Search Filters

Search Results

Found 34 results

510(k) Data Aggregation

    K Number
    K253047

    Validate with FDA (Live)

    Device Name
    PRESSONE™
    Date Cleared
    2026-03-24

    (183 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Not Found

    Device Description

    Not Found

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K260533

    Validate with FDA (Live)

    Device Name
    ELISIO™-H
    Date Cleared
    2026-03-19

    (30 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ELISIO™-H dialyzers are intended for hemodialysis, hemodiafiltration, hemofiltration, and isolated ultrafiltration in patients with acute kidney injury or chronic kidney disease when conservative therapy is judged to be inadequate.

    Device Description

    The ELISIO™-H is a single-use, high-flux hemodialyzer intended for the extracorporeal treatment of patients with renal failure. The device is constructed of a plastic cylindrical housing comprised of two compartments, one for blood and one for dialysate, separated by a semi-permeable membrane. Blood is circulated through the blood compartment, while dialysate flows counter currently through the dialysate compartment. This flow pattern enables the removal of uremic toxins, excess fluid, and electrolytes via diffusion and/or convection, depending on the selected treatment modality.

    The dialyzer interfaces with the patient via an ISO 8637 compliant blood tubing set and must be used with dialysis machines equipped with an ultrafiltration controller or accurate fluid balancing system. The ELISIO™-H features the POLYNEPHRON™ membrane (Polyethersulfone), for high clearance efficiency and a polypropylene housing not made with BPA or DEHP.

    The device is available in eight sizes (surface area): 0.9, 1.1, 1.3, 1.5, 1.7, 1.9, 2.1, and 2.5 m². It is sterile (gamma radiation) and non-pyrogenic with a three-year shelf life. No integrated components or accessories are included.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K222852

    Validate with FDA (Live)

    Date Cleared
    2023-05-25

    (246 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nipro SafeTouch Needle is indicated for general purpose injection of fluid from vials, and to and from parts of the body below the surface of the skin. It includes a needle shield to prevent against accidental needlestick injuries. The needle is compatible with standard Luer-lock syringes.

    The Nipro Syringe with SafeTouch Needle is indicated for general purpose injection of fluid from vials, and to and from parts of the body below the surface of the skin. It includes a needle shield to prevent against accidental needlestick injuries.

    Device Description

    The Nipro SafeTouch Needle is a disposable hypodermic single lumen needle with attached safety device. The safety device is a needle shield that covers the needle post treatment eliminating accidental needlestick injuries. It is provided with and without a piston syringe.

    The Nipro SafeTouch Needle is available in gauges of 18-30, lengths of ½". The needles are available individually or attached to a luer lock syringe in volumes of 1, 3, 5 and 10mL.

    The device is sterile, single use only, non-toxic, and non-pyrogenic. It is sterilized by Ethylene Oxide gas and has a shelf-life of 5 years.

    AI/ML Overview

    The provided text describes specific acceptance criteria and study details for the Nipro SafeTouch Needle and Nipro Syringe with SafeTouch Needle, but it does not describe a study proving the device meets acceptance criteria in the context of an AI/ML device. Instead, it outlines a 510(k) premarket notification for a traditional medical device, comparing it to a predicate device.

    The document discusses non-clinical tests and a simulated clinical study, but these are for a physical medical device, not an AI/ML diagnostic or therapeutic system. Therefore, I cannot provide information on items such as multi-reader multi-case studies, standalone algorithm performance, ground truth establishment for AI training sets, or sample sizes for AI training sets, as these concepts are not relevant to the provided text.

    However, I can extract the acceptance criteria and performance related to the physical medical device and the simulated clinical study.

    Here's the information based on the provided text, focusing on the device's physical and functional properties:

    Table of Acceptance Criteria and Reported Device Performance

    Item for ComparisonAcceptance Criteria (Predicate Device K162081 BD Luer-Lok™ Hypodermic Syringe with BD Eclipse™ Hypodermic Needle performance or recognized standards)Reported Device Performance (Subject device: K222852 SafeTouch Needle / Syringe with SafeTouch Needle)Conclusion
    1. Device Description
    1.1. Indications for UseGeneral purpose injection and aspiration of fluid; mechanism to cover needlepoint after use to prevent accidental needle sticks.General purpose injection and aspiration of fluid; needle shield to prevent accidental needlestick injuries.Similar - no new issues of safety and effectiveness
    1.2. Mechanism of ActionMechanical/ManualMechanical/ManualSame
    1.3. Technology OverviewGraduated disposable piston syringe with safety hypodermic needle (Stainless-Steel cannula, plastic hub, needle shield).Graduated disposable piston syringe with safety hypodermic needle (Stainless-Steel cannula, plastic hub, needle shield).Same
    1.4. Anatomical LocationIntramuscular, intravascular, or subcutaneous injectionIntramuscular, intravascular, or subcutaneous injectionSame
    2. Material Components
    2.1. Syringe barrelStyrene acrylic copolymerPolypropyleneSimilar - no new issues of safety and effectiveness
    2.2. LubricantSiliconeSiliconeSame
    2.3. PlungerPolypropylenePolypropyleneSame
    2.4. StopperPolyisoprene RubberPolyisoprene RubberSame
    2.5. Needle CannulaStainless SteelStainless SteelSame
    2.6. Needle Hub with integrated safety mechanism (needle shield)PolypropylenePolypropyleneSame
    3. Physical Characteristics
    3.1. Needle Gauge20 - 3018 - 30Nipro provides a larger gauge needle. Testing results demonstrate no new issues of safety and effectiveness
    3.2. Needle hub ColorColor-coded to the appropriate gauge needle per ISO 6009Color-coded to the appropriate gauge needle per ISO 6009Same
    3.3. Needle Length1/2" - 1"1/2" - 1 1/2"Nipro provides a longer length needle. Testing results demonstrate no new issues of safety and effectiveness
    3.4. Syringe Volume1 - 10 mL1 - 10 mLSame
    4. Sterility
    4.1. MethodGammaEO (Ethylene Oxide)Both methods are recognized as Category A.
    4.2. Single-UseYesYesSame
    4.3. SAL (Sterility Assurance Level)10^-610^-6Same
    4.4. Shelf Life5 years5 yearsSame
    5. Biological Specifications
    5.1. Systemic InjectionNo abnormalities or mice deathNo abnormalities or mice deathWithin Criteria; Same
    5.2. Intracutaneous reaction testNo erythema, edema, or necrosisNo erythema, edema, or necrosisWithin Criteria; Same
    5.3. Hemolysis testHemolytic index shall not exceed 2%Hemolytic index shall not exceed 2%Within Criteria; Same
    5.4. Bacterial Endotoxin<= 20 EU/device<= 20 EU/deviceWithin Criteria; Same
    5.5. Microbial barrier testNo microbial growth was observedNo microbial growth was observedWithin Criteria; Same

