K Number
K170900

Validate with FDA (Live)

Date Cleared
2017-11-09

(227 days)

Product Code
Regulation Number
880.5860
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DASH 6® NRFit Lock syringe and DASH® NRFit Slip Syringe are intended to be used with ISO 80369-6 NRFit neuraxial compliant devices for administration of neuraxial medication.

The DASH 6® NRFit Plastic LOR Device is intended to be used with ISO 80369-6 NRFit neuraxial compliant epidural needle for locating the epidural space.

Accessories:

DASH 6® Introducer Needle is intended to be used with ISO 80369-6 NRFit neuraxial compliant spinal needle for guiding the placement of the spinal needle into the arachnoid/epidural space.

DASH 6@ NRFit Syringe Cap is intended to be used with ISO 80369-6 NRFit neuraxial syringes for sealing the tip of the syringe.

DASH 6® NRFit Drawing Up Filter Straw are intended to be used with ISO 80369-6 NRFit neuraxial syringe for the drawing up of neuraxial medication and anesthetic.

DASH 6@ NRFit Blunt Drawing Up Needle (with and without filter) is intended to be used with ISO 80369-6 NRFit neuraxial syringe for the drawing up of neuraxial medication and anesthetic.

DASH 6® NRFit Bacterial Disc Filter is intended to be used with ISO 80369-6 NRFit neuraxial compliant devices to ensure aseptic administration of neuraxial medication and anesthetic.

DASH 6® NRFit Epidural Flat Filter is intended to be used with ISO 80369-6 NRFit neuraxial compliant devices to ensure aseptic administration of neuraxial medication and anesthetic.

DASH 6® NRFit Tuohy-Borst Adapter is intended to be used with a epidural catheter to provide an ISO 80369-6 NRFit compliant connection.

DASH 6® NRFit Syringe to Syringe Adapter is intended to be used with ISO 80369-6 NRFit neuraxial syringe to allow for mixing/transfering medication between two syringes.

DASH 6® NRFit Needle Hub Cap is intended for sealing the hub of an NRFit compliant needle.

DASH 6@ NRFit Epidural Catheter Feeder is intended to be used with ISO 80369-6 NRFit neuraxial compliant devices to assist the insertion of an epidural needle into the epidural space.

Device Description

The DASH 6® NRFit Syringes (Lock and Slip Syringes) are Single Use, in-hospital devices. They are provided in sizes ranging from 1 mL to 60 mL. The devices incorporate a male NRFit connector for connection to a female NRFit port. The syringes are designed according to ISO 7886-1 & ISO 80369-6 standard. The syringes are used for neuraxial purposes (ISO 80369-6 NRFit connection) and functions exactly the same as any hypodermic syringes that are on the market (ISO 7886-1).

The syringes will be supplied as individually packed (sterile) and the bulk packed (Nonsterile). The Sterile packed items will be supplied directly to the user and the bulk packed will be supplied to Anesthetic Conduction Kit manufacturers to be packaged into kit.

The syringe packaging will indicate the volume/size, the connector type (Lock or Slip variants) that the syringe contains and the sterility/sterilization method the syringe have been processed.

The DASH 6 ° Plastic LOR devices are Single Use, in-hospital devices. It is provided with a 10 mL size/volume and incorporates a male ISO 80369-6 NRFit slip connector to connect to a female ISO 80369-6 NRFit Port.

The Plastic LOR devices are designed according to ISO 80369-6 standard. The syringes are used for neuraxial purposes (ISO 80369-6 NRFit connection) and functions exactly the same as the LOR devices on the market.

The device will be supplied as individually packed (sterile) and the bulk packed (Nonsterile). The Sterile packed items will be supplied directly to the user and the bulk packed will be supplied to Anesthetic Conduction Kit manufacturers to be packaged into kit.

The devices packaging will indicate the volume/size, the connector type (Slip) that the device contains and the sterility/sterilization method the device have been processed.

Accessories:
DASH 6® NRFit Syringe Caps: Single Use, in-hospital devices with a female ISO 80369-6 NRFit connector.
DASH 6® NRFit Drawing up Filter Straw: Single Use, in-hospital devices with a female ISO 80369-6 NRFit connector and filter.
DASH 6 ° NRFit Blunt Drawing up Needle (with and without filter): Single Use, in-hospital devices with a female ISO 80369-6 NRFit connector and blunt needle (with or without filter).
DASH 6® NRFit Bacterial Disc Filter: Single Use, in-hospital devices with male & female ISO 80369-6 NRFit connectors and a 0.22µm filter.
DASH 6® NRFit Epidural Flat Filter: Single Use, in-hospital devices with male & female ISO 80369-6 NRFit connectors and a 0.22um filter.
DASH 6 ° NRFit Tuohy Borst Adapter: Single Use, in-hospital devices with a female ISO 80369-6 NRFit connector and open catheter port.
DASH 6® NRFit Needle Hub Caps: Single Use, in-hospital devices with a Male ISO 80369-6 NRFit connector.
DASH 6® Epidural Catheter Feeder: Single Use, in-hospital devices with a compatible male ISO 80369-6 NRFit connector.

