Search Filters

Search Results

Found 18 results

510(k) Data Aggregation

    K Number
    K232781

    Validate with FDA (Live)

    Date Cleared
    2024-03-07

    (178 days)

    Product Code
    Regulation Number
    862.1675
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Safety Sliding Blood Collection Set is intention into a patient's vascular system for blood collection or short term (up to 2 hours) IV administration, by a trained medical professional. It aids in the prevention of accidental needle stick through the use of an active safety feature.

    Device Description

    Components of the subject device include: (1) Needle cap, (2) Butterfly wings, (3) Safety shield, (4) Needle, (5) Flexible tube, (6) Female luer (adapter), (7) Male luer hub, (8) Puncture needle, (9) Rubber sleeve, (10) Puncture needle protect cover, (11) Holder, (12) Protective cap.

    The subject device has three configurations: with Holder, with Holder, without Male Adapter. Each configuration has all of the (1) - (6) components mentioned above. The difference between the three configurations is that they have only one or more of the (7) – (12) components mentioned above. The proposed device is available various needle size and flexible tube length. The range of needle size is from 25G (0.5mm) to 20G (0.9mm). The range of flexible tube length is from 10cm (4") to 30cm (12").

    The subject device has sharps injury protection features. It aids in the prevention of accidental needle stick through the use of an active safety feature.

    The subject device is a sterile, single use device. It is sterilized by Ethylene Oxide Gas (EtO) to achieve a SAL of 10-6 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of three years.

    No DEHP, BPA and Natural Rubber Latex are added in the proposed device.

    AI/ML Overview

    The provided document describes the Jiangsu Caina Medical Co., Ltd.'s "Safety Sliding Blood Collection Set" (K232781) and its substantial equivalence to a predicate device (K980414), but it does not contain acceptance criteria for device performance or a study demonstrating the device meets such criteria.

    The document primarily focuses on:

    • Regulatory Clearance: Announcing FDA clearance (510(k)) based on substantial equivalence.
    • Indications for Use: Describing the intended purpose of the device.
    • Device Description: Outlining the components and configurations.
    • Comparison to Predicate Device: Detailing similarities and differences between the subject device and the predicate.
    • Non-Clinical Testing: Listing the international standards and specific tests conducted (e.g., biocompatibility, sterility, seal integrity, sharps injury protection) to support the substantial equivalence claim.

    Therefore, it is not possible to provide the requested information regarding acceptance criteria and a study proving their fulfillment based on this document.

    The document explicitly states:

    • 10. Clinical Test Conclusion: "Clinical studies are not required to demonstrate substantial equivalence to the predicate device." This indicates that the regulatory clearance was based on non-clinical data and comparison to an existing device, rather than a clinical trial with specific performance acceptance criteria.

    The "Non-Clinical Testing" section lists various tests performed (e.g., ISO, ASTM standards, extraction force, simulated clinical study for safety feature, sharps injury protection testing). These tests would have their own internal pass/fail criteria as defined by the respective standards or the manufacturer's internal specifications. However, these specific criteria and the detailed results demonstrating achievement are not provided in this publicly available 510(k) summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233190

    Validate with FDA (Live)

    Device Name
    ENFit Adaptor
    Date Cleared
    2024-01-18

    (112 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ENFit Adaptor is intended for connecting an enteral giving set with an ENFit connector to an enteral catheter with funnel.

    Device Description

    To facilitate the interconnection between newly-released feeding sets, which are equipped with female ENFit connectors, and the funnel feeding ports of previously-designed enteral feeding tubes, the proposed device is designed with the mating component of an ENFit female connector, designed and tested in accordance with ISO 80369-3 and ISO 80369-20, at one of its ends, and the mating component of a non-ISO 80369-1 compliant funnel feeding port, a stepped tip connector, at its other end. It can prevent an interruption or delay in patient care by allowing fitment of enteral feeding sets with ISO 80369-3 compliant ENFit female connectors to existing enteral feeding tubes with non-ISO 80369-1 compliant end connectors, referred to as "funnel feeding ports" which are still available for use in healthcare setting.

    The proposed device is sterile or non-sterile device is sterilized by ethylene oxide to achieve a SAL 10- and supplied in sterility maintenance package which maintain the sterility of the device during the shelf life 5 years.

    No DEHP, BPA and Natural Rubber Latex are added in the device.

    AI/ML Overview

    The provided text is a 510(k) summary for the ENFit Adaptor, a medical device. It focuses on demonstrating substantial equivalence to a predicate device rather than detailing a study that proves the device meets specific acceptance criteria in the context of an AI/ML medical device. Therefore, I cannot extract the requested information regarding acceptance criteria, device performance, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, training set details, or ground truth establishment relevant to an AI/ML device from this document.

    The document discusses non-clinical tests to verify design specifications and compliance with various ISO standards for mechanical performance, biocompatibility, sterilization, shipping, and shelf-life of the physical adaptor. These tests are performed to demonstrate that the device itself is safe and effective as a physical connector, not as an AI-powered diagnostic or therapeutic tool.

    For instance, the document mentions:

    • Performance tests: Fluid Leakage, Stress Cracking, Resistance to Separation from Axial Load, Resistance to Separation from Unscrewing, Resistance to Overriding, Disconnection by Unscrewing, Connector Incompatibility Test, Materials Used for Small-bore Connectors, Dimensional analysis, Compatibility and Usability Test.
    • Biocompatibility tests: Cytotoxicity, Irritation, Skin Sensitization.
    • Sterilization, Shipping and Shelf-Life tests: Ethylene oxide sterilization residuals, seal leak detection, seal strength, seal integrity, Bacterial Endotoxins Test.

    These are all physical and chemical property tests for a medical device connector, not performance metrics of an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230635

    Validate with FDA (Live)

    Device Name
    Pen Needle
    Date Cleared
    2023-06-02

    (87 days)

    Product Code
    Regulation Number
    880.5570
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Pen Needle is intended for use with pen injector devices for the subcutaneous injection of drugs.

