K Number
K170371

Validate with FDA (Live)

Date Cleared
2017-10-20

(256 days)

Product Code
Regulation Number
876.5980
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DASH 3TM ENFit Syringe is a sugle use syninge indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids or nututional formula into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral synnges are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups.

The DASH 3TM Eccentric ENFit Syringe is a single use syringe indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids or nutritional formula into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syninges are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups.

The DASH 3TM Low Volume Tip ENFit Syringe is a sungle use syringe indicated for use as a dispenser, a measuning device, and a fluid transfer device. It is used to deliver fluids or nutritional formula into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syninges are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinicians, in all age groups.

The DASH 3M ENFit Syringe Cap allows the advance preparation and secure storage and transport of medication fluids or mutritional formula. The synnge cap will fit any size of the Single Use Synnges (DASH 3TM)

Device Description

The DASH 3™ ENFit Syringe device family is a Single Use, in-hospital and home care (DASH 3TM ENFit Syringe, DASH 3TM Eccentric ENFit Syringe, DASH 3TM ENFit Low Dose Tip Syringe) device.

It is provided in sizes from 1 mL to 100 mL. The device incorporates a female ENFit connector for connection to an enteral access device with a male ENFit port specified in ISO 80369-3. The Low Dose Tip contains the low dose design feature specified in (Draft) ISO 20695.

The DASH 3TM ENFit accessories are designed to be compatible with ENFit connector devices. The DASH 3TM ENFit Syringe Caps fit any size of the Single Use ENFit Syringes (DASH 3™), which allows the advance preparation and secure storage and transport of medication/fluids or nutritional formula.

AI/ML Overview

The provided text describes a medical device, the DASH 3™ ENFit Syringe family and its accessories, and outlines the testing conducted to demonstrate its substantial equivalence to predicate devices, rather than a study proving the device meets specific acceptance criteria in the context of diagnostic performance. Therefore, I cannot extract information related to sample sizes for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, or training set details as these are not relevant to this type of device submission.

However, I can provide the acceptance criteria and confirmed performance based on the engineering and functional testing described in the document.

Acceptance Criteria and Reported Device Performance

The device performance is demonstrated through various tests designed to confirm compliance with ISO standards and other design specifications. The document states that the DASH 3™ Syringe device family and its accessories "met all acceptance criteria, as described in Section 17." The table below summarizes the listed tests, which serve as the acceptance criteria, and the reported outcome.

Acceptance Criteria (Individual Test Defined in ISO 80369-3 or other specified standard)Reported Device Performance
ISO 80369-3 and 80369-20 ENFit Connector Tests:
Fluid Leakage (Test Method Defined in ISO 80369-20 Annex C)Met acceptance criteria
Stress Cracking (Test Method Defined in ISO 80369-20 Annex E)Met acceptance criteria
Resistance to separation from axial load (Test Method Defined in ISO 80369-20 Annex F)Met acceptance criteria
Resistance to separation from unscrewing (Test Method Defined in ISO 80369-20 Annex G)Met acceptance criteria
Resistance to overriding (Test Method Defined in ISO 80369-20 Annex H)Met acceptance criteria
Disconnection by unscrewing (Test Method Defined in ISO 80369-20 Annex I)Met acceptance criteria
Dimensional Analysis:
Conformity to critical dimensions of ISO 80369-3Met acceptance criteria
Device Verification Testing for Piston Syringe (ISO 7886-1):
LubricantMet acceptance criteria
Limits of extractable metalsMet acceptance criteria
Limits for acidity/alkalinityMet acceptance criteria
Capacity Tolerance (Graduated Scale)Met acceptance criteria
Dead SpaceMet acceptance criteria
Air and liquid leakage Testing (Side Force/Axial Force)Met acceptance criteria
Biocompatibility Testing:
Meeting requirements for biocompatibilityMet acceptance criteria
Stability Testing:
Evaluation of properties after accelerated aging to establish 5-year shelf life for syringes and 3-5-year for accessoriesMet acceptance criteria
Confirmation of strength and integrity characteristics of sterile device packaging and sterile barrier maintenance for labeled shelf lifeMet acceptance criteria
Risk Management:
Assessment of risk associated with misconnection of the ENFit connectorRisk assessed and process captured
Additional testing for Low Dose Syringe tip design:
Low Dose Enteral Connector Misconnection AssessmentMet acceptance criteria
Usability Assessment for the Low Dose Syringe Tip DesignMet acceptance criteria
Low Dose Syringe Design Dose Accuracy TestingMet acceptance criteria

