Search Filters

Search Results

Found 62 results

510(k) Data Aggregation

    K Number
    K242002
    Manufacturer
    Date Cleared
    2024-11-22

    (136 days)

    Product Code
    Regulation Number
    884.4530
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LKF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FemVue MINI is intended to instill a consistent alternating pattern of saline and air as a continuous stream of contrast media into the uterus and fallopian tubes to be used in conjunction with an intrauterine catheter for performance of sono-hysterosalpingogram (Sono HSG).

    Device Description

    Fem Vue® MINI is a dual-barrel contrast media syringe that can be connected to an intrauterine catheter for controlled delivery of saline-air contrast media during sono-hysterosalpingogram (Sono HSG) procedures. Sono HSG consists of an ultrasound evaluation of the fallopian tubes with or without assessment of the uterine cavity. The device operates by retraction of the plunger, which simultaneously fills the two syringes in the device with either air or saline. The device is then connected to a compatible uterine catheter, and when the plunger is depressed, a consistent stream of saline and air is delivered into the uterus and fallopian tubes. The device is provided sterilization and is intended for single-use.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the studies mentioned for the FemVue MINI Saline-Air Device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance metrics. Instead, it states that "all predetermined acceptance criteria were met" for various tests. Therefore, the table below is constructed by inferring the acceptance criteria from the type of test conducted.

    Acceptance Criterion (Inferred from Test)Reported Device Performance
    Sterility (achieves sterility after Ethylene Oxide sterilization)Met (per ISO 11135:2014, AAMI TIR 28:2016, ISO 10993-7: 2008)
    Package Integrity (maintains sterility barrier)Met (per Visual inspection, ASTM F2096-11, ASTM F88/ F88M-23, ASTM F1886/F1886-16)
    Withstand Transportation (maintains integrity after transport simulation)Met (per ASTM D4169-22)
    Biocompatibility (non-cytotoxic, non-sensitizing, non-irritating)Met (non-cytotoxic, non-sensitizing, and non-irritating per ISO 10993-5: 2009, ISO 10993-10: 2021)
    Visual Inspection (meets specified visual appearance)Met (before and after accelerated aging)
    Functional/Cycle Testing (operates as intended through its lifecycle)Met (before and after accelerated aging)
    Saline-Air Quantification (delivers consistent alternating pattern of saline and air)Met (before and after accelerated aging)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for the individual bench performance tests. It only states that these tests were conducted. It also does not explicitly state the data provenance (e.g., country of origin) or if the studies were retrospective or prospective, though bench testing is inherently prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The tests described are primarily physical, chemical, and functional bench tests, rather than clinical studies requiring expert ground truth for interpretation.

    4. Adjudication Method for the Test Set

    This information is not provided in the document. Given the nature of the bench tests, it's unlikely a formal adjudication method (like 2+1, 3+1) would be applicable. The tests likely have objective pass/fail criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document describes non-clinical performance data for a device, not a diagnostic algorithm that would typically require such a study to assess human reader improvement with AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone performance study (algorithm only) was not done. This device is a manual instrument, not an AI or software algorithm.

    7. The Type of Ground Truth Used

    For the non-clinical performance studies described, the "ground truth" refers to established engineering standards, material specifications, and functional requirements for a medical device. This is determined by the specific ASTM and ISO standards cited (e.g., ISO 11135 for sterility, ISO 10993 for biocompatibility, ASTM F1980-16 for accelerated aging). The concept of expert consensus, pathology, or outcomes data as ground truth is not applicable to these types of engineering and functional tests.

    8. The Sample Size for the Training Set

    This information is not applicable and therefore not provided. This device is a hardware product, not a machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable and therefore not provided, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240364
    Date Cleared
    2024-10-16

    (253 days)

    Product Code
    Regulation Number
    884.4530
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LKF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Use of the RELIEEV Uterine Manipulator Injector is indicated in Diagnostic Laparoscopy, Minilaparotomy, Fertility Exams, and Salpingoplasty procedures where manipulation of the uterus is required.

    Device Description

    The "RELIEEV" Uterine Manipulator Injector is intended to be used during medical procedures where manipulation of the uterus or injection of fluids into the uterine lumen are needed. This device is a sterile, single use product consisting of a plastic (PVC) insertion tube that includes two lumens, one for inflation of a 10 mL intrauterine balloon (cuff) at the distal end of the insertion tube via the inflation valve using the provided 10 mL syringe and the other for injection of fluid through a distal end-port using a user-provided syringe via the Leur fitting. The device also includes a movable/removable rigid handle that allows alteration of the device insertion depth and includes a cervical stop, laser etched depth markings to aid in setting the insertion depth, and a pilot balloon to assess maintenance of cuff inflation. The device is curved to facilitate forward uterine manipulation. The device has an overall length of 36 cm and an outer diameter of the catheter component of 5.0 mm. When inflated with 10 mL of air, the balloon has a diameter of 23.0-24.9 mm.

    AI/ML Overview

    This document is an FDA 510(k) summary for the "RELIEEV" Uterine Manipulator Injector. It describes the device and claims substantial equivalence to a predicate device, the Panpac Uterine Manipulator Injector, Model Umi 4.5 (K092980).

    Based on the provided text, the device in question is a medical instrument, not an AI/software device. Therefore, the questions related to AI/software performance, ground truth, human readers, and training/test sets are not applicable to this document.

    The acceptance criteria and study proving the device meets them are related to the physical and biological safety and performance of the Uterine Manipulator Injector, rather than an AI algorithm.

    Here's the information extracted from the document, tailored to the type of device described:


    Acceptance Criteria and Device Performance for "RELIEEV" Uterine Manipulator Injector

    This device is a physical medical instrument, and its performance evaluation focuses on its physical characteristics, sterility, biocompatibility, and mechanical integrity, comparing it to a legally marketed predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document details various tests performed to ensure the device meets safety and performance standards for a physical medical device. The "acceptance criteria" are implied by the successful completion of these tests per the referenced standards and the statement that "all predetermined acceptance criteria were met."