    Device Description

    The proposed device, Pen Needle, is designed for use with pen injectors for subcutaneous injection of a desired dose of drugs approved for delivery using a pen needle. It is provided sterile, single use. The proposed device consists of four components, which are 1) Needle Container 2) Needle Shield 3) Needle Tube and 4) Needle Hub. Needle Container, together with Sealed Paper can forming the unit packaging of Pen Needle, that maintains the sterility of Pen Needle Hub can be connected screwed onto the pen injectors.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device (Pen Needle) and focuses on demonstrating substantial equivalence to a predicate device. It is not an AI/ML medical device submission. Therefore, the document does not contain information typically found in an AI/ML medical device acceptance criteria and study section, such as:

    • A table of acceptance criteria and reported device performance for an AI/ML algorithm.
    • Sample sizes used for a test set or data provenance for an AI/ML algorithm.
    • Number of experts or qualifications of experts used for AI ground truth.
    • Adjudication methods for AI test sets.
    • Multi-Reader Multi-Case (MRMC) comparative effectiveness studies.
    • Standalone AI performance.
    • Type of ground truth used for AI/ML.
    • Sample size or ground truth establishment for an AI/ML training set.

    The document discusses non-clinical tests performed to verify the Pen Needle met design specifications and complied with ISO standards, focusing on manufacturing quality, sterility, and biocompatibility rather than an AI algorithm's performance.

    Therefore, I cannot extract the requested information as it is not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221406

    Validate with FDA (Live)

    Date Cleared
    2022-10-28

    (165 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vented Vial Transfer Pin is intended for the transfer and mixing of drugs contained in a vial.

    Device Description

    The proposed device consists of five components: (1) Protective cap, (2) Piercing spike, (3) Filter medium, (4) Filter shell, (5) Luer connector cap. The piercing spike contains the dual lumen channel for liquid and air. A 0.2um hydrophobic air filter medium is assembled to the end of air channel. This enables keeping an equilibrium pressure between the drug vial and the ambient pressure, filtering the inserted/released air through filter medium. There is a female Luer lock connector in the end of liquid channel, it can be attached a device with a male Luer connector (e.g., standard syringe). The proposed device is available 2 specifications according to design, with security clip and without security clip. The Vented Vial Transfer Pin with Security Clip designed which can be fixed to a 13mm drug vial to prevent accidental separation. The proposed device is a sterile, single use device. It is sterilized by Ethylene Oxide Gas (EtO) to achieve a SAL of 106 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of five years. No DEHP, BPA and Natural Rubber Latex are added in the proposed device.

    AI/ML Overview

    This document is a 510(k) Summary for a medical device called the "Vented Vial Transfer Pin." It mostly focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study proving performance against acceptance criteria for the new device as an AI or diagnostic tool.

    However, it does list several non-clinical tests that were performed to verify that the proposed device met design specifications and was substantially equivalent to the predicate device. These tests, along with comments on the differences between the proposed and predicate devices, indirectly describe acceptance criteria and how the device met them.

    Here's a breakdown of the information requested, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria alongside reported performance in the general sense of a diagnostic or AI device. Instead, it lists standards with which non-clinical tests complied and states that the proposed device "met all design specifications" and "complies with the following standards." The comments section addresses differences between the proposed and predicate device and assures that these differences do not affect substantial equivalence on safety and effectiveness.

    Here's an interpretation based on the provided text, focusing on the non-clinical tests:

    Acceptance Criteria (Inferred from Standards & Test Objectives)Reported Device Performance (Compliance)
    Biocompatibility
    No cytotoxicity (per ISO 10993-5:2009)Achieved: "no cytotoxicity per ISO 10993-5"
    No irritation (per ISO 10993-10:2010)Achieved: "no irritation per ISO 10993-10"
    No sensitization (per ISO 10993-10:2010)Achieved: "no sensitization per ISO 10993-10"
    No acute systemic toxicity (per ISO 10993-11:2017)Achieved: "no acute systemic per ISO 10993-11"
    No pyrogenicity (per USP <151>)Achieved: "no pyrogen per USP"
    No hemolysis (per ASTM F756-17)Achieved: "no hemolysis per ASTM F756"
    Sterility & Shelf Life
    Sterility Assurance Level (SAL) of 10-6Achieved: "Sterility Assurance Level (SAL) of 10-6"
    Ethylene Oxide Sterilization Residuals within limits (per ISO 10993-7:2008 AMD.1:2019)Achieved: "Examination of the Ethylene Oxide (EO) and Ethylene Chlorohydrin (ECH) residuals have met with ISO 10993-7AMD1:2019"
    Maintenance of sterility for 5 years (packaging integrity per ISO 11607 and ISTA 3A)Achieved: "demonstrated that the aged samples also complied with the requirements of ISO 8536-4 and ISO 80369-7. The ability of immediate package of the proposed device to maintain the device in a sterile state for a period of 5 years has been validated in accordance with ISO 11607 and ISTA 3A."
    Packaging Integrity
    No seal leaks (per ASTM F1929-15)Not explicitly stated "achieved," but compliance with standard for packaging integrity is implied by being listed.
    Seal strength (per ASTM F88/F88M-15)Not explicitly stated "achieved," but compliance with standard for packaging integrity is implied.
    Seal integrity (visual inspection per ASTM F1886/F1886M-16)Not explicitly stated "achieved," but compliance with standard for packaging integrity is implied.
    Bacterial Endotoxins
    Within limits (per USP <85>)Not explicitly stated "achieved," but compliance with standard is implied by being listed.
    Performance Testing
    Product Functionality Performance (ISO 8536-4, 7.4 and ISO 80369-7)Achieved: "Non-clinical tests were conducted to verify that the proposed device met all design specifications" and "complies with the following standards".
    Filter Bursting Pressure (ISO 8536-4, 7.5)Achieved generally as part of "met all design specifications" and compliance.
    Flow Rate Testing (ISO 8536-4, 7.10)Achieved generally as part of "met all design specifications" and compliance.
    Cap/valve-housing detachment from body Test (ISO 8536-4, 7.3)Achieved generally as part of "met all design specifications" and compliance.
    Internal stress level for assembled product (ISO 8536-4, 7.13)Achieved generally as part of "met all design specifications" and compliance.
    Small-bore connectors compliance (ISO 80369-7:2021)Achieved: "demonstrated that the aged samples also complied with the requirements of ISO 8536-4 and ISO 80369-7"
    Mechanical Strength (Cap Assembly)
    Tensile strength of parts 15N for 15sAchieved: "The tensile strength of parts have been tested at 15N for 15s. Both of them meet acceptable criteria."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document relates to a physical medical device (Vented Vial Transfer Pin), not an AI/diagnostic software. Therefore, there is no "test set" in the context of data for an algorithm. The non-clinical tests would have involved physical samples of the device. The document does not specify the number of units tested for each non-clinical test.

