K Number
K220381

Validate with FDA (Live)

Date Cleared
2022-05-20

(99 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Diode Laser Therapy Systems is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

Device Description

Diode Laser Therapy Systems is a surgical device, which is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI); The Diode Laser Therapy Systems utilize a semiconductor diode with invisible infrared radiation as a laser source to emit 808nm wavelength laser which is absorbed by melanin. The laser power is delivered to the treatment area via laser handpiece. The emission laser is activated by a foot-switch. The Diode Laser Therapy Systems utilize the principle of photoepilation for hair removal. The Photoepilation is a technique for removal of unwanted hair by thermal destruction of the hair follicle and its reproductive system (stems cells). The heat is caused by selective absorption of electromagnetic radiation emitted by laser light sources. As melanin is the main chromophore existing in hair follicles, so the melanin could absorb the energy from the laser, which would result in temperature rapid increase, then the hair follicle and its reproductive system (stems cells) would be destroyed by increased high temperature without damage epidermis and the surrounding normal tissue.

AI/ML Overview

The provided text is a 510(k) summary for a medical device (Diode Laser Therapy Systems) and does not contain information about acceptance criteria and a study proving the device meets those criteria, as typically found in clinical study reports with specific performance metrics (e.g., sensitivity, specificity, accuracy).

Instead, this document focuses on demonstrating substantial equivalence to a predicate device based on non-clinical testing and general safety standards.

Therefore, many of the requested fields cannot be directly populated from the provided text. However, I can extract information related to the device's technical specifications and the non-clinical tests performed to support its safety and performance.

Here's a breakdown of what can and cannot be answered based on the provided document:

1. A table of acceptance criteria and the reported device performance

The document does not specify formal "acceptance criteria" in terms of clinical performance metrics (like sensitivity, specificity, etc.) that the device must meet. Instead, the acceptance is based on demonstrating substantial equivalence to a predicate device through:

  • Matching intended use and fundamental technology.
  • Complying with recognized safety and performance standards (IEC, ISO).
  • Showing that any differences in technical specifications do not raise new questions of safety or effectiveness.
Criteria Type (based on substantial equivalence)Acceptance/Comparison GoalProposed Device Performance/Finding
Intended UseSame as predicateHair removal, permanent hair reduction on all skin types (Fitzpatrick I-VI), including tanned skin.
Regulatory InformationSame Classification, Regulation, Product Code as predicateClass II, 21 CFR 878.4810, GEX, General & Plastic Surgery Review Panel.
Laser TypeSame as predicate (Diode Laser)Diode Laser
Laser ClassificationSame as predicate (Class IV)Class IV
Laser WavelengthSame as predicate (808nm)808nm ± 2nm
FluenceSame as predicate (1-100 J/cm²)1-100 J/cm²
EMC, Electrical & Laser SafetyComply with relevant IEC standards (60601-1, 60601-1-2, 60601-2-22, 60825)Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1.
BiocompatibilityComply with ISO 10993 series (Cytotoxicity, Sensitization, Irritation) for patient contact materials.No Cytotoxicity, No evidence of sensitization, No evidence of irritation (for 6061 Aluminum, Sapphire, ABS). Complies with ISO 10993-5, ISO 10993-10.
Spot SizeNon-clinical tests show differences do not affect therapeutic effect and meet requirements.15mm × 15mm (different from predicate 1.44 cm² but tested to meet requirements).
FrequencyWithin predicate’s range; bench tests support substantial equivalence and no new safety issues.1-10Hz (compared to predicate's 1-20 Hz, within range).
Pulse DurationSlight difference, within predicate’s range, no new safety issues, bench tests support substantial equivalence.1-300ms (compared to predicate's 10-400ms, slight difference deemed safe).
Power SupplyComplies with IEC 60601-1 and 60601-1-2, no effect on effectiveness/safety.110 Va.c. 60Hz 2000W (different from predicate but tested safe).
Dimension/WeightComplies with IEC 60601-1 for mechanical performance, no effect on substantial equivalence.Different from predicate (460x390x1230mm, 75Kg vs 560x380x1180mm, 60Kg). Tested to be mechanically safe.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • The document explicitly states: "No clinical study is included in this submission."
  • Therefore, there is no "test set" in the context of clinical data, no sample size, and no data provenance relevant to clinical performance. The testing mentioned (non-clinical) refers to bench testing on the device hardware and materials.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. No clinical test set with ground truth established by experts is mentioned.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. No clinical test set.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is a laser therapy system, not an AI-assisted diagnostic or imaging device. No MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This device is not an algorithm. Performance was assessed through non-clinical bench testing of the physical device.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • For the non-clinical tests, the "ground truth" or reference for compliance is established by international standards (IEC, ISO) for electrical safety, laser safety, electromagnetic compatibility, and biocompatibility, as well as the technical specifications of the predicate device.

