Search Filters

Search Results

Found 9 results

510(k) Data Aggregation

    K Number
    K240609

    Validate with FDA (Live)

    Date Cleared
    2024-07-17

    (135 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System Zygomatic implants are intended for placement in the maxillary arch to provide support for fixed or removable dental prostheses in patients with partially or fully edentulous maxillae. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System Zygomatic implants are intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    Device Description

    The purpose of this submission is to add components to the S.I.N. Dental Implant System. This submission includes zygomatic dental implants and mating abutments for screw-retained, multi-unit prostheses.

    The subject device Zygomatic Plus Morse Taper implants have a body diameter of 4.0 mm that tapers to approximately 2.3 mm at the apex. The implants have a Morse taper ("CM") abutment interface connection, with an internal 16° cone taper. The abutment platform is 4.0 mm. All implants are threaded starting at the apex and extending approximately 17 mm in the coronal direction, with a major thread diameter of 4 mm; the thread tapers over the final 6 mm of the implant to the apex. The implants are provided with overall lengths of 30, 32.5. 35. 37.5. 40. 42.5.45. 47.5. 50. 52.5.57.5. and 60 mm. The surface of all subject device implants is acid-etched from the implant platform to the apex, followed by application of a hydroxyapatite coating (HA nano).

    The subject device abutments are multi-unit, indexed abutments for use only with the subject device Zygomatic Plus Morse Taper implants for maxillary full-arch restorations. The subject device abutments have a prosthetic platform diameter of 4.8 mm, and gingival heights of 1.5, 2.0, and 2.5 mm. The prosthetic post is angled 52° or 60° to the long axis of the implant. These components are used with previously cleared abutment screws.

    All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67. All implants have an acid-etched HA100 surface treatment, identical to that cleared in K231127. All subject device abutments are manufactured from titanium alloy (Ti-6A1-4V) conforming to ASTM F136. All subject device implants and abutments are provided sterile to the end user.

    AI/ML Overview

    The provided text is a 510(k) Summary for a dental implant system. It explicitly states that no clinical data were included in this submission to support substantial equivalence (see "No clinical data were included in this submission." on page 6 of 9).

    Therefore, it is not possible to provide specific details about acceptance criteria, device performance, sample sizes, expert ground truth establishment, adjudication methods, or MRMC studies, as这些 types of studies were not conducted or reported in this 510(k) submission.

    The "Performance Data" section (pages 6-7 of 9) refers to non-clinical data only:

    • mechanical testing
    • gamma irradiation sterilization
    • bacterial endotoxin testing
    • MR environment evaluation
    • sterile barrier shelf life data
    • biocompatibility data

    These tests are typically conducted to assess the physical and biological characteristics of the device, not its clinical performance in patients as would be measured in an AI/MRMC study.

    In summary, based on the provided document, no study was conducted to prove the device meets acceptance criteria related to AI/MRMC performance as outlined in your request. The submission relies on substantial equivalence to predicate devices, supported by non-clinical performance data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241378

    Validate with FDA (Live)

    Date Cleared
    2024-07-12

    (58 days)

    Product Code
    Regulation Number
    880.6850
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

    The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

    Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

    S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

    S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

    The combined weight of the Epikut PTG Surgical Kit and associated instruments is 632 grams.

    The weight of the empty Epikut PTG Surgical Kit is 353 grams.

    Device Description

    The subject device is a reusable rigid container, comprising a base (bottom), a removable inner tray, and a lid (cover). The subject device tray is to be used to organize and protect the instruments that are sterilized by the healthcare provider. The base, inner tray, and lid components are designed to be integrated into a single unit which contains and protects the interior components during sterilization. The tray is perforated to allow for penetration of the sterilant, is to be used with moist heat (steam) and requires the use of an FDA cleared wrap to maintain sterility. The subject device components are manufactured from injection molded polysulfone (PSU), and holders of various geometries to position instruments in the kits are manufactured from silicone.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for a medical device (S.I.N. Instrument Kits, an instrument sterilization tray). This document describes the nonclinical testing and acceptance criteria for mechanical and sterilization performance of the device itself, rather than the performance of an AI/ML algorithm.

    Therefore, the specific information requested in your prompt regarding acceptance criteria and study design for an AI/ML device (e.g., sample sizes, expert ground truth, MRMC studies, standalone algorithm performance, training set details) cannot be extracted from this document. This document describes a traditional medical device, not an AI/ML enabled one.

    However, I can extract the acceptance criteria and performance data for the device itself based on the provided "Summary of Nonclinical Testing" table.


    Acceptance Criteria and Device Performance (Non-AI/ML)

    The S.I.N. Instrument Kits are sterilization trays. The studies performed are to ensure the trays can be properly cleaned, sterilized, and maintain their integrity over a specified lifespan.

    1. Table of acceptance criteria and the reported device performance:

    Test MethodologyPurposeAcceptance CriteriaReported Device Performance
    Manual Cleaning ValidationTo validate that the cleaning instructions appropriately clean the tray, and to ensure the sterilization cycle will be effective.Protein assay: No color change = absence of protein residue (Sensitivity = 1 µg protein residue) Total organic carbon assay: results < limit of quantification = 0.1 ppm; Assay limit of detection = 0.015 ppmPass
    Bacterial Endotoxin Testing, USP <85>To validate that the cleaning instructions appropriately clean the tray, and to ensure the BET level meets FDA expectation (≤ 20 EU/device).Endotoxin results: ≤ 20 EU/devicePass
    Sterilization Validation (including sterilant penetration and dry time validation)To validate that the sterilization instructions appropriately sterilize the tray and contents.3 consecutive half-cycles performed for each of 3 sizes of trays demonstrate complete inactivation of all biologic indicators. A minimum SAL (Sterility Assurance Level) of 10^-6 is achieved if the Instructions for Use are followed.Pass
    Dry TimeTo validate that the sterilization instructions appropriately dry the wrapped tray for storage.Using pre-cycle and post-cycle weights, the weight gain after drying will be ≤ 3%.Pass
    Life Cycle / Simulated Use-life ValidationTo validate the service life of the trays as stated in the Instructions for Use.Visual inspection, component dimensional fit verification, functional closure/latch verification for 250 use cycles.Pass
    Biocompatibility of Subject Device (Cytotoxicity testing)To evaluate the cytotoxicity potential of the test article using an in vitro cell culture assay.Non-cytotoxic if: ≤ 50% of L-929 cells exposed to extract of device show rounding and lysis.Pass

