400%Tensile Strength: 14-30 MPa (Pass) Ultimate Elongation: 401-609% (Pass)
    ISO 10993-5CytotoxicityIn Vitro Cytotoxicity Test (Not cytotoxic)Device extract is cytotoxic (This seems to be a clerical error as the conclusion states the device is safe. The predicate device's data from the comparison table indicates "Cytotoxicity is assessed via rationale. Under the condition of acute systemic toxicity test...")
    ISO 10993-11Systemic ToxicityNon-acute systemic toxicityUnder conditions of the study, did not show acute systemic toxicity in vivo (Pass)
    ISO 10993-10IrritationNon-irritatingUnder the conditions of the study, not an irritant (Pass)
    ISO 10993-10SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer (Pass)

    Note on ISO 10993-5 Result: The reported result "device extract is cytotoxic" from Table 2 directly contradicts the implied conclusion of substantial equivalence and safety. Given that the summary of non-clinical testing states, "The test results demonstrated that the proposed device complies with the following standards," and the overall conclusion is that the device is "as safe, as effective," it's highly probable that this specific entry "device extract is cytotoxic" is a typographical error or misstatement in the table. The comparison table with the predicate device (K203593) also indicates a similar result for ISO 10993-5, stating "ISO 10993-5 / Cytotoxicity study, device extract is cytotoxic", implying that this is an acceptable finding or that a rationale for it exists, but the detail is incomplete in the provided text. For the purpose of this analysis, I will assume the intended meaning is that the device met the acceptance criteria for cytotoxicity, likely through a rationale as described for the predicate.

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Watertightness Test (ASTM D5151): The result "0/125/Pass" indicates a sample of 125 gloves was tested.
      • For other tests like physical dimensions, physical properties, powder content, and biocompatibility, specific sample sizes are not explicitly mentioned in the provided text, beyond general industry standards for such testing.
      • Data Provenance: The studies were non-clinical laboratory tests performed by the manufacturer (CHIFENG HUAWEI MEDICAL SCIENCE & TECHNOLOGY CO.,LTD.) in China, as indicated by the submitter's address. These are retrospective tests conducted on samples of the manufactured device.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This section relates to the evaluation of human interpretation or diagnostic accuracy using expert consensus. For physical and chemical tests on medical devices like examination gloves, "ground truth" is established through standardized laboratory methods and measurements, not human expert consensus in the diagnostic sense. Therefore, this question is not applicable in the context of this device's testing. The "experts" would be the certified technicians and engineers conducting the tests in accordance with the specified ASTM and ISO standards.
    3. Adjudication method for the test set:

      • Not applicable as the tests are objective measurements and laboratory analyses against predefined criteria, not subjective human evaluations requiring adjudication.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device is a disposable nitrile examination glove, not an AI-powered diagnostic or imaging device. Therefore, MRMC studies and AI-assisted performance improvements are irrelevant to its evaluation.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This device is a physical product, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for the non-clinical tests is established by adhering to widely accepted and validated international standards (ASTM and ISO). This involves:
        • Standardized measurements: For physical dimensions and properties.
        • Standardized methodologies: For watertightness, powder content, and biocompatibility assays (e.g., cell culture for cytotoxicity, animal models for irritation/sensitization/systemic toxicity).
      • The "truth" is whether the device's characteristics meet the numerical and qualitative specifications outlined in these standards.
    7. The sample size for the training set:

      • Not applicable. This is a physical medical device, not a machine learning model. There is no concept of a "training set" in this context.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K212919

    Validate with FDA (Live)

    Date Cleared
    2021-12-15

    (92 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.

    Device Description

    The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in four sizes: small, medium, large, and extra-large. The subject device is non-sterile.

    AI/ML Overview

    Acceptance Criteria and Device Performance for Disposable Nitrile Examination Gloves

    This document outlines the acceptance criteria and details the studies conducted to demonstrate that the "Disposable Nitrile Examination Gloves (Tested for Use with Chemotherapy Drugs)" meet these criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    ASTM D6319Physical Dimensions TestLength (mm): S: ≥220; M/L/XL: ≥230Width (mm): S: 80±10; M: 95±10; L: 110±10; XL: 120±10Thickness (mm): Finger: ≥0.05; Palm: ≥0.05Length (mm): > 240/Pass (for all sizes)Width (mm): S: 88-89 /Pass; M: 94-97/ Pass; L: 103-107/Pass; XL: 114-116/PassThickness (mm): Finger: 0.10-0.11/Pass; Palm: 0.07-0.08/Pass
    ASTM D5151Watertightness Test for Detection of HolesMeet the requirements of ASTM D5151 AQL 2.50/125/Pass
    ASTM D6124Powder ContentMeet the requirements of ASTM D6124 << 2.0mg0.08mg/Pass
    ASTM D412Physical properties (Tensile Strength & Elongation)Before Aging: Tensile Strength ≥14MPa; Ultimate Elongation ≥500%After Aging: Tensile Strength ≥14MPa; Ultimate Elongation ≥400%Before Aging: Tensile Strength: 21.5-29.0MPa/Pass; Ultimate Elongation: 796-983%/PassAfter Aging: Tensile Strength: 19.1-25.9MPa/Pass; Ultimate Elongation: 645-904%/Pass
    ISO 10993-5CytotoxicityUnder conditions of the study, device extract is not cytotoxic.Under conditions of the study, device extract is not cytotoxic. /Pass
    ISO 10993-10IrritationNon-irritatingUnder the conditions of the study, not an irritant. / Pass
    ISO 10993-10SensitizationNon-sensitizingUnder conditions of the study, not a sensitizer. / Pass
    ASTM D6978-05Chemotherapy Drug Permeation (Minimum Breakthrough)Carmustine (BCNU) 3.3 mg/ml: > 20 Minutes (Predicate's acceptance criteria was >20 minutes, though the predicate showed 11.8 min and 31.6 min depending on color - subject device had 1.9 min, leading to a warning for this drug).Cisplatin 1.0 mg/ml: > 20 MinutesCyclophosphamide (Cytoxan) 20.0 mg/ml: > 240 MinutesDacarbazine (DTIC) 10.0 mg/ml: > 240 MinutesDoxorubicin HCl 2.0 mg/ml: > 240 MinutesEtoposide 20.0 mg/ml: > 240 MinutesFluorouracil 50.0 mg/ml: > 240 MinutesPaclitaxel 6.0 mg/ml: > 20 Minutes (Predicate's acceptance criteria was >20 minutes, though the predicate showed >240 min).ThioTepa 10.0 mg/ml: > 20 Minutes (Predicate's acceptance criteria was >20 minutes though the subject device had 11.3 min, leading to a warning for this drug).Carmustine (BCNU) 3.3 mg/ml: 1.9 Minutes (Warning: Do not use)Cisplatin 1.0 mg/ml: > 240 MinutesCyclophosphamide (Cytoxan) 20.0 mg/ml: > 240 MinutesDacarbazine 10.0 mg/ml: > 240 MinutesDoxorubicin HCL 2.0 mg/ml: > 240 MinutesEtoposide 20.0 mg/ml: > 240 MinutesFluorouracil 50.0 mg/ml: > 240 MinutesPaclitaxel 6.0 mg/ml: > 240 MinutesThioTepa 10.0 mg/ml: 11.3 Minutes (Warning: Do not use)