Details of the Study:

The studies conducted are primarily engineering and functional bench tests, and risk assessments.

  • Sample sized used for the test set and the data provenance: The document does not specify exact sample sizes for each test. The data provenance is implied to be from internal testing conducted by the manufacturer, Intervene Group Limited, likely in the UK, given the address. The studies are prospective in the sense that they were conducted for the purpose of this 510(k) submission.
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for these types of engineering tests is typically defined by adherence to published international standards (e.g., ISO) and the device's design specifications, not by expert consensus on observational data.
  • Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. Performance is measured against predefined objective standards and specifications.
  • If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/diagnostic device.
  • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI/diagnostic device.
  • The type of ground truth used (expert consensus, pathology, outcomes data, etc): The "ground truth" for these tests is the adherence to the requirements outlined in the cited ISO standards (ISO 80369-3, ISO 80369-20, ISO 7886-1, Draft ISO 20695) and the internal design specifications of the device.
  • The sample size for the training set: Not applicable. This is not a machine learning device; therefore, there is no "training set."
  • How the ground truth for the training set was established: Not applicable.

In conclusion, the study supporting the device's substantial equivalence consists of a series of bench tests and assessments confirming that the DASH 3™ ENFit Syringe family and its accessories meet established engineering and performance standards, thereby demonstrating that they perform as intended and do not raise new questions of safety or effectiveness compared to predicate devices.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 20, 2017

Intervene Group Limited Homer Trieu Regulatory Affairs Manager Russell Building, Brunel Science Park, Kingston Lane Uxbridge, UB8 3PQ GB

Re: K170371

Trade/Device Name: Dash 3 ENFit Syringe, Dash 3 Eccentric ENFit Syringe, Dash 3 ENFit Low Dose Tip Syringe, Dash 3 ENFit Syringe Cap. Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: Class II Product Code: PNR Dated: September 11, 2017 Received: September 25, 2017

Dear Homer Trieu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Joyce M. Whang -S

Benjamin R. Fisher, Ph.D. for Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170371

Device Name

DASH 374 ENFit Syringe, DASH 374 Eccentric ENFit Low Dose Tip Syninge, DASH 374 ENFit Syringe Cap

Indications for Use (Describe)

The DASH 3TM ENFit Syringe is a sugle use syninge indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids or nututional formula into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral synnges are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups.

The DASH 3TM Eccentric ENFit Syringe is a single use syringe indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids or nutritional formula into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syninges are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups.

The DASH 3TM Low Volume Tip ENFit Syringe is a sungle use syringe indicated for use as a dispenser, a measuning device, and a fluid transfer device. It is used to deliver fluids or nutritional formula into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syninges are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinicians, in all age groups.

Accessories

The DASH 3M ENFit Syringe Cap allows the advance preparation and secure storage and transport of medication fluids or mutritional formula. The synnge cap will fit any size of the Single Use Synnges (DASH 3TM)

Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including thetime to review instructions, search existing data sources, gather and maintain the data needed and completeand review the collection of information. Send comments regarding this burden estimate or any other aspectof this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SECTION 5 510(K) SUMMARY

This summary of the 510(k) premarket notification for the DASH 3™ is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

{4}------------------------------------------------

5.1 General Information

Applicant:

INTERVENE Group Ltd. Russell Building Brunel Science Park Kingston Lane Uxbridge, Middlesex UB8 3PQ UK Tel: + 44 (0)207 841 2734

Contact Person:

Homer Trieu Regulatory Affairs Manager, INTERVENE Group Limited Phone: +44 (0) 0207 841 2735 E-mail: htrieu@ivltd.co.uk Date Prepared: February 6, 2017

5.2 Device Information

Trade/Proprietary Name:

DASH 3™ ENFit Syringe, DASH 3™ Eccentric ENFit Syringe, DASH 3™ ENFit Low Dose Tip Syringe, DASH 3TM ENFit Syringe Cap.

