K Number
K122658

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-12-20

(111 days)

Product Code
Regulation Number
884.1050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Femasys FemCerv is a sterile, disposable endocervical sampler indicated for single patient use in obtaining tissue samples from the endocervical canal for histological analysis. Clinical indications include: further evaluation of an abnormal Pap smear; cervical lesions extending into the endocervical canal; undiagnosed uterine bleeding.

Device Description

The FemCerv Endocervical Sampler (FemCerv) is a sterile, disposable device that collects an endocervical tissue sample for histological evaluation.

AI/ML Overview

Here's a breakdown of the requested information based on the provided text:

Acceptance Criteria and Device Performance for FemCerv™ Endocervical Sampler

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
FunctionalityBench Testing:- Device functionality and integrity over cycle testing of opening and closing the sample collection chamber.- Insertion capability in a model of the endocervical canal.- Sheath opening functionality.- Rotational functionality.- Sheath closure.- Sample release into a specimen container."A sample set of 29 devices was tested and passed all criteria."
Shelf LifeShelf life integrity up to 1 year accelerated aging:- Same test methodology and acceptance criteria as bench testing."A sample set of 29 devices was tested and passed all criteria."
BiocompatibilityISO 10993 standards:- Cytotoxicity.- Irritation.- Sensitization."Biocompatibility of components according to ISO 10993 standards: cytotoxicity, irritation, and sensitization was conducted and passed all criteria."
Clinical PerformanceClinical Study:- Insertion.- Device opening and closing.- Adequate sampling capability based upon a pathologist's evaluation."An 85.7% sample adequacy rate was based on sampling with 7 devices." (This is the reported performance for "adequate sampling capability")

2. Sample size used for the test set and the data provenance:

  • Bench Testing: 29 devices
  • Shelf Life Testing: 29 devices
  • Clinical Study: 7 devices
  • Data Provenance: The document does not explicitly state the country of origin. The clinical study involved subjects prior to scheduled hysterectomy, which implies a prospective collection of samples for the purpose of the study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • The clinical study's sample adequacy was "based upon a pathologist's evaluation." The document does not specify the number of pathologists or their specific qualifications (e.g., years of experience).

4. Adjudication method for the test set:

  • The document states that sample adequacy was based on "a pathologist's evaluation." It does not mention any specific adjudication method (e.g., 2+1, 3+1, none) among multiple experts, suggesting a single pathologist's assessment or an unstated internal process.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an endocervical sampler, not an AI-assisted diagnostic tool for image interpretation. Therefore, the concept of human readers improving with AI assistance is not applicable.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • No, this is a physical medical device for collecting tissue samples, not an algorithm. Therefore, the concept of standalone algorithm performance is not applicable.

7. The type of ground truth used:

  • Clinical Study: The ground truth for sample adequacy was established by pathological evaluation of the collected tissue samples.

8. The sample size for the training set:

  • The document does not describe a "training set" in the context of machine learning or AI. The tests mentioned are for verifying the physical properties and performance of the device itself.

9. How the ground truth for the training set was established:

  • Not applicable, as there is no mention of a training set for an algorithm or AI model.

{0}------------------------------------------------

K122658

510(k) Summary for the FemCerv™ Endocervical Sampler

(Models REF FCV-009, FCV-011, and FCV-013)

Date:December 18, 2012 DEC 2 0 2012
510(k) Submitter and Primary Contact:Lisa PeacockVice President, Regulatory and Clinical AffairsFemasys Inc.5000 Research CourtSuite 100Suwanee, GA 30024Tel: 770-500-3910Fax: 770-500-3980LPEACOCK@FEMASYS.COM
Device Common Name:endocervical sampler
FDA Device Classification Name:endocervical aspirator
Product Code:PCF
Classification Regulation:21 CFR 884.1050
Device Class:2
Panel:Obstetrics/Gynecology
Indication for Use:The Femasys FemCerv is a sterile, disposable endocervical sampler indicated for single patient use in obtaining tissue samples from the endocervical canal for histological analysis. Clinical indications include:further evaluation of an abnormal Pap smear;cervical lesions extending into the endocervical canal;undiagnosed uterine bleeding.
Device Description:The FemCerv Endocervical Sampler (FemCerv) is a sterile, disposable device that collects an endocervical tissue sample for histological evaluation.
Predicate Device:K060320 FemECC™ Endocervical CuretteThe FemCerv is substantially equivalent to the predicate in intended use, sample collection method using a single-pass

{1}------------------------------------------------

Summary of Testing:

rotational scraping mode within the endocervical canal, sample transfer method, and in its sterile, disposable, non-metal design.

Final, finished, sterilized FemCerv devices were tested by the following methods for verification of functionality, safety, and effectiveness.

  • · Bench testing was conducted to evaluate device functionality and integrity over cycle testing of opening and closing the device's sample collection chamber. In addition, the bench testing included simulated use testing in a model of the endocervical canal evaluating insertion capability, sheath opening functionality, rotational functionality, sheath closure, and sample release into a specimen container. A sample set of 29 devices was tested and passed all criteria.
  • · Shelf life integrity up to 1 year accelerated aging was evaluated using the same test methodology and acceptance criteria in the point above. A sample set of 29 devices was tested and passed all criteria.
  • · Biocompatibility of components according to ISO 10993 standards: cytotoxicity, irritation, and sensitization was conducted and passed all criteria.
  • · A clinical study was conducted involving endocervical sampling on subjects prior to scheduled hysterectomy to demonstrate insertion, device opening and closing, and adequate sampling capability based upon a pathologist's evaluation. An 85.7% sample adequacy rate was based on sampling with 7 devices.

Conclusions of Substantial Equivalence Demonstrations:

  • · FemCerv is substantially equivalent to its proposed predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 20, 2012

Femasys, Inc. % Ms. Lisa Peacock Vice President, Regulatory and Clinical Affairs 5000 Research Court, Suite 100 SUWANEE GA 30024

Re: K122658

Trade/Device Name: FemCerv™ Endocervical Sampler Regulation Number: 21 CFR§ 884.1050 Regulation Name: Endocervical aspirator Regulatory Class: II Product Code: PCF Dated: December 11, 2012 Received: December 12, 2012

Dear Ms. Peacock:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{3}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Herbert R. Lerner

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Femasys Inc.

Page 1 of 1

Indications for Use

510(k) Number (if known): K122658

Device Name: FemCerv™ Endocervical Sampler (Models REF FCV-009, FCV-011, and FCV-013)

Indications for Use:

The Femasys FemCerv is a sterile, disposable endocervical sampler indicated for single patient use in obtaining tissue samples from the endocervical canal for histological analysis. Clinical indications include:

  • further evaluation of an abnormal Pap smear; ●
  • cervical lesions extending into the endocervical canal; 미
  • undiagnosed uterine bleeding. 8

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert R. Lerner

(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K122658 510(k) Number

§ 884.1050 Endocervical aspirator.

(a)
Identification. An endocervical aspirator is a device designed to remove tissue from the endocervix (mucous membrane lining the canal of the cervix of the uterus) by suction with a syringe, bulb and pipette, or catheter. This device is used to evaluate endocervical tissue to detect malignant and premalignant lesions.(b)
Classification. Class II (performance standards).