K Number
K201688

Validate with FDA (Live)

Date Cleared
2020-10-21

(121 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

S.I.N. Instrument Kits are intended to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

Device Description

The subject device includes a total of 14 instrument trays. The subject device trays are reusable rigid containers, comprising a base (bottom), one or more removable inner trays, and a lid (cover). The subject device trays are to be used to organize and protect the instruments that are sterilized by the healthcare provider. The base, inner tray, and lid components are designed to be integrated into a single unit which contains and protects the interior components during sterilization. The trays are perforated to allow for penetration of the sterilant, are to be used with moist heat (steam), and require the use of an FDA cleared wrap to maintain sterility. The subject device components are manufactured from injection molded polysulfone (PSU), and holders of various geometries to position instrument in the kits are manufactured from silicone. The subject device includes a total of 5 sizes (same lid, base, and enclosed volume), and a total of 14 tray configurations.

AI/ML Overview

The provided text describes the acceptance criteria and a non-clinical study to prove that the S.I.N. Instrument Kits meet these criteria for their intended use as sterilization wraps/containers.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

Test MethodologyPurposeAcceptance CriteriaReported Device Performance
Manual Cleaning Validation (FDA Guidance: Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, March 2015)To validate that the cleaning instructions provided in the Instructions for Use appropriately clean the tray, and to ensure the sterilization cycle will be effective.Protein assay: No color change = absence of protein residue (Sensitivity = 1 µg protein residue)Total organic carbon assay: results < limit of quantification = 0.1 ppm; (Assay limit of detection = 0.015 ppm)Pass
Bacterial Endotoxin Testing (USP <85>)To validate that the cleaning instructions provided in the Instructions for Use appropriately clean the tray, and to ensure the BET level meets FDA expectation (≤ 20 EU/device).Endotoxin results ≤ 20 EU/devicePass
Sterilization Validation (including sterilant penetration and dry time validation) (ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO 17665-2)To validate that the sterilization instructions listed in the Instructions for Use appropriately sterilize the tray and contents.3 consecutive half-cycles performed for each of 3 sizes of trays demonstrate complete inactivation of all biologic indicators; A minimum SAL of 10^-6 is achieved if the Instructions for Use are followed.Pass
Dry time (part of Sterilization Validation)To validate that the sterilization instructions listed in the Instructions for Use appropriately dry the wrapped tray for storage.Using pre-cycle and post-cycle weights, the weight gain after drying will be ≤ 3%.Pass
Life Cycle / Simulated Use Life Validation (FDA Guidance: Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, March 2015)To validate the service life of the trays as stated in the Instructions for Use.Visual inspection, component dimensional fit verification, functional closure/latch verification for 250 use cycles.Pass
Biocompatibility of Subject Device (by cytotoxicity testing) (ANSI/AAMI/ISO 10993-5, ANSI/AAMI/ISO 10993-12)To evaluate the cytotoxicity potential of the test article using an in vitro cell culture assay.Non-cytotoxic if ≤ 50% of L-929 cells exposed to extract of device show rounding and lysis.Pass

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for Test Set:
    • For Sterilization Validation, the test set involved "3 sizes of trays". The document lists 5 overall tray dimensions and 14 configurations, suggesting that not all configurations were tested for sterilization, but rather representative sizes.
    • For Life Cycle / Simulated Use Life Validation, the testing involved simulating 250 use cycles.
    • Specific quantities for other tests (Manual Cleaning, Bacterial Endotoxin, Biocompatibility) are not explicitly stated, but the "Pass" results imply adequate testing was performed.
  • Data Provenance: The document states the manufacturer is S.I.N. - Sistema de Implante Nacional S.A. based in São Paulo, Brazil. The testing described is non-clinical, implying laboratory-based studies rather than patient data. The document does not specify if the testing was performed retrospectively or prospectively, but standard validation studies like these are typically conducted prospectively.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

This information is not applicable to this study. The study involves non-clinical performance testing of a medical device (sterilization instrument kits), not an AI algorithm or diagnostic device that requires human expert review to establish ground truth from patient data. The "ground truth" here is based on objective, quantifiable measures against established industry standards for sterility, cleanliness, and material properties.

