Search Filters

Search Results

Found 8 results

510(k) Data Aggregation

    K Number
    K242002

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-11-22

    (136 days)

    Product Code
    Regulation Number
    884.4530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FemVue MINI is intended to instill a consistent alternating pattern of saline and air as a continuous stream of contrast media into the uterus and fallopian tubes to be used in conjunction with an intrauterine catheter for performance of sono-hysterosalpingogram (Sono HSG).

    Device Description

    Fem Vue® MINI is a dual-barrel contrast media syringe that can be connected to an intrauterine catheter for controlled delivery of saline-air contrast media during sono-hysterosalpingogram (Sono HSG) procedures. Sono HSG consists of an ultrasound evaluation of the fallopian tubes with or without assessment of the uterine cavity. The device operates by retraction of the plunger, which simultaneously fills the two syringes in the device with either air or saline. The device is then connected to a compatible uterine catheter, and when the plunger is depressed, a consistent stream of saline and air is delivered into the uterus and fallopian tubes. The device is provided sterilization and is intended for single-use.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the studies mentioned for the FemVue MINI Saline-Air Device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance metrics. Instead, it states that "all predetermined acceptance criteria were met" for various tests. Therefore, the table below is constructed by inferring the acceptance criteria from the type of test conducted.

    Acceptance Criterion (Inferred from Test)Reported Device Performance
    Sterility (achieves sterility after Ethylene Oxide sterilization)Met (per ISO 11135:2014, AAMI TIR 28:2016, ISO 10993-7: 2008)
    Package Integrity (maintains sterility barrier)Met (per Visual inspection, ASTM F2096-11, ASTM F88/ F88M-23, ASTM F1886/F1886-16)
    Withstand Transportation (maintains integrity after transport simulation)Met (per ASTM D4169-22)
    Biocompatibility (non-cytotoxic, non-sensitizing, non-irritating)Met (non-cytotoxic, non-sensitizing, and non-irritating per ISO 10993-5: 2009, ISO 10993-10: 2021)
    Visual Inspection (meets specified visual appearance)Met (before and after accelerated aging)
    Functional/Cycle Testing (operates as intended through its lifecycle)Met (before and after accelerated aging)
    Saline-Air Quantification (delivers consistent alternating pattern of saline and air)Met (before and after accelerated aging)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for the individual bench performance tests. It only states that these tests were conducted. It also does not explicitly state the data provenance (e.g., country of origin) or if the studies were retrospective or prospective, though bench testing is inherently prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The tests described are primarily physical, chemical, and functional bench tests, rather than clinical studies requiring expert ground truth for interpretation.

    4. Adjudication Method for the Test Set

    This information is not provided in the document. Given the nature of the bench tests, it's unlikely a formal adjudication method (like 2+1, 3+1) would be applicable. The tests likely have objective pass/fail criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document describes non-clinical performance data for a device, not a diagnostic algorithm that would typically require such a study to assess human reader improvement with AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone performance study (algorithm only) was not done. This device is a manual instrument, not an AI or software algorithm.

    7. The Type of Ground Truth Used

    For the non-clinical performance studies described, the "ground truth" refers to established engineering standards, material specifications, and functional requirements for a medical device. This is determined by the specific ASTM and ISO standards cited (e.g., ISO 11135 for sterility, ISO 10993 for biocompatibility, ASTM F1980-16 for accelerated aging). The concept of expert consensus, pathology, or outcomes data as ground truth is not applicable to these types of engineering and functional tests.

    8. The Sample Size for the Training Set

    This information is not applicable and therefore not provided. This device is a hardware product, not a machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable and therefore not provided, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241693

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-09-06

    (86 days)

    Product Code
    Regulation Number
    884.4530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FemChec® is intended to instill a consistent alternating pattern of saline and air as a continuous stream of contrast media into the uterus and fallopian tubes to be used in conjunction with an intrauterine catheter for performance of sono-hysterosalpingogram (Sono HSG).