    Data provenance is not applicable here as it refers to typical AI/diagnostic data (e.g., patient images, medical records).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This device is a physical medical device, not an AI or diagnostic tool that requires ground truth established by medical experts for performance evaluation. The "ground truth" for this device's performance would be derived from the physical and chemical properties and test outcomes according to the cited standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not an AI/diagnostic software study involving human readers or adjudication of results.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For this physical device, the "ground truth" is established by:

    • Compliance with international and national standards (e.g., ISO, ASTM, USP) for materials, sterility, packaging, and functional performance.
    • Material properties and test results from laboratory analyses (e.g., cytotoxicity, hemolytic properties, tensile strength).
    • Physical measurements and functional tests directly assessing the device's ability to perform its intended function (e.g., flow rate, filter bursting pressure).

    8. The sample size for the training set

    Not applicable. This is a physical device, not an AI/ML algorithm.

    9. How the ground truth for the training set was established

    Not applicable. This is a physical device, not an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222155

    Validate with FDA (Live)

    Date Cleared
    2022-09-16

    (58 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Single use O-ring gasket syringe with ENFit connector (provided sterile and non-sterile):
    The device is indicated for use as a dispenser, a measuring device and a fluid transfer device. It is used to deliver fluids into the body orally or enterally. It is intended to be used in clinical settings by users ranging from clinicians to laypersons in all age groups.

    Reusable O-ring gasket syringe with ENFit connector-(provided non-sterile):
    The device is indicated for use as a dispenser, a measuring device. It is used to deliver fluids into the body orally or enterally. It is intended to be used multiple times in non-clinical settings by users ranging from clinicians to laypersons in all age groups. The device is indicated for single patient use only.

    Device Description

    The proposed device is manual feeding syringe, used as a dispenser, a measuring device, and a fluid transfer device. And it is used to deliver fluids into the body orally or enterally in clinical and non-clinical settings.

    There are 2 types of connector for proposed device: Low dose tip ENFit connector(1ml to 6ml) and Standard ENFit connector(12ml to 60ml). The proposed device is standard piston syringe consisting of a plastic hollow barrel with female ENFit connector of ISO 80369-3, plunger, 0-ring piston, Connector of 1ml and 3ml syringe situated centrally, connector of other syringe situated eccentrically. All connectors are compatible only with enteral access or accessories having male ENFit connector of ISO 80369-3. The proposed syringes are available in transparent barrel and amber barrel.

    The proposed device is provided sterile or non-sterile, single use or reusable. The sterile device is sterilized by Ethylene Oxide Gas (EtO) to achieve a SAL of 10-6 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life. The reusable device is for single patient use only.

    The proposed device is made with polypropylene and Silica gel. No DEHP, BPA and Natural Rubber Latex are added in the proposed device. The proposed device also has an optional syringe tip cap.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for a medical device (O-ring gasket syringe with ENFit connector). This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than proving a device meets specific acceptance criteria through a large-scale clinical study or AI performance evaluation.

    Therefore, the information typically requested in your prompt regarding AI/algorithm performance, ground truth establishment for training and test sets, multi-reader multi-case (MRMC) studies, expert opinions, and sample sizes for AI model development is not present in this document.

    The document demonstrates compliance with performance standards for a physical medical device. Here's what can be extracted and inferred based on the document's content:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't present a table of "acceptance criteria" in a quantitative sense as might be found for algorithm performance (e.g., minimum accuracy). Instead, it lists the standards the device was tested against and states that the test results "demonstrated that the proposed device complies with the following standards." Compliance with these standards is the acceptance criterion for a physical device in this context.

    Acceptance Criteria (Standards Complied With)Reported Device Performance
    ISO 10993-7:2008 Biological evaluation of medical devices Part 7: Ethylene oxide sterilization residuals, AMENDMENT 1: Applicability of allowable limits for neonates and infants (2019)Complies
    ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Test for in vitro cytotoxicityComplies
    ISO 10993-6:2016 Biological evaluation of medical devices -- Part 6: Tests for local effects after implantationComplies
    ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationComplies
    ISO 10993-11:2017 Biological evaluation of medical devices -- Part 11: Tests for systemic toxicityComplies
    USP-NF <151> Pyrogen TestComplies
    ASTM F1929-15 Standard test method for detecting seal leaks in porous medical packaging by dye penetration.Complies
    ASTM F88/F88M-21 standard method for seal strength of flexible barrier materialsComplies
    ASTM F1886/F1886M-16 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual InspectionComplies
    USP-NF <85> Bacterial Endotoxins TestComplies
    ISO 7886-1:2017 Sterile hypodermic syringes for single use - Part 1: Syringes for manual useComplies
    ISO 80369-3:2016 Small-bore connectors for liquids and gases in healthcare applications - Part 3: Connectors for enteral applicationsComplies
    ISO 80369-20:2015 Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methodsComplies
    Microbial Enumeration and Specified Microorganisms Tests reportComplies
    Cleaning instructions verificationComplies
    Re-use cycle parameters studyComplies
    Microbial study in simulation usingComplies
    Risk assessment in accordance to ISO 14971Complies
    Dosing Accuracy TestingComplies
    Additional Acceptance Criteria for Reusable Devices (implied):
    Ability of immediate package to maintain sterility for 5 yearsValidated (in accordance with ASTM F1886/F1886M-16, ASTM F88/F88M-21, ASTM F1929-15)
    Performance after 5 years accelerated agingComplies with ISO7886-1 and ISO 80369-3