    Regarding the study proving the device meets acceptance criteria:

    The document describes a series of non-clinical bench tests and a simulated clinical study for the physical device.

    1. Sample size used for the test set and the data provenance:

      • Simulated Clinical Study: 500 subject devices were evaluated. The provenance of the data (country of origin, retrospective/prospective) is not specified, but it's a prospective study since it's a simulated clinical evaluation of new devices.
      • Bench Testing: The document states "Bench testing was performed on the subject device and the predicate device" but does not specify the exact sample sizes for each type of bench testing.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Simulated Clinical Study: 10 "trained professionals" evaluated the devices. Their specific qualifications (e.g., nurses, doctors) or years of experience are not specified.
      • Predicate Device Comparison: The predicate device serves as a benchmark for comparison. Its "ground truth" (i.e., its established safety and efficacy) is derived from its prior 510(k) clearance (K162081).
      • Bench Testing: Standards (e.g., ISO, ASTM, USP) define the acceptance criteria and methods, and accredited labs conduct these tests. The "experts" in this context are the testing personnel skilled in performing the specified tests.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Simulated Clinical Study: The document states "There were no device failures, no user-related adverse events or safety device related adverse events." This implies a pass/fail assessment based on observed events, not an adjudication process requiring multiple expert opinions on a specific outcome (like in image interpretation).
      • Bench Testing: Adjudication is not typically applicable for these types of objective physical and chemical tests, where results are measured against defined criteria in standards.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not Applicable. This document is for a physical medical device (needle/syringe with safety feature), not an AI/ML diagnostic or therapeutic system. Therefore, MRMC studies and concepts like "human readers improving with AI assistance" are not relevant.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not Applicable. This document is for a physical medical device, not an AI/ML algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Simulated Clinical Study: User observation and event reporting (device failures, adverse events). The "ground truth" is the absence or presence of these events as observed by the trained professionals.
      • Bench Testing: Quantitative measurements (e.g., dimensional specifications, hemolytic index, bacterial endotoxin levels) and qualitative assessments (e.g., visual inspection, microbial growth) against predefined criteria in recognized standards (ISO, ASTM, USP).
    7. The sample size for the training set:

      • Not Applicable. This document describes a physical medical device. There is no AI/ML "training set" in this context. The manufacturing and design process relies on engineering principles and quality control, not machine learning model training.
    8. How the ground truth for the training set was established:

      • Not Applicable. As there is no AI/ML training set, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K212677

    Validate with FDA (Live)

    Device Name
    aboNT SYRINGE
    Date Cleared
    2022-04-15

    (234 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indicated for intramuscular administration of Dysport® for the temporary improvement of moderate to severe glabellar lines in adults <65 years of age.

    Device Description

    The aboNT Syringe is a piston syringe with needle consisting of graduated barrel, plunger rod, and gasket. It is provided with a permanently attached hypodermic needle that is 31G x 5/16″. The syringe barrel is 0.3 mL in volume, with a graduated scale labeled in both volumetric (mL) and corresponding unit dosing for Dysport® (60 units/0.3mL). The syringe is sterile, single use only, non-toxic, non-pyrogenic and sterilized by E-beam radiation. The shelf-life has been determined to be 5 years.

    AI/ML Overview

    The provided text describes the submission of a medical device (aboNT Syringe) for FDA clearance (K212677) and details its non-clinical testing for substantial equivalence to a predicate device. It explicitly states that no clinical testing was performed or provided, and therefore, a multi-reader multi-case (MRMC) comparative effectiveness study with human readers assisted by AI was not conducted, and a standalone algorithm performance study was not conducted.

    Here's an analysis of the provided information regarding acceptance criteria and studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding "reported device performance" in the format one might expect for a study evaluating an AI algorithm's diagnostic performance. Instead, it compares the subject device (aboNT Syringe) to a predicate device (UniTox® Syringe) across various technological characteristics. The "performance" in this context refers to meeting established standards for syringe manufacturing and function, rather than diagnostic accuracy.

    However, we can extract some "acceptance criteria" based on the comparisons and testing outlined. The "reported device performance" is implicitly that the subject device meets these standards and is "within specification."