    Device Description

    FemChec® is a dual-barrel contrast media syringe that can be connected to an intrauterine catheter for controlled delivery of saline-air contrast media during sono-hysterosalpingogram (Sono HSG) procedures. Sono HSG consists of an ultrasound evaluation of the fallopian tubes with or without assessment of the uterine cavity. The device operates by retraction of the plunger, which simultaneously fills the two syringes in the device with either air or saline. The device is then connected to a compatible uterine catheter, and when the plunger is depressed, a consistent stream of saline and air is delivered into the uterus and fallopian tubes. The device is provided sterile via EO sterilization and is intended for single-use.

    AI/ML Overview

    The provided text is a 510(k) summary for the FemChec Controlled Saline-Air Device (FCD-250), which is a medical device and not an AI or software device. Therefore, the questions related to AI/algorithm performance, training sets, test sets, ground truth establishment by experts, adjudication methods, and MRMC studies are not applicable to this document. The document focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance data for a physical medical device.

    However, I can extract the acceptance criteria and a summary of the studies performed from the provided text for the FemChec device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Tests/CriteriaReported Device Performance
    SterilizationEthylene Oxide Sterilization Validation per ISO 11135-1:2014, AAMI TIR 28:2016, ISO 10993-7:2008Not explicitly stated "met criteria," but implied that testing was performed to support substantial equivalence.
    PackagingVisual inspection, Bubble Leak test per ASTM F2096-11, Seal Strength testing per ASTM F88/F88M-15Not explicitly stated "met criteria," but implied that testing was performed to support substantial equivalence.
    TransportationTransportation Simulation testing per ASTM D4169-14Not explicitly stated "met criteria," but implied that testing was performed to support substantial equivalence.
    BiocompatibilityCytotoxicity per ISO 10993-5:2009, Sensitization ISO 10993-10:2021, Irritation per ISO 10993-10:2021Tested to be non-cytotoxic, non-sensitizing, and non-irritating.
    Bench PerformanceVisual Inspection, Functional/Cycle Testing, Saline-Air Quantification (before and after accelerated aging to 12 months real-time aging per ASTM F1980-16)All predetermined acceptance criteria were met.

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not applicable as the studies conducted are non-clinical bench performance and material compatibility studies for a physical device, not a performance study involving a test set of data points or images.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    This information is not applicable. Ground truth establishment by experts is relevant for AI/software devices requiring interpretation of data (e.g., medical images), which is not the case for this physical medical device.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods are relevant for studies where multiple readers/experts assess cases, which is not applicable to the non-clinical studies for this physical device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document describes non-clinical performance studies for a physical device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a physical medical instrument, not an algorithm.

    7. The Type of Ground Truth Used

    This information is not applicable in the context of AI/software performance. For the described bench tests, the "ground truth" would be established by the specifications and standards (e.g., flow rates, material properties, sterility levels) defined in the test protocols themselves, which are based on recognized international and ASTM standards.

    8. The Sample Size for the Training Set

    This information is not applicable. The device is a physical medical instrument, not an AI/software device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable. The device is a physical medical instrument, not an AI/software device that requires a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231730

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-09-22

    (101 days)

    Product Code
    Regulation Number
    884.6110
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FemaSeed Intratubal Insemination is intended to introduce washed sperm or in vitro fertilized (IVF) embryos into the uterine ostium via ultrasound guidance

    Device Description

    The FemaSeed Intratubal Insemination consists of polymer curved transfer catheter with balloon, polymer guide catheter and separate balloon inflation syringe. The balloon inflation syringe is a 3 mL syringe fitted with clips that allow for the piston to lock in place to maintain balloon inflation. When the clips are pinched, the piston releases, deflating the balloon. The device is intended to deliver washed sperm or in vitro fertilized embryos into the uterine ostium via ultrasound guidance by a licensed healthcare provider specifically trained in women's health.

