K Number
K251626

Validate with FDA (Live)

Date Cleared
2025-11-03

(159 days)

Product Code
Regulation Number
872.6640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Midmark Dental Delivery Systems are intended to provide dental professionals with low voltage electric, air, and water to operate dental handpieces, syringes, and accessories during dental examinations and procedures.

Midmark Assistant's Delivery Systems are intended to provide dental professionals with air, water, and suction to operate dental handpieces, syringes, and accessories during dental examinations and procedures.

Device Description

The Midmark Dental Delivery System encompasses two types of dental units: Dental Delivery Systems and Assistant's Delivery Systems.

The Dental Delivery System is a workstation for the dental clinical space that provides air, water and electrical power for the operation of dental instruments. The system has two models with three mounting options that can provide instrument positioning for right-handed and left-handed clinicians practicing to the patient's side or for clinicians practicing at the 12 o'clock position. The DS-1100-001 model mounts to the base of the Midmark Dental Chair. This model has a combination of swing arms and a flex arm providing both horizontal and vertical positioning of the delivery unit. The DS-1200-001 model mounts to a cabinet or column base and also includes an Assistant's Delivery System. The cabinet mount has a combination of swing arms providing horizontal positioning of the delivery unit to either side of a clinician practicing behind the patient at the 12 o'clock position. The adjustable column mount option that connects to a cabinet adds vertical positioning adjustment to the DS-1200-001.

The Dental Delivery System contains an air and water control manifold designed to route coolant air, drive air, and water to the active dental handpiece. This manifold consists of a pneumatically actuated diaphragm mechanism and an integrated air and water distribution block. When handpieces are docked in their respective instrument holders, air pressure in the handpiece line applies force to the diaphragm within the manifold, closing the associated air ports and preventing air and water flow. When a handpiece is undocked from its holder, a valve within the holder releases the air pressure against the diaphragm. When the foot control is activated, drive and coolant air pressures overcome the diaphragm, opening the previously closed air port. The foot control activation simultaneously sends a pneumatic signal to open the corresponding water valve within the manifold assembly. As a result, the drive air, coolant air, and coolant water are able to flow through the distribution pathways to the undocked handpiece. Returning the handpiece to its holder triggers the holder's valve mechanism to re-pressurize the diaphragm, re-closing the air and water distribution pathways.

The Assistant's Delivery System is a workstation for the dental clinical space that provides air, water, and vacuum for the operation of dental instruments. The ADS-1100-001 model mounts to the head end of the Midmark Dental Chair. A junction box located at the foot end of the chair base provides a contained space for the connection of the Assistant's Delivery System to air, water, and vacuum sources. The ADS-1200-001 mounts to a Midmark cabinet base. The DS-1200-001 Assistant's Delivery System mounts on a movable arm which is common to the Dental Delivery System. The Assistant's Delivery System mounts to the lower side of its supporting arm and the Dental Delivery System mounts to the upper side of the arm allowing both to pivot independently for positioning of the instruments to the clinician. The cabinetry base has a space for the connection of the Assistant's Delivery System to air, water, and vacuum sources.

AI/ML Overview

N/A

U.S. Food & Drug Administration 510(k) Clearance Letter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

November 3, 2025

Midmark Corporation
Rachel Schwieterman
Regulatory Affairs Specialist
60 Vista Drive
Versailles, Ohio 45380

Re: K251626
Trade/Device Name: Midmark Dental Delivery System
Regulation Number: 21 CFR 872.6640
Regulation Name: Dental Operative Unit And Accessories
Regulatory Class: Class I, reserved
Product Code: EIA
Dated: October 2, 2025
Received: October 3, 2025

Dear Rachel Schwieterman:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K251626 - Rachel Schwieterman Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K251626 - Rachel Schwieterman Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

MICHAEL E. ADJODHA -S

Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K251626

Please provide the device trade name(s).
Midmark Dental Delivery System

Please provide your Indications for Use below.