    Area of PerformanceSpecific Test/EvaluationAcceptance Criteria (Implied by standard and successful outcome)Reported Device Performance
    SterilizationEO sterilization validationConformance to ISO 11135-1:2014 & ISO 10993-7:2008 for residual EtOMet (testing performed)
    PackagingVisual inspectionConformance to ASTM F1886/F1886M-16Met (testing performed)
    Seal Strength testingConformance to ASTM F88/F88M-23Met (testing performed)
    Dye Penetration testConformance to ASTM F1929-15Met (testing performed)
    TransportationTransportation Simulation testingConformance to ASTM D4169-23Met (testing performed)
    BiocompatibilityCytotoxicityNon-cytotoxic per ISO 10993-5:2009Non-cytotoxic
    SensitizationNon-sensitizing per ISO 10993-10:2021Non-sensitizing
    IrritationNon-irritating per ISO 10993-23:2021Non-irritating
    Acute Systemic ToxicityNon-systemically toxic per ISO 10993-11:2017Non-systemically toxic
    Material Mediated PyrogenicityNon-pyrogenic per USPNon-pyrogenic
    Bench Performance (after accelerated aging)Dimensional specificationsMet pre-determined criteriaAll "predetermined acceptance criteria were met"
    Bend test (Force to tip deflection)Met pre-determined criteriaAll "predetermined acceptance criteria were met"
    Cuff Burst TestMet pre-determined criteriaAll "predetermined acceptance criteria were met"
    Tensile strength testing (all joints/connections)Met pre-determined criteriaAll "predetermined acceptance criteria were met"
    Repeated cuff inflationMet pre-determined criteriaAll "predetermined acceptance criteria were met"
    Prolonged inflationMet pre-determined criteriaAll "predetermined acceptance criteria were met"
    Insertion depth marker integrity testingMet pre-determined criteriaAll "predetermined acceptance criteria were met"
    Handle locking mechanism testing (force to move from set position)Met pre-determined criteriaAll "predetermined acceptance criteria were met"

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the sample sizes for the various tests. It refers to established testing standards (e.g., ISO, ASTM, USP), which would specify appropriate sample sizes for each test type. The data provenance is from bench testing and laboratory evaluations conducted by the manufacturer (LI Medical Corporation LTD.) to support regulatory submission. This is a prospective assessment of newly manufactured devices under controlled conditions.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This concept is not applicable to the assessment of a physical medical device like a uterine manipulator. "Ground truth" in this context refers to the defined acceptable parameters for physical and biological performance, which are established by international standards and regulatory guidelines, not by individual experts evaluating images or clinical data.

    4. Adjudication Method for the Test Set

    Not applicable for a physical medical device. Performance is judged against specified engineering, material, and biological safety standards, not through adjudication of expert opinions.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable, as this is not an AI/software device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable, as this is not an AI/software device.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is based on established industry standards and regulatory requirements for medical device safety and efficacy. This includes:

    • Physical and mechanical properties (e.g., dimensions, tensile strength, inflation capabilities).
    • Sterility assurance (e.g., EtO residuals).
    • Biocompatibility (e.g., cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity).
    • Packaging integrity.

    These are quantitative metrics or pass/fail criteria derived from validated test methods.

    8. The Sample Size for the Training Set

    Not applicable, as this is not an AI/software device. There is no "training set" in the context of a physical medical device's performance validation; rather, there are samples tested according to predefined standards.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as this is not an AI/software device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241693
    Manufacturer
    Date Cleared
    2024-09-06

    (86 days)

    Product Code
    Regulation Number
    884.4530
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LKF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FemChec® is intended to instill a consistent alternating pattern of saline and air as a continuous stream of contrast media into the uterus and fallopian tubes to be used in conjunction with an intrauterine catheter for performance of sono-hysterosalpingogram (Sono HSG).

    Device Description

    FemChec® is a dual-barrel contrast media syringe that can be connected to an intrauterine catheter for controlled delivery of saline-air contrast media during sono-hysterosalpingogram (Sono HSG) procedures. Sono HSG consists of an ultrasound evaluation of the fallopian tubes with or without assessment of the uterine cavity. The device operates by retraction of the plunger, which simultaneously fills the two syringes in the device with either air or saline. The device is then connected to a compatible uterine catheter, and when the plunger is depressed, a consistent stream of saline and air is delivered into the uterus and fallopian tubes. The device is provided sterile via EO sterilization and is intended for single-use.

    AI/ML Overview

    The provided text is a 510(k) summary for the FemChec Controlled Saline-Air Device (FCD-250), which is a medical device and not an AI or software device. Therefore, the questions related to AI/algorithm performance, training sets, test sets, ground truth establishment by experts, adjudication methods, and MRMC studies are not applicable to this document. The document focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance data for a physical medical device.

    However, I can extract the acceptance criteria and a summary of the studies performed from the provided text for the FemChec device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Tests/CriteriaReported Device Performance
    SterilizationEthylene Oxide Sterilization Validation per ISO 11135-1:2014, AAMI TIR 28:2016, ISO 10993-7:2008Not explicitly stated "met criteria," but implied that testing was performed to support substantial equivalence.
    PackagingVisual inspection, Bubble Leak test per ASTM F2096-11, Seal Strength testing per ASTM F88/F88M-15Not explicitly stated "met criteria," but implied that testing was performed to support substantial equivalence.
    TransportationTransportation Simulation testing per ASTM D4169-14Not explicitly stated "met criteria," but implied that testing was performed to support substantial equivalence.
    BiocompatibilityCytotoxicity per ISO 10993-5:2009, Sensitization ISO 10993-10:2021, Irritation per ISO 10993-10:2021Tested to be non-cytotoxic, non-sensitizing, and non-irritating.
    Bench PerformanceVisual Inspection, Functional/Cycle Testing, Saline-Air Quantification (before and after accelerated aging to 12 months real-time aging per ASTM F1980-16)All predetermined acceptance criteria were met.

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not applicable as the studies conducted are non-clinical bench performance and material compatibility studies for a physical device, not a performance study involving a test set of data points or images.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    This information is not applicable. Ground truth establishment by experts is relevant for AI/software devices requiring interpretation of data (e.g., medical images), which is not the case for this physical medical device.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods are relevant for studies where multiple readers/experts assess cases, which is not applicable to the non-clinical studies for this physical device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document describes non-clinical performance studies for a physical device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The device is a physical medical instrument, not an algorithm.

    7. The Type of Ground Truth Used

    This information is not applicable in the context of AI/software performance. For the described bench tests, the "ground truth" would be established by the specifications and standards (e.g., flow rates, material properties, sterility levels) defined in the test protocols themselves, which are based on recognized international and ASTM standards.