    Since this document does not pertain to an AI/ML device, the following points typical for AI/ML device evaluations are not applicable and thus, no information can be extracted for them:

    • 2. Sample size used for the test set and the data provenance: Not applicable as this is not an AI/ML study.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    • 7. The type of ground truth used: Not applicable (for AI/ML, but for a physical device, the "ground truth" is established through standardized testing procedures and specified acceptance criteria, as shown in the table).
    • 8. The sample size for the training set: Not applicable.
    • 9. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K232821

    Validate with FDA (Live)

    Date Cleared
    2024-05-31

    (261 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    Unitite Slim, Unitite, Unitite Compact, Strong SW CM S.I.N Dental Implant System implants with lengths less than 7 mm are intended for delayed loading only.

    Epikut CM, Epikut Plus CM, Epikut S, Epikut S Plus S.I.N. Dental Implant System implants with lengths of 18, 20, 22, or 24 mm may be tilted up to 30°. When used in the mandible or maxilla with implants with lengths of 18, 20, 22, or 24 mm at an angulation of 30°, a minimum of four implants must be used and must be splinted. When placed in the maxilla with lengths of 18, 20, 22, or 24 mm at angulations between 0° and less than 30°, the S.I.N. Dental Implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants.

    Epikut CM, Epikut Plus CM, Strong SW CM Plus, Strong SW HE Plus, Strong SW HI Plus All digitally-designed custom abutments for use with Interface CAD-CAM abutments are to be sent to a S.I.N.-validated milling center for manufacture.

    Device Description

    The purpose of this submission is to add labeling claims for a surface treatment designated HA™®, which is applied to implants of the S.I.N. Dental Implant System, as previously cleared in K170392. K193096. K200992, K211921, K221453, and K222231. The new labeling claims and the information supporting the claims are shown below in Performance Data.

    The following information is a summary of the designs of the subject device implants, as they have been previously cleared with the HA"" surface treatment. Note that there are no changes to any aspect of the design, nor to the HA" surface treatment, as described in the respective 510(k) Premarket Notifications referenced above.

    The previously cleared implants that are the subject of this submission includes eleven (11) product lines from six (6) 510(k)s. Implant body diameters, platform diameters and lengths for each product line are shown below in Table 10.1 Summary of Subject Device Designs and Sizes.

    K170392: Unitite Slim, Unitite, Unitite Compact K193096: Strong SW CM Plus, Strong SW HE Plus, Strong SW HI Plus K200992: Strong SW Plus K211921: Epikut Plus CM, Epikut Plus HE K221453: Epikut Plus CM (additional lengths) K22231: Epikut S Plus

    All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67. The HA1000 surface treatment is identical to that cleared in K211921.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the U.S. FDA for the S.I.N. Dental Implant System. It primarily focuses on demonstrating substantial equivalence to previously cleared predicate devices by comparing their indications for use, designs, materials, and manufacturing processes, with an emphasis on adding new labeling claims related to a surface treatment.

    Crucially, this document states: "No clinical data were included in this submission." This means that the device's performance was not proven through a study involving human subjects or artificial intelligence (AI). The document relies on non-clinical data such as in vitro testing, animal testing referenced from published literature, sterilization, endotoxin, shelf-life, biological evaluation, MR compatibility, and fatigue testing to support the claims and substantial equivalence.

    Therefore, most of the requested information regarding acceptance criteria, study details, sample sizes, expert involvement, and specific performance metrics for an AI/device study cannot be extracted from this document.

    However, I can extract the information related to the non-clinical performance data presented to meet the criteria for substantial equivalence to predicates.

    Here's a breakdown of what can be extracted and an explanation of why other requested information is not present:

    Information that CANNOT be extracted from this document (and why):

    • A table of acceptance criteria and the reported device performance (for clinical/AI studies): Not applicable as no clinical or AI study was performed. The acceptance criteria relate to demonstrating substantial equivalence through non-clinical data, not specific performance metrics against a ground truth for a diagnostic claim.
    • Sample sizes used for the test set and the data provenance: Not applicable.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    • Adjudication method for the test set: Not applicable.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable (not an AI device).
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.) for test set: Not applicable as no clinical study or diagnostic AI was evaluated.
    • The sample size for the training set: Not applicable (not an AI device).
    • How the ground truth for the training set was established: Not applicable (not an AI device).

    What CAN be extracted/inferred regarding "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the context of this 510(k) submission:

    The "acceptance criteria" for a 510(k) are typically met by demonstrating substantial equivalence to existing legally marketed devices (predicates). The "study that proves the device meets the acceptance criteria" in this case refers to the non-clinical tests and comparisons performed to show this equivalence and support the new labeling claims.

