S.I.N. Instrument Kits
K241378 · S.I.N. - Sistema DE Implante Nacional S.A. · KCT · Jul 12, 2024 · General Hospital
Device Facts
| Record ID | K241378 |
| Device Name | S.I.N. Instrument Kits |
| Applicant | S.I.N. - Sistema DE Implante Nacional S.A. |
| Product Code | KCT · General Hospital |
| Decision Date | Jul 12, 2024 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.6850 |
| Device Class | Class 2 |
Intended Use
S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization. The combined weight of the Epikut PTG Surgical Kit and associated instruments is 632 grams. The weight of the empty Epikut PTG Surgical Kit is 353 grams.
Device Story
Reusable rigid container system; comprises polysulfone base, removable inner tray, and lid; includes silicone instrument holders. Used by healthcare providers to organize, protect, and contain non-porous medical instruments during steam sterilization. Perforated design allows sterilant penetration; requires external FDA-cleared sterilization wrap to maintain sterility post-process. Device protects instruments during handling and sterilization cycles; supports 250 reuse cycles. Benefits include organized instrument management and effective sterilization maintenance.
Clinical Evidence
Bench testing only. Validated via manual cleaning, bacterial endotoxin testing (USP <85>), sterilization cycle efficacy (ANSI/AAMI/ISO 17665-1/2), dry time, 250-cycle simulated use-life, and biocompatibility (ISO 10993-5). All tests passed.
Technological Characteristics
Rigid container; materials: polysulfone (base, tray, lid), medical-grade silicone (grommets). Perforated for steam penetration. Reusable (up to 250 cycles). Sterilization: gravity displacement steam (121°C, 30 min exposure, 30 min dry). Requires FDA-cleared sterilization wrap. Dimensions: 215 x 100 x 35.5 mm.
Indications for Use
Indicated for use by healthcare providers to enclose and sterilize non-porous medical devices via gravity displacement steam sterilization (121°C for 30 minutes). Requires use of an FDA-cleared sterilization wrap.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
- S.I.N. Instrument Kits (K201688)
Reference Devices
- S.I.N. Instrument Kits (K212404)
Related Devices
- K243317 — NUVENTUS NV.C Surgical Cassette and Tray · Infinitum Eta, Ltd. · Feb 11, 2025
- K203618 — Neodent EasyGuide Kit Cases · Jjgc Industria E Comercio DE Materiais Dentarios S.A. · Mar 9, 2021
- K222005 — S.I.N. Instrument Kits · S.I.N. - Sistema DE Implante Nacional S.A. · Sep 29, 2022
- K214051 — Neodent Implant System-Helix NGM Compact Surgical Kit Cases · JJGC Indústria e Comércio de Materiais Dentários S.A. · Apr 13, 2022
- K223662 — Neodent Implant System - Helix Short Surgical Kit Cases · Jjgc Industria E Comercio DE Materiais Dentarios S.A. · Apr 20, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 12, 2024
S.I.N. - Sistema de Implante Nacional S.A. Denise Domiciano Quality and Regulatory Manager Rua Soldado Ocimar Guimarães da Silva, 421 São Paulo, São Paulo 03348-060 Brazil
Re: K241378
Trade/Device Name: S.I.N. Instrument Kits Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: May 15, 2024 Received: May 15, 2024
Dear Denise Domiciano:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Digitally signed by Stephen Stephen A. A. Anisko -S Date: 2024.07.12 19:48:56 Anisko -S -04'00' for: Christopher Dugard Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices
OHT4: Office of Surgical
{2}------------------------------------------------
and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241378
Device Name S.I.N. Instrument Kits
#### Indications for Use (Describe)
Indications for Use for Epikut PTG Surgical Kit
S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.
The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:
Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.
S.I.N. Instrument Kits are intended for sterilization of non-porous loads.
S.I.N. Instrument Kits are recommended not to be stacked during sterilization.
The combined weight of the Epikut PTG Surgical Kit and associated instruments is 632 grams.
