Found 15 results

510(k) Data Aggregation

    K Number
    K251626

    Validate with FDA (Live)

    Date Cleared
    2025-11-03

    (159 days)

    Product Code
    Regulation Number
    872.6640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Midmark Dental Delivery Systems are intended to provide dental professionals with low voltage electric, air, and water to operate dental handpieces, syringes, and accessories during dental examinations and procedures.

    Midmark Assistant's Delivery Systems are intended to provide dental professionals with air, water, and suction to operate dental handpieces, syringes, and accessories during dental examinations and procedures.

    Device Description

    The Midmark Dental Delivery System encompasses two types of dental units: Dental Delivery Systems and Assistant's Delivery Systems.

    The Dental Delivery System is a workstation for the dental clinical space that provides air, water and electrical power for the operation of dental instruments. The system has two models with three mounting options that can provide instrument positioning for right-handed and left-handed clinicians practicing to the patient's side or for clinicians practicing at the 12 o'clock position. The DS-1100-001 model mounts to the base of the Midmark Dental Chair. This model has a combination of swing arms and a flex arm providing both horizontal and vertical positioning of the delivery unit. The DS-1200-001 model mounts to a cabinet or column base and also includes an Assistant's Delivery System. The cabinet mount has a combination of swing arms providing horizontal positioning of the delivery unit to either side of a clinician practicing behind the patient at the 12 o'clock position. The adjustable column mount option that connects to a cabinet adds vertical positioning adjustment to the DS-1200-001.

    The Dental Delivery System contains an air and water control manifold designed to route coolant air, drive air, and water to the active dental handpiece. This manifold consists of a pneumatically actuated diaphragm mechanism and an integrated air and water distribution block. When handpieces are docked in their respective instrument holders, air pressure in the handpiece line applies force to the diaphragm within the manifold, closing the associated air ports and preventing air and water flow. When a handpiece is undocked from its holder, a valve within the holder releases the air pressure against the diaphragm. When the foot control is activated, drive and coolant air pressures overcome the diaphragm, opening the previously closed air port. The foot control activation simultaneously sends a pneumatic signal to open the corresponding water valve within the manifold assembly. As a result, the drive air, coolant air, and coolant water are able to flow through the distribution pathways to the undocked handpiece. Returning the handpiece to its holder triggers the holder's valve mechanism to re-pressurize the diaphragm, re-closing the air and water distribution pathways.

    The Assistant's Delivery System is a workstation for the dental clinical space that provides air, water, and vacuum for the operation of dental instruments. The ADS-1100-001 model mounts to the head end of the Midmark Dental Chair. A junction box located at the foot end of the chair base provides a contained space for the connection of the Assistant's Delivery System to air, water, and vacuum sources. The ADS-1200-001 mounts to a Midmark cabinet base. The DS-1200-001 Assistant's Delivery System mounts on a movable arm which is common to the Dental Delivery System. The Assistant's Delivery System mounts to the lower side of its supporting arm and the Dental Delivery System mounts to the upper side of the arm allowing both to pivot independently for positioning of the instruments to the clinician. The cabinetry base has a space for the connection of the Assistant's Delivery System to air, water, and vacuum sources.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K233026

    Validate with FDA (Live)

    Date Cleared
    2024-04-04

    (195 days)

    Product Code
    Regulation Number
    880.6880
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Midmark Smart M9® and Smart M11® Sterilizers can be used in medical and dental offices, hospitals, clinics, nursing homes, laboratories, and other facilities to sterilize heat and moisture stable items (including dental handpieces) that are compatible with steam sterilization. Refer to Standard Cycle Parameters for detailed information.

    Device Description

    The Midmark Smart M9® and Smart M11® Sterilizers utilize steam flush pressure pulse (SFPP) technology to achieve sterilization on heat and moisture stable items (including dental handbieces) that are compatible with steam sterilization.

    The Midmark Smart M9® and Smart M11® Sterilizers are compact, self-contained portable units. They can be placed on any level support surface where an electrical outlet is available with no other installation required. The M9 sterilizer models are the smaller of the models with a 9 in (22.9 cm) diameter x 15 in (38.1 cm) deep stainless-steel chamber. The M 1 sterilizer models have an 11" (27.9 cm) diameter x 18" (45.7 cm) deep stainless-steel chamber.

    The Midmark Smart M9® and Smart M11® Sterilizers use saturated steam at high pressure and temperature as the sterilizing agent to kill infectious bio-organisms on items placed in the chamber for processing. They use the dynamic Steam Flush Pressure Pulse (SFPP) cycle type for the pre-set cycles that include a 4-minute 270°F and a 3-minute 275°F cycle for wrapped or pouched instruments (including dental handpieces) and cassettes, a 30-minute 250°F cvcle for textiles and instrument packs requiring a lower temperature. and a 3-minute 270°F cycle for unwrapped instruments. The 3-minute 275°F is added to align with Table B.2 in AAMI TIR12:2020. In the SFPP type cycle, residual air is removed from the chamber and its contents by a series of controlled pressure pulses and steam flushes.

    The Midmark Smart M9® and Smart M11® Sterilizers are designed to automate the sterilization process, to the extent possible, and the user interface on the subject models extends this capability to include sterilization record keeping. To use the sterilizer, the operator fills the water reservoir with water (distilled or purified) and loads the included trays with properly cleaned and prepped instruments for sterilization. The loaded trays are then placed inside the chamber, and the chamber door is manually closed by the operator. Based on the cycle parameters that are appropriate for the type of load being processed, the operator then selects the appropriate sterilization cycle on the user interface. On the subject units there are options for the user to enter load type and indicator information. Once the cycle is selected and the operator presses "Start," there is an option to capture the identification of the operator. The sterilizer then automatically performs all the operations necessary to complete the sterilization process without further interaction from the operator. The sterilization cycle is composed of several phases which include Filling, Heating, Sterilization, Venting, and Drying, Audible signals indicate cycle initiation, completion, and/or interruption, and the user interface provides visual communication of device status, operator instructions, and troubleshooting information. The LED light bar also provides an estimate of the cycle progression. At the conclusion of the cycle there is another option to capture the identification of the operator that unloads and approves or rejects the results of the sterilization cycle.

    All sterilizer cycle and user maintenance (Routine Care) records are stored internally on an SD card. Midmark has developed a Digital Ecosystem Connectivity Module that is incorporated into the Midmark Smart M9® and Smart M11® Sterilizers which adds the ability to transfer electronic sterilization records to the Midmark cloud and to remotely view the current status of, or cycle history for multiple sterilizers. Optional features and settings that may be distributed to multiple sterilizers from the cloud include distribution of software updates to the connectivity module, compliance settings, and entering the results of biological indicator tests. Connectivity is not needed to perform the sterlizer's intended use, and connecting a sterilizer does not change the intended use.