    Prior to insertion, the slider is moved back until the curved transfer catheter is fully retracted and contained within guide catheter. The preassembled guide catheter is then passed transvaginally through the cervix to the fundus, under ultrasound guidance. Once in position, the preloaded transfer catheter is advanced approximately 2 cm into the uterine ostium and the balloon is inflated with the provided syringe. The balloon location in the uterine ostium is confirmed using ultrasound. The samples are then instilled into the uterine ostium of the fallopian tube, followed by deflating the balloon and discarding the device. The recommended volume for loading washed sperm should not exceed 1 mL and 50-100 µL when loading embryos.

    AI/ML Overview

    The provided text describes the FemaSeed Intratubal Insemination device and its comparison to a predicate device, along with a summary of non-clinical performance data. However, it does not include detailed acceptance criteria or a study proving the device meets them in the format requested. The document is a 510(k) summary for FDA clearance, which focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing. It lists various tests performed and their general outcomes (e.g., "all predetermined acceptance criteria were met"), but it doesn't quantify those criteria or the specific performance metrics for each, nor does it describe specific studies with sample sizes, expert involvement, or ground truth establishment for a device performance acceptance criteria table as requested.

    Therefore, the following response will extract the available information and highlight the missing details based on your request.


    1. Table of Acceptance Criteria and Reported Device Performance

    Based on the provided text, specific quantified acceptance criteria and reported device performance are generally not detailed. The document often states that "predetermined acceptance criteria were met" without specifying what those criteria were. However, some specific acceptance criteria and performance are mentioned:

    Acceptance Criteria CategorySpecific Acceptance Criteria (if stated)Reported Device Performance (if stated)
    BiocompatibilityN/A - Refer to ISO 10993 seriesNon-cytotoxic, non-sensitizing, non-irritating
    Endotoxin< 20 EU/device< 20 EU/device
    Human Sperm Survival Assay (HSSA)≥80% of control motility at 24 hours following 30-minute exposure≥80% of control motility at 24 hours following 30-minute exposure
    Mouse Embryo Assay (MEA)≥ 80% embryos developed to expanded blastocyst at 72 hours≥ 80% embryos developed to expanded blastocyst at 72 hours
    Mechanical PerformancePredetermined acceptance criteria metAll predetermined acceptance criteria were met
    Balloon IntegrityPredetermined acceptance criteria metAll predetermined acceptance criteria were met
    Maximum Burst PressurePredetermined acceptance criteria metAll predetermined acceptance criteria were met
    Maximum Cycles for Stop Lock ClipsPredetermined acceptance criteria metAll predetermined acceptance criteria were met
    Maximum Extension ForcePredetermined acceptance criteria metAll predetermined acceptance criteria were met
    Depth of Insertion & Transfer Catheter AdvancementPredetermined acceptance criteria metAll predetermined acceptance criteria were met
    Aspiration & DeliveryPredetermined acceptance criteria metAll predetermined acceptance criteria were met
    Minimize Lost SpecimenPredetermined acceptance criteria metAll predetermined acceptance criteria were met
    Indicate Full Advancement/RetractionPredetermined acceptance criteria metAll predetermined acceptance criteria were met
    Guide Catheter Insertion (no kinking, minimize discomfort/injury)Predetermined acceptance criteria metAll predetermined acceptance criteria were met

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes non-clinical performance studies (e.g., sterilization, package integrity, biocompatibility, bench testing). However, it does not specify the sample sizes used for these tests. For instance, for mechanical tests or HSSA/MEA, the number of devices or biological samples tested is not provided. The data provenance is implied to be from laboratory testing performed by or for the device manufacturer (Femasys, Inc.), which is typically considered prospective data generation for regulatory submission. There is no mention of country of origin for any data beyond the general context of U.S. FDA submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. The studies mentioned are non-clinical, benchtop, and biological compatibility tests, which typically do not involve human "experts" establishing a ground truth in the way described (e.g., radiologists interpreting images). The "ground truth" for these tests would be the established scientific and engineering standards and methods against which the device performance is measured (e.g., acceptable cytotoxicity levels, sperm motility percentages, embryo development rates).