AI/ML Overview

The provided text is a 510(k) Premarket Notification for the DASH 6® NRFit System, which includes syringes and various accessories. This document primarily focuses on demonstrating substantial equivalence to predicate devices through design verification and compliance with specific ISO standards. It does not describe an AI medical device or a study involving human readers or AI assistance. Therefore, I cannot provide information on acceptance criteria and a study that proves the device meets those criteria in the context of AI performance, MRMC studies, standalone algorithm performance, number of experts, or ground truth for AI model development.

However, I can extract the acceptance criteria and the type of study conducted to demonstrate the device's technical specifications and substantial equivalence, as presented in the document:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the DASH 6® NRFit System are primarily defined by compliance with several ISO standards. The studies conducted were design verification tests demonstrating compliance with these standards.

Acceptance Criteria (Defined by ISO Standards)Reported Device Performance (as demonstrated by testing)
ISO 80369-6: NRFit Connectors
- Fluid Leakage (per ISO 80369-20 Annex C)Met all acceptance criteria. Functions verified to operate as designed.
- Stress Cracking (per ISO 80369-20 Annex E)Met all acceptance criteria. Functions verified to operate as designed.
- Resistance to separation from axial load (per ISO 80369-20 Annex F)Met all acceptance criteria. Functions verified to operate as designed.
- Resistance to separation from unscrewing (per ISO 80369-20 Annex G)Met all acceptance criteria. Functions verified to operate as designed.
- Resistance to overriding (per ISO 80369-20 Annex H)Met all acceptance criteria. Functions verified to operate as designed.
- Dimensional analysis for critical dimensionsMet all acceptance criteria for critical dimensions in accordance with ISO 80369-6.
ISO 7886-1: Sterile hypodermic syringes for single use (for DASH 6 Syringes)
- Capacity ToleranceDevice Verification Testing conducted according to ISO 7886-1. Functions verified to operate as designed.
- Leakage Testing Side ForceDevice Verification Testing conducted according to ISO 7886-1. Functions verified to operate as designed.
- Leakage Testing Axial ForceDevice Verification Testing conducted according to ISO 7886-1. Functions verified to operate as designed.
ISO 10993 Series (Biocompatibility)
- Cytotoxicity (ISO 10993-5)Testing done per ISO 10993-5. (Implied compliance as part of "All necessary bench testing was conducted... to support a determination of substantial equivalence").
- Sensitization (ISO 10993-10)Testing done per ISO 10993-10. (Implied compliance as part of "All necessary bench testing was conducted...").
- Extractables and Leachables (ISO 10993-18 and 10993-19)Testing done per ISO 10993-18 and 10993-19. (Implied compliance as part of "All necessary bench testing was conducted...").
ISO 7864-1: Sterile hypodermic needles for single use (for DASH 6 NRFit needles - in certain respects)
- Needle to Needle Hub bondingDesign verification testing conducted in accordance with ISO 7864-1 (in certain respects). Functions verified to operate as designed.
- Dimensional tolerancesDesign verification testing conducted in accordance with ISO 7864-1 (in certain respects). Functions verified to operate as designed.
- Blockages within the needleDesign verification testing conducted in accordance with ISO 7864-1 (in certain respects). Functions verified to operate as designed.
Filter Device Specific Tests
- Bubble point testDesign verification testing completed to ensure functioning according to requirement.
- Flow-rateDesign verification testing completed to ensure functioning according to requirement.
- Filtration efficiencyDesign verification testing completed to ensure functioning according to requirement.
Other General Tests
- Sterilization ValidationConducted. (Implied compliance as part of "All necessary bench testing was conducted...").
- Shipping and PackagingConducted. (Implied compliance as part of "All necessary bench testing was conducted...").
- Accelerated AgingConducted. (Implied compliance as part of "All necessary bench testing was conducted...").
- Risk Assessment (FMEA, Risk Management Report)Risk associated with misconnection of the NRFit connector assessed and captured. (Implied that mitigation efforts and assessment meet acceptable risk levels).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document describes bench testing for physical characteristics, dimensional verification, and biocompatibility, not a clinical study with a "test set" in the context of AI or diagnostic performance. Therefore, typical sample size or data provenance details for medical image analysis are not applicable. The testing was conducted to verify compliance with engineering standards (ISO standards). The document does not specify the exact number of units tested for each specific test, but mentions "All necessary bench testing was conducted."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable as the device is a physical medical device (syringes, needles, etc.) and its performance is evaluated against engineering and biological standards, not diagnostic interpretations from experts. Ground truth is established by objective measurements against specified engineering and material science criteria.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This device is not an AI diagnostic tool and does not involve human adjudication of results in the traditional sense of a clinical study or image interpretation.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The DASH 6® NRFit System is a medical device (syringes, needles, etc.) and not an AI-powered diagnostic or assistive tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This device is not an algorithm or AI system. Its performance is evaluated intrinsically based on its physical properties and adherence to manufacturing and safety standards.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's evaluation is primarily the specifications and requirements defined by international standards (e.g., ISO 80369-6, ISO 7886-1, ISO 10993 series) and the manufacturer's own design specifications. It is based on objective, quantifiable physical and chemical measurements (e.g., fluid leakage, dimensions, material compatibility).