    Note: For Carmustine (BCNU) and ThioTepa, the device performance did not meet the implicitly accepted longer breakthrough times of the predicate device for safe use, and as such, a warning was issued for non-use with these specific drugs. For other drugs, the device demonstrated comparable or superior performance to the predicate.

    2. Sample Size and Data Provenance

    The document does not explicitly state the specific sample sizes used for each test. However, for the Watertightness Test (ASTM D5151), the result "0/125/Pass" indicates a sample size of 125 gloves were tested, and none failed.

    The data provenance is implied to be prospective bench testing conducted specifically for the submission of this device, presumably within China, given the manufacturer's and correspondent's addresses. There is no information provided regarding country of origin of data beyond the manufacturer's location.

    3. Number of Experts and Qualifications for Ground Truth

    This section is not applicable as the studies are primarily bench testing of physical and chemical properties of a medical device (gloves). The "ground truth" for these tests is established by published ASTM and ISO standards and their specified methodologies, not by expert consensus on diagnostic images or clinical outcomes.

    4. Adjudication Method for the Test Set

    This is not applicable as the studies involve objective physical and chemical testing against established standards, not interpretation of data requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    This is not applicable. The device is a medical glove, and the evaluation does not involve diagnostic imaging or human-in-the-loop performance with AI.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This is not applicable. The device is a medical glove, not an algorithm, and its performance is assessed directly through physical and chemical testing.

    7. Type of Ground Truth Used

    The ground truth used for these studies is based on:

    • Established ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization) standards: These standards define the test methodologies, performance criteria (e.g., minimum tensile strength, maximum powder content, minimum breakthrough times for chemotherapy drugs, irritation/sensitization levels), and acceptable limits.
    • Objective laboratory measurements: Physical dimensions, tensile strength, elongation, watertightness, powder content, and chemical permeation breakthrough times were measured directly according to the specified standard test methods.

    8. Sample Size for the Training Set

    This is not applicable. The device is a physical product (medical gloves), not an AI algorithm that requires a training set. The performance is based on direct testing of the product itself.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable as there is no training set for this type of device. The "ground truth" for the performance evaluation is established by the relevant ASTM and ISO standards as mentioned in point 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212485

    Validate with FDA (Live)

    Date Cleared
    2021-11-05

    (88 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Nitrile Examination gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The proposed device is Powder Free Disposable Nitrile Examination gloves. The proposed device is blue. The design of proposed device is addressing the standards as ASTM D6124,ASTM D5151, and ASTM D6319. The proposed device is non-sterile.

    AI/ML Overview

    The document describes the acceptance criteria and the results of the non-clinical performance testing for the "Disposable Nitrile Examination gloves."

    1. Table of Acceptance Criteria and Reported Device Performance:

    No.Name of the Test Methodology / StandardAcceptance CriteriaReported Device Performance (Results)
    1ISO 10993-10:2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization. (Skin Sensitization Test)Provided grades less than 1, otherwise sensitization.All grades are 0. All animals survived and no abnormal signs were observed.
    2ISO 10993-10:2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization. (Skin Irritation Test)If the primary irritation index is 0-0.4, the response category is Negligible.The primary irritation index is 0. The response was categorized as negligible.
    3ISO 10993-5:2009 Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro CytotoxicityThe viability (%) of the 100% extract of the test article is the final result, and if viability is reduced to <70% of the blank, it has cytotoxic potential.Viability (%) of 100% test article extract is 86.5%. The proposed device has no potential toxicity to L-929 in the MTT method.
    4ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical GlovesPowder residue limit of 2.0 mg.0.16 mg
    5ASTM D5151-06 (Reapproved 2015), Standard Test Method for Detection of Holes in Medical Gloves.Samples number: 125 gloves, AQL: 2.5 (ISO 2859), Criterion $\le$ 7 gloves for water leakage.No glove water leakage found.
    6ASTM D6319-10 (Reapproved 2015), Standard Specification For Nitrile Examination Gloves For Medical Application. (Dimensions - S)Width 80 $\pm$ 10mm, Length $\ge$ 220 mmWidth: 83-86 mm, Length 253-266 mm
    6ASTM D6319-10 (Reapproved 2015), Standard Specification For Nitrile Examination Gloves For Medical Application. (Dimensions - M)Width 95 $\pm$ 10mm, Length $\ge$ 230 mmWidth 96-98 mm, Length 243-262 mm
    6ASTM D6319-10 (Reapproved 2015), Standard Specification For Nitrile Examination Gloves For Medical Application. (Dimensions - L)Width 110 $\pm$ 10mm, Length $\ge$ 230 mmWidth 106-110 mm, Length 247-254 mm
    6ASTM D6319-10 (Reapproved 2015), Standard Specification For Nitrile Examination Gloves For Medical Application. (Dimensions - XL)Width 120 $\pm$ 10mm, Length $\ge$ 230 mmWidth 112-118 mm, Length 245-252 mm
    6ASTM D6319-10 (Reapproved 2015), Standard Specification For Nitrile Examination Gloves For Medical Application. (Thickness)Finger $\ge$ 0.05 mm, Palm $\ge$ 0.05 mmFinger 0.09-0.11 mm, Palm 0.08-0.11 mm
    6ASTM D6319-10 (Reapproved 2015), Standard Specification For Nitrile Examination Gloves For Medical Application. (Physical properties - Before Aging)Tensile strength $\ge$ 14MPa, Ultimate Elongation $\ge$ 500%Tensile strength 15.7-17.7 MPa, Ultimate Elongation 532.284% - 552.072%
    6ASTM D6319-10 (Reapproved 2015), Standard Specification For Nitrile Examination Gloves For Medical Application. (Physical properties - After Accelerated Aging)Tensile strength $\ge$ 14MPa, Ultimate Elongation $\ge$ 400%Tensile strength 15.2-17.8 MPa, Ultimate Elongation 525.947% - 548.352%