The weight of the empty Epikut PTG Surgical Kit is 353 grams.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary K241378
# S.I.N. - Sistema de Implante Nacional S.A. S.I.N. Instrument Kits
JUNE 25, 2024
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | S.I.N. - Sistema de Implante Nacional S.A. |
|-------------------|--------------------------------------------------|
| | Rua Soldado Ocimar Guimarães da Silva, |
| | 421 São Paulo, São Paulo |
| | 03348-060 BRAZIL |
| Telephone | +55 -11-21693000 ext 3236 |
| Official Contact | Denise Domiciano, Quality and Regulatory Manager |
| | Email denise.domiciano@sinimplantsystem.com |
#### Representative
Floyd G. Larson, MS, MBA PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, CA 92130 + 858 792- 1235 Telephone Fax + 858 792-1236 Email flarson@paxmed.com
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | S.I.N. Instrument Kits |
|------------------------------|---------------------------------------------------------------------|
| Common Name | Instrument sterilization trays |
| Regulation Number | 21 CFR 880.6850 |
| Regulation Name | Sterilization Wrap Containers, Trays, Cassettes & Other Accessories |
| Regulatory Class | Class II |
| Product Code | KCT |
| Classification Panel | General Hospital |
| Reviewing Office | Office of Surgical and Infection Control Devices (OHT4) |
| Reviewing Division | Division of Infection Control and Plastic Surgery Devices (DHT4B) |
| PREDICATE DEVICE INFORMATION | |
Primary predicate device is: K201688, S.I.N. Instrument Kits, S.I.N. - Sistema de Implante Nacional S.A. Reference device: K212404, S.I.N. Instrument Kits, S.I.N. – Sistema de Implante Nacional S.A.
{5}------------------------------------------------
#### SUBJECT DEVICE INDICATION FOR USE STATEMENT
Indications for Use for Epikut PTG Surgical Kit
S.I.N. Instruments Kits are intended to be used to enclose other medical devices that are to be sterlized by a health care provider.
S.I.N. Instruments Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instruments Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.
The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:
Gravity displacement - Exposure at 121°C°C for 30 minutes and 30 minutes dry time.
S.I.N. Instruments Kits are intended for sterilization of non-porous loads.
S.I.N. Instruments Kits are recommended not to be stacked during sterilization.
The combined weight of the Epikut PTG Surgical Kit and the associated instruments is 632 grams. The weight of the empty Epikut PTG Surgical Kit is 353 grams.
#### SUBJECT DEVICE DESCRIPTION
The subject device is a reusable rigid container, comprising a base (bottom), a removable inner tray, and a lid (cover). The subject device tray is to be used to organize and protect the instruments that are sterilized by the healthcare provider. The base, inner tray, and lid components are designed to be integrated into a single unit which contains and protects the interior components during sterilization. The tray is perforated to allow for penetration of the sterilant, is to be used with moist heat (steam) and requires the use of an FDA cleared wrap to maintain sterility. The subject device components are manufactured from injection molded polysulfone (PSU), and holders of various geometries to position instruments in the kits are manufactured from silicone.
#### TECHNOLOGICAL CHARACTERISTICS COMPARISON TABLE
The comparison of the Indications for Use Statements and technological characteristics for the subject device, the predicate device, and the reference device are provided at the end of this summary on pages 5-8.
The subject device is provided in one (1) size and one (1) configuration; the primary predicate device K201688 is provided in 5 sizes and 14 configurations. The subject device and the predicate device have similar overall dimensions, enclose similar volumes, and have similar vent to volume ratios. Differences in the dimensions and vent to volume ratios between the subject device and the predicate device are mitigated by the sterilization validation performed.
{6}------------------------------------------------
### SUMMARY OF NONCLINICAL TESTING
Provided below are the nonclinical test methodologies performed to demonstrate the subject device meets the acceptance criteria of the standard.