    1. A table of acceptance criteria and the reported "device performance" (in terms of non-clinical criteria):

    The acceptance criteria for this 510(k) submission are implicitly met by demonstrating that the subject device is as safe and effective as the predicate devices and that the new labeling claims are supported by scientific evidence. The "performance" here refers to meeting specific engineering, biological, and material standards, not diagnostic accuracy.

    Acceptance Criterion (Implicitly required for 510(k) & specific claims)Reported Device "Performance" (Evidence provided)
    Equivalence in Intended UseSubject device's Indications for Use (IFUS) are identical or substantially similar to those of predicate and reference devices (K170392, K193096, K200992, K211921, K221453, K222231). This includes general placement, loading conditions, specific length limitations for delayed loading, and angulation requirements for longer implants.
    Equivalence in Design/Physical CharacteristicsThe design (body/platform diameters, lengths, implant/abutment interface) of the subject device implants (including the new HA™® surface) is identical to those previously cleared in predicate/reference 510(k)s (K170392, K193096, K200992, K211921, K221453, K222231).
    Equivalence in MaterialsSubject device is manufactured from unalloyed titanium conforming to ASTM F67, identical to predicate devices. The HA™® surface treatment is identical to that cleared in K211921.
    Sterilization EfficacyDemonstrated by gamma irradiation sterilization to a sterility assurance level of 10^-6 by VDmax25 method, according to ISO 11137-1 and ISO 11137-2 (referenced from K211921).
    Absence of Pyrogens/EndotoxinsBacterial endotoxin testing (LAL test) according to ANSI/AAMI ST72 on manufacturing water (weekly) and sterilized product (quarterly) demonstrated all sterile product meets a limit of < 20 EU/device (referenced from K211921).
    Shelf-Life StabilityShelf life testing, including testing of samples after 4 years of real-time aging according to ASTM F1929 and F88/F88M (packaging sterile barrier) and sterility testing of product (referenced from K211921).
    BiocompatibilityBiological evaluation performed according to ISO 10993-1, with test results leveraged from K170392 for ISO 10993-3 (genotoxicity), ISO 10993-5 (cytotoxicity), ISO 10993-6 (implantation), ISO 10993-10 (sensitization, irritation), and ISO 10993-11 (systemic toxicity) to support biocompatibility.
    MRI Safety and CompatibilityNon-clinical analysis and testing according to ASTM F2052 (magnetically induced displacement force), ASTM F2213 (magnetically induced torque), ASTM F2182 (RF induced heating), and ASTM F2119 (image artifact), and the FDA guidance document "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment" (issued May 2021). No specific results are provided in the summary.
    Mechanical Strength/FatigueFatigue testing according to ISO 14801 referenced from K170392, K193096, K200992, K211921, K221453, and K222231 is applicable to the subject device (the specific results or "performance" against a benchmark are not provided in the summary, just that such testing was leveraged).
    Support for HAnano Surface ClaimsSurface Wettability (Hydrophilicity): Referenced Bezerra et al., 2017. Page 6: "Water contact angle was significantly less on HAnano treated titanium surface than on dual acid-etched titanium surface."Osteoblast Behavior/Differentiation: Referenced Bezerra et al., 2017 (Pages 4, 10) and da Silva et al., 2020 (Page 7). Claims include greater transition to differentiation-related signaling, changes in osteogenic gene marker expression, and spreading morphology different from machined/hydrophilic dual acid-etched surfaces. Results of in vitro testing are not necessarily predictive of human clinical performance.

    2. Sample sized used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Sample Size: Not applicable. The "test set" here would refer to the samples used in the non-clinical tests (e.g., in vitro cell cultures, physical test samples, animal models where applicable for referenced studies). Exact sample sizes for these referenced non-clinical studies are not provided in this summary.
    • Data Provenance:
      • Country of Origin: Not specified for the referenced studies (Bezerra et al., da Silva et al.). The manufacturing country for the device is Brazil (S.I.N. - Sistema de Implante Nacional S.A., São Paulo, Brazil).
      • Retrospective or Prospective: Not applicable as these were non-clinical, in vitro, or animal studies, not human clinical trials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as this is a non-clinical submission for a dental implant, not an AI/diagnostic device requiring expert ground truth for imaging interpretation. The "ground truth" for the non-clinical properties is established by the methods, standards, and scientific literature cited (e.g., ASTM, ISO standards, and published research papers).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. There was no expert reader study.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC study was not done. This is not an AI-enabled device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No, a standalone algorithm performance study was not done. This is not an AI-enabled device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for the claims made are based on non-clinical in vitro and animal studies, and adherence to recognized standards for material properties, sterilization, biocompatibility, and mechanical performance. Specifically for the HAnano claims, the ground truth is derived from the experimental results reported in the referenced scientific publications (e.g., measurements of water contact angle, observation of cell morphology, assessment of gene marker expression).

    8. The sample size for the training set:

    • Not applicable. This is not an AI device trained on a dataset.

    9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K231127

    Validate with FDA (Live)

    Date Cleared
    2023-12-07

    (231 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System Zygomatic implants are intended for placement in the maxillary arch to provide support for fixed or removable dental prostheses in patients with partially or fully edentulous maxillae. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System Zygomatic implants are intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    Device Description

    The purpose of this submission is to add components to the S.I.N. Dental Implant System. This submission includes zygomatic dental implants, mating abutments for screw-retained, multi-unit prostheses, and the corresponding abutment screws.

    The subject device Zygomatic Plus Morse Taper implants have a body diameter of 4.0 mm that tapers to 1.65 mm at the apex. The implants have a Morse taper ("CM") abutment interface connection, with an internal 16° cone taper. The abutment platform is 4.0 mm. All implants are threaded starting at the apex and extending approximately 17 mm in the coronal direction, with a major thread diameter of 4 mm; the thread tapers over the final 6 mm of the implant to the apex. The implants are provided with overall lengths of 30, 32.5, 35, 37.5, 40, 42.5, 45, 47.5, 50, 52.5, 57.5, and 60 mm. The surface of all subject device implants is acid-etched from the implant platform to the apex, followed by application of a hydroxyapatite coating (HA nano).