    8. The Sample Size for the Training Set

    This information is not applicable. The device is a physical medical instrument, not an AI/software device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable. The device is a physical medical instrument, not an AI/software device that requires a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222798
    Manufacturer
    Date Cleared
    2022-12-16

    (91 days)

    Product Code
    Regulation Number
    884.4530
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LKF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Rejoni Intrauterine Catheter is for the delivery of contrast media or saline into the uterine cavity during Hysterosalpingography (HSG) or Sonohysterography (SHG) for examination of the uterus and fallopian tubes.

    When used for HSG. the Rejoni Intrauterine Catheter can be used for evaluation of tubal patency.

    The Rejoni Intrauterine Catheter is used to access the uterine cavity for the delivery of saline for Sonohysterography (SHG).

    Device Description

    The Rejoni Intrauterine Catheter is a sterile, single lumen uterine catheter that facilitates access to the uterus to deliver fluids during Hysterosalpingography (HSG) or Sonohysterography (SHG). The Rejoni Intrauterine Catheter consists of an inner shaft and an additional support sheath over the proximal portion of the catheter shaft to provide additional stiffness. A repositionable stopper ("acom") is connected to the support sheath, which is placed on the catheter shaft, and can be repositioned from 0 cm up to 10 cm from the distal tip of the catheter. The stopper maintains placement of the catheter against the external cervical os during the procedure, and a luer lock-style adapter on the proximal end is available for compatibility with syringes. The distal end contains a straight through hole for delivery of fluids during the procedure.

    AI/ML Overview

    The Rejoni Intrauterine Catheter is a sterile, single-lumen uterine catheter intended for delivering contrast media or saline into the uterine cavity during Hysterosalpingography (HSG) or Sonohysterography (SHG) for examination of the uterus and fallopian tubes. It can also be used for evaluation of tubal patency during HSG.

    Here's an analysis of the acceptance criteria and the study that proves the device meets those criteria:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test CategorySpecific Test / CriterionAcceptance CriteriaReported Device Performance
    Mechanical TestingPerformance - MechanicalIn accordance with device performance specificationsPass
    Performance - FunctionalIn accordance with device performance specificationsPass
    Dimensional VerificationMet design specificationsPass
    Visual AssessmentMet design specificationsPass
    Tensile Testing (luer to inner shaft; acorn to outer shaft)Met design specificationsPass
    Kink TestingMet design specificationsPass
    Leak TestingMet design specificationsPass
    BiocompatibilityCytotoxicity (ISO 10993-5:2009)Non-cytotoxicPass
    Sensitization (ISO 10993-10:2021)Non-sensitizingPass
    Intracutaneous Reactivity (ISO 10993-10:2021)Non-irritatingPass
    SterilitySterility Assurance Level (SAL)10$^{-6}$10$^{-6}$ (validated)
    Sterilization Process ValidationConforms to ISO 11137-1, -2, -3Validated
    Shelf LifeDevice performance over shelf lifeMaintains performance as per mechanical testingPass
    Sterile barrier integrity over shelf lifeMaintains sterile barrierPass
    Package integrity after accelerated aging (ASTM 1980-16)No gross leaks, adequate seal strength, visual integrityPass
    Package integrity after simulated shipping (ISTA-3A:2018, ASTM D4169-16)No gross leaks, adequate seal strength, visual integrityPass

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes used for each specific test (e.g., how many devices were used for tensile testing, biocompatibility, etc.). However, it indicates these were pre-clinical tests (lab-based) rather than clinical studies with patient data. Therefore, data provenance such as country of origin of data or retrospective/prospective characteristics is not applicable in this context.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    Not applicable. The described studies are pre-clinical performance and safety tests (mechanical, biocompatibility, sterility, shelf life) rather than studies that require expert-established ground truth on diagnostic accuracy, for example. The "ground truth" for these tests is defined by established industry standards (ISO, ASTM) and the device's own design specifications.

    4. Adjudication Method:

    Not applicable. As noted above, these are pre-clinical tests following defined protocols and standards, not studies requiring expert adjudication of results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. The device is a physical catheter, not an AI-assisted diagnostic tool requiring evaluation of human reader improvement.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    No, a standalone (algorithm only) study was not done. This device is a manual instrument and does not involve an algorithm.

    7. Type of Ground Truth Used:

    The ground truth for these tests is implicitly or explicitly defined by:

    • Design Specifications: For mechanical and functional performance, the device's own engineering and design requirements serve as the ground truth.
    • International Standards (ISO, ASTM): For biocompatibility, sterility, and packaging/shelf life, the ground truth is established by the specified ISO and ASTM standards and their acceptance criteria (e.g., non-cytotoxic, SAL of 10$^{-6}$).

    8. Sample Size for the Training Set:

    Not applicable. This is a physical medical device, not an AI/machine learning algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223064
    Date Cleared
    2022-10-28

    (28 days)

    Product Code
    Regulation Number
    884.4530
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LKF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ALLY II UPS™ (Uterine Positioning System) is intended to assist the surgical staff in mounting, positioning and holding uterine manipulators during gynecological laparoscopic surgical procedures. It is intended for use by trained operating room personnel in an operating room environment.

    Device Description

    The ALLY II UPS™ (Uterine Positioning System) attaches to the operating room table and enables the bed-side assistant to readily mount, hold, and position the manipulator during laparoscopic surgical procedures. The ALLY II UPS enables access and provides the ability to maneuver and maintain the manipulator in a desired position. The ALLY II UPS consists of the ALLY II UPS and a manipulator adapter with built-in sterile drape, known as the adapter drape.

    The CooperSurgical ALLY II UPS is a non-patient contacting, electromechanical device that consists of a single, multi-segmented, articulated arm. The ALLY II UPS can be attached to the standard operating room bed rail, and a separate, sterile, disposable Adapter Drape that is used to attach a uterine manipulator to the ALLY II UPS. When unlocked, the flexible arm allows the attached manipulator to be positioned by the user. The arm can then be locked in the desired position by releasing a foot pedal, activating a linear actuator that applies tension to an internal cable, drawing the segments together and thus locking the arm. The segmented design of the arm allows lateral/medial movement from a single point to position the uterine manipulator.

    The purpose of this submission is to gain clearance for minor modifications made to the ALL Y Uterine Positioning System to meet user needs and allow for user convenience. The overall functionality and interface of the ALLY II UPS for the user remains the same.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for a medical device called the ALLY II UPS™ (Uterine Positioning System). This type of document is for medical devices, not AI/ML algorithms. Therefore, the questions related to AI/ML specific aspects like training data, ground truth establishment, MRMC studies, and standalone algorithm performance are not applicable to this submission.

    The acceptance criteria and performance data described in this document relate to the physical and mechanical performance of the revamped medical device compared to its predicate.