    2. Sample sizes used for the test set and the data provenance:

    • Sample Size: Not explicitly stated for each test. These would be determined by the specific requirements of each ISO/ASTM standard (e.g., how many units are tested for seal integrity or biocompatibility).
    • Data Provenance: The tests were "Non-clinical tests" conducted to verify compliance with international standards. The country of origin for the testing itself is not specified, but the applicant (Jiangsu Caina Medical Co., Ltd.) is from China. These are laboratory-based compliance tests, not data collected from human subjects.
    • Retrospective or Prospective: Not applicable, as these are technical performance tests, not clinical data collection.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable. The "ground truth" for the device's physical performance is established by the definitions and methodologies within the cited international standards (ISO, ASTM, USP-NF). The tests are designed to objectively measure properties and assess compliance according to these universally accepted procedures, not through expert consensus in the way a medical image diagnosis might be.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. Adjudication methods are typically used for subjective assessments (like medical image interpretation). For compliance testing against engineering and biological standards, the results are typically quantitative measurements that either meet or fail a predefined specification.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This type of study (MRMC, human readers, AI assistance) is relevant for AI-powered diagnostic devices. The device in question is a physical syringe, not an AI diagnostic tool.
    • The document explicitly states: "No clinical study is included in this submission."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For this device, the "ground truth" for proving performance is adherence to established international and national engineering and biological standards. This involves:
      • Validated test methods: e.g., the specific procedures for pyrogen testing (USP-NF <151>), seal integrity (ASTM F88/F88M-21), or material biocompatibility (ISO 10993 series).
      • Defined acceptance limits/specifications: e.g., allowable residuals of ethylene oxide, specified ranges for dosing accuracy, or criteria for passing cytotoxicity tests.

    8. The sample size for the training set:

    • Not applicable. This is not an AI model requiring a training set.

    9. How the ground truth for the training set was established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K220184

    Validate with FDA (Live)

    Date Cleared
    2022-07-21

    (178 days)

    Product Code
    Regulation Number
    880.5860
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pump Alignment Syringe is intended for use by healthcare professionals for injection of fluids for medical purposes into patients of all ages, including neonates in a controlled manner with compatible infusion pumps. The Pump Alignment Syringe can also be used for manual injection of fluids for medical purposes.

    Device Description

    The Pump Alignment Syringe is available in various specifications, including 3ml, 5ml, 10ml, 30ml and 60ml. The proposed device consists of three components: (1) plunger, (2) piston, (3) barrel. The connector of barrel have two types of luer-lock and luer-slip. The proposed device is provided sterile. The product is sterilized by Ethylene Oxide Gas to achieve a SAL of 10-6 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of five years.

    AI/ML Overview

    The provided text describes the submission of a 510(k) premarket notification for a Pump Alignment Syringe (K220184) and its comparison to a predicate device (K980987). However, it does not contain information regarding traditional acceptance criteria, performance metrics, or study details (like sample size, ground truth establishment, expert qualifications, adjudication methods, or MRMC studies) typically associated with AI/ML-based medical device studies.

    The document focuses on demonstrating substantial equivalence based on non-clinical tests and a comparison of technical characteristics. The "performance testing" mentioned refers to compliance with ISO standards for syringes, not a study evaluating specific performance metrics against acceptance criteria in the context of AI/ML.

    Here's an analysis based on the information provided in the document:

    1. A table of acceptance criteria and the reported device performance:

    The document does not explicitly state "acceptance criteria" and "reported device performance" in the context of AI/ML. Instead, it lists standards that the device conforms to. For the purpose of this request, I will interpret "reported device performance" as the device's conformity to these standards.

    Acceptance Criteria (Standard Compliance)Reported Device Performance
    Biocompatibility
    ISO 10993-7:2008 (Ethylene Oxide Sterilization Residuals)Complies
    ISO 10993-5:2009 (Test for in vitro cytotoxicity)Complies
    ISO 10993-10:2010 (Test for irritation and delayed-type hypersensitivity)Complies
    ISO 10993-11:2017 (Tests for systemic toxicity)Complies
    USP <151> (Pyrogen Test)Complies
    ASTM F756-17 (Assessment of Hemolytic Properties of Materials)Complies
    ISO 10993-4:2017 (Selection of tests for interactions with blood)Complies
    Packaging/Sterility
    ASTM F1929-15 (Detecting seal leaks in porous medical packaging by dye penetration)Complies
    ASTM F88/F88M-15 (Seal strength of flexible barrier materials)Complies
    ASTM F1886/F1886M-16 (Determining Integrity of Seals for Flexible Packaging by Visual Inspection)Complies
    USP <85> (Bacterial Endotoxins Test)Complies
    Physical and Functional Performance
    USP <788> (Particulate Matter in Injections)Complies
    ISO 7886-1:2017 (Sterile hypodermic syringe for single use - Part 1: Syringes for manual use)Complies
    ISO 7886-2:2020 (Sterile hypodermic syringe for single use - Part 2: Syringes for use with power-driven syringe pumps)Complies
    ISO 80369-7:2021 (Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications)Complies
    Sterility Assurance Level (SAL)
    10-6Achieved 10-6

    Study Proving Device Meets Acceptance Criteria:

    The study that proves the device meets the acceptance criteria consists of non-clinical tests conducted to verify that the proposed device met all design specifications and was substantially equivalent to the predicate device. These tests assessed biocompatibility, packaging integrity, sterility, and physical/functional performance according to the listed standards.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    This information is not provided in the document. The document primarily reports adherence to engineering and biological safety standards, not a clinical or AI/ML performance study involving test sets of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not applicable/not provided. The device is a physical medical device (syringe), and its performance evaluation involves compliance with established engineering and biological standards, not interpretation by experts to establish a "ground truth" on a test set (as would be typical for an AI/ML diagnostic or prognostic device).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not applicable/not provided. Adjudication methods are relevant for studies where expert disagreement needs a resolution mechanism, which is not the case for the type of testing described (compliance with physical and biological standards).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI/ML devices that assist human interpretation, which is not the function of this Pump Alignment Syringe.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    No, a standalone algorithm-only performance assessment was not done. The device is a physical syringe and does not involve an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for this device is its adherence to established international and national standards (e.g., ISO, ASTM, USP) for materials, sterility, package integrity, and functional performance. It is not based on expert consensus, pathology, or outcomes data in the traditional sense of a clinical diagnostic study.

    8. The sample size for the training set:

    This information is not applicable/not provided. The device is not an AI/ML product developed using training data. Its development involves engineering design and manufacturing processes, followed by testing against standards.

    9. How the ground truth for the training set was established:
    This information is not applicable/not provided as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220060

    Validate with FDA (Live)

    Date Cleared
    2022-06-10

    (151 days)

    Product Code
    Regulation Number
    880.5200
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pre-filled Syringe with Saline is intended for use in flushing compatible intravenous administration sets and indwelling intravenous access devices. Use according to the recommendations of the manufacturer for the appropriate device.

    Device Description

    Not Found

    AI/ML Overview

    This document is an FDA 510(k) clearance letter for a "Pre-filled Syringe with Saline." It does not contain any information about acceptance criteria or a study proving device performance as would be relevant for an AI/ML or diagnostic device.

    The provided text deals with:

    • The FDA's determination of substantial equivalence for a medical device (a pre-filled saline syringe).
    • Regulatory classifications and requirements for medical devices.
    • Indications for Use for the specific device.

    Therefore, I cannot provide the requested information about acceptance criteria and a study proving device performance because the document does not discuss these aspects for an AI/ML or diagnostic device. The device in question is a simple medical device (a saline syringe), not one that would typically undergo the kind of performance studies involving AI, ground truth, expert readers, or MRMC studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211593

    Validate with FDA (Live)

    Date Cleared
    2021-09-30

    (129 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Enteral Pump Syringe delivers nutritional formula to the gastrointestinal system of a patient who is physically unable to eat and swallow.

    Enteral Pump Syringe is intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to health care professionals, to administer nutritional formula. In particular, these syringes can be connected with enteral infusion pump.

    Enteral Pump Syringe is intended for neonatal, pediatric and adult use.

    Enteral Pump Syringe is single use device.

    Device Description

    The proposed device is modified device of K190502(ENFit Oral/Enteral Syringe,cleared July 2, 2019) according to 21CFR 820.30. The change is a operation mode change from manual use to use with infusion pump.

    The proposed device is a disposable enteral feeding syringe provided in a variety of sizes from 1ml to 60ml. The ENFit enteral feeding syringe consists of moveable plunger, piston, calibrated hollow barrel and tip cap(not necessary). At the distal end of the barrel is a female connector, designed according to ISO 80369-3 (often referred to as an ENFit connection). The tip is designed to mate only with enteral administration devices and is incompatible with Luer connectors. The device is used to deliver fluids into the body enterally by connected to an enteral access device with male ENFit connector. The syringes (size 6ml to 60ml) incorporate a female Standard ENFit connector; the syringes (size 1ml and 6ml) incorporate a female low dosing ENFit connector.

    There are 2 types of ENFit connector ENFit syringe and central connector ENFit syringe. The sizes of the Side connector ENFit syringe range from 12ml to 60ml; and The sizes of the central connector ENFit syringe range from 1ml to 60ml. The syringes are provided with transparent or amber color barrel.

    The material used of deviec are Polypropylene. Polyisoprene rubber, Polydimethysiloxane(inside surface coating of barrel ). Amber additive(for used amber color barrel only), which are also patient-contacting materials for the proposed devices. The time of contact duration is less than 24 hours. The contact level is mucosal membrane, indirect.

    The proposed device is sterile or non-sterile, single use. The sterilized by Ethylene Oxide Gas to achieve a SAL of 10-6 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of five years.

    The proposed device intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to health care professionals. The device has no energy in itself. But the syringe works with infusion pump and extension sets.The syringe is aspirated nutritional formula liquid, connected the extension set. Then user installed it on the infusion pump, and pump fluids into a body enterally in a controlled manner for administer enteral nutrition. The accuracy of graduations for enteral syringes is +/- 4% of expelled volume.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for an "Enteral Pump Syringe." It focuses on demonstrating substantial equivalence to a predicate device, rather than proving a medical device meets specific performance acceptance criteria through the kind of study described in the prompt. This device is a syringe, not an AI or imaging device, so the requested information regarding AI performance metrics, expert adjudication, MRMC studies, and ground truth establishment is not applicable.

    Therefore, I cannot provide the requested information for an AI/imaging device. However, I can extract the information relevant to the performance and testing of this specific medical device, the Enteral Pump Syringe, as presented in the document:

    Device: Enteral Pump Syringe

    Context: This document is an FDA 510(k) submission, aiming to demonstrate "substantial equivalence" of the proposed device to an already legally marketed predicate device. It is not a clinical study report for an AI or imaging device.


    Information that can be extracted from the provided text regarding this device's performance and testing:

    1. A table of acceptance criteria and the reported device performance:

    The document primarily refers to compliance with established international standards rather than specific quantitative acceptance criteria and reported performance values in a table format. The "acceptance criteria" here are the requirements of these standards.