4. Adjudication Method for the Test Set:

This information is not applicable as the study involves objective, non-clinical laboratory testing, not subjective human-in-the-loop performance evaluation that would require an adjudication method.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

No, an MRMC comparative effectiveness study was not conducted. This type of study is relevant for diagnostic devices (especially AI-assisted models) where human readers (e.g., radiologists) interpret images or data, and their performance is compared with and without AI assistance. This submission describes the non-clinical performance of a sterilization instrument kit, which does not involve human interpretation in the same way. The document explicitly states: "No clinical data were included in this submission."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This question is not applicable as the device is a physical instrument kit for sterilization, not an AI algorithm or software. The testing performed is to validate the physical and functional aspects of the kit itself.

7. The Type of Ground Truth Used:

The ground truth used for this device is based on objective, quantifiable measurements and adherence to established industry standards for medical device reprocessing and biocompatibility. Specifically:

  • Sterility: Sterility Assurance Level (SAL) of 10^-6, complete inactivation of all biologic indicators.
  • Cleanliness: Absence of protein residue (no color change in protein assay), Total Organic Carbon (TOC) assay results below limit of quantification.
  • Endotoxin Levels: Endotoxin results ≤ 20 EU/device.
  • Dryness: Weight gain after drying ≤ 3%.
  • Physical integrity and functionality: Visual inspection, component dimensional fit verification, and functional closure/latch verification after simulated use cycles.
  • Biocompatibility: Non-cytotoxic according to L-929 cell exposure.

These are not "expert consensus", "pathology", or "outcomes data" in the typical sense applied to diagnostic AI models.

8. The Sample Size for the Training Set:

This information is not applicable. This is a physical medical device, not an AI model that requires a training set of data.

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable as there is no AI model or training set involved.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 21, 2020

S.I.N. - Sistema de Implante Nacional S.A. % Kevin Thomas Vice President and Director of Regulatory Affairs PaxMed International. LLC 12264 El Camino Real, Suite 400 San Diego, California 92130

Re: K201688

Trade/Device Name: S.I.N. Instrument Kits Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: September 15, 2020 Received: September 17, 2020

Dear Kevin Thomas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CAPT Elizabeth F. Claverie, MS Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known)

K201688

Device Name

S.I.N. Instrument Kits

Indications for Use (Describe)

Indications for Use for Safe Drill Unitite Kit

S.I.N. Instrument Kits are intended to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Safe Drill Unitite Kit and the associated instruments is 304 grams.

The weight of the empty Safe Drill Unitite Kit is 150 grams.

Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Paperwork Reduction Act (PRA) Staff

PRAStaff @fda.hhs.gov

{3}------------------------------------------------

510(k) Number (if known)

K201688

Device Name

S.I.N. Instrument Kits

Indications for Use (Describe)

Indications for Use for Safe Drill SW Kit

S.I.N. Instrument Kits are intended to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Safe Drill SW Kit and the associated instruments is 278 grams.

The weight of the empty Safe Drill SW Kit is 138 grams.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Offic

Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov

{4}------------------------------------------------

510(k) Number (if known)

K201688

Device Name

S.I.N. Instrument Kits

Indications for Use (Describe)

Indications for Use for Prosthetic Kit

S.I.N. Instrument Kits are intended to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Prosthetic Kit and the associated instruments is 332 grams.

The weight of the empty Prosthetic Kit is 160 grams.

Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) X Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
X Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov

{5}------------------------------------------------

510(k) Number (if known)

K201688

Device Name

S.I.N. Instrument Kits

Indications for Use (Describe)

Indications for Use for Rotatory Expanding Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Rotatory Expanding Kit and the associated instruments is 276 grams.

The weight of the empty Rotatory Expanding Kit is 133 grams.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

PRAStaff @fda.hhs.gov

{6}------------------------------------------------

510(k) Number (if known)

K201688

Device Name

S.I.N. Instrument Kits

Indications for Use (Describe)

Indications for Use for Bone Expander Kit

S.I.N. Instrument Kits are intended to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Bone Expander Kit and the associated instruments is 974 grams.

The weight of the empty Bone Expander Kit is 367 grams.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of

information unless it displays a currently valid OMB number."

{7}------------------------------------------------

510(k) Number (if known)

K201688

Device Name

S.I.N. Instrument Kits

Indications for Use (Describe)

Indications for Use for Sinus Lift Kit

S.I.N. Instrument Kits are intended to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Sinus Lift Kit and the associated instruments is 808 grams.

The weight of the empty Sinus Lift Kit is 370 grams.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov

{8}------------------------------------------------

510(k) Number (if known)

K201688

Device Name

S.I.N. Instrument Kits

Indications for Use (Describe)

Indications for Use for Osteotome Kit

S.I.N. Instrument Kits are intended to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Osteotome Kit and the associated instruments is 957 grams.