| Test Methodology | Purpose | Acceptance Criteria | Results |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Manual Cleaning Validation FDA<br>Guidance<br>Reprocessing Medical Devices in<br>Health Care Settings: Validation<br>Methods and Labeling<br>(issued March 2015<br>Referenced from K201688 | The purpose of this test is to<br>validate that the cleaning<br>instructions provided in the<br>Instructions for Use appropriately<br>clean the tray, and to ensure the | Protein assay Acceptance criterion:<br>No color change = absence of protein<br>residue<br>Sensitivity = 1 µg protein residue | Pass |
| | sterilization cycle will be effective. | Total organic carbon assay Acceptance<br>criterion:<br>results < limit of quantification = 0.1 ppm;<br>Assay limit of detection = 0.015 ppm | Pass |
| Bacterial Endotoxin Testing, USP<br><85><br>Referenced from K201688 | The purpose of this test is to<br>validate that the cleaning<br>instructions provided in the<br>Instructions for Use appropriately<br>clean the tray, and to ensure the<br>BET level meets FDA expectation<br>(≤ 20 EU/device) | Acceptance criterion:<br>Endotoxin results : ≤ 20 EU/device | Pass |
| Sterilization Validation<br>including sterilant penetration and<br>dry time validation<br>ANSI/AAMI/ISO 17665-1<br>ANSI/AAMI/ISO 17665-2<br>Referenced from K201688 | The purpose of this test is to<br>validate that the sterilization<br>instructions listed in the<br>Instructions for Use appropriately<br>sterilize the tray and contents. | Acceptance criterion:<br>3 consecutive half-cycles performed for<br>each of 3 sizes of trays demonstrate<br>complete inactivation of all biologic<br>indicators.<br>A minimum SAL of 10-6 is achieved if the<br>Instructions for Use are followed | Pass |
| Dry time<br>Referenced from K201688 | The purpose of this test is to<br>validate that the sterilization<br>instructions listed in the<br>Instructions for Use appropriately<br>dry the wrapped tray for storage. | Acceptance criterion:<br>Using pre-cycle and post-cycle weights, the<br>weight gain after drying will be ≤ 3% | Pass |
| Life Cycle / Simulated Use-life<br>Validation FDA Guidance<br>Reprocessing Medical Devices in<br>Health Care Settings: Validation<br>Methods and Labeling<br>(issued March 2015)<br>Referenced from K201688 | The purpose of this test is to<br>validate the service life of the trays<br>as stated in the Instructions for Use. | Acceptance criteria:<br>Visual inspection, component dimensional<br>fit verification, functional closure/latch<br>verification for 250 use cycles | Pass |
| Biocompatibility of Subject<br>Device<br>Cytotoxicity testing<br>ANSI/AAM.T/.ISO 10993-5<br>ANSI/AAM.T/ ISO 10993- 12<br>Referenced from K201688<br>and K212404 | The purpose of this test is to<br>evaluate the cytotoxicity potential<br>of the test article using an in vitro<br>cell culture assay. | Acceptance criterion:<br>Non-cytotoxic if :≤ 50% of L-929 cells<br>exposed to extract of device show rounding<br>and lysis | Pass |
### Summary of Nonclinical Testing Table
In summary, the nonclinical testing provided for the subject device met the acceptance criteria for each standard and test methodology used to evaluate the subject device as shown in the table above.
No clinical data was included in this submission.
{7}------------------------------------------------
## CONCLUSION
The conclusions drawn from the nonclinical tests demonstrate that the subject device in this 510(k) submission, K241378, S.I.N. Instrument Kit, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K201688.
