Search Filters

Search Results

Found 44 results

510(k) Data Aggregation

    K Number
    K250159

    Validate with FDA (Live)

    Date Cleared
    2025-10-17

    (269 days)

    Product Code
    Regulation Number
    866.5510
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Immunoglobulin Isotypes (GAM) for the EXENT Analyser:

    The Immunoglobulin Isotypes (GAM) for the EXENT Analyser is a MALDI-TOF mass spectrometry immunoassay that is used in conjunction with the Binding Site Optilite IgG, IgA and IgM assays for the semi-quantitative in vitro measurement of monoclonal IgG, IgA, and IgM as a reflex test in serum for patients with a result suggestive of the presence of monoclonal immunoglobulins by serum protein electrophoresis (gel or capillary zone electrophoresis), or with an abnormal serum free light chain concentration and free light chain ratio result.

    The assay is intended for use as an aid in the evaluation of monoclonal gammopathy of undetermined significance (MGUS); and as an aid in the diagnosis of smouldering multiple myeloma (SMM), multiple myeloma (MM), Waldenström's macroglobulinaemia, and AL amyloidosis.

    Assay results should be used in conjunction with other laboratory and clinical findings.

    EXENT Analyser:

    The EXENT Analyser is an automated analyser intended for the qualitative and quantitative in vitro measurement of analytes in human body fluids used in conjunction with the EXENT assays. The device is designed for professional in vitro diagnostic use only and it is not a device for self-testing.

    Device Description

    The system consisting of the EXENT® Analyser and the EXENT® assays are intended for the in vitro measurements of analytes in human body fluids. It is designed to provide automation and integration of all the analytical steps (including liquid handling and MALDI-ToF mass spectrometry). The EXENT Analyser is designed to be used solely in combination with EXENT assays to measure a variety of analytes depending on the reagents. The device is designed for professional use only and it is not a device for self-testing.

    The EXENT Analyser combines automated immunoassay with readout by MALDI-ToF mass spectrometry. It is a modular analyser, and the major components are described in Table 3 and illustrated schematically in Figure 1.

    ComponentDescriptionFunction
    EXENT-iP® 500Automated liquid handlerPreparation of patient samples by magnetic bead immunoprecipitation assays for subsequent analysis by MALDI-ToF Mass spectrometry
    EXENT-iX® 500MALDI-ToF mass spectrometerAnalysis of prepared patient samples by MALDI-ToF mass spectrometry
    EXENT-iQ® softwareWorkflow and data management softwareManagement of the workflow between the EXENT-iP500 and EXENT-iX500 instruments. Data management including processing and results release.

    The EXENT-iP500 component is an automated liquid handler that prepares human body fluids using the EXENT assay specific reagents. The samples are prepared using magnetic beads that are coated with isotype-specific antibodies. Any unbound material is washed away during the sample preparation process. The EXENT-iP500 also manages the transfer of the prepared patient sample to the MALDI plate.

    The EXENT-iX500 component is a MALDI-ToF mass spectrometer. Signals are produced by ionizing the compound or biological material under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass-to-charge ratios. The EXENT-iX500 is used to read samples prepared by the EXENT-iP500.

    The EXENT-iQ software integrates sample preparation and MALDI-ToF mass spectrometry and is used for data storage and processing. It is the primary user interface used by the user to review and release results.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250549

    Validate with FDA (Live)

    Date Cleared
    2025-05-23

    (87 days)

    Product Code
    Regulation Number
    866.5550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Optilite® Freelite Mx Kappa Free Kit:
    The Optilite Freelite Mx Kappa Free Kit is intended for the quantitative in vitro measurement of Kappa free light chains in serum and urine using the Binding Site Optilite analyser. Measurement of free light chains in serum and urine aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenström's macroglobulinaemia, AL amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus (SLE); and in serum aids in the evaluation of monoclonal gammopathy of undetermined significance (MGUS). Results of the free light chain measurements should always be interpreted in conjunction with other laboratory and clinical findings.

    Optilite® Freelite Mx Lambda Free Kit:
    The Optilite Freelite Mx Lambda Free Kit is intended for the quantitative in vitro measurement of Lambda free light chains in serum and urine using the Binding Site Optilite analyser. Measurement of free light chains in serum and urine aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenström's macroglobulinaemia, AL amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus (SLE); and in serum aids in the evaluation of monoclonal gammopathy of undetermined significance (MGUS). Results of the free light chain measurements should always be interpreted in conjunction with other laboratory and clinical findings.

    Device Description

    The determination of soluble antigen concentration by turbidimetric methods involves the reaction with specific antiserum to form insoluble complexes. When light is passed through the suspension formed a portion of the light is transmitted and focused onto a photodiode by an optical lens system. The amount of transmitted light is indirectly proportional to the specific protein concentration in the test sample. Concentrations are automatically calculated by reference to a calibration curve stored within the instrument.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and its associated 510(k) Summary pertains to the Optilite Freelite Mx Kappa Free Kit and Optilite Freelite Mx Lambda Free Kit. The submission sought to add a claim for the evaluation of Monoclonal Gammopathy of Undetermined Significance (MGUS) to the intended use statement (in serum).

    Here's an analysis of the acceptance criteria and the study proving the device meets these criteria, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly list acceptance criteria in terms of pre-defined numerical thresholds for sensitivity, specificity, or positive/negative rates that were required to be met for the MGUS claim. Instead, it describes clinical performance studies designed to demonstrate the utility of the device for MGUS evaluation. The conclusion states that "The studies generated successful results where all pre-defined acceptance criteria were met," implying these criteria were internal to the manufacturer's study design and not explicitly detailed in the public document as numerical targets.

    However, we can infer the de facto "reported device performance" based on the results of Study 1 and Study 2.

    Inferred Performance/Results:

    Performance MetricFreelite Mx Kappa & Lambda Kits (for MGUS evaluation)Commentary
    Study 1: Determination of Positive Rate Ratio in MGUSThis study aimed to show the "positive rate" of the test in clinically confirmed MGUS patients. The "positive rate" here refers to the percentage of MGUS patients whose FLC ratio was abnormal based on the device's reference intervals. It is not a traditional sensitivity as it's not classifying all MGUS cases, but rather indicating how often the FLC ratio is abnormal within the MGUS population.
    All MGUS positive56.3% (n=129/N=229)Implies that 56.3% of confirmed MGUS patients had an abnormal FLC ratio according to the device's criteria.
    Light chain MGUS100.0% (n=10/N=10)Shows that the device identified all LC-MGUS cases as "abnormal" (test-positive). This is a strong indicator for this specific subtype.
    Non-light chain only MGUS54.3% (n=119/N=219)Shows the positive rate for MGUS cases that are not LC-MGUS.
    Study 1: Determination of Negative Rate Ratio in Disease Controls (non-MGUS)This study aimed to show the "negative rate" of the test in non-MGUS patients with other conditions. The "negative rate" here refers to the percentage of non-MGUS patients whose FLC ratio was within the normal reference interval. It is not a traditional specificity as it's not classifying all non-MGUS cases as negative, but rather indicating how often the FLC ratio is normal in a disease control group.
    Disease Controls (Non-MGUS)91.9% (n=125/N=136)Implies that 91.9% of patients with polyclonal hypergammaglobulinemia (non-MGUS disease controls) had a normal FLC ratio.
    Study 2: Evaluation of MGUS Progression (Stable vs. Progressive)This study retrospectively evaluated the ability of the device to track stable versus progressive MGUS based on FLC changes. The acceptance criteria here would likely revolve around the consistency of FLC results with the clinical status (stable or progressive).
    Stable MGUS test positive patients95.6% (n=43/N=45)This indicates that 95.6% of patients with clinically stable MGUS showed FLC results consistent with stability (i.e., less than 25% increase in involved FLC).
    Progressive MGUS test positive patients50.0% (n=2/N=4)This indicates that 50.0% of patients with clinically progressive MGUS (converting to MM) showed FLC results consistent with progression (i.e., FLC ratio outside reference interval AND >= 25% increase in involved FLC from baseline). The small sample size (N=4) for progressive MGUS patients should be noted for this metric.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Study 1 (Evaluation of MGUS and disease controls at single time points):

      • MGUS Test Set: 229 samples from patients with clinically confirmed MGUS.
      • Disease Control Test Set (non-MGUS): 136 samples from patients with polyclonal hypergammaglobulinemia.
      • Provenance: Retrospective study. The country of origin is not explicitly stated, but the submitter (The Binding Site Ltd) is based in the United Kingdom.
    • Study 2 (Evaluation of MGUS progression):

      • Total Patients: 49 MGUS patients.
        • Stable MGUS Cohort: 45 patients.
        • Progressive MGUS Cohort: 4 patients (who progressed from MGUS to MM).
      • Total Samples: 185 samples (up to 4 individual draws for stable, up to 6 for progressive).
      • Provenance: Retrospective study. Country of origin not explicitly stated.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • The document states that the clinical diagnostic criteria for MGUS and disease controls were "confirmed with each site" and "as practiced clinically."
    • For the progressive MGUS study, changes were defined as clinical progression from MGUS to MM.
    • No specific number of experts or their qualifications (e.g., "Radiologist with 10 years of experience") are mentioned in the provided text. The ground truth relies on "clinical diagnosis" or "clinical truth," which typically implies the consensus opinion of treating clinicians or established diagnostic criteria within the medical community (e.g., IMWG guidelines for MGUS/MM).