8. The sample size for the training set

  • Not applicable. No AI model or training set is mentioned as this is a physical medical device.

9. How the ground truth for the training set was established

  • Not applicable. No AI model or training set is mentioned.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 20, 2022

Beijing LaserTell Medical Co., Ltd. Mr. Ray Wang Official Correspondent Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District Beijing, Beijing 102401 China

Re: K220381

Trade/Device Name: Diode Laser Therapy Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: April 18, 2022 Received: April 22, 2022

Dear Ray Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K220381

Device Name Diode Laser Therapy Systems

Indications for Use (Describe)

The Diode Laser Therapy Systems is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

The assigned 510(k) Number: K220381

510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

    1. Date of Preparation:2022/05/18
    1. Sponsor Identification

Beijing LaserTell Medical Co., Ltd.

Block1, No.12 Jingsheng South 2nd Road, JQ Science Park, Tongzhou District, Beijing, China.

Contact Person: Zeng Xun Position: General Manager Tel: +86 13810934930 Fax: +86 10 56370036 Email: alex(@lasertell.com

    1. Designated Submission Correspondent
      Mr. Ray Wang

Beijing Believe-Med Technology Service Co., Ltd.

Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, 102401, China

Tel: +86-18910677558 Fax: +86-10-56335780 Email: information(@believe-med.com

{4}------------------------------------------------

4. Identification of Proposed Device

Trade Name: Diode Laser Therapy Systems Common Name: Powered Laser Surgical Instrument Model(s): AlexMED Pro

Regulatory Information: Classification Name: Powered Laser Surgical Instrument Classification: II; Product Code: GEX; Regulation Number: 21 CFR 878.4810; Review Panel: General & Plastic Surgery;

Indication For Use Statement:

The Diode Laser Therapy Systems is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.

Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

న్. Device Description

Diode Laser Therapy Systems is a surgical device, which is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI);

The Diode Laser Therapy Systems utilize a semiconductor diode with invisible infrared radiation as a laser source to emit 808nm wavelength laser which is absorbed by melanin. The laser power is delivered to the treatment area via laser handpiece. The emission laser is activated by a foot-switch.

The Diode Laser Therapy Systems utilize the principle of photoepilation for hair removal. The Photoepilation is a technique for removal of unwanted hair by thermal destruction of the hair follicle and its reproductive system (stems cells). The heat is caused by selective absorption of electromagnetic radiation emitted by laser light sources. As melanin is the main chromophore existing in hair follicles, so the melanin could absorb the energy from the laser, which would result in temperature rapid increase, then the hair follicle and its reproductive system (stems cells) would be destroyed by increased high temperature without damage epidermis and the surrounding normal tissue.

    1. Identification of Predicate Device(s)
      510(k) Number: K180353 Product Name: Diode Laser Hair Removal Device Manufacturer: Zhengzhou PZ Laser Slim Technology Co., Ltd

{5}------------------------------------------------

7. Substantially Equivalent (SE) Comparison

ItemProposed DevicePredicate DeviceRemark
Device NameDiode Laser Therapy SystemsDiode Laser Hair Removal / Device
ClassificationRegulation21 CFR 878.481021 CFR 878.4810SAME
Classification PanelGeneral & PlasticSurgeryGeneral & PlasticSurgerySAME
ClassIIIISAME
Product CodeGEXGEXSAME
Indication for useThe Diode Laser Therapy Systems is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.The Diode Laser Hair Removal device is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.SAME