    AI/ML Overview

    The provided document is a 510(k) summary for the Midmark Smart M9® and Smart M11® Sterilizers. It details the device's characteristics and its non-clinical performance data to demonstrate substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and study proving the device meets them, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this device are based on its ability to perform sterilization effectively and safely according to established standards. The reported device performance is indicated by its compliance with these standards.

    Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standards Used)Reported Device Performance (Results)
    Performance: ANSI/AAMI ST55:2016 Table-Top Steam Sterilizers (FDA Recognition Number 14-518)Passed; the results of the evaluation demonstrate compliance of the device to the standards.
    Safety: ANSI/UL 61010-1 Third Edition 5/12/2012 / Revised 7/19/2019, Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use; Part 1: General Requirements (FDA Recognition Number 19-41)Passed; the results of the evaluation demonstrate compliance of the device to the standards.
    Safety: IEC 61010-2-040:2020 Edition 3.0 Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use - Part 2-040: Particular Requirements for Sterilizers and Washer-Disinfectors Used to Treat Medical MaterialsPassed; the results of the evaluation demonstrate compliance of the device to the standards.
    EMC (Electromagnetic Compatibility): IEC 60601-1-2 Edition 4.1 2020-09 Consolidated Version Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests (FDA Recognition Number 19-36)Passed; the results of the evaluation demonstrate compliance of the device to the standards.
    Pressure Vessel: American Society of Mechanical Engineers (ASME) Boiler and Pressure Vessel Code, Section VIII, Division 1: 2023 EditionPassed; the results of the evaluation demonstrate compliance of the device to the standards.

    Study Details:

    The document states that non-clinical evaluations were performed for the Midmark Smart M9® and Smart M11® Sterilizer. The purpose of these evaluations was to demonstrate compliance with recognized standards for safety and performance, thereby proving substantial equivalence to a legally marketed predicate device.

    1. Sample size used for the test set and the data provenance:

      • The document does not specify a sample size for the test set (e.g., number of sterilization cycles tested, number of units tested).
      • The data provenance is not explicitly stated in terms of country of origin or whether it was retrospective or prospective. However, given that it concerns a medical device seeking FDA clearance, it's implied that the testing was conducted under controlled conditions to regulatory standards.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not applicable to this type of device and study. The ground truth for a sterilizer is established through physical and biological indicators demonstrating the elimination of microorganisms, not through expert consensus on image interpretation or diagnosis. The "experts" implicit in this context are the testing laboratories and their personnel following the specified scientific and engineering standards.
    3. Adjudication method for the test set:

      • This is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation (e.g., radiology reads) to resolve discrepancies. For device performance testing against engineering standards, the results are objective measurements (e.g., temperature, pressure, biological indicator kill) that either pass or fail the predefined criteria.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

      • No, an MRMC comparative effectiveness study was not performed. This type of study is relevant for diagnostic imaging AI/CAD systems that assist human readers. The Midmark Smart M9® and Smart M11® Sterilizers are steam sterilizers, and their performance is evaluated directly against physical, chemical, and biological sterilization parameters, not through human interpretation of cases.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • This question is not applicable in the conventional sense of AI/algorithm performance. The device is a physical sterilizer with automated cycles. Its "performance" is inherently "standalone" in that it performs its function without human intervention during the sterilization cycle once initiated. The evaluations listed are standalone device performance tests.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth for a sterilizer's performance is objectively defined by the complete kill of specific biological indicators (e.g., bacterial spores known to be highly resistant to sterilization processes) under specified conditions (temperature, pressure, time). It also includes measured physical parameters (e.g., temperature and pressure profiles over time) matching the standards, and the successful completion of cycles without errors. This is a scientific and engineering ground truth, not based on human consensus or pathology from patient outcomes.
    7. The sample size for the training set:

      • The document does not mention a "training set" in the context of machine learning or AI. This device is a traditional medical device (steam sterilizer), not a deep learning or AI-based diagnostic tool that requires a training set. The design and validation of the sterilizer are based on established engineering principles and extensive physical and biological testing.
    8. How the ground truth for the training set was established:

      • This question is not applicable as there is no mention of a training set for an AI/ML algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K201667

    Validate with FDA (Live)

    Date Cleared
    2020-08-06

    (48 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the Midmark EOIS is to provide dental radiographic examination to aid in the diagnosis of diseases of the teeth, jaw and oral structures. When the system is equipped with the cephalometric option, the system will provide cephalometric radiographic examinations for use in orthodontic treatment planning and evaluation. When the system is equipped with the CBCT option, the system will also provide volumetric and tomographic images of the oral and maxillofacial region, for diagnostic examination of teeth, jaws, oral structures, and some cranial bones.

    Device Description

    The Midmark Extraoral Imaging System (EOIS) is a diagnostic X-ray system using an extraoral source intended for dental radiographic examination and diagnosis of diseases for the following regions of the patient's anatomy:

    • . dental (teeth)
    • . oral and maxillofacial region (mouth and jaw)
      It produces the following radiographic images:
    • . Panoramic (PAN) images used for diagnostic examination of dentition (teeth), jaws and oral structures.
    • Cephalometric (CEPH) images of maxillofacial region and parts of the skull for CEPH examination, if equipped with the CEPH feature.
    • . Carpal images assisting in estimating bone-age, if equipped with the CEPH feature, and when the carpal attachment is used.
    • . Cone Beam Computed Tomography (CBCT) volumetric and tomographic images of the oral and maxillofacial region for diagnostic examination of dentition (teeth), jaws, oral structures, and some cranial bones, if equipped with the CBCT small or medium feature.
      The main modality is limited to PAN radiographic images but can be upgraded by installing the CEPH imaging feature, allowing the device to provide CEPH radiographic images for use in orthodontic treatment planning and evaluation. This extension is also intended to allow carpal imaging to assist in estimating bone-age.
      The Midmark EOIS can be upgraded by installing a 3D option allowing the device to produce computed volumetric and tomographic images of the listed above anatomy regions by computed tomographic reconstruction of the digital X-ray data. The X-ray data is captured by two dimensional images taken by exposing the patient anatomy at different anales with a cone shaped X-ray beam. Reconstruction software converts these two-dimensional images into a three-dimensional data set.
    AI/ML Overview

    The provided text focuses on the K201667 510(k) submission for the Midmark Extraoral Imaging System (EOIS), detailing its indications for use, technological characteristics, and non-clinical performance data.

    However, it does not contain detailed information about acceptance criteria for an AI device, nor does it describe a study that explicitly proves an AI device meets such criteria. The document describes a traditional medical imaging device (X-ray system) and its performance, not an AI/ML medical device.

