K Number
K180099

Validate with FDA (Live)

Date Cleared
2018-10-12

(269 days)

Product Code
Regulation Number
866.5270
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Optilite High Sensitivity C-Reactive Protein Kit is intended for the quantitative in vitro measurement of C-Reactive Protein in serum using the Binding Site Optilite analyser for evaluation of conditions thought to be associated with inflammation, in otherwise healthy individuals. This test should be used in conjunction with other laboratory and clinical findings.

Device Description

The Optilite High Sensitivity C-Reactive Protein Kit is comprised of latex reagent, single calibrator, controls (high and low) and reaction buffer in liquid form. The reagents contain 0.099% sodium azide as preservative.

AI/ML Overview

Acceptance Criteria and Study for Optilite High Sensitivity C-Reactive Protein Kit

Here's a breakdown of the acceptance criteria and the study results for the Optilite High Sensitivity C-Reactive Protein Kit, based on the provided FDA 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance:

Performance CharacteristicAcceptance Criteria (Implicit from Study Design)Reported Device Performance (Optilite Kit)
PrecisionTotal Precision: %CV < 10%Sample 1 (0.98 mg/L): 5.6%
Within-Run Precision: %CV < 5%Sample 2 (1.55 mg/L): 5.0%
Between-Run Precision: %CV < 8%Sample 3 (2.99 mg/L): 4.5%
Between-Day Precision: %CV < 8%Sample 4 (5.41 mg/L): 4.4%
Sample 5 (8.50 mg/L): 3.1%
Individual Within-Run, Between-Run, and Between-Day %CVs were all within their respective implicit criteria (e.g., Within-Run %CVs ranged from 1.5% to 2.9%).
Linearity/Reportable RangeDeviation from linearity ≤ 10%Linear over 0.44 - 12.16 mg/L with deviation from linearity ≤ 10%
TraceabilityTraceable to an international reference standardERM-DA474
Kit Stability (Shelf-life)Sufficient to support a claim (e.g., 12+ months)Currently supports a 13-month stability claim (on-going study)
Open-vial StabilitySufficient for practical useUp to 3 months at 2-8°C
On-board StabilitySufficient for practical useAt least 30 days on Optilite Analyser
Detection Limit (LoQ)Clinically relevant lower limit of quantification0.5 mg/L
Analytical Specificity (Interference)Mean results from spiked samples within 10% of control samplesNo significant interference from hemoglobin (5g/L), conjugated bilirubin (200mg/L), unconjugated bilirubin (200mg/L), rheumatoid factor (25lU/mL), triglyceride (1000mg/dL), intralipid (2000mg/dL), and 6 common therapeutic drugs.
Method Comparison with PredicateRegression analysis (e.g., Passing & Bablok slope and intercept) indicating substantial agreement with predicate device (implicit acceptance of close to 1 for slope and close to 0 for intercept with narrow CIs)Passing & Bablok: y = 0.99x + 0.09 (95% CI Slope: 0.97 to 1.01, 95% CI Intercept: 0.05 to 0.11) - PASS
Weighted Deming: y = 0.97x + 0.10 (95% CI Slope: 0.94 to 0.99, 95% CI Intercept: 0.07 to 0.13)
Weighted Linear: y = 0.95x + 0.12 (95% CI Slope: 0.93 to 0.98, 95% CI Intercept: 0.10 to 0.15)
Expected Values/Reference Range VerificationVerification of stated reference intervalVerified by testing 50 adult donor samples; Consensus reference interval for adults: <3 mg/L

2. Sample Size Used for the Test Set and Data Provenance:

  • Precision Study: 5 serum samples (targeted at specific concentrations: 1mg/L, 1-3mg/L, 3mg/L, 5mg/L, and 9mg/L).
  • Linearity Study: Serially diluted serum samples (number not explicitly stated, but sufficient to cover 0.5 – 10.0 mg/L).
  • Detection Limit (LoB, LoD, LoQ) Studies:
    • LoB: 4 serum samples
    • LoD: 4 serum samples
    • LoQ: 4 serum samples
  • Analytical Specificity (Interference) Study: Serum samples with CRP concentrations targeted at <1mg/L, 1-3mg/L, 3mg/L, and >3mg/L.
  • Method Comparison with Predicate Device: 391 serum samples (302 reported results within the measuring range). These included 87 normal donors and 304 clinical samples.
  • Expected Values/Reference Range Verification: 50 adult donor samples.

