Found 9 results

510(k) Data Aggregation

    K Number
    K252599

    Validate with FDA (Live)

    Date Cleared
    2026-02-24

    (190 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode Laser Therapy Systems is intended for use in dermatological procedures requiring coagulation.

    Indications for Use:

    Mixed Wavelength Handpiece (755 nm + 808 nm + 1064 nm):

    • Temporary hair reduction (Hair-Removal function).
    • Treatment of benign vascular and vascular-dependent lesions (Vascular-Lesion function).

    Single Wavelength Handpieces (755 nm, 808 nm, or 1064 nm):

    • Hair removal and permanent hair reduction in all skin types (Fitzpatrick skin types I-VI).

    Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after completion of a treatment regimen.

    Device Description

    The Diode Laser Therapy Systems, Model WLA-02, is a medical device intended for dermatological procedures requiring coagulation. It includes mixed and single wavelength handpieces designed for temporary and permanent hair reduction, as well as the treatment of benign vascular lesions. The system delivers controlled laser energy to target melanin or vascular structures while integrated cooling helps protect surrounding skin tissue. The system comprises:

    • Main Console: Includes power switch, handpiece holder, LCD touch screen, emergency shut-off, hand-grip grab bar, handpiece connector, access door, observation window, power input, inlet, air switch, gate interlock connector, foot switch connector, vent, drain, and cooling fans.
    • Treatment Handpiece: Incorporates laser aperture, laser emission indicator, laser emission system, and handpiece display screen.
    • Foot Switch: Activates laser emission.
    • Embedded Software: Diode Laser Therapy Systems Control Software.
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K231896

    Validate with FDA (Live)

    Date Cleared
    2024-02-20

    (237 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode Laser Therapy System (Model: FG2000-B/FG2000-B Pro) is indicated for temporary hair reduction.

    The Diode Laser Therapy System(Model: FG2000-D+Pro/FG2000-D) is indicated for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs re-growing when measured at 6,9 and 12 months after the completion of a treatment regimen. It is suitable for all skin types (Fitzpatrick skin type I-VI), including tanned skin.

    Device Description

    The Diode Laser Therapy System consists of main unit, handpiece, and its accessories. Diode Laser Therapy System is intended for hair removal, mainly based on the principle of selective photothermal interaction, which means that lasers of specific wavelengths can only be selectively absorbed by the target color base.

    The Diode Laser Therapy System(Model: FG2000-B/FG2000-B Pro) are desktop devices, which combines 3 wavelengths (755+808+1064 nm) into a single handpiece to achieve purpose for temporary hair reduction.

    The Diode Laser Therapy System (Model: FG2000-D+Fro/ FG2000-E) are vertical device, which is a single wavelength (808 nm only) device.

    AI/ML Overview

    This document describes the FDA's 510(k) premarket notification for the "Diode Laser Therapy System" by Beijing ADSS Development Co., Ltd. A 510(k) submission aims to demonstrate that a new medical device is substantially equivalent to a legally marketed predicate device, rather than proving its safety and effectiveness de novo through clinical trials, as would be required for a PMA (Premarket Approval) device.

    Therefore, the provided document does not contain information typically found in a study proving a device meets acceptance criteria derived from a clinical trial for AI/ML-based medical devices. The device in question is a laser therapy system, not an AI/ML-based device that would require complex performance metrics like sensitivity, specificity, or AUC, nor would it involve human readers, expert ground truth adjudication, or MRMC studies.

    The "acceptance criteria" discussed in this document relate to the substantial equivalence to predicate devices and compliance with relevant performance and safety standards for laser devices.

    Here's an analysis based on the provided text, addressing the points you raised where applicable to a non-AI/ML device:

    1. A table of acceptance criteria and the reported device performance

    The document does not present a formal table of "acceptance criteria" in terms of performance metrics like sensitivity/specificity for an AI/ML device. Instead, "acceptance criteria" are implied by the demonstration of "substantial equivalence" to predicate devices and compliance with recognized international standards for medical electrical equipment and laser products.

    The tables provided in Section 6, "Summary of technological characteristics of device compared to the predicate device," serve as the primary comparison metrics to show substantial equivalence.

    Acceptance Criteria (Implied by Substantial Equivalence and Standards Compliance) and Reported Device Performance:

    ItemAcceptance Criteria (Predicate Characteristics / Standard Compliance)Reported Device Performance (Subject Device Characteristics)Comparison / Outcome
    Intended Use / Indications for UsePredicate K230371 (Alma Soprano Titanium): Temporary hair reduction; permanent reduction in hair regrowth (6, 9, 12 months, all skin types I-VI, including tanned skin) for 810nm and 755nm applicators. Also includes dermatology procedures requiring coagulation, benign vascular and vascular dependent lesions, and topical heating for temporary relief of muscle pain, etc.Subject Device FG2000-B/FG2000-B Pro: Indicated for temporary hair reduction.Subject Device FG2000-D+/FG2000-D+Pro/FG2000-E: Indicated for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9 and 12 months after the completion of a treatment regimen. Suitable for all skin types (Fitzpatrick skin type I-VI), including tanned skin.SE. Subject device's indications are covered by the predicate device (K230371, and K220381 for permanent hair reduction).
    Laser TypeDiode laserDiode laserSE
    Laser ClassificationClass IVClass IVSE
    Laser WavelengthPredicate K230371: combination of 755nm/810nm/1064nmSubject Device FG2000-B/FG2000-B Pro: combination of 755nm/808nm/1064nm (SE)Subject Device FG2000-D+/FG2000-D+Pro/FG2000-E: 808nm (SE)SE (808nm is common in diode lasers and considered equivalent to 810nm for these purposes)
    Spot SizePredicate K230371: 2010 mmmm (2cm²)Subject Device FG2000-B/FG2000-B Pro: 12×16 mm*mm(1.92cm²)Different. Deemed not to raise new questions of safety/effectiveness during bench testing.
    Predicate K220381: 15x15mm*mmSubject Device FG2000-D+/FG2000-D+Pro/FG2000-E: 10×10mm*mmDifferent. Deemed not to raise new questions of safety/effectiveness during bench testing.
    Pulse WidthPredicate K230371: Up to 200msSubject Device FG2000-B/FG2000-B Pro: up to 200msSE
    Predicate K220381: 1-300msSubject Device FG2000-D+/FG2000-D+Pro/FG2000-E: 2~100msDifferent. Deemed not to raise new questions of safety/effectiveness during bench testing.
    Energy FluencePredicate K230371: 2 to 8J/cm2Subject Device FG2000-B/FG2000-B Pro: 2~8J/cm2SE
    Predicate K220381: 1~100J/cm2Subject Device FG2000-D+/FG2000-D+Pro/FG2000-E: 2~100J/cm2Different (Lower minimum). Deemed not to raise new questions of safety/effectiveness during bench testing.
    Pulse FrequencyUp to 10 Hz / 1~10Hz1~10HzSE
    User InterfaceLCD Color TouchscreenLCD Color TouchscreenSE
    Electrical SafetyComply with ANSI/AAMA ES 60601-1, IEC 60601-2-22Comply with ANSI/AAMA ES 60601-1, IEC 60601-2-22SE
    EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2SE
    Laser SafetyComply with IEC 60601-2-22, IEC 60825-1Comply with IEC 60601-2-22, IEC 60825-1SE
    Handpiece Tip MaterialSapphireQuartz crystalSE (Functionally equivalent for intended purpose)
    Biocompatibility (Cytotoxicity, Sensitization, Irritation)Comply with ISO 10993-5, ISO 10993-10 (older versions for predicates), ISO 10993-23Comply with ISO 10993-5:2009, ISO 10993-10:2021, ISO 10993-23:2021SE (newer standards complied with, indicating continued safety)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • No clinical test set was used. Section 8 explicitly states: "There was no clinical testing performed."
    • The "test set" for this submission consists of non-clinical bench testing for performance and compliance with relevant standards. The data provenance is implied to be from Beijing ADSS Development Co., Ltd. (China), given their location and the submission coming from them.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. As there was no clinical testing or AI/ML component, there was no need for experts to establish ground truth in the context of diagnostic or treatment efficacy. The "ground truth" for substantial equivalence is the predicate device's proven characteristics and compliance with recognized standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No clinical data requiring adjudication was generated.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This relates to AI/ML device assessment, which this product is not.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This relates to AI/ML device assessment, which this product is not.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" in this context is the established safety and performance of the legally marketed predicate devices and compliance with internationally recognized performance and safety standards (e.g., IEC 60601 series, ISO 10993 series for biocompatibility, IEC 60825-1 for laser safety).

    8. The sample size for the training set

    • Not applicable. This device does not use a training set as it is not an AI/ML device.

    9. How the ground truth for the training set was established

    • Not applicable. No training set was used.

    In summary:

    This document is a 510(k) clearance letter for a diode laser therapy system. The "study" proving the device met "acceptance criteria" was primarily a non-clinical assessment based on:

    • Bench testing: To confirm the device's technical specifications and performance (e.g., laser wavelength, pulse characteristics, energy fluence).
    • Comparison to predicate devices: Demonstration of "substantial equivalence" in terms of intended use, technological characteristics, and safety features to legally marketed devices.
    • Compliance with recognized standards: Adherence to national and international safety and performance standards for medical electrical equipment, laser devices, and biocompatibility.

    The "acceptance criteria" are implied by meeting these requirements and showing that any differences from predicate devices do not raise new questions of safety or effectiveness. Clinical testing, human reader studies, and AI/ML specific evaluations (like those mentioned in your prompt) were not part of this clearance process for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220381

    Validate with FDA (Live)

    Date Cleared
    2022-05-20

    (99 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode Laser Therapy Systems is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    Diode Laser Therapy Systems is a surgical device, which is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI); The Diode Laser Therapy Systems utilize a semiconductor diode with invisible infrared radiation as a laser source to emit 808nm wavelength laser which is absorbed by melanin. The laser power is delivered to the treatment area via laser handpiece. The emission laser is activated by a foot-switch. The Diode Laser Therapy Systems utilize the principle of photoepilation for hair removal. The Photoepilation is a technique for removal of unwanted hair by thermal destruction of the hair follicle and its reproductive system (stems cells). The heat is caused by selective absorption of electromagnetic radiation emitted by laser light sources. As melanin is the main chromophore existing in hair follicles, so the melanin could absorb the energy from the laser, which would result in temperature rapid increase, then the hair follicle and its reproductive system (stems cells) would be destroyed by increased high temperature without damage epidermis and the surrounding normal tissue.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (Diode Laser Therapy Systems) and does not contain information about acceptance criteria and a study proving the device meets those criteria, as typically found in clinical study reports with specific performance metrics (e.g., sensitivity, specificity, accuracy).