    Standard / CharacteristicAcceptance Criteria (implied by standard compliance)Reported Device Performance (as stated in document)
    ISO 80369-3:2016 (Small-Bore Connectors for Enteral Applications)Compliance with specifications for enteral connectors"The proposed device complies with the following standards: ISO 80369-3:2016" and "These product models were tested according to ISO 80369-3."
    ISO 80369-20:2015 (Common Test Methods for Small-Bore Connectors)Compliance with common test methods"The proposed device complies with the following standards: ISO 80369-20:2015"
    ISO 7886-1:2017 (Syringes for Manual Use)Compliance with specifications for sterile hypodermic syringes for manual use"The proposed device complies with the following standards: ISO 7886-1:2017" and "These product models were tested according to ISO 7886-1."
    ISO 7886-2:2020 (Syringes for use with power-driven syringe pumps)Compliance with specifications for sterile hypodermic syringes for use with power-driven syringe pumps"The proposed device complies with the following standards: ISO 7886-2:2020" and "These product models were tested according to ISO 7886-2."
    ISO 10993-5:2009 (Biocompatibility - in Vitro Cytotoxicity)No cytotoxicity"Cytotoxicity: No Cytotoxicity"
    ISO 10993-10:2010 (Biocompatibility - Irritation and Skin Sensitization)No irritation, no sensitization"Skin Irritation: No Irritation," "Sensitization: No Sensitization"
    ISO 10993-7:2008 (Biocompatibility - Ethylene oxide sterilization residuals)Compliance with residuals limits"ISO 10993-7:2008" listed as a complied standard.
    ISO 14971:2007 (Medical Devices - Application of Risk Management)Application of risk management principles"ISO 14971:2007" listed as a complied standard.
    ASTM F88/F88M-15 (Seal Strength of Flexible Barrier Materials)Compliance with seal strength requirements"ASTM F88/F88M-15" listed as a complied standard.
    ASTM F1929-15 (Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration)Compliance with seal leak detection"ASTM F1929-15" listed as a complied standard.
    ASTM F1886F1886M-16 (Determining Integrity of Seals for Flexible Packaging by Visual Inspection)Compliance with visual inspection for seal integrity"ASTM F1886F1886M-16" listed as a complied standard.
    Graduation Accuracy (for enteral syringes)+/- 4% of expelled volume"The accuracy of graduations for enteral syringes is +/- 4% of expelled volume." (This is stated as a characteristic, implying it meets this.)
    SAL (Sterility Assurance Level)10^-6"EO Sterilized to achieve a SAL of 10^-6"
    Shelf Life5 years"maintain the sterility of the device during the shelf life of five years."

    2. Sample sized used for the test set and the data provenance:

    • Sample Size: Not explicitly stated in the provided text. The document refers to "non-clinical tests" and "bench tests" to verify compliance with standards, but not the number of units tested.
    • Data Provenance: Not specified, but likely from the manufacturer's internal testing facilities in China (Jiangsu Caina Medical Co., Ltd.). The tests are "non-clinical" (bench tests) and therefore not from patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable as this is a medical device (syringe) and not an AI or imaging device that requires expert-established ground truth from images or clinical data. The "ground truth" for this device's performance is adherence to engineering and biocompatibility standards measured through bench tests.

    4. Adjudication method for the test set:

    Not applicable. As above, this is not an AI/imaging device study requiring human reader adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is not an AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device does not involve an algorithm.

    7. The type of ground truth used:

    The "ground truth" for this device's performance is established by the international and national standards it is required to meet (e.g., ISO and ASTM standards for sterility, biocompatibility, connector compatibility, and physical performance). Compliance is shown through bench tests and material characterization, not expert consensus, pathology, or outcomes data in the traditional sense for diagnostic or prognostic devices.

    8. The sample size for the training set:

    Not applicable. This device does not involve machine learning or a "training set."

    9. How the ground truth for the training set was established:

    Not applicable. This device does not involve machine learning.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200463

    Validate with FDA (Live)

    Date Cleared
    2021-04-07

    (407 days)

    Product Code
    Regulation Number
    880.5570
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Huber Needle Infusion Set: The Huber Needle Infusion Set is a device with a non-coring right angle needle intended for insertion into the septum of a subcutaneously implanted port and for the infusion of fluids and drugs, as well as blood sampling into the port. The 19G-22G needles of device are also suitable for power injection of contrast media to a maximum pressure of 325psi. When used with ports indicated for power injection, the maximum recommended infusion rate is approximately 5ml/s for 19G and 20G, 2ml/s for 22G.

    Safety Huber Needle Infusion Set: The Safety Huber Needle Infusion Set is a device with a non-coring right angle needle intended for insertion into the septum of a subcutaneously implanted port and for the infusion of fluids and drugs, as well as blood sampling into the port. The safety feature is manually activated during needle removal, and in the prevention of accidental needle sticks. The 19G-22G needles of device are also suitable for power injection of contrast media to a maximum pressure of 325psi. When used with ports indicated for power injection, the maximum recommended infusion rate is approximately 5ml/s for 19G and 20G, 2ml/s for 22G.

    Device Description

    The Huber Needle Infusion Set and Safety Huber Needle Infusion Set non-coring angled needle are single use, sterile and non-pyrogenic device intended for insertion into the septum of a subcutaneously implanted port for the infusion of fluids and drugs, as well as blood sampling through the port. The Huber Needle Infusion Set and Safety Huber Needle Infusion Set are available in different gauge sizes (25G, 24G, 22G, 20G, 19G) and length (0.5", 0.75", 1.0", 1.24", 1.5"). The 19G-22G needles of device are suitable for power injection of contrast media into the central venous system only through an implanted port that is indicated for power injection at a maximum pressure of 325psi. For power injection of contrast media, the maximum recommended infusion rate is approximately 5ml/s for 19G and 20G and 2ml/s for 22G.The 24G and 25G needles of device should not be used with power injectors. The Huber Needle Infusion Set is available in five models (S-1, S-2, Y-1, Y-2, D-1), and the Safety Huber Needle Infusion Set is available in four models (S-1s, S-2s, Y-1s, Y-2s). The safety feature is manually activated during needle removal, and is designed to aid in the prevention of accidental needlestick. After activation, the non-coring needle is locked in the safety mechanism-plate. The device also have Y-sites with Y-needleless adapters or Y-needle adapters for secondary access. The needleless adapters eliminate the risk of needlestick.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the Huber Needle Infusion Set and Safety Huber Needle Infusion Set. This document is submitted to the FDA to demonstrate that a new device is substantially equivalent to a legally marketed predicate device, meaning it is as safe and effective.