    Acceptance Criteria (Derived from Standards/Predicate Comparison)Reported Device Performance (Subject Device)
    Intended Use: Injection of botulinum toxin (Dysport®)Meets (Indicated for intramuscular administration of Dysport®)
    Syringe Type: Piston syringe with attached needleMeets (Piston syringe with attached needle)
    Design: Permanently attached needle, low-dead space, specific toxin dose graduation markingsMeets (Permanently attached needle, low-dead space, specific toxin dose graduation markings)
    Operational Principles: ManualMeets (Manual)
    Biocompatibility: Compliant to ISO 10993-1Compliant to ISO 10993-1
    Barrel Transparency: No particle and extraneous matterNo particle and extraneous matter
    Delivery Accuracy (Less than half nominal cap.): Within specified range (Predicate: Min: 0.0957 mL, Max: 0.1008 mL)Min: 0.0968 mL, Max: 0.1024 mL (Within specification)
    Delivery Accuracy (Equal to or greater than half nominal cap.): Within specified range (Predicate: Min: 0.485 mL, Max: 0.496 mL)Min: 0.269 mL, Max: 0.300 mL (Within specification, noting volume difference)
    Reuse Durability: NA – single use onlyNA – single use only
    Needle Cover Strength: Within specified range (Predicate: Min: 0.52 kgf, Max: 1.57 kgf)Min: 0.32 kgf, Max: 0.51 kgf (Within specification)
    Hub/Needle Bond Strength: Within specified range (Predicate: Min: 42.65 N)Min: 44.72 N (Within specification)
    Sterilization Method: E-Beam or equivalentE-Beam (Similar, no new issues)
    Shelf-life: 5 years5 years
    Visual Inspection: Meets ISO 7886-1, ISO 7864Meets standards
    Dimensional Specifications: Meets ISO 7886-1Meets standards
    Mechanical & Performance Characteristics: Meets ISO 7886-1, ISO 7864, ISO 8537, ISO 11608-1Meets standards
    Chemical & Biological Characteristics: Meets ISO 7886-1, ISO 8537Meets standards
    Sterilization Validation: Meets ISO 11137-1, -2, -3Meets standards
    Shelf-life Testing: Meets ASTM F1980-16, ASTM D4169-16Meets standards
    Dye Penetration & Seal Strength: Meets ASTM F1929-15, ASTM F88/F88M-1Meets standards
    Biocompatibility: Meets ISO 10993-1 methods (Cytotoxicity, Sensitization, Intracutaneous Irritation, Pyrogen, Extractables/Leachables)Meets standards
    Particulate Matter: Meets USP <788>Meets standard

    2. Sample size used for the test set and the data provenance

    The document does not specify a distinct "test set" in the context of an AI algorithm evaluation with sample sizes. The testing conducted is primarily non-clinical, involving material and performance assessments according to established standards. For these physical and chemical tests, sample sizes would typically be defined by the specific standard, but are not detailed in this summary. The data provenance is internal to the manufacturer's testing processes as required for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a non-clinical evaluation of a medical device (syringe), not an AI algorithm requiring expert ground truth for diagnostic accuracy.

    4. Adjudication method for the test set

    Not applicable. This is a non-clinical evaluation of a medical device (syringe), not an AI algorithm requiring expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not conducted. The document explicitly states: "This submission does not warrant any clinical testing, therefore no clinical testing performed for or provided in this submission." This device is a syringe, not an AI-powered diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    No, a standalone algorithm performance study was not done. The device is a syringe, not an AI algorithm.

    7. The type of ground truth used

    For the non-clinical tests, the "ground truth" is defined by the specifications and acceptable ranges established by recognized international and national standards (e.g., ISO 7886-1, ISO 7864, ASTM F1980-16, USP <788>, etc.). The device's performance is compared against these scientifically established benchmarks.

    8. The sample size for the training set

    Not applicable. This is a non-clinical evaluation of a syringe; there is no "training set" in the context of machine learning.

    9. How the ground truth for the training set was established

    Not applicable. There is no "training set" or corresponding ground truth establishment for an AI algorithm in this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190037

    Validate with FDA (Live)

    Date Cleared
    2019-10-24

    (289 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cronus HP PTA Balloon Catheter is indicated for Percutaneous Transluminal Angioplasty in the following vessel areas:

    • Femoral arteries
    • Popliteal arteries
    • Iliac arteries
    • Renal arteries
    • For the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
      These catheters are not intended for use in coronary arteries or the neurovasculature.
    Device Description

    The Cronus™ HP PTA Balloon Catheter is intended for PTA (Percutaneous Transluminal Angioplasty) procedures. It is an over the wire (OTW) 0.035" dual lumen catheter with a distally mounted semi- compliant inflatable balloon and a flush cut tip. The catheter manifold includes two lumens. A y-connector adaptation is located at the proximal part of the catheter and provides access to the two different lumens. The lumen marked "GW" is the central lumen of the catheter which terminates at the distal tip. This lumen is used to pass the catheter over a guidewire with a maximum outer diameter of 0.035 inches. The lumen, marked "BALLOON" is used to inflate and deflate the dilatation balloon with a solution of contrast medium and saline. The balloon has two radiopaque markers for positioning the balloon relative to the stenosis. The balloon segment expands to a known diameter at a specific inflation pressure.
    The Cronus™ HP PTA Balloon Catheter is sterilized using ethylene oxide gas and satisfies a minimum Sterility Assurance Level (S.A.L.) of 10 °. Shelf-life is determined to be 4 years.
    The device is available in multiple configurations ranging in sizes of balloon diameter 4-10mm, balloon lengths 20-80mm, and catheter lengths of 45 and 80cm

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding the Cronus HP PTA Balloon Catheter. It focuses on demonstrating substantial equivalence to a predicate device, rather than proving a device meets specific acceptance criteria based on performance study results.

    Therefore, the requested information regarding acceptance criteria and a study proving the device meets them cannot be extracted from the provided text. The document explicitly states:

    • "This submission does not warrant any clinical testing, therefore no clinical testing was performed for or provided in this submission."

    This indicates that clinical performance studies to establish specific acceptance criteria and prove the device meets them were not part of this 510(k) submission. The submission relies on non-clinical (bench) testing and comparison of technological characteristics to a legally marketed predicate device to demonstrate substantial equivalence.

    Here's what can be inferred from the document regarding the device's evaluation (though it doesn't directly address the acceptance criteria and study proving performance as you've requested):

    The 510(k) process for this device, as detailed, focuses on demonstrating substantial equivalence to an existing predicate device (Cronus HP (K151141)). This means the manufacturer is asserting their new device is as safe and effective as the predicate device, not necessarily that it meets a set of isolated performance criteria from a clinical trial.