    Instead, this document focuses on demonstrating substantial equivalence to a predicate device based on non-clinical testing and general safety standards.

    Therefore, many of the requested fields cannot be directly populated from the provided text. However, I can extract information related to the device's technical specifications and the non-clinical tests performed to support its safety and performance.

    Here's a breakdown of what can and cannot be answered based on the provided document:

    1. A table of acceptance criteria and the reported device performance

    The document does not specify formal "acceptance criteria" in terms of clinical performance metrics (like sensitivity, specificity, etc.) that the device must meet. Instead, the acceptance is based on demonstrating substantial equivalence to a predicate device through:

    • Matching intended use and fundamental technology.
    • Complying with recognized safety and performance standards (IEC, ISO).
    • Showing that any differences in technical specifications do not raise new questions of safety or effectiveness.
    Criteria Type (based on substantial equivalence)Acceptance/Comparison GoalProposed Device Performance/Finding
    Intended UseSame as predicateHair removal, permanent hair reduction on all skin types (Fitzpatrick I-VI), including tanned skin.
    Regulatory InformationSame Classification, Regulation, Product Code as predicateClass II, 21 CFR 878.4810, GEX, General & Plastic Surgery Review Panel.
    Laser TypeSame as predicate (Diode Laser)Diode Laser
    Laser ClassificationSame as predicate (Class IV)Class IV
    Laser WavelengthSame as predicate (808nm)808nm ± 2nm
    FluenceSame as predicate (1-100 J/cm²)1-100 J/cm²
    EMC, Electrical & Laser SafetyComply with relevant IEC standards (60601-1, 60601-1-2, 60601-2-22, 60825)Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1.
    BiocompatibilityComply with ISO 10993 series (Cytotoxicity, Sensitization, Irritation) for patient contact materials.No Cytotoxicity, No evidence of sensitization, No evidence of irritation (for 6061 Aluminum, Sapphire, ABS). Complies with ISO 10993-5, ISO 10993-10.
    Spot SizeNon-clinical tests show differences do not affect therapeutic effect and meet requirements.15mm × 15mm (different from predicate 1.44 cm² but tested to meet requirements).
    FrequencyWithin predicate’s range; bench tests support substantial equivalence and no new safety issues.1-10Hz (compared to predicate's 1-20 Hz, within range).
    Pulse DurationSlight difference, within predicate’s range, no new safety issues, bench tests support substantial equivalence.1-300ms (compared to predicate's 10-400ms, slight difference deemed safe).
    Power SupplyComplies with IEC 60601-1 and 60601-1-2, no effect on effectiveness/safety.110 Va.c. 60Hz 2000W (different from predicate but tested safe).
    Dimension/WeightComplies with IEC 60601-1 for mechanical performance, no effect on substantial equivalence.Different from predicate (460x390x1230mm, 75Kg vs 560x380x1180mm, 60Kg). Tested to be mechanically safe.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • The document explicitly states: "No clinical study is included in this submission."
    • Therefore, there is no "test set" in the context of clinical data, no sample size, and no data provenance relevant to clinical performance. The testing mentioned (non-clinical) refers to bench testing on the device hardware and materials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. No clinical test set with ground truth established by experts is mentioned.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No clinical test set.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a laser therapy system, not an AI-assisted diagnostic or imaging device. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is not an algorithm. Performance was assessed through non-clinical bench testing of the physical device.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • For the non-clinical tests, the "ground truth" or reference for compliance is established by international standards (IEC, ISO) for electrical safety, laser safety, electromagnetic compatibility, and biocompatibility, as well as the technical specifications of the predicate device.

    8. The sample size for the training set

    • Not applicable. No AI model or training set is mentioned as this is a physical medical device.

    9. How the ground truth for the training set was established

    • Not applicable. No AI model or training set is mentioned.
    Ask a Question

    Ask a specific question about this device

    K Number
    K212978

    Validate with FDA (Live)

    Date Cleared
    2021-11-10

    (54 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode laser therapy system is intended for hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.

    Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    The proposed device, Diode laser therapy system, is a surgical device. It utilizes a semiconductor diode as a laser source (808nm). The laser power is delivered to the treatment area via a laser handle. The emission laser is activated by handle and footswitch linkage.

    The treatment can be applied on different Fitzpatrick skin type, including I (White), II (White with pigment), III (Yellow), IV (Yellow with pigment), V (Brown) and VI (Black).

    AI/ML Overview

    The provided text describes a 510(k) submission for a medical device, the "Diode laser therapy system," which is intended for permanent hair reduction.