    The subject device abutments are multi-unit, indexed abutments for use only with the subject device Zygomatic Plus Morse Taper implants for maxillary full-arch restorations. The subject device abutments have a prosthetic platform diameter of 4.8 mm, and gingival heights of 2 mm and 3 mm. The prosthetic post is angled 45° to the long axis of the implant. These components are used with subject device abutment and prosthetic screws, or previously cleared screws.

    All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67. All implants have an acid-etched HA1200 surface treatment, identical to that cleared in K200992. All subject device abutments are manufactured from titanium alloy (Ti-6A1-4V) conforming to ASTM F136. All subject device implants, abutments, and abutment screws are provided sterile to the end user.

    AI/ML Overview

    I am sorry, but based on the provided text, I cannot provide a comprehensive description of the acceptance criteria and the study that proves the device meets them, as requested. The document is an FDA 510(k) summary for a dental implant system. While it mentions performance data, it primarily focuses on establishing substantial equivalence to predicate devices rather than providing detailed acceptance criteria and a full study report of the subject device's performance against those criteria.

    Here's why I cannot fulfill each requested point:

    1. A table of acceptance criteria and the reported device performance: The document does not explicitly state acceptance criteria for the new device in a quantifiable way beyond citing compliance with standards like ISO 11137-1, ISO 11137-2, ASTM F2052, ASTM F2213, ASTM F2182, ASTM F2119, ASTM F67, and ASTM F136. It mentions mechanical testing was done "to demonstrate that the subject device abutments... have sufficient strength for the intended use," but it doesn't provide specific numerical acceptance limits or the reported performance values against them.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): The text states "Non-clinical data provided in this submission...mechanical testing conducted using a method modified from ISO 14801 and engineering analysis," but it does not specify the sample size used for this mechanical testing. There is no information about data provenance (country of origin, retrospective/prospective). No clinical data was included in this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): This information is not relevant to the type of non-clinical, mechanical, and sterilization testing described in the document. No expert review of a "test set" in the context of clinical interpretation or diagnosis is mentioned.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable, as no expert "test set" for clinical evaluation is mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. The device is a dental implant system, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. The device is a dental implant system, not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): The "ground truth" here would be the physical properties and performance characteristics determined by established engineering and sterilization standards (e.g., sterility assurance level, mechanical strength, biocompatibility, MR compatibility). These are not expert consensus in a clinical sense, pathology, or outcomes data.

    8. The sample size for the training set: Not applicable. This is a physical medical device, not an AI/machine learning model that requires a training set.

    9. How the ground truth for the training set was established: Not applicable, for the same reason as point 8.

    The document focuses on demonstrating that the S.I.N. Dental Implant System (specifically the new zygomatic implants and abutments) is substantially equivalent to legally marketed predicate devices through non-clinical data, including:

    • Gamma irradiation sterilization validation.
    • Mechanical testing (modified from ISO 14801).
    • MR compatibility testing (referenced from K222231 and standards ASTM F2052, ASTM F2213, ASTM F2182, ASTM F2119).
    • Sterile barrier shelf life data (referenced from K203725).
    • Biocompatibility data (referenced from K200992 and standards ASTM F67, ASTM F136).

    The substantial equivalence argument highlights similarities in indications for use, design principles, materials, manufacturing, sterilization, and shelf life to various predicate and reference devices. However, it does not provide detailed acceptance criteria and specific performance results for the subject device beyond qualitative statements like "sufficient strength."

    Ask a Question

    Ask a specific question about this device

    K Number
    K222231

    Validate with FDA (Live)

    Date Cleared
    2022-10-27

    (94 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    S.I.N. Dental Implants with lengths of 18, 20, 22, or 24 mm be tilted up to 30°. When used in the mandible or maxilla with implants with lengths of 18, 20, 22, or 24 mm at an angulation of 30°, a minimum of four implants must be used and must be splinted. When placed in the maxilla with lengths of 18, 20, 22, or 24 mm at angulations between 0° and less than 30°, the S.I.N. Dental Implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants.

    Device Description

    The purpose of this submission is to add components to the S.I.N. Dental Implant System, which includes components cleared previously in K211921, K200992, and K170392.

    This submission includes dental implants Epikut S with a Morse taper (CM) abutment interface and an acid-etched endosseous surface, and Epikut S Plus implants with an endosseous surface produced by acidetching followed by application of a hydroxyapatite coating (HA™M). The implant design and endosseous surfaces are nearly identical to the Epikut Plus CM implants cleared in K211921, with the exception of the implant-abutment connection (subject device 16° Morse taper, K211921 11.5° Morse taper), the additional body/platform diameter (4.0 mm), and the longer lengths (18, 20, 22, and 24 mm).

    All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67. The acid etching procedure is applied to all subject device dental implants. The acid etching process in this submission is identical to the process used to manufacture the dental implants cleared in K211921, and the HA100 surface treatment is identical to that cleared in K211921.

    AI/ML Overview

    This FDA 510(k) summary (K222231) describes a dental implant system and focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than on presenting a study with specific acceptance criteria, test sets, or performance metrics in the way these terms are typically used for AI/ML-based diagnostic devices.