8. The sample size for the training set

Not applicable. There is no "training set" as this is a physical medical device, not a machine learning model.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 9, 2017

INTERVENE Group Limited Homer Trieu Regulatory Affairs Manager Russell Building, Brunel Science Park Kington Lane, Uxbridge Middlesex UB8 3PQ UNITED KINGDOM

Re: K170900

Trade/Device Name: DASH 6 ° NRFit: Lock Syringe, Plastic LOR Device, Introducer Needle, Syringe Caps, Drawing Up Filter Straw, Blunt Drawing Up Needle (with and without filter), Bacterial Disc Filter, Epidural Flat Filter, Tuohy Borst Adapter, Syringe-Syringe adapter, Needle Hub Cap, Epidural Catheter Feeder Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Svringe Regulatory Class: Class II Product Code: FMF, BSN, BSO, BSP Dated: October 6, 2017 Received: October 10, 2017

Dear Homer Trieu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Tina
Kiang-S

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170900

Device Name

DASH 6@ NRFit: Lock Syringe. Plastic LOR Device. Introducer Needle. Svringe Caps, Drawing Up Filter Straw, Blunt Drawing Up Needle (with and without filter), Bacterial Disc Filter, Tuohy Borst Adapter, Syringe-Syringe adapter, Needle Hub Cap, Epidural Catheter Feeder.

Indications for Use (Describe)

The DASH 6® NRFit Lock syringe and DASH® NRFit Slip Syringe are intended to be used with ISO 80369-6 NRFit neuraxial compliant devices for administration of neuraxial medication.

The DASH 6® NRFit Plastic LOR Device is intended to be used with ISO 80369-6 NRFit neuraxial compliant epidural needle for locating the epidural space.

Accessories:

DASH 6® Introducer Needle is intended to be used with ISO 80369-6 NRFit neuraxial compliant spinal needle for guiding the placement of the spinal needle into the arachnoid/epidural space.

DASH 6@ NRFit Syringe Cap is intended to be used with ISO 80369-6 NRFit neuraxial syringes for sealing the tip of the syringe.

DASH 6® NRFit Drawing Up Filter Straw are intended to be used with ISO 80369-6 NRFit neuraxial syringe for the drawing up of neuraxial medication and anesthetic.

DASH 6@ NRFit Blunt Drawing Up Needle (with and without filter) is intended to be used with ISO 80369-6 NRFit neuraxial syringe for the drawing up of neuraxial medication and anesthetic.

DASH 6® NRFit Bacterial Disc Filter is intended to be used with ISO 80369-6 NRFit neuraxial compliant devices to ensure aseptic administration of neuraxial medication and anesthetic.

DASH 6@ NRFit Epidural Flat Filter is intended to be used with ISO 80369-6 NRFit neuraxial compliant devices to ensure aseptic administration of neuraxial medication and anesthetic.

DASH 6® NRFit Tuohy-Borst Adapter is intended to be used with a epidural catheter to provide an ISO 80369-6 NRFit compliant connection.

DASH 6® NRFit Syringe to Syringe Adapter is intended to be used with ISO 80369-6 NRFit neuraxial syringe to allow for mixing/transfering medication between two syringes.

DASH 6® NRFit Needle Hub Cap is intended for sealing the hub of an NRFit compliant needle.