Midmark Dental Delivery Systems are intended to provide dental professionals with low voltage electric, air, and water to operate dental handpieces, syringes, and accessories during dental examinations and procedures.

Midmark Assistant's Delivery Systems are intended to provide dental professionals with air, water, and suction to operate dental handpieces, syringes, and accessories during dental examinations and procedures.

Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

Midmark Dental Delivery System Page 11 of 65

Page 5

Midmark Dental Delivery System 510(k) Summary

Midmark Corporation
60 Vista Drive
Versailles, Ohio 45380
1-800-MIDMARK
midmark.com

Designing better care.™

In accordance with 21 CFR §807.92, a 510(k) summary is provided below with the required information.

1. Administrative Information

Date Summary PreparedOctober 30, 2025
510(k) Sponsor AddressMidmark Corporation60 Vista DriveVersailles, Ohio 45380
Contact PersonRachel SchwietermanRegulatory Affairs SpecialistEmail: rtschwieterman@midmark.comPhone: 937-526-8567Fax: 937-526-8482
Trade NameMidmark Dental Delivery System
Common NameDental Operative Unit
Classification NameDental Operative Unit and Accessories(21 CFR 872.6640, Product Code EIA)
Device ClassificationClass I
Review PanelDental
510(k) NumberK251626

2. Equivalent Predicate Comparators

Manufacturer NameTrade Name510(k) No.Decision Date
Midmark CorporationElevance Delivery UnitK120239August 24, 2012

Page 6

Midmark Dental Delivery System 510(k) Summary

3. Indications for Use

Midmark Dental Delivery Systems are intended to provide dental professionals with low voltage electric, air, and water to operate dental handpieces, syringes, and accessories during dental examinations and procedures.

Midmark Assistant's Delivery Systems are intended to provide dental professionals with air, water, and suction to operate dental handpieces, syringes, and accessories during dental examinations and procedures.

4. Models

Model NumberModel Name
DS-1100-001Midmark Left Right - Chair Mounted Dental Delivery System
DS-1200-001DS-1200-CABINETMidmark 12 O'clock - Dental and Assistant's Cabinet-Mounted Delivery System
DS-1200-COLUMNMidmark 12 O'clock Dental and Assistant's Column-Mounted Delivery System
ADS-1100-001Midmark Chair Mounted - Assistant's Delivery System
ADS-1200-001Midmark 12 O'clock - Cabinet Mounted- Assistant's Delivery System

5. Device Description

The Midmark Dental Delivery System encompasses two types of dental units: Dental Delivery Systems and Assistant's Delivery Systems.

The Dental Delivery System is a workstation for the dental clinical space that provides air, water and electrical power for the operation of dental instruments. The system has two models with three mounting options that can provide instrument positioning for right-handed and left-handed clinicians practicing to the patient's side or for clinicians practicing at the 12 o'clock position. The DS-1100-001 model mounts to the base of the Midmark Dental Chair. This model has a combination of swing arms and a flex arm providing both horizontal and vertical positioning of the delivery unit. The DS-1200-001 model mounts to a cabinet or column base and also includes an Assistant's Delivery System. The cabinet mount has a combination of swing arms providing horizontal positioning of the delivery unit to either side of a clinician practicing behind the patient at the 12 o'clock position. The adjustable column mount option that connects to a cabinet adds vertical positioning adjustment to the DS-1200-001.

Page 7

Midmark Dental Delivery System 510(k) Summary

The Dental Delivery System contains an air and water control manifold designed to route coolant air, drive air, and water to the active dental handpiece. This manifold consists of a pneumatically actuated diaphragm mechanism and an integrated air and water distribution block. When handpieces are docked in their respective instrument holders, air pressure in the handpiece line applies force to the diaphragm within the manifold, closing the associated air ports and preventing air and water flow. When a handpiece is undocked from its holder, a valve within the holder releases the air pressure against the diaphragm. When the foot control is activated, drive and coolant air pressures overcome the diaphragm, opening the previously closed air port. The foot control activation simultaneously sends a pneumatic signal to open the corresponding water valve within the manifold assembly. As a result, the drive air, coolant air, and coolant water are able to flow through the distribution pathways to the undocked handpiece. Returning the handpiece to its holder triggers the holder's valve mechanism to re-pressurize the diaphragm, re-closing the air and water distribution pathways.

