K Number
K242002

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-11-22

(136 days)

Product Code
Regulation Number
884.4530
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

FemVue MINI is intended to instill a consistent alternating pattern of saline and air as a continuous stream of contrast media into the uterus and fallopian tubes to be used in conjunction with an intrauterine catheter for performance of sono-hysterosalpingogram (Sono HSG).

Device Description

Fem Vue® MINI is a dual-barrel contrast media syringe that can be connected to an intrauterine catheter for controlled delivery of saline-air contrast media during sono-hysterosalpingogram (Sono HSG) procedures. Sono HSG consists of an ultrasound evaluation of the fallopian tubes with or without assessment of the uterine cavity. The device operates by retraction of the plunger, which simultaneously fills the two syringes in the device with either air or saline. The device is then connected to a compatible uterine catheter, and when the plunger is depressed, a consistent stream of saline and air is delivered into the uterus and fallopian tubes. The device is provided sterilization and is intended for single-use.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the studies mentioned for the FemVue MINI Saline-Air Device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of acceptance criteria with corresponding performance metrics. Instead, it states that "all predetermined acceptance criteria were met" for various tests. Therefore, the table below is constructed by inferring the acceptance criteria from the type of test conducted.

Acceptance Criterion (Inferred from Test)Reported Device Performance
Sterility (achieves sterility after Ethylene Oxide sterilization)Met (per ISO 11135:2014, AAMI TIR 28:2016, ISO 10993-7: 2008)
Package Integrity (maintains sterility barrier)Met (per Visual inspection, ASTM F2096-11, ASTM F88/ F88M-23, ASTM F1886/F1886-16)
Withstand Transportation (maintains integrity after transport simulation)Met (per ASTM D4169-22)
Biocompatibility (non-cytotoxic, non-sensitizing, non-irritating)Met (non-cytotoxic, non-sensitizing, and non-irritating per ISO 10993-5: 2009, ISO 10993-10: 2021)
Visual Inspection (meets specified visual appearance)Met (before and after accelerated aging)
Functional/Cycle Testing (operates as intended through its lifecycle)Met (before and after accelerated aging)
Saline-Air Quantification (delivers consistent alternating pattern of saline and air)Met (before and after accelerated aging)

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample sizes used for the individual bench performance tests. It only states that these tests were conducted. It also does not explicitly state the data provenance (e.g., country of origin) or if the studies were retrospective or prospective, though bench testing is inherently prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. The tests described are primarily physical, chemical, and functional bench tests, rather than clinical studies requiring expert ground truth for interpretation.

4. Adjudication Method for the Test Set

This information is not provided in the document. Given the nature of the bench tests, it's unlikely a formal adjudication method (like 2+1, 3+1) would be applicable. The tests likely have objective pass/fail criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document describes non-clinical performance data for a device, not a diagnostic algorithm that would typically require such a study to assess human reader improvement with AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No, a standalone performance study (algorithm only) was not done. This device is a manual instrument, not an AI or software algorithm.

7. The Type of Ground Truth Used

For the non-clinical performance studies described, the "ground truth" refers to established engineering standards, material specifications, and functional requirements for a medical device. This is determined by the specific ASTM and ISO standards cited (e.g., ISO 11135 for sterility, ISO 10993 for biocompatibility, ASTM F1980-16 for accelerated aging). The concept of expert consensus, pathology, or outcomes data as ground truth is not applicable to these types of engineering and functional tests.