    The document is a regulatory submission for a medical device (S.I.N. Dental Implant System), not a study evaluating algorithm performance. Therefore, it does not contain information about:

    • A table of acceptance criteria and reported device performance in the context of an algorithm's diagnostic accuracy.
    • Sample sizes, data provenance, number of experts, adjudication methods for a test set.
    • Multi-Reader Multi-Case (MRMC) comparative effectiveness studies.
    • Standalone algorithm performance.
    • Sample size for training sets or how ground truth for training sets was established.

    Instead, the document focuses on non-clinical performance data to show that the new components of the S.I.N. Dental Implant System (subject device) are as safe and effective as existing legally marketed devices.

    Here's an breakdown of the acceptance criteria (or rather, the demonstration of equivalence) and the studies (non-clinical tests and analyses) that prove the device meets these criteria, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria here are based on demonstrating that the subject device's design, materials, manufacturing, sterilization, and performance characteristics are substantially equivalent to predicate devices already cleared by the FDA. The performance data presented are non-clinical and relate to the physical and biological properties of the implant.

    Acceptance Criteria (Demonstrated Equivalence to Predicate Devices)Reported Device Performance (as demonstrated by non-clinical studies/analysis)
    Intended Use (Functional and esthetic rehabilitation of edentulous mandible or maxilla)Equivalent to K211921, K203725, K170392, and K050406. The subject device's Indications for Use statement is largely identical or similar to the predicate devices, with specific language related to tilting angles and required number of implants/splinting justified by K050406 and K203725.
    Material Composition (Unalloyed titanium, ASTM F67)Equivalent to K211921, K203725, K170392. All subject device implants are manufactured from the same material.
    Endosseous Surface Treatment (Acid-etched; HA™ for Epikut S Plus)Equivalent to K211921 and K170392 for acid-etching. The HA™ treatment is identical to that cleared in K211921. Characterization (SEM, XPS, TEM, XRD, adherence) from K211921 leveraged.
    Sterilization Method & Sterility Assurance Level (SAL) (Gamma irradiation to 10⁻⁶ SAL at 25 kGy)Equivalent to K211921 and K203725. Sterilization method selected and substantiated according to ISO 11137-1 and ISO 11137-2.
    Bacterial Endotoxin Limit (< 20 EU/device)Equivalent to K211921. Demonstrated via Limulus amebocyte lysate (LAL) testing according to ANSI/AAMI ST72 on samples from manufacturing water (weekly) and sterilized product (quarterly).
    Shelf Life/Packaging Integrity (4 years real-time aging)Equivalent to K211921. Testing of samples after 4 years of real-time aging according to ASTM F1929 and F88/F88M (packaging sterile barrier) and sterility testing of product.
    BiocompatibilityEquivalent to K211921. Biological evaluation performed according to ISO 10993-1, with test results leveraged from K211921.
    MR Safety (Magnetically induced displacement force, torque, RF heating, image artifact)Non-clinical analysis and testing performed according to ASTM F2052, ASTM F2213, ASTM F2182, ASTM F2119, and FDA guidance. Results demonstrate safety in the MR environment.
    Mechanical Performance / Implant-Abutment Connection (Engineering analysis of new components)Equivalent to K200992 (for mechanical testing applicability). Engineering analysis demonstrated that the subject device implants, with compatible previously-cleared abutments, do not create a new worst-case construct, thus making previous mechanical testing applicable. (Note: Specific mechanical test results/values are not provided in this summary, but the applicability of prior testing is the demonstration of equivalence here).
    Physical Dimensions (Body/platform diameters and lengths, including new 4.0mm diameter and 18-24mm lengths)Equivalent or within the range of previously cleared devices. New 4.0mm diameter is within the range of K170392. 18-24mm lengths are similar to K203725.

    2. Sample Size Used for the Test Set and Data Provenance

    This is not applicable as this document describes a 510(k) submission for a physical medical device, not an algorithm's test set. The "test sets" here refer to batches of physical devices or materials undergoing non-clinical laboratory testing. The provenance of the data is the manufacturer (S.I.N. - Sistema de Implante Nacional S.A.) or its testing facilities, and the studies are non-clinical evaluations.

    3. Number of Experts Used to Establish the Ground Truth and Qualifications

    This is not applicable for a non-clinical device submission. Ground truth, in this context, would relate to objective measurements against established standards (e.g., ASTM, ISO standards) rather than expert consensus on medical images.

    4. Adjudication Method

    Not applicable. Adjudication methods are typically associated with resolving discrepancies in expert interpretations during clinical or algorithm performance studies. The non-clinical tests described have defined protocols and objective pass/fail criteria.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This type of study is relevant for evaluating the impact of an AI algorithm on human reader performance, which is not the subject of this 510(k) submission for a dental implant.

    6. Standalone Performance Study

    No. This refers to the performance of an algorithm without human intervention. This document details the performance characteristics of a physical implant device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical tests is established by compliance with recognized consensus standards (e.g., ISO, ASTM, AAMI) and the device's adherence to its design specifications, materials testing, and performance against these standards. For example:

    • Sterility: Achieved according to ISO 11137-1 and ISO 11137-2.
    • Bacterial Endotoxin: Meets limits (< 20 EU/device) as per ANSI/AAMI ST72.
    • Packaging: Integrity verified against ASTM F1929 and F88/F88M.
    • Biocompatibility: Evaluated against ISO 10993-1.
    • MR Safety: Assessed against ASTM F2052, F2213, F2182, F2119.
    • Mechanical Performance: Demonstrated through engineering analysis applying previous mechanical testing (K200992).
    • Material Composition: Conforms to ASTM F67.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This is not an AI/ML device.