{8}------------------------------------------------
## Indications for Use Statements
| | Indications for Use Statements |
|-------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device – K241378<br>S.I.N. Instrument Kits<br>S.I.N.- Sistema de Implante Nacional S.A.<br>Epikut PTG Surgical Kit | S.I.N. Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instruments Kits are intended to allow ster<br>use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:<br>Gravity displacement - Exposure at 121°C for 30 minutes and 30 minutes dry time.<br>S.I.N. Instruments Kits are intended for sterilization of non-porous loads. S.I.N.<br>Instruments Kits are recommended not to be stacked during sterilization.<br>The combined weight of the Epikut PTG Surgical Kit and the associated instruments is 632grams.<br>The weight of the empty Epikut PTG Surgical Kit is 353 grams. |
| Primary Predicate Device – K201688<br>S.I.N. Instrument Kits<br>S.I.N. – Sistema de Implant Nacional S.A.<br>Safe Drill Unitite Kit | S.I.N. Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instruments Kits are intended to allow steri<br>use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:<br>Gravity displacement - Exposure at 121°C for 30 minutes and 30 minutes dry time.<br>S.I.N. Instruments Kits are intended for sterilization of non-porous loads. S.I.N.<br>Instruments Kits are recommended not to be stacked during sterilization.<br>The combined weight of the Safe Drill Unitite Kit and the associated instruments is 304 grams.<br>The weight of the empty Safe Drill Unitite Kit is 150 grams. |
| Safe Drill SW Kit | S.I.N. Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instruments Kits are intended to allow steri<br>use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:<br>Gravity displacement - Exposure at 121°C for 30 minutes and 30 minutes dry time.<br>S.I.N. Instruments Kits are intended for sterilization of non-porous loads. S.I.Ν.<br>Instruments Kits are recommended not to be stacked during sterilization.<br>The combined weight of the Safe Drill SW Kit and the associated instruments is 278 grams.<br>The weight of the empty Safe Drill SW Kit is 138 grams. |
| Prosthetic Kit | S.I.N. Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instruments Kits are intended to allow steri<br>use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:<br>Gravity displacement - Exposure at 121°C for 30 minutes and 30 minutes dry time.<br>S.I.N. Instruments Kits are intended for sterilization of non-porous loads. S.I.N.<br>Instruments Kits are recommended not to be stacked during sterilization.<br>The combined weight of the Prosthetic Kit and the associated instruments is 332 grams.<br>The weight of the empty Prosthetic Kit is 160 grams. |
| Rotatory Expanding Kit | S.I.N. Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instruments Kits are intended to allow steri<br>use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:<br>Gravity displacement - Exposure at 121°C for 30 minutes and 30 minutes dry time.<br>S.I.N. Instruments Kits are intended for sterilization of non-porous loads.<br>S.I.N. Instruments Kits are recommended not to be stacked during sterilization.<br>The combined weight of the Rotatory Expanding Kit and the associated instruments is 276 grams.<br>The weight of the empty Rotatory Expanding Kit is 133 grams. |
| Bone Expander Kit | S.I.N. Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instruments Kits are intended to allow steri<br>use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:<br>Gravity displacement - Exposure at 121°C for 30 minutes and 30 minutes dry time.<br>S.I.N. Instruments Kits are intended for sterilization of non-porous loads. S.I.N.<br>Instruments Kits are recommended not to be stacked during sterilization.<br>The combined weight of the Bone Expander Kit and the associated instruments is 974 grams.<br>The weight of the empty Bone Expander Kit is 367 grams. |
| | |
| Sinus Lift Kit | S.I.N. Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instruments Kits are intended to allow sterilize<br>use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:<br>Gravity displacement — Exposure at 121°C °C for 30 minutes and 30 minutes dry time.<br>S.I.N. Instruments Kits are intended for sterilization of non-porous loads. S.I.N.<br>Instruments Kits are recommended not to be stacked during sterilization.<br>The combined weight of the Sinus Lift Kit and the associated instruments is 808 grams.<br>The weight of the empty Sinus Lift Kit is 370 grams. |