Generic/Common Name:Enteral syringes with enteral specific connectors (syringes)Gastrointestinal tubes with enteral specific connectors (accessories)
Product Code:PNR (syringes)
Device Class and Panel:Class II, Gastroenterology and Urology
Classification Regulation:21 CFR§876.5980 - Gastrointestinal tube and accessories.

{5}------------------------------------------------

Predicate Device/Reference devices in the market 5.3

5.3.1 Predicate for Syringes

Primary Predicate:Centered Tip Syringeand Low Dose TipSyringeReference Device:Eccentric Syringe Tip
Trade Name:Monoject EnteralFeeding Syringe withENFit ConnectorNeoConnectOral/Enteral Syringeswith ENFit Connectors
Manufacturer andClearance Number:CovidienK161963K161045NeoMed, Inc.K161039
Product Code:PNRPNR
Classification Name:Enteral Syringes withEnteral SpecificConnectorsEnteral Syringes withEnteral SpecificConnectors

5.3.2 Predicate/Reference for Accessories

Predicate: Syringe Cap
Trade Name:NeoMed NeoConnect™ EnteralSyringes with ENFit™ Connectorand compatible NeoSecure™ TipCaps
Manufacturer andClearance Number:NeoMed, Inc.K152857
Product Code:PIF
ClassificationName:Gastrointestinal Tube and accessories

{6}------------------------------------------------

5.4 Indications for Use

Svringes

The DASH 3TM ENFit Syringe is a single use syringe indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids or nutritional formula into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syringes are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups.

The DASH 3TM Eccentric ENFit Syringe is a single use syringe indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids or nutritional formula into the gastrointestinal system who is physically unable to eat and swallow. The enteral syringes are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups.

The DASH 3™ Low Dose Tip ENFit Syringe is a single use syringe indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids or nutritional formula into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syringes are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups.

Accessories

The DASH 3TM ENFit Syringe Cap allows the advance preparation and secure storage and transport of medication/fluids or nutritional formula. The syringe cap will fit any size of the Single Use Syringes (DASH 3TM).

{7}------------------------------------------------

5.5 Product Description

The DASH 3™ ENFit Syringe device family is a Single Use, in-hospital and home care (DASH 3TM ENFit Syringe, DASH 3TM Eccentric ENFit Syringe, DASH 3TM ENFit Low Dose Tip Syringe) device.

It is provided in sizes from 1 mL to 100 mL. The device incorporates a female ENFit connector for connection to an enteral access device with a male ENFit port specified in ISO 80369-3. The Low Dose Tip contains the low dose design feature specified in (Draft) ISO 20695.

The DASH 3TM ENFit accessories are designed to be compatible with ENFit connector devices. The DASH 3TM ENFit Syringe Caps fit any size of the Single Use ENFit Syringes (DASH 3™), which allows the advance preparation and secure storage and transport of medication/fluids or nutritional formula.

5.6 Substantial Equivalence

The indications for use for the (primary) predicate and reference devices are substantially equivalent to the proposed indications for use for the DASH 3TM ENFit Syringe, DASH 3TM Eccentric ENFit Syringe, DASH 3TM ENFit Low Dose Tip Syringe and DASH 3TM ENFit Syringe Cap. Both device families have the same intended use and similar technological characteristics. Any differences in the technological characteristics between the devices do not raise any new issues of safety or effectiveness. These differences can be from non-critical dimensions (i.e. barrel wings, plunger dimension), additional sizes (i.e. 100ml Syringe), the color of the devices and packaging arrangements. Thus, the DASH 3TM ENFit Syringe device family and accessories are substantially equivalent to the predicate and reference devices.