    Therefore, I cannot extract the specific information requested in the prompt regarding AI device acceptance criteria and study details. The document describes:

    • Non-clinical Performance Data (Section 5.6): This includes medical electrical safety testing, imaging performance testing (against IEC 61223-3-4), dose comparison evaluation, and CBCT modality testing.
    • Clinical Performance Data (Section 5.7): This mentions an "Image Performance study" (evaluating image quality of phantom images by dental practitioners for diagnostic acceptability) and a "Usability study" (examining use-related hazards with simulated use involving clinicians).

    None of these describe an AI study with its specific acceptance criteria as detailed in the prompt's numbered requests. The "Reconstruction Engines" and "iterative reconstruction" mentioned in Section 5.4 are image reconstruction algorithms, not AI algorithms designed for diagnosis or specific clinical tasks requiring the kind of rigorous AI validation outlined in the prompt.

    Given the provided text, I must state that the information required to answer the prompt's specific questions about AI device acceptance criteria and a study proving it meets them is NOT present.

    The document discusses imaging system performance and safety, but not AI model performance against explicit clinical acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163337

    Validate with FDA (Live)

    Date Cleared
    2017-08-31

    (276 days)

    Product Code
    Regulation Number
    880.6880
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Midmark and Ritter M9 and M11 UltraClave® Automatic Sterilizers and the Ritter M9D AutoClave® Automatic Sterilizer can be used in medical and dental offices, hospitals, clinics, nursing homes, laboratories, and other facilities to sterilize heat and moisture stable items (including dental handpieces) that are compatible with steam sterilization. Refer to Standard Cycle Parameters on the following page for detailed information.

    Device Description

    The Midmark and Ritter M9 and M11 UltraClave and Ritter M9D AutoClave Self-Contained steam sterilizers are designed to sterilize medical, surgical, laboratory, and dental items, using pressurized steam. The M9 UltraClave models include a variation designated as the M9D Autoclave, the only difference being that the M9 UltraClave includes a motor to automatically open the door for drying and the M9D Autoclave is a manual model which requires the operator to open the door for drying. They are self-contained devices that are electrically powered with automatic controls for controlling time and temperature that are designed for safe, easy operation with minimal operator training.

    AI/ML Overview

    The provided document describes the Midmark and Ritter M9 and M11 UltraClave Automatic Sterilizers and the Ritter M9D AutoClave Automatic Sterilizer, which are self-contained steam sterilizers used in medical and dental offices, hospitals, clinics, nursing homes, laboratories, and other facilities to sterilize heat and moisture stable items.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device is a steam sterilizer, and its performance is evaluated based on its ability to achieve sterility For medical devices, the common performance criteria for sterilization are the achievement of a certain temperature and pressure for a specified duration, resulting in a Sterility Assurance Level (SAL) of 10^-6, meaning the probability of a single viable microorganism remaining after sterilization is one in a million.

    The document implicitly defines acceptance criteria through the "Standard Cycle Parameters" and confirms the device's performance through validation studies.

    Acceptance Criterion (Standard Cycle Parameter)Reported Device Performance (Achieved during validation)
    Unwrapped Cycle:
    Sterilizing Temperature: 270°F (132°C)Achieved (implicit through successful biological testing and adherence to standard)
    Sterilizing Time: 3 min.Achieved (implicit through successful biological testing and adherence to standard)
    Pressure (Ref): 27.1 psi (186 kPa)Achieved (implicit through successful biological testing and adherence to standard)
    Drying Time: 30 min.Achieved
    Pouches Cycle:
    Sterilizing Temperature: 270°F (132°C)Achieved (implicit through successful biological testing and adherence to standard)
    Sterilizing Time: 4 min.Achieved (implicit through successful biological testing and adherence to standard)
    Pressure (Ref): 27.1 psi (186 kPa)Achieved (implicit through successful biological testing and adherence to standard)
    Drying Time: 30 min.Achieved
    Packs Cycle:
    Sterilizing Temperature: 250°F (121°C)Achieved (implicit through successful biological testing and adherence to standard)
    Sterilizing Time: 30 min.Achieved (implicit through successful biological testing and adherence to standard)
    Pressure (Ref): 15 psi (104 kPa)Achieved (implicit through successful biological testing and adherence to standard)
    Drying Time: 30 min.Achieved
    Handpieces Cycle:
    Sterilizing Temperature: 270°F (132°C)Achieved (implicit through successful biological testing and adherence to standard)
    Sterilizing Time: 4 min.Achieved (implicit through successful biological testing and adherence to standard)
    Pressure (Ref): 27.1 psi (186 kPa)Achieved (implicit through successful biological testing and adherence to standard)
    Drying Time: 30 min.Achieved
    Microbiological Performance (SAL):
    Sterility Assurance Level (SAL): 10⁻⁶Achieved (demonstrated through biological indicator overkill method)
    Physical Performance Requirements:All aspects of the standard met.

    The document states: "Testing showed that the M9 and M11 UltraClave and the M9D AutoClave Self-Contained Steam Sterilizers meet all aspects of the standard, including physical and microbiological performance requirements."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: For the physical performance validation, temperature sensors were used to determine the thermal profile of the empty chamber for "three (3) full chamber runs for each pre-programmed cycle parameter" (Unwrapped, Pouches, Packs, Handpieces). For microbiological testing, "biological testing was conducted on fully loaded chambers." No specific number of "fully loaded chambers" or individual biological indicators (BIs) is explicitly stated for the test set, beyond the statement that BIs were inoculated in "most difficult to sterilize locations."
    • Data Provenance: The studies were conducted by "MIDMARK Corporation." The country of origin for the data is not explicitly stated, but Midmark Corporation is based in Versailles, Ohio, USA, suggesting the testing was likely conducted in the USA. The data is prospective, generated specifically for these validation studies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of device (steam sterilizer) does not typically involve human experts in establishing the "ground truth" in the same way an AI diagnostic tool would. The ground truth for sterilization efficacy is established by scientific standards and laboratory methods.

    • Number of Experts: Not applicable in the context of human interpretation. The "experts" are the scientific and technical personnel who designed and executed the validation protocols based on established sterilization standards.
    • Qualifications of Experts: Not explicitly stated, but implied to be qualified personnel in sterilization science, microbiology, and engineering, familiar with relevant standards (e.g., ANSI/AAMI ST55).

    4. Adjudication Method for the Test Set

    Not applicable. Sterilization testing is based on objective, quantifiable measures (temperature, pressure, time, and biological indicator kill rates) rather than subjective expert opinion requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This device is a standalone medical device (sterilizer), not an AI algorithm intended for interpretation by human readers. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant.

    6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance Study

    Yes, a standalone performance study was conducted. The entire validation process, as described, evaluates the device's intrinsic mechanical and microbiological performance. The sterilizer operates autonomously through software-controlled microcontrollers (as described in the "Device Description") to achieve the specified sterilization cycles. The human operator initiates the cycle and loads/unloads the device, but the sterilization process itself is algorithmic (pre-programmed cycles and control loops) and standalone.