    Here's an analysis of the provided information regarding acceptance criteria and supporting studies:

    1. A table of acceptance criteria and the reported device performance

      The document does not explicitly present a table of acceptance criteria and reported device performance in the way a diagnostic AI would have with metrics like sensitivity, specificity, or AUC. Instead, the "acceptance criteria" are implied by the comparison to a predicate device (K210168, Diode Laser Therapy Systems by Beijing Kes Biology Technology CO., LTD.) and compliance with relevant non-clinical standards.

      The "performance" is primarily demonstrated through this substantial equivalence comparison and through non-clinical testing.

      Here's a summation of the comparison table provided, which implicitly serves as the performance assessment against the predicate:

      ItemAcceptance Criteria (Predicate K210168)Reported Device Performance (Proposed Device K212978)Remark
      Product CodeGEXGEXSame
      Regulation No.21 CFR 878.481021 CFR 878.4810Same
      ClassIIIISame
      Indications for UsePermanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Defined as long-term, stable reduction in hair regrowing at 6, 9, and 12 months after treatment.Permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Defined as long-term, stable reduction in hair regrowing at 6, 9, and 12 months after treatment.Same
      ConfigurationMain Unit, Handpiece, Foot ControlMain Unit, Handle, Foot ControlSame
      Laser TypeDiode LaserDiode LaserSame
      Laser ClassificationClass IVClass IVSame
      Laser Wavelength808 nm808 nmSame
      Spot Size12 mm × 12 mm = 1.44 cm²1.2 cm × 2.4 cm = 2.88 cm²Different
      Fluence10-125 J/cm²10-120 J/cm²Different
      Frequency1-10 Hz1-10 HzSame
      Pulse Duration10-400 ms10-300 msDifferent
      Power Supply99V-121V, 50/60Hz, 1400VAAC 100~230V/50/60Hz, 2000VADifferent
      Dimension450 × 430 × 1000 mm430 × 500 × 1030 mmDifferent
      Weight52 kg65 kgDifferent
      Patient contact materialAluminum alloy, ABSAluminum alloy, SapphireDifferent
      CytotoxicityNo Cytotoxicity (Comply with ISO 10993-5:2009)No Cytotoxicity (Tested according to ISO 10993-5:2009)Same
      SensitizationNo evidence of Sensitization (Comply with ISO 10993-10:2010)No evidence of Sensitization (Tested according to ISO 10993-10:2010)Same
      IrritationNo evidence of Irritation (Comply with ISO 10993-10:2010)No evidence of Irritation (Tested according to ISO 10993-10:2010)Same
      Electrical SafetyComply with ANSI/AAMI ES60601-1:2005, IEC 60601-2-22Comply with ANSI/AAMI ES60601-1:2005/(R)2012 and C1:2009/(R)2012 and A2:2010/(R) 2012 and IEC 60601-2-22:2012, IEC 60825-1:2014Same
      EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2:2014Same
      Software Validation & VerifN/A (implied by predicate clearance)Software verification & validation conducted as per FDA guidance for "Moderate" level of concern.N/A (New test)

      Explanation for "Different" remarks: The document states that despite the differences, non-clinical tests were conducted to demonstrate that these differences do not affect the safety and effectiveness of the proposed device. For example:

      • Spot Size: Non-clinical tests were done, and the results show the proposed device meets requirements.
      • Fluence: The proposed device's fluence range is within the predicate's range.
      • Pulse Duration: The proposed device's pulse duration range can be covered by the predicate's range.
      • Power Supply: Electrical safety and EMC tests were conducted, and the device works normally.
      • Dimension and Weight: These are physical specifications and do not raise safety/effectiveness issues.
      • Patient contact material: Biocompatibility tests met ISO 10993 requirements.
    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

      The document explicitly states: "No clinical study is included in this submission."
      Therefore, there is no clinical test set, sample size, or data provenance related to clinical performance. The evaluation is based on non-clinical testing and substantial equivalence to a predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

      Since no clinical study was performed, there was no clinical "test set" and thus no experts or ground truth established for clinical performance as would be the case for a diagnostic AI device.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

      Not applicable, as no clinical test set or human interpretation was involved.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      Not applicable. The device is a laser therapy system, not a diagnostic AI system that assists human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

      Not applicable. The device is a physical laser therapy system, not a standalone AI algorithm. It operates as a medical instrument for treatment.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

      For the non-clinical tests, the "ground truth" or reference was compliance with international standards (e.g., ISO 10993 for biocompatibility, IEC 60601 series for electrical safety and electromagnetic compatibility, IEC 60825-1 for laser safety) and demonstrating that parameter differences did not negatively impact safety and effectiveness compared to the predicate.

    8. The sample size for the training set

      Not applicable. This is not an AI/ML device that requires a training set in the conventional sense. The "training" for such a device would be its design and manufacturing process to meet specifications and standards.

    9. How the ground truth for the training set was established

      Not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210168

    Validate with FDA (Live)

    Date Cleared
    2021-05-10

    (108 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode Laser Therapy Systems are intended for hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.

    Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    The proposed device, Diode Laser Therapy System, is a surgical device. It utilizes a semiconductor diode as a laser source (808nm). The laser power is delivered to the treatment area via a laser handpiece. The emission laser is activated by a footswitch.