DASH 6@ NRFit Epidural Catheter Feeder is intended to be used with ISO 80369-6 NRFit neuraxial compliant devices to assist the insertion of an epidural needle into the epidural space.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) SUMMARY SECTION 5.

This summary of the 510(k) premarket notification for the DASH 6° is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

{5}------------------------------------------------

510(k) Notification K

5.1 GENERAL INFORMATION

Applicant:

INTERVENE Group Limited Russell Building Brunel Science Park Kingston Lane Uxbridge, Middlesex UB8 3PQ UK Tel: + 44 (0)1895 347 568

Contact Person:

Homer Trieu Regulatory Affairs Manager, INTERVENE Group Ltd Phone: +44 (0)1895 347 568 E-mail: htrieu@jvltd.co.uk

Date Prepared: March 27, 2017

5.2 DEVICE INFORMATION

Trade/Proprietary Name:

DASH 6® NRFit: Lock Syringe, Slip Syringe, Plastic LOR Device, Introducer Needle, Syringe Caps, Drawing Up Filter Straw, Blunt Drawing Up Needle (with and without filter), Bacterial Disc Filter, Epidural Flat Filter, Tuohy Borst Adapter, Syringe-Syringe adapter, Needle Hub Cap, Epidural Catheter Feeder

Generic/Common Name:Syringe Piston
Product Code:FMF
Device Class and Panel:Class II, General Hospital
Classification Regulation:21 CFR§880.5860
Generic/Common Name:Catheter, Conduction, Anesthetic
Product Code:BSO
Device Class and Panel:Class II, Anesthesiology

{6}------------------------------------------------

INTERVENE Group Limited DASH 6® NRFit SYSTEM (DASH 6®) 510(k) Premarket Notification

Generic/Common Name:Needle, Conduction, Anesthetic (W/WO Introducer)
Product Code:BSP
Device Class and Panel:Class II, Anesthesiology
Generic/Common Name:Filter, Conduction, Anesthetic
Product Code:BSN
Device Class and Panel:Class II, Anesthesiology

5.3 PREDICATE DEVICE(S) SYRINGE COMPONENTS

Primary Predicate:Lock/Slip SyringeReference Device:NRFit ConnectorReference Device:Lock/Syringe
Trade NameBecton DickinsonHypodermic SyringeTuohy NRFitCorrectInject safety System
Manufacturer andClearance NumberBecton, Dickensonand CompanyK151766PAJUNK GmbHMedizintechnologieK160297 {Modificationof PAJUNK TuohyAnesthesia ConductionNeedles [K040965] toinclude NRFit connect-ors}Smiths MedicalK110053
Product CodeFMFBSPCAZ
ClassificationNameSyringe, PistonAnesthesia NeedleAnesthesia Conduction Kit
Primary Predicate:LOR SyringeReference Device:NRFit ConnectorReference Device:Plastic LOR
Trade NameSpectra-LOR SyringeTuohy NRFitBusse Plastic LOR Syringe
Manufacturer andClearance NumberSpectra medicaldevice, Inc.K081524PAJUNK GmbHMedizintechnologieK160297 {Modificationof PAJUNK TuohyAnesthesia ConductionNeedles [K040965] toinclude NRFit connect-ors}Smiths MedicalK110053
Product CodeFMFBSPCAZ
ClassificationNameSyringe, PistonAnesthesia NeedleAnesthesia Conduction Kit

{7}------------------------------------------------

INTERVENE Group Limited DASH 6® NRFit SYSTEM (DASH 6®) 510(k) Premarket Notification

5.4 PREDICATE DEVICE(S) ACCESSORIES

Predicate Device:Bacterial Disc FilterPredicate Device:Drawing Up Straw with FilterPredicate Device:Epidural Catheter FeederPredicate Device:Introducer Needle 19GPredicate Device: TuohyBorst Adapter
Trade NamePortex Epidural Filter5 micron straw filtercontained in the ArrowFlextip Plus Closed TipEpidural Catheter KitThread assist guide containedin the Busse Epidural CatheterKitSpinal Introducer Needlecontained in the B BraunSpinal Needle TrayBD Durasafe VariableExtension Combined SpinalEpidural CSE Needle Set/Kit
Manufacturerand ClearanceNumberSmiths Medical ASD, Inc.K083451ARROW International, Inc.K161075, K143581Busse Hospital DisposablesK093920B. Braun Medical Inc.K112515Beckton, Dickinson andCompanyK954953
Product CodeBSNCAZBSOBSPCAZ
ClassificationNameAnesthesia Conduction Kit1) Needle, Aspiration andinjection, disposable2) Anesthesia Conduction KitAnesthesia ConductionCatheterNeedle, conduction,anesthetic (w/wointroducer)Anesthesia Conduction kit