    4. Adjudication Method for the Test Set

    • The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for establishing the clinical ground truth. It relies on "clinical diagnosis" or "clinical truth" established by the sites providing the samples.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC study was done. This device is an in vitro diagnostic (IVD) kit for quantitative measurement of biomarkers, not an image-based AI device that would typically involve human readers interpreting images. The closest analog would be a clinical utility study comparing outcomes with and without the FLC test, but that is beyond the scope of a 510(k) for an IVD kit unless explicitly requested for a novel claim.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • This is a standalone study. The Optilite Freelite Mx Kappa and Lambda Free Kits are laboratory assays. The "performance data" presented (positive rates, negative rates, tracking of progression) are derived solely from the measurements made by the device (Optilite Analyser) on patient samples, compared against established clinical diagnoses (ground truth). There is no "human-in-the-loop" aspect to the device's direct measurement and result generation. The results are then interpreted by clinicians in conjunction with other findings, but the device's output itself is standalone.

    7. Type of Ground Truth Used

    • The ground truth used was clinical diagnosis/clinical truth.
      • For Study 1: "clinically confirmed MGUS" and "polyclonal hypergammaglobulinemia confirmed by study testing... with supporting clinical information." The diagnostic criteria for MGUS were stated to align with or be similar to those outlined by the International Myeloma Working Group (IMWG).
      • For Study 2: "clinically determined stable or progressive status" for MGUS patients, with progression defined as conversion from MGUS to Multiple Myeloma (MM) based on clinical diagnosis. FLC results criteria adapted from IMWG guidelines were used for evaluation of the device's performance, but the patient's status (stable/progressive) was the "clinical diagnosis."

    8. Sample Size for the Training Set

    • The document states, "The devices in this submission have not materially changed since originally cleared under K150658." This implies that the core analytical performance studies (e.g., analytical measuring range, precision, accuracy) were previously established and referenced from prior 510(k) submissions (K173732 and K150658).
    • For the new MGUS claim, the performance data presented are for clinical validation on the test sets described in points 1 and 2.
    • The document does not describe a distinct "training set" for the clinical claim of MGUS evaluation, nor is it typical for IVD kits to have a "training set" in the machine learning sense for their clinical claims. The performance studies demonstrate the device's ability to measure FLCs in patient populations relevant to MGUS, and these measures are compared against clinical ground truth. The assay itself (reagents, instrument) would have been "trained" (i.e., optimized and validated analytically) during its initial development and previous clearances.

    9. How the Ground Truth for the Training Set Was Established

    • As a "training set" for the clinical claim is not explicitly mentioned or relevant for this type of IVD 510(k), this question is not directly applicable. For the device itself, analytical validation and calibration would have established its operational parameters, but this isn't "ground truth" in the diagnostic performance sense. The clinical ground truth for the test sets was derived from established clinical diagnoses and diagnostic criteria.
    Ask a Question

    Ask a specific question about this device

    K Number
    K192116

    Validate with FDA (Live)

    Date Cleared
    2019-09-04

    (29 days)

    Product Code
    Regulation Number
    866.5510
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This kit is intended for the quantitative in vitro determination of human IgA in serum, lithium heparin or EDTA plasma, using the Binding Site SPAPLUS turbidimetric analyser. Measurement of IgA aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents. The test results are to be used in conjunction with other clinical and laboratory findings.

    Device Description

    Human IqA liquid reagent kit for use on SPAPLUS® comprises the following reagents:

    Antiserum: Monospecific goat anti IgA supplied in stabilised liquid form. lt contains 0.099% sodium azide, 0.1% E-amino-n-caproic acid (EACA), 0.5% BSA and 0.01% benzamidine as preservatives.

    Calibrator and Controls: These consist of pooled human serum and are supplied in stabilised liquid form. The concentration of IgA given on the quality control certificate has been obtained by comparison with European Reference Material ERM-DA470k. They contain 0.099% sodium azide, 0.1% EACA and 0.01% benzamidine as preservatives.

    Reaction Buffer: Containing 0.099% sodium azide as a preservative.

    AI/ML Overview

    The provided document describes the "Human IgA liquid reagent kit for use on SPAPLUS" by The Binding Site Group Ltd. This device is intended for the quantitative in vitro determination of human IgA in serum, lithium heparin, or EDTA plasma using the Binding Site SPAPLUS turbidimetric analyzer. The submission is a special 510(k) for a modification to an existing device (K103824), specifically a change in the source of the detection antibody from sheep to goat.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present acceptance criteria in a dedicated table format with corresponding reported device performance for all aspects. Instead, acceptance criteria are described within the context of each study and then a conclusion about meeting those criteria is stated. I will extract and present this information in a table format where possible.

    Test Parameter / CharacteristicAcceptance Criteria (Implicit/Explicit)Reported Device PerformanceConclusion on Meeting Criteria
    Precision (Repeatability & Within Lab)No explicit numerical acceptance criteria given, but the expectation is "no change in performance compared to the device cleared in K103824".CV% ranges from 1.7% to 7.1%.Do not indicate any change in performance compared to K103824.
    Precision (Between Instrument)No explicit numerical acceptance criteria given, but the expectation is "no change in performance compared to the device cleared in K103824".CV% ranges from 0.0% to 1.8%.Do not indicate any change in performance compared to K103824.
    Precision (Between Lot)No explicit numerical acceptance criteria given, but the expectation is "no change in performance compared to the device cleared in K103824".CV% ranges from 0.0% to 3.2%.Do not indicate any change in performance compared to K103824.
    Linearity/Assay Reportable RangeNo explicit numerical acceptance criteria given, but the expectation is "comparable to those currently presented in the product insert" and "no change in performance compared to K103824".Linear regression equation: y=0.993x - 0.230 g/L with an R value of 0.996.Comparable to product insert, do not indicate any change in performance compared to K103824.
    Accelerated Kit StabilityMaximum allowable difference of ±15% to verify the stability claim of 18 months.All controls, internal reference, and samples for two batches (455406 and 455407) passed, achieving an equivalent stability of 561 days at 4ºC, which exceeds the required 395 days (18 months).Pass for all tested parameters and batches.
    On-Board StabilityNo explicit numerical acceptance criteria given, but the expectation is "no difference in the cleared on-board stability claim".Studies showed no difference in the cleared on-board stability claim.No difference observed.
    Limit of Quantitation (LoQ)Allowable CV of 8%.LoQ claim was validated by all samples reporting within the acceptance criteria.LoQ claim validated.
    Limit of Detection (LoD) & Limit of Blank (LoB)No explicit numerical acceptance criteria given, but the expectation is "no change in performance compared to the device cleared in K103824".LoD estimated as 0.003 g/L, LoB as 0.001 g/L. No change in performance observed after antisera change.Do not indicate any change in performance compared to K103824.
    Method Comparison with Predicate Device (Bland Altman)No explicit numerical acceptance criteria given, but the expectation is "no change in performance compared to K103824".Mean Bias: -2.18%, 95% Limits of Agreement: 0.55% to 3.17%.Do not indicate any change in performance compared to K103824.
    Method Comparison with Predicate Device (Passing Bablok)No explicit numerical acceptance criteria given, but the expectation is "no change in performance compared to K103824".Equation: y=1.017x + 0.002, Slope 95% CI: 1.004 to 1.029, Intercept 95% CI: -0.029 to 0.026. Correlation coefficient: 0.998.Do not indicate any change in performance compared to K103824.
    Expected Values/Reference Range Transfer≤2 samples falling outside of the limits of the reference interval to be transferred.All 20 samples tested gave results within the reference interval (1.553 to 4.840 g/L).Meets acceptance criteria, reference interval can be transferred.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Precision Studies:
      • Repeatability and Within Laboratory: 4 different samples, 80 data points for each level (total 320 data points).
      • Between Instrument: 4 different samples, 24 data points for each level (total 96 data points).
      • Between Lot: 4 different samples, 24 data points for each level (total 96 data points).
    • Linearity Study: A high pool (8.19g/L) and a low pool (0.12 g/L) were used to create a dilution series. Each diluted sample was tested in 3 replicates.
    • Accelerated Stability: 6 replicates of controls, internal reference, and 3 different samples were tested.
    • Limit of Quantitation (LoQ) Validation: 4 samples were tested using two reagent lots.
    • Method Comparison with Predicate Device: 102 serum samples and 42 plasma samples (total 144 samples).
    • Expected Values/Reference Range Transfer: 20 samples from "apparently healthy US donors".

    Data Provenance:

    • For the Expected Values/Reference Range Transfer study, samples were from "apparently healthy US donors".
    • For other studies, the country of origin or whether the data was retrospective or prospective is not explicitly stated. The studies were carried out by The Binding Site Group Ltd., which is based in the UK, so it's likely testing was conducted there unless otherwise specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This product is an in vitro diagnostic (IVD) reagent kit for quantitative Immunoglobulin A (IgA) determination. The "ground truth" in this context refers to the accurate measurement of IgA concentration. The document mentions traceability to European Reference Material ERM-DA470k/IFCC for calibration. This implies that the standard itself serves as the "ground truth" reference, rather than independent expert consensus on clinical interpretation. There are no mentions of experts establishing ground truth in the sense of clinical diagnosis or image interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Adjudication methods like 2+1 or 3+1 are typically used in studies involving subjective interpretation by multiple human readers (e.g., in radiology studies) to resolve discrepancies and establish a consensus "ground truth." Since this device is a quantitative IVD assay and its performance is validated against analytical standards, reference materials, and comparative measurements, such adjudication methods are not applicable and therefore not used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic tools where human readers (e.g., radiologists) use AI assistance to improve their diagnostic accuracy. This device is a standalone quantitative laboratory test kit, not an AI or imaging diagnostic aid, so an MRMC study is not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the performance studies described are for the standalone device (reagent kit on the SPAPLUS analyzer) without human-in-the-loop performance influencing the measurement part of the device's function. The analytical performance, stability, and comparison studies evaluate the kit's ability to accurately measure IgA concentrations. The device's output is a quantitative value, not an interpretation of data that requires human feedback for performance assessment.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for this quantitative assay is primarily based on:

    • Reference Materials: Calibration traceability to ERM-DA470k/IFCC. This is the international standard for immunoglobulin measurements.
    • Reference Methods: The predicate device itself (K103824) serves as a comparative "ground truth" for method comparison studies, demonstrating substantial equivalence.
    • Defined Concentrations: For linearity and precision studies, samples with known or precisely diluted concentrations are used.
    • Clinical Reference Intervals: The ability to transfer established reference intervals (based on a healthy population) is also part of the "ground truth" for clinical usability.