    The treatment can be applied on different Fitzpatrick skin type, including I (White), II (White with pigment), III (Yellow), IV (Yellow with pigment), V (Brown) and VI (Black); in addition, the treatment can also be applied to different parts of the body, such as arm, chest, leg, underarm etc.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called "Diode Laser Therapy Systems," which is intended for hair reduction. It details the device, its indications for use, and a comparison to predicate and reference devices.

    However, the document does not contain information about acceptance criteria, the study design, ground truth establishment, or performance metrics typically associated with proving a device meets acceptance criteria through clinical or standalone studies for an AI/algorithm-based medical device.

    Specifically, the document states: "No clinical study is included in this submission." This means there is no clinical performance data for this specific device in this submission to draw from to describe acceptance criteria or study results as requested in the prompt. The "study" mentioned in the prompt refers to a performance evaluation (often a clinical trial or a robust standalone study for AI devices).

    The comparison table (Table 1: Comparison of Technology Characteristics) on page 6 primarily focuses on technical specifications (e.g., laser wavelength, spot size, fluence, electrical safety) and regulatory compliance with standards like ISO and IEC, rather than clinical performance metrics or AI algorithm performance.

    Therefore, I cannot fulfill most of the request based on the provided text, as the information regarding clinical performance studies, acceptance criteria for such studies, and all related details (sample size, experts, ground truth, MRMC, standalone performance, training set details) is not present.

    The document concludes that the device is "Substantially Equivalent (SE)" based on non-clinical tests and technological comparisons to predicate devices, not on the basis of a clinical performance study with defined acceptance criteria and tested results.

    To answer your request based on the provided text, I must state that the information is absent.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192569

    Validate with FDA (Live)

    Date Cleared
    2019-12-13

    (86 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode Laser Therapy System is intended for hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.

    Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    The proposed device, Diode Laser, is a surgical device, which is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI);

    Function module description
    a. Power System
    b. Microprocessor Control System
    c. Operation Display System
    d. Cooling System
    e. Handpiece module

    AI/ML Overview

    The provided text is a 510(k) premarket notification for a medical device, specifically a "Diode Laser Therapy System" intended for hair reduction. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than providing extensive clinical study data to prove the device meets specific acceptance criteria in the way a new, novel device might.

    Therefore, the document explicitly states: "No clinical study is included in this submission." As such, there is no study described that proves the device meets specific acceptance criteria in terms of clinical performance metrics. The acceptance criteria for this submission are primarily focused on "substantial equivalence" to a predicate device based on technical characteristics and safety standards.

    However, I can extract the information relevant to the assessment requested, based on the non-clinical tests performed and the comparison to an existing predicate device.

    Here's the information structured as requested, noting the limitations due to the nature of a 510(k) submission:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria here are derived from the comparison to the predicate device and the adherence to recognized safety standards. Performance is indicated by matching the predicate's technical specifications and safety compliance.

    Acceptance Criterion (Based on Substantial Equivalence and Safety Standards)Reported Device Performance (Diode Laser Therapy System)
    Intended Use Equivalence: Hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin, with permanent reduction measured at 6, 9, and 12 months post-treatment.Meets: Intended use is identical to the predicate device.
    Product Code: GEX (Laser Surgical Instrument)Meets: Product Code is GEX.
    Regulation Number: 21 CFR 878.4810Meets: Regulation Number is 21 CFR 878.4810.
    Principle of Operation: Diode LaserMeets: Operated by Diode Laser.
    Laser Type: Diode LaserMeets: Diode Laser.
    Laser Classification: Class IVMeets: Class IV.
    Laser Wavelength: 808 nmMeets: 808 nm.
    Fluence Range: 5-40 J/cm²Meets: 5-40 J/cm². (It is explicitly stated "They are exactly same").
    Patient Contact Materials Biocompatibility: No cytotoxicity, no evidence of sensitization, no evidence of irritation from handpiece tip.Meets: Handpiece tip materials show no cytotoxicity, sensitization, or irritation, complying with ISO 10993-5 and ISO 10993-10.
    Electrical Safety: Compliance with IEC 60601-1 and IEC 60601-2-22.Meets: Complies with IEC 60601-1 and IEC 60601-2-22.
    EMC (Electromagnetic Compatibility): Compliance with IEC 60601-1-2.Meets: Complies with IEC 60601-1-2.
    Laser Safety: Compliance with IEC 60601-2-22 and IEC 60825-1.Meets: Complies with IEC 60601-2-22 and IEC 60825-1.
    Other Parameter Equivalence: (Spot Size, Pulse Duration, Frequency, Power Supply, Dimension, Weight)Discussion: Differences exist but are deemed not to affect safety and effectiveness, based on non-clinical testing and comparison to reference devices for safety margins (e.g., maximum pulse duration).

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: No clinical test set data is provided as "No clinical study is included in this submission." The "test set" for this 510(k) relies on the general comparison of technical specifications and safety test results against established standards and predicate devices.
    • Data Provenance: The data provenance for non-clinical tests (electrical safety, EMC, laser safety, biocompatibility) would be from laboratory testing performed on the Diode Laser Therapy System (LFS-K8) by San He Lefis Electronics Co., Ltd. The document does not specify the country of origin for these specific test results, but the manufacturer is based in China. These non-clinical tests are inherent to the device itself and are essentially prospective in the sense that they are performed on the device being submitted for clearance to demonstrate compliance with standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. As no clinical study was performed, there was no "test set" in the sense of patient data requiring expert ground truth establishment for clinical endpoints.