    7. Type of Ground Truth Used

    The ground truth used is based on established microbiological and physical standards for sterilization, specifically:

    • Biological Indicator Overkill Method: This is a gold standard for demonstrating sterilization efficacy. Geobacillus stearothermophilus spores are highly resistant, and their complete inactivation (achieving a 10⁻⁶ SAL) confirms the sterilization process is effective.
    • Physical Parameters: Measurement of temperature, pressure, and time profiles within the chamber ensure that the physical conditions required for sterilization were met.
    • Reference Standards: The studies were conducted "in accordance with ANSI/AAMI ST55: 2010/(R)2014," a recognized standard for tabletop steam sterilizers, which defines the ground truth for performance requirements.

    8. Sample Size for the Training Set

    Not applicable. This is a physical device (sterilizer) with an embedded operating algorithm, not an AI model that undergoes "training" on a dataset in the typical machine learning sense. The device's control system is developed through engineering design and robust testing, not statistical training on large datasets.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable for the same reasons as point 8. The "ground truth" for the device's operational parameters (e.g., control logic for heating, venting, sterilizing timing) is established through engineering principles, design specifications, and adherence to regulatory and industry standards for steam sterilization.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161909

    Validate with FDA (Live)

    Device Name
    IQvitals Zone
    Date Cleared
    2016-11-17

    (128 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IQvitals Zone is intended to be used by clinicians and medically qualified personnel for measuring and monitoring:

    · Noninvasive blood pressure for adult and pediatric patients (3 years and above)

    • · Pulse rate for adult and pediatric patients
    • · Noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2) for adult and pediatric patients
    • · Body temperature measured at Temporal Artery for adult and pediatric patients

    The most likely location for IQvitals Zone device to be used is for monitoring patient in general medical locations, hospitals and alternative care environments.

    Device Description

    The IQvitals Zone is designed to be used for measuring and monitoring systolic and diastolic blood pressure, pulse rate, temperature and oxygen saturation (SpO2) for adult and pediatric patients. All functions of the device are performed via the touch screen display, except the power on/off function, which is a separate button on the back of the device.

    The IQvitals Zone can be wirelessly connected with low energy Bluetooth to mobile computers or be connected with a USB cable to computers and has the ability to send the measurement results to the computers.

    The device has a rechargeable lithium ion battery and two mounting options: a mobile cart and a wall mount. All vitals parameters can be simultaneously measured and easily viewed on the touch screen display or the connected computer. Temperature is measured at Temporal Artery and the Temperature Probe is connected serially to the IQvitals Zone.

    AI/ML Overview

    The IQvitals Zone is a multi-parameter vital signs monitor intended for use by clinicians and medically qualified personnel for measuring and monitoring noninvasive blood pressure, pulse rate, noninvasive functional oxygen saturation of arteriolar hemoglobin (SpO2), and body temperature at the temporal artery for adult and pediatric patients (3 years and above).

    1. Table of Acceptance Criteria and Reported Device Performance & 6. Standalone Performance:

    The provided document presents the performance characteristics of the IQvitals Zone and compares them to predicate devices. This comparison implicitly serves as the acceptance criteria for the device, demonstrating that its performance is substantially equivalent to legally marketed devices. Since the device is a vital signs monitor designed to directly measure physiological parameters, the "algorithm only" or "standalone" performance is inherent in these functional specifications.

    CharacteristicAcceptance Criteria (Predicate Device K072516, K101445, K060576, K110028)Reported Device Performance (IQvitals Zone)
    NIBP Monitor Module
    Measurement Accuracy± 5 mmHg (standard deviation < 8 mmHg) Meets ANSI.AAMI SP10:2002 standard accuracy range± 5 mmHg (standard deviation < 8 mmHg) Meets ANSI.AAMI SP10:2002 standard accuracy range
    Measurement Resolution1 mmHg1 mmHg
    Pulse Rate RangeK072516: 30 – 240 BPM; K101445: 30 – 200 BPM (adult, pediatric) / 35 – 220 BPM (neonate)40 – 200 BPM
    Pulse Rate AccuracyK072516: ± 2 bpm or ± 2%, whichever is greater; K101445: ±3 bpm or ±5%, whichever is greater±3 bpm or ±5%, whichever is greater
    Pulse Rate Resolution1 bpm1 bpm
    Initial Cuff PressureK072516: User-selectable; K101445: AutomaticAutomatic or User-selectable
    Maximum Cuff PressureK072516: 290 mmHg (adult mode), 145 mmHg (neonatal mode); K101445: 280 mmHg (adult, pediatric), 130 mmHg (neonate)280 ± 5 mmHg
    Overpressure CutoffK072516: Adult Mode: 290 mmHg, Neonatal Mode: 145 mmHg; K101445: 300 mmHg ±15 mmHg (adult, pediatric), 150 mmHg maximum (neonate)300 ± 30 mmHg
    Pulse Oximetry Specifications
    SpO2 Measurement Range0-100%0-100%
    SpO2 Measurement AccuracyAdult: 70 - 100% (± 2%); Pediatric: 70 - 100% (± 3%)Adult: 70 - 100% (± 2%); Pediatric: 70 - 100% (± 3%)
    Pulse Rate Measurement Range (from SpO2)K060576: 20 - 240 BPM; K110028: 25 - 240 BPM20 - 240 BPM (Nellcor NELL-1) / 25 - 240 BPM (Masimo)
    Pulse Rate Measurement Accuracy (from SpO2)± 3 BPM (no motion); ± 5 BPM (with motion)± 3 BPM (no motion); ± 5 BPM (with motion)
    SpO2 / Pulse Rate ResolutionSpO2: 1%; Pulse Rate: 1 bpmSpO2: 1%; Pulse Rate: 1 bpm
    Thermometer Specifications
    Measurement RangeK011291: 15.5°C to 43°C (61°F to 110°F)15.5°C to 43°C (61°F to 110°F)
    Measurement AccuracyK011291: +/-0.1°C (+/- 0.2°F) [within the +/- 0.6°C (+/-1.0°F) range specified by ASTM E1112 standard]+/-0.1°C (+/- 0.2°F)
    Resolution0.1 °C / 0.1 °F0.1 °C / 0.1 °F
    Response TimeK011291: ~0.04 seconds~0.04 seconds

    2. Sample Size Used for the Test Set and Data Provenance:

    The document states that the clinical validation of the Midmark Non-invasive Blood Pressure (NIBP) module was conducted per ANSI / AAMI / ISO 81060-2:2013. This standard "specifies requirements for the clinical investigation of all automated sphygmomanometers, with the exception of those for neonates." It typically requires a minimum of 85 participants for NIBP clinical investigations to ensure statistically meaningful data for accuracy and precision against a reference measurement. The document does not explicitly state the exact number of participants in their study or the country of origin of the data. Given it's a 510(k) submission to the FDA, it is likely the studies either followed international standards that are acceptable to the FDA or were conducted in the US or in a region that meets FDA requirements. The clinical validation would be prospective.