    Here's an analysis based on the provided text, focusing on the device's acceptance criteria and proven performance:

    1. A table of acceptance criteria and the reported device performance

    The document doesn't provide a direct table of acceptance criteria with corresponding performance values for the ALLY II UPS™. Instead, it broadly states that "The ALLY II UPS met the pre-determined acceptance criteria for each intended output." It lists the types of tests performed.

    Implied Acceptance Criteria and Performance (based on Section VII. Performance Data):

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance (Summary)
    Durability (Useful Life)Device sustains intended function, material integrity, and performance over its expected lifespan."was performed" (details not provided, but implies met criteria given substantial equivalence conclusion)
    Holding ForceDevice can consistently and securely hold uterine manipulators in desired positions without slippage or movement beyond acceptable tolerances."was performed" (details not provided, but implies met criteria given substantial equivalence conclusion)
    Range of MotionDevice can manipulate to designated 6 points of extension in anteverted hand position and designated 5 points of extension in retroverted hand position.Subject device has "greater Range of Motion than the predicate" and met the specified extension points.
    Link SoakMaterials of the flexible arm (17-4 stainless steel, PEEK inserts) maintain integrity and function after exposure to relevant cleaning/sterilization protocols."was performed" (details not provided, but implies met criteria given substantial equivalence conclusion)
    Foot Pedal Cable RetentionThe detachable foot pedal cable remains securely connected during use and does not pose a safety risk."Justification" submitted (details not provided, but implies met criteria given substantial equivalence conclusion)
    Electrical Safety (IEC 60601-1)Compliance with general requirements for basic safety and essential performance of medical electrical equipment.Device is "compliant with the following standards: IEC 60601-1: 2005+AMD1:2012+AMD2:2020 CSV"
    EMC (IEC 60601-1-2)Compliance with requirements for electromagnetic compatibility for medical electrical equipment.Device is "compliant with the following standards: IEC 60601-1-2: 2014"

    2. Sample size used for the test set and the data provenance

    • Sample Size: The document does not specify the sample size for any of the tests (Durability, Holding force, Range of motion, Link Soak, Foot Pedal Cable Retention).
    • Data Provenance: The data provenance is internal testing performed by CooperSurgical, Inc. It is a prospective test in the sense that newly designed units of the ALLY II UPS™ were manufactured and then tested to demonstrate their performance. The country of origin for the testing is not explicitly stated but is implied to be within the company's operational context (likely US, given the FDA submission).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable as this is a mechanical device, not an AI/ML system that requires expert interpretation for ground truth. The "ground truth" for this device would be defined by engineering specifications and objective measurements (e.g., force in Newtons, angles in degrees, material integrity).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable for this type of device. Adjudication methods are typically used in clinical studies or for establishing ground truth from human interpretations (e.g., radiology reads). The performance of this device is assessed through engineering and mechanical testing against predefined specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a mechanical uterine positioning system, not an AI-assisted diagnostic or therapeutic tool with human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a purely mechanical device; there is no AI algorithm to test for "standalone" performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" in this context is defined by engineering specifications, mechanical measurements, and compliance with recognized standards. For example:

    • Holding Force: Measured in units of force (e.g., Newtons or pounds) against a specified minimum.
    • Range of Motion: Measured in degrees or specific positions achievable, defined by design.
    • Durability: Performance over a specified number of cycles or operating hours.
    • Material Integrity: Assessed through material science tests and visual inspection post-soak/cycle.
    • Electrical Safety/EMC: Compliance verified against the specified IEC standards (60601-1 and 60601-1-2).

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As stated above, this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220202
    Date Cleared
    2022-03-31

    (66 days)

    Product Code
    Regulation Number
    884.4530
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LKF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Uterine ElevatOR PRO with OccludOR Balloon™ is indicated for manipulation of the uterus, and injection of fluids or gases during laparoscopic procedures including total laparoscopic hysterectomy, laparoscopic assisted vaginal hysterectomy, laparoscopic tubal occlusion, and diagnostic laparoscopy. The Uterine ElevatOR PRO with OccludOR Balloon™ maintains pneumoperitoneum during laparoscopic procedures by sealing the vagina once colpotomy is performed.

    Device Description

    The Uterine ElevatOR PRO with OccludOR Balloon™ is a sterile, single-use medical device used to manipulate the uterus and cervix in surgical and diagnostic procedures. The device includes an insulated and anatomically designed stainless steel shaft with an intrauterine balloon and reference graduations (centimetres) at the proximal [to patient] end and an external handle at the distal end. The intrauterine balloon is inflated with air using a standard syringe (not included). The device has a vaginal cup that is pushed along the shaft to support a green cervical cup. The cervical cup provides a guide for colpotomy. There are four sizes of cervical cups available, 32mm, 37mm, and 40mm, which correspond to the different model numbers. A locking assembly secures the position of the vaginal cup and therefore the cervical cup. Sites for suturing are located on the cervical cup. An Occluder Balloon further reduces leakage of pneumoperitoneum during laparoscopic surgery. A port at the distal end of handle provides a channel for injection of fluids or gases.

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device called the "Uterine ElevatOR PRO with OccludOR Balloon™", not an AI/ML device. Therefore, the request for acceptance criteria and study details related to AI/ML device performance (such as sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth types) is not applicable.

    The document describes performance data for a physical medical device, which includes:

    1. A table of acceptance criteria and the reported device performance:

    The document broadly states that "In all testing, the subject device met the pre-specified acceptance criteria." However, it does not provide a specific table detailing each acceptance criterion and the corresponding reported device performance. It lists the types of tests conducted:

    Test CategorySpecific Tests PerformedReported Performance
    Biocompatibility- Cytotoxicity per ISO 10993-5:2009
    • Irritation per ISO 10993-10:2010
    • Sensitization per ISO 10993-5:2010
    • Acute Systemic toxicity per ISO 10993-11:2017
    • Pyrogenicity per ISO 10993-11 and USP | "The subject device is considered externally communicating device & surface contact device that may come in contact with tissue and breached or compromised surface respectively for a duration of less than 24 hours."
      "met the pre-specified acceptance criteria." |
      | Performance & Mechanical | - Intrauterine balloon integrity & concentric inspection
    • Intrauterine balloon inflation stability test for a defined time period.
    • Intrauterine balloon burst test
    • Intrauterine balloon fatigue test
    • Intrauterine balloon adhesion strength test
    • OccludOR balloon outer diameter inspection & leakage free test
    • OccludOR balloon deflation time test
    • OccludOR balloon capacity test- reliability/safety
    • OccludOR balloon fatigue durability test
    • OccludOR balloon burst test
    • Vaginal cup air leakage test
    • Cervical cup temperature test
    • Air inflation valve test
    • Dye-injection port compliance & leakage test
    • Thumb screw strength test
    • Locking force test of vaginal cup
    • Ease of movement of vaginal cup along the shaft
    • Cervical cup detachment force test
    • Cervical cup stopper adhesion strength test
    • Device tensile strength
    • Handle joint strength
    • Tip deflection
    • Ultrasonic welding strength | "In all testing, the subject device met the pre-specified acceptance criteria." |

