Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K162233
    Manufacturer
    Date Cleared
    2016-12-23

    (137 days)

    Product Code
    Regulation Number
    862.1675
    Reference & Predicate Devices
    Why did this record match?
    Reference Devices :

    K980414, K923090

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Kurin Blood Culture Collection Set is a winged blood collection needle with flexible tubing intended for venipuncture to obtain blood samples. It is provided with a safety shield for covering the used venipuncture needle prior to disposal to aid in the prevention of needlestick injury if manually activated after the blood collection, the set also includes a safety shield and apparatus for connection to vacuum based collection vials.

    Device Description

    The Kurin device is a sterile, single use blood culture collection set. The Kurin includes a winged needle with flexible tubing and an attached vial adapter intended for venipuncture to obtain blood culture samples. Kurin is identical in every way to existing sets used to collect blood culture samples except for the insertion into the tubing of the Kurin blood capture chamber. The Kurin blood capture device sequesters the initial draw of blood upon initial venipuncture. The set is provided with a safety shield for covering the used needle prior to disposal. The amount of blood diverted is very small, estimated at a fraction of 1 ml.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called the "Kurin Blood Culture Collection Set." This document focuses on demonstrating substantial equivalence to predicate devices through non-clinical testing rather than proving performance against specific acceptance criteria for diagnostic accuracy or effectiveness in a clinical context. Therefore, the information typically found in acceptance criteria and detailed study descriptions for AI/ML medical devices (like sensitivity, specificity, F1 score, sample size for test sets, expert qualifications, etc.) is not present in this document.

    However, I can extract the non-clinical testing performed and present them as "acceptance criteria" and "reported performance" based on how they are described in the document as meeting specifications.

    Here's the analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Derived from Non-Clinical Testing)Reported Device Performance
    Sterilization: Compliance with ISO 11135-1:2006 for Ethylene Oxide sterilization.The Kurin device is sterilized using a validated EO sterilization process which complies with ISO 11135-1:2006.
    Aging/Shelf Life: Achieve a 1-year shelf life for sterility and performance. Protocols for up to 3 years.The Kurin device was validated to achieve 1-year shelf life with protocols for up to 3 years of shelf life for sterility and performance.
    Functional, Leakage, and Tensile Test: No compromise to the function of the blood collection device, no leakage, and passing tensile testing in accordance with ISO 1135-3 sections 5.2 and 5.3.The Kurin set confirmed the addition of the Kurin device shows no compromise to the function of the blood collection device in regards to functionality. With the addition of the device, the product does not leak and passed tensile testing in accordance with sections 5.2 and 5.3 of ISO 1135-3.
    Packaging Integrity and Shipping Test: Compliance with ISO 11607 and ASTM D4169-14.This test was completed, and the Kurin device passed all tests in accordance with ISO 11607 and ASTM D4169-14.
    Biocompatibility Tests: No leachables (MEM Elution Test), no adverse effect on blood constituents (IVH Blood Count Test).The Kurin device passed two types of biocompatibility tests: the MEM Elution Test demonstrated that no leachables are present from the system, and the IVH Blood Count Test which demonstrated that the blood path did not adversely affect the constituents of blood exposed to the systems fluid path.
    User Verification Test: Instructions for use are easy to understand, and device functionality is acceptable.Testing was conducted to evaluate if the instructions for use were easy to understand and the functionality of the device. (Implied success as it "passed")
    Flow Rate Test: Addition of the Kurin Device does not change the flow rate of liquid compared to the predicate.Testing verified that the addition of the Kurin Device into the tubing of the FDA cleared BD Vacutainer® brand blood collection set and Safety-Lok™ Blood Collection set did not change the flow rate of liquid passing through the device.

    2. Sample size used for the test set and the data provenance

    The document describes non-clinical engineering and bench testing, not a clinical test set with patient data. No specific sample sizes for these tests (e.g., number of devices tested for flow rate, leakage, etc.) are provided in the summary. Data provenance is not applicable in the context of clinical populations as it's not a clinical study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable as this document describes non-clinical, bench testing, and engineering verification, not a study evaluating expert performance or ground truth established by experts. The "ground truth" here is determined by the specifications of the standards (e.g., ISO, ASTM) and the physical properties of the device.