The Assistant's Delivery System is a workstation for the dental clinical space that provides air, water, and vacuum for the operation of dental instruments. The ADS-1100-001 model mounts to the head end of the Midmark Dental Chair. A junction box located at the foot end of the chair base provides a contained space for the connection of the Assistant's Delivery System to air, water, and vacuum sources. The ADS-1200-001 mounts to a Midmark cabinet base. The DS-1200-001 Assistant's Delivery System mounts on a movable arm which is common to the Dental Delivery System. The Assistant's Delivery System mounts to the lower side of its supporting arm and the Dental Delivery System mounts to the upper side of the arm allowing both to pivot independently for positioning of the instruments to the clinician. The cabinetry base has a space for the connection of the Assistant's Delivery System to air, water, and vacuum sources.

Device specifications are provided in Table 4.

Table 4. Midmark Dental Delivery System Specifications

SpecificationValue
Electrical Ratings36V 5.5A
Fuse Ratings3.15 A
Compressed Air Input80 – 100 psig
Water Input40 – 60 psig
Vacuum250 Nl/min
Drive Air Output55 Nl/min @ 45psig
Coolant Air Output1.5 Nl/min @ 35psig
Coolant Water Output50ml/min @ 18psig
DS-1100-001 length x width x height59" x 19" x 36"
ADS-1100-001 length x width x height38" x 11" x 31"
DS-1200-001 length x width x height45" x 19" x 32"

Page 8

Midmark Dental Delivery System 510(k) Summary

Table 4. Midmark Dental Delivery System Specifications (continued)

SpecificationValue
ADS-1200-001 length x width x height44" x 17" x 22"
DS-1100-001 Weight76 lbs
ADS-1100-001 Weight17 lbs
DS-1200-001 Weight98 lbs
ADS-1200-001 Weight43 lbs
Operating EnvironmentTemperature: 59 to 95°FHumidity: 10 to 90% (non-condensing)Pressure: 10 to 15.3 psi
Work Surface/Tray SizeThe worksurface is inlaid with a removeable silicone rubber mat and will accommodate a 9.75" x 13.50" tray

6. Technological Characteristics

Refer to Table 5 for a comparison of the Midmark Dental Delivery System technological characteristics with those of its predicate. Changes made to the Midmark Dental Delivery System do not raise new issues of safety or effectiveness. The Intended Use of the Midmark Dental Delivery System remains identical to that of the predicate device.

Table 5. Comparison of Technological Characteristics

CharacteristicSubject Device: Midmark Dental Delivery SystemPredicate Device: Elevance Delivery UnitsComparison
FDA 510(k) ClearanceK251626K120239N/A
Product CodeEIAEIAEquivalent
ClassClass IClass IEquivalent
Regulation872.6640872.6640Equivalent

Page 9

Midmark Dental Delivery System 510(k) Summary

Table 5. Comparison of Technological Characteristics (continued)

CharacteristicSubject Device: Midmark Dental Delivery SystemPredicate Device: Elevance Delivery UnitsComparison
Indications for UseMidmark Dental Delivery Systems are intended to provide dental professionals with low voltage electric, air, and water to operate dental handpieces, syringes, and accessories during dental examinations and procedures.Midmark Assistant's Delivery Systems are intended to provide dental professionals with air, water, and suction to operate dental handpieces, syringes, and accessories during dental examinations and procedures.The Midmark Elevance Instrument Delivery System is intended for use by professional dental practitioners in providing treatment to dental patients in a dental operatory. The system is designed to deliver air, water, vacuum and low voltage electricity to hand-held dental instruments.Similar
Midmark Dental Delivery System ComponentsDoctor's instrument head and mount arms, assistant's instrument head and mount arms, air, water, and vacuum distribution, low voltage power supplyDoctor's instrument head and mount arms, assistant's instrument head and mount arms, air, water, and vacuum distribution, low voltage power supplyEquivalent
Operatory EnvironmentDental Chair, Dental Light, Cuspidor, 12 o'clock CabinetDental Chair, Dental Light, Cuspidor, 12 o'clock CabinetEquivalent