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    The document does not specify sample sizes for any of the tests, nor does it provide information on data provenance (country of origin, retrospective/prospective). This type of detail is typically found in full study reports, not a 510(k) summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    Not applicable, as this is a physical medical device and not an AI/ML device relying on "ground truth" derived from expert consensus. The "truth" for this device's performance is determined by meeting physical and biological specifications through various standard engineering and biocompatibility tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable for a physical medical device as described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable, as this is a physical medical device and not an AI/ML device involving human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable, as this is a physical medical device without an algorithm component.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    Not applicable for a physical medical device. The "ground truth" here is adherence to engineering specifications and biocompatibility standards.

    8. The sample size for the training set:

    Not applicable, as this is a physical medical device and not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set or AI/ML model for this physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212505
    Device Name
    DUMI ManipulatOR
    Date Cleared
    2022-01-10

    (154 days)

    Product Code
    Regulation Number
    884.4530
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LKF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The use of the DUMI ManipulatOR is indicated in diagnostic laparoscopy, mini laparotomy, fertility exams, and salpingoplastic procedures where manipulation of the uterus is required.

    Device Description

    The DUMI ManipulatOR is a sterile, single use, disposable medical device designed for the intra-operative manipulation of both the anteverted and retroverted uterus as well as facilitating simple and effective dye studies. The device includes an anatomically designed balloon at the distal end that is injected with air by syringe through the luer connector of inflation valve. The three-way hub can accommodate fluid and can be used for dye studies.

    AI/ML Overview

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document outlines various tests conducted to demonstrate the safety and effectiveness of the DUMI ManipulatOR, primarily by establishing substantial equivalence to a predicate device. The acceptance criteria are generally "successfully met predetermined acceptance criteria" or demonstrating performance "at least the same as the predicate device." Specific numerical criteria are not always explicitly stated in this summary, but the types of tests and their successful outcome are.

    Test CategorySpecific TestAcceptance Criteria (as stated or implied)Reported Device Performance
    General ComplianceCompliance with ISO 13485:2016, 21 CFR 820Requirements for regulatory purposes metSuccessfully met
    ISO 14971:2019 (Risk Management)Application of risk management to medical devicesSuccessfully met
    SterilizationEO Sterilization ValidationRequirements for development, validation, and routine control of sterilization processSuccessfully met
    EO ResiduesAcceptable residue levelsSuccessfully met
    Bacteriostasis and FungistatisNo unacceptable growthSuccessfully met
    BioburdenAcceptable bioburden levelsSuccessfully met
    Sterility Test (ISO 11737-2)Demonstrated sterilitySuccessfully met
    Shelf LifePackaging integrity post 5 year accelerated agingMaintain integritySuccessfully met
    Performance post 5 year accelerated agingMaintain intended performanceSuccessfully met
    PackagingSeal Strength (ASTM F88)Acceptable seal strengthSuccessfully met
    Dye Penetration (ASTM F1929)No leaksSuccessfully met
    Gross Leaks (ASTM F2096)No gross leaksSuccessfully met
    Packaging for terminally sterilized medical devices (ISO 11607-1, ISO 11607-2)Requirements for materials, sterile barrier systems, packaging systems, formation, sealing, assemblySuccessfully met
    TransportationHandling, Loose Load Vibration, Low Pressure (High Altitude) Hazard, Vehicle VibrationWithstand transportation stresses without damage or compromise to deviceSuccessfully met
    BiocompatibilityCytotoxicity (ISO 10993-5)Non-cytotoxicFound to be non-cytotoxic
    Sensitization (ISO 10993-10)Non-sensitizingFound to be non-sensitizing
    Acute Systemic Toxicity (ISO 10993-11)Non-toxicFound to be non-toxic
    Irritation (ISO 10993-10)Non-irritatingFound to be non-irritating
    General Biocompatibility (ISO 10993-1)Evaluation and testing within a risk management processSuccessfully met
    Performance (Device Specific)Intrauterine balloon capacity and diameter testPerformance at least the same as predicate (10cc volume for balloon)Performed at least the same as predicate (same volume)
    Intrauterine balloon fatigue test (repeat inflation)Performance at least the same as predicatePerformed at least the same as predicate
    Intrauterine balloon burst testPerformance at least the same as predicatePerformed at least the same as predicate
    Air leakage test for prolonged timePerformance at least the same as predicatePerformed at least the same as predicate
    Spring load locking mechanism force testPerformance at least the same as predicatePerformed at least the same as predicate
    Component joint strength testPerformance at least the same as predicatePerformed at least the same as predicate
    Tip deflection testPerformance at least the same as predicatePerformed at least the same as predicate
    Balloon adhesion strength testPerformance at least the same as predicatePerformed at least the same as predicate
    Luer connector leakage test (ISO 80369-7)Performance at least the same as predicate; compliant with standard connectorsPerformed at least the same as predicate; compliant with standard
    REACH Regulation (EC) No 1907/2006- Non-DEHP (EN 14372:2004)Non-DEHPSuccessfully met

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state specific sample sizes for each individual test. However, it references several standards that typically dictate sample size requirements for testing and quality control:

    • ISO 2859-1:1999: "Sampling procedures for inspection by attribute - Part 1: Sampling schemes indexed by acceptance quality limit (AQL) for lot-by-lot inspection." This standard would have guided the sample sizes for attribute inspection tests.
    • ASTM F3172-15: "Design verification device size and sample selection for Endovascular devices." While the DUMI ManipulatOR is not an endovascular device, this standard indicates that sample size selection for design verification was a considered aspect of the testing strategy.

    Given the nature of the device (uterine manipulator) and the tests performed (physical, mechanical, biocompatibility, sterilization, packaging), the data would be prospective as the tests were conducted specifically for this submission.