The weight of the empty Osteotome Kit is 350 grams.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

{9}------------------------------------------------

510(k) Number (if known)

K201688

Device Name

S.I.N. Instrument Kits

Indications for Use (Describe)

Indications for Use for Unitite Surgical Kit

S.I.N. Instrument Kits are intended to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Unitite Surgical Kit and the associated instruments is 1126 grams.

The weight of the empty Unitite Surgical Kit is 515 grams.

Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Paperwork Reduction Act (PRA) Staff

PRAStaff @fda.hhs.gov

{10}------------------------------------------------

510(k) Number (if known)

K201688

Device Name

S.I.N. Instrument Kits

Indications for Use (Describe)

Indications for Use for Strong SW Surgical Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Strong SW Surgical Kit and the associated instruments is 698 grams.

The weight of the empty Strong SW Surgical Kit is 130 grams.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff @fda.hhs.gov

{11}------------------------------------------------

510(k) Number (if known)

K201688

Device Name

S.I.N. Instrument Kits

Indications for Use (Describe)

Indications for Use for Tryon Surgical Kit KCHE 03

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Tryon Surgical Kit KCHE 03 and the associated instruments is 1127 grams.

The weight of the empty Tryon Surgical Kit KCHE 03 is 520 grams.

Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

{12}------------------------------------------------

510(k) Number (if known)

K201688

Device Name

S.I.N. Instrument Kits

Indications for Use (Describe)

Indications for Use for Tryon Surgical Kit KCHE 04

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Tryon Surgical Kit KCHE 04 and the associated instruments is 1138 grams.

The weight of the empty Tryon Surgical Kit KCHE 04 is 523 grams.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

{13}------------------------------------------------

510(k) Number (if known)

K201688

Device Name

S.I.N. Instrument Kits

Indications for Use (Describe)

Indications for Use for Unitite Surgical Guided Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Unitite Surgical Guided Kit and the associated instruments is 1434 grams.

The weight of the empty Unitite Surgical Guided Kit is 650 grams.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)_ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including thetime to review instructions, search existing data sources, gather and maintain the data needed and completeand review the collection of information. Send comments regarding this burden estimate or any other aspectof this information collection, including suggestions for reducing this burden, to:
Food and Drug AdministrationOffice of Chief Information OfficerPaperwork Reduction Act (PRA) StaffPRAStaff@fda.hhs.govDepartment of Health and Human Services

{14}------------------------------------------------

510(k) Number (if known)

K201688

Device Name

S.I.N. Instrument Kits

Indications for Use (Describe)

Indications for Use for Strong SW Surgical Guided Kit

Tyne of Use /Select one or hoth as annlicable /

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Strong SW Surgical Guided Kit and the associated instruments is 1399 grams.

The weight of the empty Strong SW Surgical Guided Kit is 647 grams.

1 ype of Oos (Ourout One of Wom, us upplivable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including thetime to review instructions, search existing data sources, gather and maintain the data needed and completeand review the collection of information. Send comments regarding this burden estimate or any other aspectof this information collection, including suggestions for reducing this burden, to:
Food and Druq AdministrationOffice of Chief Information OfficerDepartment of Health and Human ServicesPaperwork Reduction Act (PRA) Staff

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

PRAStaff @fda.hhs.gov

{15}------------------------------------------------

510(k) Number (if known)

K201688

Device Name

S.I.N. Instrument Kits

Indications for Use (Describe)

Indications for Use for Zygomatic Surgical Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Zygomatic Surgical Kit and the associated instruments is 1150 grams.

The weight of the empty Zygomatic Surgical Kit is 464 grams.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.


This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

{16}------------------------------------------------

510(k) Summary

K201688

S.I.N. - Sistema de Implante Nacional S.A.