| Osteotome Kit | S.I.N. Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instruments Kits are intended to allow sterilize<br>use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:<br>Gravity displacement - Exposure at 121°C °C for 30 minutes and 30 minutes dry time.<br>S.I.N. Instruments Kits are intended for sterilization of non-porous loads. S.I.N.<br>Instruments Kits are recommended not to be stacked during sterilization.<br>The combined weight of the Osteotome Kit and the associated instruments is 957 grams.<br>The weight of the empty Osteotome Kit is 350 grams. |
| Unitite Surgical Kit | S.I.N. Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instruments Kits are intended to allow sterilize<br>use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:<br>Gravity displacement - Exposure at 121°C °C for 30 minutes and 30 minutes dry time.<br>S.I.N. Instruments Kits are intended for sterilization of non-porous loads. S.I.N.<br>Instruments Kits are recommended not to be stacked during sterilization.<br>The combined weight of the Unitite Surgical Kit and the associated instruments is 1126<br>grams. The weight of the empty Unitite Surgical Kit is 515grams. |
| Strong SW Surgical Kit | S.I.N. Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instruments Kits are intended to allow sterilize<br>use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:<br>Gravity displacement - Exposure at 121°C °C for 30 minutes and 30 minutes dry time.<br>S.I.N. Instruments Kits are intended for sterilization of non-porous loads. S.I.N.<br>Instruments Kits are recommended not to be stacked during sterilization.<br>The combined weight of the Strong SW Surgical Kit and the associated instruments is 698 grams.<br>The weight of the empty Strong SW Surgical Kit is 130 grams. |
| Tryon Surgical Kit KCHE 03 | S.I.N. Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instruments Kits are intended to allow sterilize<br>use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:<br>Gravity displacement - Exposure at 121°C °C for 30 minutes and 30 minutes dry time.<br>S.I.N. Instruments Kits are intended for sterilization of non-porous loads. S.I.N.<br>Instruments Kits are recommended not to be stacked during sterilization.<br>The combined weight of the Tryon Surgical Kit KCHE 03 and the associated instruments is 1127 grams.<br>The weight of the empty Tryon Surgical Kit KCHE 03 is 520 grams. |
| Tryon Surgical Kit KCHE 04 | S.I.N. Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instruments Kits are intended to allow sterilize<br>use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:<br>Gravity displacement - Exposure at 121°C °C for 30 minutes and 30 minutes dry time.<br>S.I.N. Instruments Kits are intended for sterilization of non-porous loads. S.I.N.<br>Instruments Kits are recommended not to be stacked during sterilization.<br>The combined weight of the Tryon Surgical Kit KCHE 04 and the associated instruments is 1138 grams.<br>The weight of the empty Tryon Surgical Kit KCHE 04 is 523 grams. |
| Unitite Surgical Guided Kit | S.I.N. Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instruments Kits are intended to allow sterilize<br>use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:<br>Gravity displacement - Exposure at 121°C °C for 30 minutes and 30 minutes dry time.<br>S.I.N. Instruments Kits are intended for sterilization of non-porous loads. S.I.N.<br>Instruments Kits are recommended not to be stacked during sterilization.<br>The combined weight of the Unitite Surgical Guided Kit and the associated instruments is 1434 grams.<br>The weight of the empty Unitite Surgical Guided Kit is 650 grams. |
| | |
| Strong SW Surgical Guided Kit | S.I.N. Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instruments Kits are intended to allow ster<br>use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:<br>Gravity displacement - Exposure at 121°C °C for 30 minutes and 30 minutes dry time.<br>S.I.N. Instruments Kits are intended for sterilization of non-porous loads. S.I.N.<br>Instruments Kits are recommended not to be stacked during sterilization.<br>The combined weight of the Strong SW Surgical Guided Kit and the associated instruments is 1399 grams.<br>The weight of the empty Strong SW Surgical Guided Kit is 647 grams. |