    The provenance of the data is not explicitly stated per test but is implied to be from testing labs associated with The O.R. Company Pty Ltd, an Australian company with offices in Australia, North America, and Hong Kong. The testing was conducted to meet international and US FDA standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. The tests described are primarily objective, quantitative engineering, microbiological, and chemical tests. Clinical expert "ground truth" (e.g., radiologists, pathologists) is typically relevant for diagnostic devices or AI algorithms that interpret medical data. For a manual surgical instrument, the "ground truth" for performance is established through defined acceptance criteria in engineering standards, material science, and regulatory guidelines.

    4. Adjudication Method for the Test Set:

    This information is not applicable/provided. Adjudication methods (like 2+1, 3+1) are usually relevant in clinical trials or studies where human interpretation or consensus is required to establish a gold standard for disease presence or other clinical outcomes. The tests performed for the DUMI ManipulatOR are laboratory-based and objective, with outcomes determined by measurement against predetermined standards.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not conducted and is not applicable to this device. The DUMI ManipulatOR is a manual surgical instrument, not an AI-powered diagnostic or assistive tool for image interpretation. Therefore, there is no concept of "human readers" or "AI assistance" in its evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    No, a standalone algorithm-only performance assessment was not done. The DUMI ManipulatOR is a physical medical instrument, not an algorithm or AI system.

    7. The Type of Ground Truth Used:

    The "ground truth" for this device's performance is based on:

    • Engineering and Material Standards: Defined specifications and performance limits derived from international and national standards (e.g., ISO, ASTM).
    • Predicate Device Equivalence: The performance of the predicate device (Panpac Uterine Manipulator Injector, Model UMI 4.5) serves as a benchmark for comparison in several performance tests. The new device must perform "at least the same as" the predicate.
    • Regulatory Requirements: Compliance with quality management systems (ISO 13485, 21 CFR 820), risk management (ISO 14971), and biocompatibility guidelines (ISO 10993 series).

    8. The Sample Size for the Training Set:

    This information is not applicable. The DUMI ManipulatOR is not an AI/machine learning device, so there is no concept of a "training set" for an algorithm. The device itself is manufactured, and its performance is verified through testing.

    9. How the Ground Truth for the Training Set was Established:

    This information is not applicable as there is no "training set" for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181770
    Manufacturer
    Date Cleared
    2019-03-29

    (269 days)

    Product Code
    Regulation Number
    884.4530
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LKF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intrauterine Access Balloon Catheter is used for delivery of contrast medium into the uterine cavity during hysterosalpingography (HSG) for examination of the uterus and fallopian tubes, and evaluation of tubal patency. The catheter lumen accepts catheters up to 6 French in diameter to access the uterine cavity for assessment of tubal patency.

    The Selective Salpingography Catheter is intended to be used through a uterine access device and positioned at the tubal ostium for injection of contrast medium for fluoroscopic evaluation of tubal patency.

    Device Description

    The IAB and SSG Catheters facilitate access to the uterus and tubal ostium to deliver contrast media during diagnostic procedures.

    The IAB Catheter is 9 French (Fr) in diameter and has a 23 cm working length. It is designed with a polyurethane shaft and a silicone distal balloon with a 2-cc inflation volume. The catheter includes two lumens, one for balloon inflation and the other can be used to inject contrast medium into the uterus for hysterosalpingography (HSG) and as an access channel that accepts devices up to 6 Fr in diameter for use in assessing tubal patency. The Intrauterine Access Balloon Catheter is packaged with a shipping stylet and a 3 mL syringe that is used to inflate the balloon with sterile media.

    The SSG Catheter is manufactured from braided polyethylene and is designed with an angled distal tip with a radiopaque tungsten band. The catheter is 5.5 Fr in diameter and has a 40 cm working length. The SSG Catheter is intended to be used through a dimensionally compatible access catheter (IAB Catheter) to gain access to the uterine cavity for a more selective visualization of the fallopian tubes when positioned at the tubal ostium.

    Both devices are supplied sterile (ethylene oxide) and are intended for single-use only.

    AI/ML Overview

    The provided document focuses on the K181770 510(k) submission for the Intrauterine Access Balloon Catheter and the Selective Salpingography Catheter, demonstrating substantial equivalence to predicate devices. It does not contain information about a study proving device meets acceptance criteria in the manner described in the request, such as a clinical study with a test set, expert ground truth, multi-reader, multi-case study, or standalone algorithm performance.

    However, the document does list performance data from bench studies that were conducted to ensure reliable design and performance, with predetermined acceptance criteria being met. This implies that the device does meet specific acceptance criteria based on these bench tests.

    Here is the information extracted from the document regarding the acceptance criteria and the "studies" (bench performance tests) that prove the device meets them:

    Acceptance Criteria and Reported Device Performance

    Note: The document describes "predetermined acceptance criteria" rather than specifying precise quantitative values for each. The reported performance indicates that these criteria were met.

    Acceptance Criteria CategorySpecific Test/Performance CharacteristicReported Device Performance
    GeneralSterilization validationMet per ISO 11135-1:2014
    Ethylene oxide residualMet per ISO 10993-7:2008(R)2012
    Transportation simulationMet per ASTM D4169-16
    Package integrity (dye penetration)Met per ASTM F1929-15 after 3 years real-time aging
    Package integrity (seal strength)Met per ASTM F88/F88M-15 after 3 years real-time aging
    Biocompatibility (cytotoxicity)Met per ISO 10993-5:2009
    Biocompatibility (sensitization)Met per ISO 10993-10:2010
    Biocompatibility (irritation)Met per ISO 10993-10:2010
    IAB Catheter SpecificShelf-lifeSupported for two years
    Compatibility (with other components)Met (dimensional compatibility)
    Compatibility (with SSG Catheter)Met (dimensional compatibility)
    Tensile strength (junctions)Peak load value greater than predetermined criterion
    Lumen patencyLumen was patent; predetermined acceptance criteria met
    Balloon leakage and integrityMet predetermined acceptance criteria; no leakage
    Balloon burst volumeMet predetermined acceptance criteria; burst above labeled volume
    Dimensional accuracy (critical dimensions)Met acceptance criteria
    Check-Flo valve leak reductionMet predetermined acceptance criteria for leakage reduction
    SSG Catheter SpecificShelf-lifeSupported for three years
    Compatibility (with IAB Catheter)Dimensional compatibility assessed and met
    Compatibility (with 0.039-inch wire guides)Dimensional compatibility assessed and met
    Tensile strength (junctions)Peak load value greater than predetermined criteria
    Lumen patencyLumen was patent; predetermined acceptance criteria met
    Torque responseMet acceptance criterion (per Cook's internal procedures)
    Radiopacity of marker bandDetermined to be radiopaque in a simulated clinical setting
    Kink radiusAdequately characterized
    Dimensional accuracy (critical dimensions)Met acceptance criteria