    8. The sample size for the training set

    This document describes a special 510(k) submission for a modification to an existing device, which mostly involves re-validation studies to ensure the modification (change in antibody source from sheep to goat) does not alter performance compared to the cleared predicate. It does not describe the development of a novel algorithm or AI model that requires a distinct "training set." The studies performed are validation studies, not training. Therefore, a "training set" in the context of machine learning is not applicable here.

    9. How the ground truth for the training set was established

    As there is no "training set" in the machine learning sense for this device submission, this question is not applicable. The device's underlying principles (immunoturbidimetry) are well-established, and its performance is validated against analytical standards rather than learned from a dataset.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191985

    Validate with FDA (Live)

    Device Name
    Optilite IgA Kit
    Date Cleared
    2019-08-19

    (25 days)

    Product Code
    Regulation Number
    866.5510
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Optilite IgA Kit is intended for the quantitative in vitro measurement of IgA in serum, lithium heparin or EDTA plasma using the Binding Site Optilite analyser. Measurement of IgA aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents. This test should be used in conjunction with other laboratory and clinical findings.

    Device Description

    The Optilite IgA Kit comprises the following reagents: Antiserum: Goat anti IgA supplied in stabilised liquid form. Preservatives: 0.099% sodium azide, 0.1% E-amino-n-caproic acid (EACA), 0.5% BSA and 0.01% benzamidine. Calibrator and Controls: Pooled human serum, supplied in stabilised liquid form. Contain 0.099% sodium azide, 0.1% EACA and 0.01% benzamidine as preservatives. The concentration given on the quality control certificate has been obtained by comparison with the DA470k international reference material. Reaction Buffer: Containing 0.099% sodium azide as a preservative.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study detailed in the provided document for the Optilite IgA Kit, organized according to your requested information.

    It's important to note that this document is a 510(k) summary for a modification to an existing device (Optilite IgA Kit, K103824). Therefore, the study presented here primarily focuses on confirming that the modified device (with a change from sheep to goat antibody) performs comparably to the predicate device (the original cleared kit) and that the performance characteristics detailed in the original submission still hold true. This is not a study to establish initial performance characteristics, but rather to demonstrate equivalence after a modification.


    Acceptance Criteria and Device Performance Study for Optilite IgA Kit (K191985)

    This study was conducted to demonstrate that the modified Optilite IgA Kit, with a change in the source of the detection antibody from sheep to goat, maintains comparable performance to the predicate device (Optilite IgA Kit, K103824). The acceptance criteria were primarily based on demonstrating no significant change in performance compared to the previously cleared device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (Relative to Predicate Device's Performance Claims)Reported Device Performance (Modified Kit)Conclusion
    PrecisionNo change in performance compared to predicate. Data should support existing precision claims in product insert.Repeatability & Within Lab: Consistent low CV% (1.0-6.3%) across 5 levels. Between Instrument: Low CV% (0.0-2.9%) across 5 levels. Between Lot: Low CV% (0.8-5.2%) across 5 levels.Meets Criteria: The results do not indicate any change in performance compared to the cleared device (K103824).
    Linearity/Assay Reportable RangeNo change in performance compared to predicate. Data should support existing linearity claims in product insert.Linear regression equation: y=1.007x + 0.159 g/L with R value of 0.999.Meets Criteria: Results are comparable to those presented in the original product insert, indicating no change in performance.
    Kit Stability (Accelerated)Verified stability in accordance with ISO 23640:2015, with maximum allowable difference of ±15% compared to baseline. Should support 18-month stability claim.All tested parameters (IR, Controls, Samples 1, 2, 3) passed the accelerated stability criteria, achieving or exceeding the required stability days (e.g., sample 1 achieved 561 equivalent days at 4°C, required 395 days).Meets Criteria: All parameters passed, verifying the stability claim.
    Detection Limit (LoD, LoQ, LoB)No change in performance compared to predicate. Data should support existing claims in product insert.LoQ validated at 0.02 g/L (within 8% CV acceptance criteria). LoD estimated at 0.007 g/L; LoB estimated at 0.005 g/L. No change observed after antisera change.Meets Criteria: No change in performance observed, supporting existing LoB, LoD, and LoQ claims.
    Method Comparison (vs. Predicate)Bland Altman Mean Bias close to 0%, 95% Limits of Agreement indicating good concordance. Passing Bablok slope close to 1, intercept close to 0. High Correlation Coefficient. No indication of change in performance vs. predicate.Bland Altman Mean Bias: 2.38%, 95% Limits of Agreement: -10.6% to 15.37%. Passing Bablok: y=1.012x + 0.011 (Slope 95% CI: 1.001 to 1.027, Intercept 95% CI: -0.006 to 0.044). Correlation coefficient: 0.998.Meets Criteria: Results do not indicate any change in performance compared to the cleared device (K103824).
    Reference Range Transfer≤2 samples falling outside of the limits of the original reference interval from 20 tested samples.19 out of 20 samples gave results within the reference interval (0.947 to 4.043 g/L). One sample was at the lower boundary (<0.021 g/L).Meets Criteria: Only 1 sample fell outside the quantitative range for a healthy individual, indicating successful transferability of the reference interval.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Precision (Repeatability & Within Laboratory): 5 different samples, N=80 per level (total 400 replicates).
    • Precision (Between Instrument): 5 different samples, N=24 per level (total 120 replicates).
    • Precision (Between Lot): 5 different samples, N=24 per level (total 120 replicates).
    • Linearity: High pool (8.59 g/L) and low pool (0.10 g/L) for dilution series. Each diluted sample tested in 3 replicates.
    • Kit Stability (Accelerated): 6 replicates of controls, internal reference, and samples.
    • Detection Limit (LoQ): 4 samples tested using 2 reagent lots.
    • Method Comparison: 102 serum samples, 42 plasma samples (total 144 samples). The document does not specify county of origin, but it is implied to be for US market. The study appears to be prospective as it's conducted to support a new submission for a modified device.
    • Reference Range Transfer: 20 samples from apparently healthy US donors.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of in-vitro diagnostic device (measuring Immunoglobulin A levels in blood samples) does not typically involve human expert readers or a "ground truth" derived from expert consensus in the same way an imaging AI device would. The "ground truth" for the performance characteristics (precision, linearity, limits, comparison) is derived from:

    • Reference materials: Traceability to ERM-DA470k/IFCC for calibration.
    • Established analytical methods: Results from the predicate device serve as the comparative "reference."
    • Pre-defined specifications: Acceptance criteria for analytical performance established based on regulatory guidelines (e.g., CLSI standards) and the performance of the predicate device.

    Therefore, there is no mention of experts establishing ground truth for the test set in the conventional sense of human readers for an AI imaging study.

    4. Adjudication Method for the Test Set

    Not applicable for this type of in-vitro diagnostic analytical performance study. Adjudication by human experts is typically performed in clinical studies involving interpretation of data (e.g., imagery, patient symptoms) where subjective judgment might be involved. Here, the measurements are quantitative and based on instrument readings.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size

    No, an MRMC study was not done. This is an in-vitro diagnostic device that quantifies an analyte (IgA) in blood samples. MRMC studies are specific to imaging devices and human interpretation. This device does not involve human readers in its direct use or interpretation of results beyond a lab technician operating the instrument and a clinician interpreting the quantitative IgA values.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the analytical performance studies (precision, linearity, detection limits, method comparison) represent the "standalone" performance of the analytical instrument and reagents. The device provides a direct quantitative measurement, so there isn't a human-in-the-loop directly assisting the measurement process itself, only interpreting the final numerical result.

    7. The Type of Ground Truth Used

    The ground truth for this device is based on:

    • Quantitative Reference Materials: The calibration is traceable to an international reference material (ERM-DA470k/IFCC).
    • Comparative Performance to Predicate Device: The performance of the predicate device (Optilite IgA Kit, K103824) serves as the established "truth" against which the modified device is compared to prove substantial equivalence.
    • Analytical Standards: Established methods and guidelines (e.g., CLSI standards) define how analytical accuracy, precision, and other parameters are measured and what constitutes acceptable performance.
    • Known Concentrations in Quality Control (QC) Materials: Pooled human sera with known concentrations are used for controls and calibrators, providing a known value against which measurements are assessed.

    8. The Sample Size for the Training Set

    This document describes a pre-market notification for a modified in-vitro diagnostic device; as such, it does not detail a "training set" in the context of machine learning or AI algorithm development. The "training" for such a system would typically refer to the initial development and optimization of the assay reagents and instrument parameters. The studies presented here are primarily verification and validation studies to demonstrate that the performance of the modified device is equivalent to the predicate. The specifics of the original assay development/optimization (which could be considered analogous to "training") are not provided here, as this is a 510(k) for a modification.