    For the SpO2 and temperature modules, the document relies on the prior clearance of the OEM modules (Nellcor NELL-1, Masimo MX-5 OEM board, and Exergen TemporalScanner) and their established performance characteristics, indicating that these components have already undergone their own clinical validations. Therefore, specific new clinical studies for these modules within the IQvitals Zone device are not detailed in this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    For the NIBP clinical validation conducted according to ANSI / AAMI / ISO 81060-2:2013, the ground truth is established by two or more trained observers who simultaneously take conventional auscultatory blood pressure measurements (using a mercury or other validated sphygmomanometer and stethoscope) without knowledge of the device's readings. These observers are highly trained healthcare professionals (e.g., physicians, nurses, or technicians) with demonstrated proficiency in indirect blood pressure measurement. Their qualifications and training for consistency across readings are crucial and typically outlined in the study protocol. The document does not specify the exact number of observers or their detailed qualifications, but adherence to the standard implies these criteria were met.

    For the SpO2 and temperature modules, the ground truth would have been established during their original clinical validations using appropriate reference methods (e.g., CO-oximetry for SpO2, and calibrated thermometers for temperature).

    4. Adjudication Method for the Test Set:

    For NIBP validation under ANSI / AAMI / ISO 81060-2:2013, if there are discrepancies between the two trained observers' readings that exceed predefined limits, a third independent adjudicator or reading typically intervenes, or the measurement is discarded. This "2+1" or similar method ensures the accuracy of the reference blood pressure. The document does not explicitly state the adjudication method, but adherence to the standard implies a method for resolving observer discrepancies was used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, the provided documentation describes a 510(k) premarket notification for a vital signs monitor. This type of device directly measures physiological parameters and is not an AI-assisted diagnostic imaging tool that would typically undergo MRMC comparative effectiveness studies to evaluate human reader improvement with AI assistance. Therefore, this section is not applicable.

    7. The Type of Ground Truth Used:

    • Noninvasive Blood Pressure (NIBP): The ground truth for NIBP was established through concurrent auscultatory measurements by trained human observers, as required by the ANSI / AAMI / ISO 81060-2:2013 standard. This is a form of expert consensus/reference measurement.
    • SpO2 and Temperature: The ground truth for the SpO2 and temperature modules would have been established during their original clinical validations. For SpO2, this typically involves CO-oximetry as a reference method. For temperature, it involves highly accurate calibrated reference thermometers.

    8. The Sample Size for the Training Set:

    The document does not explicitly state a "training set" as it would for a machine learning algorithm. For vital sign monitors, the core algorithms (e.g., oscillometric for NIBP, pulse oximetry principles) are typically developed and refined using extensive engineering and physiological data, rather than being "trained" on a discreet dataset in the same way modern AI algorithms are. The "training" in this context refers to the development and internal validation of the algorithms and hardware, which predates the clinical validation detailed. The NIBP clinical validation study serves as the primary external validation of the device's accuracy.

    9. How the Ground Truth for the Training Set Was Established:

    As elaborated in point 8, the concept of a "training set" in the context of this traditional medical device is different from that of modern AI. The ground truth for initial algorithm development and internal validation would involve:

    • NIBP: Use of reference pressure manometers and simulated pulse waveforms, as well as testing on human subjects where reference auscultatory and/or invasive blood pressure measurements are simultaneously obtained.
    • SpO2: In vitro testing against known oxygen saturation levels and in vivo testing with reference CO-oximetry.
    • Temperature: Calibration against known temperature sources and clinical studies comparing to reference thermometers (e.g., rectal, internal core temperature).

    This ground truth is established through rigorous engineering and physiological testing, often involving controlled environments and direct comparison to highly accurate reference standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122643

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-01-07

    (131 days)

    Product Code
    Regulation Number
    872.1800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The indications for use of the Progeny Vantage Panoramic X-Ray System is to provide dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. When the system is equipped with the cephalometric option, the system will also provide cephalometic radiographic examinations for use in orthodontic treatment planning and evaluation.

    Device Description

    The Progeny Vantage Panoramic X-Ray System is an extraoral radiographic imaging system for producing digital radiographs in panoramic and cephalometric views of the teeth, jaw and oral structure. The Progeny Vantage Panoramic Extraoral Radiographic Imaging System consists of the following main components: X-ray tubehead with integrated collimation, Digital Image Receptor, Rotating C-Arm for tubehead and image receptor mounting, Overhead Arm, Elevating Column, Patient Positioning Table, Electronic Control Unit, Computer Display Workstation, 8 ft. coil cord with exposure switch, Optional cephalometric extension arm.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study performed for the Progeny Vantage Panoramic X-Ray System, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state quantitative acceptance criteria for image quality. Instead, the acceptance is based on a determination of "at least as effective as" predicate devices by expert consensus.

    Acceptance Criteria (Implied)Reported Device Performance
    Image quality deemed acceptable for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures.Two licensed dentists deemed image quality acceptable and effective.
    Image quality deemed at least as effective as predicate devices (Instrumentarium OP200D and Planmeca ProMax).Two licensed dentists concluded Progeny Vantage images were at least as effective as Instrumentarium and Planmeca images.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The clinical test involved images taken of "a skull phantom". This suggests a sample size of one skull phantom.
    • Data Provenance: The images were specifically generated for the purpose of this K122643 submission through a clinical test using the Progeny Vantage device itself, as well as images from predicate devices. The text doesn't specify a country of origin beyond the device manufacturer being based in Lincolnshire, IL, USA. This appears to be a prospective test conducted specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Two licensed dentists.
    • Qualifications of Experts: They are described as "licensed dentists." No further details regarding their years of experience or specialization (e.g., oral radiologist) are provided in the document.

    4. Adjudication Method for the Test Set

    The document states: "Both dentists deemed that the image quality was acceptable..." and "...concluded that the Progeny Vantage images were at least as effective as the Instrumentarium and Planmeca images."
    This indicates an unspecified consensus or individual agreement model. There's no mention of a formal adjudication process like 2+1 or 3+1, but rather a collective conclusion from both reviewers.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done.
    • No AI assistance is mentioned in the context of this device. The device is an X-ray system, not an AI-powered diagnostic tool. Therefore, there is no discussion of human readers improving with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not applicable. This device is an X-ray imaging system. Its "performance" is the quality of the image it produces, which is then interpreted by a human (dentist). It does not have an autonomous algorithm that would perform a standalone diagnostic function.