    The remaining requested information pertains to clinical performance studies, which are not detailed in this 510(k) submission as no such studies were needed for substantial equivalence based on the provided document. The document lists bench performance studies as the primary evidence.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable / Not provided for the device. The document describes bench performance studies (e.g., tensile testing, lumen patency, balloon integrity). These are laboratory tests on device units, not tests on clinical patient data. Therefore, concepts like "test set" in the clinical sense, "data provenance," "country of origin," or "retrospective/prospective" do not apply.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable / Not provided. The device performance was assessed against engineering specifications and industry standards in bench testing, not against expert clinical ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable / Not provided. Adjudication methods are typically relevant for clinical studies involving human interpretation or subjective assessments, which were not part of the described bench performance tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC study was not done. This type of study is for AI/human-in-the-loop performance, which is not relevant to this medical device submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No, a standalone algorithm performance study was not done. This submission is for a physical medical device, not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Engineering specifications, industry standards, and internal procedures served as the "ground truth" or reference for the bench performance tests. For instance, material properties (tensile strength), dimensional tolerances, and functional requirements (balloon integrity, lumen patency) were assessed against predefined, objective criteria.

    8. The sample size for the training set

    • Not applicable / Not provided. No "training set" in the context of machine learning or clinical data was mentioned or used for this device. The bench tests involved samples of manufactured devices.

    9. How the ground truth for the training set was established

    • Not applicable / Not provided. As there was no training set in the clinical or AI context, no ground truth needed to be established for it. The "ground truth" for the device's acceptable performance was established by engineering design specifications, relevant ISO standards, ASTM standards, and Cook's internal procedures.
    Ask a Question

    Ask a specific question about this device

    K Number
    K180300
    Manufacturer
    Date Cleared
    2018-10-26

    (266 days)

    Product Code
    Regulation Number
    884.4530
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LKF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Margolin HSG Cannula is for the delivery of contrast media or saline into the uterine cavity during Hysterosalpingography (HSG) or Sonohysterography (SHG) for examination of the uterus and fallopian tubes. When used for HSG, the Margolin HSG Cannula can be used for evaluation of tubal patency.

    The Goldstein Sonohysterography Catheter is used to access the uterine cavity for the delivery of saline for Sonohysterography (SHG).

    Device Description

    The Margolin HSG Cannula and Goldstein Sonohysterography Catheter are single lumen uterine catheters that facilitate access to the uterus to deliver fluid (contrast media or saline) during diagnostic procedures. Both feature a silicone "acom" wedge that seals the cervical canal and maintains the position of the catheter during the procedure.

    AI/ML Overview

    This is a 510(k) summary for medical devices (Margolin HSG Cannula and Goldstein Sonohysterography Catheter) and, as such, focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed clinical study for novel performance metrics with acceptance criteria, expert ground truth, or statistical power.

    Therefore, many of the requested data points (e.g., effect size of human readers with/without AI, sample size for training set, specific study design for clinical performance against pre-defined acceptance criteria, detailed ground truth establishment for a test set) are not presented in this document because they are not typically required for a 510(k) submission for these types of devices. The "performance data" section rather lists engineering and biocompatibility tests.

    However, based on the provided text, I can extract the following information:

    1. A table of acceptance criteria and the reported device performance

    The document states: "All predetermined acceptance specifications were met in the following tests." However, the specific numerical acceptance criteria for each test are not explicitly detailed in this summary. The performance is reported as meeting these unspecified specifications.

    Test CategoryAcceptance Criteria (Not Explicitly Stated Numerically)Reported Device Performance
    Sterilization ValidationPer ISO 11135-1:2007 requirementsMet applicable design and performance requirements
    Transportation SimulationPer ASTM D4169-16 requirementsMet applicable design and performance requirements
    Package IntegrityPer ASTM F2096-11 (Bubble Leak), ASTM F88-09 (Seal Strength), ASTM F1866-09 (Visual Inspection) requirements after aging to three yearsMet applicable design and performance requirements
    BiocompatibilityPer ISO 10993-5:2009 (Cytotoxicity), ISO 10993-10:2010 (Sensitization), ISO 10993-10:2010 (Irritation)Met applicable design and performance requirements
    Dimensional VerificationConformance with Cook's predetermined acceptance specifications for catheter, stylet, ink marker bands, silicone positioner dimensionsVerified conformance to predetermined acceptance specifications
    Lumen Patency and Liquid Leakage TestFluid path of catheter patent and no leakage from joints/parts at predetermined injection pressureFluid path was patent and did not leak
    Tensile TestingPeak load value greater than predetermined acceptance specification (for hub to shaft (both devices), hub to stylet wire (Margolin HSG Cannula), and tip to shaft (Margolin HSG Cannula))Peak load value was greater than the predetermined acceptance specification
    Positioner Force to DisplaceConformance to Cook requirements for force needed to displace movable silicone positioner (Goldstein Sonohysterography Catheter)Predetermined acceptance specification was met
    Stylet Compatibility TestingStylet fits within catheter lumen and does not extend beyond tip (Margolin HSG Cannula)Predetermined acceptance specification was met

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample sizes for the "performance data" tests. The data provenance is not mentioned, as these are primarily engineering and bench tests, not clinical studies in a specific geographical location. The studies are prospective in the sense that they are conducted on manufactured devices to test design specifications.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. These are engineering performance tests, not clinical performance studies requiring expert ground truth for interpretation (e.g., image analysis, diagnosis).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This concept applies to clinical studies where multiple reviewers' opinions need to be reconciled to establish a ground truth, which is not the case for these bench tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. These are medical devices (cannulas and catheters) used for fluid delivery, not AI-powered diagnostic or assistive tools. Therefore, an MRMC study related to human readers improving with AI assistance is entirely irrelevant to this document.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. As stated above, these are physical medical devices, not algorithms.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the performance tests listed, the "ground truth" is defined by the technical specifications and standards (e.g., ISO 11135-1:2007 for sterilization, ASTM D4169-16 for transportation, ASTM F2096-11 for bubble leak, ISO 10993 series for biocompatibility, and "Cook's predetermined acceptance specifications" for dimensional, patency, tensile, and force tests). There's no clinical ground truth established by experts or pathology for these types of tests.