    The "tests" mentioned are primarily bench tests and biocompatibility tests, which establish the basic safety and physical characteristics of the device, and ensure it functions as intended in a laboratory setting. These are not clinical studies designed to show performance against specific clinical acceptance criteria.

    To answer your specific questions in the context of this document's limitations:

    1. A table of acceptance criteria and the reported device performance:

      • Not applicable / Not provided. The document lists various non-clinical tests (e.g., packaging burst, balloon compliance, balloon rated burst pressure) and states they were "conducted successfully." However, it does not provide specific quantitative acceptance criteria or reported performance values for these tests in a table, nor does it present the results of a clinical study proving performance against acceptance criteria.
    2. Sample size used for the test set and the data provenance:

      • Not applicable / Not provided for clinical data. Since no clinical testing was performed, there's no test set for clinical data. For the bench tests, sample sizes are not explicitly stated in this summary. Data provenance would be laboratory testing in an ISO-certified facility likely, but specifics are missing.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. No clinical test set where "ground truth" would be established by experts is mentioned.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. No clinical test set requiring adjudication is mentioned.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device is a PTA Balloon Catheter, not an AI-assisted diagnostic tool.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a physical medical device, not an algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Not applicable. For the bench tests, physical measurements and engineering standards serve as the "truth," but this isn't analogous to clinical ground truth.
    8. The sample size for the training set:

      • Not applicable. This device does not involve machine learning or a "training set."
    9. How the ground truth for the training set was established:

      • Not applicable.

    In summary, this FDA document is a 510(k) summary for a physical medical device that achieved clearance through demonstrating substantial equivalence to a predicate device, primarily via non-clinical testing. It did not involve or report on a clinical study designed to prove the device met specific acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191359

    Validate with FDA (Live)

    Device Name
    Nipro Syringe
    Date Cleared
    2019-10-09

    (141 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To inject fluids into or withdraw fluids from the body.

    Device Description

    The NIPRO Syringe is a piston syringe consisting of graduated barrel, plunger rod, and gasket. It is provided in luer slip, luer lock and eccentric luer slip tips, with and without attached hypodermic needles are available in ranges of 20-27G, and 3/8 to 1 ½" in length. Gauges are color coded for easy differentiation. Its function is mechanical. The syringe is sterile, single use only, non-toxic, non-pyrogenic and sterilized by Ebeam radiation. The shelf-life has been determined to be 5 years.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device, specifically a Nipro Syringe. It describes a conventional medical device, not an AI/ML-driven device. Therefore, the information requested in the prompt regarding acceptance criteria, study details, expert involvement, and ground truth for an AI/ML device is not present in the provided document.

    The document discusses:

    • Device Description: A piston syringe with varied sizes, gauges, and tips.
    • Indications for Use: To inject or withdraw fluids from the body.
    • Technological Characteristics: Materials, design, sterilization, expiry dating, and dimensions.
    • Non-Clinical Performance Testing: A comprehensive list of biocompatibility, visual inspection, dimensional, mechanical, chemical, sterilization, transportation, pyrogen, and shelf-life tests performed to demonstrate substantial equivalence to a predicate device.
    • Clinical Study Conclusion: Explicitly states "No clinical Study is included in this submission."

    Since the prompt is specifically about the acceptance criteria and study that proves the device meets the acceptance criteria (implying an AI/ML device context), and the document describes a physical medical device (syringe), it does not contain the information necessary to answer the questions about AI/ML device performance, ground truth, expert consensus, or MRMC studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182940

    Validate with FDA (Live)

    Date Cleared
    2019-07-19

    (269 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SURDIAL DX Hemodialysis System is indicated for implementing hemodialysis treatments in patients with acute and chronic kidney (renal) failure. The system can be used in hospitals, healthcare facilities and outpatient dialysis centers when prescribed by a physician and used by qualified operators. The SURDIAL DX accommodates the use of both low flux and high flux hemodialysis.

    Device Description

    The SURDIAL DX Hemodialysis System is a single patient dialysis unit that incorporates the mechanism of mixing and producing of dialysate by quantitative mixing method and the mechanism of ultrafiltration control by use of a piston pump. The SURDIAL DX consists of 4 sections: Panel operating section, Blood circuit control section, Liquid circuit control section and Power source control section. CAN communication is used for communication between each section.

    AI/ML Overview

    This document is a 510(k) Summary for the NIPRO SURDIAL DX Hemodialysis System, demonstrating its substantial equivalence to a predicate device. It details performance specifications, but it does not describe a study involving a test set with ground truth established by experts, MRMC studies, or standalone algorithm performance, as these are typically associated with artificial intelligence/machine learning (AI/ML) device submissions.

    The "performance data" section in this document refers to engineering and quality assurance testing (biocompatibility, electrical safety, software verification, environmental testing, human factors) to ensure the device functions as intended and meets established safety standards for a hemodialysis machine. It does not describe a study that would involve a "test set" and "ground truth" derived from expert consensus in the context of diagnostic accuracy, which is what your questions imply.

    Therefore, for aspects related to AI/ML device evaluations (e.g., test set sample size, expert consensus, MRMC studies, standalone performance, training set details), the information is not available in the provided text.

    Here's a breakdown of the available information based on your request, focusing on the device's engineering and safety performance rather than diagnostic accuracy:


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the comparative table against the predicate device, where "Same" or "Similar" indicates meeting equivalence. Specific quantitative acceptance values are embedded within the "Subject Device: SURDIAL DX" column, and its performance is those reported values, as the testing confirmed it met those specifications.