    The provided text does not contain information about an AI/ML powered medical device. Therefore, the specific details requested regarding acceptance criteria and study proving an AI/ML device meets acceptance criteria (such as sample size for test sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, training set details, and ground truth establishment) are not present in this document.

    The document focuses on the physical and functional equivalence of the Huber Needle Infusion Set to its predicate device through a series of non-clinical tests (performance testing and biocompatibility testing).

    Here's what can be extracted, interpreted based on the provided document, and why the other AI/ML related questions cannot be answered:

    1. Table of Acceptance Criteria and Reported Device Performance (Focus: Non-AI/ML Physical Device)

    The document primarily outlines performance tests and biocompatibility tests to prove substantial equivalence of a physical medical device (Huber needle). It doesn't explicitly list "acceptance criteria" in a table format with "reported device performance" in the way one might for an AI/ML algorithm's effectiveness metrics (e.g., sensitivity, specificity thresholds). Instead, it states that tests were conducted to verify that the proposed device met all design specifications and test results demonstrated compliance with related standards requirements.

    However, we can infer some "acceptance criteria" based on the tests performed:

    Acceptance Criterion (Inferred from Tests)Reported Device Performance (Summary from document)
    Wing Flexibility: Withstand bending without crackingVerified as able to withstand bending without cracking.
    Penetration ForceDemonstrated compliance with ISO 10555-6:2015.
    Coring TestingDemonstrated compliance with ASTM F3212-16.
    Occlusion TestingVerified to withstand 45 psi pressure; liquid path free from leakage.
    Leakage, Joint Integrity, Tubing Testing, Tensile StrengthDemonstrated compliance with ISO 8536-4:2010.
    Safety Mechanism Activation and FunctionDemonstrated compliance with ISO 23908:2011 and FDA Guidance "Medical Devices with Sharps Injury Prevention Features". Test results showed safety mechanism protects needle tip from exposure and provides audible click.
    Ability to withstand power injection pressuresWatertight and resistant up to 330 psi.
    Power Injection Flow rateConfirmed maximum flow rates with a power injector.
    Needle performanceDemonstrated compliance with ISO 7864:2016 and ISO 9626:2016.
    Luer Connector performanceDemonstrated compliance with ISO 80369-7:2016 and ISO 80369-20:2015.
    MRI CompatibilityDemonstrated compliance with ASTM F2182-09, ASTM F2052-06, ASTM 2119-07, ASTM 2213-06 R11.
    ParticulatesDemonstrated compliance with USP <788>.
    SterilizationDemonstrated compliance with ISO 10993-7:2008, ASTM F1886/F1886M-16, ASTM F88/F88M-15, ASTM F1929-15.
    Biocompatibility (Cytotoxicity, Skin Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Pyrogen, Hemolytic Property, Partial Thromboplastin, Complement Activation, Bacterial Endotoxin Limit)Tested according to ISO 10993-1:2018 and specific sub-parts/related standards; results indicate compliance.

    Regarding AI/ML specific questions:

    2. Sample size used for the test set and data provenance:

    • Not applicable for AI/ML. This document describes testing of a physical medical device. The "test set" here refers to physical samples of the device and test media (e.g., liquids for flow rate, materials for penetration/coring). No medical imaging or patient data is involved.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable for AI/ML. Ground truth for a physical device is established through verified measurement techniques and standardized testing protocols, not expert consensus on medical images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable for AI/ML. Adjudication relates to resolving disagreements among experts for ground truth labeling in AI/ML studies. For physical device performance, tests are objective and measurable against established standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This device is a physical infusion set, not an AI/ML algorithm intended to assist human readers (e.g., radiologists). Therefore, no MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable for AI/ML diagnostic/interpretive purposes. For this physical device, "ground truth" is defined by the established physical, chemical, and biological properties of the materials and function of the device as measured by validated laboratory methods and adherence to international standards (e.g., ISO, ASTM, USP).

    8. The sample size for the training set:

    • Not applicable for AI/ML. This refers to a physical medical device, not an AI model requiring a training set.

    9. How the ground truth for the training set was established:

    • Not applicable for AI/ML. See point 8.

    In summary: The provided document is a 510(k) submission for a non-AI/ML medical device (Huber Needle Infusion Set). The "acceptance criteria" discussed are performance and material standards for the physical device, and the "study" consists of a comprehensive suite of non-clinical, laboratory-based tests to demonstrate compliance with these standards and substantial equivalence to a predicate device. None of the questions specifically pertaining to AI/ML device validation are addressed because the device is not an AI/ML product.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201356

    Validate with FDA (Live)

    Date Cleared
    2021-02-12

    (267 days)

    Product Code
    Regulation Number
    868.5140
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Plastic LOR Syringe is intended for use in conjunction with an epidural needle, to verify the needle tip placement in the epidural space by the loss of Resistance technique, it will be filled with air and/or saline during use. The loss of Resistance Syringe is not intended for injection or aspiration. The syringe will be sold sterile individually packaged, and as part of a sterile kit.