Data Provenance: The document does not explicitly state the country of origin for the data or whether it was retrospective or prospective. Given "The Binding Site Group, Ltd." is based in the UK, it's highly probable that the studies were conducted there. The nature of the studies (e.g., precision, linearity, method comparison) generally implies prospective testing of samples for analytical performance, though the 304 "clinical samples" in the method comparison could be retrospective or a mix.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

Not applicable. This device is an in vitro diagnostic (IVD) for quantitative measurement of a biomarker (C-Reactive Protein). The "ground truth" for its performance is established through comparisons with reference methods, calibrated materials, and statistical analysis of assay performance (precision, linearity, detection limits), rather than through expert consensus on qualitative interpretation.

4. Adjudication Method for the Test Set:

Not applicable, as this is a quantitative IVD device. Test results are numerical and assessed against predefined analytical performance criteria, not subjective interpretations requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for AI-powered diagnostic imaging devices where human readers interpret images. The Optilite Kit is an automated immunoassay system, not an imaging device requiring human interpretation.

6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

Yes, a standalone study of the algorithm (the Optilite High Sensitivity C-Reactive Protein Kit and Optilite analyser) was performed. All the analytical performance characteristics (precision, linearity, detection limits, analytical specificity) and method comparison studies described are standalone evaluations of the device's performance without human intervention in the result generation process.

7. The Type of Ground Truth Used:

The ground truth for this device is established through:

  • International Reference Standards: The assay is traceable to the international reference standard ERM-DA474 for calibration.
  • Statistical Definitions: For precision (CV%), linearity (deviation from linearity), detection limits (LoB, LoD, LoQ), established statistical methodologies (CLSI EP5-A3, EP6-A, EP17-A2) define the acceptable "ground truth" or performance targets.
  • Reference Methods/Predicate Device: For method comparison, the results from the legally marketed predicate device (C-Reactive Protein High Sensitive Test System for Cobas Integra Instruments) serve as the comparative "ground truth" for demonstrating substantial equivalence.
  • Clinical Relevance: The reference interval (<3 mg/L) is derived from consensus and published literature (Macy et al., 1997), serving as a clinically accepted range.

8. The Sample Size for the Training Set:

Not applicable in the context of this device. This is a laboratory diagnostic kit, not an AI or machine learning algorithm that requires a "training set" in the conventional sense. The "training" of the device involves its design, optimization, and calibration using standardized materials and methods.

9. How the Ground Truth for the Training Set Was Established:

Not applicable. As mentioned above, there isn't a "training set" for this type of device in the AI sense. The development and optimization of the reagent and instrument involved:

  • Chemical and Biological Principles: Based on well-established immunoturbidimetry principles and antibody-antigen reactions.
  • Calibration: Using a single calibrator diluted on the analyser, traceable to the international reference standard ERM-DA474, to establish the relationship between optical signal and CRP concentration.
  • Reagent Formulation and Instrument Design: Iterative development and testing of reagent components (latex reagent, calibrator, controls, buffer) and instrument parameters to achieve desired analytical performance characteristics.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services logo. The logo on the right is the FDA logo, which includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 12, 2018

The Binding Site Group Ltd Andrea Thomas, Regulatory Affairs Manager 8 Calthorpe Road Edgbaston Birmingham B15 10T UK

Re: K180099

Trade/Device Name: Optilite High Sensitivity C-Reactive Protein Kit Regulation Number: 21 CFR 866.5270 Regulation Name: C-reactive protein immunological test system Regulatory Class: Class II Product Code: DCN Dated: August 29, 2018 Received: September 4, 2018

Dear Andrea Thomas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kellie B. Kelm -S

for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180099

Device Name

Optilite High Sensitivity C-Reactive Protein Kit

Indications for Use (Describe)

The Optilite High Sensitivity C-Reactive Protein Kit is intended for the quantitative in vitro measurement of C-Reactive Protein in serum using the Binding Site Optilite analyser for evaluation thought to be associated with inflammation, in otherwise healthy individuals. This test should be used in conjunction with other laboratory and clinical findings.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Optilite High Sensitivity C-Reactive Protein Kit 510(k) Summary

Date prepared: 11th October 2018

A. 510(k) Number:

K180099

B. Purpose for Submission:

New device

C. Measurand:

C-Reactive Protein

D. Type of Test:

Quantitative immunoturbidimetry

E. Applicant:

The Binding Site Group, Ltd. Contact person: Andrea Thomas Address: 8 Calthorpe Road, Birmingham, B15 1QT, United Kingdom Contact number: +44(0)121 456 9500

F. Proprietary and Established Names:

Optilite® High Sensitivity C-Reactive Protein Kit

G. Regulatory Information:

    1. Regulation section:
      21 CFR 866.5270, C-reactive protein immunological test system.
    1. Classification:
      Class II
    1. Product code:
      DCN, System, Test, C-Reactive Protein
    1. Panel:
      Immunology (82)

{4}------------------------------------------------

H. Intended use:

1. Intended use(s):

The Optilite High Sensitivity C-Reactive Protein Kit is intended for the quantitative in vitro measurement of C-Reactive Protein in serum using the Binding Site Optilite analyser for evaluation of conditions thought to be associated with inflammation, in otherwise healthy individuals. This test should be used in conjunction with other laboratory and clinical findings.