    4. Adjudication Method for the Test Set

    This is not applicable and not provided. Adjudication methods are typically used in clinical studies or studies involving human readers/evaluators where there might be inter-observer variability in assessment. The non-clinical tests described do not involve such a process.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done or mentioned. The studies described are non-clinical hardware performance and biological compatibility tests, not clinical or reader performance studies.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a "standalone" study in the context of an algorithm or AI without human intervention was not done or mentioned. The device is a physical medical instrument (catheter) used by a healthcare provider.

    7. The type of Ground Truth Used

    For the biological tests:

    • Mouse Embryo Assay (MEA): The ground truth is the biological standard of embryo development (≥ 80% embryos developed to expanded blastocyst at 72 hours).
    • Human Sperm Survival Assay (HSSA): The ground truth is the biological standard of sperm motility (≥80% of control motility).
    • Endotoxin: The ground truth is the chemical standard of acceptable endotoxin levels (< 20 EU/device).

    For mechanical/physical tests:

    • The "ground truth" or reference is the predetermined acceptance criteria based on engineering standards, design specifications, and regulatory requirements, ensuring functionality, safety, and compatibility.

    For sterilization, package integrity, and biocompatibility:

    • The "ground truth" is adherence to recognized international standards and FDA guidance documents (e.g., ISO 11135-1:2014, ASTM F2096-11, ISO 10993 series).

    8. The Sample Size for the Training Set

    This information is not provided and is not applicable as the device is a physical catheter, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    This information is not provided and is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122658

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-12-20

    (111 days)

    Product Code
    Regulation Number
    884.1050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Femasys FemCerv is a sterile, disposable endocervical sampler indicated for single patient use in obtaining tissue samples from the endocervical canal for histological analysis. Clinical indications include: further evaluation of an abnormal Pap smear; cervical lesions extending into the endocervical canal; undiagnosed uterine bleeding.

    Device Description

    The FemCerv Endocervical Sampler (FemCerv) is a sterile, disposable device that collects an endocervical tissue sample for histological evaluation.

    AI/ML Overview

    Here's a breakdown of the requested information based on the provided text:

    Acceptance Criteria and Device Performance for FemCerv™ Endocervical Sampler

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    FunctionalityBench Testing:- Device functionality and integrity over cycle testing of opening and closing the sample collection chamber.- Insertion capability in a model of the endocervical canal.- Sheath opening functionality.- Rotational functionality.- Sheath closure.- Sample release into a specimen container."A sample set of 29 devices was tested and passed all criteria."
    Shelf LifeShelf life integrity up to 1 year accelerated aging:- Same test methodology and acceptance criteria as bench testing."A sample set of 29 devices was tested and passed all criteria."
    BiocompatibilityISO 10993 standards:- Cytotoxicity.- Irritation.- Sensitization."Biocompatibility of components according to ISO 10993 standards: cytotoxicity, irritation, and sensitization was conducted and passed all criteria."
    Clinical PerformanceClinical Study:- Insertion.- Device opening and closing.- Adequate sampling capability based upon a pathologist's evaluation."An 85.7% sample adequacy rate was based on sampling with 7 devices." (This is the reported performance for "adequate sampling capability")

    2. Sample size used for the test set and the data provenance:

    • Bench Testing: 29 devices
    • Shelf Life Testing: 29 devices
    • Clinical Study: 7 devices
    • Data Provenance: The document does not explicitly state the country of origin. The clinical study involved subjects prior to scheduled hysterectomy, which implies a prospective collection of samples for the purpose of the study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The clinical study's sample adequacy was "based upon a pathologist's evaluation." The document does not specify the number of pathologists or their specific qualifications (e.g., years of experience).