    Aspect of Device PerformanceAcceptance Criteria (from Subject Device Specifications)Reported Device Performance (from Subject Device Specifications)
    Water Supply
    Pressure range0.5 – 7.4 bar0.5 – 7.4 bar (report indicates acceptable)
    Temperature range5 - 30 °C / 41- 86 °F5 - 30 °C / 41- 86 °F (report indicates acceptable)
    Alarm "No water"Internally monitored by pressure sensor and flow regulatorInternally monitored by pressure sensor and flow regulator (report indicates acceptable)
    Extracorporeal Circulation
    Blood pump pumping rate0, 10 to 600 mL/min (5 mL/min increment)0, 10 to 600 mL/min (5 mL/min increment) (report indicates acceptable)
    Tolerance interval (Blood Pressure)< 10 % for blood pressure up to -200 mmHg< 10 % for blood pressure up to -200 mmHg (report indicates meeting stricter tolerance than predicate)
    Working pressure range (Inlet)Inlet pressure up to: -500 mmHgInlet pressure up to: -500 mmHg (report indicates acceptable)
    Working pressure range (Outlet)Max. outlet pressure: +500 mmHgMax. outlet pressure: +500 mmHg (report indicates acceptable)
    Heparin pump pumping rate0.0 to 10.0 mL/h (0.1 mL/h increment)0.0 to 10.0 mL/h (0.1 mL/h increment), with specific bolus rates for 10, 20, 30 mL syringes (report indicates acceptable)
    Heparin pump flow accuracyFlow accuracy +/- 10 %Flow accuracy +/- 10 % (report indicates acceptable)
    Pressure range (Heparin pump)0 to 450 mmHg0 to 450 mmHg (report indicates acceptable)
    Dialysate System
    Flow (dialysate)100 to 800 mL/min (100 mL/min increment)100 to 800 mL/min (100 mL/min increment) (report indicates acceptable)
    DF tolerance at dialysis machine0 to +10 % from the set value0 to +10 % from the set value (report indicates compliance with IEC standard)
    Bridging time of protection systemNot bridgeable during dialysisNot bridgeable during dialysis (report indicates acceptable)
    Ultrafiltration
    Operating range0 to 4,000 mL/h0 to 4,000 mL/h (report indicates acceptable)
    Overall accuracyFbal ± 0.15 mL/chamber cycle, FUF Ultrafiltration pump tolerance < 1 %Fbal ± 0.15 mL/chamber cycle, FUF Ultrafiltration pump tolerance < 1 % (report indicates acceptable)
    Dialysate Temperature
    Temperature setting range34 - 39 °C/93 - 102 °F34 - 39 °C/93 - 102 °F (report indicates compliance with IEC standard)
    Temperature tolerance-0.5 °C to +31 °F to 33 °F-0.5 °C to +31 °F to 33 °F (report indicates acceptable)
    Heating power1800 W1800 W (report indicates acceptable)
    Dialysate Conductivity
    Operating range (Bicarbonate)2.4 – 7.0 mS/cm2.4 – 7.0 mS/cm (report indicates acceptable)
    Operating range (Total)12.5 – 16.5 mS/cm12.5 – 16.5 mS/cm (report indicates acceptable)
    Tolerance±0.2 mS/cm or less±0.2 mS/cm or less (report indicates acceptable)
    ABPM Blood Pressure Monitoring
    Cuff pressure range0 - 300 mmHg0 - 300 mmHg (report indicates acceptable)
    AccuracyWithin ±3 mmHgWithin ±3 mmHg (report indicates acceptable)
    Blood pressure determination timeTypical 25 secondsTypical 25 seconds (report indicates acceptable)
    Pulse rate determination40 – 200 bpm40 – 200 bpm (report indicates acceptable)
    Pulse rate accuracyThe smaller of ±2% or ±2 bpmThe smaller of ±2% or ±2 bpm (report indicates acceptable)
    Overpressure cut-offIn normal operation: 300 mmHg; In single failure: 330 mmHgIn normal operation: 300 mmHg; In single failure: 330 mmHg (report indicates acceptable)
    Time of alarm suppression< 1 second< 1 second (report indicates acceptable)
    Deactivation of acoustic signal120 seconds120 seconds (report indicates acceptable)
    Transmembrane Pressure
    Alarm limit (max. TMP)0 to +700 mmHg0 to +700 mmHg (report indicates acceptable)
    Alarm limit (min. TMP)-200 to +200 mmHg-200 to +200 mmHg (report indicates acceptable)
    Bridging time of protection systemNot bridgeable during dialysisNot bridgeable during dialysis (report indicates acceptable)
    Deactivation of acoustic alarm120 seconds120 seconds (report indicates acceptable)
    Degassing System
    Tolerance± 20 mmHg± 20 mmHg (report indicates acceptable, tighter than predicate)
    Arterial and Venous Pressure Monitoring
    Arterial inlet pressure (PA) measurement tolerance+/- 10 mmHg+/- 10 mmHg (report indicates acceptable)
    Venous return pressure (PV) measurement tolerance+/- 10 mmHg+/- 10 mmHg (report indicates acceptable)
    Bridging time of protection systemNot bridgeable during dialysisNot bridgeable during dialysis (report indicates acceptable)
    Bridging time of protective system (PV)Not bridgeable during dialysisNot bridgeable during dialysis (report indicates acceptable)
    Deactivation of acoustic alarm (PA)120 seconds120 seconds (report indicates acceptable)
    Deactivation of acoustic alarm (PV)120 seconds120 seconds (report indicates acceptable)
    Dialysate Conductivity Monitor
    Limit±5 % (of set value)±5 % (of set value) (report indicates acceptable)
    Bridging time of protection systemcannot be bridged during dialysiscannot be bridged during dialysis (report indicates acceptable)
    Deactivation of acoustic alarm120 seconds120 seconds (report indicates acceptable)
    Ultrafiltration Controller
    Protection systemYesYes (report indicates acceptable)
    Bridging time of protection systemNot bridgeable during dialysisNot bridgeable during dialysis (report indicates acceptable)
    Deactivation of acoustic alarm120 seconds120 seconds (report indicates acceptable)
    Air Detector System
    SensitivityAlarms for single bubble > 10 µL or accumulated 0.3 µL+ over 1 min into volume of 1, 10, 50 or 300 µL; Default 10 µLSpecified sensitivity met (report indicates acceptable)
    Protection systemUltrasonic transmission methodUltrasonic transmission method (report indicates acceptable)
    Bridging time of protective systemNot bridgeable during dialysisNot bridgeable during dialysis (report indicates acceptable)
    Blood Leak Detector
    MeasurementOpticalOptical (report indicates acceptable)
    Tolerance+/- 100 ppm+/- 100 ppm (report indicates acceptable)
    Alarm thresholdSetting range: 50 to 500 ppm (10 ppm increment); Default: 280 ppm (Ht 32 %)Specified alarm thresholds met (report indicates acceptable)
    Bridging time of protective systemNot bridgeable during dialysisNot bridgeable during dialysis (report indicates acceptable)
    Deactivation of acoustic alarm120 seconds120 seconds (report indicates acceptable)
    Temperature Monitor – Dialysate
    Excessive-temperature protectionLower limit 33°C/91°F; Upper limit 40°C/104 °FLower limit 33°C/91°F; Upper limit 40°C/104 °F (report indicates acceptable)
    Protection systemTemperature sensorTemperature sensor (report indicates acceptable)
    Bridging time of protection systemCannot be bridged during dialysisCannot be bridged during dialysis (report indicates acceptable)
    Deactivation of acoustic alarm120 seconds120 seconds (report indicates acceptable)
    Fail-safe design (power failure)
    BatteryPower failure backup batteryPower failure backup battery (report indicates acceptable)

