Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K251185

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-10-22

    (189 days)

    Product Code
    Regulation Number
    870.5150
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Recana Thrombectomy Catheter System is indicated for:

    • The non-surgical removal of thrombi and emboli from veins.
    • Injection, infusion, and/or aspiration of contrast media and other fluids into or from a vein.

    The Recana Thrombectomy Catheter System is indicated for use in peripheral venous vasculature.

    Device Description

    The Recana Thrombectomy Catheter System is designed to remove thrombus and emboli from native vessels or stented veins. The system is comprised of the debulking catheter, collection basket(s) with delivery sheaths, and a family of sheaths. Together, the Recana Thrombectomy Catheter System functions to capture and remove obstructive / occlusive thrombus from the venous vasculature.

    The debulking catheter utilizes an adjustable diameter stainless steel coring element to remove thrombus from the venous vasculature. Handle controls enable expansion and collapse of the coring element to achieve the desired working diameter. The debulking catheter is 0.035" guidewire and 0.060" basket shaft compatible and provides a working length of 80cm.

    18mm and 30mm Collection basket(s) are self-expanding nitinol wire-form and braided structures designed to ensure capture and removal of thrombi and emboli. The basket shaft has a 0.035" guidewire compatible lumen for over the wire delivery through a 9Fr Delivery Sheath with a working length of 95cm.

    Sheaths for introduction of the debulking catheter and collection baskets into the vasculature and for the recapture and removal of thrombi and emboli via the collection basket(s) are offered in four (4) size configurations (13Fr x 30cm, 13Fr x 90cm, 16Fr x 30cm, and 16Fr x 90cm) providing optionality based on access vessel size and location.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K222185

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-04-13

    (265 days)

    Product Code
    Regulation Number
    870.1210
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    In selective areas of peripheral veins between 7 and 16 mm in diameter, the Stratus Infusion Catheter is intended for the infusion of diagnostic and therapeutic agents into the vein wall or perivascular area, or intraluminally.

    Device Description

    The Stratus™ Infusion Catheter is a peripheral infusion catheter designed to access an intraluminal, intramural, or perivascular vein wall layer and infuse fluid diagnostic and/or therapeutic agents. The Stratus Infusion Catheter is a non-implantable device that is provided sterile and is intended for single patient use. The Stratus Infusion Catheter is designed to accommodate individual vein anatomies or disease progression (e.g., thrombosis), by the catheter platform allowing multiple infusions and infusion locations per device.

    AI/ML Overview

    I am sorry, but the provided text does not contain information about acceptance criteria or a study proving that a device meets those criteria. The document is an FDA 510(k) clearance letter for the "Stratus Infusion Catheter," which focuses on establishing substantial equivalence to a predicate device.

    While it mentions "design verification and validation testing" including bench testing, biocompatibility, sterilization validation, and animal testing, it does not provide:

    • A table of acceptance criteria and reported device performance.
    • Sample sizes used for a test set or data provenance.
    • Details about experts used for ground truth or their qualifications.
    • Adjudication methods.
    • Results of a multi-reader multi-case (MRMC) comparative effectiveness study.
    • Details of a standalone (algorithm only) performance study.
    • The specific type of ground truth used (beyond implying general animal study outcomes).
    • Sample size for a training set or how its ground truth was established.

    The document primarily focuses on a comparison of the characteristics of the Stratus Infusion Catheter to its predicate device (Bullfrog Micro-Infusion Device) to demonstrate substantial equivalence, rather than detailing a study that measures its performance against specific acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DASH 6® NRFit Lock syringe and DASH® NRFit Slip Syringe are intended to be used with ISO 80369-6 NRFit neuraxial compliant devices for administration of neuraxial medication.

    The DASH 6® NRFit Plastic LOR Device is intended to be used with ISO 80369-6 NRFit neuraxial compliant epidural needle for locating the epidural space.