{8}------------------------------------------------

INTERVENE Group Limited DASH 6® NRFit SYSTEM (DASH 6®) 510(k) Premarket Notification

Predicate Device:Syringe Cap/Needle HubCapsPredicate Device:Blunt Drawing up Needle(with and without filter)
Trade NameCorrectInject SafetySystem(Caps)1) CorrectInject SafetySystem (Filter NeedleAssembly)2) B-D Twinpack Syringefilling device.
Manufacturerand ClearanceNumberSmiths Medical ASD, Inc.K1100531) Smiths Medical ASD,Inc. K1100532) Becton Dickinson.K974006
Product CodeCAZ1) CAZ2) FMI
ClassificationNameAnesthesia Conduction Kit1) Anesthesia ConductionKit2) Needle, Hyperdermic,Single Lumen.

{9}------------------------------------------------

ર.5 INDICATIONS FOR USE

5.5.1 DASH 6 Primary Device

5.5.1.1 DASH 6 Syringe

The DASH 60 NRFit Lock Syringe and DASH 60 NRFit Slip Syringes are intended to be used with ISO 80369-6 NRFit™ neuraxial compliant devices for administration of neuraxial medication or anesthetic.

5.5.1.2 DASH 6 LOR Device

The DASH 6 ° NRFit Plastic LOR Device is intended to be used with ISO 80369-6 NRFit " neuraxial compliant epidural needle for locating the epidural space.

DASH® 6 Accessories 5.5.2

DASH 6 Introducer Needle 5.5.2.1

The DASH 6® Introducer Needle is intended to be used with ISO 80369-6 NRFit™ neuraxial compliant spinal needle for guiding the placement of the spinal needle into the arachnoid/epidural space.

5.5.2.2 DASH 6® NRFit Svringe Cap

The DASH 6 ° NRFit Syringe Cap is intended to be used with ISO 80369-6 NRFit™ neuraxial syringes for sealing the tip of the syringe.

DASH 60 NRFit Drawing Up Filter Straw 5.5.2.3

DASH 6® NRFit Drawing up Straws are intended to be used with ISO 80369-6 NRFit™ neuraxial syringe for the drawing up of neuraxial medication and anesthetic.

DASH 6 % NRFit Blunt Drawing Up Needle (with and without filter) 5.5.2.4

DASH 6 ° NRFit Blunt Drawing Up Needle is intended to be used with ISO 80369-6 NRFit™ neuraxial syringe for the drawing up of neuraxial medication and anesthetic.

5.5.2.5 DASH 6 ° NRFit Bacterial Disc Filter

The DASH 6® Bacterial Filter is intended to be used with ISO 80369-6 NRFit™ neuraxial compliant devices to ensure aseptic administration of neuraxial medication and anesthetic.

5.5.2.6 DASH 6 NRFit Epidural Flat Filter

The DASH 6® Bacterial Filter is intended to be used with ISO 80369-6 NRFit™ neuraxial compliant devices to ensure aseptic administration of neuraxial medication and anesthetic.

{10}------------------------------------------------

DASH 6® NRFit Tuohy-Borst Adapter 5.5.2.7

The DASH 6 Tuohy-Borst Adapter is intended to be used with a epidural catheter to provide an ISO 80369-6 NRFit" compliant connection.

5.5.2.8 DASH 6° NRFit Syringe to Syringe Adapter

The DASH 6 syringe to syringe adapter is intended to be used with ISO 80369-6 NRFit™ neuraxial syringe to allow for mixing/transfering medication between two syringes.

DASH 6® NRFit Needle Hub Cap 5.5.2.9

The DASH 6® Needle Hub Cap is intended for sealing the hub of an NRFit™ compliant needle.

5.5.2.10 DASH 6° NRFit Epidural Catheter Feeder

The DASH 6 Epidural Catheter Feeder is intended to be used with ISO 80369-6 NRFit™ neuraxial compliant devices to assist the insertion of an epidural needle into the epidural space.

PRODUCT DESCRIPTION 5.6

Primary Devices

DASH 6 Syringes

The DASH 6® NRFit Syringes (Lock and Slip Syringes) are Single Use, in-hospital devices. They are provided in sizes ranging from 1 mL to 60 mL. The devices incorporate a male NRFit connector for connection to a female NRFit port. The syringes are designed according to ISO 7886-1 & ISO 80369-6 standard. The syringes are used for neuraxial purposes (ISO 80369-6 NRFit connection) and functions exactly the same as any hypodermic syringes that are on the market (ISO 7886-1).