    2. Sample Size for Test Set and Data Provenance

    This document does not describe a "test set" in the context of an AI/ML diagnostic study with patient data. The "performance data" refers to validation testing of the device itself (e.g., electrical, mechanical, software functionality, biocompatibility). Therefore, there is no sample size of patient data or data provenance (country, retrospective/prospective) relevant to a diagnostic AI model study.

    3. Number of Experts and Qualifications for Ground Truth

    Not applicable. Ground truth from experts is not established for this type of device performance testing. The "ground truth" here is the engineering specification that the device must meet, tested through controlled experimental conditions in a lab.

    4. Adjudication Method for Test Set

    Not applicable. There is no expert adjudication for the engineering performance and safety tests described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This type of study is for evaluating the impact of AI assistance on human reader performance, typically in diagnostic imaging. The SURDIAL DX Hemodialysis System is a treatment device, not a diagnostic AI system, and this document does not refer to any such studies.

    6. Standalone Performance (Algorithm Only)

    Not applicable. This device is a hemodialysis machine. While it contains software, its "performance" is its ability to conduct hemodialysis treatments safely and effectively, not to perform a standalone diagnostic algorithm. The software verification and validation section confirms software functions as intended for controlling the machine.

    7. Type of Ground Truth Used

    The "ground truth" for this device's performance testing is adherence to engineering specifications, recognized international standards (e.g., ISO, IEC), and documented functional requirements. For biocompatibility, it's the results of specific laboratory tests against established biological safety criteria. For software, it's the functionality of the code against its design specifications and requirements.

    8. Sample Size for Training Set

    Not applicable. This document does not describe an AI/ML model that would have a "training set" of data.

    9. How Ground Truth for Training Set was Established

    Not applicable, as there is no AI/ML training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173029

    Validate with FDA (Live)

    Device Name
    Nipro Syringe
    Date Cleared
    2018-03-30

    (183 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    to inject fluids into or withdraw fluids from the body

    Device Description

    The NIPRO Syringe is a piston syringe consisting of graduated barrel, plunger rod, and gasket. It is provided in luer slip, luer lock and eccentric luer slip tips, with and without attached hypodermic needles are available in ranges of 20-25G, and 3/8 to 1 ½" in length. Gauges are color coded for easy differentiation. Its function is mechanical. The syringe is sterile, single use only, non-toxic, non-pyrogenic and sterilized by Ebeam radiation. The shelf-life has been determined to be 5 years. This device is not intended for patients less than 2 years of age.

    AI/ML Overview

    The Nipro Syringe is a medical device and thus the provided document is a 510(k) summary, which outlines its substantial equivalence to a predicate device rather than a study proving unmet acceptance criteria in the context of AI/ML performance. Therefore, most of the requested information regarding AI/ML clinical study specifics (such as expert involvement, adjudication, MRMC studies, standalone performance, and ground truth establishment for training and test sets) is not applicable or cannot be extracted from this document.

    However, I can extract the acceptance criteria related to the device's physical and performance characteristics, and how the device's performance was evaluated against these criteria.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes non-clinical tests performed to ensure the Nipro Syringe meets established standards. The acceptance criteria are implicit in the "specifications of the standards listed," and the reported device performance is that these tests were "performed within the specifications" and that the device "performs equivalent to the predicate device."

    Acceptance Criteria (Standards Met)Reported Device Performance
    Visual Inspection (ISO 7886-1, ISO 7864-1)Performed within specifications of the standards; no significant differences.
    Dimensional Specifications (ISO 7886-1)Performed within specifications of the standards; slight dimensional changes from predicate, but both devices remain in compliance with all aspects of ISO 7886-1.
    Gauging (ISO 594-1)Performed within specifications of the standards.
    Mechanical and Performance Characteristic (ISO 7886-1, ISO 7864-1, ISO 594-1, ISO 594-2)Performed within specifications of the standards; results demonstrate equivalent performance to the predicate device.
    Chemical and Biological Characteristic (ISO 7886-1, ISO 11137-1, and specific biological tests)Performed within specifications of the standards. Includes: Cytotoxicity, Sensitization, Intracutaneous Irritation, Systemic Toxicity, Pyrogen testing (Material Mediated & LAL), Hemolysis, Particulate Analysis.
    Sterilization Validation (ISO 11137-1, ISO 11137-2, ISO 11137-3)Performed within specifications of the standards; Ebeam sterilization validated.
    Transportation Tests (ASTM-D4169-09)Performed within specifications of the standards.
    Biocompatibility (ISO 10993)Performed within specifications; tests conducted: MEM Elution (Cytotoxicity), Kligman Maximization (Sensitization), Intracutaneous Irritation, ISO Systemic Toxicity Study in Mice, USP Rabbit Pyrogen Study, ASTM Hemolysis Study, USP Limulus Amebocyte Lysate (LAL) testing, Particulate Analysis Device fluid pathway.
    Pyrogen Testing (ISO 10993-11, USP 39-NF34: 2016, <151>)Performed within specifications of the standards.
    Shelf-life Testing (ASTM F1980-07)Performed within specifications of the standards; 5-year shelf-life determined.