Dental Delivery System Instrument Components and Controls:

CharacteristicSubject Device: Midmark Dental Delivery SystemPredicate Device: Elevance Delivery UnitsComparison
Delivery Unit Mount OptionsMidmark Dental Chair or 12 o'clock Column/Cabinet mounted swing armsElevance Dental Chair or 12 o'clock Cabinet mounted swing armsSimilar
Delivery Unit Height AdjustmentSpring-loaded flex armSpring-loaded flex armEquivalent
Flex Arm ReleaseNot requiredE-field sensor and pilot air switchDifferent
Delivery Unit Instrument HoldersFour or six positions are available, up to five instruments may be controlled by a pneumatic foot controlFour or six positions are available, up to five instruments may be controlled by a pneumatic foot controlEquivalent
System Master SwitchPneumatic valvePneumatic valveEquivalent

Page 10

Midmark Dental Delivery System 510(k) Summary

Table 5. Comparison of Technological Characteristics (continued)

CharacteristicSubject Device: Midmark Dental Delivery SystemPredicate Device: Elevance Delivery UnitsComparison
Instrument DetectionPneumatic pilot valve and air actuated electronic signalPneumatic pilot valve and optical sensorSimilar
Instrument Air and Water Line ActuationMidmark proprietary diaphragm moduleMidmark proprietary kink valvesDifferent
Instrument Drive Air VolumeManually adjustable pinch valvesProportional solenoid valveDifferent
Instrument Coolant Water VolumeManually adjustable needle valvesProportional solenoid valveDifferent
Instrument Coolant Air VolumeManually adjustable needle valveManually adjustable needle valveEquivalent
Handpiece Line FlushManually actuated pilot valveElectronic solenoid pilot valveDifferent
Syringe Air and Water VolumeManually adjustable pinch valvesManually adjustable needle valvesSimilar
Air and Water Shutoff for Bien Air Endodontics ModeNot availableElectronic solenoid valveDifferent
Air and Water External PortsPanel mounted couplers with automatic shutoffPanel mounted couplers with automatic shutoff (included on the assistant's unit)Equivalent
Foot ControlProvides variable drive air to engage and operate doctor's instruments; includes a wet/dry switchProvides variable drive air to engage and operate doctor's instruments; includes a wet/dry switchEquivalent
Instrument Lubrication Oil Collection SystemPlastic housing with gauzePlastic housing with gauzeEquivalent
Midmark Dental Chair and Light Remote ControlTouchpad membraneTouchpadEquivalent
Software UpdatesSoftware updates via a wired CAN message from a network connected Midmark Dental Chair.Software updates via USB on the display control PCBADifferent
USB PortPass-through USB extender and USB portPass-through USB extender and USB portEquivalent

Dental Delivery System Instrument Options (Non-Midmark Manufactured Devices):

Page 11

Midmark Dental Delivery System 510(k) Summary

Table 5. Comparison of Technological Characteristics (continued)

