Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K130618

    Validate with FDA (Live)

    Device Name
    SFI-ANCHOR
    Date Cleared
    2014-03-03

    (361 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SFI-Anchor® D20 is intended for the fixation of (dental) prostheses to corresponding dental implants. The compatible implant systems are specified in Table 1 below. The SFI-Anchor® CD20 is intended as an additional retaining element on CAD / CAM milled dental bars.

    Device Description

    The SFI-Anchor® D20 is an endosseous dental implant abutment for the fixation of (dental) prostheses to the following ITI Dental Implant System® / Straumann® Dental Implant Systems: Standard Ø 4.1 mm Regular Neck (RN) / Standard Ø 4.8 mm Regular Neck (RN) / Standard Plus Ø 4.1 mm Regular Neck (RN) / Standard Plus Ø 4.8 mm Regular Neck (RN) / Tapered Effect Ø 4.1 mm Regular Neck (RN) / Bone Level Ø 4.1 mm, Regular CrossFit® (RC) / Bone Level Ø 4.8 mm, Regular CrossFit® (RC). The SFI-Anchor® CD 20 is intended as an additional retaining element on dental bars. The SFI-Anchor® D20 are straight abutments with gingival heights ranging from 1mm to 5 mm. The SFI-Anchor® CD 20 is a straight bar attachment. The SFI-Anchor® Housing and the SFI-Anchor® Retention Inserts can correct for up to 10° of divergence (20° of divergence between two implants). The SFI-Anchor® is offered in a diameter 4.8 mm. The SFI-Anchor® D20 is connected to the corresponding implants via a mating screw thread and interface geometry. The SFI-Anchor® CD20 is directly screwed on a dental bar construct. The SFI-Anchor® Housing is cemented in the dental prostheses. The SFI-Anchor® Housing, along with an SFI-Anchor® Retention Insert, allow connection of the prosthesis to the abutment / bar attachment by retention-grip. The abutments / bar attachment are machined from Titanium Alloy conforming to ASTM F-136. Other SFI-Anchor® key functional parts are made out of Elitor® (conforming to standard ISO 22674) and of Pekkton® conforming to standard ASTM D3418, ASTM D638 and ASTM D790.

    AI/ML Overview

    The provided text describes a dental implant abutment, SFI-Anchor® D20 and SFI-Anchor® CD20, and its substantial equivalence to predicate devices. It does not contain information about acceptance criteria and a study proving the device meets those criteria in the context of AI/ML performance.

    Therefore, I cannot provide the requested information. The document focuses on the mechanical and material characteristics of the dental abutments, their compatibility with implant systems, and a comparison with existing predicate devices to establish substantial equivalence for regulatory approval. It is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131526

    Validate with FDA (Live)

    Device Name
    SFI-BAR
    Date Cleared
    2013-09-23

    (118 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SFI-Bar® is intended to be used with the implant manufacturer's implant (Table 1) to provide support for fixation of overdentures.

    Device Description

    The SFI-Bar® provides the connection between compatible dental implant systems for the fixation of removable overdentures. The SFI-Bar@ consists of an implant adapter (abutinent) and a stress-free bar for the fixation of removable overdentures. The implant adapter is screwed into the dental implant.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the SFI-Bar® based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Performance Characteristics)Reported Device Performance
    Torque testsMet acceptance criteria
    Application testingMet acceptance criteria
    Functional testingMet acceptance criteria
    Performance equivalence to predicate SFI-Bar®Equivalent

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a separate "test set" and its sample size in the context of clinical data for the SFI-Bar®. The performance evaluation relies on non-clinical test data.

    Data Provenance: The testing was conducted as part of a 510(k) submission, originating from Switzerland (Cendres + Métaux SA, Biel/Bienne, Switzerland). The data is retrospective in the sense that it's an evaluation against established performance metrics and comparison to predicate devices, rather than a prospective clinical trial.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided as the ground truth for this device's performance evaluation was based on engineering and performance testing rather than expert-derived clinical ground truth.