    9. How the Ground Truth for the Training Set was Established

    Not applicable in the context of this 510(k) for a modified in-vitro diagnostic device. As explained above, there isn't a "training set" in the machine learning sense. The "ground truth" for the development of the original assay would have involved rigorous biochemical and analytical characterization of the antibodies and their reaction with IgA, optimization of assay conditions (e.g., reagent concentrations, incubation times), and calibration against international reference standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191635

    Validate with FDA (Live)

    Device Name
    Optilite IgM Kit
    Date Cleared
    2019-07-15

    (26 days)

    Product Code
    Regulation Number
    866.5510
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Optilite IgM Kit is intended for the quantitative in vitro measurement of IgM in human serum, lithium heparin or EDTA plasma using the Binding Site Optilite analyser. Measurement of IgM aids in the diagnosis of abnomal protein metabolism and the body's lack of ability to resist infectious agents. The test results are to be used in conjunction with other clinical and laboratory findings.

    Device Description

    The Optilite IgM Kit comprises the following reagents: Antiserum: Goat anti IgM supplied in stabilised liquid form. Preservatives: 0.099% sodium azide, 0.1% E-amino-n-caproic acid (EACA), 0.5% BSA and 0.01% benzamidine. Calibrator and Controls: Pooled human serum, supplied in stabilised liquid form. Contain 0.099% sodium azide, 0.1% EACA and 0.01% benzamidine as preservatives. The concentration given on the quality control certificate has been obtained by comparison with the DA470k international reference material. Reaction Buffer: Containing 0.099% sodium azide as a preservative.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Optilite IgM Kit, which is a quantitative in vitro measurement system for IgM in human serum, lithium heparin, or EDTA plasma. The submission is a modification to an existing device (K082129). The core of the submission revolves around demonstrating that the modified device maintains performance comparable to the previously cleared device, rather than proving initial efficacy or diagnostic accuracy.

    Based on the provided text, here's an analysis of the acceptance criteria and study proving the device meets them:

    Key Takeaway: This 510(k) submission is for a modification to an existing device (change of antibody source and buffer). Therefore, the "acceptance criteria" and "study" are primarily focused on demonstrating that this modification does not negatively impact the performance compared to the original, already cleared device. There is no new clinical effectiveness study to establish diagnostic accuracy against a disease state from scratch.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally framed as "no change in performance compared to the device cleared in K082129" or meeting specific pre-defined quantitative thresholds (e.g., allowable CV, limits of agreement).

    Acceptance Criteria CategorySpecific Acceptance Criteria (implicit/explicit)Reported Device Performance and Conclusion
    Precision/ReproducibilityNo significant change in repeatability, within-laboratory, between-instrument, and between-lot precision compared to the predicate (K082129). (Implicit, demonstrated through comparison of observed CVs and SDs to historical or expected performance, as guided by CLSI EP5-A3). For example, meeting specific CV targets for different levels.Repeatability and Within Laboratory: • Level 1: Total CV 4.9% • Level 5: Total CV 7.0%. Between Instrument: • Level 1: CV 5.2% • Level 5: CV 6.9%. Between Lot: • Level 1: CV 0.7% • Level 5: CV 1.5%. Conclusion: "The above results do not indicate any change in performance compared to the device cleared in K082129. The precision claims in the product insert therefore still accurately represent the performance of the modified kit and do not need to be amended."
    Linearity/Assay Reportable RangeNo significant change in linearity compared to the predicate (K082129). Demonstrated by strong correlation coefficient (r value) of the linear regression. (Implicit, based on historical performance).Linear Regression Equation: y = 1.029 x – 0.1752 g/L r value: 0.998. Conclusion: "These results are comparable to those currently presented in the product insert and therefore do not indicate any change in performance compared to the device cleared in K082129. The linearity claims in the product insert therefore still accurately represent the performance of the modified kit and do not need to be amended."
    Kit Stability (Accelerated)Stability claim of 12 months (365 days) at 4°C verified with a maximum allowable difference of ±15% (per EP25-A).Achieved Stability (equivalent at 4°C): • IR: 507 days • Low Control: 531 days • High Control: 878 days • Samples 1-3: 561 days. Decision: All "Pass." Conclusion: Verified stability claim. (Real-time study ongoing for further support).
    Detection Limit (LoQ, LoD, LoB)LoQ validated by all samples reporting within an allowable CV of 8%. No significant change in LoD or LoB from the predicate (0.01 g/L and 0.007 g/L respectively). (Implicit, based on historical performance as per EP17-A2).LoQ: All tested samples were within the acceptance criteria of allowable CV of 8%. LoD: No change in performance observed after antisera change. LoB: No change in performance observed after antisera change. Conclusion: "The results generated do not indicate any change in performance compared to the device cleared in K082129. The LoB, LoD and LoQ claims in the product insert therefore still accurately represent the performance of the modified kit and do not need to be amended."
    Method Comparison with Predicate DeviceAgreement between the modified and predicate device, evidenced by low bias, tight limits of agreement, high correlation coefficient, and appropriate Passing Bablok regression results to demonstrate equivalence.Bland Altman Mean Bias: -2.13% 95% Limits of Agreement: -9.63% to 5.36% Passing Bablok: y= 0.9775x + 0.009 (Slope 95% CI: 0.971 to 0.983; Intercept 95% CI: 0.001 to 0.016) Correlation coefficient: 0.999. Conclusion: "The above results do not indicate any change in performance compared to the device cleared in K082129. The comparison claims in the product insert therefore still accurately represent the performance of the modified kit and do not need to be amended."
    Expected Values/Reference RangeTransferred reference interval accepted if ≤2 samples fall outside the limits of the reference interval.Of 20 samples, 18 gave results within the reference interval (0.35 – 2.42 g/L). Two samples were slightly below the lower boundary (0.305 g/L and 0.319 g/L). Conclusion: "The results of this study therefore meet the acceptance criteria and indicate that the reference interval can be transferred from the originally cleared device."

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Precision Studies (Repeatability and Within Laboratory):
      • Sample Size: 5 different samples. For each level, N=80 for Repeatability/Within Lab Summary (20 working days * 2 runs/day * 2 reps/run).
      • Data Provenance: Not explicitly stated, but typically internal lab data (e.g., from the manufacturer's R&D or QC labs).
    • Precision Studies (Between Instrument):
      • Sample Size: 5 different samples. For each level, N=24 (6 working days * 2 runs/day * 2 reps/run).
      • Data Provenance: Not explicitly stated, typically internal lab data.
    • Precision Studies (Between Lot):
      • Sample Size: 5 different samples. For each level, N=24 (6 working days * 2 runs/day * 2 reps/run).
      • Data Provenance: Not explicitly stated, typically internal lab data.
    • Linearity Study:
      • Sample Size: Dilution series from a high pool (8.44 g/L) and a low pool (0.17 g/L). Each diluted sample tested in 3 replicates.
      • Data Provenance: Not explicitly stated, typically internal lab data.
    • Kit Stability (Accelerated):
      • Sample Size: 6 replicates of controls, internal reference, and samples.
      • Data Provenance: Not explicitly stated, typically internal lab data.
    • Detection Limit (LoQ) Study:
      • Sample Size: Four samples.
      • Data Provenance: Not explicitly stated, typically internal lab data.
    • Method Comparison with Predicate Device:
      • Sample Size: 120 serum samples and 60 plasma samples (total 180 samples).
      • Data Provenance: Not explicitly stated, but typically clinical samples or commercial samples purchased for the study. Given the medical context, these would be human samples. The study was conducted as per pre-submission meeting Q171503, implying a prospective design for this specific comparative testing.
    • Expected Values/Reference Range:
      • Sample Size: 20 samples from apparently healthy US donors.
      • Data Provenance: Prospective collection of samples from US donors.

    Overall Data Provenance: The document explicitly mentions "20 samples from apparently healthy US donors" for the reference range study. For other analytical performance studies (precision, linearity, stability, detection limit), the provenance is not specified, but these are typically conducted in-house by the manufacturer (retrospective analysis vs. new data collection is not detailed, but likely new data generation for the modified kit).


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This device is an in vitro diagnostic (IVD) for quantitative measurement of IgM. The "ground truth" for IVDs of this nature is established by the reference measurement procedure or the assigned/target values of calibrators and controls used in the analytical performance studies, not by clinical expert consensus in the way a diagnostic imaging AI might establish ground truth from radiologists.

    • For Analytical Studies (Precision, Linearity, LoD/LoQ, Stability): The ground truth for these samples are their known or reference concentrations, which are determined by highly accurate laboratory methods or traceable standards (e.g., ERM-DA470k/IFCC for IgM). No clinical experts are involved in establishing this type of ground truth.
    • For Method Comparison: The "ground truth" in this context is the result obtained from the predicate device (the unmodified K082129 kit), as the goal is to show agreement between the modified and unmodified kit.
    • For Reference Range Study: The "ground truth" is the established reference interval for IgM in a healthy population. The study assessed whether the modified kit's results for healthy individuals fell within this known range.

    Therefore, the concept of "experts establishing ground truth" as it applies to reader studies for imaging devices does not directly apply here.


    4. Adjudication Method for the Test Set

    Given that this is an IVD kit for quantitative measurement and not an imaging device requiring human interpretation, there is no mention of an adjudication method among multiple human readers. The analytical performance is evaluated against quantitative measurements, established standards, or the performance of the predicate device.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. MRMC studies are specific to diagnostic performance, typically for imaging devices where multiple readers interpret cases with and without AI assistance. This submission is for an IVD kit that provides a quantitative measurement, not an interpretative diagnosis requiring human readers in the loop. The "comparison" done was a method comparison between the modified kit and the predicate kit.


    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, in essence. The "device" here is the Optilite IgM Kit used on the Optilite analyser. Its performance is evaluated analytically (precision, linearity, limits, stability) and comparatively against its predicate. This is a standalone performance assessment of the kit, which automatically provides quantitative results. There is no human interpretative step that the kit is assisting or replacing within its intended use.