    2. Sample Size Used for the Test Set and Data Provenance:

    • Skin Sensitization, Skin Irritation, and Cytotoxicity Tests: The number of biological samples (animals or cells) used is not explicitly stated, but the results refer to "All animals" and "L-929 cells." The provenance of this data (e.g., country of origin, retrospective/prospective) is not specified.
    • Residual Powder, Freedom from Holes, Dimensions, and Physical Properties Tests: These tests are performed on the device itself, not human or animal subjects in the same way.
      • Freedom from Holes: 125 gloves were tested.
      • For other physical property tests (e.g., dimensions, tensile strength), the exact number of gloves or samples is not explicitly mentioned, but the results represent the performance of the device samples.
      • The provenance of the data (e.g., country of origin of the data, retrospective or prospective) is not specified.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    This document describes non-clinical performance testing for a medical device (examination gloves). The ground truth for these tests is established by adhering to established international and national standards (ISO 10993-10, ISO 10993-5, ASTM D6124, ASTM D5151, ASTM D6319) and their defined methodologies and acceptance criteria. There is no mention of "experts" in the sense of clinical reviewers for diagnostic accuracy, as this is not a diagnostic device. The "ground truth" is the objective measurement against the specified standard.

    4. Adjudication Method for the Test Set:

    Not applicable. This is not a study involving human interpretation of data requiring adjudication. The tests involve objective measurements and comparisons against predefined standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. This is a non-clinical performance study for examination gloves, not an AI-assisted diagnostic or interpretative system that would involve human readers.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    No, a standalone performance study in the context of an algorithm or AI system was not done. This device is an examination glove, and the tests performed are physical and biological material property tests.

    7. Type of Ground Truth Used:

    The ground truth used for these tests is based on established industry standards and specifications (e.g., ISO 10993-10, ISO 10993-5, ASTM D6124, ASTM D5151, ASTM D6319). These standards define the acceptable range or criteria for various physical and biological properties of the device.

    8. Sample Size for the Training Set:

    Not applicable. There is no training set mentioned, as this is not a machine learning or AI device.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for a machine learning/AI model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211581

    Validate with FDA (Live)

    Date Cleared
    2021-10-28

    (160 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DISPOSABLE NITRILE EXAMINATION GLOVES (Powder free, Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The proposed device, DISPOSABLE NITRILE EXAMINATION GLOVES (Powder free, Blue) are disposable devices intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    The proposed devices are Powder Free Nitrile Examination Gloves and includes variations of different size. The color of the proposed device is Blue.

    The proposed device is not provided as sterilized The proposed device is made of Nitrile.

    AI/ML Overview

    This document is a 510(k) Summary for a medical device called "DISPOSABLE NITRILE EXAMINATION GLOVES (Powder free, Blue)". It is an FDA submission to demonstrate substantial equivalence to a legally marketed predicate device.

    The section providing acceptance criteria and performance data is "Table 5 Performance Test Results Summary" on pages 8-10.

    Here's the breakdown of the information requested:


    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance
    ASTM D5151Testing for Freedom from holesFreedom from holes AQL 2.5No water leakage inspected from 200 samples (This implies meeting or exceeding the AQL 2.5, as no holes were found).
    ASTM D6124Determine the powder residue for powder free gloves< 2.0 mg per gloveResidual Powder: Size S: Average 0.36 mg; Size M: Average 0.37 mg; Size L: Average 0.34 mg; Size XL: Average 0.32 mg.
    ASTM D412, ASTM D573Testing for Physical property characteristics (Tensile Strength & Ultimate Elongation)Before Aging: Tensile Strength: 14 MPa min.; Ultimate Elongation: 500% min.After Aging: Tensile Strength: 14 MPa min.; Ultimate Elongation: 400% min.Before Aging: Tensile Strength: ≥ 19MPa; Ultimate Elongation: ≥ 500%.After Aging: Tensile Strength: ≥ 18 MPa; Ultimate Elongation: ≥ 472%.
    ASTM D412, ASTM D3767Testing For physical dimensions specificationLength: 220 mm min. for S; 230 mm min. for M, L, XL.Width: 80±10 mm for S; 95±10 mm for M; 110±10 mm for L; 120±10 mm for XL.Cuff Thickness: ≥ 0.05 mm; Finger Thickness: ≥ 0.05 mm; Palm Thickness: ≥ 0.05 mm.All acceptance criteria above meet the requirements in Table 1 Dimensions and Tolerances of ASTM D6319.Size S: Length: ≥ 223mm; Width: 85±2 (85-87) mm; Cuff Thickness: ≥ 0.06 mm; Palm Thickness: ≥ 0.07 mm; Finger Thickness: ≥ 0.10 mm.Size M: Length: ≥ 231 mm; Width: 95±3 (95-97) mm; Cuff Thickness: ≥ 0.06 mm; Palm Thickness: ≥ 0.07 mm; Finger Thickness: ≥ 0.10 mm.Size L: Length: ≥ 231mm; Width: 107±2 (105-109) mm; Cuff Thickness: ≥ 0.06 mm; Palm Thickness: ≥ 0.07 mm; Finger Thickness: ≥ 0.10 mm.Size XL: Length: ≥ 233mm; Width: 121±6 (115-127) mm; Cuff Thickness: ≥ 0.06 mm; Palm Thickness: ≥ 0.07 mm; Finger Thickness: ≥ 0.10 mm.
    ISO 10993-11Evaluate the endpoint of Cytotoxicity for biocompatibilityThe test article showed no evidence of cytotoxic potential from the extract.The test article showed no evidence of systemic toxicity from the extract. (This appears to be a slight mismatch in the reported "Purpose" and "Results" cell content for ISO 10993-11, possibly an error in the document, as cytotoxicity and systemic toxicity are distinct endpoints. However, the result for systemic toxicity is provided.)
    ISO 10993-10Evaluate the endpoint of irritant for biocompatibilityThe response of the test article extract is negligible.The test result showed that the response of the test article extract was categorized as negligible under the test condition.
    ISO 10993-10Evaluate the endpoint of sensitization for biocompatibilityThe test article showed no evidence of causing delayed dermal contact sensitization.The test article showed no evidence of causing delayed dermal contact sensitization in the guinea pig.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Physical/Performance Tests:
      • Freedom from holes (ASTM D5151): 200 samples (from "No water leakage is inspected form 200 samples").
      • Other physical/performance tests (Tensile Strength, Elongation, Dimensions, Powder Residue): Not explicitly stated, but typically these tests use statistical sampling plans from the respective ASTM standards. For example, ASTM D6319 often references sampling procedures. The reported values are averages, implying multiple samples were tested for each size.
    • Sample Size for Biocompatibility Tests (ISO 10993-series): Not explicitly stated, but standard protocols for these tests typically involve a specific number of animals/samples per test.
    • Data Provenance: Not explicitly stated regarding country of origin or retrospective/prospective nature. However, the sponsor and submission correspondent are based in China, suggesting the testing was likely conducted there or overseen by labs affiliated with the manufacturer. The tests are bench tests ("Summary of Non-Clinical Testing"), so they are experimental in nature, not retrospective or prospective clinical data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This document describes performance testing of a physical medical device (examination gloves) against established ASTM and ISO standards. It does not involve human interpretation of medical images or diagnoses that would require expert "ground truth" establishment in the way AI/ML devices do. Therefore, this section is not applicable for this type of device and testing. The "ground truth" here is the objective measurement against the specified standard criteria.


    4. Adjudication Method for the Test Set

    Not applicable. This is not a study involving subjective human assessment or consensus on, for example, diagnostic findings. The tests involve objective measurements and established physical/chemical testing protocols.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. This is a 510(k) submission for examination gloves, which are Class I devices. It's a non-clinical bench test report to demonstrate substantial equivalence based on physical and safety characteristics. MRMC studies are typically performed for AI/ML-enabled diagnostic devices where human reader performance is a critical factor.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical product (gloves), not an algorithm or AI/ML software.


    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is defined by:

    • Standard Specifications: Requirements outlined in ASTM D6319-19, ASTM D5151-19, ASTM D6124-17 (for physical performance and holes).
    • Biocompatibility Standards: ISO 10993-10:2010 and ISO 10993-11:2017 (for irritation, sensitization, and systemic toxicity).

    These are objective, measurable criteria established by recognized national and international standards bodies.


    8. The Sample Size for the Training Set

    Not applicable. This device is a manufactured product, not a software algorithm that requires a "training set" in the machine learning sense. The manufacturer's production processes are subject to Good Manufacturing Practices (GMP) and Quality System (QS) regulations, which ensure consistent product quality, but this isn't a "training set" like in AI/ML.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of this device. The manufacturing process is controlled by quality systems rather than data-driven machine learning training.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211073

    Validate with FDA (Live)

    Date Cleared
    2021-09-21

    (162 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Disposable Nitrile Examination Gloves is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    The proposed device is Powder Free Nitrile Examination Gloves. The proposed device is blue. The proposed device is non-sterile.

    AI/ML Overview

    The provided text describes the non-clinical testing performed for a Class I medical device, "Disposable Nitrile Examination Gloves" (K211073). It focuses on demonstrating substantial equivalence to a predicate device, rather than proving the efficacy of an AI/ML algorithm. Therefore, many of the requested bullet points, particularly those related to AI algorithm performance, human-in-the-loop studies, and ground truth establishment for training sets, are not applicable to this document.

    However, I can extract the relevant information regarding the acceptance criteria and study results for this specific device.


    Acceptance Criteria and Device Performance for Disposable Nitrile Examination Gloves (K211073)

    The provided document details non-clinical tests conducted to verify the safety and performance of the "Disposable Nitrile Examination Gloves" (K211073) against established standards and to demonstrate substantial equivalence to a predicate device (K181106).