S.I.N Instrument Kits

October 14, 2020

ADMINISTRATIVE INFORMATION

Manufacturer NameS.I.N. – Sistema de Implante Nacional S.A.
Avenida Vereador Abel Ferreira, 1100
São Paulo, São Paulo
03340-000 Brazil
Telephone +55-11-21693000 ext 3236
Official ContactDenise Domiciano, Quality and Regulatory Manager
Representative/ConsultantKevin A. Thomas, PhD
Floyd G. Larson, MS, MBA
PaxMed International, LLC 12264
El Camino Real, Suite 400 San
Diego, CA 92130
Telephone +1 858-792-1235
Fax +1 858-792-1236
Email kthomas@paxmed.com
flarson@paxmed.com

DEVICE NAME AND CLASSIFICATION

Trade/Proprietary NameS.I.N. Instrument Kits
Common NameInstrument sterilization trays
Regulation Number21 CFR 880.6850
Regulation NameSterilization Wrap Containers, Trays, Cassettes & Other Accessories
Regulatory ClassClass II
Product CodeKCT
Classification PanelGeneral Hospital
Reviewing OfficeOffice of Surgical and Infection Control Devices (OHT4) Reviewing
DivisionDivision of Infection Control and Plastic Surgery Devices (DHT4B)

PREDICATE DEVICE INFORMATION The primary predicate device is: K182865, Neodent Instrument Kits, JJGC Indústria e Comércio de Materiais Dentários S.A.

{17}------------------------------------------------

SUBJECT DEVICE INDICATIONS FOR USE STATEMENTS

Indications for Use for Safe Drill Unitite Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 ℃ for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Safe Drill Unitite Kit and the associated instruments is 304 grams. The weight of the empty Safe Drill Unitite Kit is 150 grams.

Indications for Use for Safe Drill SW Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Safe Drill SW Kit and the associated instruments is 278 grams.

The weight of the empty Safe Drill SW Kit is 138 grams.

Indications for Use for Prosthetic Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement – Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Prosthetic Kit and the associated instruments is 332 grams.

The weight of the empty Prosthetic Kit is 160 grams.

Indications for Use for Rotatory Expanding Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement – Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Rotatory Expanding Kit and the associated instruments is 276 grams.

The weight of the empty Rotatory Expanding Kit is 133 grams.

{18}------------------------------------------------

Indications for Use for Bone Expander Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Bone Expander Kit and the associated instruments is 974 grams.

The weight of the empty Bone Expander Kit is 367 grams.

Indications for Use for Sinus Lift Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Sinus Lift Kit and the associated instruments is 808 grams.

The weight of the empty Sinus Lift Kit is 370 grams.

Indications for Use for Osteotome Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 ℃ for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Osteotome Kit and the associated instruments is 957 grams.

The weight of the empty Osteotome Kit is 350 grams.

Indications for Use for Unitite Surgical Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Unitite Surgical Kit and the associated instruments is 1126 grams.

The weight of the empty Unitite Surgical Kit is 515 grams.

Indications for Use for Strong SW Surgical Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

{19}------------------------------------------------

The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Strong SW Surgical Kit and the associated instruments is 698 grams.

The weight of the empty Strong SW Surgical Kit is 130 grams.

Indications for Use for Tryon Surgical Kit KCHE 03

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement – Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Tryon Surgical Kit KCHE 03 and the associated instruments is 1127 grams.

The weight of the empty Tryon Surgical Kit KCHE 03 is 520 grams.

Indications for Use for Tryon Surgical Kit KCHE 04

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Tryon Surgical Kit KCHE 04 and the associated instruments is 1138 grams. The weight of the empty Tryon Surgical Kit KCHE 04 is 523 grams.

Indications for Use for Unitite Surgical Guided Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Unitite Surgical Guided Kit and the associated instruments is 1434 grams.

The weight of the empty Unitite Surgical Guided Kit is 650 grams.

Indications for Use for Strong SW Surgical Guided Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

{20}------------------------------------------------

The combined weight of the Strong SW Surgical Guided Kit and the associated instruments is 1399 grams. The weight of the empty Strong SW Surgical Guided Kit is 647 grams.

Indications for Use for Zygomatic Surgical Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement – Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Zygomatic Surgical Kit and the associated instruments is 1150 grams.

The weight of the empty Zygomatic Surgical Kit is 464 grams.

SUBJECT DEVICE DESCRIPTION

The subject device includes a total of 14 instrument trays. The subject device trays are reusable rigid containers, comprising a base (bottom), one or more removable inner trays, and a lid (cover). The subject device trays are to be used to organize and protect the instruments that are sterilized by the healthcare provider. The base, inner tray, and lid components are designed to be integrated into a single unit which contains and protects the interior components during sterilization. The trays are perforated to allow for penetration of the sterilant, are to be used with moist heat (steam), and require the use of an FDA cleared wrap to maintain sterility. The subject device components are manufactured from injection molded polysulfone (PSU), and holders of various geometries to position instrument in the kits are manufactured from silicone. The subject device includes a total of 5 sizes (same lid, base, and enclosed volume), and a total of 14 tray configurations.

