K Number
K192569

Validate with FDA (Live)

Date Cleared
2019-12-13

(86 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Diode Laser Therapy System is intended for hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

Device Description

The proposed device, Diode Laser, is a surgical device, which is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI);

Function module description
a. Power System
b. Microprocessor Control System
c. Operation Display System
d. Cooling System
e. Handpiece module

AI/ML Overview

The provided text is a 510(k) premarket notification for a medical device, specifically a "Diode Laser Therapy System" intended for hair reduction. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than providing extensive clinical study data to prove the device meets specific acceptance criteria in the way a new, novel device might.

Therefore, the document explicitly states: "No clinical study is included in this submission." As such, there is no study described that proves the device meets specific acceptance criteria in terms of clinical performance metrics. The acceptance criteria for this submission are primarily focused on "substantial equivalence" to a predicate device based on technical characteristics and safety standards.

However, I can extract the information relevant to the assessment requested, based on the non-clinical tests performed and the comparison to an existing predicate device.

Here's the information structured as requested, noting the limitations due to the nature of a 510(k) submission:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria here are derived from the comparison to the predicate device and the adherence to recognized safety standards. Performance is indicated by matching the predicate's technical specifications and safety compliance.

Acceptance Criterion (Based on Substantial Equivalence and Safety Standards)Reported Device Performance (Diode Laser Therapy System)
Intended Use Equivalence: Hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin, with permanent reduction measured at 6, 9, and 12 months post-treatment.Meets: Intended use is identical to the predicate device.
Product Code: GEX (Laser Surgical Instrument)Meets: Product Code is GEX.
Regulation Number: 21 CFR 878.4810Meets: Regulation Number is 21 CFR 878.4810.
Principle of Operation: Diode LaserMeets: Operated by Diode Laser.
Laser Type: Diode LaserMeets: Diode Laser.
Laser Classification: Class IVMeets: Class IV.
Laser Wavelength: 808 nmMeets: 808 nm.
Fluence Range: 5-40 J/cm²Meets: 5-40 J/cm². (It is explicitly stated "They are exactly same").
Patient Contact Materials Biocompatibility: No cytotoxicity, no evidence of sensitization, no evidence of irritation from handpiece tip.Meets: Handpiece tip materials show no cytotoxicity, sensitization, or irritation, complying with ISO 10993-5 and ISO 10993-10.
Electrical Safety: Compliance with IEC 60601-1 and IEC 60601-2-22.Meets: Complies with IEC 60601-1 and IEC 60601-2-22.
EMC (Electromagnetic Compatibility): Compliance with IEC 60601-1-2.Meets: Complies with IEC 60601-1-2.
Laser Safety: Compliance with IEC 60601-2-22 and IEC 60825-1.Meets: Complies with IEC 60601-2-22 and IEC 60825-1.
Other Parameter Equivalence: (Spot Size, Pulse Duration, Frequency, Power Supply, Dimension, Weight)Discussion: Differences exist but are deemed not to affect safety and effectiveness, based on non-clinical testing and comparison to reference devices for safety margins (e.g., maximum pulse duration).

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: No clinical test set data is provided as "No clinical study is included in this submission." The "test set" for this 510(k) relies on the general comparison of technical specifications and safety test results against established standards and predicate devices.
  • Data Provenance: The data provenance for non-clinical tests (electrical safety, EMC, laser safety, biocompatibility) would be from laboratory testing performed on the Diode Laser Therapy System (LFS-K8) by San He Lefis Electronics Co., Ltd. The document does not specify the country of origin for these specific test results, but the manufacturer is based in China. These non-clinical tests are inherent to the device itself and are essentially prospective in the sense that they are performed on the device being submitted for clearance to demonstrate compliance with standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. As no clinical study was performed, there was no "test set" in the sense of patient data requiring expert ground truth establishment for clinical endpoints.

4. Adjudication Method for the Test Set

Not applicable. No clinical test set and no expert adjudication for clinical endpoints.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a laser therapy system, not an AI-powered diagnostic or assistive tool for human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This is a non-AI hardware medical device.

7. The Type of Ground Truth Used

The "ground truth" in this 510(k) submission relates to:

  • Established Standards: Ground truth for safety and performance parameters is defined by recognized international standards (e.g., IEC 60601 series, ISO 10993 series, IEC 60825-1).
  • Predicate Device Specifications: The technical specifications and established safety profile of the legally marketed predicate device (K181019) serve as a comparative ground truth for evaluating substantial equivalence.
  • There is no pathology, outcomes data, or expert consensus used for clinical ground truth in this submission because clinical studies were explicitly excluded.

8. The Sample Size for the Training Set

Not applicable. No AI/machine learning component requiring a training set. This is a physical medical device.

9. How the Ground Truth for the Training Set was Established

Not applicable. No AI/machine learning component requiring a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 13, 2019

San He Lefis Electronics Co., Ltd. % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd. FangShan District Beijing, 102401 Cn

Re: K192569

Trade/Device Name: Diode Laser Therapy System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: September 16, 2019 Received: September 18, 2019

Dear Ray Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Purva Pandya Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K192569

Device Name Diode Laser Therapy System

Indications for Use (Describe)

The Diode Laser Therapy System is intended for hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: K 1 9 2 5 6 9

  • Date of Preparation 1.
    09/16/2019

  • Sponsor 2.