    7. The Type of Ground Truth Used

    • For Analytical Performance (Precision, Linearity, LoD/LoQ, Stability):
      • Traceability: ERM-DA470k/IFCC (a certified reference material for immunoglobulins). This is a highly robust form of ground truth based on international standards.
      • Known Concentrations: Samples and controls with established or assigned target values.
    • For Method Comparison:
      • Predicate Device Results: The results from the previously cleared Optilite IgM Kit (K082129) served as the reference for comparison to demonstrate equivalence.
    • For Reference Range Study:
      • Established Reference Interval: The pre-determined reference range (0.35 – 2.42 g/L) for IgM in a healthy population served as the ground truth against which results from healthy donors were compared.

    There is no pathology confirmation or outcomes data mentioned, as these are typically used for assessing clinical accuracy or impact on patient management, which is beyond the scope of this 510(k) for a modified quantitative IVD.


    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of an AI/ML algorithm. This submission is for an immunoturbidimetric test kit, not an AI-powered diagnostic system. The "training" of such a system would refer to the development and optimization of the reagent formulation and methods, which involves extensive R&D and internal testing, often with many samples. However, the document provided focuses on the validation testing required for regulatory submission.

    The section M. Performance Characteristics details the validation experiments for the modified kit. These are the test sets to prove performance, not training sets.


    9. How the Ground Truth for the Training Set Was Established

    As noted above, there is no "training set" in the AI/ML sense described. For the development of the original device and its subsequent modifications, the "ground truth" would be established through:

    • Reference materials: Calibrating the assay against international reference standards (e.g., ERM-DA470k/IFCC).
    • Internal validation: Testing against well-characterized in-house samples and controls with known concentrations.
    • Method optimization: Iteratively adjusting reagent concentrations and reaction parameters to achieve desired analytical performance (sensitivity, specificity, precision, linearity) based on these known samples and reference materials.

    The provided document focuses on the verification and validation studies for a pre-defined modified product, demonstrating its performance against regulatory requirements, rather than the developmental phase of the original or modified kit's creation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191465

    Validate with FDA (Live)

    Date Cleared
    2019-06-27

    (24 days)

    Product Code
    Regulation Number
    866.5510
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This kit is intended for the quantitative in vitro determination of human IgM in human serum, lithium heparin or EDTA plasma, using the Binding Site SPAPLUS turbidimetric analyser. Measurement of IgM aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents. The test results are to be used in conjunction with other clinical and laboratory findings.

    Device Description

    The SPAPlus IgM Kit comprises the following reagents:

    Antiserum: Goat Anti-IgM is supplied in stabilised liquid form. Preservatives: 0.099% sodium azide, 0.1% E-amino-n-caproic acid (EACA), 0.5% BSA and 0.01% benzamidine.

    Calibrator and Controls: Pooled human serum, supplied in stabilised liquid form. Contain 0.099% sodium azide. 0.1% EACA and 0.01% benzamidine as preservatives. The concentration given on the quality control certificate has been obtained by comparison with the DA470k international reference material.

    Reaction Buffer: Containing 0.099% sodium azide as a preservative.

    AI/ML Overview

    The provided document (K191465 - Human IgM Kit for use on SPAPlus Special 510(k) Submission Summary) describes the analytical and performance characteristics of a quantitative in vitro diagnostic device, not an AI/ML-based medical device. Therefore, many of the questions regarding acceptance criteria for AI/ML performance, ground truth establishment by experts, MRMC studies, and training/test set details are not applicable.

    This submission is for a modification to an existing device (Human IgM Kit for Use on SPAPlus, K082129), specifically changing the source of the detection antibody from sheep to goat and altering the antibody resting buffer. The primary goal of the studies presented is to demonstrate that these changes do not adversely affect the device's performance compared to the original cleared device and that the new kit is substantially equivalent.

    Here's an attempt to answer the questions based on the available information, noting when a question is not applicable to this type of device:

    Acceptance Criteria and Device Performance

    1. A table of acceptance criteria and the reported device performance

    Since this is a chemistry assay device, the "acceptance criteria" are more about demonstrating comparable performance to the predicate device and meeting established analytical performance standards for in vitro diagnostics rather than an AI/ML output metric like accuracy or AUC.

    Performance CharacteristicAcceptance Criteria / GoalReported Device Performance
    PrecisionComparable to the predicate device (K082129) and consistent with established CLSI guidelines (EP5-A3).Repeatability and Within Laboratory:Level 1 (0.344 mg/L): Total CV% = 6.1Level 2 (2.949 mg/L): Total CV% = 4.0Level 3 (5.975 mg/L): Total CV% = 2.9Between Instrument:Level 1 (0.340 mg/L): CV% = 2.4Level 2 (2.994 mg/L): CV% = 5.4Level 3 (6.184 mg/L): CV% = 4.8Between Lot:Level 1 (0.318 mg/L): CV% = 3.0Level 2 (3.007 mg/L): CV% = 3.0Level 3 (6.236 mg/L): CV% = 4.7 Conclusion: No change in performance compared to K082129.
    Linearity/Assay RangeMaintain linearity and reportable range comparable to the predicate (K082129).Linear regression equation: y = 0.9904x - 0.1583 g/L with an r value of 0.999. Conclusion: Comparable to predicate, linearity claims unchanged.
    Kit StabilityMaintain stability claims (12 months at 4ºC) in accordance with ISO 23640:2015 and CLSI EP25-A.Accelerated Stability:All parameters (IR, Control Low, Control High, Samples 1, 2, 3) passed acceptance criteria for 365 days stability at 4ºC based on accelerated testing (39-47.3 days at 37ºC).Conclusion: Stability claim of 12 months verified. Real Time Stability: Ongoing. On Board Stability: No difference from original 510(k) submission.
    Detection Limit (LoD/LoQ)Maintain LoD, LoB, and LoQ claims comparable to the predicate (K082129) and conform to CLSI EP17-A2.LoQ: Validated as 0.2 g/L (at standard 1/20 dilution) with all samples reporting within 10% allowable CV.LoD: Estimated 0.004 g/L.LoB: Estimated 0.001 g/L.Conclusion: No change in performance observed after antisera change; claims unchanged.
    Method ComparisonDemonstrate substantial equivalence to the predicate device.Bland Altman Mean Bias: -2.18% (95% Limits: -16.55% to 12.18%)Passing Bablok: y = 0.964x + 0.008 (Slope 95% CI: 0.947 to 0.986, Intercept 95% CI: -0.008 to 0.028)Correlation coefficient: 0.996 (for sample ranges 0.107 - 11.780 g/L for predicate; 0.116 - 11.066 g/L for test device).Conclusion: No change in performance compared to K082129.
    Reference Range Transfer≤2 out of 20 samples fall outside the established reference interval (0.35 - 2.42 g/L) for US donors.19 out of 20 samples were within the reference interval (0.364 to 1.736 g/L). One sample was 0.348 g/L (lower boundary 0.35 g/L).Conclusion: Met acceptance criteria, indicating reference interval can be transferred.

    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Precision Studies:
      • Repeatability & Within Laboratory: 3 different samples, 80 replicates each (20 working days, 2 runs/day, 2 reps/run).
      • Between Instrument: 3 different samples, 24 replicates each (6 working days, 2 instruments/day, 2 reps/instrument).
      • Between Lot: 3 different samples, 24 replicates each (6 working days, 2 lots/day, 2 reps/lot).
      • Provenance: Not explicitly stated, but likely retrospective lab testing from the manufacturer (The Binding Site Group Ltd. in UK). Dates of testing are implied by "over 20 working days" etc.
    • Linearity Study: High and low pools, dilution series, 6 replicates per diluted sample.
      • Provenance: Not explicitly stated, likely retrospective lab testing.
    • Stability Studies:
      • Accelerated Stability: 6 replicates of controls, internal reference, and samples.
      • Provenance: Not explicitly stated, likely retrospective lab testing.
    • Detection Limit (LoQ) Study: Four samples, two reagent lots.
      • Provenance: Not explicitly stated, likely retrospective lab testing.
    • Method Comparison Study: 89 serum samples and 44 plasma samples.
      • Provenance: Performed "in accordance with pre-submission meeting Q171503." Not explicitly stated, but likely retrospective from clinical laboratories or biobanks.
    • Reference Range Transfer Study: 20 samples from "apparently healthy US donors."
      • Provenance: Prospective collection from apparently healthy US donors appears to be implied given the mention of "US donors" and the purpose of transferring the reference interval.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is an in vitro diagnostic device that measures a quantitative analyte (IgM). "Ground truth" for this type of device is established by the analytical measurement itself, traceable to an international reference material (ERM-DA470k/IFCC). Clinical expert consensus is not a method for establishing the "ground truth" concentration of an analyte in a sample for this type of device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing ground truth in image interpretation or clinical diagnosis, often when human expert variability is a factor. This submission is for a turbidimetric assay, where the "ground truth" is a quantitative measurement traceable to a reference standard.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device and does not involve human readers interpreting images or data.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML device. The "performance" is the analytical performance of the assay itself.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The "ground truth" for the quantitative measurement of IgM in this IVD is established by its traceability to a recognized international reference material (ERM-DA470k/IFCC). This standard provides the authoritative value for IgM concentration against which the device's measurements are calibrated and compared.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device that requires a "training set" in the machine learning sense. The device is a "kit" of reagents and controls used on an analyser.