8. The Sample Size for the Training Set

This information is not applicable and therefore not provided. This device is a hardware product, not a machine learning model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable and therefore not provided, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 22, 2024

Femasys Inc. Carrie Engleman Associate Director of Regulatory 3950 Johns Creek Ct. Suite 100 Suwanee, Georgia 30024

Re: K242002

Trade/Device Name: FemVue MINI Saline-Air Device Regulation Number: 21 CFR 884.4530 Regulation Name: Obstetric-Gynecologic Specialized Manual Instrument Regulatory Class: II Product Code: LKF Dated: July 9, 2024 Received: July 9, 2024

Dear Carrie Engleman:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

{2}------------------------------------------------

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Monica D. Garcia -S

Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K242002

Device Name

FemVue MINI Saline-Air Device

Indications for Use (Describe)

FemVue MINI is intended to instill a consistent alternating pattern of saline and air as a continuous stream of contrast media into the uterus and fallopian tubes to be used in conjunction with an intrauterine catheter for performance of sono-hysterosalpingogram (Sono HSG).

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary - K242002

I. SUBMITTER

Applicant:Femasys Inc.
Applicant Address:3950 Johns Creek Court, Suite 100Suwanee, GA 30024
Contact Primary:Carrie EnglemanAssociate Director, RegulatoryPhone: 770-500-3910 x 138Email: Cengleman@femasys.com
Contact Secondary:Christine ThomasSenior VP, Regulatory and ClinicalPhone: 770-500-3910 x 137Email: Cthomas@femasys.com
Date Prepared:November 20, 2024
II. DEVICE
Trade Name:FemVue® MINI Saline-Air Device
Common Name:Contrast media syringe
Regulation Name:Obstetric-gynecologic specialized manual instrument
Regulation Number:21 CFR 884.4530
Regulatory Class:II
Product Code:LKF (Cannula Manipulator/Injector Uterine)

III. PREDICATE DEVICE

FemVue® Saline-Air Device (K110288), manufactured by Femasys Inc.

The predicate device has not been subject to a design-related recall.

IV. DEVICE DESCRIPTION

Fem Vue® MINI is a dual-barrel contrast media syringe that can be connected to an intrauterine catheter for controlled delivery of saline-air contrast media during sono-hysterosalpingogram (Sono HSG) procedures. Sono HSG consists of an ultrasound evaluation of the fallopian tubes with or without assessment of the uterine cavity.

{5}------------------------------------------------

The device operates by retraction of the plunger, which simultaneously fills the two syringes in the device with either air or saline. The device is then connected to a compatible uterine catheter, and when the plunger is depressed, a consistent stream of saline and air is delivered into the uterus and fallopian tubes. The device is provided sterilization and is intended for single-use.

INDICATIONS FOR USE V.

Fem Vue® MINI is intended to instill a consistent alternating pattern of saline and air as a continuous stream of contrast media into the uterus and fallopian tubes to be used in conjunction with an intrauterine catheter for performance of sono-hysterosalpingogram (Sono HSG).

COMPARISON OF THE INTENDED USE AND TECHNOLOGICAL VI. CHARACTERISTICS OF THE SUBJECT AND PREDICATE DEVICES

The intended use and key technological characteristics of the subject and predicate device are compared in the table below.

ComparisonItemFemVue® MINI Saline-Air DeviceSubject DeviceFemVue® Saline-Air DevicePredicate DeviceComparison
510(k)K242002K110288N/A
RegulationNumber884.4530884.4530Same
Device ClassClass IIClass IISame
Product CodeLKFLKFSame
DeviceClassificationNameCannula, Manipulator/Injector, UterineCannula, Manipulator/Injector, UterineSame
RegulationDescriptionObstetric-gynecologic specializedmanual instrumentObstetric-gynecologic specializedmanual instrumentSame
RegulationMedicalSpecialtyObstetrics/GynecologyObstetrics/GynecologySame
Indications forUseFemVue MINI is intended to instill aconsistent alternating pattern of salineand air as a continuous stream ofcontrast media into the uterus andfallopian tubes to be used inconjunction with an intrauterinecatheter for performance of sono-hysterosalpingogram (Sono HSG).The FemVue Saline-Air Device instillsa consistent alternating pattern of salineand air as a continuous stream ofcontrast media into the uterus andfallopian tubes to be used inconjunction with an intrauterinecatheter for performance of sono-hysterosalpingogram (Sono HSG).Same
DesignFeaturesDual-barrel contrast media syringe thatcan be connected to an intrauterinecatheter.Dual-barrel contrast media syringe thatcan be connected to an intrauterinecatheter.Same
Principle ofOperationSterile dual-barrel contrast mediasyringe that is connected to anintrauterine catheter to instill saline-aircontrast medium as part of a sono-Sterile dual-barrel contrast mediasyringe that is connected to anintrauterine catheter to instill saline-aircontrast medium as part of a sono-Same

Table 1. Intended use and technological characteristics comparison.