    7. The Type of Ground Truth Used

    The ground truth for evaluating the device's image quality was established through expert consensus/opinion by the two licensed dentists. They reviewed the images and made a subjective judgment on their acceptability and comparative effectiveness against predicate devices.

    8. The Sample Size for the Training Set

    • Not applicable. This document describes a traditional medical device (an X-ray system), not a machine learning or artificial intelligence algorithm that would typically require a training set. The device's operation is based on established physics and engineering principles, not on learned patterns from a dataset.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As stated above, there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K120239

    Validate with FDA (Live)

    Device Name
    ELEVANCE
    Manufacturer
    Date Cleared
    2012-08-24

    (211 days)

    Product Code
    Regulation Number
    872.6640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Elevance is intended for use by professional dental practitioners in providing treatment to dental patients in a dental operatory. The system is designed to deliver air, water, vacuum and low-voltage electricity to hand-held dental instruments.

    Device Description

    The Elevance Delivery Unit includes components to deliver air, water, electrical power, and vacuum to dental handpieces and accessories. The controls are contained in a Doctor's Unit, an Assistant's Unit, and a Cuspidor. Additional parts include mount arms, console mount housing, and a junction box that houses a power supply and air/water regulators. Handpiece accessories or instruments can be added to the Unit, Midmark does not manufacture these accessories but, do provide means to connect them into the Unit. These include high and low speed pneumatic handpiece tubing, electric handpiece motors, scalers, intraoral cameras, curing lights, air/water syringe, and SE and HVE vacuum instruments.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary regarding the Midmark Elevance dental delivery unit, focusing on acceptance criteria and supporting studies:

    This 510(k) document is for a dental delivery unit, not an AI or imaging device. Therefore, the typical "acceptance criteria" and "study" framework involving performance metrics like sensitivity, specificity, F1-score, and ground truth established by experts is not applicable in the same way it would be for an AI-powered diagnostic tool.

    Instead, the acceptance criteria for this type of device revolve around demonstrating substantial equivalence to a predicate device, primarily through engineering and regulatory compliance testing.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (from 510(k) context)Reported Device Performance (from 510(k) summary)
    Technological Characteristics:
    - Indications for Use (consistent)Found to have similar technological characteristics
    - Function of Handpiece Accessories (consistent)Found to have similar technological characteristics
    - Performance of Handpiece Accessories (consistent)Found to have similar technological characteristics
    Safety and Efficacy:
    - No adverse impact from minor differencesMinor differences (centrally located controls, touchpad, proportional solenoid valves for air/water adjustment) determined to not have any impact on the safety or efficacy of the Elevance.
    Non-Clinical Performance:
    - Design Verification (Qualification Run evaluation)Conducted
    - Electrical Safety (e.g., IEC 60601-1)EN 60601-1-2:2007 Part 1-2 conducted; IEC 60601-1 Part 1 conducted
    - Electromagnetic Compatibility (EMC) (e.g., EN 60601-1-2)EN 60601-1-2:2007 Part 1-2 conducted
    - Harmonic Current Emissions (e.g., EN 61000-3-2)EN 61000-3-2:2006 +A1:2009 +A2:2009 Part 3-2 conducted
    - Dental Equipment General Requirements (ISO 7494-1)ISO 7494-1:2004 conducted
    - Dental Equipment Water/Air Supply Requirements (ISO 7494-2)ISO 7494-2:2003 conducted
    Conclusion:The data shows that the Elevance demonstrates substantial equivalence to the predicate as a Dental Delivery Unit.

    For the remainder of the questions, the answers will reflect the nature of device clearance (510(k) for a medical device that is not an AI/imaging diagnostic tool) rather than an AI performance study.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not applicable in the context of an AI/imaging diagnostic test set. The "testing" here refers to engineering and regulatory compliance tests on the physical device itself, not a dataset of patient images or conditions. The number of units tested for electrical safety or functional verification is not specified in the summary but would be determined by internal quality and regulatory standards for device manufacturing.
    • Data Provenance: Not applicable in the context of clinical data provenance. The "data" comes from the physical testing of the Elevance device, likely performed in a lab setting by Midmark Corporation or a contracted testing facility.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: Not applicable. "Ground truth" in the AI/imaging sense does not apply here. The "truth" for this device relies on its adherence to engineering specifications and international standards (like ISO and EN).
    • Qualifications of Experts: Not applicable. The "experts" would be the engineers and technicians who designed, built, and tested the device, along with regulatory specialists who ensured compliance with standards. Their qualifications would be in engineering, quality assurance, and regulatory affairs, not clinical medical expertise for diagnostic ground truth.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. There is no ambiguous result or expert discrepancy requiring adjudication for this type of device testing. Tests either pass or fail based on predetermined criteria outlined in the relevant standards (e.g., specific voltage limits, pressure ranges, insulation resistance).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This is not applicable as the Elevance is a dental delivery unit, not an AI or imaging device that would involve human readers interpreting cases.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This is not applicable as the Elevance is a dental delivery unit, not an algorithm.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: Not applicable in the AI/imaging diagnostic sense. The "ground truth" for this device is its adherence to established engineering specifications, safety standards (e.g., electrical, EMC), and functional performance criteria as defined by international standards (ISO, EN, IEC). It's about meeting an objective, defined physical state or performance level, rather than correlating with a clinical diagnosis or outcome.

    8. The Sample Size for the Training Set

    • Training Set Sample Size: Not applicable. The Elevance is a physical medical device, not an AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • How Ground Truth Was Established: Not applicable. As there is no AI model or training set, there is no "ground truth" to establish for a training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K112380

    Validate with FDA (Live)

    Device Name
    CLEARVISION
    Manufacturer
    Date Cleared
    2011-11-22

    (97 days)

    Product Code
    Regulation Number
    872.1800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ClearVision is intended to be used by dentists and other qualified professionals for producing diagnostic x-ray radiographs of dentition, jaws and other oral structures.

    Device Description

    ClearVision is a digital imaging system for dental radiographic application. The product is to be used for routine dental radiographic examinations such as bitewings, periapicals, etc. Two different sized sensors (size 1 and size 2) are utilized to image different anatomy and for different patient sizes. The CMOS sensor connects directly to a USB connection in a PC without the need for an intermediate electrical interface. ClearVision works with a standard dental intraoral x-ray source without any connection to the x-ray source. ClearVision captures an image automatically upon sensing the production of x-ray and after the x-ray is complete, transfers the image to an imaging software program on the PC. Disposable sheaths are used with each use to prevent cross-contamination between patients.