    In summary, this 510(k) pertains to a physical dental implant and uses non-clinical testing and engineering analysis to establish substantial equivalence to previously cleared predicate devices, rather than performing clinical studies or AI/ML algorithm validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222005

    Validate with FDA (Live)

    Date Cleared
    2022-09-29

    (84 days)

    Product Code
    Regulation Number
    880.6850
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

    The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

    Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

    S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

    S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

    Device Description

    The subject device includes a total of three (3) instrument trays. The subject device trays are reusable rigid containers, comprising a base (bottom), one or more removable inner trays, and a lid (cover). The subject device trays are to be used to organize and protect the instruments that are sterilized by the healthcare provider. The base, inner tray, and lid components are designed to be integrated into a single unit which contains and protects the interior components during sterilization. The trays are perforated to allow for penetration of the sterilant, are to be used with moist heat (steam), and require the use of an FDA cleared wrap to maintain sterility. The subject device components are manufactured from injection molded polysulfone (PSU), and holders of various geometries to position instruments in the kits are manufactured from silicone. The subject device includes two (2) sizes (lid, base, and enclosed volume), and a total of three (3) tray configurations.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the S.I.N. Instrument Kits.

    Acceptance Criteria and Reported Device Performance

    Test MethodologyPurposeAcceptance CriteriaReported Performance
    Manual Cleaning Validation (FDA Guidance Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling)Validate cleaning instructions and ensure sterilization cycle effectiveness.Protein assay: No color change = absence of protein residue (Sensitivity = 1 µg protein residue). Total organic carbon assay: Results < limit of quantification = 0.1 ppm; Assay limit of detection = 0.015 ppm.Pass (For both assays)
    Bacterial Endotoxin Testing (USP <85>)Validate cleaning instructions and ensure BET level meets FDA expectation (≤ 20 EU/device).Endotoxin results ≤ 20 EU/device.Pass
    Sterilization Validation (ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO 17665-2)Validate sterilization instructions and ensure appropriate sterilization of tray and contents.3 consecutive half-cycles performed for each of 3 sizes of trays demonstrate complete inactivation of all biologic indicators; A minimum SAL of 10^-6 is achieved if the Instructions for Use are followed.Pass
    Dry Time (Referenced from K201688)Validate that sterilization instructions appropriately dry the wrapped tray for storage.Using pre-cycle and post-cycle weights, the weight gain after drying will be ≤ 3%.Pass
    Life Cycle / Simulated Use Life Validation (FDA Guidance Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling)Validate the service life of the trays as stated in the Instructions for Use.Visual inspection, component dimensional fit verification, functional closure/latch verification for 250 use cycles.Pass
    Biocompatibility of Subject Device (Cytotoxicity testing - ANSI/AAMI/ISO 10993-5, ANSI/AAMI/ISO 10993-12)Evaluate the cytotoxicity potential of the test article using an in vitro cell culture assay.Non-cytotoxic if ≤ 50% of L-929 cells exposed to extract of device show rounding and lysis.Pass

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Sample Size for Test Set: The document mentions "3 consecutive half-cycles performed for each of 3 sizes of trays" for Sterilization Validation. For Biocompatibility, it states "L-929 cells exposed to extract of device." The exact number of units or iterations per test beyond what is explicitly stated (e.g., how many trays were tested for cleaning, how many units for biocompatibility extract) is not explicitly detailed in the provided text.
      • Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective. It only lists the standards followed (e.g., FDA Guidance, USP, ANSI/AAMI/ISO).
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided in the document. The tests performed are laboratory-based and follow established standards, implying no human expert consensus was required to establish ground truth for these specific performance criteria.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • This information is not applicable as the tests are non-clinical, objective validations against predetermined criteria, not assessments requiring expert adjudication of subjective findings.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done. This device (S.I.N. Instrument Kits, which are instrument sterilization trays) is a physical medical device, not an AI-powered diagnostic tool. Therefore, the concept of "human readers improve with AI assistance" is not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No standalone (algorithm-only) performance study was done. As mentioned above, this is a physical medical device, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth for these non-clinical tests is based on objective measurements and adherence to established regulatory standards and protocols (e.g., protein residue levels, endotoxin levels, sterility assurance level (SAL), weight change, visual inspection, dimensional verification, cell viability).
    7. The sample size for the training set:

      • There is no training set as this is a physical medical device validation, not a machine learning model.
    8. How the ground truth for the training set was established:

      • This question is not applicable as there is no training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K212404

    Validate with FDA (Live)

    Date Cleared
    2021-11-03

    (93 days)

    Product Code
    Regulation Number
    880.6850
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

    Device Description

    The subject device includes a total of 4 instrument trays. The subject device trays are reusable rigid containers, comprising a base (bottom), one or more removable inner travs, and a lid (cover). The subject device trays are to be used to organize and protect the instruments that are sterilized by the healthcare provider. The base, inner tray, and lid components are designed to be integrated into a single unit which contains and protects the interior components during sterilization. The trays are perforated to allow for penetration of the sterilant, are to be used with moist heat (steam), and require the use of an FDA cleared wrap to maintain sterility. The subject device components are manufactured from injection molded polysulfone (PSU), and holders of various geometries to position instrument in the kits are manufactured from silicone. The subject device includes a total of 3 sizes (same lid, base, and enclosed volume), and a total of 4 tray configurations.

    AI/ML Overview

    The document provides information on the nonclinical testing performed for the S.I.N. Instrument Kits to demonstrate they meet acceptance criteria.