TECHNOLOGICAL CHARACTERISTICS COMPARISON TABLE

The comparison of the Technological characteristics for the subject devices and the predicate device are provided at the end of the 510k summary on pages 7 - 10.

The subject device is provided in 5 sizes and 14 configurations; the predicate device K182865 is provided in 4 sizes and 4 configurations. The subject device and the predicate device have similar overall dimensions, enclose similar volumes, and have similar vent to volume ratios. Differences in the dimensions and vent to volume ratios between the subject device and the predicate device are mitigated by the sterilization validation performed.

SUMMARY OF NONCLINICAL TESTING

Provided below are the non-clinical test methodologies performed to demonstrate the subject devices met the acceptance criteria of the standard.

{21}------------------------------------------------

Test MethodologyPurposeAcceptance CriteriaResults
Manual Cleaning Validation FDAGuidanceReprocessing Medical Devices inHealth Care Settings: ValidationMethods and Labeling(issued March 2015)The purpose of this test is tovalidate that the cleaninginstructions provided in theInstructions for Use appropriatelyclean the tray, and to ensure thesterilization cycle will beeffective.Protein assayAcceptance criterion:No color change = absence of proteinresidueSensitivity = 1 µg protein residuePass
Total organic carbon assayAcceptance criterion:results < limit of quantification = 0.1 ppm;Assay limit of detection = 0.015 ppmPass
Bacterial Endotoxin Testing,USP <85>The purpose of this test is tovalidate that the cleaninginstructions provided in theInstructions for Use appropriatelyclean the tray, and to ensure theBET level meets FDA expectation(≤ 20 EU/device)Acceptance criterion:Endotoxin results ≤ 20 EU/devicePass
Sterilization Validationincluding sterilant penetration anddry time validationANSI/AAMI/ISO 17665-1ANSI/AAMI/ISO 17665-2The purpose of this test is tovalidate that the sterilizationinstructions listed in theInstructions for Use appropriatelysterilize the tray and contents.Acceptance criterion:3 consecutive half-cycles performed foreach of 3 sizes of trays demonstratecomplete inactivation of all biologicindicators;A minimum SAL of 10-6 is achieved if theInstructions for Use are followedPass
Dry timeThe purpose of this test is tovalidate that the sterilizationinstructions listed in theInstructions for Use appropriatelydry the wrapped tray for storage.Acceptance criterion:Using pre-cycle and post-cycle weights,the weight gain after drying will be ≤ 3%Pass
Life Cycle / Simulated Use LifeValidation FDA GuidanceReprocessing Medical Devices inHealth Care Settings: ValidationMethods and Labeling(issued March 2015)The purpose of this test is tovalidate the service life of thetrays as stated in the Instructionsfor Use.Acceptance criteria:Visual inspection, component dimensionalfit verification, functional closure/latchverification for 250 use cyclesPass
Biocompatibility of SubjectDevice (by cytotoxicity testing)ANSI/AAMI/ISO 10993-5ANSI/AAMI/ISO 10993-12The purpose of this test is toevaluate the cytotoxicity potentialof the test article using an in vitrocell culture assay.Acceptance criterion:Non-cytotoxic if ≤ 50% of L-929 cellsexposed to extract of device showrounding and lysisPass

Summary of Nonclinical Testing Table

In summary, the nonclinical testing provided for these devices met the acceptance criteria for each standard and test methodology used to evaluate the devices as shown in the table above.

No clinical data were included in this submission.

CONCLUSION

The conclusions drawn from the nonclinical tests demonstrate that the subject devices are as safe, as effective, and performs as well as or better than the legally marketed predicate device.