    AI/ML Overview

    This provides an analysis of the provided text regarding the ClearVision Digital Sensor System.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the ClearVision Digital Sensor System are derived from its comparison to predicate devices (Schick CDR and Gendex GXS-700) and general engineering requirements. The reported device performance indicates equivalency or superiority to these predicates.

    Acceptance Criteria CategorySpecific Criteria/TestPredicate Device A (Schick CDR) PerformancePredicate Device B (Gendex GXS-700) PerformanceClearVision Sensor PerformanceMet?
    Imaging PerformanceImage Line Pair Phantom--Equivalent to GXS-700, Superior to Schick CDRYes
    Image Aluminum Step Wedge--Equivalent to GXS-700, Superior to Schick CDRYes
    Image Tooth Phantom--Equivalent to GXS-700, Superior to Schick CDRYes
    Electrical SafetyIEC 60601-1 compliance--Meets requirementsYes
    EMI/EMCIEC 60601-1-2 compliance--Meets requirementsYes
    DurabilitySensor housing and cable mechanical testing--Met all specified requirementsYes
    ReliabilityConsistent image capture and transfer over extended life--Completely reliableYes
    Image Quality ConsistencyConsistent over expected lifetime exposures to radiation--Meets requirementsYes
    Hermetic ClassificationIP67 per IEC 60529--Meets requirementsYes

    Note: The document states "found to be equivalent to the Gendex GXS-700 in all three tests and superior to the Schick sensor in all three imaging tests," implying that the performance level of the GXS-700 served as the primary benchmark for "equivalency" for the ClearVision's imaging performance.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a specific numerical sample size for the test set used in the imaging performance comparison. It mentions "each sensor to image a line pair phantom, an aluminum step wedge, and a tooth phantom." This implies at least one instance of imaging each of these phantoms per sensor.

    The data provenance is not explicitly stated. Given the context of a 510(k) submission and the nature of the tests (imaging phantoms), it is highly likely that this was prospective data generated in a controlled laboratory or engineering setting, likely within the United States where the company is based. There is no mention of patient data or clinical trials.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not describe the use of human experts to establish ground truth for the test set. The tests performed ("imaging a line pair phantom, an aluminum step wedge, and a tooth phantom") are objective, physical measurements against established standards for image quality and resolution (e.g., line pairs, step wedge density differences). Therefore, the "ground truth" would be inherent in the physical phantoms themselves and the objective metrics used to evaluate the images.

    4. Adjudication Method for the Test Set

    No adjudication method is described, as the evaluation methods appear to be objective and quantitative (e.g., measuring line pairs, density differences). Human interpretation or consensus for ground truth was not mentioned or implied.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study described is a technical performance comparison of the device against predicate devices using physical phantoms, not a clinical study involving human readers or patient cases. Therefore, there is no effect size related to human reader improvement with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the described imaging performance comparison of the ClearVision sensor is a standalone assessment. The evaluation focuses solely on the device's ability to capture and produce images from phantoms, without any involvement of a human interpreter in the loop for diagnostic decision-making during the testing process. The device itself is a digital sensor, not an AI algorithm.

    7. The Type of Ground Truth Used

    The ground truth used for the imaging performance tests was objective, physical standards provided by the phantoms:

    • Line pair phantom: Provides known spatial frequencies (lines per millimeter) to assess resolution.
    • Aluminum step wedge: Provides known material thicknesses/densities to assess contrast and dynamic range.
    • Tooth phantom: Likely provides a realistic but standardized representation of dental anatomy to assess overall image quality and detail capture.

    8. The Sample Size for the Training Set

    The document describes a physical medical device (a digital X-ray sensor), not an AI/machine learning algorithm. Therefore, there is no training set in the context of an algorithm or AI model. The device's "training" or development would have involved engineering design, prototyping, and iterative testing to meet specifications, but not a dataset for training an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Since this is a physical device and not an AI algorithm, the concept of a "training set" and establishing ground truth for it is not applicable. The device's "ground truth" during its development would have been established through engineering specifications, material properties, and performance targets derived from scientific principles and a comparison to existing technologies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K091866

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-07-20

    (26 days)

    Product Code
    Regulation Number
    872.1800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The indications for use of the Progeny Vantage Extra-Oral Panoramic X-Ray System is to provide dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures.

    Device Description

    The Progeny Vantage Panoramic X-ray System is an extraoral radiographic imaging system for producing digital radiographs in a panoramic view of the teeth, jaw, and oral structure. The Progeny Vantage Panoramic Extraoral Radiographic Imaging System consists of the following main components: X-ray tubehead with integrated collimation. Digital Image Receptor Rotating C-Arm for tubehead and image receptor mounting Overhead arm Elevating Column Patient Positioning Table Electronic Control Unit Computer Display Workstation 8 ft. coil cord with exposure switch

    AI/ML Overview

    The provided 510(k) summary for the Progeny Vantage Panoramic X-Ray System primarily focuses on demonstrating substantial equivalence to predicate devices through comparison of technical specifications and indications for use. It does not contain information about specific acceptance criteria related to device performance in terms of diagnostic effectiveness that would typically be evaluated in a study with a ground truth, human readers, or quantitative performance metrics like sensitivity and specificity.

    Therefore, many of the requested sections regarding acceptance criteria, study details, sample sizes, expert qualifications, and ground truth establishment cannot be extracted from this document, as they are not present.

    Here's a breakdown of what can and cannot be answered based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in terms of diagnostic performance (e.g., sensitivity, specificity, accuracy). Instead, it relies on demonstrating substantial equivalence to predicate devices by matching their characteristics. The "reported device performance" in this context refers to the device's technical specifications.

    CharacteristicAcceptance Criteria (Implied by Predicate Devices)Reported Device Performance (Progeny Vantage)
    kVp60-84 kVp (within range of predicates)60-84 kVp
    mA1-16 mA (within range of predicates)4-10 mA
    Digital SensorYesYes
    Image Pixel size66 µm - 96 µm (within range of predicates)96 µm
    Exposure Time2-17s (within range of predicates)8-10s
    Image ProfilesPan, TMJ, Ortho (matching predicates)Pan, TMJ, Ortho
    Operator Exposure ControlDeadman SwitchDeadman Switch
    User InterfaceColor Touch Screen / Keypad & LED DisplayColor Touch Screen
    X-Ray tube focal spot0.4mm² - 0.5mm² (within range of predicates)0.5mm²
    Magnification1.2 - 1.3 (within range of predicates)1.2
    ColumnTelescoping / FixedTelescoping
    ConstructionAluminum castings/extrusions with plastic/metalAluminum castings with plastic/metal covers
    Indications For UseTo provide dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures (matching predicates)To provide dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures
    Safety & Effectiveness (General)Performance testing, software testing, hazard analysis, same indications as predicatesPerformance testing and verification to meet product specifications, Software testing to validate software design and performance, Hazard analysis and risk level assessment, Same indications for use as predicate devices

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not explicitly provided in the document. The filing mentions "Performance testing and verification to meet product specifications" and "Software testing to validate software design and performance" but does not detail the size or nature of the test sets used for these evaluations, nor their provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided. The application focuses on technical equivalence rather than a clinical study requiring expert assessment of ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This type of study was not done, nor is it applicable, as the device is a panoramic x-ray system, not an AI-powered diagnostic tool for interpretation. Its function is to produce images, not to analyze them with AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not provided, and it is not applicable given the nature of the device as an imaging system rather than an algorithm for standalone diagnosis.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    This information is not provided. For an imaging device like this, ground truth for image quality assessments would typically involve objective measurements (e.g., spatial resolution, contrast-to-noise ratio) and potentially subjective evaluation by clinicians to confirm diagnostic utility. However, the document does not elaborate on how these were established.