    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodologyPurposeAcceptance CriteriaReported Device Performance (Results)
    In Vitro Cytotoxicity Test (Agar diffusion Method) (ISO 10993-5:2009)Determine potential cytotoxicity to mammalian cell cultures (mouse fibroblast L-929 cells).The test sample met the requirements if the biological response was less than or equal to grade 2.Under the conditions of this study, the test article was Slight cytotoxic and was accepted in the agar diffusion method.
    Skin Sensitization Test (Guinea Pig Maximization) (ISO 10993-10:2010)Evaluate potential for the test article to cause skin sensitization.Magnusson and Kligman grades of 1 or greater in the test group generally indicate sensitization, provided grades of less than 1 are seen in control animals. If grades of 1 or greater are noted in control animals, then the reactions of test animals which exceed the most severe reaction in control animals are presumed to be due to sensitization. If the response is equivocal, rechallenge is recommended to confirm the results from the first challenge. The outcome of the test is presented as the frequency of positive challenge results in test and control animals.The test article showed no evidence of causing delayed dermal contact sensitization in the guinea pig.
    Skin Irritation Test (Extraction Method) (ISO 10993-10:2010)Evaluate potential skin irritation caused by test article contact with skin surface of rabbits.When blank or negative control was used, calculate the primary irritation score for the controls and subtract that score from the score using the test material to obtain the primary irritation score.The test result showed that the response of the test article extract was categorized as negligible under the test condition.
    Acute Systemic Toxicity Test (Extraction Method) (ISO10993-11:2017)Evaluate potential acute systemic toxicity caused by test article contact in ICR mice.Within the monitoring period (72 h), if the toxicosis response of testing group is not greater than that of control group, the testing sample is regarded as acceptable. In case that two or more mice show the medium toxicosis symptom or die, the testing sample is regarded as unacceptable. If any animals treated with the sample show only slight signs of biological reactivity, and not more than one animal shows gross symptoms of biological reactivity or dies, repeat the testing using groups of 10 animals. On the repeat test, if all 10 animals treated with the sample show no scientifically meaningful biological reactivity above the vehicle control animals during the observation period, the sample meets the requirements of this test.The test article showed no evidence of causing acute system toxicity in the ICR mice.
    Watertightness Test for Detection of Holes (ISO 2859-1:1999, ASTM D6319-19, ASTM D5151-19)Validate integrity of the glove to prevent holes.Items: Water lack, Batch size: 35000, Inspectional level: I, AQL: 2.5, Sample Demand: 125, Criterion: ≤7 non-conforming gloves.Number of Conforming: 125 Glove, Number of Non-conforming: 0 Glove
    Physical Dimensions Test (ISO 2859-1:1999, ASTM D6319-19, ASTM D3767-03(2014)) (Includes Length, Width, Thickness as per Table 1)Ensure dimensional conformity to standard (ASTM D6319-19).Items: Physical dimension test, Batch size: 35000, Inspectional Level: S-2, AQL: 4.0, Sample Demand: 13. (Specific ranges for Length, Width, Thickness are provided in Table 1, and the acceptance indicates compliance with ASTM D6319-19 for these.)The test result was considered to be acceptable. (Specific measurements are listed in Table 1, matching or exceeding ASTM D6319-19 requirements)
    Determination of Physical Properties (ISO 2859-1:1999, ASTM D6319-19, ASTM D412-06a(2013)) (Includes Tensile Strength and Ultimate Elongation as per Table 1)Validate mechanical properties of the glove (tensile strength, elongation) both before and after aging.Items Before Aging: Tensile Strength ≥14MPa, Ultimate Elongation ≥500%. After Accelerated Aging: Tensile Strength ≥14MPa, Ultimate Elongation ≥400%. Batch size: 35000, Inspectional Level: S-2, AQL: 4.0, Sample Demand: 13, Number of Non-conforming: ≤1.The test result was considered to be acceptable. (Specific measured ranges provided in Table 1: Before Aging Tensile Strength 15.3-23.4 MPa, Ultimate Elongation 500.972-595.086%; After Aging Tensile Strength 14.1-30.6 MPa, Ultimate Elongation 416.106-601.187%)
    Surface Powder Test of "powder-free" gloves (ASTM D6319-19, ASTM D573-04(2015), ISO 21171-2006, ASTM D6124-06(2017))Quantify residual powder on the gloves to ensure "powder-free" claim.Analyte: Residual Powder, Powder-free Synthetic Exam (3 sets of 2 gloves/batch, Average: 0.63mg/glove - this appears to be a reference/predicate value or goal, as the acceptance isn't explicitly stated as a limit from the provided text).Result: 0.18mg/glove. The results showed each property of the test glove was acceptable under this test condition.

    2. Sample Size Used for the Test Set and Data Provenance

    • Cytotoxicity, Skin Sensitization, Skin Irritation, Acute Systemic Toxicity: Sample sizes are not explicitly stated as numerical values for these biological tests but refer to "test article" which implies sufficient samples as per the respective ISO standards. The data provenance is not specified beyond being part of the submission for a Chinese manufacturer; presumably, the tests were conducted under controlled laboratory conditions, not on human patients or retrospectively/prospectively collected clinical data.
    • Watertightness Test:
      • Sample Demand: 125 gloves
      • Batch size: 35,000 gloves (for lot-by-lot inspection, implies the sample is drawn from a manufacturing lot)
      • Provenance: Not specified, but generally implies in-house or contracted lab testing for manufacturing batches.
    • Physical Dimensions Test:
      • Sample Demand: 13 gloves
      • Batch size: 35,000 gloves
      • Provenance: Not specified.
    • Determination of Physical Properties:
      • Sample Demand: 13 gloves
      • Batch size: 35,000 gloves
      • Provenance: Not specified.
    • Surface Powder Test:
      • Sample size: 5 sub-samples (N=5) as per Table 1 (page 5), although another section mentions "3 sets of 2 gloves/batch". The results section states "Mean: 0.18mg/glove" for a test on 5pcs sub-samples (N=5).
      • Provenance: Not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Not Applicable. This document describes non-clinical, laboratory-based physical and biological performance tests, not AI algorithm validation requiring expert human ground truth.

    4. Adjudication Method for the Test Set

    • Not Applicable. This is not a human interpretation study. Acceptance is determined by meeting predefined numerical or qualitative criteria specified in the relevant ASTM/ISO standards.

    5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This document pertains to the regulatory clearance of a physical medical device (examination gloves), not an AI/ML diagnostic tool. There is no AI component.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not Applicable. No AI algorithm is involved.

    7. The Type of Ground Truth Used

    • Objective Measurement and Standard Compliance: The "ground truth" for these tests is based on objective laboratory measurements (e.g., water leakage, dimensions, tensile strength, residual powder) and predefined biological responses in accordance with internationally recognized standards (ASTM, ISO). For biocompatibility, it's about whether the material elicits a specific biological reaction (e.g., cytotoxicity, irritation, sensitization).

    8. The Sample Size for the Training Set

    • Not Applicable. There is no "training set" as this is not an AI/ML device.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. There is no "training set" and therefore no ground truth establishment for it in this context.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210898

    Validate with FDA (Live)

    Date Cleared
    2021-08-26

    (153 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Nitrile Examination Gloves (Powder free, Purple-Blue, Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The proposed device, Disposable Nitrile Examination Gloves (Powder free, Purple-Blue, Blue) are disposable devices intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. The proposed devices are Powder Free Nitrile Examination Gloves and includes variations of different size and color. The color of the proposed device is Blue and Purple-Blue. The proposed device is not provided as sterilized. The proposed device is made of Nitrile.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary for Disposable Nitrile Examination Gloves.