CharacteristicSubject Device: Midmark Dental Delivery SystemPredicate Device: Elevance Delivery UnitsComparison
Air-Driven Handpiece Connector OptionsHigh-speed or low-speed handpiece connector with or without integrated illuminationHigh-speed or low-speed handpiece connector with or without integrated illuminationEquivalent
Bien Air Electric MotorNot availableSingle or dual system including Bien Air control module. Motor settings provided via a touchpad and display screen.Different
Piezoelectric ScalerIncludes scaler control module with setting adjusted manually via potentiometer.Includes scaler control module with setting provided via a touchpad and display screen.Similar
Magneto-Strictive ScalerIncludes scaler control module with setting adjusted manually via potentiometer.Includes scaler control module with setting provided via a touchpad and display screen.Similar
Syringe options3-way air and water syringe or air only syringe; accepts autoclavable or disposable tips3-way air and water syringe or air only syringe; accepts autoclavable or disposable tipsEquivalent

Assistant's Delivery System Components and Controls:

CharacteristicSubject Device: Midmark Dental Delivery SystemPredicate Device: Elevance Delivery UnitsComparison
Assistant's Unit Mount OptionsMidmark Dental Chair or 12 o'clock Cabinet/Column mounted swing armsElevance Dental Chair or 12 o'clock Cabinet mounted swing armsSimilar
Assistant Instrument HoldersFour holders are standardFour holders are standardEquivalent
Midmark Dental Chair and Light Remote ControlTouchpad membraneTouchpadEquivalent
Vacuum Solids CollectorSolids collector built into housing, accepts disposable basketsSolids collector built into housing, accepts disposable basketsEquivalent
Syringe Water and Air ControlsManual needle valvesManual pinch valvesSimilar

Page 12

Midmark Dental Delivery System 510(k) Summary

Table 5. Comparison of Technological Characteristics (continued)

CharacteristicSubject Device: Midmark Dental Delivery SystemPredicate Device: Elevance Delivery UnitsComparison
Air-Driven Handpiece Connector OptionsHigh-speed or low-speed handpiece connector with or without integrated illuminationHigh-speed or low-speed handpiece connector with or without integrated illuminationEquivalent
Bien Air Electric MotorNot availableSingle or dual system including Bien Air control module. Motor settings provided via a touchpad and display screen.Different
Piezoelectric ScalerIncludes scaler control module with setting adjusted manually via potentiometer.Includes scaler control module with setting provided via a touchpad and display screen.Similar
Magneto-Strictive ScalerIncludes scaler control module with setting adjusted manually via potentiometer.Includes scaler control module with setting provided via a touchpad and display screen.Similar
Syringe options3-way air and water syringe or air only syringe; accepts autoclavable or disposable tips3-way air and water syringe or air only syringe; accepts autoclavable or disposable tipsEquivalent

Assistant's Delivery System Components and Controls:

CharacteristicSubject Device: Midmark Dental Delivery SystemPredicate Device: Elevance Delivery UnitsComparison
Assistant's Unit Mount OptionsMidmark Dental Chair or 12 o'clock Cabinet/Column mounted swing armsElevance Dental Chair or 12 o'clock Cabinet mounted swing armsSimilar
Assistant Instrument HoldersFour holders are standardFour holders are standardEquivalent
Midmark Dental Chair and Light Remote ControlTouchpad membraneTouchpadEquivalent
Vacuum Solids CollectorSolids collector built into housing, accepts disposable basketsSolids collector built into housing, accepts disposable basketsEquivalent
Syringe Water and Air ControlsManual needle valvesManual pinch valvesSimilar

Assistant's Delivery System Instrument Options (Non-Midmark Manufactured Devices):

CharacteristicSubject Device: Midmark Dental Delivery SystemPredicate Device: Elevance Delivery UnitsComparison
High Volume Evacuator (HVE)HVE housing with shutoff, accepts disposable tipsHVE housing with shutoff, accepts disposable tipsEquivalent
Illuminated HVEZyris Isolite illuminated HVE optionNot availableDifferent
Saliva Ejector (SE)SE housing with shutoff, accepts disposable tipsSE housing with shutoff, accepts disposable tipsEquivalent
Syringe Options3-way air/water syringe or air only syringe; accepts autoclavable or disposable tips3-way air/water syringe or air only syringe; accepts autoclavable or disposable tipsEquivalent