    4. Adjudication method for the test set

    Not applicable. Adjudication methods are typically relevant for studies involving human interpretation or subjective assessments, which are not described here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a blood collection set, not an AI/ML diagnostic tool, and no MRMC studies are mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical medical device, not an algorithm, and its performance is assessed via non-clinical testing.

    7. The type of ground truth used

    The "ground truth" for the non-clinical tests described in this document is based on engineering specifications, compliance with international standards (ISO, ASTM), and predefined performance benchmarks (e.g., no leakage, specific flow rates, successful sterilization).

    8. The sample size for the training set

    Not applicable. As this is not an AI/ML device, there is no training set.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Reference Devices :

    K980414, K090426

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IMPROSAFE® Blood Collection Sets with Pre-attached to be used with evacuated blood collection tube for the collection of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.
    The IMPROSAFE® Multi-Sample Needles are intended to be used with evacuated blood collection tube for the collection of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.
    The IMPROVACUTER® Blood Collection Sets are intended to be used with evacuated blood collection tube for the collection of venous blood.
    The IMPROSAFE® Blood Collection Sets are intended to be used with evacuated blood collection tube for the collection of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.
    The IMPROVACUTER® Blood Collection Sets with Pre-attached holder are intended to be used with evacuated blood collection tube for the collection of venous blood.
    The IMPROSAFE® Multi-Sample Needles (Flashback) are intended to be used with evacuated blood collection tube for the collection of venous blood. The safety shield is intended to aid in the protection against accidental needle stick injury.
    The IMPROVACUTER® Multi-Sample Needles are used with holder and vacuum blood collection tube for the venous blood collection.

    Device Description

    The proposed devices are blood collection sets form a channel between patient's vein and the evacuated blood collection tube intended for collection of blood. They are available in different configurations as follows:
    IMPROSAFE® Blood Collection Set with Pre-attached holder: Patient-End Needle, Non-Patient End Needle, Flexible Tubing, Pre-attached Holder, Safety Mechanism, Visible Flash Back
    IMPROSAFE® Blood Collection Set: Patient-End Needle, Non-Patient End Needle, Flexible Tubing, N.A., Safety Mechanism, Visible Flash Back
    IMPROVACUTER® Blood Collection Set with Pre-attached Holder: Patient-End Needle, Non-Patient End Needle, Flexible Tubing, Pre-attached Holder, N.A., Visible Flash Back
    IMPROVACUTER® Blood Collection Set: Patient-End Needle, Non-Patient End Needle, Flexible Tubing, N.A., N.A., Visible Flash Back
    IMPROSAFE® Multi Sample Needle: Patient-End Needle, Non-Patient End Needle, N.A., Pre-attached Holder, Safety Mechanism, N.A.
    IMPROSAFE® Multi Sample Needle (Flashback): Patient-End Needle, Non-Patient End Needle, N.A., Pre-attached Holder, Safety Mechanism, Visible Flash Back
    IMPROVACUTER® Multi Sample Needle: Patient-End Needle, Non-Patient End Needle, N.A., N.A., N.A., N.A.

    AI/ML Overview

    The provided text is a 510(k) Summary for several blood collection devices. It aims to demonstrate substantial equivalence to previously cleared predicate devices, rather than proving a device meets specific acceptance criteria through an independent clinical study for AI software.

    Therefore, many of the requested categories (e.g., sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, type of ground truth, training set information) are not applicable or not provided in this document, as it pertains to medical hardware (needles and collection sets) and not AI software.

    However, I can extract information related to non-clinical tests and the general equivalence claims.

    Here's the breakdown of the information provided in the document:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly state "acceptance criteria" in a quantitative, pre-defined manner for performance metrics. Instead, it refers to conformity with established international and ASTM standards. The "reported device performance" is essentially that the device "conforms" to these standards.