| Zygomatic Surgical Kit | S.I.N. Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instruments Kits are intended to allow ster<br>of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:<br>Gravity displacement - Exposure at 121°C °C for 30 minutes and 30 minutes dry time.<br>S.I.N. Instruments Kits are intended for sterilization of non-porous loads. S.I.N.<br>Instruments Kits are recommended not to be stacked during sterilization.<br>The combined weight of the Zygomatic Surgical Kit and the associated instruments is 1150 grams.<br>The weight of the empty Zygomatic Surgical Kit is 464 grams. |
| Reference Device – K212404<br>S.I.N. Instrument Kits<br>S.I.N. – Sistema de Implante Nacional S.A. | |
| Safe Drill Epikut Kit | S.I.N. Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instruments Kits are intended to allow ster<br>of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:<br>Gravity displacement - Exposure at 121°C °C for 30 minutes and 30 minutes dry time.<br>S.I.N. Instruments Kits are intended for sterilization of non-porous loads. S.I.N.<br>Instruments Kits are recommended not to be stacked during sterilization.<br>The combined weight of the Safe Drill Epikut Kit and the associated instruments is 154 grams.<br>The weight of the empty Safe Drill Epikut Kit is 138 grams. |
| Unitite Surgical Kit | S.I.N. Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instruments Kits are intended to allow ster<br>of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:<br>Gravity displacement - Exposure at 121°C °C for 30 minutes and 30 minutes dry time.<br>S.I.N. Instruments Kits are intended for sterilization of non-porous loads. S.I.N.<br>Instruments Kits are recommended not to be stacked during sterilization.<br>The combined weight of the Unitite Surgical Kit and the associated instruments is 620 grams.<br>The weight of the empty Unitite Surgical Kit is 520 grams. |
| Epikut Surgical Kit | S.I.N. Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instruments Kits are intended to allow ster<br>of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:<br>Gravity displacement - Exposure at 121°C °C for 30 minutes and 30 minutes dry time.<br>S.I.N. Instruments Kits are intended for sterilization of non-porous loads. S.I.N.<br>Instruments Kits are recommended not to be stacked during sterilization.<br>The combined weight of the Epikut Surgical Kit and the associated instruments is 605 grams.<br>The weight of the empty Epikut Surgical Kit is 520 grams. |
| Epikut Surgical Guided Kit | S.I.N. Instruments Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instruments Kits are intended to allow ster<br>of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.<br>The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:<br>Gravity displacement - Exposure at 121°C °C for 30 minutes and 30 minutes dry time.<br>S.I.N. Instruments Kits are intended for sterilization of non-porous loads. S.I.N.<br>Instruments Kits are recommended not to be stacked during sterilization.<br>The combined weight of the Epikut Surgical Guided Kit and the associated instruments is 808 grams.<br>The weight of the empty Epikut Surgical Guided Kit is 650 grams. |
zation of the enclosed medical devices. S.I.N. Instruments Kits require the
ization of the enclosed medical devices. S.I.N. Instruments Kits require the
zation of the enclosed medical devices. S.I.N. Instruments Kits require the
zation of the enclosed medical devices. S.I.N. Instruments Kits require the
zation of the enclosed medical devices. S.I.N. Instruments Kits require the
zation of the enclosed medical devices. S.I.N. Instruments Kits require the
{9}------------------------------------------------
zation of the enclosed medical devices. S.I.N. Instruments Kits require the
zation of the enclosed medical devices. S.I.N. Instruments Kits require the
zation of the enclosed medical devices. S.I.N. Instruments Kits require the
ization of the enclosed medical devices. S.I.N. Instruments Kits require the
zation of the enclosed medical devices. S.I.N. Instruments Kits require the
zation of the enclosed medical devices. S.I.N. Instruments Kits require the
zation of the enclosed medical devices. S.I.N. Instruments Kits require the
{10}------------------------------------------------
ization of the enclosed medical devices. S.I.N. Instruments Kits require the
ization of the enclosed medical devices. S.I.N. Instruments Kitsrequire the use
ization of the enclosed medical devices. S.I.N. Instruments Kitsrequire the use