{22}------------------------------------------------

Indications for Use Statements
K201688 Subject DeviceS.I.N. Instrument KitsS.I.N. - Sistema de Implante Nacional S.A.
Safe Drill Unitite KitS.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allthe use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.S.I.N. Instrument Kits are intended for sterilization of non-porous loads.S.I.N. Instrument Kits are recommended not to be stacked during sterilization.The combined weight of the Safe Drill Unitite Kit and the associated instruments is 304 grams.The weight of the empty Safe Drill Unitite Kit is 150 grams.
Safe Drill SW KitS.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allthe use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.S.I.N. Instrument Kits are intended for sterilization of non-porous loads.S.I.N. Instrument Kits are recommended not to be stacked during sterilization.The combined weight of the Safe Drill SW Kit and the associated instruments is 278 grams.The weight of the empty Safe Drill SW Kit is 138 grams.
Prosthetic KitS.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allthe use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.S.I.N. Instrument Kits are intended for sterilization of non-porous loads.S.I.N. Instrument Kits are recommended not to be stacked during sterilization.The combined weight of the Prosthetic Kit and the associated instruments is 332 grams.The weight of the empty Prosthetic Kit is 160 grams.
Rotatory Expanding KitS.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allthe use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.S.I.N. Instrument Kits are intended for sterilization of non-porous loads.S.I.N. Instrument Kits are recommended not to be stacked during sterilization.The combined weight of the Rotatory Expanding Kit and the associated instruments is 276 grams.The weight of the empty Rotatory Expanding Kit is 133 grams.
Bone Expander KitS.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allthe use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.S.I.N. Instrument Kits are intended for sterilization of non-porous loads.S.I.N. Instrument Kits are recommended not to be stacked during sterilization.The combined weight of the Bone Expander Kit and the associated instruments is 974 grams.The weight of the empty Bone Expander Kit is 367 grams.
Sinus Lift KitS.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allthe use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.S.I.N. Instrument Kits are intended for sterilization of non-porous loads.S.I.N. Instrument Kits are recommended not to be stacked during sterilization.The combined weight of the Sinus Lift Kit and the associated instruments is 808 grams.The weight of the empty Sinus Lift Kit is 370 grams.
Indications for Use Statements
Osteotome KitS.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosedFDA cleared to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement –Exposure at 121 °C for 30 minutes and 30 minutes dry time.S.I.N. Instrument Kits are intended for sterilization of non-porous loads.S.I.N. Instrument Kits are recommended not to be stacked during sterilization.The combined weight of the Osteotome Kit and the associated instruments is 957 grams. The weight of the emptyOsteotome Kit is 350 grams.
Unitite Surgical KitS.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosedFDA cleared to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement -Exposure at 121 °C for 30 minutes and 30 minutes dry time.S.I.N. Instrument Kits are intended for sterilization of non-porous loads.S.I.N. Instrument Kits are recommended not to be stacked during sterilization.The combined weight of the Unitite Surgical Kit and the associated instruments is 1126 grams. The weight of the emptyUnitite Surgical Kit is 515 grams.
Strong SW Surgical KitS.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosedFDA cleared to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement -Exposure at 121 °C for 30 minutes and 30 minutes dry time.S.I.N. Instrument Kits are intended for sterilization of non-porous loads.S.I.N. Instrument Kits are recommended not to be stacked during sterilization.The combined weight of the Strong SW Surgical Kit and the associated instruments is 698 grams. The weight of the emptyStrong SW Surgical Kit is 130 grams.
Tryon Surgical Kit KCHE03S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosedFDA cleared to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement -Exposure at 121 °C for 30 minutes and 30 minutes dry time.S.I.N. Instrument Kits are intended for sterilization of non-porous loads.S.I.N. Instrument Kits are recommended not to be stacked during sterilization.The combined weight of the Tryon Surgical Kit KCHE 03 and the associated instruments is 1127 grams. The weight of the emptyTryon Surgical Kit KCHE 03 is 520 grams.
Tryon Surgical Kit KCHE04S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosedFDA cleared to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement -Exposure at 121 °C for 30 minutes and 30 minutes dry time.S.I.N. Instrument Kits are intended for sterilization of non-porous loads.S.I.N. Instrument Kits are recommended not to be stacked during sterilization.The combined weight of the Tryon Surgical Kit KCHE 04 and the associated instruments is 1138 grams. The weight of the emptyTryon Surgical Kit KCHE 04 is 523 grams.
Unitite Surgical Guided KitS.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosedFDA cleared to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement -Exposure at 121 °C for 30 minutes and 30 minutes dry time.S.I.N. Instrument Kits are intended for sterilization of non-porous loads.S.I.N. Instrument Kits are recommended not to be stacked during sterilization.The combined weight of the Unitite Surgical Guided Kit and the associated instruments is 1434 grams. The weight of the emptyUnitite Surgical Guided Kit is 650 grams.
Strong SW Surgical GuidedKitS.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosedFDA cleared to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement -Exposure at 121 °C for 30 minutes and 30 minutes dry time.S.I.N. Instrument Kits are intended for sterilization of non-porous loads.S.I.N. Instrument Kits are recommended not to be stacked during sterilization.The combined weight of the Strong SW Surgical Guided Kit and the associated instruments is 1399 grams. The weight of the empty
Indications for Use Statements
Zygomatic Surgical KitS.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allowthe use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.S.I.N. Instrument Kits are intended for sterilization of non-porous loads.S.I.N. Instrument Kits are recommended not to be stacked during sterilization.The combined weight of the Zygomatic Surgical Kit and the associated instruments is 1150 grams.The weight of the empty Zygomatic Surgical Kit is 464 grams.
Predicate Device
K182865Neodent Instrument KitsJJGC Indústria e Comércio de Materiais Dentários S.A.Indications for Use for GM/WS Surgical Kit Case:Neodent Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Neodent Instrument Kits are intended to allrequire the use of FDA cleared wrap to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles:Fractionated vacuum (pre-vacuum) – Exposure at 132 °C for 4 minutes, 20 minute dry timeGravity displacement - Exposure at 132 °C for 15 minutes, 20 minute dry timeNeodent Instrument Kits are intended for sterilization of non-porous loads.The combined weight of the GM/WS Surgical Kit Case and the associated instruments is 674.5 g. The weight of the empty Kit Case is 507 grams.Neodent Instrument Kits are recommended not to be stacked during sterilization.Indications for Use for GM Prosthetic Kit Case:Neodent Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Neodent Instrument Kits are intended to allrequire the use of FDA-cleared wrap to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles:Fractionated vacuum (pre-vacuum) – Exposure at 132 °C for 4 minutes, 20-minute dry timeGravity displacement - Exposure at 132 °C for 15 minutes, 20-minute dry time.Neodent Instrument Kits are intended for sterilization of non-porous loads.The combined weight of the GM Prosthetic Kit Case and the associated instruments is 250.5 g. The weight of the empty Kit Case is 210 grams.Neodent Instrument Kits are recommended not to be stacked during sterilization.Indications for Use for GM Try-In Kit Case:Neodent Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Neodent Instrument Kits require the use ofThe kits are to be enclosed in a sterilizable wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles:Fractionated vacuum (pre-vacuum) – Exposure at 132 °C for 4 minutes, 20-minute dry time.Gravity displacement - Exposure at 132 °C for 15 minutes, 20-minute dry time.Neodent Instrument Kits are intended for sterilization of non-porous loads.The combined weight of the GM Try-In Kit Case and the associated instruments is 212.6 g. The weight of the empty Kit Case is 195 grams.Neodent Instrument Kits are recommended not to be stacked during sterilization.Indications for Use for GM Guided Surgery Kit Case:Neodent Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. Neodent Instrument Kits are intended to allrequire the use of FDA-cleared wrap to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA-cleared for the indicated cycles, and moist heat (steam) sterilized using one of the following cycles:Fractionated vacuum (pre-vacuum) - Exposure at 132 °C for 4 minutes, 20-minute dry timeGravity displacement - Exposure at 132 °C for 15 minutes, 40-minute dry time.Neodent Instrument Kits are intended for sterilization of non-porous loads.The combined weight of the GM Guided Surgery Surgical Kit Case and the associated instruments is 728.4 g. The weight of the empty Kit Case is 567 grams.Neodent Instrument Kits are recommended not to be stacked during sterilization.