    Device Description

    The Plastic LOR syringe is a piston syringe of three pieces. It consists of barrel, plunger and piston. The barrel is available with Luer slip, NRFit lock, NRFit Slip male connector. The Plastic LOR syringe is provided in combinations of four connectors and four syringe capacities (5ml, 7ml, 10ml, 20ml). The device is a sterile finished disposable device, supplied sterile to the end user and non-sterile will be supplied to Anesthetic Conduction Kit manufacturers to be sterile and packaged into kit. The Plastic LOR syringe is a single use device which will be supplied as sterile individually packaged and the non-sterile bulk packed. The sterilized by Ethylene Oxide Gas to achieve a SAL of 10° and supplied sterility maintenance package which could maintain the sterility of the device during the s helf life of five years.

    AI/ML Overview

    This document is a 510(k) Summary for a Plastic LOR Syringe. The purpose of this summary is to demonstrate that the device is substantially equivalent to legally marketed predicate devices.

    Here's the breakdown of the acceptance criteria and the study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily derived from various ISO and ASTM standards for syringes, neuraxial, and Luer connectors, as well as biocompatibility and sterilization standards. The reported device performance indicates compliance with these standards.

    Acceptance Criteria CategorySpecific Standard/ClauseReported Device Performance
    Physical/MechanicalClause 5 of ISO 7886-1:2017 (General requirements)Conforms to requirements
    Clause 6 of ISO 7886-1:2017 (Extraneous matter)Conforms to requirements
    Clause 7 of ISO 7886-1:2017 (Lubricant)Conforms to requirements
    Clause 8 of ISO 7886-1:2017 (Tolerance on graduated capacity)Conforms to requirements
    Clause 9 of ISO 7886-1:2017 (Graduated scale)Conforms to requirements
    Clause 10 of ISO 7886-1:2017 (Barrel)Conforms to requirements
    Clause 11 of ISO 7886-1:2017 (Piston stopper/plunger assembly)Conforms to requirements
    Clause 12 of ISO 7886-1:2017 (Nozzle)Conforms to requirements
    Clause 13 of ISO 7886-1:2017 (Performance)Conforms to requirements
    Neuraxial Small-Bore ConnectorClause 5 of ISO 80369-6:2016 (Dimensional requirements)Conforms to requirements
    Clause 6.1 of ISO 80369-6:2016 (Fluid leakage)Conforms to requirements
    Clause 6.2 of ISO 80369-6:2016 (Sub-atmospheric pressure air leakage)Conforms to requirements
    Clause 6.3 of ISO 80369-6:2016 (Stress cracking)Conforms to requirements
    Clause 6.4 of ISO 80369-6:2016 (Resistance to separation from axial load)Conforms to requirements
    Clause 6.5 of ISO 80369-6:2016 (Resistance to separation from unscrewing)Conforms to requirements
    Clause 6.6 of ISO 80369-6:2016 (Resistance to overriding)Conforms to requirements
    Luer ConnectorClause 5 of ISO 80369-7:2016 (Dimensional requirements)Conforms to requirements
    Clause 6.1 of ISO 80369-7:2016 (Fluid leakage)Conforms to requirements
    Clause 6.2 of ISO 80369-7:2016 (Sub-atmospheric pressure air leakage)Conforms to requirements
    Clause 6.3 of ISO 80369-7:2016 (Stress cracking)Conforms to requirements
    Clause 6.4 of ISO 80369-7:2016 (Resistance to separation from axial load)Conforms to requirements
    Clause 6.5 of ISO 80369-7:2016 (Resistance to separation from unscrewing)Conforms to requirements
    Clause 6.6 of ISO 80369-7:2016 (Resistance to overriding)Conforms to requirements
    Sterile Barrier PackagingASTM F1886/F1886M-16 (Visual inspection)Maintained integrity
    ASTM F88/F88-15 (Seal strength)Maintained integrity
    ASTM F1929-15 (Dye penetration test)Maintained integrity
    Sterilization & Shelf LifeISO 10993-7:2008 (EO residue)EO ECH residue did not exceed the limit
    USP 39-NF 34 <85> (Bacteria Endotoxin Limit)Endotoxin limit did not exceed 2.15 EU/device
    Shelf Life Evaluation (Physical, Mechanical, Chemical, Package Tests on aging samples)Maintained performance during claimed shelf life (5 years)
    Biocompatibility(Implicitly based on ISO 10993 for indirect contact, <24 hours)No negative impacts from materials: Cytotoxicity (Passed), Sensitization (Passed), Intracutaneous reactivity (Passed), Acute Systemic Toxicity (Passed), Pyrogen (Passed), Particulate testing (Passed).
    Connector Compatibility(Implicitly tested against incompatible connectors)Would not be connected with incompatible connector (verified)

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each non-clinical test (e.g., how many syringes were tested for fluid leakage, or how many samples for biocompatibility tests). It simply states that "Non clinical tests were conducted to verify that the proposed device met all design specifications."

    The data provenance is from non-clinical tests conducted by Jiangsu Caina Medical Co., Ltd. (the manufacturer). The country of origin of the testing data is presumably China, where the manufacturer is located. The tests are retrospective in the sense that they were performed to support the 510(k) submission for an already developed device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as this is a non-clinical submission for a physical medical device (syringe), not an AI/software device that requires expert-established ground truth for a test set. The "ground truth" here is compliance with established international and national standards.

    4. Adjudication method for the test set

    This information is not applicable for the same reasons as point 3. Testing against standards does not typically involve adjudication in the way AI/software studies do.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This is a 510(k) submission for a physical medical device (Plastic LOR Syringe), which does not involve human readers or AI assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm performance study was not done. This device is a physical medical instrument, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for this submission is compliance with recognized consensus standards (ISO, ASTM, USP) for medical devices, materials, and sterilization, as well as the demonstration of substantial equivalence to predicate devices based on these tests.

    8. The sample size for the training set

    This information is not applicable as this is a physical medical device. There is no AI model or "training set" involved in its development or regulatory submission.

    9. How the ground truth for the training set was established

    This information is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2