    4. Adjudication Method for the Test Set

    This information is not applicable as the evaluation did not involve human interpretation or adjudication processes like those in clinical image analysis studies. The adjudication for performance was based on meeting technical specifications from engineering tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices, which is not the nature of the SFI-Bar®.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This concept is not applicable to the SFI-Bar®. The device is a physical dental abutment, not an algorithm. The performance evaluation was of the physical device's mechanical properties and function.

    7. The Type of Ground Truth Used

    The ground truth for evaluating the SFI-Bar®'s performance was based on:

    • Engineering specifications and standards: Adherence to established mechanical properties (e.g., torque, application, functional requirements).
    • Performance characteristics of predicate devices: The SFI-Bar® was compared for substantial equivalence to legally marketed predicate devices, meaning their established safe and effective performance served as a benchmark.

    8. The Sample Size for the Training Set

    This information is not applicable. The SFI-Bar® is a physical medical device, not an AI/ML algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable as there is no "training set" for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123638

    Validate with FDA (Live)

    Device Name
    PEKKTON IVORY
    Date Cleared
    2013-02-21

    (87 days)

    Product Code
    Regulation Number
    872.3770
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Pekkton® ivory is intended to be used for long-term, temporary dental crowns and bridges frameworks.

    Device Description

    Pekkton® ivory is a semi-crystalline thermoplastic supplied in blanks. Pressing the blank gives a framework, from which the dental technician manufactures the final veneered dental crown and / or temporary dental bridge.

    AI/ML Overview

    This document describes the premarket notification for the Pekkton® ivory dental device. The information provided focuses on demonstrating substantial equivalence to predicate devices and verifying the biological safety and performance of the device.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance metrics in a clear, summarized format. However, it lists several biological evaluation tests and states that the "acceptance criteria were met."

    Test TypeAcceptance Criteria (Implied)Reported Device Performance
    Biocompatibility Tests:
    Cytotoxicity (EN ISO 10993-5)No cytotoxic potentialMeets requirements: Investigated material Pekkton® ivory can be evaluated to have no cytotoxic potential.
    Irritation (EN ISO 10993-10)AcceptableMeets requirements: Biological risks evaluated; clinical evaluation and testing demonstrated safety.
    Delayed Type Hypersensitivity (EN ISO 10993-10)AcceptableMeets requirements: Biological risks evaluated; clinical evaluation and testing demonstrated safety.
    Acute Systemic Toxicity (EN ISO 10993-11)AcceptableMeets requirements: Biological risks evaluated; clinical evaluation and testing demonstrated safety.
    Subchronic and chronic systemic toxicity (EN ISO 10993-11)AcceptableMeets requirements: Biological risks evaluated; clinical evaluation and testing demonstrated safety.
    Implantation (EN ISO 10993-6)AcceptableMeets requirements: Biological risks evaluated; clinical evaluation and testing demonstrated safety.
    Genotoxicity (EN ISO 10993-3)No genotoxic potentialMeets requirements: Biological risks evaluated; clinical evaluation and testing demonstrated safety.
    Carcinogenicity (EN ISO 10993-3)No carcinogenic potentialMeets requirements: Biological risks evaluated; clinical evaluation and testing demonstrated safety.
    Chemical characterization (EN ISO 10993-18)No potentially organic toxic leachable substancesMeets requirements: None of the investigated extracts exhibited any product-related peaks, indicating no potentially organic toxic leachable substances were released.
    USP Plastic Class VIMeets USP 34 (88) requirementsMeets requirements: Systemic injection tests exhibited no adverse clinical signs. Concluded that the test material Pekkton® ivory meets the requirements of the USP Plastic Class VI.
    Physical/Mechanical Tests:
    Chemical and physical propertiesAccording to material data sheet (Predicate benchmarks)Meets requirements: Tested and shown to be equivalent in performance characteristics to the predicate devices.
    Bond strength to veneering compositesAcceptable levelsMeets requirements: Tested and shown to be equivalent in performance characteristics to the predicate devices.
    Static fracture loadAcceptable levelsMeets requirements: Tested and shown to be equivalent in performance characteristics to the predicate devices.
    Fatigue testsAcceptable levelsMeets requirements: Tested and shown to be equivalent in performance characteristics to the predicate devices.
    Simulated use testsAcceptable performanceMeets requirements: Tested and shown to be equivalent in performance characteristics to the predicate devices.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify the exact sample sizes for each test in the test set. It mentions that the studies were performed "in compliance with international GLP regulations," which implies controlled laboratory settings. The data provenance is from Cendres+Métaux SA, Switzerland, and the testing was retrospective (meaning conducted specifically for the 510(k) submission, not as part of an ongoing clinical trial).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not provided. The testing described is primarily laboratory-based (biocompatibility, chemical, physical) rather than relying on human expert interpretation for ground truth.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    This information is not applicable/provided. The tests are objective laboratory measurements, not reliant on expert adjudication of results.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    This information is not applicable/provided. The device is a dental material (Pekkton® ivory) and not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study is not relevant to this submission.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable/provided. The device is a material, not an algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the test set was established through standardized laboratory testing protocols and established scientific methods, as defined by international standards (e.g., ISO 10993 series, USP 34 <88>) for biocompatibility, and material science methodologies for chemical and physical properties. For example:

    • Biocompatibility: The "ground truth" for cytotoxicity was the observable impact on L929 mouse fibroblasts. For chemical characterization, it was the absence of product-related peaks in GG/MS / ICP analyses.
    • Physical/Mechanical Properties: The "ground truth" was based on quantitative measurements (e.g., bond strength, fracture load, fatigue) compared against benchmarks from predicate devices or established material data sheets.

    8. The Sample Size for the Training Set

    This information is not applicable/provided. The submission describes a device validation through physical and biological testing, not a machine learning model that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable/provided as there is no training set for a machine learning model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K111390

    Validate with FDA (Live)

    Date Cleared
    2011-09-12

    (117 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SFI-Bar® is intended to be used with the implant manufacturer's implant (TABLE a) to provide support for fixation of overentures.

    Device Description

    The SFI-Bar® provides the connection between compatible dental implant systems for the fixation of removable overdentures. The SFI-Bar® consists of an implant adapter (abutment) and a stress-free bar for the fixation of removable overdentures. The implant adapter is screwed into the dental implant.

    AI/ML Overview

    Here's an analysis of the provided text to extract information about acceptance criteria and the study proving compliance for the SFI-Bar® device:

    The provided document describes the 510(k) premarket notification for the SFI-Bar® device, a dental implant abutment. It seeks to establish substantial equivalence to predicate devices, rather than proving performance against novel acceptance criteria in a comprehensive clinical study.

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission states that the SFI-Bar® is equivalent in performance characteristics to the predicate SFI-Bar® and that acceptance criteria were met. However, the document does not explicitly list specific numerical acceptance criteria for performance metrics. Instead, it relies on demonstrating equivalence to the predicate device.

    The "Performance Data" section mentions:

    • Validation of Sterilization for metallic components.
    • Evaluation of the intended manual disinfection procedure with SFI-Bar® Retention inserts G.
    • Application and functional testing based on literature research.

    The key acceptance criterion seems to be: "The SFI-Bar® is equivalent in performance characteristics to the predicate SFI-Bar®."

    Acceptance Criteria (Implied)Reported Device Performance
    Sterilization validated for metallic components.Validation of Sterilization has been conducted.
    Intended manual disinfection procedure evaluated.Evaluation of the intended manual disinfection procedure with SFI-Bar® Retention inserts G. has been conducted.
    Application and functional testing consistent with predicate.Application and functional testing based on literature research have been conducted. The test methods used were the same as in the predicate device.
    Overall performance equivalent to the predicate SFI-Bar®.Testing has shown that the SFI-Bar® is equivalent in performance characteristics to the predicate SFI-Bar®.
    Device is safe and effective for its intended use.The summary of technological characteristics indicates that the device is safe and effective for its intended use and performs as well or better than the predicate devices.
    Ability to provide support for fixation of overdentures with compatible implant systems.The SFI-Bar® is intended to be used with the implant manufacturer's (TABLE A) implant to provide support for fixation of overdentures. (Supported by technical compatibility and material equivalence.)

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "Non-clinical test data was used to support the substantially equivalence claim. Clinical testing was not necessary."