    Accessories:

    DASH 6® Introducer Needle is intended to be used with ISO 80369-6 NRFit neuraxial compliant spinal needle for guiding the placement of the spinal needle into the arachnoid/epidural space.

    DASH 6@ NRFit Syringe Cap is intended to be used with ISO 80369-6 NRFit neuraxial syringes for sealing the tip of the syringe.

    DASH 6® NRFit Drawing Up Filter Straw are intended to be used with ISO 80369-6 NRFit neuraxial syringe for the drawing up of neuraxial medication and anesthetic.

    DASH 6@ NRFit Blunt Drawing Up Needle (with and without filter) is intended to be used with ISO 80369-6 NRFit neuraxial syringe for the drawing up of neuraxial medication and anesthetic.

    DASH 6® NRFit Bacterial Disc Filter is intended to be used with ISO 80369-6 NRFit neuraxial compliant devices to ensure aseptic administration of neuraxial medication and anesthetic.

    DASH 6® NRFit Epidural Flat Filter is intended to be used with ISO 80369-6 NRFit neuraxial compliant devices to ensure aseptic administration of neuraxial medication and anesthetic.

    DASH 6® NRFit Tuohy-Borst Adapter is intended to be used with a epidural catheter to provide an ISO 80369-6 NRFit compliant connection.

    DASH 6® NRFit Syringe to Syringe Adapter is intended to be used with ISO 80369-6 NRFit neuraxial syringe to allow for mixing/transfering medication between two syringes.

    DASH 6® NRFit Needle Hub Cap is intended for sealing the hub of an NRFit compliant needle.

    DASH 6@ NRFit Epidural Catheter Feeder is intended to be used with ISO 80369-6 NRFit neuraxial compliant devices to assist the insertion of an epidural needle into the epidural space.

    Device Description

    The DASH 6® NRFit Syringes (Lock and Slip Syringes) are Single Use, in-hospital devices. They are provided in sizes ranging from 1 mL to 60 mL. The devices incorporate a male NRFit connector for connection to a female NRFit port. The syringes are designed according to ISO 7886-1 & ISO 80369-6 standard. The syringes are used for neuraxial purposes (ISO 80369-6 NRFit connection) and functions exactly the same as any hypodermic syringes that are on the market (ISO 7886-1).

    The syringes will be supplied as individually packed (sterile) and the bulk packed (Nonsterile). The Sterile packed items will be supplied directly to the user and the bulk packed will be supplied to Anesthetic Conduction Kit manufacturers to be packaged into kit.

    The syringe packaging will indicate the volume/size, the connector type (Lock or Slip variants) that the syringe contains and the sterility/sterilization method the syringe have been processed.

    The DASH 6 ° Plastic LOR devices are Single Use, in-hospital devices. It is provided with a 10 mL size/volume and incorporates a male ISO 80369-6 NRFit slip connector to connect to a female ISO 80369-6 NRFit Port.

    The Plastic LOR devices are designed according to ISO 80369-6 standard. The syringes are used for neuraxial purposes (ISO 80369-6 NRFit connection) and functions exactly the same as the LOR devices on the market.

    The device will be supplied as individually packed (sterile) and the bulk packed (Nonsterile). The Sterile packed items will be supplied directly to the user and the bulk packed will be supplied to Anesthetic Conduction Kit manufacturers to be packaged into kit.

    The devices packaging will indicate the volume/size, the connector type (Slip) that the device contains and the sterility/sterilization method the device have been processed.