Air/Water/Electric Supply Components:

CharacteristicSubject Device: Midmark Dental Delivery SystemPredicate Device: Elevance Delivery UnitsComparison
Water DistributionSelf-contained water bottle system or city water regulator and shutoff valve.Self-contained water bottle system or city water regulator and shutoff valve.Equivalent
Air DistributionAir regulator and shutoff valveAir regulator and shutoff valveEquivalent
Power Supply36Vdc Low voltage output with 120Vac input24Vac Low voltage output with 120Vac or 240Vac inputSimilar

7. Non-Clinical Test Data

Non-clinical testing and evaluations were performed for the Midmark Dental Delivery System. The non-clinical tests demonstrated that the Midmark Dental Delivery System meets the applicable requirements of the referenced standards, and the results support that the device performs as safe and as effective as the legally marketed predicate device for its intended use.

Electrical Safety and Electromagnetic Compatibility:

The Midmark Dental Delivery System was evaluated to IEC 60601-1, ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 80601-2-60, and IEC TS 60601-4-2 to ensure compliance with electrical safety standards.

Performance Testing:

The Midmark Dental Delivery System was evaluated to ISO 7494-1, ISO 7494-2, and ISO 9168 to verify performance of the device.

Page 13

Midmark Dental Delivery System 510(k) Summary

Reprocessing Testing:

The Midmark Dental Delivery System had reprocessing validations conducted in accordance with the FDA guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling", issued March 17, 2015. Additional verification testing was completed on the dental unit water lines according to ISO 16954.

Software Verification:

The software of the Midmark Dental Delivery System was verified in accordance with the FDA guidance "Content of Premarket Submissions for Device Software Functions", issued June 14, 2023.

Biocompatibility:

Biocompatibility testing was completed for the Midmark Dental Delivery System utilizing ISO 10993-1, ISO 10993-5, ISO 10993-10, and ISO 10993-23.

A summary of evaluations are shown in Table 6. Based on the results of studies performed, the Midmark Dental Delivery System does not raise new issues of safety or effectiveness.

Table 6. Non-Clinical Performance Data

StandardStandard Name
IEC 60601-1-2:2020 Ed. 4.1Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
ANSI/AAMI ES60601-1:2005+A2:2010(R2012)+A1:2012+ A2:2021Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]
IEC 60601-1:2005/A1:2012/A2:2020Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 80601-2-60:2019Medical electrical equipment - Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment
IEC TS 60601-4-2:2024Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
ISO 7494-1:2018Dentistry - Stationary dental units and dental patient chairs - Part 1: General requirements

Page 14

Midmark Dental Delivery System 510(k) Summary

Table 6. Non-Clinical Performance Data (continued)

StandardStandard Name
ISO 7494-2:2022Dentistry - Stationary dental units and dental patient chairs - Part 2: Air, water, suction and wastewater systems
ISO 9168:2009Dentistry - Hose connectors for air driven dental handpieces
ISO 10993-1:2018Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
ISO 10993-5:2009Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2021Biological evaluation of medical devices - Part 10: Tests for skin sensitization
ISO 10993-23:2021Biological evaluation of medical devices - Part 23: Tests for irritation
ISO 16954:2015Dentistry — Test methods for dental unit waterline biofilm treatment

8. Clinical Performance Data

Clinical data was not required.

9. Conclusion

In terms of Intended Use, design, material, principles of operation, energy source, and performance in nonclinical testing, Midmark has determined that the subject Midmark Dental Delivery System is substantially equivalent and is as safe and effective and performs as well as or better than the legally marketed predicate device, the Elevance Delivery Unit (K120239).

§ 872.6640 Dental operative unit and accessories.

(a)
Identification. A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.(b)
Classification. Class I (general controls). Except for dental operative unit, accessories are exempt from premarket notification procedures in subpart E of part 807 of this chapter subject to § 872.9.