{6}------------------------------------------------

ComparisonItemFemVue® MINI Saline-Air DeviceSubject DeviceFemVue® Saline-Air DevicePredicate DeviceComparison
hysterosalpingogram (Sono HSG)procedure.hysterosalpingogram (Sono HSG)procedure.
PrescriptionOnlyYesYesSame
Target PatientPopulationPatient undergoing Sono HSGprocedure.Patient undergoing Sono HSGprocedure.Same
Shelf Life12 months60 monthsDifferent.Thedifferencesdo not raisedifferentquestions ofsafety andeffectiveness.
Syringe5 mL10 mLDifferent.The differentsize of thesyringes doesnot raisedifferentquestions ofsafety andeffectiveness.

The subject and predicate device have the same intended use - to instill a consistent alternating pattern of saline and air as a continuous stream of contrast media through a connected intrauterine catheter into the uterus and fallopian tubes for performance of sonohysterosalpingogram (Sono HSG) procedure. The subject and predicate device have the same principle of operation (dual-barrel syringe device that delivers saline and air contrast through an intrauterine catheter). However, minor modifications have been made to the device dimensions and the materials to reduce the overall size of the FemVue MINI device in comparison to the predicate device. These differences do not raise different questions of safety and effectiveness.

SUMMARY OF NON-CLINICAL PERFORMANCE DATA VII.

The following studies have been performed to support substantial equivalence to the predicate device:

· Ethylene Oxide Sterilization Validation testing per:

-ISO 11135:2014 -AAMI TIR 28:2016 -ISO 10993-7: 2008

{7}------------------------------------------------

  • · Package Integrity testing:
    -Visual inspection -Bubble Leak test per ASTM F2096-11 -Seal Strength testing per ASTM F88/ F88M-23 -Seal Visual Inspection per ASTM F1886/F1886-16

  • · Transportation Simulation testing per ASTM D4169-22

  • · Biocompatibility studies conducted in accordance with the 2023 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process". Testing included the following assessments:

-Cytotoxicity per ISO 10993-5: 2009 -Sensitization ISO 10993-10: 2021 -Irritation per ISO 10993-10: 2021

Testing showed the device material to be non-cytotoxic, non-sensitizing, and non- irritating.

  • · Bench performance studies before and after accelerated aging to the equivalent of 12-months of real-time aging in accordance with ASTM F1980-16 demonstrated that all predetermined acceptance criteria were met in the following tests:
    -Visual Inspection -Functional/Cycle Testing -Saline-Air Quantification

VIII. CONCLUSIONS

The results of the testing demonstrate that the Fem Vue MINI Saline-Air Device are as safe and effective as the predicate device and supports a determination of substantial equivalence.

§ 884.4530 Obstetric-gynecologic specialized manual instrument.

(a)
Identification. An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes.
(2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant.
(3) An umbilical clamp is an instrument used to compress the umbilical cord.
(4) A uterine curette is an instrument used to scrape and remove material from the uterus.
(5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix.
(6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus.
(7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination.
(8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix.
(9) A gynecological cerclage needle is a looplike instrument used to suture the cervix.
(10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus.
(11) A gynecological fibroid screw is an instrument used to hold onto a fibroid.
(12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity.
(13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina.
(14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures.
(15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus.
(16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis.
(17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina.
(18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
(b)
Classification. (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.