The syringes will be supplied as individually packed (sterile) and the bulk packed (Nonsterile). The Sterile packed items will be supplied directly to the user and the bulk packed will be supplied to Anesthetic Conduction Kit manufacturers to be packaged into kit.

The syringe packaging will indicate the volume/size, the connector type (Lock or Slip variants) that the syringe contains and the sterility/sterilization method the syringe have been processed.

DASH 6 Plastic LOR Devices

The DASH 6 ° Plastic LOR devices are Single Use, in-hospital devices. It is provided with a 10 mL size/volume and incorporates a male ISO 80369-6 NRFit slip connector to connect to a female ISO 80369-6 NRFit Port.

{11}------------------------------------------------

The Plastic LOR devices are designed according to ISO 80369-6 standard. The syringes are used for neuraxial purposes (ISO 80369-6 NRFit connection) and functions exactly the same as the LOR devices on the market.

The device will be supplied as individually packed (sterile) and the bulk packed (Nonsterile). The Sterile packed items will be supplied directly to the user and the bulk packed will be supplied to Anesthetic Conduction Kit manufacturers to be packaged into kit.

The devices packaging will indicate the volume/size, the connector type (Slip) that the device contains and the sterility/sterilization method the device have been processed.

Accessories

DASH 6® NRFit Syringe Caps

The DASH 6® NRFit Syringe Cap are Single Use, in-hospital devices. It incorporates a female ISO 80369-6 NRFit connector to be connected on an ISO 80369-6 compliant male port.

The device will be supplied as individually packed (sterile) and the bulk packed (Nonsterile). The Sterile packed items will be supplied directly to the user and the bulk packed will be supplied to Anesthetic Conduction Kit manufacturers to be packaged into kit.

The devices packaging will indicate it is a syringe cap and the sterility/sterilization method the device have undergone.

DASH 6® NRFit Drawing up Filter Straw

The DASH 6 ° NRFit Drawing up Filter Straw are Single Use, in-hospital devices. It incorporates a female ISO 80369-6 NRFit connector to be connected on an ISO 80369-6 compliant NRFit syringe to draw up neuraxial medication and anesthetic.

The device will be supplied as individually packed (sterile) and the bulk packed (Nonsterile). The Sterile packed items will be supplied directly to the user and the bulk packed will be supplied to Anesthetic Conduction Kit manufacturers to be packaged into kit.

The devices packaging will indicate it is a drawing up filter straw and the sterility/sterilization method the device have undergone.

DASH 6 ° NRFit Blunt Drawing up Needle (with and without filter)

The DASH 6 ° NRFit Blunt Drawing up Needle (with and without filter) are Single Use, in-hospital devices. It incorporates a female ISO 80369-6 NRFit connector to be connected on an ISO 80369-6 compliant NRFit syringe to draw up neuraxial medication and anesthetic.

The device will be supplied as individually packed (sterile) and the bulk packed (Nonsterile). The Sterile packed items will be supplied directly to the user and the bulk packed will be supplied to Anesthetic Conduction Kit manufacturers to be packaged into kit.

{12}------------------------------------------------

The devices packaging will indicate it is a Blunt Drawing up Needle (with or without filter) and the sterility/sterilization method the device have undergone.

DASH 6® NRFit Bacterial Disc Filter

The DASH 6® NRFit Bacterial Disc Filter is Single Use, in-hospital devices. It incorporates a male & female ISO 80369-6 NRFit connector on the device and a 0.22µm filter in the middle of the housing. The device is to be connected to a female & male ISO 80369-6 NRFit compliant device to ensure aseptic administration of neuraxial medication and anesthetic.

The device will be supplied as individually packed (sterile). The Sterile packed items will be supplied directly to the hospitals.

The devices packaging will indicate it is a bacterial disc filter and the sterility/sterilization method the device have undergone.

DASH 6® NRFit Epidural Flat Filter

The DASH 6® NRFit Epidural Flat Filter is Single Use, in-hospital devices. It incorporates a male & female ISO 80369-6 NRFit connector on the device and a 0.22um filter in the middle of the housing. The device is to be connected to a female & male ISO 80369-6 NRFit compliant device to ensure aseptic administration of neuraxial medication and anesthetic.

The device will be supplied as individually packed (sterile) and the bulk packed (Nonsterile). The Sterile packed items will be supplied directly to the user and the bulk packed will be supplied to Anesthetic Conduction Kit manufacturers to be packaged into kit.