    Accessories:
    DASH 6® NRFit Syringe Caps: Single Use, in-hospital devices with a female ISO 80369-6 NRFit connector.
    DASH 6® NRFit Drawing up Filter Straw: Single Use, in-hospital devices with a female ISO 80369-6 NRFit connector and filter.
    DASH 6 ° NRFit Blunt Drawing up Needle (with and without filter): Single Use, in-hospital devices with a female ISO 80369-6 NRFit connector and blunt needle (with or without filter).
    DASH 6® NRFit Bacterial Disc Filter: Single Use, in-hospital devices with male & female ISO 80369-6 NRFit connectors and a 0.22µm filter.
    DASH 6® NRFit Epidural Flat Filter: Single Use, in-hospital devices with male & female ISO 80369-6 NRFit connectors and a 0.22um filter.
    DASH 6 ° NRFit Tuohy Borst Adapter: Single Use, in-hospital devices with a female ISO 80369-6 NRFit connector and open catheter port.
    DASH 6® NRFit Needle Hub Caps: Single Use, in-hospital devices with a Male ISO 80369-6 NRFit connector.
    DASH 6® Epidural Catheter Feeder: Single Use, in-hospital devices with a compatible male ISO 80369-6 NRFit connector.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for the DASH 6® NRFit System, which includes syringes and various accessories. This document primarily focuses on demonstrating substantial equivalence to predicate devices through design verification and compliance with specific ISO standards. It does not describe an AI medical device or a study involving human readers or AI assistance. Therefore, I cannot provide information on acceptance criteria and a study that proves the device meets those criteria in the context of AI performance, MRMC studies, standalone algorithm performance, number of experts, or ground truth for AI model development.

    However, I can extract the acceptance criteria and the type of study conducted to demonstrate the device's technical specifications and substantial equivalence, as presented in the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the DASH 6® NRFit System are primarily defined by compliance with several ISO standards. The studies conducted were design verification tests demonstrating compliance with these standards.

    Acceptance Criteria (Defined by ISO Standards)Reported Device Performance (as demonstrated by testing)
    ISO 80369-6: NRFit Connectors
    - Fluid Leakage (per ISO 80369-20 Annex C)Met all acceptance criteria. Functions verified to operate as designed.
    - Stress Cracking (per ISO 80369-20 Annex E)Met all acceptance criteria. Functions verified to operate as designed.
    - Resistance to separation from axial load (per ISO 80369-20 Annex F)Met all acceptance criteria. Functions verified to operate as designed.
    - Resistance to separation from unscrewing (per ISO 80369-20 Annex G)Met all acceptance criteria. Functions verified to operate as designed.
    - Resistance to overriding (per ISO 80369-20 Annex H)Met all acceptance criteria. Functions verified to operate as designed.
    - Dimensional analysis for critical dimensionsMet all acceptance criteria for critical dimensions in accordance with ISO 80369-6.
    ISO 7886-1: Sterile hypodermic syringes for single use (for DASH 6 Syringes)
    - Capacity ToleranceDevice Verification Testing conducted according to ISO 7886-1. Functions verified to operate as designed.
    - Leakage Testing Side ForceDevice Verification Testing conducted according to ISO 7886-1. Functions verified to operate as designed.
    - Leakage Testing Axial ForceDevice Verification Testing conducted according to ISO 7886-1. Functions verified to operate as designed.
    ISO 10993 Series (Biocompatibility)
    - Cytotoxicity (ISO 10993-5)Testing done per ISO 10993-5. (Implied compliance as part of "All necessary bench testing was conducted... to support a determination of substantial equivalence").
    - Sensitization (ISO 10993-10)Testing done per ISO 10993-10. (Implied compliance as part of "All necessary bench testing was conducted...").
    - Extractables and Leachables (ISO 10993-18 and 10993-19)Testing done per ISO 10993-18 and 10993-19. (Implied compliance as part of "All necessary bench testing was conducted...").
    ISO 7864-1: Sterile hypodermic needles for single use (for DASH 6 NRFit needles - in certain respects)
    - Needle to Needle Hub bondingDesign verification testing conducted in accordance with ISO 7864-1 (in certain respects). Functions verified to operate as designed.
    - Dimensional tolerancesDesign verification testing conducted in accordance with ISO 7864-1 (in certain respects). Functions verified to operate as designed.
    - Blockages within the needleDesign verification testing conducted in accordance with ISO 7864-1 (in certain respects). Functions verified to operate as designed.
    Filter Device Specific Tests
    - Bubble point testDesign verification testing completed to ensure functioning according to requirement.
    - Flow-rateDesign verification testing completed to ensure functioning according to requirement.
    - Filtration efficiencyDesign verification testing completed to ensure functioning according to requirement.
    Other General Tests
    - Sterilization ValidationConducted. (Implied compliance as part of "All necessary bench testing was conducted...").
    - Shipping and PackagingConducted. (Implied compliance as part of "All necessary bench testing was conducted...").
    - Accelerated AgingConducted. (Implied compliance as part of "All necessary bench testing was conducted...").
    - Risk Assessment (FMEA, Risk Management Report)Risk associated with misconnection of the NRFit connector assessed and captured. (Implied that mitigation efforts and assessment meet acceptable risk levels).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document describes bench testing for physical characteristics, dimensional verification, and biocompatibility, not a clinical study with a "test set" in the context of AI or diagnostic performance. Therefore, typical sample size or data provenance details for medical image analysis are not applicable. The testing was conducted to verify compliance with engineering standards (ISO standards). The document does not specify the exact number of units tested for each specific test, but mentions "All necessary bench testing was conducted."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as the device is a physical medical device (syringes, needles, etc.) and its performance is evaluated against engineering and biological standards, not diagnostic interpretations from experts. Ground truth is established by objective measurements against specified engineering and material science criteria.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This device is not an AI diagnostic tool and does not involve human adjudication of results in the traditional sense of a clinical study or image interpretation.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The DASH 6® NRFit System is a medical device (syringes, needles, etc.) and not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is not an algorithm or AI system. Its performance is evaluated intrinsically based on its physical properties and adherence to manufacturing and safety standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's evaluation is primarily the specifications and requirements defined by international standards (e.g., ISO 80369-6, ISO 7886-1, ISO 10993 series) and the manufacturer's own design specifications. It is based on objective, quantifiable physical and chemical measurements (e.g., fluid leakage, dimensions, material compatibility).