    4. Adjudication Method for the Test Set

    Not applicable. No clinical test set and no expert adjudication for clinical endpoints.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a laser therapy system, not an AI-powered diagnostic or assistive tool for human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a non-AI hardware medical device.

    7. The Type of Ground Truth Used

    The "ground truth" in this 510(k) submission relates to:

    • Established Standards: Ground truth for safety and performance parameters is defined by recognized international standards (e.g., IEC 60601 series, ISO 10993 series, IEC 60825-1).
    • Predicate Device Specifications: The technical specifications and established safety profile of the legally marketed predicate device (K181019) serve as a comparative ground truth for evaluating substantial equivalence.
    • There is no pathology, outcomes data, or expert consensus used for clinical ground truth in this submission because clinical studies were explicitly excluded.

    8. The Sample Size for the Training Set

    Not applicable. No AI/machine learning component requiring a training set. This is a physical medical device.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. No AI/machine learning component requiring a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191611

    Validate with FDA (Live)

    Date Cleared
    2019-07-31

    (44 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode Laser Therapy Systems are intended for hair reduction on all skin types (Fitzpatrick skin type I-VI).

    Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

    Device Description

    The Diode Laser Therapy Systems has one model WLA-01, which mainly consist of the main console, treatment hand piece and a foot switch. The treat hand piece consist of the laser aperture, laser emission indicator, laser emission and hand piece display screen. The main console consist of power switch, hand piece holder, LCD touch screen, emergency shut off, hand-grip grab bar, connector hand piece, access door, observation Window, power input, inlet, air switch, gate interlock connector, foot switch connector, vent, drain and fans. The device has an embedded software named Diode Laser Therapy Systems Control Software.

    The principle of laser hair removal is selective photothermolysis, 808nm wavelength effectively penetrates deep into and absorbed by the target chromophore. The adequate pulse duration, energy density and epidermal cooling ensure an adequate thermal damage to the target tissue without damaging the surrounding tissue to achieve effective hair removal.

    The laser window of the hand piece will contact the skin directly and the laser output will contact the skin through the window. The Diode Laser Therapy System is suitable for use in healthcare facility/hospital.

    AI/ML Overview

    Here's an analysis based on the provided text, outlining the acceptance criteria and study information for the Diode Laser Therapy Systems:

    1. Table of Acceptance Criteria and Reported Device Performance

    Criteria CategoryAcceptance Criteria (Set by device manufacturer or regulatory body)Reported Device Performance (as stated in the document)
    BiocompatibilityNo cytotoxic, sensitization, or irritation effects for tissue-contacting parts (laser aperture).The device successfully passed cytotoxicity, sensitization, and irritation tests according to FDA Blue Book Memorandum #G95-1 and ISO 10993-1. (The laser aperture is considered tissue contacting for less than 24 hours).
    Electrical Safety (ES)Compliance with ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text), IEC 60601-2-22 Edition 3.1 2012-10, and IEC 60825-1: 2014.The system complies with ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text), IEC 60601-2-22 Edition 3.1 2012-10, and IEC 60825-1: 2014 standards for safety.
    Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2:2014.The system complies with the IEC 60601-1-2:2014 standard for EMC.
    Software Verification & ValidationCompliance with FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" for a "moderate" level of concern software.Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered a "moderate" level of concern, indicating successfully met criteria for this risk level.
    Clinical Performance (Hair Reduction)(Implied by substantial equivalence) Achieve "permanent hair reduction," defined as a long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime, for all skin types (Fitzpatrick I-VI).No new clinical studies were conducted for this submission. The device's equivalency was established by comparing its technological characteristics (e.g., wavelength, energy density, pulse width, frequency, cooling system) and indications for use to legally marketed predicate devices (K152898 and K141063), which are presumed to have met this clinical performance standard. The document explicitly states: "no clinical studies were needed to support this 510(k) Premarket Notification."

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: Not applicable. The submission states, "no clinical studies were needed to support this 510(k) Premarket Notification." The performance data submitted was for non-clinical aspects (biocompatibility, electrical safety, EMC, software V&V). These tests were conducted on the device itself or its components.
    • Data Provenance: The origin of the non-clinical test data (biocompatibility, electrical safety, EMC, software V&V) is not specified as a country of origin in the provided text. It is assumed to be internal testing or conducted by accredited labs on behalf of the manufacturer, Wingderm Electro-Optics Ltd. (China). The data is described as "performance data were provided in support of the substantial equivalence determination," indicating it's retrospective relative to the 510(k) submission date.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable for the non-clinical tests.
    • For clinical performance (hair reduction), the ground truth was established by prior studies or general acceptance of the predicate devices. The document does not specify experts or their qualifications for the predicate devices' studies.