w sterilization of the enclosed medical devices. S.I.N. Instrument Kits require

w sterilization of the enclosed medical devices. S.I.N. Instrument Kits require

w sterilization of the enclosed medical devices. S.I.N. Instrument Kits require

w sterilization of the enclosed medical devices. S.I.N. Instrument Kits require

w sterilization of the enclosed medical devices. S.I.N. Instrument Kits require

w sterilization of the enclosed medical devices. S.I.N. Instrument Kits require

{23}------------------------------------------------

nedical devices. S.I.N. Instrument Kits require the use of a wrap that is

nedical devices. S.I.N. Instrument Kits require the use of a wrap that is

nedical devices. S.I.N. Instrument Kits require the use of a wrap that is

nedical devices. S.I.N. Instrument Kits require the use of a wrap that is

nedical devices. S.I.N. Instrument Kits require the use of a wrap that is

nedical devices. S.I.N. Instrument Kits require the use of a wrap that is

nedical devices. S.I.N. Instrument Kits require the use of a wrap that is

{24}------------------------------------------------

sterilization of the enclosed medical devices. S.I.N. Instrument Kits require

w sterilization of the enclosed medical devices. Neodent Instrument Kits

w sterilization of the enclosed medical devices. Neodent Instrument Kits

FDA-cleared wrap to maintain the sterility of the enclosed devices.