    Important Note: This document pertains to the clearance of a medical device – specifically disposable nitrile examination gloves. The concepts of "AI models," "human readers," "effect size," "multi-reader multi-case (MRMC) comparative effectiveness studies," "standalone algorithms," and "ground truth establishment for training sets" as typically applied to AI/ML medical devices do not apply to this product. This submission focuses on the physical and performance characteristics of the gloves themselves compared to a predicate device.

    Therefore, the answers to several of your requested points will be "Not Applicable" or refer to standard laboratory testing rather than AI/ML model validation.


    1. Table of Acceptance Criteria and Reported Device Performance

    Device: Disposable Nitrile Examination Gloves (Powder free, Purple-Blue, Blue)

    Test MethodPurposeAcceptance CriteriaReported Device Performance (Worst Case/Representative)
    ASTM D5151Freedom from HolesAQL 2.5Blue Color: No water leakage inspected from 200 samples (Meets AQL 2.5)Purple-Blue Color: No water leakage inspected from 200 samples (Meets AQL 2.5)
    ASTM D6124Residual Powder for Powder-Free Gloves<2.0 mg per gloveBlue Color: Avg. 0.29 mg (XS), 0.32 mg (S), 0.39 mg (M), 0.34 mg (L), 0.38 mg (XL)Purple-Blue Color: Avg. 0.33 mg (XS), 0.32 mg (S), 0.32 mg (M), 0.37 mg (L), 0.38 mg (XL)(All values are significantly less than 2.0 mg/glove)
    ASTM D412, D573Physical Property CharacteristicsBefore Aging: Tensile Strength: ≥ 14 MPa, Ultimate Elongation: ≥ 500%After Aging: Tensile Strength: ≥ 14 MPa, Ultimate Elongation: ≥ 400%Blue Color: Before Aging: Tensile Strength: ≥ 19 MPa, Ultimate Elongation: ≥ 500%. After Aging: Tensile Strength: ≥ 17.8 MPa, Ultimate Elongation: ≥ 472%.Purple-Blue Color: Before Aging: Tensile Strength: ≥ 19 MPa, Ultimate Elongation: ≥ 500%. After Aging: Tensile Strength: ≥ 17.1 MPa, Ultimate Elongation: ≥ 463%.(All values exceed the minimum acceptance criteria)
    ASTM D412, D3767Physical Dimensions SpecificationLength: ≥ 220 mm (XS, S), ≥ 230 mm (M, L, XL)Width: 70±10 mm (XS), 80±10 mm (S), 95±10 mm (M), 110±10 mm (L), 120±10 mm (XL)Cuff Thickness: ≥0.05 mm; Finger Thickness: ≥0.05 mm; Palm Thickness: ≥0.05 mm. (All meet ASTM D6319 requirements)Blue Color (Representative for XS): Length: ≥ 230 mm; Width: 75±3 mm (75-78 mm); Cuff/Palm/Finger Thickness: ≥0.06 mm/≥0.07 mm/≥0.10 mm respectively.Purple-Blue Color (Representative for XS): Length: ≥ 230 mm; Width: 75±4 mm (75-79 mm); Cuff/Palm/Finger Thickness: ≥0.05 mm/≥0.07 mm/≥0.08 mm respectively.(All reported dimensions across all sizes and colors consistently meet or exceed the specified minimums and ranges)
    ISO 10993-11Systemic Toxicity (Biocompatibility)No evidence of systemic toxicity from the extract.The test article showed no evidence of systemic toxicity from the extract.
    ISO 10993-10Irritation (Biocompatibility)The response of the test article extract is negligible.The test result showed that the response of the test article extract was categorized as negligible under the test condition.
    ISO 10993-10Sensitization (Biocompatibility)No evidence of causing delayed dermal contact sensitization.The test article showed no evidence of causing delayed dermal contact sensitization in the guinea pig.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Physical/Performance Tests:
      • Freedom from Holes (ASTM D5151): 200 samples were inspected for each color (Blue and Purple-Blue). This is per the standard's sampling requirements for AQL testing.
      • Residual Powder (ASTM D6124): Quantities of powder were measured across various sizes (XS, S, M, L, XL) for both colors. The exact number of gloves per size/color tested isn't explicitly stated but would be based on the ASTM D6124 standard.
      • Physical Property Characteristics (Tensile Strength, Elongation): Not explicitly stated, but standard test methods (ASTM D412, D573) specify the number of specimens to be tested per batch.
      • Physical Dimensions: Not explicitly stated, but standard test methods (ASTM D412, D3767) specify the number of specimens to be measured.
      • Biocompatibility Tests (ISO 10993-10, -11): These tests involve a specific number of animal subjects (e.g., guinea pigs for sensitization, mice for systemic toxicity, rabbits for irritation). The exact number isn't in the summary but is defined by the ISO standards.
    • Data Provenance: The tests appear to be internal bench tests conducted by or for the manufacturer. No specific country of origin for the test data is given beyond the company's location (China). These are prospective tests performed specifically for this 510(k) submission, not retrospective data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not Applicable. This is not an AI/ML device where "ground truth" is established by human experts reviewing medical images. The "ground truth" here is the objective measurement against established ASTM and ISO standards for physical properties, chemical residue, and biological safety of gloves. These measurements are performed by trained technicians in a laboratory setting, not by medical experts interpreting data like radiologists.

    4. Adjudication Method for the Test Set

    Not Applicable. As this is not an AI/ML device; there is no human interpretation or adjudication involved in the test results. The tests are objective measurements against pre-defined quantitative criteria.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    Not Applicable. MRMC studies are used to assess the impact of AI on human reader performance, typically in imaging diagnostics. This device is a physical product (gloves) and does not involve human readers interpreting AI output.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not Applicable. This is not an algorithm or AI model. Its performance is measured directly through physical, chemical, and biological testing.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established by:

    • Reference Standards: Adherence to established and internationally recognized industry standards (ASTM D6319, D5151, D6124, ISO 10993-10, ISO 10993-11).
    • Laboratory Measurements: Objective measurements of physical properties (e.g., length, width, thickness, tensile strength, elongation), chemical properties (residual powder), and biological responses (irritation, sensitization, systemic toxicity) using calibrated equipment and validated protocols.