    This indicates that a "test set" in the sense of a patient cohort or a large sample for a clinical trial was not used for this submission. The testing performed was non-clinical.

    • Sample Size for Test Set: Not applicable in the context of clinical or large-scale performance testing. The "test samples" mentioned in the context of "analysis of platforms to identify worst-case test samples" would likely refer to physical samples or configurations chosen for non-clinical engineering tests (e.g., sterilization, disinfection). The specific number of these non-clinical worst-case samples is not provided.
    • Data Provenance: The submission is from Cendres+Metaux SA, Biel/Bienne, Switzerland. The non-clinical testing was conducted by the manufacturer, but the location where the tests were performed is not specified. The data is retrospective in the sense that it's based on existing designs and literature, and performed prior to this submission for regulatory clearance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. As clinical testing was not performed and the submission relies on demonstrating substantial equivalence through non-clinical data and comparison to predicate devices, there was no "ground truth" to be established by experts in a clinical context for a test set. The validation and evaluation mentioned primarily concern manufacturing processes and materials.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set requiring expert adjudication was used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This device is a dental implant abutment, a physical medical device, not an AI or imaging diagnostic tool. Therefore, MRMC studies and AI assistance are irrelevant to this submission.

    6. If a Standalone (i.e., algorithm only without human-in-the loop performance) Was Done

    Not applicable. This is a physical medical device, not an algorithm or software.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    Given the nature of the device and the submission, the "ground truth" relied upon is primarily:

    • Engineering and material standards (e.g., ASTM F 136 for Titanium alloy).
    • Performance characteristics of legally marketed predicate devices (K073628, K083876, K102382) as established through their prior regulatory clearances.
    • Scientific literature research supporting application and functional principles.
    • Validation of manufacturing processes (sterilization, cleaning).

    There isn't "ground truth" in the sense of clinical outcomes or expert consensus on diagnostic interpretations.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set, there is no ground truth establishment for it.

    In summary: The submission for the SFI-Bar® focuses on demonstrating substantial equivalence to existing predicate devices, leveraging non-clinical testing, material standards, and literature review. It does not involve new clinical trials, expert consensus, or AI performance metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K102382

    Validate with FDA (Live)

    Date Cleared
    2010-12-03

    (102 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SFI-Bar® is intended to be used with the implant manufacturer's . (Table 1) implant to provide support for fixation of overdentures.

    Device Description

    The SFI-Bar® provides the connection between compatible dental implant systems for the fixation of removable overdentures. The SFI-Bar® consists of an implant adapter (abutment) and a stress-free bar for the fixation of removable overdentures. The implant adapter is screwed into the dental implant.

    AI/ML Overview

    The provided text describes a 510(k) submission for the SFI-Bar® Abutment, a dental device. The document primarily focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing. Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly list specific numerical acceptance criteria for each test. Instead, it states that "The acceptance criteria were met" and that the "SFI-Bar® is equivalent in performance characteristics to the predicate SFI-Bar®." The performance characteristics evaluated are generally described.

