Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K212978
    Date Cleared
    2021-11-10

    (54 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode laser therapy system is intended for hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.

    Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    The proposed device, Diode laser therapy system, is a surgical device. It utilizes a semiconductor diode as a laser source (808nm). The laser power is delivered to the treatment area via a laser handle. The emission laser is activated by handle and footswitch linkage.

    The treatment can be applied on different Fitzpatrick skin type, including I (White), II (White with pigment), III (Yellow), IV (Yellow with pigment), V (Brown) and VI (Black).

    AI/ML Overview

    The provided text describes a 510(k) submission for a medical device, the "Diode laser therapy system," which is intended for permanent hair reduction.

    Here's an analysis of the provided information regarding acceptance criteria and supporting studies:

    1. A table of acceptance criteria and the reported device performance

      The document does not explicitly present a table of acceptance criteria and reported device performance in the way a diagnostic AI would have with metrics like sensitivity, specificity, or AUC. Instead, the "acceptance criteria" are implied by the comparison to a predicate device (K210168, Diode Laser Therapy Systems by Beijing Kes Biology Technology CO., LTD.) and compliance with relevant non-clinical standards.

      The "performance" is primarily demonstrated through this substantial equivalence comparison and through non-clinical testing.

      Here's a summation of the comparison table provided, which implicitly serves as the performance assessment against the predicate:

      ItemAcceptance Criteria (Predicate K210168)Reported Device Performance (Proposed Device K212978)Remark
      Product CodeGEXGEXSame
      Regulation No.21 CFR 878.481021 CFR 878.4810Same
      ClassIIIISame
      Indications for UsePermanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Defined as long-term, stable reduction in hair regrowing at 6, 9, and 12 months after treatment.Permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Defined as long-term, stable reduction in hair regrowing at 6, 9, and 12 months after treatment.Same
      ConfigurationMain Unit, Handpiece, Foot ControlMain Unit, Handle, Foot ControlSame
      Laser TypeDiode LaserDiode LaserSame
      Laser ClassificationClass IVClass IVSame
      Laser Wavelength808 nm808 nmSame
      Spot Size12 mm × 12 mm = 1.44 cm²1.2 cm × 2.4 cm = 2.88 cm²Different
      Fluence10-125 J/cm²10-120 J/cm²Different
      Frequency1-10 Hz1-10 HzSame
      Pulse Duration10-400 ms10-300 msDifferent
      Power Supply99V-121V, 50/60Hz, 1400VAAC 100~230V/50/60Hz, 2000VADifferent
      Dimension450 × 430 × 1000 mm430 × 500 × 1030 mmDifferent
      Weight52 kg65 kgDifferent
      Patient contact materialAluminum alloy, ABSAluminum alloy, SapphireDifferent
      CytotoxicityNo Cytotoxicity (Comply with ISO 10993-5:2009)No Cytotoxicity (Tested according to ISO 10993-5:2009)Same
      SensitizationNo evidence of Sensitization (Comply with ISO 10993-10:2010)No evidence of Sensitization (Tested according to ISO 10993-10:2010)Same
      IrritationNo evidence of Irritation (Comply with ISO 10993-10:2010)No evidence of Irritation (Tested according to ISO 10993-10:2010)Same
      Electrical SafetyComply with ANSI/AAMI ES60601-1:2005, IEC 60601-2-22Comply with ANSI/AAMI ES60601-1:2005/(R)2012 and C1:2009/(R)2012 and A2:2010/(R) 2012 and IEC 60601-2-22:2012, IEC 60825-1:2014Same
      EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2:2014Same
      Software Validation & VerifN/A (implied by predicate clearance)Software verification & validation conducted as per FDA guidance for "Moderate" level of concern.N/A (New test)

      Explanation for "Different" remarks: The document states that despite the differences, non-clinical tests were conducted to demonstrate that these differences do not affect the safety and effectiveness of the proposed device. For example:

      • Spot Size: Non-clinical tests were done, and the results show the proposed device meets requirements.
      • Fluence: The proposed device's fluence range is within the predicate's range.
      • Pulse Duration: The proposed device's pulse duration range can be covered by the predicate's range.
      • Power Supply: Electrical safety and EMC tests were conducted, and the device works normally.
      • Dimension and Weight: These are physical specifications and do not raise safety/effectiveness issues.
      • Patient contact material: Biocompatibility tests met ISO 10993 requirements.
    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

      The document explicitly states: "No clinical study is included in this submission."
      Therefore, there is no clinical test set, sample size, or data provenance related to clinical performance. The evaluation is based on non-clinical testing and substantial equivalence to a predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

      Since no clinical study was performed, there was no clinical "test set" and thus no experts or ground truth established for clinical performance as would be the case for a diagnostic AI device.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

      Not applicable, as no clinical test set or human interpretation was involved.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      Not applicable. The device is a laser therapy system, not a diagnostic AI system that assists human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

      Not applicable. The device is a physical laser therapy system, not a standalone AI algorithm. It operates as a medical instrument for treatment.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

      For the non-clinical tests, the "ground truth" or reference was compliance with international standards (e.g., ISO 10993 for biocompatibility, IEC 60601 series for electrical safety and electromagnetic compatibility, IEC 60825-1 for laser safety) and demonstrating that parameter differences did not negatively impact safety and effectiveness compared to the predicate.

    8. The sample size for the training set

      Not applicable. This is not an AI/ML device that requires a training set in the conventional sense. The "training" for such a device would be its design and manufacturing process to meet specifications and standards.

    9. How the ground truth for the training set was established

      Not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1