    2. Sample Size Used for the Test Set and Data Provenance:

    This information is not explicitly provided in the document. The document refers to non-clinical tests performed to specific standards, but it doesn't detail the sample sizes used for each test or the provenance of any data (e.g., country of origin, retrospective/prospective). As this is a physical medical device, not an AI/ML system, "test set" in the context of data for model evaluation is not directly applicable.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    Not applicable. This device is a physical syringe, not an AI/ML system requiring expert consensus for ground truth establishment. The "ground truth" for this device's performance is determined by meeting the specifications of established engineering and biological standards.

    4. Adjudication Method for the Test Set:

    Not applicable for a physical device. The compliance with standards and their specifications serves as the evaluation basis, not human adjudication of AI output.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    Not applicable. This is not an AI-assisted diagnostic or treatment device, so there is no human-in-the-loop performance improvement with or without AI to measure.

    6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study:

    Not applicable. This is a physical medical device, not an algorithm. Its performance is inherent to its design and manufacturing, evaluated through non-clinical testing against established standards.

    7. Type of Ground Truth Used:

    The "ground truth" for the Nipro Syringe is compliance with internationally recognized and industry-accepted standards (ISO, ASTM, USP) for medical devices, specifically for piston syringes and needles. This includes performance benchmarks for:

    • Physical dimensions and integrity
    • Mechanical function
    • Sterility
    • Biocompatibility
    • Chemical characteristics
    • Shelf-life

    8. Sample Size for the Training Set:

    Not applicable. This is a physical device, not an AI/ML model, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable. As there is no training set for an AI/ML model, the concept of establishing ground truth for it is irrelevant in this context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161681

    Validate with FDA (Live)

    Date Cleared
    2017-03-08

    (264 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nipro SafeTouch IV Catheter is a passive anti-needle stick intra-vascular catheter that is inserted into the patient's vascular system for short term use for the infusion of fluids, and/or blood components. 16-24 gauge catheters may be used with power injectors for which the maximum pressure setting is 300psi

    Device Description

    Nipro SafeTouch IV Catheter consists of an over-the-needle, catheter made of radiopaque polyurethane with a passive safety mechanism that will engage upon removal from the patient in order to prevent accidental needlestick injuries. The device is available in gauges 16-24 and in catheter lengths of ¾" – 2". It is also available with or without wings on the hub. Gauges are color coded for easy differentiation. Nipro SafeTouch IV Catheter is sterilized by Ethylene Oxide gas (EtO). It is packaged in a peel blister that ensures the sterility of the device until package is opened and labeled for single use only. A sleeve of SafeTouch Cath contains 5 units of blister packages. An inner box contains 10 sleeves (50 units). An outer box contains 40 inner boxes (3000 units). Its shelf life period is 5 years.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the "Nipro SafeTouch IV Catheter". Since it's a 510(k) submission, the primary goal is to demonstrate substantial equivalence to a legally marketed predicate device, not necessarily to prove optimal performance against specific clinical acceptance criteria in the same way a new drug or novel high-risk device might.

    Based on the provided information, the device is an Intravascular Catheter with a passive anti-needle stick safety mechanism. The acceptance criteria and supporting studies are focused on its non-clinical performance and a simulated clinical study for the safety mechanism.

    Here's the breakdown of the information requested:

    1. Table of Acceptance Criteria and the Reported Device Performance

    The acceptance criteria are not explicitly stated with quantitative thresholds in the provided document beyond "All of the above tests were performed within the specifications of the standards listed above." The performance is reported as "passed with zero failures" for the simulated clinical study and "performs equivalent to the predicate device" for other tests.

    Acceptance Criteria CategorySpecific Test/EvaluationReported Device Performance
    Mechanical/Physical1. Disengagement force between componentsPerformed within specifications of internal test method
    2. Penetration forcePerformed within specifications of internal test method
    3. Air leakagePerformed within specifications of internal test method
    4. Flow ratePerformed within specifications of internal test method
    8. Corrosion resistance (ISO 10555-1:1995)Performed within specifications of ISO 10555-1:1995
    Safety Mechanism5. Safety mechanism performance (simulated clinical study)Passed with zero failures
    Sterilization & Packaging6. Package integrity/microbial barrier (ISO 11607-2)Performed within specifications of ISO 11607-2
    9. Sterilization validation (ISO 11135-1:2007, ISO 11135-2:2008)Performed within specifications of ISO 11135-1/11135-2
    10. EtO residuals (ISO 10993-7)Performed within specifications of ISO 10993-7
    11. Transportation tests (ASTM-d4169-09)Performed within specifications of ASTM-d4169-09
    14. Shelf-life testing (ASTM F1980-07)Performed within specifications of ASTM F1980-07
    Biocompatibility7. Chemical testing / acid, alkalinity and extractable metals (ISO 7864:1993)Performed within specifications of ISO 7864:1993
    12. Biocompatibility (ISO 10993, various sub-parts)Performed within specifications of ISO 10993
    13. Pyrogen testing (USP 35-NF30:2012, <85>)Performed within specifications of USP 35-NF30:2012, <85>
    Particulate matter testingPerformed within specifications