The devices packaging will indicate it is an Epidural Flat Filter and the sterility/sterilization method the device have undergone.

DASH 6® NRFit Tuohy Borst Adapter

The DASH 6 ° NRFit Tuohy Borst Adapter is Single Use, in-hospital devices. It incorporates a female ISO 80369-6 NRFit connector to be connected on an ISO 80369-6 NRFit compliant male port and open catheter port, which will allow a epidural catheter to provide an ISO 80369-6 NRFit™ compliant connection.

The device will be supplied as bulk packed (Non-sterile). The bulk packed will be supplied to Anesthetic Conduction Kit manufacturers to be packaged into kit.

The devices packaging will indicate it is a Tuohy borst Adapter and the sterility/sterilization method the device have undergone.

DASH 6® NRFit Needle Hub Caps

{13}------------------------------------------------

The DASH 6® NRFit Needle Hub Cap are Single Use, in-hospital devices. It incorporates a Male ISO 80369-6 NRFit connector to be connected on an ISO 80369-6 compliant female port.

The device will be supplied as individually packed (sterile) and the bulk packed (Nonsterile). The Sterile packed items will be supplied directly to the user and the bulk packed will be supplied to Anesthetic Conduction Kit manufacturers to be packaged into kit.

The devices packaging will indicate it is a Needle Hub cap and the sterility/sterilization method the device have undergone.

DASH 6® Epidural Catheter Feeder.

The DASH 6® Epidural Catheter Feeder are Single Use, in-hospital devices. It incorporates a compatible male ISO 80369-6 NRFit connector to be connected on an ISO 80369-6 compliant female port.

The device will be supplied as bulk packed (Non-sterile). The bulk packed will be supplied to Anesthetic Conduction Kit manufacturers to be packaged into kit.

The devices packaging will indicate it is an Epidural Catheter Feeder and the sterility/sterilization method the device have undergone.

5.6.1 Substantial Equivalence

The indications for use for the (primary) predicates and associated accessories are substantially equivalent to the proposed indications for use for the DASH 6 ° range of devices and accessories.

The DASH 6 NRFit syringes, BD hypodermic syringes and the syringes placed in the CorrectIniect Safety System Kit have the same intended use, operate similarly and have similar materials. The only difference is the technological characteristics of the connection system. The DASH 60 NRFit Syringe is in accordance to ISO 80369-6 compliance and is specified for Neuraxial purposed only. Pajunk Tuohy NRFit Anesthetic Conduction needle is reference to show that the DASH 6° NRFit syringes are developed to comply with ISO 80369-6 NRfit Standard and for this used within the regulation of Anesthesiology.

The DASH 6 NRFit LOR Devices are equivalent to that of the Specta-LOR syringes and Busse Plastic LOR syringe. The LOR device operates the same way as the predicate devices and has the same intended use. The only difference is the technological characteristics of the connection system.

{14}------------------------------------------------

The DASH 6® NRFit Syringe Cap and DASH 6® NRFit Needle Hub Cap are equivalent to that of the Caps supplied in the CorrectInject kit. The accessory operates the same way as the predicate devices and has the same intended use. The only difference is the connection system on the caps.

The DASH 6® NRFit Drawing Up Filter Straw are equivalent to that of B.Braun Filter Straw is currently sold in the US market. Please find the attached link (https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRL/r1.cfm?lid=186755&lpcd=GAA). The filter straw is a class I device itself are exempt from 510k. The devices operates exactly the same way as the predicate device. The 5 Micron Straw Filter listed in the Arrow Epidural Catheter Kit is also used as a reference predicate as the filter straw will be packaged in to an Anesthetic conduction kit. The only difference is the connection system on the connector of the straw hub.

The DASH 6® NRFit Blunt Drawing Up Needle (with and without filter) are equivalent to that of the filter needle supplied in the CorrectInject kit and the twin Pack Syringe filling device supplied by BD. The blunt needle operates the same way as the predicate devices and has the same intended use. The only difference is the connection system on the connector of the needle hub.

The DASH 6° NRFit Bacterial Disc Filter and Epidural Flat Filter are equivalent to that of the Portex supplied in the CorrectInject kit. The filters operated exactly the same way as the predicate devices and has the same intended use. The only difference is the connection system on the housing.

The DASH 6 ° NRFit Tuohy-Borst Adapter are equivalent to that of the Tuohy borst adaptor supplied in the EpiLong set for Anesthesiology, supplied by Pajunk. The Tuohy borst adapter operated exactly the same way as the predicate devices and has the same intended use. The only difference is the connection system.