2. Indication(s) for use:

Same as Intended use(s).

3. Special conditions for use statement(s):

Prescription use only

4. Special instrument requirements:

The Binding Site Optilite analyser

l. Device Description:

The Optilite High Sensitivity C-Reactive Protein Kit is comprised of latex reagent, single calibrator, controls (high and low) and reaction buffer in liquid form. The reagents contain 0.099% sodium azide as preservative.

J. Substantial equivalence information:

1. Predicate device name(s) and 510(k) number(s):

C-Reactive Protein High Sensitive Test System for Cobas Integra Instruments (K053603)

{5}------------------------------------------------

2. Comparison with predicate:

Similarities

ltemDevicePredicate
Intended useQuantitative in vitro measurement ofC-reactive protein (CRP)Same
Reference Interval<3mg/L<5mg/L
Sample typeSerumSerum (also Li-heparin and EDTAplasma)
Reagent typeLatex enhancedSame
Open vial stabilityThree months at 2 to 8°C12 weeks (on-board)
MethodTurbidimetrySame

Differences

ItemDevicePredicate
Measuring Range0.5-10mg/L (neat)0.1-20.0mg/L1.5-300mg/L (1/15 dilution)
AnalyserOptiliteIntegra
Antibodyanti- human CRP (sheep)anti-human CRP (mouse)
On board stability30 days (refrigerated on analyser)12 weeks
Calibration6 point, single calibrator diluted on analyser5 point, calibrator set
TraceabilityDA474ERM-DA470 (CRM470)

K. Standards and Guidance documents referenced:

Guidance for Industry - Review Criteria for Assessment of C - Reactive Protein (CRP), High Sensitivity C-Reactive Protein (hsCRP) and Cardiac C-Reactive Protein (cCRP) Assays Guidance for Industry and FDA Staff

CLSI EP17-A2 Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline - Second Edition

CLSI EP7-A2 Interference Testing in Clinical Chemistry, Approved Guideline - Second Edition

CLSI EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach

CLSI EP5-A3 Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline - Third Edition

{6}------------------------------------------------

L. Test Principle:

The determination of soluble antigen (CRP) concentration by turbidimetric methods involves the reaction with specific antiserum (anti-CRP) to form insoluble complexes. When light is passed through the suspension formed a portion of the light is transmitted and focused onto a photodiode by an optical lens system. The amount of transmitted light is indirectly proportional to the specific protein concentration in the test sample. Concentrations are automatically calculated by reference to a calibration curve stored within the instrument.

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

Precision/Reproducibility: a.

The studies were based on CLSI EP5-A3, where 5 sera samples (targeted at 1mg/L, 1-3mg/L, 3mg/L, 5mg/L, and 9mg/L) were tested in 2 runs per day (each of the 2 runs in duplicate) over 21 days using 3 reagent lots on 3 analysers. Results met the acceptance criteria for total precision (%CV<10%), within-run precision (%CV<5%), between-run precision (%CV<8%), and between-day precision (%CV<8%).

Sample12345
Mean concentration (mg/L)0.981.552.995.418.50
Total precisionSD0.0550.0770.1340.2370.260
%CV5.65.04.54.43.1
Within-run precisionSD0.0290.0240.0470.1150.168
%CV2.91.51.62.12.0
Between-run precisionSD0.0000.0210.0690.0810.000
%CV0.01.42.31.50.0
Between-day precisionSD0.0470.0700.1040.1910.199
%CV4.84.53.53.52.3
Between-batch precisionSD0.0110.0220.0570.1310.137
%CV1.11.41.92.41.6
Between-instrument precisionSD0.0470.0540.0680.1130.100
%CV4.73.52.32.11.2

b. Linearity/assay reportable range:

A linearity study was performed following CLSI EP6-A. The linearity of the high sensitivity CRP assay has been confirmed using serially diluted serum samples to cover the standard measuring range of 0.5 – 10.0 mg/L. The results demonstrated that the high sensitivity CRP assay is linear over the ranges of 0.44 - 12.16 mg/L at 1+0 dilution with deviation from linearity ≤ 10%.