    4. Adjudication method for the test set:

    • The document states that sample adequacy was based on "a pathologist's evaluation." It does not mention any specific adjudication method (e.g., 2+1, 3+1, none) among multiple experts, suggesting a single pathologist's assessment or an unstated internal process.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an endocervical sampler, not an AI-assisted diagnostic tool for image interpretation. Therefore, the concept of human readers improving with AI assistance is not applicable.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No, this is a physical medical device for collecting tissue samples, not an algorithm. Therefore, the concept of standalone algorithm performance is not applicable.

    7. The type of ground truth used:

    • Clinical Study: The ground truth for sample adequacy was established by pathological evaluation of the collected tissue samples.

    8. The sample size for the training set:

    • The document does not describe a "training set" in the context of machine learning or AI. The tests mentioned are for verifying the physical properties and performance of the device itself.

    9. How the ground truth for the training set was established:

    • Not applicable, as there is no mention of a training set for an algorithm or AI model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K110993

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-10-12

    (187 days)

    Product Code
    Regulation Number
    884.4530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FemChec™ Pressure Management Device limits the maximum applied pressure while instilling contrast media into the uterus and fallopian tubes. It is to be used in conjunction with an intrauterine catheter for performance of hysterosalpingogram (HSG), such as tubal occlusion confirmation tests in women who have had tubal procedures for permanent female sterilization. The FemChec limits the applied intrauterine pressure to 200 mmHg.

    The FemChec is provided with a fluid collector, stopcock, and intrauterine catheter for performing the HSG procedure.

    Device Description

    The FemChec Pressure Management Device (FemChec) is a pressure-limiting contrast media syringe that is connected to an intrauterine catheter for low-pressure instillation of contrast media during HSG procedures.

    AI/ML Overview

    Here's an analysis of the FemChec™ Pressure Management Device's acceptance criteria and the study used to demonstrate its performance, based on the provided 510(k) summary:

    Acceptance Criteria and Device Performance for FemChec™ Pressure Management Device

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Functional/Performance Aspects)Reported Device Performance (as demonstrated by testing)
    Fluid instilling function of syringeConfirmed (implied by "fluid instilling function of syringe" test)
    Attachment functionalityConfirmed (implied by "attachment functionality" test)
    Single-hand actuationConfirmed (implied by "single-hand actuation" test)
    Pressure testing < 200 mm HgLimits applied intrauterine pressure to 200 mmHg
    Biocompatibility (cytotoxicity, irritation, sensitization, acute systemic toxicity)Confirmed (according to ISO 10993 standards)

    2. Sample Size and Data Provenance

    The provided document describes non-clinical methods used for testing. This means the testing was likely conducted in a laboratory setting or on simulated systems, not on human subjects.

    • Sample Size for Test Set: The document does not specify the exact number of devices or components tested for each criterion. It only lists the types of tests performed.
    • Data Provenance: The data is from non-clinical testing, meaning it's not derived from human patients. The country of origin for the data is not explicitly stated, but given the 510(k) submission to the FDA, it would typically be generated by the manufacturer's own internal testing or by a contracted lab, likely in the US or a country with comparable regulatory standards. The testing is prospective in the sense that it was conducted specifically to support this regulatory submission.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: Not applicable. Since the testing was non-clinical, there were no human experts establishing a "ground truth" for the performance characteristics in the way they would for a diagnostic accuracy study. The "ground truth" for the functional tests (e.g., pressure limiting) would have been defined by the device's design specifications and accepted engineering principles, verified through instrument measurements.
    • Qualifications of Experts: Not applicable for establishing ground truth in this context. The "experts" involved would be engineers and lab technicians conducting the tests according to established protocols.

    4. Adjudication Method

    • Adjudication Method: Not applicable. Non-clinical performance testing typically involves objective measurements against predefined specifications, rather than subjective interpretation requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? No. The submission describes non-clinical performance testing, not a clinical study involving human readers.
    • Effect size of human readers improve with AI vs without AI assistance: Not applicable, as no MRMC study was conducted, and the device is a mechanical pressure management system, not an AI-powered diagnostic tool.