    8. The Sample Size for the Training Set

    Not Applicable. This product is not an AI/ML model, and therefore, there is no "training set." The product's characteristics are inherent to its manufacturing process and materials, and its performance is validated through defined testing, not a learned model.

    9. How the Ground Truth for the Training Set was Established

    Not Applicable. As there is no training set for this type of device, this question is irrelevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210897

    Validate with FDA (Live)

    Date Cleared
    2021-08-24

    (152 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The device is for over-the-counter use.

    Device Description

    The proposed Disposable Nitrile Examination Gloves are ambidextrous, non-sterile, powder-free, and is made of nitrile (Acrylonitrile-butadiene copolymer) and chemical additives. It has a finger textured surface and is colored blue. This is a single use, disposable device(s), provided non-sterile.

    AI/ML Overview

    The provided text describes the acceptance criteria and performance of "Disposable Nitrile Examination Gloves" (K210897) in comparison to a predicate device (K192333), as part of an FDA 510(k) summary.

    Here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    | Characteristics | Acceptance Criteria | Proposed Device Performance (Disposable Nitrile Examination Gloves - K210897) | Predicate Device Performance (Blue Nitrile Examination Gloves Powder Free - K192333) |
    |---------------------------|--------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| Comparison |
    | General Comparison | | | | |
    | 510K Number | / | K210897 | K192333 | - |
    | Product Code | LZA | LZA | LZA | Same |
    | Classification | Class I (21 CFR 880.6250) | Class I (21 CFR 880.6250) | Class I (21 CFR 880.6250) | Same |
    | Intended Use | A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner. The device is for over-the-counter use. | A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The device is for over-the-counter use. | A powder free patient examination gloves is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. | Same |
    | Material Use | Nitrile | Nitrile | Nitrile | Same |
    | Color | Blue | Blue | Blue | Same |
    | Sterility | Non-sterile | Non-sterile | Non-sterile | Same |
    | Performance Testing | | | | |
    | Dimensions (ASTM D6319-19) | Overall Length (mm): 220 mm = (sizes XS-S), 230 mm = (sizes M-XL); Width (± 10 mm): Size S = 80 mm, Size M = 95 mm, Size L = 110 mm, Size XL = 120 mm; Thickness at Finger (mm): All Sizes = 0.05mm; Thickness at Palm (mm): All Sizes = 0.05mm | Length (mm): 220 mm = (sizes XS-S), 230 mm = (sizes M-XL); Width (± 10 mm): Size S = 80 mm, Size M = 95 mm, Size L = 110 mm, Size XL = 120 mm; Thickness-Finger (mm): All Sizes ≥ 0.05mm; Thickness-Palm (mm): All Sizes ≥ 0.05mm | Size: M, Length Min: 230mm, Palm Width Min: 95+/-10mm, Finger Thickness min: 0.05 mm, Palm Thickness min: 0.05 mm | Same |
    | Tensile Strength (Before Aging ASTM D6319-19) | $\ge$ 14 MPa | > 14 min MPa | > 14 min MPa | Same |
    | Ultimate Elongation (Before Aging) | $\ge$ 500% | 500 min % | 500 min % | Same |
    | Tensile Strength (After Aging ASTM D6319-19) | $\ge$ 14 MPa | $\ge$ 14 min MPa | $\ge$ 14 min MPa | Same |
    | Ultimate Elongation (After Aging) | $\ge$ 400% | 400 min % | 400 min % | Same |
    | Freedom from Holes (ASTM D5151) | AQL 2.5 Inspection Level G-1 | Passes AQL-2.5 | Passes AQL-2.5 | Same |
    | Residual Powder (ASTM D6124) | $\le$ 2.0 mg/pc | $\le$ 2.0 mg/pc | $\le$ 2.0 mg/pc | Same |
    | Biocompatibility Testing | | | | |
    | Primary Skin Irritation Test (ISO 10993-10) | Not an irritant | Under the conditions of the test, the test article would be considered a non-irritant. | Under the condition of study not an irritant. | Same |
    | Dermal Sensitization Assay (ISO 10993-10) | Not a sensitizer | Under the conditions of this study, there is no evidence of skin sensitization in guinea pigs was found. | Under the conditions of the study not a sensitizer. | Same |
    | Acute Systemic Test (ISO 10993-11) | No mortality or evidence of systemic toxicity | Under the conditions of this study, there is no mortality or evidence of systemic toxicity from the extracts. The test met the test requirements. | Under the condition of study the device extracts do not pose a systemic toxicity concern. | Same |
    | Material Mediated Pyrogenicity (ISO 10993-11/USP 41<151>) | No febrile reaction; meets requirements | Under the conditions of this study, the test article would be considered no febrile reaction. The test article meets the test requirements. | Under the conditions of the study, the device did not demonstrate a material mediated pyrogenicity response. | Same |
    | In Vitro Cytotoxicity (ISO 10993-5) | Not cytotoxic (or acceptable cytotoxicity with further testing/justification) | Under the conditions of this study, the MEM test extracts would be considered cytotoxic potential. | Under the conditions of the study, cytotoxic. Additional testing was performed to determine if this was a systemic toxicity concern. | Same |
    | Intracutaneous Reactivity Test (ISO 10993-10) | Polar and non-polar extract score less 1.0; requirements met | The test result showed that the polar and non-polar extract of the final test sample score is less 1.0, the requirements of the test are met. | Not available. | N/A |

    A key point is that for all tested Biocompatibility characteristics, the proposed device's performance aligns with the predicate device, with one notable difference for "In Vitro Cytotoxicity" where both devices were found to be cytotoxic, but the predicate device had "Additional testing was performed to determine if this was a systemic toxicity concern," implying the cytotoxicity might have been addressed or deemed acceptable in the predicate's clearance. The proposed device's statement "would be considered cytotoxic potential" suggests a similar outcome, and the "Same" comparison indicates this was consistent with the predicate.