    Acceptance Criteria (Implied by Standards)Reported Device Performance (Conformity)
    Material (Stainless steel needle tubing): Conforms to ISO 9626:1991+AMENDMENT 1 2001Conforms to ISO 9626:1991 AMD 2001 (and ISO 7864:1993 for sterile hypodermic needles)
    Sterile Hypodermic Needles: Conforms to ISO 7864: 1993Conforms to ISO 7864: 1993 (and ISO 9626:1991 AMD 2001 for needle tubing)
    Ethylene Oxide Sterilization residuals: Conforms to ISO 10993-7:2008Conforms with ISO 10993
    Seal Strength of Flexible Barrier Materials: Conforms to ASTM F 88/F88M-09Conforms to ASTM F 88/F88M-09
    Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration: Conforms to ASTM F1929-12Conforms to ASTM F1929-12
    Bacterial Endotoxins Limit: Conforms to USP38-NF33Conforms to USP38-NF33
    Tests for in vitro cytotoxicity: Conforms to ISO 10993-5:2009Conform with ISO 10993
    Tests for irritation and delayed-type hypersensitivity: Conforms to ISO 10993-10:2010Conform with ISO 10993
    Tests for systemic toxicity: Conforms to ISO 10993-11:2006Conform with ISO 10993
    Assessment of hemolytic properties of materials: Conforms to ASTM F 756-13Conform to ASTM F 756-13
    Sharps Injury Prevention Features: Conforms to Guidance for Industry and FDA Staff: Medical Devices with Sharps Injury Prevention FeaturesConforms to Guidance for Industry and FDA Staff: Medical Devices with Sharps Injury Prevention Features
    Labeling Requirements: Conforms to 21 CFR Part 801Conform with 21 CFR Part 801
    Sterility: EO sterilizedEO sterilized
    Single Use: Single UseSingle Use

    2. Sample sized used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Not provided. This document does not detail specific sample sizes for non-clinical tests or data provenance. The tests are general compliance tests for medical devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    • Not applicable/Not provided. This is not an AI device or a diagnostic device relying on expert interpretation for "ground truth." The tests are for physical/biological properties and compliance with engineering standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable/Not provided. This concept is not relevant to the non-clinical testing of blood collection devices described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a blood collection set/needle, not an AI-powered diagnostic tool, so MRMC studies are irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable. The "ground truth" for this type of device is compliance with specific material, sterility, performance (e.g., seal strength, endotoxin limits), and biocompatibility standards, as measured by established laboratory test methods.

    8. The sample size for the training set:

    • Not applicable. There is no "training set" as this is not a machine learning device.

    9. How the ground truth for the training set was established:

    • Not applicable. There is no "training set" or corresponding ground truth for this type of device.

    Summary of the Study (Non-Clinical Test Conclusion) from the document:

    The non-clinical tests were conducted to verify that the proposed devices met all design specifications and were Substantially Equivalent (SE) to the predicate devices. The test results demonstrated that the proposed device complies with the following standards:

    • ISO 9626:1991+AMENDMENT 1 2001 (Stainless steel needle tubing)
    • ISO 7864: 1993 (Sterile hypodermic needles for single use)
    • ISO 10993-7:2008 (Biological evaluation - Ethylene oxide sterilization residuals)
    • ASTM F 88/F88M-09 (Seal strength of flexible barrier materials)
    • ASTM F1929-12 (Detecting seal leaks in porous medical packaging by dye penetration)
    • USP38-NF33 (Bacterial Endotoxins Limit)
    • ISO 10993-5:2009 (Biological evaluation - Tests for in vitro cytotoxicity)
    • ISO 10993-10:2010 (Biological evaluation - Test for irritation and delayed-type hypersensitivity)
    • ISO 10993-11:2006 (Biological evaluation - Tests for systemic toxicity)
    • ASTM F 756-13 (Assessment of hemolytic properties of materials)
    • Guidance for Industry and FDA Staff: Medical Devices with Sharps Injury Prevention Features

    Clinical Test Conclusion:
    "No clinical study is included in this submission."

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1