    6. Standalone (Algorithm Only) Performance Study

    • Was a standalone study done? Yes, in a sense. The described testing focuses solely on the performance of the device itself (e.g., its ability to limit pressure, its biocompatibility, its functionality) in a laboratory setting, independent of human interaction beyond operating the device. However, "standalone" in this context typically refers to AI algorithm performance without human intervention, which is not relevant here.

    7. Type of Ground Truth Used

    • Type of Ground Truth: The ground truth for the non-clinical tests was based on engineering specifications and objective measurements. For instance, the "pressure testing < 200 mm Hg" would involve measuring the pressure exerted by the device and verifying it does not exceed the 200 mmHg limit. Biocompatibility relies on adherence to established international standards (ISO 10993).

    8. Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. This device is a mechanical medical device, not a machine learning or AI-based system. Therefore, there is no "training set" in the context of algorithm development.

    9. How Ground Truth for the Training Set Was Established

    • How Ground Truth for Training Set Was Established: Not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K110288

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-04-28

    (86 days)

    Product Code
    Regulation Number
    884.4530
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FemVue™ Saline-Air Device instills a consistent alternating pattern of saline and air as a continuous stream of contrast media into the uterus and fallopian tubes to be used in conjunction with an intrauterine catheter for performance of sono-hysterosalpingogram (Sono HSG).

    Device Description

    The FemVue™ Saline-Air Device is a dual-barrel contrast media syringe that can be connected to an intrauterine catheter to instill saline-air contrast media during sono-hysterosalpingogram (Sono HSG) procedures. Sono HSG consists of an ultrasound evaluation of the fallopian tubes with or without assessment of the uterine cavity.

    AI/ML Overview

    The FemVue™ Saline-Air Device, as described in the K110288 510(k) submission, is a physical medical device, specifically a contrast media syringe, rather than an AI/ML powered medical device. Therefore, many of the requested categories related to AI/ML device studies (e.g., sample size for test set, data provenance, ground truth establishment, MRMC studies, standalone algorithm performance, training set details) are not applicable to this submission.

    However, based on the provided text, the acceptance criteria and the study performed can be described as follows:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance (Summary of Testing)
    Fluid and air instillation function of syringeDemonstrated
    Catheter attachmentDemonstrated
    Single-hand actuationDemonstrated
    Delivers continuous and consistent stream of an alternating pattern of saline and airDemonstrated

    2. Sample size used for the test set and the data provenance

    Not applicable for a physical device where the "test set" would typically refer to a dataset for an AI/ML algorithm. The testing described is non-clinical performance testing of the physical hardware itself. The document does not specify a numerical sample size for the devices tested, nor does it refer to data provenance in the context of clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a non-clinical performance study of a physical device. Ground truth, in the context of expert review, is not relevant here. The "ground truth" for these tests is the physical and mechanical functionality of the device against predefined performance targets.

    4. Adjudication method for the test set

    Not applicable. Adjudication methods are typically used when subjective assessments by multiple experts need to be reconciled for establishing ground truth in clinical or image-based studies. This submission describes non-clinical performance testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device, and thus no MRMC study or AI assistance evaluation was performed or is relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the non-clinical tests would have been the pre-defined engineering specifications and performance targets for the device's mechanical and fluidic functionalities. For example, "continuous and consistent stream of an alternating pattern of saline and air" would have specific measurable parameters that the device was designed to meet.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not an AI/ML algorithm requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable. As above, this is a physical device, and there is no training set as understood in AI/ML development.

    Ask a Question

    Ask a specific question about this device

    K Number
    K100662

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-04-07

    (30 days)

    Product Code
    Regulation Number
    884.4530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FemVue™ Cornual Balloon Catheter is intended for the delivery of contrast media during hysterosalpingogram (HSG) and saline infusion hysterosonography (SIS) for the evaluation of the fallopian tube(s) selectively and/or the uterus. The following are some clinical indications: suspected polyps, fibroids, adhesions, or endometrial thickening, and/or the selective evaluation of fallopian tube patency.