w sterilization of the enclosed medical devices. Neodent Instrument Kits

{25}------------------------------------------------

Technological Characteristics

Subject DevicePrimary Predicate DeviceComparison
AttributeK201688K182865
S.I.N. Instrument KitsNeodent Instrument Kits
S.I.N. - Sistema de Implante Nacional S.A.JJGC Indústria e Comércio de Materiais Dentários S.A.
Product CodeKCTKCTSame
DesignRigid polymer base, lid, and removable inner trayRigid polymer base, lid, and removable inner traySame
MaterialsPolysulfone (base, tray, lid)Polysulfone (base, tray)Similar
Medical grade silicone (grommets)Polyphenylsulfone (Radel R5000) (lid)
Medical grade silicone (grommets)Titanium alloy (retention fixtures)
Materials Compatible withSterilization Method described below.YesYesSame
PerforatedYes; allows moist heat (steam) penetration to achieve sterilizationYes; allows moist heat (steam) penetration to achieve sterilizationSame
ReusableYesYesSame
Number of Overall Tray Dimensions54Similar
Total Number of Configurations144Similar
Dimensions andVent to Volume RatioTrayLength x Width x Height, mmVent to Volume RatioSetLength x Width x Height, mmVent to Volume RatioSimilar
COUSD 02, Safe Drill Unitite Tray113.7 x 75.7 x 29.50.0089 cm²/cm³Set 110.287264 x 163 x 540.0102 cm²/cm³
COWSD 02, Safe Drill SW Tray113.7 x 75.7 x 29.50.0089 cm²/cm³Set 110.294195 x 90 x 540.0191 cm²/cm³
COTMEC, Prosthetic Tray113.7 x 75.7 x 29.50.0089 cm²/cm³Set 110.295195 x 90 x 440.0247 cm²/cm³
COER, Rotatory Expanding Tray113.7 x 75.7 x 29.50.0089 cm²/cm³Set 110.296264 x 163 x 580.0093 cm²/cm³
COEXP, Bone Expander Tray113.7 x 75.7 x 29.50.0100 cm²/cm³
COLEV, Sinus Lift Tray215 x 100 x 33.50.0100 cm²/cm³
COOST, Osteotome Tray113.7 x 75.7 x 29.50.0100 cm²/cm³
COSU 03, Unitite Surgical Tray165 x 190 x 550.0086 cm²/cm³
COSW 02, Strong SW Surgical Tray165 x 190 x 550.0086 cm²/cm³
COHE 03, Tryon Surgical Tray KCHE 03165 x 190 x 550.0086 cm²/cm³
COHE 04, Tryon Surgical Tray KCHE 04165 x 190 x 550.0086 cm²/cm³
COSUG 02, Unitite Surgical Guided Tray142 x 206 x 720.0083 cm²/cm³
COSWG 02, Strong SW Surgical Guided Tray165 x 190 x 550.0083 cm²/cm³
COKZ, Zygomatic Surgical Tray113.7 x 75.7 x 29.50.0131 cm²/cm³
ReusableYesYesSame
Use Life TestingReusable up to 250 cyclesDisassembled, cleaned, assembled,sterilized Visual inspectionComponent dimensional fit verificationFunctional closure (lid-base latch)verificationReusable up to 100 cyclesAssembled/disassembled, cleaned,sterilized Visual inspectionComponent dimensional fit verificationFunctional closure (lid-base latch)verificationSimilar
Sterilization Method
SterilantMoist heat (steam)Moist heat (steam)Same
CyclesGravity displacementGravity displacementSame (SAL)
30 minute exposure at 121 oC (250 oF), with a 30 minute drying time; results in SAL of 10-615 minute exposure at 132 oC (250 oF), with a 20 minute drying time
(No other cycle validated)Fractionated vacuum (pre-vacuum)Notapplicable
4 minute exposure at 132 oC (250 oF), with a 20 minute drying time
Sterile BarrierSterilization wrap, FDA cleared for indicated method and cyclesSterilization wrap, FDA cleared for indicated method and cyclesSame

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).