    8. The sample size for the training set

    This information is not provided. There is no mention of a "training set" as the device is an imaging system, not a machine learning model requiring a training phase for its core function.

    9. How the ground truth for the training set was established

    This information is not provided and is not applicable for the reasons mentioned above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090670

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-07-07

    (116 days)

    Product Code
    Regulation Number
    880.6880
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Midmark M3 UltraFast™ Automatic Sterilizer can be used in medical, dental, and veterinary offices, hospitals, clinics, nursing homes, laboratories, and other facilities to sterilize heat and moisture stable reusable items (including dental handpieces) that are compatible with steam sterilization. Refer to Standard Cycle Parameters, in Table-1, on the following page, for detailed information:

    Device Description

    The M3 UltraFast™ Automatic Sterilizer is a small table-top steam sterilizer that uses saturated steam at high temperature and pressure to kill infectious bio-organisms.

    The device is composed of a water reservoir, an oscillatory pump, a small electric boiler, an ASME certified pressure vessel, surface heaters for drying, a removable door/tray assembly, an electric gear motor for opening and closing the door/tray, and a solenoid valve for venting. The action of all of these components is coordinated by an integral proprietary design electronic control system, which includes a PC board, an LCD display, and a user interface membrane switch. The entire device is enclosed by a plastic and steel case. External to this device is a condensation tank, which is connected to the back of the unit, via a plastic tube.

    Three different fully automatic pre-programmed sterilization cycles with parameters specific to the different load characteristics provide fast and easy use for efficient instrument processing. This sterilizer provides both audible and visual notification upon cycle completion and will dry the load in accordance with the CDC guidelines. A programmable dry cycle allows the user to customize the dry times from 20-60 minutes. This sterilizer has integrated technology to let the operator know if they have low water in the reservoir, or a full external condensing tank eliminating the need to continually monitor water levels.

    AI/ML Overview

    1. Table of Acceptance Criteria and Reported Device Performance:

    The acceptance criteria for the Midmark M3 UltraFast™ Automatic Sterilizer are a Sterility Assurance Level (SAL) of 10⁻⁶ for all three sterilization cycles. The reported device performance demonstrates that the device successfully meets this criterion for each cycle.

    Acceptance Criteria (Sterility Assurance Level - SAL)Reported Device Performance
    Unwrapped Cycle: 10⁻⁶Validated to a SAL of 10⁻⁶ for the Unwrapped Cycle
    Pouches Cycle: 10⁻⁶Validated to a SAL of 10⁻⁶ for the Pouches Cycle
    Low Temp Cycle: 10⁻⁶Validated to a SAL of 10⁻⁶ for the Low Temp Cycle
    Thermal ConditionsAchieves and maintains steady state thermal conditions
    Total Kill Endpoint TimeEstablished as a starting point for cycle development

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the specific "sample size" in terms of number of sterilization runs for the test set. However, it mentions "Repetitive testing" for the validation of each cycle, implying multiple runs were conducted.

    The data provenance indicates that the studies were conducted by a "third party, SPSmedical Supply Corporation," on behalf of Midmark Corp. This suggests an independent laboratory performed the testing. The country of origin of the data is not specified, but given the submission to the FDA, it is highly likely to be the United States. The studies are prospective in nature, as they involve actively testing the device's performance to establish its efficacy.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. The ground truth for sterilization efficacy (total kill endpoint and sterility assurance level) is established through standardized microbiological testing using biological indicators (Geobacillus stearothermophilus spores) rather than human expert opinion. Therefore, the concept of "experts" in the traditional sense of medical image interpretation (e.g., radiologists) is not applicable here. The experts involved would be microbiologists and validation scientists from SPSmedical Supply Corporation, but their specific number and qualifications are not detailed.

    4. Adjudication Method for the Test Set:

    Adjudication methods like "2+1" or "3+1" are typically used in studies involving human interpretation of data where consensus is needed (e.g., image reading). This is not relevant for the type of testing performed for a sterilizer. The "adjudication" is based on objective scientific criteria: the complete kill of a specified number of bacterial spores, as measured by microbiological techniques, to achieve a defined Sterility Assurance Level (SAL).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. MRMC studies are used to evaluate the diagnostic performance of human readers, typically in medical imaging, and are not applicable to the performance evaluation of a sterilizer. The device's effectiveness is measured against objective microbiological and thermal criteria, not against human interpretation performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    Yes, the studies conducted were standalone performance evaluations of the Midmark M3 UltraFast™ Automatic Sterilizer. The tests focused solely on the device's ability to achieve sterilization without human intervention during the sterilization cycle itself. The device is an automatic sterilizer, performing its function independently once initiated.

    7. The Type of Ground Truth Used:

    The ground truth used for the sterilizer's performance evaluation is microbiologically established sterility, specifically a 10⁻⁶ Sterility Assurance Level (SAL), based on the complete inactivation of Geobacillus stearothermophilus spores.

    This includes:

    • Biological Indicators: The use of Geobacillus stearothermophilus spores as a challenge organism. The "total kill endpoint study" directly measures the time required to inactivate these spores.
    • Thermal Profiles: Measurement of temperature throughout the chamber to ensure consistent thermal conditions.
    • Validation of Sterilization Cycles: Confirmation that specific cycles (Unwrapped, Pouches, Low Temp) consistently achieve the desired SAL under defined load conditions.

    8. The Sample Size for the Training Set:

    This information is not applicable to this type of device and study. "Training set" refers to data used to train a machine learning algorithm. The Midmark M3 UltraFast™ Automatic Sterilizer is a physical device, not an AI or software algorithm. Its performance is validated through defined testing protocols, not through a training/testing split of a dataset.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable, as there is no "training set" for a physical sterilizer device. The device's operational parameters (e.g., temperature, pressure, time for each cycle) are designed and engineered by Midmark based on scientific principles of steam sterilization and validated through the performance studies described in the document.

    Ask a Question