    1. A table of acceptance criteria and the reported device performance

    Test MethodologyAcceptance CriteriaResults
    Manual Cleaning ValidationProtein assay Acceptance criterion: No color change = absence of protein residue. Sensitivity = 1 µg protein residue.Total organic carbon assay Acceptance criterion: results < limit of quantification = 0.1 ppm; Assay limit of detection = 0.015 ppm.Pass
    Bacterial Endotoxin Testing (USP <85>)Endotoxin results ≤ 20 EU/device.Pass
    Sterilization Validation (ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO 17665-2)3 consecutive half-cycles performed for each of 3 sizes of trays demonstrate complete inactivation of all biologic indicators. A minimum SAL of 10-6 is achieved if the Instructions for Use are followed.Pass
    Dry TimeUsing pre-cycle and post-cycle weights, the weight gain after drying will be ≤ 3%.Pass
    Life Cycle / Simulated Use Life ValidationVisual inspection, component dimensional fit verification, functional closure/latch verification for 250 use cycles.Pass
    Biocompatibility of Subject Device (Cytotoxicity testing - ANSI/AAMI/ISO 10993-5, ANSI/AAMI/ISO 10993-12)Non-cytotoxic if ≤ 50% of L-929 cells exposed to extract of device show rounding and lysis.Pass

    2. Sample sizes used for the test set and the data provenance

    • Sterilization Validation: 3 consecutive half-cycles were performed for each of the 3 sizes of trays. This indicates a sample size of at least 9 sterilization cycles in total (3 sizes * 3 half-cycles).
    • Other tests: The exact sample sizes for Manual Cleaning Validation, Bacterial Endotoxin Testing, Dry Time, Life Cycle / Simulated Use Life Validation, and Biocompatibility are not explicitly stated by number, but the results indicate that sufficient samples were tested to meet the acceptance criteria.
    • Data Provenance: The document does not explicitly state the country of origin of the data or whether it was retrospective or prospective. Given that the manufacturer is S.I.N. - Sistema de Implante Nacional S.A. from Brazil, it is likely the testing was conducted in Brazil or at an international testing facility. The nature of the tests (laboratory testing of medical devices) implies a prospective test setup.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This section is not applicable. The tests performed are laboratory-based performances of a medical device (instrument kits for sterilization) against established regulatory standards and guidance, not diagnostic evaluations requiring expert consensus on "ground truth."

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This section is not applicable for this type of nonclinical, laboratory-based device performance testing. Ground truth is established by the predefined acceptance criteria of the test methodologies.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This type of study is not applicable as the device is a medical instrument kit for sterilization, not an AI-based diagnostic tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No. This concept is not applicable as the device is a physical medical instrument kit.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is based on established scientific and regulatory standards and methods for medical device sterilization, cleaning, biocompatibility, and durability. These include:

    • Predefined acceptance criteria based on standard test methods (e.g., absence of protein residue, endotoxin levels, sterility assurance level (SAL), weight gain after drying, visual inspection for integrity, cytotoxicity).
    • Compliance with recognized standards such as USP <85>, ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO 17665-2, ANSI/AAMI/ISO 10993-5, and ANSI/AAMI/ISO 10993-12.
    • FDA Guidance documents (e.g., "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling").

    8. The sample size for the training set

    This section is not applicable as the device is a physical medical instrument kit, not an AI/machine learning model that requires training data.

    9. How the ground truth for the training set was established

    This section is not applicable as the device is a physical medical instrument kit and does not involve a training set or AI model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211921

    Validate with FDA (Live)

    Date Cleared
    2021-10-20

    (121 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

    Device Description

    The purpose of this submission is to add components to the S.I.N. Dental Implant System components cleared in K200992, K193096, and K170398.

    This submission includes dental implants Epikut CM with a Morse taper (CM) abutment interface and a double acid-etched endosseous surface, and Epikut Plus CM implants with an endosseous surface produced by double acid-etching followed by application of a hydroxyapatite coating (HA™Y). These implant endosseous surfaces are identical to those cleared in K200992.

    This submission also includes dental implants Epikut HE with an external hexagon (HE) abutment interface and a double acid-etched endosseous surface, and Epikut Plus HE implants with an endosseous surface produced by double acid-etching followed by application of a hydroxyapatite coating (HA1000). These implant endosseous surfaces also are identical to those cleared in K200992.

    All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67. The acid etching procedure is applied to all subject device dental implants. The acid etching process in this submission is identical to the process used to manufacture the dental implants cleared in K200992. K193096, and K170398. The Epikut Plus CM and Epikut Plus HE implants have the identical HA™ surface treatment cleared in K200992 and K193096.

    The subject device dental implants are compatible with abutments and prosthetic components cleared previously in K200992. K193096. K170398. and K051859. Epikut HE and Epikut Plus HE dental implants are not for use with angled abutments.

    AI/ML Overview

    The provided document (K211921 510(k) Summary for S.I.N. Dental Implant System) is a 510(k) premarket notification summary submitted to the FDA. It does not describe an AI medical device or a clinical study intended to prove its performance against acceptance criteria for such a device.

    Instead, this document describes a dental implant system and its substantial equivalence to previously cleared predicate devices. The performance data presented are non-clinical bench tests (sterilization, bacterial endotoxin, shelf-life, biocompatibility, and coating characterization) to demonstrate the device's physical and biological properties.

    Therefore, I cannot directly answer your prompt using the provided text, as the information regarding acceptance criteria, AI device performance, sample sizes for test sets, expert ground truth, adjudication methods, MRMC studies, standalone performance, and training set details are not applicable to the content of this document.