DASH 6 ° NRFit Syringe to Syringe Adapter are a Class I accessory. The connection system contains two female ISO 80369-6 NRFit connectors.

DASH 6 ° NRFit Epidural Catheter Feeder are equivalent to that of the Thread assist guide contained in the Busse Epidural Catheter Kit. The Epdirual operated exactly the same way as the predicate devices and has the same intended use. The only difference is the connection system on the accessory.

{15}------------------------------------------------

The differences in the technological characteristics between the devices and accessories do not raise any new issues of safety or effectiveness. Thus, the DASH 6° NRFit Syringe device family and accessories (DASH 6 ° NRFit Syringe Cap, DASH 6 ° NRFit Drawing Up Straw with Filter, DASH 6® NRFit Blunt Drawing Up Needle (with and without filter), DASH 6° NRFit Bacterial Disc Filter, DASH 6° NRFit Epidural Flat Filter, DASH 68 NRFit Tuohy-Borst Adapter, DASH 66 NRFit Syringe to Syringe Adapter, DASH 6® NRFit Needle Hub Cap, DASH 6® NRFit Epidural Catheter Feeder) are substantially equivalent to their predicate devices.

5.6.2 Testing in Support of Substantial Equivalence Determination

All necessary bench testing was conducted on the DASH 6° Family of devices to support a determination of substantial equivalence to the (primary) predicate and the associated accessories. Testing included biocompatibility, sterilization validation, shipping and packaging, accelerated aging, and design verification testing. Design verification testing included the following:

  • Visual inspection
  • Dimensional verification
  • Physical testing

Performance testing was conducted to confirm compliance to the design specifications of ISO 80369-6; all functions have been verified to operate as designed. The DASH 6 Syringe device family, its male NRFit connector and its accessories (with female NRFit connectors) have met all acceptance criteria, as described in Section 18.

5.6.3 Tests Defined in ISO 80369-6 and 80369-20 for NRFit Connectors

Individual Test Defined inISO 80369-6Test Method Definedin ISO 80369-20
Fluid LeakageAnnex C
Stress CrackingAnnex E
Resistance to separation from axialloadAnnex F
Resistance to separation fromunscrewingAnnex G
Resistance to overridingAnnex H

{16}------------------------------------------------

Dimensional analysis was conducted for critical dimensions of the DASH 6® NRFit Syringe device family and its accessories (with NRFit connectors), in accordance with the criteria in ISO 80369-6.

Device Verification Testing for the piston syringe was conducted in accordance with ISO 7886-1 (Capacity Tolerance, Leakage Testing Side Force, Leakage Testing Axial Force). Cytotoxicity, Sensitization, and Extractables and Leachables testing was also done per ISO 10993-5, 10993-10, and 10993-18 and 10993-19, respectively.

The DASH 6 NRFit plastic LOR device itself did not undergo testing according to due to its intended use. The LOR device is a low friction device and it is used to locate the epidural space. The requirement for low friction and integrity of the seal from the gasket to the barrel is therefore compromised to produce the required clinical function to enable the physician to locate the epidural space. As such, to test the performance for "leakage and resistance" for the LOR device will not be testable under the regime of ISO 7886-1 as it is not applicable.

Design verification testing for the needles were conducted in accordance with ISO 7864-1 but only in certain respects (i.e. Needle to Needle Hub bonding, dimensions tolerances and blockages within the needle). The reason for this is that the DASH 6 NRFit needles are not intend for administering medication. Therefore, certain characteristic aspect of specified in ISO 7864-1, were taken into account for these needles.

Design verification testing were completed on the filter device to ensure that the device is also functioning according to it requirement. This includes the bubble point test, flow-rate and filtration efficiency.

The risk associated with misconnection of the NRFit connector has been assessed and the process captured in the following documents:

  • FMEA (Design, User, Process)
  • Risk Management Report

The collective results of the testing demonstrate that the DASH 6® NRFit Syringe device family and accessories meet specifications and perform as intended. In addition, the collective bench testing demonstrates that the DASH 6® NRFit Syringe device family and accessories do not raise new questions of safety or effectiveness as compared to the predicate or the associated accessories.

5.7 CONCLUSION

The DASH 6 ° NRFit Syringe family and accessories have the same intended use and similar technological characteristics as the predicate devices. The DASH 6 NRFit Syringe family and accessories have been tested to ensure that they perform as intended and do not raise new questions of safety or effectiveness. As such, the DASH 6 NRFit Syringe device family and accessories are substantially equivalent to the predicate and the associated accessories.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).