    8. The sample size for the training set

    Not applicable. There is no "training set" as this is a physical medical device, not a machine learning model.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DASH 3TM ENFit Syringe is a sugle use syninge indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids or nututional formula into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral synnges are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups.

    The DASH 3TM Eccentric ENFit Syringe is a single use syringe indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids or nutritional formula into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syninges are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups.

    The DASH 3TM Low Volume Tip ENFit Syringe is a sungle use syringe indicated for use as a dispenser, a measuning device, and a fluid transfer device. It is used to deliver fluids or nutritional formula into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syninges are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinicians, in all age groups.

    The DASH 3M ENFit Syringe Cap allows the advance preparation and secure storage and transport of medication fluids or mutritional formula. The synnge cap will fit any size of the Single Use Synnges (DASH 3TM)

    Device Description

    The DASH 3™ ENFit Syringe device family is a Single Use, in-hospital and home care (DASH 3TM ENFit Syringe, DASH 3TM Eccentric ENFit Syringe, DASH 3TM ENFit Low Dose Tip Syringe) device.

    It is provided in sizes from 1 mL to 100 mL. The device incorporates a female ENFit connector for connection to an enteral access device with a male ENFit port specified in ISO 80369-3. The Low Dose Tip contains the low dose design feature specified in (Draft) ISO 20695.

    The DASH 3TM ENFit accessories are designed to be compatible with ENFit connector devices. The DASH 3TM ENFit Syringe Caps fit any size of the Single Use ENFit Syringes (DASH 3™), which allows the advance preparation and secure storage and transport of medication/fluids or nutritional formula.

    AI/ML Overview

    The provided text describes a medical device, the DASH 3™ ENFit Syringe family and its accessories, and outlines the testing conducted to demonstrate its substantial equivalence to predicate devices, rather than a study proving the device meets specific acceptance criteria in the context of diagnostic performance. Therefore, I cannot extract information related to sample sizes for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, or training set details as these are not relevant to this type of device submission.