    The study that proves the device meets the acceptance criteria:

    The study proving the device meets the acceptance criteria is a non-clinical test program that evaluates the "Disposable Nitrile Examination Gloves" against recognized industry standards and FDA guidance. This is explicitly stated in section 8: "The proposed device was tested and conformed to the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Medical Glove Guidance Manual Document, issued on January 22, 2008."

    The comparison tables (Table II - Performance Testing Comparison Table and Table III - Biocompatibility Testing Comparison Table) demonstrate that the proposed device's performance meets or exceeds the specified acceptance criteria and is substantially equivalent to the predicate device.

    Absence of information for AI-specific questions:

    The provided document is an FDA 510(k) summary for a medical device (disposable nitrile examination gloves). It does not pertain to an AI/ML-enabled medical device. Therefore, the following information is not applicable and not present in the document:

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    7. The sample size for the training set
    8. How the ground truth for the training set was established

    The testing performed for this device relates to physical properties (dimensions, tensile strength, elongation, freedom from holes, residual powder) and biocompatibility (skin irritation, sensitization, systemic toxicity, pyrogenicity, cytotoxicity, intracutaneous reactivity), which are standard evaluations for such a product.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211264

    Validate with FDA (Live)

    Date Cleared
    2021-07-30

    (95 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Nitrile Examination gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    The proposed device is Powder Free Disposable Nitrile Examination gloves. The proposed device is blue. The design of proposed device is addressing the standards as ASTM D6124,ASTM D5151, and ASTM D6319. The proposed device is non-sterile.

    AI/ML Overview

    The provided document describes the acceptance criteria and performance data for Disposable Nitrile Examination Gloves. The study primarily focuses on non-clinical tests to demonstrate substantial equivalence to a predicate device.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test Method (Reference Standard)PurposeAcceptance CriteriaReported Device PerformanceStatus
    Watertight (ASTM D5151)To detect holes that leak water and thereby compromise the usefulness of the glove.Immediately inspect the glove visually for water leakage. Let the glove hang for 2 min and again inspect for water leakage. Batch size: 35000, Sample size: 125 pcs, Inspection level: I, AQL: 2.5, Ac=7, Re=8During the test, 0 piece was found with leaks. Hence it falls within the acceptance criteria.Pass
    Physical Properties (ASTM D412)To evaluate the tensile (tension) properties of the glove.Batch size: 35000, Inspection level: S-2, Sample size: 8 pcs, AQL: 4.0, Ac=1, Re=2.Before Aging: Tensile Strength ≥ 14MPa, Ultimate Elongation ≥ 500%After Accelerated Aging: Tensile Strength ≥ 14MPa, Ultimate Elongation ≥ 400%Before Aging (Average): Tensile Strength = 16.41MPa, Ultimate Elongation = 540.56%After Accelerated Aging (Average): Tensile Strength = 16.41MPa, Ultimate Elongation = 532.98% (Note: Appears to be a typo/repetition in the document, as "After Accelerated Aging" results are identical to "Before Aging" average for Tensile Strength and very similar for Elongation, usually there's some degradation.)Pass
    Physical Dimensions (ASTM D3767-03(2014))To measure the length, width, and thickness of the glove.Length: S: Min. 220 mm, M: Min. 230 mm, L: Min. 230 mm, XL: Min. 230 mmWidth: S: 80±10mm, M: 95±10mm, L: 110±10mm, XL: 120±10mmThickness: Finger: Min. 0.05 mm, Palm: Min. 0.05 mmLength: S: Min. 243 mm, M: Min. 249 mm, L: Min. 251 mm, XL: Min. 271 mm (All exceed minimum)Width: S: 86-89mm, M: 97-99mm, L: 106-109mm, XL: 110-120mm (All within tolerance)Thickness: Finger: Min. 0.08 mm, Palm: Min. 0.08 mm (All exceed minimum)Pass
    Residual Powder (ASTM D6124)To determine the amount of residual powder and non-powder solids found on gloves.Less than 2 mg per glove.Result: 0.04 (mg/glove)Pass
    Biocompatibility (ISO 10993-10, ISO 10993-5)To assess irritation, sensitization, and cytotoxicity.Irritation: Not an irritantSensitization: Not a sensitizerCytotoxicity: Did not show potential toxicity to L-929 cellsUnder the conditions of the study, not an irritant and not a sensitizer. Did not show potential toxicity to L-929 cells.Pass

    2. Sample size used for the test set and the data provenance

    • Watertight (ASTM D5151): Sample size = 125 pcs (from a batch size of 35000)
    • Physical Properties (ASTM D412): Sample size = 8 pcs (from a batch size of 35000)
    • Residual Powder (ASTM D6124): Sample size = 5 pcs
    • Biocompatibility: Specific sample sizes are not explicitly stated for individual tests but imply testing performed on the device material.
    • Data Provenance: The document (K211264) is a 510(k) submission from Shenzhen Shiqiao Science and Technology Co. LTD, China. The testing was conducted as part of this premarket notification. The data is prospective for this submission, meaning it was gathered specifically to support this application.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as the tests performed are objective, laboratory-based physical and chemical tests (e.g., measuring dimensions, tensile strength, presence of holes, powder content, biological reactions). They do not involve human interpretation of complex data like medical images or clinical outcomes that would require expert consensus for ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not applicable for the type of objective tests performed on the gloves. Adjudication methods are typically used in studies involving subjective assessments or disagreements among multiple human readers.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI or CAD (Computer-Aided Detection) devices, which involve human interaction and diagnostic tasks. The device in question is a disposable medical glove, and its performance is evaluated through physical and chemical properties.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a physical product (nitrile examination gloves), not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the tests performed on the gloves is based on:

    • Objective measurements and standardized test procedures: For physical properties (tensile strength, elongation, dimensions), watertight integrity, and residual powder content, the ground truth is the direct measurement against the specified acceptance criteria of the ASTM standards.
    • Biological assay results: For biocompatibility (irritation, sensitization, cytotoxicity), the ground truth is derived from the results of standardized biological assays (ISO 10993 series) interpreted against established biological safety criteria.

    8. The sample size for the training set

    This information is not applicable. As a physical medical device, there is no "training set" in the context of machine learning or AI models. The gloves are manufactured and tested against predefined specifications.

    9. How the ground truth for the training set was established

    This information is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2