    8. The sample size for the training set

    Not applicable. There is no AI or machine learning model being "trained" for these medical devices.

    9. How the ground truth for the training set was established

    Not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180291
    Manufacturer
    Date Cleared
    2018-10-26

    (267 days)

    Product Code
    Regulation Number
    884.4530
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    LKF

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cook® Silicone Balloon HSG Catheter is used for delivery of contrast medium or saline into the uterine cavity and fallopian tubes for the evaluation of tubal patency using hysterosalpingography, or to access the uterine cavity for saline infusion sonohysterography.

    Device Description

    The Cook® Silicone Balloon HSG Catheter is a dual lumen polyurethane catheter that is manufactured in 5.0 or 7.0 French (Fr) sizes and is 30 cm in length. The Cook® Silicone Balloon HSG Catheter is available with an open-end tip, or a closed-end tip with a single side port. The silicone balloon is designed with a balloon volume of 1.0 mL (5.0 Fr) or 1.5 mL (7.0 Fr). Depending upon the device a 1 or 3 mL syringe is provided with the balloon catheter.

    The Cook® Silicone Balloon HSG Catheter will be supplied sterile (ethylene oxide) and is intended for one-time use. The set is packaged in a peel-open pouch. The 5.0 Fr size catheter has a three-year shelf-life, and the 7.0 Fr size catheter has a two-year shelf-life.

    AI/ML Overview

    This document describes the safety and effectiveness testing for the Cook® Silicone Balloon HSG Catheter, but it is not an AI/ML medical device submission. Therefore, it does not contain the specific information requested about acceptance criteria, study details (sample size, data provenance, expert ground truth, adjudication, MRMC studies, standalone performance, training set details), which are typically required for AI/ML device submissions.

    The provided document is a 510(k) Premarket Notification for a physical medical device. It demonstrates substantial equivalence to a predicate device through bench performance studies and other non-clinical tests.

    However, I can extract and present the available information in a format that addresses some aspects of your request, even if it doesn't align with an AI/ML context:


    Acceptance Criteria and Device Performance (Based on available information for a physical device)

    Test CategoryAcceptance Criteria (Pre-defined Specifications)Reported Device Performance
    Sterilization ValidationPer ISO 11135-1:2007 requirementsAll predetermined acceptance criteria were met.
    BiocompatibilityPer ISO 10993-5:2009 (Cytotoxicity), ISO 10993-10:2010 (Sensitization, Intracutaneous Irritation) requirementsAll predetermined acceptance criteria were met for Cytotoxicity, Guinea Pig Maximization Sensitization, and Intracutaneous Irritation testing.
    Transportation SimulationPer ASTM D4169-16 requirementsAll predetermined acceptance criteria were met.
    Package IntegrityPer ASTM F2096-11 (Bubble Leak), ASTM F88-09 (Seal Strength), ASTM F1866-09 (Visual Inspection) after aging (2 or 3 years shelf-life)All predetermined acceptance specifications were met for Bubble Leak test, Seal Strength testing, and Visual Inspection after aging.
    Simulated UseCatheter able to be inserted through cervical canal of nulliparous cervix model, syringe able to mate with check valve, inflate balloon to labeled volume, and deflate fully.Testing demonstrated that the catheter was able to be inserted through the cervical canal of a model of a nulliparous cervix, the syringe was able to mate with catheter check valve, inflate balloon to labeled volume, and deflate balloon fully. (All predetermined acceptance specifications were met).
    Balloon Integrity TestCatheter and balloon show no abnormalities (breakage, leakage, cracking at joint/material) when filled to labeled volume. No leakage after 10 mins in 37°C water bath.Testing demonstrated that the catheter and balloon did not show any abnormalities such as breakage, leakage, or cracking at the balloon joint or balloon material when filled to the labeled volume. It also demonstrated that the catheter and balloon did not leak after being filled to the labeled volume and submerging in a 37°C water bath for a minimum of 10 minutes. (All predetermined acceptance specifications were met).
    Balloon Diameter & Max. Balloon VolumeTest results meet pre-defined acceptance specifications for balloon diameter with max fill, and max balloon pressure/volume at catastrophic failure.Test results met the pre-defined acceptance specifications for these devices (balloon diameter when filled with maximum fill volume, and maximum balloon pressure and volume at catastrophic failure). (All predetermined acceptance specifications were met).
    Lumen Patency & Liquid LeakageFluid path is patent and does not leak under a predetermined injection pressure.Testing demonstrated that the fluid path of the catheter was patent and did not leak under a predetermined injection pressure. (All predetermined acceptance specifications were met).
    Tensile TestingPeak load value greater than predetermined acceptance specification for hub to shaft, shaft, shaft to manifold, extension tubing to manifold.Testing demonstrated that the peak load value was greater than the predetermined acceptance specification. (All predetermined acceptance specifications were met).

    Summary of Study Details (for a physical medical device, not AI/ML):

    1. Sample Size used for the test set and data provenance:

      • The document implies that an adequate number of units were tested to demonstrate statistical significance and meet pre-defined acceptance criteria for bench performance and other non-clinical tests.
      • Specific numerical sample sizes for each test are not provided in this summary.
      • Data Provenance: The tests are described as "Performance Data" and "Bench Performance studies," indicating the data was generated from laboratory testing of the physical catheter devices. Country of origin for data or retrospective/prospective nature is not applicable as this is not a clinical study on patient data.
    2. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

      • Not applicable. This is not an AI/ML device that generates diagnostic outputs requiring expert interpretation for ground truth. The "ground truth" for this physical device is defined by the physical properties and performance specifications of the device itself (e.g., balloon inflates, no leaks, proper tensile strength).
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. This is not an AI/ML device or a clinical study requiring adjudication of diagnostic interpretations.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is a physical medical device, not an AI/ML software.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a physical medical device, not an AI/ML software.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for this device's performance is based on engineering specifications and testing standards (e.g., ISO, ASTM standards, pre-defined internal acceptance specifications).
    7. The sample size for the training set:

      • Not applicable. This is not an AI/ML device that requires a training set.
    8. How the ground truth for the training set was established:

      • Not applicable. This is not an AI/ML device that requires a training set.

    Conclusion from the document:

    "The results of the performance testing described above demonstrates that the Cook® Silicone Balloon HSG Catheter is as safe and effective as the predicate device and supports a determination of substantial equivalence." (Page 6)

    Ask a Question

    Ask a specific question about this device

    Page 1 of 7