San He Lefis Electronics Co., Ltd.

No. 1 Production Room., ChaoBai Street, West Yiangbin Road, YanJiao Development Zone, Sanhe County, LangFang, HeBei Province, China 300000

Contact Person: Position: Ning Li Title: General Manager Assistant Tel: +86-18811392951 Fax: +86-316-3096027 Email: 1091118421@qq.com

    1. Submission Correspondent
      Mr. Ray Wang

Beijing Believe-Med Technology Service Co., Ltd.

Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, BeiJing, China 102401

Tel: +86-18910677558 Fax: +86-10-56335780 Email: ray.wang@believe-med.com

{4}------------------------------------------------

4. Identification of Proposed Device

Trade Name: Diode Laser Therapy System Common Name: Powered Laser Surgical Instrument Model(s): LFS-K8

Regulatory Information: Classification Name: Powered Laser Surgical Instrument Classification: II; Product Code: GEX ; Regulation Number: 21 CFR 878.4810; Review Panel: General & Plastic Surgery;

Indication For Use:

The Diode Laser Therapy System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

5. Device Description

The proposed device, Diode Laser, is a surgical device, which is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI);

Function module description

a. Power System

Power system including network power supply, laser power supply system, refrigeration power supply system and auxiliary power supply system, as shown in Fig 9-1.

The power lines and rocker switch are used to control the on -off of hair removal instrument with the outside power supply. The key switch, emergency stop switch are used to control the on-off of network power, and then through the auxiliary power supply, provide the work current for the microprocessor control system, touch screen operation display system and other functional components, meanwhile, the on-off of power system for refrigeration and laser are realized by the pick-up and switching off of the relay that controlled by micro - processor. The power-on, energizing and the light emission process of main unit are subject to microprocessor control, in order to achieve controlling the hardware with software through microprocessor, so that the appliance operation are more stable and reliable.

Laser power system adopt the power devices to constitute current regulation loop, thus can change the volume of output current through system regulation power devices of microprocessor, thereby

{5}------------------------------------------------

controlling the volume of output laser energy and work pulse width, through the feedback loop, export the current sampling signal and send to the micro - processor for real-time monitoring.

Various power supply components from the power system of laser hair removal instrument mostly adopt the safe and reliable switching power supply to achieve AC-DC conversion, these components passed through CE- certification, so that the safe effectiveness of the system get adequately guarantee, not only ensure the personal safety of users, but also ensures the long-term and reliable work of the devices itself.

The power system has the perfect self-protection function, it can continuously feed a number of important signal back to the microprocessor, once the important signal has an error, the microprocessor will promptly notify the users through the friendly man-appliance interface (Touch Screen) and take automatic protection treatment .After the error sigmal is released, the power system can automatically return to normal state.

b. Microprocessor Control System

Diode Laser device adopt the microprocessor control system. The microprocessor control system is composed of microprocessor, control loop and detection circuit, see Fig 9-1.

Control module of hair removal instrument is highly automated, the user just needs to turn on rocker switch and key switch and eject emergency stop switch, after power-on, the system will automatically perform the self-check for the control system (include all the preset checks against some possible damages to the appliance and the calling of treatment parameters and other a series of actions.

c. Operation Display System

Operation Display System is composed of touch screen, indicator and buzzer.

Among them, the touch screen connect directly to the host microprocessor, the operator can operate the touch screen to control the microprocessor to work, and through the information that is displayed on the screen to understand the main unit state and the relevant parameters, the hair removal instrument adopt the safe and reliable touch screen that passed through CE certification, also, the manufacturer designed a simple and intuitive operational interface, which can be easy to realize man-machine dialogue.

The indicators are used to display the work state of main unit; meanwhile, the buzzer is used as the prompt for laser output and fault alarm.

d. Cooling System

Cooling system consists of cooling fan, cooling water re-circulation system and water temperature regulation system.

After the laser hair removal instrument starting, the cooling system begins to work and adjust the water temperature to the preset temperature scale. The cooling fan give air cooling for the electric

{6}------------------------------------------------

parts and radiator in the main unit, through air flow in the appliance to take away the generated heat.

When the laser is running, internal circulating water can displace and takes away the heat and deliver to the water temperature regulation system, where the heat can be displaced and sent to the radiator for dispersion. Therefore, the circulation water temperature is maintained within a dynamic balance range, this ensures the normal working temperature of laser. By monitoring cooling water temperature in real time through temperature sampling, the control system can start the cooling components from the water temperature regulation system in due time to adjust water temperature, also, it can automatically judges whether the temperature is normal. If the temperature exceeds the early warning value that preset, the system will automatically disable laser output and give alarm so as to avoid damage to the laser.

e. Handpiece module

Handpiece module is the heart of the device, which is the execution unit of the device and completes the laser emission function. The semiconductor laser emits light to output energy.