    9. How the ground truth for the training set was established

    Not applicable (as above, no "training set" in the AI/ML sense). The calibration of the assay (which could be conceptually analogous to "training" a traditional assay) is traceable to ERM-DA470k/IFCC, an international reference material.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190686

    Validate with FDA (Live)

    Date Cleared
    2019-05-28

    (71 days)

    Product Code
    Regulation Number
    866.5510
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Optilite IgM CSF Kit is intended for the quantitative in vitro measurement of IgM in cerebrospinal fluid (CSF) samples using the Optilite analyser.

    Device Description

    The Optilite IgM CSF Kit comprises the following reagents:
    Latex Reagent: Supplied in stabilised liquid form. Preservatives: 0.025% sodium azide, 0.1% E-amino-n-caproic acid (EACA) and 0.01% benzamidine, 0.05% ProClin.
    Calibrator and Controls: Pooled human serum, supplied in stabilised liquid form. Containing 0.099% sodium azide, 0.1% EACA and 0.01% benzamidine as preservatives. The concentration given on the quality control certificate has been obtained by comparison with the DA470k international reference material.
    Reaction Buffer: Containing 0.099% sodium azide as a preservative.

    AI/ML Overview

    The provided text describes the 510(k) submission for the Optilite IgM CSF Kit, an in vitro diagnostic device, not an AI/ML-based device. Therefore, the questions related to AI/ML specific criteria, such as expert adjudication, MRMC studies, standalone algorithm performance, and training set details, are not applicable to this submission.

    The acceptance criteria and device performance evaluation for this diagnostic kit are centered on analytical performance characteristics, which are standard for laboratory assays.

    Here's a breakdown of the applicable information from the provided text:

    1. Table of acceptance criteria and the reported device performance:

    Performance CharacteristicAcceptance Criteria (from CLSI Guidelines and Internal Standards)Reported Device Performance
    Precision
    Total Precision (%CV)<10%Level 1: 7.2%Level 2: 5.9%Level 3: 3.6%Level 4: 4.2%
    Within-run Precision (%CV)<5%Level 1: 3.9%Level 2: 2.8%Level 3: 2.2%Level 4: 2.3%
    Between-run Precision (%CV)<8%Level 1: 1.5%Level 2: 2.5%Level 3: 2.6%Level 4: 2.9%
    Between-day Precision (%CV)<8%Level 1: 5.9%Level 2: 4.6%Level 3: 1.3%Level 4: 2.0%
    Linearity/Assay Reportable RangeDeviation from linearity <10% over the measuring rangeConfirmed over 0.07 - 4.55 mg/L with deviation from linearity <10%
    Detection Limit (LoQ)Not explicitly stated as a numerical criterion, but defined as the bottom of the measuring range.0.11 mg/L (bottom of measuring range)
    Analytical Specificity (Interference)Mean results from spiked samples within 10% of control samples.Not affected by Acetaminophen (1324µmol/L), Acetylsalicylic Acid (3.62mmol/L), Bilirubin (80mg/L), Haemoglobin (1g/L)
    Method Comparison (Regression Analysis)Not explicitly stated as a numerical criterion for slope/intercept, but implied to show substantial equivalence.N=155 samples, Slope: 1.02, 95% CI: 1.00 to 1.04; Intercept: 0.07, 95% CI: 0.03 to 0.09; Pearson's r: 0.997

    2. Sample size used for the test set and the data provenance:

    • Precision Study: 4 sample preparations tested (details on the specific number of aliquots or patient samples composing these preparations are not given). The study design involved 2 runs per day (each in duplicate) over 5 days using 3 analyzers.
    • Linearity Study: A serially diluted sample was used. Specific N is not provided.
    • Detection Limit (LoQ) Study: The study was based on CLSI EP17-A2. Specific N is not provided.
    • Analytical Specificity (Interference) Study: Samples at different IgM concentrations were spiked with interfering substances and tested. Specific N is not provided.
    • Method Comparison Study: A total of 239 samples were initially tested, including 219 native CSF samples and 20 patient samples spiked with purified IgM. 155 of these samples were within the measuring range and used for regression analysis.
    • Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, the samples would be human CSF.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is a quantitative immunodiagnostic assay, not an imaging device requiring expert interpretation for ground truth. The "ground truth" for this device's performance is established by reference methods, international standards (ERM-DA470k/IFCC), and the inherent accuracy and precision of the measurements themselves validated against established statistical methodologies (CLSI guidelines).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable, as this is a quantitative diagnostic assay, not a subjective interpretation task that requires adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable, as this is not an AI/ML-based device that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device is an automated turbidimetric analyzer (Optilite) performing a quantitative assay. Its performance is inherently "standalone" in the sense that the instrument provides a numerical result without human interpretation being part of the result generation. The performance metrics presented (precision, linearity, detection limit, method comparison) are its standalone performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The ground truth or reference for this quantitative assay is:

    • Traceability: Established through calibration traceable to the ERM-DA470k/IFCC international reference material.
    • Method Comparison: Comparison against a legally marketed predicate device (Human IgM CSF Kit for use on SPAPLUS K120750) using patient samples. The predicate device itself serves as a reference point for substantial equivalence.
    • Statistical Methodologies: Adherence to CLSI (Clinical and Laboratory Standards Institute) guidelines for evaluating precision, linearity, detection capability, and interference.

    8. The sample size for the training set:

    Not applicable, as this is not an AI/ML device that requires a "training set."

    9. How the ground truth for the training set was established:

    Not applicable, as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183151

    Validate with FDA (Live)

    Date Cleared
    2019-01-23

    (70 days)

    Product Code
    Regulation Number
    866.5510
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Optilite IgA CSF Kit is intended for the quantitative in vitro measurement of IgA in cerebrospinal fluid (CSF) using the Optilite analyser.

    Device Description

    The Optilite IgA CSF Kit comprises the following reagents: Latex Reagent, Calibrator and Controls, and Reaction Buffer.

    AI/ML Overview

    The provided text describes the performance characteristics of the Optilite IgA CSF Kit. Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance
    Precision (Total CV%)Not explicitly stated, but generally <10% for clinical assaysLevel 1 (2.24 mg/L): 4.0%
    Level 2 (3.53 mg/L): 5.6%
    Level 3 (4.59 mg/L): 4.4%
    Level 4 (28.90 mg/L): 2.9%
    Linearity/Assay Reportable RangeDeviation from linearity <10%Confirmed over the range of 1.0-44.8 mg/L with deviation from linearity <10%
    TraceabilityTraceable to international reference materialTraceable to ERM-DA470k/IFCC
    Kit Stability (Shelf life)Sufficient for practical useUp to 18 months
    Open-Vial StabilitySufficient for practical useUp to 3 months (at 2-8°C)
    On-Board StabilitySufficient for practical useUp to 30 days (on Optilite Analyser)
    Limit of Quantitation (LoQ)Defined as the bottom of the measuring range0.91 mg/L
    Analytical Specificity (Interference)No significant interference from common substancesNo significant interference observed with hemoglobin (2.5g/L), bilirubin (100mg/L), acetaminophen (1324μmol/L), or acetylsalicylic acid (3.63mmol/L)
    Method Comparison (Correlation)High correlation with a predicate device (e.g., correlation coefficient > 0.95, slope near 1, intercept near 0)Correlation coefficient 0.984; Passing Bablok: y = 1.05x - 0.02 (Slope 95% Cl: 1.03 to 1.07, Intercept 95% Cl: -0.07 to 0.05)

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision/Reproducibility: 4 different samples were used. The number of runs (two per day for 5 days) and analysts (two) suggest multiple measurements for each sample. Data provenance is not explicitly stated but is implicitly from an internal laboratory study (prospective).
    • Linearity: One serially diluted sample was used. Data provenance is implicitly from an internal laboratory study (prospective).
    • Analytical Specificity: A CSF sample close to the medical decision point and an elevated CSF sample were tested. Data provenance is implicitly from an internal laboratory study (prospective).
    • Method Comparison: 130 CSF samples (including 40 samples with analyte levels within the reference interval) were used. Data provenance is not explicitly stated, but it's likely from a collection of clinical samples. It is retrospective in nature as these are "samples" analyzed, not patients prospectively enrolled.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This device is an in vitro diagnostic (IVD) for quantitative measurement of a biomarker. The "ground truth" for the test set is established by the reference methods or the true concentration of the analyte, not by expert interpretation in the same way an imaging or pathology device would involve human experts.

    • For precision, the ground truth is the true concentration of IgA in the sample, measured repeatedly to assess variation.
    • For linearity, the ground truth is the expected concentration based on the serial dilution.
    • For traceability, the ground truth is the value assigned by the international reference material (ERM-DA470k/IFCC).
    • For method comparison, the ground truth is the measurement obtained from the alternative commercially available assay (predicate device).

    Therefore, the concept of "number of experts used" or their "qualifications" for establishing ground truth is not directly applicable in the context of this type of quantitative IVD performance study.

    4. Adjudication Method for the Test Set

    Not applicable. As a quantitative IVD, the results are numerical values, and adjudication by experts is not a standard practice for assessing performance metrics like precision, linearity, or method correlation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation (e.g., radiologists reading images) to assess the impact of AI assistance on human performance. The Optilite IgA CSF Kit is a standalone automated quantitative assay, not an AI-assisted interpretation tool.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Yes, the performance studies described are intrinsically standalone performance evaluations of the Optilite IgA CSF Kit (the "algorithm/device only" in this context) without human-in-the-loop performance. The device provides a quantitative measurement directly.

    7. The Type of Ground Truth Used

    • Precision and Linearity: The ground truth is based on the known or reference concentrations of the samples used in the study.
    • Traceability: The ground truth is the value assigned by the international reference material (ERM-DA470k/IFCC).
    • Method Comparison: The ground truth for comparative purposes is the result obtained from an alternative commercially available assay (predicate device).