Table 7-1 General Comparison

Table 7-2 Performance Comparison

ITEMProposed DevicePredicate DeviceRemark
Laser TypeDiode LaserDiode LaserSAME
LaserClassificationClass IVClass IVSAME
LaserWavelength808nm±2nm808nmSAME
Spot Size15mm×15mm1.44cm²Analysis 1
Fluence1-100J/cm²1-100J/cm²SAME
Frequency1-10Hz1-20 HzAnalysis 2
Pulse Duration1-300ms10~400msAnalysis 3
Power Supply110 Va.c.60Hz2000WAC 110V-230V/50-60Hz2000VAAnalysis 4
Dimension460mm×390mm×1230mm560mm×380mm×1180mmAnalysis 5
Weight75Kg60KgAnalysis 5

{6}------------------------------------------------

Analvsis :

Analysis 1: #Spot size

The proposed device is different in Spot Size from the predicate, Spot size only affects the area of treatment, not affect the therapeutic effect. Non-clinical tests about the spot size have been conducted on the proposed device and the results show that the proposed device meets the requirements. Therefore, this difference will not affect the substantially equivalency.

Analysis 2: #Frequency

The frequency for the proposed device is different from the predicate device. However, the frequency of the proposed device is within the range of that of the predicate device, which can justify that the difference in the parameter of frequency will not raise new safety issues of the proposed device. And the bench tests conducted on the proposed device is same with the predicate device, the results of which could support the substantially equivalency with predicate device.

Analysis 3: #Pulse Duration

The proposed device only has slight difference in pulse duration with the predicate device. The frequency of the proposed device is within the range of that of the predicate device, which can justify that the difference in the parameter of Pulse duration will not raise new safety issues of the proposed device. So the slight difference is considered to have no effectiveness and safety. And the bench tests conducted on the proposed device is same with the predicate device, the results of which could support the substantially equivalency with predicate device.

Analysis 4: # Power Supply

The Power Supply of proposed device is different with the predicate device, but the proposed device has been conducted as IEC 60601-1 and IEC 60601-1-2, the results shown that, there is no effect the effectiveness and safety.

Analysis 5: #Dimension/ Weight

The proposed device is different in dimension and weight from the predicate device. By complying with IEC 60601-1, the mechanical performance of the proposed device is determined to be accepted, therefore, this difference will not affect the substantially equivalency.

ItemProposed DevicePredicate DeviceRemark
EMC, Electrical and Laser Safety

Table 7-3 Safety Comparison

{7}------------------------------------------------

Electrical SafetyComply with IEC 60601-1, IEC 60601-2-22Comply with IEC 60601-1, IEC 60601-2-22SAME
EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2SAME
Laser SafetyComply with IEC 60601-2-22, IEC 60825Comply with IEC 60601-2-22, IEC 60825SAME
Patient Contact Materials and Biocompatibility
Patient Indirect Contact MaterialsTip of Handle (6061 Aluminum & Sapphire)Plastic shell (Acrylonitrile Butadiene Styrene)SapphireAnalysis 6
CytotoxicityNo CytotoxicityNo CytotoxicitySAME
SensitizationNo evidence of sensitizationNo evidence of sensitizationSAME
IrritationNo evidence of irritationNo evidence of irritationSAME

Analysis 6 # Patient Indirect Contact Materials

The patient contact material for the proposed device is different from the predicate device. However, biocompatibility test has been conducted on the proposed device and the test result can meet the requirements of ISO 10993 series standard. Therefore, this difference will not affect safety and effectiveness of the proposed device.

Non-Clinical Test Conclusion 8.

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

IEC 60601-1-2:2014 Medical Electrical Equipment-Part 1-2: General Requirements For Basic Safety And Essential Performance Collateral Standard: Electromagnetic Compatibility-Requirements And Tests

IEC 60601- 1:2005+CORR.1:2006+CORR.2:2007+A1:2012 Medical Electrical Equipment - Part 1: General Requirements for Safety

IEC 60601-2-22:2007 + A1:2012 +A2:2019 Medical Electrical Equipment - Part 2: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment.

IEC 60825-1:2014 (Third Edition) Safety of laser products - Part 1: Equipment classification, and requirements

ISO 10993-5: 2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity

ISO 10993-10: 2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

{8}------------------------------------------------

9. Clinical Testing

No clinical study is included in this submission.

  • Conclusion 10.
    The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, the Diode Laser Therapy Systems (AlexMED Pro) is as safe, as effective, and performs as well as or better than the legally marketed predicate device, Diode Laser Hair Removal Device cleared under K180353.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.