    Device Description

    The FemVue Cornual Balloon Catheter ("FCBC") is a latex-free Balloon Catheter within a transcervical Delivery Sheath.

    AI/ML Overview

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Implied)Reported Device Performance
    Biocompatibility (cytotoxicity, irritation, sensitization, acute systemic toxicity)Met (tested by ISO 10993 methods)
    Tensile strength of joint bondsMet (tensile testing performed)
    Component functionalityMet (tested)
    Inflation balloon integrityMet (tested)
    Substantially equivalent in functionality, safety, and effectiveness to predicate device (K083690)Achieved (based on non-clinical methods listed above and FDA's substantial equivalence determination)

    Note: The document explicitly lists the types of tests performed to demonstrate substantial equivalence but does not provide specific numerical acceptance criteria (e.g., "tensile strength of X Newtons") or the exact numerical results for the device performance. The FDA's clearance indicates that the device met the necessary criteria for substantial equivalence.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document describes non-clinical testing only. Therefore, there is no human subject test set in the traditional sense of a clinical study described in this submission. The tests performed are engineering and bench-top tests of the physical device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This submission relies solely on non-clinical, bench-top testing. There were no human experts establishing ground truth for a clinical test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There was no clinical test set requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a catheter for delivering contrast media, not an AI-powered diagnostic tool. No MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical device (catheter), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable for a clinical ground truth. For the engineering tests, the "ground truth" was established by the specifications and performance characteristics of the predicate device and relevant international standards (e.g., ISO 10993 for biocompatibility).

    8. The sample size for the training set

    Not applicable. This device is not an AI/machine learning algorithm that requires a training set. The tests performed are for a physical medical device.

    9. How the ground truth for the training set was established

    Not applicable. As noted above, there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K083690

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-06-23

    (193 days)

    Product Code
    Regulation Number
    884.4530
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FemVue™ Catheter System is intended for the delivery of contrast media during hysterosalpingography (HSG) and saline infusion hysterosonography (SIS) for the evaluation of the fallopian tube(s) selectively and/or the uterus. The following are some clinical indications: suspected polyps, fibroids, adhesions, or endometrial thickening, and/or the selective evaluation of fallopian tube patency.

    Device Description

    The FemVue Catheter System ("FCS") is comprised of two components: a Sheath and a latex free balloon Catheter. The Sheath is used to deliver the Catheter into the uterus.

    AI/ML Overview

    The provided text for the FemVue™ Catheter System (K083690) is a 510(k) summary and FDA clearance letter. It does not contain information about acceptance criteria, device performance metrics, or study details such as sample sizes, ground truth establishment, or expert qualifications.

    The summary focuses on demonstrating substantial equivalence to predicate devices through:

    • Equivalent device type, indication for use, principle of operation, and device description.
    • Similar principles of operation for device placement, device description, and safety features.
    • Substantially equivalent indication for use, placement location, and principle of operation for evaluation of the fallopian tubes.

    The only "testing" mentioned is biocompatibility testing (cytotoxicity, irritation, and sensitization) in compliance with ISO 10993. This type of testing ensures the material safety of the device when in contact with tissues, but it does not evaluate the device's functional performance in terms of diagnostic accuracy or effectiveness for its intended use (e.g., evaluating fallopian tube patency or uterine abnormalities).

    Therefore, based solely on the provided text, I cannot complete the requested tables and sections related to performance acceptance criteria and a study demonstrating that the device meets those criteria. Such information is typically found in the clinical study section of a 510(k) submission, which is not included here.

    The document indicates that the device was cleared based on substantial equivalence to existing devices, implying that its performance is presumed to be similar to its predicates, rather than requiring a de novo performance study with specific acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1