Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K223791

    Validate with FDA (Live)

    Device Name
    TalWire
    Date Cleared
    2023-07-14

    (207 days)

    Product Code
    Regulation Number
    870.1330
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TalWire is intended for percutaneous entry of peripheral vessels using the Seldinger Technique. The TalWire is not intended for use in the coronary or cerebral vasculature.

    Device Description

    The TalWire is a family of vascular access mandrel guidewires. The TalWire is available in various sizes: between 40 to 80 cm in length and in the 0.018" diameter range. The TalWire is available with Nitinol (NiTi) cores and Stainless Steel coils. The TalWire consist of a solid core shaft with a ground tapered section at the distal end of the guidewire. A microcoil is wound with a lumen that is then placed over the tapered distal end of the coil is secured to the shaft via a weld and the proximal end is bonded to the shaft using adhesive. The TalWire includes an articulatable tip at the distal end of the guidewire. The finished guidewire is placed in a protective polymer dispenser hoop in a labeled Tyvek pouch. The TalWire is provided as a sterile, single-use device, and is sterilized using a validated ethylene oxide process.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called TalWire, a catheter guidewire. It compares the TalWire to a predicate device and includes performance data to support substantial equivalence.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that "All tests passed and met the predefined acceptance criteria." However, it does not explicitly list the quantitative acceptance criteria for each performance test. It only lists the types of performance tests conducted.

    TestAcceptance Criteria (Not Explicitly Stated, Assumed to be "Pass")Reported Device Performance
    Dimensional Attributes(Assumed: Meets design specifications)Passed
    Visual Inspection(Assumed: No defects)Passed
    Simulated Use(Assumed: Performs as intended in simulated environment)Passed
    Tensile Strength(Assumed: Meets strength requirements)Passed
    Torque Strength(Assumed: Meets torque requirements)Passed
    Torqueability(Assumed: Adequate torque response)Passed
    Corrosion Resistance(Assumed: No significant corrosion)Passed
    Kink Resistance(Assumed: Resists kinking)Passed
    Tip Flexibility(Assumed: Adequate flexibility)Passed
    Radiopacity(Assumed: Visible under fluoroscopy)Passed
    Flex Resistance(Assumed: Resists fatigue/damage from flexing)Passed
    Fracture Resistance(Assumed: Resists fracture)Passed
    Prolapse Force(Assumed: Meets prolapse force requirements)Passed

    Biocompatibility:

    • Evaluation in compliance with ISO 10993-1.
    • Specific tests conducted: Cytotoxicity, Guinea Pig Maximization Sensitization, Intracutaneous Irritation, Material Mediated Pyrogenicity, Acute Systemic Toxicity, ASTM Hemolysis, SC5b-9 Complement Activation Assay, Standard Thrombogenicity in Canine.
    • Reported device performance: "The device was found biocompatible."

    Sterilization, Packaging and Shelf Life Testing:

    • EtO Sterilization validation (SAL 10^-6).
    • Shelf life and packaging testing after environmental conditioning and transportation simulation.
    • Reported device performance: "All tests were successfully completed."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes for any of the performance tests.
    The data provenance is not explicitly stated beyond being "performance data." This type of testing is typically done in a lab setting rather than using patient data, so concepts like "country of origin" or "retrospective/prospective" studies are not applicable in the usual sense for this kind of device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This information is not applicable as the tests described are performance tests of the physical properties and functionality of the guidewire, not assessments of diagnostic images or clinical outcomes that would require expert human interpretation as "ground truth." These tests rely on standardized methods and measurement equipment.

    4. Adjudication Method for the Test Set

    This information is not applicable for the same reasons as point 3. Performance tests are typically objectively measured against predefined criteria, not adjudicated by experts.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The document focuses on demonstrating the substantial equivalence of the device through performance testing against established standards and comparison to a predicate device, not on assessing human reader performance with or without AI assistance. This device is a physical medical device (guidewire), not an AI-driven diagnostic tool.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    No, a standalone algorithm performance study was not done. This device is a physical guidewire, not an algorithm.

    7. The Type of Ground Truth Used

    For the performance and biocompatibility testing, the "ground truth" is established by:

    • Standardized test methods: Based on FDA guidance "Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling Guidance for Industry and Food and Drug Administration Staff" and ISO 10993-1 for biocompatibility.
    • Predefined acceptance criteria: While not explicitly listed, these criteria would typically be quantitative thresholds or qualitative observations (e.g., "no visible corrosion," "no fracture").

    8. The Sample Size for the Training Set

    This information is not applicable. This is a physical medical device, not an AI model that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1