    8. The Sample Size for the Training Set

    The document does not mention a "training set" in the context of machine learning or AI. This device is an in vitro diagnostic kit based on immunoturbidimetry, which is a chemical/biological reaction and measurement system, not a machine learning model that requires a training set. The term "training set" is not applicable here.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180099

    Validate with FDA (Live)

    Date Cleared
    2018-10-12

    (269 days)

    Product Code
    Regulation Number
    866.5270
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Optilite High Sensitivity C-Reactive Protein Kit is intended for the quantitative in vitro measurement of C-Reactive Protein in serum using the Binding Site Optilite analyser for evaluation of conditions thought to be associated with inflammation, in otherwise healthy individuals. This test should be used in conjunction with other laboratory and clinical findings.

    Device Description

    The Optilite High Sensitivity C-Reactive Protein Kit is comprised of latex reagent, single calibrator, controls (high and low) and reaction buffer in liquid form. The reagents contain 0.099% sodium azide as preservative.

    AI/ML Overview

    Acceptance Criteria and Study for Optilite High Sensitivity C-Reactive Protein Kit

    Here's a breakdown of the acceptance criteria and the study results for the Optilite High Sensitivity C-Reactive Protein Kit, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Performance CharacteristicAcceptance Criteria (Implicit from Study Design)Reported Device Performance (Optilite Kit)
    PrecisionTotal Precision: %CV < 10%Sample 1 (0.98 mg/L): 5.6%
    Within-Run Precision: %CV < 5%Sample 2 (1.55 mg/L): 5.0%
    Between-Run Precision: %CV < 8%Sample 3 (2.99 mg/L): 4.5%
    Between-Day Precision: %CV < 8%Sample 4 (5.41 mg/L): 4.4%
    Sample 5 (8.50 mg/L): 3.1%
    Individual Within-Run, Between-Run, and Between-Day %CVs were all within their respective implicit criteria (e.g., Within-Run %CVs ranged from 1.5% to 2.9%).
    Linearity/Reportable RangeDeviation from linearity ≤ 10%Linear over 0.44 - 12.16 mg/L with deviation from linearity ≤ 10%
    TraceabilityTraceable to an international reference standardERM-DA474
    Kit Stability (Shelf-life)Sufficient to support a claim (e.g., 12+ months)Currently supports a 13-month stability claim (on-going study)
    Open-vial StabilitySufficient for practical useUp to 3 months at 2-8°C
    On-board StabilitySufficient for practical useAt least 30 days on Optilite Analyser
    Detection Limit (LoQ)Clinically relevant lower limit of quantification0.5 mg/L
    Analytical Specificity (Interference)Mean results from spiked samples within 10% of control samplesNo significant interference from hemoglobin (5g/L), conjugated bilirubin (200mg/L), unconjugated bilirubin (200mg/L), rheumatoid factor (25lU/mL), triglyceride (1000mg/dL), intralipid (2000mg/dL), and 6 common therapeutic drugs.
    Method Comparison with PredicateRegression analysis (e.g., Passing & Bablok slope and intercept) indicating substantial agreement with predicate device (implicit acceptance of close to 1 for slope and close to 0 for intercept with narrow CIs)Passing & Bablok: y = 0.99x + 0.09 (95% CI Slope: 0.97 to 1.01, 95% CI Intercept: 0.05 to 0.11) - PASS
    Weighted Deming: y = 0.97x + 0.10 (95% CI Slope: 0.94 to 0.99, 95% CI Intercept: 0.07 to 0.13)
    Weighted Linear: y = 0.95x + 0.12 (95% CI Slope: 0.93 to 0.98, 95% CI Intercept: 0.10 to 0.15)
    Expected Values/Reference Range VerificationVerification of stated reference intervalVerified by testing 50 adult donor samples; Consensus reference interval for adults: <3 mg/L

    2. Sample Size Used for the Test Set and Data Provenance:

    • Precision Study: 5 serum samples (targeted at specific concentrations: 1mg/L, 1-3mg/L, 3mg/L, 5mg/L, and 9mg/L).
    • Linearity Study: Serially diluted serum samples (number not explicitly stated, but sufficient to cover 0.5 – 10.0 mg/L).
    • Detection Limit (LoB, LoD, LoQ) Studies:
      • LoB: 4 serum samples
      • LoD: 4 serum samples
      • LoQ: 4 serum samples
    • Analytical Specificity (Interference) Study: Serum samples with CRP concentrations targeted at <1mg/L, 1-3mg/L, 3mg/L, and >3mg/L.
    • Method Comparison with Predicate Device: 391 serum samples (302 reported results within the measuring range). These included 87 normal donors and 304 clinical samples.
    • Expected Values/Reference Range Verification: 50 adult donor samples.

    Data Provenance: The document does not explicitly state the country of origin for the data or whether it was retrospective or prospective. Given "The Binding Site Group, Ltd." is based in the UK, it's highly probable that the studies were conducted there. The nature of the studies (e.g., precision, linearity, method comparison) generally implies prospective testing of samples for analytical performance, though the 304 "clinical samples" in the method comparison could be retrospective or a mix.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable. This device is an in vitro diagnostic (IVD) for quantitative measurement of a biomarker (C-Reactive Protein). The "ground truth" for its performance is established through comparisons with reference methods, calibrated materials, and statistical analysis of assay performance (precision, linearity, detection limits), rather than through expert consensus on qualitative interpretation.

    4. Adjudication Method for the Test Set:

    Not applicable, as this is a quantitative IVD device. Test results are numerical and assessed against predefined analytical performance criteria, not subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for AI-powered diagnostic imaging devices where human readers interpret images. The Optilite Kit is an automated immunoassay system, not an imaging device requiring human interpretation.

    6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    Yes, a standalone study of the algorithm (the Optilite High Sensitivity C-Reactive Protein Kit and Optilite analyser) was performed. All the analytical performance characteristics (precision, linearity, detection limits, analytical specificity) and method comparison studies described are standalone evaluations of the device's performance without human intervention in the result generation process.

    7. The Type of Ground Truth Used:

    The ground truth for this device is established through:

    • International Reference Standards: The assay is traceable to the international reference standard ERM-DA474 for calibration.
    • Statistical Definitions: For precision (CV%), linearity (deviation from linearity), detection limits (LoB, LoD, LoQ), established statistical methodologies (CLSI EP5-A3, EP6-A, EP17-A2) define the acceptable "ground truth" or performance targets.
    • Reference Methods/Predicate Device: For method comparison, the results from the legally marketed predicate device (C-Reactive Protein High Sensitive Test System for Cobas Integra Instruments) serve as the comparative "ground truth" for demonstrating substantial equivalence.
    • Clinical Relevance: The reference interval (<3 mg/L) is derived from consensus and published literature (Macy et al., 1997), serving as a clinically accepted range.

    8. The Sample Size for the Training Set:

    Not applicable in the context of this device. This is a laboratory diagnostic kit, not an AI or machine learning algorithm that requires a "training set" in the conventional sense. The "training" of the device involves its design, optimization, and calibration using standardized materials and methods.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable. As mentioned above, there isn't a "training set" for this type of device in the AI sense. The development and optimization of the reagent and instrument involved:

    • Chemical and Biological Principles: Based on well-established immunoturbidimetry principles and antibody-antigen reactions.
    • Calibration: Using a single calibrator diluted on the analyser, traceable to the international reference standard ERM-DA474, to establish the relationship between optical signal and CRP concentration.
    • Reagent Formulation and Instrument Design: Iterative development and testing of reagent components (latex reagent, calibrator, controls, buffer) and instrument parameters to achieve desired analytical performance characteristics.
    Ask a Question

    Ask a specific question about this device

    K Number
    K173732

    Validate with FDA (Live)

    Date Cleared
    2018-08-23

    (260 days)

    Product Code
    Regulation Number
    866.5550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Optilite Freelite Mx Kappa Free Kit is intended for the quantitative in vitro measurement of Kappa free light chains in serum and urine using the Binding Site Optilite analyser. Measurement of free light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenström's macroglobulinaemia, AL amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings.

    The Optilite Freelite Mx Lambda Free Kit is intended for the quantitative in vitro measurement of Lambda free light chains in serum and urine using the Binding Site Optilite analyser. Measurement of free light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenström's macroglobulinaemia, AL amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings.

    Device Description

    The Optilite Freelite Mx Kappa Free Kit is comprised of the following reagents: Latex Reagent: Sheep anti-Human Kappa (polyclonal monospecific) antibody (F(ab')2 fragment) bound to 200nm polystyrene latex particles. In 0.033M Glycine buffered saline pH8 (containing Sodium Azide (0.033%), Proclin (0.05%), Benzamidine (0.01%) and EACA (0.1%) as preservative plus BSA (0.033%)). Calibrator and Controls: Pooled human serum, supplied in stabilised liquid form. Containing 0.099% sodium azide, 0.1% EACA and 0.01% benzamidine as preservatives. Reaction Buffer: Containing 0.099% sodium azide as a preservative.

    The Optilite Freelite Mx Lambda Free Kit is comprised of the following reagents: Latex Reagent: Sheep anti-Human Lambda (polyclonal monospecific) Free antibody (F(ab')2 fragment) bound to 200nm polystyrene latex particles. In 0.033M Glycine buffered saline pH8 (containing Sodium Azide (0.033%), Proclin (0.05%), Benzamidine (0.01%) and EACA (0.1%) as preservatives plus BSA (0.1665%)). Calibrator and Controls: Pooled human serum, supplied in stabilised liquid form. Containing 0.099% sodium azide, 0.1% EACA and 0.01% benzamidine as preservatives. Reaction Buffer: Containing 0.099% sodium azide as a preservative.