    4. Adjudication method for the test set

    • Not applicable for the non-clinical tests.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done. This device is a laser therapy system, not an AI-assisted diagnostic or interpretive device that would typically involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical medical device (laser system), not a standalone algorithm. Its software is embedded and controls device function, but it doesn't operate as an "algorithm only" in the sense of AI or diagnostic software.

    7. The type of ground truth used

    • For non-clinical tests (biocompatibility, electrical safety, EMC, software V&V): The ground truth is the adherence to established international and national standards (e.g., ISO 10993, ANSI AAMI ES60601-1, IEC 60601-1-2, FDA software guidance). Passing these tests constitutes meeting the "ground truth" for safety and performance in these areas.
    • For clinical performance (hair reduction): The ground truth for hair reduction efficacy is based on the performance of the predicate devices, which are legally marketed and presumed to achieve "permanent hair reduction" as defined. The substantial equivalence argument relies on the subject device operating on the same principles and having comparable technical specifications to achieve the same clinical outcome as the already-cleared predicates.

    8. The sample size for the training set

    • Not applicable. This device does not use machine learning or AI models that require a "training set" in the conventional sense. The "training" for the software would involve traditional software development and testing cycles, not data-driven machine learning.

    9. How the ground truth for the training set was established

    • Not applicable (as the device does not use a "training set" in the AI/ML context).
    Ask a Question

    Ask a specific question about this device

    K Number
    K152898

    Validate with FDA (Live)

    Date Cleared
    2016-02-09

    (131 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode Laser Therapy Systems are intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI). Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

    Device Description

    The Diode Laser Therapy Systems mainly consist of console, treatment hand piece and a foot switch. The diode laser emits wavelength of 808nm from the treatment hand piece, the wavelength effectively penetrates deep into and absorbed by the target chromophore. The adequate pulse duration, energy density and epidermal cooling ensure an adequate thermal damage to the target tissue without damaging the surrounding tissue to achieve effective hair removal.

    AI/ML Overview

    Based on the provided document, here's an analysis of the acceptance criteria and the study that proves the device meets them:

    Note: This document is a 510(k) summary for a medical device (Diode Laser Therapy Systems, Model SDL-B) and focuses on establishing substantial equivalence to a predicate device, not necessarily on demonstrating clinical efficacy through a traditional comparative effectiveness study with pre-defined acceptance criteria for performance metrics like sensitivity or specificity. The "acceptance criteria" here refer to meeting regulatory standards and demonstrating equivalence in function and safety.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/Standard MetReported Device Performance
    Intended Use EquivalenceTo be substantially equivalent to the predicate device for hair removal and permanent hair reduction on all skin types (Fitzpatrick I-VI), with a definition of permanent hair reduction as long-term, stable reduction measured at 6, 9, and 12 months post-treatment.The proposed device, Diode Laser Therapy Systems, has the same intended use as the predicate device (Emvera Diolux, K123257). Both are intended for hair removal and permanent hair reduction on all skin types (Fitzpatrick I-VI), with the same definition for permanent hair reduction.
    Technological EquivalenceDemonstrate similar technological characteristics or that differences do not raise new questions of safety or effectiveness.Wavelength: Same (808 nm) Fluence (Energy Density): Same (<120 J/cm²) Frequency: Same (1Hz-10Hz) Material: Same (Sapphire) Cooling: Same (Water cooling) Output Power: Proposed device (500W) is less than predicate (600W). Regarded as acceptable. Pulse Duration: Proposed device (5ms-400ms) has a smaller maximum pulse duration than predicate (5ms-625ms). Regarded as acceptable. Spot Size: Proposed device (12mm x 10mm) is smaller than predicate (12mm x 12mm). Regarded as acceptable. Anatomical Sites: Proposed device covers fewer anatomical sites (Axilla, Face, bikini, Upper limb and lower limb) than predicate (Axilla, Face, bikini, Upper limb and lower limb, brow, upper lip, neck, chest, back). Regarded as acceptable.
    BiocompatibilityCompliance with ISO 10993-1.Testing included Cytotoxicity, Sensitization, and Irritation for the laser output window. The evaluation was conducted in accordance with FDA Blue Book Memorandum #G95-1 and ISO 10993-1. Results are assumed to be compliant as the device was cleared.
    Electrical Safety & EMCCompliance with IEC 60601-1, IEC 60825-1, and IEC 60601-1-2.The system complies with IEC 60601-1 (safety), IEC 60825-1 (laser safety), and IEC 60601-1-2 (EMC).
    Software Verification & ValidationCompliance with FDA Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."Software V&V testing was conducted, and documentation was provided. The software was deemed a "moderate" level of concern. Results are assumed to be compliant as the device was cleared.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not report any clinical test set or data provenance for human subjects. The performance data presented are primarily non-clinical: biocompatibility testing, electrical safety and EMC testing, and software verification and validation. The claim of substantial equivalence is largely based on the device's technical specifications being similar to or safer than the predicate device.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable as no clinical test set with human subject data requiring expert ground truth establishment is described in this 510(k) summary. The "ground truth" for this submission revolves around meeting engineering and biocompatibility standards.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable as no clinical test set requiring adjudication is described.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not done. This 510(k) is for a laser therapy system (hardware device), not an AI/software as a medical device (SaMD) that assists human readers.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance study for an algorithm was not done. This device is a physical laser system for direct patient treatment, not an algorithm.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" in this context refers to established regulatory and engineering standards:

    • Biocompatibility: Adherence to ISO 10993-1 standards (e.g., negative results for cytotoxicity, sensitization, irritation).
    • Electrical Safety & EMC: Compliance with IEC 60601-1, IEC 60825-1, and IEC 60601-1-2 standards.
    • Software V&V: Adherence to FDA guidance for software in medical devices.
    • Substantial Equivalence: Comparison of technical specifications and intended use against a legally marketed predicate device.