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • The document does not specify the sample size used for any of the non-clinical tests or for the "simulated clinical study" that evaluated the safety mechanism.
    • The provenance of the data (country of origin, retrospective/prospective) is not mentioned. These are likely internal laboratory tests conducted by Nipro Medical Corporation or a contracted testing facility.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This information is not provided. Since the tests are largely non-clinical engineering and laboratory tests, "experts" in the context of clinical interpretation or diagnostic decision-making are not relevant. The "ground truth" for these tests would be defined by the specifications of the relevant international standards (e.g., ISO, ASTM, USP).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This information is not provided and is generally not applicable to the types of non-clinical, objective tests described here. Adjudication methods are typically used in clinical studies where subjective human interpretation of conditions or outcomes requires consensus.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done.
    • This device is an IV catheter, not an AI-assisted diagnostic tool or imaging device. There is no AI component, and therefore no assessment of human reader improvement with AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • This question is not applicable as the device is an IV catheter, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For the non-clinical tests, the "ground truth" is established by the specifications and requirements defined within the referenced international and national standards (e.g., ISO, ASTM, USP). The device performance is compared against these predefined physical, chemical, and biological limits or parameters.
    • For the "simulated clinical study" of the safety mechanism, the ground truth would be a binary outcome of successful activation/engagement of the safety mechanism as per its design specifications.

    8. The sample size for the training set

    • This device is not an AI/machine learning device, so the concept of a "training set" is not applicable.

    9. How the ground truth for the training set was established

    • This question is not applicable as the device does not involve a training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K160444

    Validate with FDA (Live)

    Date Cleared
    2017-02-08

    (357 days)

    Product Code
    Regulation Number
    876.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Hemodialysis with the FB-U Hemodialyzer is indicated for patients with acute or chronic renal failure when conservative therapy is judged to be inadequate. It also may be indicated in the treatment of patients intoxicated with poisons or drugs.

    Device Description

    The Nipro FB-U hemodialyzer is a medical device intended for use as an artificial kidney system for the treatment of patients with renal failure. During treatment, blood is circulated from the patient, through the extracorporeal system, into the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate.

    AI/ML Overview

    Device: Nipro FB-U Hemodialyzer

    1. Table of Acceptance Criteria and Reported Device Performance:

    Performance MetricAcceptance Criteria (Predicate)Reported Device Performance (Nipro FB-U Hemodialyzer)
    ClearancesNot explicitly stated, implied to be equivalent to predicate.Demonstrated to be equivalent to predicate in non-clinical tests.
    Ultrafiltration coefficientNot explicitly stated, implied to be equivalent to predicate.Demonstrated to be equivalent to predicate in non-clinical tests.
    Volume of blood compartmentNot explicitly stated, implied to be equivalent to predicate.Demonstrated to be equivalent to predicate in non-clinical tests.
    Pressure dropNot explicitly stated, implied to be equivalent to predicate.Demonstrated to be equivalent to predicate in non-clinical tests.
    Positive and negative pressure testsNot explicitly stated, implied to be equivalent to predicate.Passed (included in non-clinical tests).
    Blood compartment integrityNot explicitly stated, implied to be equivalent to predicate.Passed (included in non-clinical tests).
    Transportation testsNot explicitly stated, implied to be equivalent to predicate.Performed and results included in submission.
    BiocompatibilityNot explicitly stated, implied to be equivalent to predicate.Performed and results included in submission.
    Shelf-life testingNot explicitly stated, implied to be equivalent to predicate.Performed and results included in submission.
    Materials - HousingPolypropylene/polycarbonate (for predicate and/or similar devices)Polypropylene
    Materials - MembranePolyether sulfone/Cellulose Triacetate (for predicate and/or similar devices)Cellulose Triacetate
    DesignMechanical device consisting of a cylindrical rigid housing enclosing hollow fibersSame
    Intended UseAs part of an artificial kidney system for removal of toxins and excess water from bloodSame
    Energy SourceTo be used with dialysis machines equipped with an ultrafiltration controller or accurate fluid balancing systemSame

    2. Sample Size Used for the Test Set and Data Provenance:

    The document states "Non-clinical tests submitted," which include "dimensional and performance to include: positive and negative pressure tests, blood compartment integrity, clearances, Ultrafiltration coefficient, volume of blood compartment, and pressure drop." It also mentions "Other test items include: Transportation tests, biocompatibility, and shelf-life testing."

    • Sample Size: The document does not explicitly state the specific sample sizes used for each of these non-clinical tests.
    • Data Provenance: The tests are described as "non-clinical tests submitted," implying they were conducted by Nipro Medical Corporation as part of their submission. No information is provided regarding the country of origin of the data or whether it was retrospective or prospective, though non-clinical tests are generally prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    Not applicable. The study relies on objective physical and chemical measurements (e.g., clearances, pressure tests) rather than expert interpretation of data for establishing ground truth in non-clinical testing.

    4. Adjudication Method for the Test Set:

    Not applicable. As the study primarily concerns non-clinical, objective measurements, an adjudication method for a "test set" (in the sense of human interpretation) is not relevant.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. No MRMC study was conducted. This device is a hemodialyzer, not an AI-powered diagnostic or assistive technology for human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a physical hemodialyzer, and therefore the concept of a standalone algorithm or human-in-the-loop performance is not relevant. The performance evaluation is of the device's physical and functional characteristics.

    7. The Type of Ground Truth Used:

    The ground truth for the non-clinical tests was established through objective physical and chemical measurements and engineering standards. For example:

    • Clearances were measured against established methods for filtration efficiency.
    • Pressure tests confirmed resistance to specified pressures.
    • Integrity tests confirmed the physical structural soundness.
    • Biocompatibility tests likely adhered to ISO standards (e.g., ISO 10993).
    • Shelf-life tests involved stability and performance over time.

    The "ground truth" for showing substantial equivalence was the performance characteristics of the legally marketed predicate devices.

    8. The Sample Size for the Training Set:

    Not applicable. This is a physical medical device, not an AI or machine learning algorithm that requires a training set. The performance data is derived from testing of the physical device.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 4