    However, I can provide the acceptance criteria and confirmed performance based on the engineering and functional testing described in the document.

    Acceptance Criteria and Reported Device Performance

    The device performance is demonstrated through various tests designed to confirm compliance with ISO standards and other design specifications. The document states that the DASH 3™ Syringe device family and its accessories "met all acceptance criteria, as described in Section 17." The table below summarizes the listed tests, which serve as the acceptance criteria, and the reported outcome.

    Acceptance Criteria (Individual Test Defined in ISO 80369-3 or other specified standard)Reported Device Performance
    ISO 80369-3 and 80369-20 ENFit Connector Tests:
    Fluid Leakage (Test Method Defined in ISO 80369-20 Annex C)Met acceptance criteria
    Stress Cracking (Test Method Defined in ISO 80369-20 Annex E)Met acceptance criteria
    Resistance to separation from axial load (Test Method Defined in ISO 80369-20 Annex F)Met acceptance criteria
    Resistance to separation from unscrewing (Test Method Defined in ISO 80369-20 Annex G)Met acceptance criteria
    Resistance to overriding (Test Method Defined in ISO 80369-20 Annex H)Met acceptance criteria
    Disconnection by unscrewing (Test Method Defined in ISO 80369-20 Annex I)Met acceptance criteria
    Dimensional Analysis:
    Conformity to critical dimensions of ISO 80369-3Met acceptance criteria
    Device Verification Testing for Piston Syringe (ISO 7886-1):
    LubricantMet acceptance criteria
    Limits of extractable metalsMet acceptance criteria
    Limits for acidity/alkalinityMet acceptance criteria
    Capacity Tolerance (Graduated Scale)Met acceptance criteria
    Dead SpaceMet acceptance criteria
    Air and liquid leakage Testing (Side Force/Axial Force)Met acceptance criteria
    Biocompatibility Testing:
    Meeting requirements for biocompatibilityMet acceptance criteria
    Stability Testing:
    Evaluation of properties after accelerated aging to establish 5-year shelf life for syringes and 3-5-year for accessoriesMet acceptance criteria
    Confirmation of strength and integrity characteristics of sterile device packaging and sterile barrier maintenance for labeled shelf lifeMet acceptance criteria
    Risk Management:
    Assessment of risk associated with misconnection of the ENFit connectorRisk assessed and process captured
    Additional testing for Low Dose Syringe tip design:
    Low Dose Enteral Connector Misconnection AssessmentMet acceptance criteria
    Usability Assessment for the Low Dose Syringe Tip DesignMet acceptance criteria
    Low Dose Syringe Design Dose Accuracy TestingMet acceptance criteria

    Details of the Study:

    The studies conducted are primarily engineering and functional bench tests, and risk assessments.

    • Sample sized used for the test set and the data provenance: The document does not specify exact sample sizes for each test. The data provenance is implied to be from internal testing conducted by the manufacturer, Intervene Group Limited, likely in the UK, given the address. The studies are prospective in the sense that they were conducted for the purpose of this 510(k) submission.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for these types of engineering tests is typically defined by adherence to published international standards (e.g., ISO) and the device's design specifications, not by expert consensus on observational data.
    • Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. Performance is measured against predefined objective standards and specifications.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/diagnostic device.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI/diagnostic device.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc): The "ground truth" for these tests is the adherence to the requirements outlined in the cited ISO standards (ISO 80369-3, ISO 80369-20, ISO 7886-1, Draft ISO 20695) and the internal design specifications of the device.
    • The sample size for the training set: Not applicable. This is not a machine learning device; therefore, there is no "training set."
    • How the ground truth for the training set was established: Not applicable.

    In conclusion, the study supporting the device's substantial equivalence consists of a series of bench tests and assessments confirming that the DASH 3™ ENFit Syringe family and its accessories meet established engineering and performance standards, thereby demonstrating that they perform as intended and do not raise new questions of safety or effectiveness compared to predicate devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1