    Note - In Optilite Freelite kits, the latex reagent and reaction buffer are supplied in a single wedge with a chamber for each fluid. They are therefore labelled as a single component Optilite Kappa Free Mx Reagent or Optilite Lambda Free Mx Reagent.

    AI/ML Overview

    The provided text describes the performance characteristics of the Optilite Freelite Mx Kappa Free Kit and Optilite Freelite Mx Lambda Free Kit, which are devices for the quantitative measurement of Kappa and Lambda free light chains in serum and urine.

    Here's an analysis of the acceptance criteria and the study that proves the device meets these criteria, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a submission for an in-vitro diagnostic device, the "acceptance criteria" are typically inferred from the performance targets set by the manufacturer and assessed against relevant CLSI (Clinical and Laboratory Standards Institute) guidelines. The reported device performance is directly stated in the tables within the document.

    Precision/Reproducibility

    Performance MetricAcceptance Criteria (Inferred from CLSI Guidelines/Industry Standards for IVDs)Reported Device Performance (Kappa Urine)Reported Device Performance (Lambda Urine)
    Kappa
    Total %CV (Sample 1)Typically < 10% (for low concentrations)7.6%-
    Total %CV (Sample 2)Typically < 10%8.2%-
    Total %CV (Sample 3)Typically < 10%7.1%-
    Total %CV (Sample 4)Typically < 10%8.8%-
    Lambda
    Total %CV (Sample 1)Typically < 10% (for low concentrations)-7.3%
    Total %CV (Sample 2)Typically < 10%-8.7%
    Total %CV (Sample 3)Typically < 10%-8.2%
    Total %CV (Sample 4)Typically < 10%-8.1%
    Total %CV (Sample 5)Typically < 10%-8.6%

    Linearity/Assay Reportable Range

    Performance MetricAcceptance Criteria (Inferred from CLSI EP6-A)Reported Device Performance (Kappa)Reported Device Performance (Lambda)
    Linearity RangeDemonstrated across the intended measuring range with acceptable deviation2.383 - 142.687 mg/L3.87 - 160.02 mg/L
    Deviation from Linearity≤ 10%≤ 10%≤ 10%

    Detection Limit

    Performance MetricAcceptance Criteria (Inferred from CLSI EP17-A)Reported Device Performance (Kappa Serum)Reported Device Performance (Kappa Urine)Reported Device Performance (Lambda Serum)Reported Device Performance (Lambda Urine)
    Limit of Blank (LoB)Accurately determined0.195 mg/L0.210 mg/L0.285 mg/L0.080 mg/L
    Limit of Detection (LoD)Accurately determined0.224 mg/L0.232 mg/L0.393 mg/L0.184 mg/L
    Limit of Quantitation (LoQ)Accurately determined and clinically relevant0.330 mg/L0.330 mg/L0.740 mg/L0.740 mg/L

    Analytical Specificity (Interference)

    Performance MetricAcceptance Criteria (Inferred from CLSI EP7-A2)Reported Device Performance (Kappa)Reported Device Performance (Lambda)
    Interference (% Deviation from Control)≤ 10% for studied interferentsNo significant interference observed for listed substancesNo significant interference observed for listed substances

    Method Comparison with Predicate Device

    Performance MetricAcceptance Criteria (Inferred for Substantial Equivalence)Reported Device Performance (Kappa Free)Reported Device Performance (Lambda Free)
    Correlation Coefficient (r)High correlation (typically > 0.95 or 0.975 for quantitative assays)0.9920.947
    Regression Equation (y=mx+c)Slope (m) close to 1, intercept (c) close to 0y = 1.02x - 0.21y = 1.11x - 0.27

    Clinical Agreement (2x2 Table with Predicate)

    Performance MetricAcceptance Criteria (Inferred for Substantial Equivalence)Reported Device Performance (Kappa)Reported Device Performance (Lambda)
    Positive AgreementHigh, clinically acceptable90.7% (82.7% to 95.2% CI)73.7% (62.8% to 82.3% CI)
    Negative AgreementHigh, clinically acceptable98.8% (93.3% to 99.8% CI)100.0% (96.0% to 100.0% CI)
    Overall AgreementHigh, clinically acceptable94.6% (90.0% to 97.1% CI)88.1% (92.3% to 92.2% CI)

    Notes on Acceptance Criteria: The document does not explicitly state quantitative acceptance criteria or pass/fail thresholds for many of these performance metrics. The criteria listed above are inferred based on typical expectations for in-vitro diagnostic devices undergoing 510(k) clearance, often guided by CLSI guidelines. The FDA's substantial equivalence determination implies that the reported performance is acceptable in comparison to the predicate device and the recognized standards.

    2. Sample Size Used for the Test Set and Data Provenance

    • Precision/Reproducibility:

      • Kappa: Four urine sample pools were tested. The testing involved 21 days with two runs per day using one reagent lot on three analyzers.
      • Lambda: Five urine sample pools were tested. The testing involved 23 days with two runs per day using one reagent lot on three analyzers.
      • Provenance: Not explicitly stated, but likely in-house (Binding Site Group Ltd.) and conducted under controlled laboratory conditions, given it's part of analytical performance testing for an IVD.
    • Linearity/Assay Reportable Range:

      • Sample Size: "serially diluted urine samples" were used. The exact number of individual samples or dilutions is not specified.
      • Provenance: Not explicitly stated, but likely in-house.
    • Detection Limit:

      • LoB: 60 determinations of blank samples (analyte depleted serum and urine at a low analyte concentration).
      • LoQ: Five independent samples (serum and urine) tested twelve times over five days.
      • Provenance: Not explicitly stated, but likely in-house.
    • Analytical Specificity (Interference):

      • Sample Size: "a single urine sample for the Kappa Free assay and the Lambda Free assay". These were prepared with interferents.
      • Provenance: Not explicitly stated, but likely in-house.
    • Method Comparison with Predicate Device:

      • Kappa Free: 165 urine samples (including 86 with analyte levels below the reference limit).
      • Lambda Free: 115 urine samples (including 59 with analyte levels below the reference limit).
      • Provenance: Not explicitly stated, but these are patient samples tested against a predicate device. Given the context of a European manufacturer and the lack of specific geographical mention, they could be from various locations or a specific clinic associated with the manufacturer. Retrospective or prospective is not specified, but typically for method comparison, collected samples are used retrospectively.
    • Clinical Agreement (2x2 Table with Predicate):

      • Kappa: 166 samples (derived from the 2x2 table: 78+1+8+79 = 166, though the sum of listed cells is 166, the total is given as 166, implying a discrepancy in the row/column totals). The method comparison stated 165, so there might be a rounding or slight difference in sample inclusion for the clinical agreement vs. regression.
      • Lambda: 168 samples (derived from the 2x2 table totals). The method comparison stated 115. This suggests different sample sets or criteria for the 2x2 table analysis compared to the regression analysis, especially given the sample sizes below the reference limit quoted for the method comparison.
      • Provenance: Not explicitly stated, but presumed to be from clinical settings for patient samples. These are likely retrospective samples previously analyzed by the predicate.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    For an in-vitro diagnostic device like this, "ground truth" and "experts" are typically not established in the same way as for imaging devices requiring expert interpretation.

    • Ground Truth: For this device, the "ground truth" is typically the result obtained from a well-established, often reference, method or the predicate device that the new device is being compared against. In this case, the predicate device (Freelite® Human Kappa/Lambda Free Kit for use on the Siemens BN™II) serves as the comparator, from which the "Assay" results are compared.
    • Experts and Qualifications: There is no mention of external experts being used for establishing ground truth. The evaluation relies on analytical and clinical performance studies comparing the new device against the predicate or established analytical standards.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    Not applicable for this type of in-vitro diagnostic device performance study. Adjudication methods like 2+1 or 3+1 are common in studies where human readers interpret medical images and their interpretations need to be reconciled to establish a ground truth or a consensus opinion. For IVD devices, the "truth" for comparison is usually defined by the results of a reference method or legally marketed predicate.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. MRMC studies, and improvement with AI assistance, are relevant to medical imaging devices where human readers (e.g., radiologists) interpret images. This submission is for an in-vitro diagnostic assay for laboratory measurement of biomarkers, which does not involve human interpretation of complex visual data in the same way.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This device is a standalone diagnostic assay (an "algorithm only" if one considers the assay chemistry and instrument processing as the "algorithm"). It performs quantitative measurements without direct human real-time interpretation influencing the numerical result. The human element is in collecting the sample, running the test, and interpreting the numerical result in a clinical context.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for comparative studies and assessment of substantial equivalence is primarily established by:

    • Predicate Device Results: For method comparison and clinical agreement, the results from the legally marketed predicate device (Freelite® Human Kappa/Lambda Free Kit on the Siemens BN™II) are used as the reference ('ground truth') for comparison.
    • CLSI Guidelines: For analytical performance (precision, linearity, detection limits, interference), the "ground truth" is defined by adherence to the methodologies and performance expectations outlined in recognized CLSI guidelines (e.g., EP17-A, EP7-A2, EP6-A, EP5-A2).

    8. The Sample Size for the Training Set

    This document does not specify a separate "training set" sample size in the context of machine learning. For IVD devices, development involves extensive internal testing and optimization (which can be considered analogous to "training"), but the provided document focuses on the validation studies. The studies described are primarily validation studies against regulatory requirements.

    9. How the Ground Truth for the Training Set Was Established

    Given that this is an IVD kit based on turbidimetry, it doesn't involve a "training set" or "ground truth" in the machine learning sense. The "ground truth" during development (analogous to training) would involve using known concentrations of analytes, reference materials, and established analytical methods to ensure the assay performs as expected.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 5