    8. The sample size for the training set

    This information is not applicable as this submission does not describe an AI model with a training set. The "training" for this device would refer to its engineering design and validation against specified requirements.


    9. How the ground truth for the training set was established

    This information is not applicable as this submission does not describe an AI model with a training set. The "ground truth" for the device's development was established through engineering specifications, regulatory standards, and performance evaluation against those standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K102669

    Validate with FDA (Live)

    Date Cleared
    2010-11-17

    (62 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode Laser Therapy System (DenLase-810/7 & DenLase-980/7) is indicated for intraoral use for the following soft tissue applications:

    • Incision
    • · Excision
    • · Vaporization
    • · Ablation
    • · Coagulation.
    Device Description

    The DenLase is a surgical device designed with compactness, portability, reliability and user-friendliness. It provides the operator with a tool for surgical and cosmetic procedures on oral soft tissue. The Denlase utilizes a semiconductor diode with invisible infrared radiation as a laser source (810nm or 980nm). The laser power is delivered to the treatment area via a flexible fiber, which has a handpiece. The emission laser is activated by a footswitch. The Denlase adopts advanced semiconductor lasers (AIGAAs) and semiconductor refrigeration technology. By laser sent out by optical fiber, it can fast and accurately dispose tissue and lesion. The device includes three operation modes as continuous. single pulse and continuous pulse, and each operation mode is adjustable respectively by three parameter of emitting power, pulse width and pulse interval. The user can save total five groups of operation mode parameter combination under different mode.

    AI/ML Overview

    The provided text describes a 510(k) summary for a Diode Laser Therapy System, which is a medical device. This type of submission focuses on demonstrating substantial equivalence to a predicate device rather than fulfilling specific performance acceptance criteria through the kind of study described in the original request.

    Therefore, many of the requested elements (like acceptance criteria, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth methodologies) are not applicable or not explicitly detailed in this type of regulatory document for this specific device.

    The document primarily addresses non-clinical testing for safety and performance against established standards, and a comparison to predicate devices to establish substantial equivalence.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Safety and Performance Standards:The device is designed, tested, and manufactured in accordance with: - IEC 60825-1: Safety of laser products Part 1: Equipment classification, requirements and user's guide. - IEC 60601-2-22: Medical Electrical Equipment Part 2: Particular requirements for the safety of diagnostic and therapeutic laser equipment. - IEC 60601-1: Medical Electrical Equipment Part1: General requirements for safety. - IEC60601-1-2: Medical Electrical Equipment Part 1: General requirements for safety-2, Collateral Standard: Electromagnetic compatibility - Requirements and tests. - ISO 10993-5: Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity. - ISO 10993-10: Biological evaluation of medical devices Part 10: Tests for irritation and delayed-type hypersensitivity.
    Substantial Equivalence to Predicate Devices:The device "shares the similar indications for use, design features, functional features, same safety compliance" with predicate devices (Odyssey Navigator Diode Laser (K062258), KaVo GENTLEray 980 (K072262), Picasso (K083142), SiroLaser(K053161)). Differences in dimension, weight, duty cycle, and frequency of pulse are noted, but the document states these do not result in marked differences because the device meets the same safety and performance standards.

    2. Sample size used for the test set and the data provenance:

    • Not Applicable/Not Provided. The document states that "Laboratory testing was conducted to validate and verify that the proposed device... met all design specifications and was substantially equivalent to the predicate device. No Clinical Information is required." This indicates that the regulatory pathway did not necessitate a clinical study with a "test set" in the context of typical AI/diagnostic device validation. The testing focused on non-clinical performance against standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable/Not Provided. As no clinical study or specific "test set" requiring ground truth establishment by experts is mentioned, this information is not available.

    4. Adjudication method for the test set:

    • Not Applicable/Not Provided. See point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is a diode laser therapy system, not an AI-assisted diagnostic or interpretative device. Therefore, an MRMC study comparing human readers with and without AI assistance is irrelevant and was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. The device is a physical laser therapy system, not an algorithm. Its performance is inherent to its physical operation and energy delivery for therapeutic procedures, not an algorithmic output that could be tested "standalone."

    7. The type of ground truth used:

    • Not Applicable. For this device, the "ground truth" for non-clinical testing refers to compliance with engineering specifications and safety standards (e.g., laser emission parameters, biocompatibility, electrical safety), rather than a clinical ground truth like pathology or outcome data.

    8. The sample size for the training set:

    • Not Applicable/Not Provided. This is a hardware medical device, not an AI model requiring a "training set."

    9. How the ground truth for the training set was established:

    • Not Applicable/Not Provided. See point 8.