5.7 Testing in Support of Substantial Equivalence Determination

All necessary bench testing was conducted on the DASH 3™ ENFit Syringe, DASH 3™ Eccentric ENFit Syringe, DASH 3™ ENFit Low Dose Tip Syringe and DASH 3™ ENFit Syringe Cap to support a determination of substantial equivalence to the (primary) predicate and reference devices. Testing included biocompatibility, sterilization validation, shipping and packaging, accelerated aging, and design verification testing. Design verification testing included the following:

  • Visual inspection
  • Dimensional verification
  • Physical testing

Performance testing was conducted to confirm compliance to the design specifications of ISO 80369-3, and also with reference to Draft ISO 20695 for the low dose tip; all functions have been verified to operate as designed. The DASH 3TM Syringe device

{8}------------------------------------------------

family, its female ENFit connector and its accessories (with male ENFit connectors) have met all acceptance criteria, as described in Section 17.

Individual Test Defined in ISO 80369-3Test Method Definedin ISO 80369-20
Fluid LeakageAnnex C
Stress CrackingAnnex E
Resistance to separation from axial loadAnnex F
Resistance to separation from unscrewingAnnex G
Resistance to overridingAnnex H
Disconnection by unscrewingAnnex I

Tests Defined in ISO 80369-3 and 80369-20 for ENFit Connectors

Dimensional analysis was conducted for critical dimensions of the DASH 3TM ENFit Syringe device family and its accessories (with ENFit connectors), in accordance with the criteria in ISO 80369-3.

Device Verification Testing for the piston syringe was conducted in accordance with ISO 7886-1. This includes:

  • Lubricant. .
  • . Limits of extractable metals,
  • . Limits for acidity/alkalinity,
  • . Capacity Tolerance (Graduated Scale),
  • . Dead Space
  • Air and liquid leakage Testing (Side Force/Axial Force.)

Biocompatibility testing has demonstrated that the DASH 3™ Syringe and its accessories meet the requirements for biocompatibility.

Stability testing evaluated the properties of the DASH 3™ Syringe and its accessories after accelerated aging to establish a 5-year shelf life for the syringes and a 3 to 5-year shelf life for the accessories and to confirm the strength and integrity characteristics of the sterile device packaging and the ability of the packaging to maintain a sterile barrier for the duration of the labeled shelf life.

The risk associated with misconnection of the ENFit connector has been assessed and the process captured in the following documents:

  • FMEA (Design, User, Process)
  • . Risk Management Report
  • . Enteral Connector Misconnection Assessment (As recommended in FDA Enteral Connector Guidance)

{9}------------------------------------------------

Additional testing for the Low Dose Syringe tip design includes:

  • . Low Dose Enteral Connector Misconnection Assessment (As recommended in FDA Enteral Connector Guidance)
  • . Usability Assessment for the Low Dose Syringe Tip Design
  • Low Dose Syringe Design Dose Accuracy Testing

The collective results of the testing demonstrate that the DASH 3™ ENFit Syringe device family and accessories meet specifications and perform as intended. In addition, the collective bench testing demonstrates that the DASH 3™ ENFit Syringe device family and accessories do not raise new questions of safety or effectiveness as compared to the predicate or reference devices.

5.8 Conclusion

The DASH 3™ ENFit Syringe family and accessories have the same intended use and similar technological characteristics as the predicate devices. The DASH 3TM ENFit Syringe family and accessories have been tested to ensure that they perform as intended and do not raise new questions of safety or effectiveness. As such, the DASH 3™ ENFit Syringe device family and accessories are substantially equivalent to the predicate and reference devices.

§ 876.5980 Gastrointestinal tube and accessories.

(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.