{7}------------------------------------------------

  • c. Traceability, Stability, Expected values (controls, calibrators, or methods):

i) Traceability:

The calibration of the assay is traceable to the international reference standard ERM- DA474.

ii) Kit Stability:

Real-time stability - A study to establish shelf-life stability (from the date of manufacture when stored at recommended temperature 2-8°C) of the Optilite High Sensitivity CRP Kit is on-going. Currently available data supports a 13 month stability claim.

Open-vial stability - The Optilite High Sensitivity CRP Kit can be stored, opened at 2 - 8℃ for up to 3 months.

On-board stability - The Optilite High Sensitivity CRP Kit can be stored on-board the Optilite Analyser for at least 30 days.

d. Detection limit:

The analytical sensitivity was determined in accordance with CLSI EP17-A2. The Limit of Blank (LoB) was based on 4 serum samples run 5 times per day, over 3 days, on 3 reagent lots to give a total of 60 results per lot. The LoB was estimated for each lot as the 95th percentile of the distribution.

The Limit of Detection (LoD) was based on 4 serum samples run 5 times per day, over 3 days, on 3 reagent lots to give a total of 60 results per lot. The LoD calculation followed parametric analysis.

The LoQ was determined from 4 serum samples targeted to be close to the bottom of the measuring range, tested 5 times per day over 3 days on 2 reagent lots.

For LoB and LoD, the highest result obtained from the two lots tested was the final result (see below):

LoBLoDLoQ
0.15mg/L0.17mg/L0.5mg/L

e. Analytical specificity:

Interferences were assessed according to CLSI EP7-A2 by testing serum samples with CRP concentrations targeted at <1mg/L, 1-3mg/L, 3mg/L and >3mg/L. Each sample was spiked with interfering substances and tested. For non-interference to be claimed, the mean results from the spiked samples must be within 10% of the mean of the control samples.

Haemoglobin and Bilirubin results: The data demonstrated that the assay was not affected by high levels of the following substances: haemoglobin (5g/L), conjugated bilirubin (200mg/L) unconjugated bilirubin (200mg/L) and rheumatoid factor (25lU/mL).

{8}------------------------------------------------

Intralipid and Triglyceride results: The data demonstrated that the assay was not affected by triglyceride (1000mg/dL), and intralipid (2000mg/dL).

Drug Interference results: The data demonstrated that the assay was not affected by the 6 therapeutic drugs tested at the concentrations given below.

DrugConcentration tested
Acetaminophen1324µmol/L
Acetylsalicylic Acid3.62mmol/L
Caffeine308µmol/L
Enalapril Maleate0.86µmol/L
Ibuprofen2425µmol/L
Warfarin32.5µmol/L

f. Assay cut-off:

Refer to Expected Values.

2. Comparison studies:

  • a. Method comparison with predicate device:
    A total of 391 serum samples, 302 of which reported results within the measuring range, were assayed in singlicate by both the Optilite High Sensitivity C-Reactive Protein Kit and the C-Reactive Protein High Sensitive Test System for Cobas Integra. The samples included 87 normal donors and 304 clinical samples.

Regression statistics are based on the balance of the paired results.

Results (n = 302)
AnalysisResult95% CI Slope95% CI InterceptPass/Fail
Passing & Babloky = 0.99x + 0.090.97 to 1.010.05 to 0.11PASS
Weighted Demingy = 0.97x + 0.100.94 to 0.990.07 to 0.13
Weighted Lineary = 0.95x + 0.120.93 to 0.980.10 to 0.15

b. Matrix comparison:

None

{9}------------------------------------------------

3. Clinical studies:

  • a. Clinical Sensitivity:
    Not applicable

  • b. Clinical specificity:
    Not applicable

c. Other clinical supportive data (when a. and b. are not applicable):

Not applicable

4. Clinical cut-off:

Not applicable

5. Expected values/Reference range:

The reference interval was verified by testing 50 adult donor samples.

Consensus reference interval for adults: <3 mg/L

Reference: Macy EM, Hayes TE, Tracy RP. Variability in the measurement of C-Reactive protein in healthy subjects; implications for reference intervals and epidemiological applications. Clin Chem 1997; 43: 52-8.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The summary includes the conclusions drawn from the nonclinical and clinical tests (discussed above) that demonstrate that the device is as safe, as effective, and performs as well as or better than the predicate device

§ 866.5270 C-reactive protein immunological test system.

(a)
Identification. A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.(b)
Classification. Class II (performance standards).