    The document explicitly states: "No clinical data were included in this submission." and focuses on demonstrating substantial equivalence through non-clinical data and comparisons to predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201688

    Validate with FDA (Live)

    Date Cleared
    2020-10-21

    (121 days)

    Product Code
    Regulation Number
    880.6850
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    S.I.N. Instrument Kits are intended to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

    Device Description

    The subject device includes a total of 14 instrument trays. The subject device trays are reusable rigid containers, comprising a base (bottom), one or more removable inner trays, and a lid (cover). The subject device trays are to be used to organize and protect the instruments that are sterilized by the healthcare provider. The base, inner tray, and lid components are designed to be integrated into a single unit which contains and protects the interior components during sterilization. The trays are perforated to allow for penetration of the sterilant, are to be used with moist heat (steam), and require the use of an FDA cleared wrap to maintain sterility. The subject device components are manufactured from injection molded polysulfone (PSU), and holders of various geometries to position instrument in the kits are manufactured from silicone. The subject device includes a total of 5 sizes (same lid, base, and enclosed volume), and a total of 14 tray configurations.

    AI/ML Overview

    The provided text describes the acceptance criteria and a non-clinical study to prove that the S.I.N. Instrument Kits meet these criteria for their intended use as sterilization wraps/containers.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test MethodologyPurposeAcceptance CriteriaReported Device Performance
    Manual Cleaning Validation (FDA Guidance: Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, March 2015)To validate that the cleaning instructions provided in the Instructions for Use appropriately clean the tray, and to ensure the sterilization cycle will be effective.Protein assay: No color change = absence of protein residue (Sensitivity = 1 µg protein residue)Total organic carbon assay: results < limit of quantification = 0.1 ppm; (Assay limit of detection = 0.015 ppm)Pass
    Bacterial Endotoxin Testing (USP <85>)To validate that the cleaning instructions provided in the Instructions for Use appropriately clean the tray, and to ensure the BET level meets FDA expectation (≤ 20 EU/device).Endotoxin results ≤ 20 EU/devicePass
    Sterilization Validation (including sterilant penetration and dry time validation) (ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO 17665-2)To validate that the sterilization instructions listed in the Instructions for Use appropriately sterilize the tray and contents.3 consecutive half-cycles performed for each of 3 sizes of trays demonstrate complete inactivation of all biologic indicators; A minimum SAL of 10^-6 is achieved if the Instructions for Use are followed.Pass
    Dry time (part of Sterilization Validation)To validate that the sterilization instructions listed in the Instructions for Use appropriately dry the wrapped tray for storage.Using pre-cycle and post-cycle weights, the weight gain after drying will be ≤ 3%.Pass
    Life Cycle / Simulated Use Life Validation (FDA Guidance: Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, March 2015)To validate the service life of the trays as stated in the Instructions for Use.Visual inspection, component dimensional fit verification, functional closure/latch verification for 250 use cycles.Pass
    Biocompatibility of Subject Device (by cytotoxicity testing) (ANSI/AAMI/ISO 10993-5, ANSI/AAMI/ISO 10993-12)To evaluate the cytotoxicity potential of the test article using an in vitro cell culture assay.Non-cytotoxic if ≤ 50% of L-929 cells exposed to extract of device show rounding and lysis.Pass

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Test Set:
      • For Sterilization Validation, the test set involved "3 sizes of trays". The document lists 5 overall tray dimensions and 14 configurations, suggesting that not all configurations were tested for sterilization, but rather representative sizes.
      • For Life Cycle / Simulated Use Life Validation, the testing involved simulating 250 use cycles.
      • Specific quantities for other tests (Manual Cleaning, Bacterial Endotoxin, Biocompatibility) are not explicitly stated, but the "Pass" results imply adequate testing was performed.
    • Data Provenance: The document states the manufacturer is S.I.N. - Sistema de Implante Nacional S.A. based in São Paulo, Brazil. The testing described is non-clinical, implying laboratory-based studies rather than patient data. The document does not specify if the testing was performed retrospectively or prospectively, but standard validation studies like these are typically conducted prospectively.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

    This information is not applicable to this study. The study involves non-clinical performance testing of a medical device (sterilization instrument kits), not an AI algorithm or diagnostic device that requires human expert review to establish ground truth from patient data. The "ground truth" here is based on objective, quantifiable measures against established industry standards for sterility, cleanliness, and material properties.

    4. Adjudication Method for the Test Set:

    This information is not applicable as the study involves objective, non-clinical laboratory testing, not subjective human-in-the-loop performance evaluation that would require an adjudication method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No, an MRMC comparative effectiveness study was not conducted. This type of study is relevant for diagnostic devices (especially AI-assisted models) where human readers (e.g., radiologists) interpret images or data, and their performance is compared with and without AI assistance. This submission describes the non-clinical performance of a sterilization instrument kit, which does not involve human interpretation in the same way. The document explicitly states: "No clinical data were included in this submission."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This question is not applicable as the device is a physical instrument kit for sterilization, not an AI algorithm or software. The testing performed is to validate the physical and functional aspects of the kit itself.

    7. The Type of Ground Truth Used:

    The ground truth used for this device is based on objective, quantifiable measurements and adherence to established industry standards for medical device reprocessing and biocompatibility. Specifically:

    • Sterility: Sterility Assurance Level (SAL) of 10^-6, complete inactivation of all biologic indicators.
    • Cleanliness: Absence of protein residue (no color change in protein assay), Total Organic Carbon (TOC) assay results below limit of quantification.
    • Endotoxin Levels: Endotoxin results ≤ 20 EU/device.
    • Dryness: Weight gain after drying ≤ 3%.
    • Physical integrity and functionality: Visual inspection, component dimensional fit verification, and functional closure/latch verification after simulated use cycles.
    • Biocompatibility: Non-cytotoxic according to L-929 cell exposure.

    These are not "expert consensus", "pathology", or "outcomes data" in the typical sense applied to diagnostic AI models.

    8. The Sample Size for the Training Set:

    This information is not applicable. This is a physical medical device, not an AI model that requires a training set of data.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable as there is no AI model or training set involved.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1