ization of the enclosed medical devices. S.I.N. Instruments Kitsrequire the use
ization of the enclosed medical devices. S.I.N. Instruments Kitsrequire the use
ization of the enclosed medical devices. S.I.N. Instruments Kitsrequire the use
{11}------------------------------------------------
Technological Characteristics
| | Subject Device | | Primary Predicate Device | Reference Device | Comparison | | |
|---------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|-----------------------------|----------------------|---------|
| Attribute | K241378<br>S.I.N. Instrument Kits<br>S.I.N. - Sistema de Implante Nacional S.A. | | K206188<br>S.I.N. Instrument Kits<br>S.I.N. - Sistema de Implante Nacional S.A. | K212404<br>S.I.N. Instrument Kits<br>S.I.N. - Sistema de Implante Nacional S.A. | | | |
| Product Code | KCT | | KCT | KCT | Same | | |
| Design | Rigid polymer base, lid, and removable inner tray | | Rigid polymer base, lid, and removable inner tray | Rigid polymer base, lid, and removable inner tray | Same | | |
| Materials | Polysulfone (base, tray, lid); Medical grade silicone (grommets) | | Polysulfone (base, tray, lid); Medical grade silicone (grommets) | Polysulfone (base, tray, lid); Medical grade silicone (grommets) | Same | | |
| Materials Compatible with Sterilization<br>Method | Yes | | Yes | Yes | Same | | |
| Perforated | Yes; allows moist heat (steam) penetration to achieve sterilization | | Yes, allows moist heat (steam) penetration to achieve sterilization | Yes, allows moist heat (steam) penetration to achieve sterilization | Same | | |
| Reusable | Yes | | Yes | Yes | Same | | |
| Number of Sizes/Configurations | 1 size, 1 configuration | | K201688 - 5 sizes, 14 configurations | K201688 - 5 sizes, 14 configurations | Similar | | |
| Dimensions<br>and<br>Vent to Volume Ratio | Subject Device Tray | Length x Width x Height, mm | Vent to Volume Ratio | Trays in K201688 | Length x Width x Height, mm | Vent to Volume Ratio | Similar |
| | COSEP 01, Epikut PTG Surgical Tray | 215 x 100 x 35.5 | 0.0087 cm²/cm³ | COUSD 02, Safe Drill Unitite Tray | 113.7 x 75.7 x 29.5 | 0.0089cm²/cm³ | |
| | | | | COWSD 02, Safe Drill SW Tray | 113.7 x 75.7 x 29.5 | 0.0089cm²/cm³ | |
| | | | | COTMEC, Prosthetic Tray | 113.7 x 75.7 x 29.5 | 0.0089cm²/cm³ | |
| | | | | COER, Rotatory Expanding Tray | 113.7 x 75.7 x 29.5 | 0.0089cm²/cm³ | |
| | | | | COEXP, Bone Expander Tray | 113.7 x 75.7 x 29.5 | 0.0100cm²/cm³ | |
| | | | | COLEV, Sinus Lift Tray | 215 x 100 x 33.5 | 0.0100 cm²/cm³ | |
| | | | | COOST, Osteotome Tray | 113.7 x 75.7 x 29.5 | 0.0100cm²/cm³ | |
| | | | | COSU 03, Unitite Surgical Tray | 165 x 190 x 55 | 0.0086cm²/cm³ | |
| | | | | COSW 02, Strong SW Surgical Tray | 165 x 190 x 55 | 0.0086cm²/cm³ | |
| | | | | COHE 03, Tryon Surgical Tray KCHE 03 | 165 x 190 x 55 | 0.0086cm²/cm³ | |
| | | | | COHE 04, Tryon Surgical Tray KCHE 04 | 165 x 190 x 55 | 0.0086cm²/cm³ | |
| | | | | COSUG 02, Unitite Surgical Guided Tray | 142 x 206 x 72 | 0.0083 cm²/cm³ | |
| | | | | COSWG 02, Strong SW Surgical Guided Tray | 165 x 190 x 55 | 0.0083 cm²/cm³ | |
| | | | | COKZ, Zygomatic Surgical Tray | 113.7 x 75.7 x 29.5 | 0.0131 cm²/cm³ | |
| | | | | Trays in K212404 | | | |
| | | | | COESD 02, Safe Drill Epikut Tray | 118 x 78.25 x 29.2 | 0.0083 cm²/cm³ | |
| | | | | COSU 04, Unitite Surgical Tray | 165x190 x 55 | 0.0086cm²/cm³ | |
| | | | | COSE 01, Epikut Surgical Tray | 165 x190 x 55 | 0.0086cm²/cm³ | |
| | | | | COSEG 01I, Epikut Surgical Guided Tray | 142x 206 x 72 | 0.0083 cm²/cm³ | |
| Reusable | Reusable up to 250 cycles | | Reusable up to 250 cycles | Reusable up to 250 cycles | Same | | |
| Use-life Testing | Disassembled, cleaned, assembled, sterilized<br>Visual inspection.<br>Component dimensional fit verification<br>Functional closure (lid-base latch) verification | | Disassembled, cleaned, assembled, sterilized<br>Visual inspection<br>Component dimensional fit verification<br>Functional closure (lid-base latch) verification | | Same | | |
| Sterilization Method | | | | | | | |
| Sterilant | Moist heat (steam) | | Moist heat (steam) | | Same | | |
| Cycles | Gravity displacement - Exposure at 121°C °C for 30 minutes and 30 minutes dry time. | | Gravity displacement - Exposure at 121°C °C for 30 minutes and 30 minutes dry time | | Same | | |
| Sterile Barrier | Sterilization wrap, FDA cleared for indicated method and cycle | | Sterilization wrap, FDA cleared for indicated method and cycle | | Same | | |