{7}------------------------------------------------

    1. Identification of Predicate Device and Reference Device
      Primary Predicate Device 510(k) Number: K181019 Product Name: Diode Laser System Manufacturer: Guangzhou Huafei Tongda Technology Co., Ltd.

Reference Device 510(k) Number: K162659 Product Name: Diode Laser Hair Removal System Manufacturer: Shandong Huamei Technology Co.,ltd.

    1. Non-Clinical Test Conclusion
      Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
  • AAMI/ANIS/ES 60601-1:2012 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance;

  • IEC 60601-2-22 Edition 3.1 2012-10, Medical Electrical Equipment - Part 2-22: Particular Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment.

  • IEC 60825-1:2014, Safety of laser products - Part 1: Equipment classification and requirements.

  • IEC 60601-1-2:2014 , Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests

  • ISO 10993-5 Third Edition 2009-06-01, Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity. (Biocompatibility)

  • ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization. (Biocompatibility)

    1. Clinical Test Conclusion

No clinical study is included in this submission.

{8}------------------------------------------------

9. Substantially Equivalent (SE) Comparison

ItemProposed DevicePredicate DeviceReference DeviceRemark
Product CodeGEXK181019K162659SE
Regulation Number21 CFR 878.481021 CFR 878.481021 CFR 878.4810SE
Intended UseThe Diode Laser Therapy System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.The Diode Laser System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.The Diode Laser Hair Removal System is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.SE
ConfigurationMain UnitHandpieceFoot ControlMain UnitHandpieceFoot ControlMain UnitHandpieceFoot ControlSE
Principle of OperationDiode LaserDiode LaserDiode LaserSE

Table 7-1 General Comparison

{9}------------------------------------------------

ItemProposed DevicePredicate DeviceK181019Reference DeviceK162659Remark
Laser TypeDiode LaserDiode LaserDiode LaserSAME
Laser ClassificationClass IVClass IVClass IVSAME
Laser Wavelength808 nm808 nm808 nmSAME
Spot Size$2.80\text{cm}^2$$1.20\text{cm}^2$$1.44 \text{ cm}^2$Discussion
Pulse Duration3ms-400ms30ms-200ms5-400msDiscussion
Fluence2-40 J/ $\text{cm}^2$5-40 J/ $\text{cm}^2$1-120J/ $\text{ cm}^2$SAME
Frequency1-10Hz1-5Hz0.5-15HzDiscussion
Power Supply110V~50/60Hz100-240V~ 50/60HzAC 110V/60HzDiscussion
Dimension510mm×600mm×510mm×600mm×450mm×Discussion
1000mm1000mm550mm×380mm
Weight45Kg50Kg52KgDiscussion

Table 7-2 Performance Comparison

Discussion

The proposed device has same indication for use with predicate devices, the main differences are output parameters, such as spot size, frequency range, pulse duration.

For these output parameters, the Fluence and Pulse duration are most important parameters which may decides that how much and how long time the energy will deliver to the patient's skin, it may effects the safety (too much energy and/or too long action time) may burn patient's skin, and effectiveness (too low energy and/or too short action time) may make the device could not achieve it's indication for use.

From the comparison above, the fluence of proposed device is 5-40 J/cm2, the fluence of predicate device are 5-40 J/cm2. They are exactly same, that's means the proposed device has capable to achieve it's indication for use.

The Pulse duration of proposed device is 3 ms -400 ms, which is difference with the predicate device, but which is same with the reference device (K162659), since the reference device has more longer fluence (120 J/cm2), then the 400 ms maximum pulse with 40 J/cm2 of proposed device could be consider as safety.

For the other differences (Spot Size, Frequency, Power Supply, Dimension and Weight), they would not affects the safety and effectiveness of proposed device for it's indications for use. And By complying with non-clinical test conducted, the proposed device is determined to be substantially equivalency with predicate device.

{10}------------------------------------------------

ItemProposed DevicePredicate DeviceK162659Reference DeviceK162659Remark
Patient Contact Materials and Biocompatibility
Patient ContactMaterialsHandpiece tipHandpiece tipHandpiece tipSAME
CytotoxicityNo CytotoxicityNo CytotoxicityNo Cytotoxicity
SensitizationNo evidence of sensitizationNo evidence of sensitizationNo evidence of sensitizationSAME
IrritationNo evidence of irritationNo evidence of irritationNo evidence of irritation
EMC, Electrical and Laser Safety
ElectricalSafetyComply with IEC 60601-1, IEC 60601-2-22Comply with IEC 60601-1, IEC 60601-2-22Comply with IEC 60601-1, IEC 60601-2-22SAME
EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2Comply with IEC 60601-1-2SAME
Laser SafetyComply with IEC 60601-2-22, IEC 60825Comply with IEC 60601-2-22, IEC 60825Comply with IEC 60601-2-22, IEC 60825SAME

Table 7-3 Safety Comparison

10. Substantially Equivalent (SE) Conclusion

Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.