    Acceptance Criteria (Implied)Reported Device Performance
    Equivalent performance characteristics to the predicate SFI-Bar®"Testing has shown that the SFI-Bar® is equivalent in performance characteristics to the predicate SFI-Bar®." This equivalence is established through:
    (Specific torque values for mechanical stability not explicitly stated)Torque tests: Conducted to evaluate performance characteristics. The methods used were the same as in the predicate device. The device was found to be equivalent in performance.
    (Specific functional criteria not explicitly stated)Application testing: Conducted to evaluate performance characteristics. The methods used were the same as in the predicate device. The device was found to be equivalent in performance.
    (Specific functional criteria not explicitly stated)Functional testing: Conducted to evaluate performance characteristics. The methods used were the same as in the predicate device. The device was found to be equivalent in performance. Key functional characteristics mentioned (also shared with the predicate) include: primary splinting of implants, countering forces that would dislodge the denture, distribution of shear forces, resilience compensation, common cleaning procedures for patients, and common patient handling (cleaning and insertion of denture).
    (Shelf life performance equivalent to predicate)Shelf life: "95% after 10 years" (same as predicate device).
    (Manufacturing processes and materials equivalent to predicate)Packaging, materials and processes: "Produced on process orientated CNC machines. The last step is a validated cleaning process (same processes). Packaging: Dentalblister, non-sterile." The material is "Wrought Titanium-6Aluminium-4Vanadium ELI Alloy for Surgical Implant applications" conforming to ASTM F 136 (same as predicate).
    (Compatibility with specified implant systems)Compatibility: The SFI-Bar® Abutment is designed to connect with various dental implant systems from Thommen Medical, Neoss, and Institut Straumann (specifically listed in Table 1). The development and manufacturing were done "in close cooperation with the implant companies" with Quality Agreements in place (Design Control, Change Control, Complaint Handling, and Post Market Surveillance). This implies successful compatibility testing was conducted to ensure proper fit and function with these diverse implant systems, similar to the predicate device's compatibility.
    (Safety for intended use)"The summary of technological characteristics as well as the torque test, application and functional testing indicate that the device is safe and effective for its intended use and performs as well or better than the predicate devices." This is a general conclusion based on the non-clinical testing and comparison to the predicate, demonstrating that the device does not raise new questions of safety or effectiveness. This is the overarching "acceptance criterion" for a 510(k) submission seeking substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the exact sample sizes used for the torque tests, application testing, or functional testing. It only states that these tests were conducted.
    • Data Provenance: The data is non-clinical test data. The country of origin of the data is not explicitly stated, but the applicant (Cendres+Métaux SA) is located in Switzerland. The testing was conducted to support a 510(k) submission to the FDA (USA). The study is retrospective in the sense that it relies on existing testing methodologies and comparisons to a previously cleared predicate device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable (N/A) as the study described is a non-clinical, bench testing study comparing the performance of a new device to a predicate device. There is no "ground truth" derived from expert review of cases or outcomes in a clinical context. The "ground truth" for the testing would be established by validated test methods and engineering specifications.

    4. Adjudication Method for the Test Set

    This information is not applicable (N/A) for the same reasons as point 3. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving interpretation of medical images or patient outcomes, not in bench testing of a dental implant abutment.

    5. If a Multi-Reader, Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This information is not applicable (N/A). The device described, the SFI-Bar® Abutment, is a physical dental implant component, not an AI or imaging diagnostic device. Therefore, an MRMC study and AI assistance effect size are irrelevant to this submission.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable (N/A). As mentioned in point 5, this is a physical dental device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this type of non-clinical device testing is engineering specifications, validated test methods, and the performance characteristics of the legally marketed predicate device. The goal is to demonstrate that the candidate device performs "as well or better than" the predicate device under the same test conditions. There isn't an "expert consensus," "pathology," or "outcomes data" ground truth in the clinical sense for this specific evaluation. The "worst-case test samples" were identified through analysis of platforms, indicating an engineering-based approach to defining test parameters.

    8. The Sample Size for the Training Set

    This information is not applicable (N/A). There is no concept of a "training set" in the context of this non-clinical bench testing for a physical dental device. Training sets are relevant for machine learning algorithms or similar data-driven models.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable (N/A), as there is no training set for this type of device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K083876

    Validate with FDA (Live)

    Date Cleared
    2010-02-26

    (424 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SFI-Bar® is intended to be used with the SPI Element Platform 4.0 mm implant (K070007) to provide support for fixation of overdentures.

    The SFI Bar consists of the following variants:

    SFI-Bar® 2 Implant: Lower jaw: Connecting 2 or 2x2 implants in the intraforaminal region Upper jaw: Connecting 2x2 implants in the anterior/premolar region

    SFI-Bar® 4 Implant: Lower jaw: Connecting 4 implants in the intraforaminal region Upper jaw: Connecting 4 implants in the anterior/premolar region

    Device Description

    Not Found

    AI/ML Overview

    I'm sorry, but the provided text does not contain the information required to answer your request. The document is an FDA 510(k) clearance letter for a dental implant abutment (SFI-Bar®), which primarily focuses on its substantial equivalence to a predicate device. It does not include details about acceptance criteria, study methodologies, sample sizes, ground truth establishment, or expert qualifications for performance evaluation.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1