K Number
K130618

Validate with FDA (Live)

Device Name
SFI-ANCHOR
Date Cleared
2014-03-03

(361 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SFI-Anchor® D20 is intended for the fixation of (dental) prostheses to corresponding dental implants. The compatible implant systems are specified in Table 1 below. The SFI-Anchor® CD20 is intended as an additional retaining element on CAD / CAM milled dental bars.

Device Description

The SFI-Anchor® D20 is an endosseous dental implant abutment for the fixation of (dental) prostheses to the following ITI Dental Implant System® / Straumann® Dental Implant Systems: Standard Ø 4.1 mm Regular Neck (RN) / Standard Ø 4.8 mm Regular Neck (RN) / Standard Plus Ø 4.1 mm Regular Neck (RN) / Standard Plus Ø 4.8 mm Regular Neck (RN) / Tapered Effect Ø 4.1 mm Regular Neck (RN) / Bone Level Ø 4.1 mm, Regular CrossFit® (RC) / Bone Level Ø 4.8 mm, Regular CrossFit® (RC). The SFI-Anchor® CD 20 is intended as an additional retaining element on dental bars. The SFI-Anchor® D20 are straight abutments with gingival heights ranging from 1mm to 5 mm. The SFI-Anchor® CD 20 is a straight bar attachment. The SFI-Anchor® Housing and the SFI-Anchor® Retention Inserts can correct for up to 10° of divergence (20° of divergence between two implants). The SFI-Anchor® is offered in a diameter 4.8 mm. The SFI-Anchor® D20 is connected to the corresponding implants via a mating screw thread and interface geometry. The SFI-Anchor® CD20 is directly screwed on a dental bar construct. The SFI-Anchor® Housing is cemented in the dental prostheses. The SFI-Anchor® Housing, along with an SFI-Anchor® Retention Insert, allow connection of the prosthesis to the abutment / bar attachment by retention-grip. The abutments / bar attachment are machined from Titanium Alloy conforming to ASTM F-136. Other SFI-Anchor® key functional parts are made out of Elitor® (conforming to standard ISO 22674) and of Pekkton® conforming to standard ASTM D3418, ASTM D638 and ASTM D790.

AI/ML Overview

The provided text describes a dental implant abutment, SFI-Anchor® D20 and SFI-Anchor® CD20, and its substantial equivalence to predicate devices. It does not contain information about acceptance criteria and a study proving the device meets those criteria in the context of AI/ML performance.

Therefore, I cannot provide the requested information. The document focuses on the mechanical and material characteristics of the dental abutments, their compatibility with implant systems, and a comparison with existing predicate devices to establish substantial equivalence for regulatory approval. It is not an AI/ML device.

{0}------------------------------------------------

510(k) Summary: SFI-Anchor® D20, SFI-Anchor® CD20 510(k) no K130618

MAR February 24.0264

510(k) Summary

Applicant's Name and Address
Submitter:Cendres+Métaux SARue de Boujean 1222501 Biel/Bienne, SwitzerlandPhone: +41 58 360 20 00Fax: +41 58 360 20 10
Contact Person:Tanja BongniRegulatory Affairs Manager
Date of Submission:February 24, 2014
Name of the Device
Trade Names:SFI-Anchor® D20, SFI-Anchor® CD20
Common Name:Abutment, implant, Dental, Endosseous
Classification Name:Endosseous dental implant abutment
Regulation Number:21 CFR 872.3630
Product Code:NHA
Device Class:II

Legally Marketed Device to which Equivalence is Claimed (Predicate Device)

Trade Name510(k) no
SFI-Bar®K111390, K102382, K083876
Locator Implant AnchorK994257
P.004 RC/NC Bar and Bridge Abutments LineK080239

Figure 1 Predicate Devices

Description of the Device

Device Description:

The SFI-Anchor® D20 is an endosseous dental implant abutment for the fixation of (dental) prostheses to the following ITI Dental Implant System® / Straumann® Dental Implant Systems:

Standard Ø 4.1 mm Regular Neck (RN) / Standard Ø 4.8 mm Regular Neck (RN) / Standard Plus Ø 4.1 mm Regular Neck (RN) / Standard Plus Ø 4.8 mm Regular Neck (RN) / Tapered Effect Ø 4.1 mm Regular Neck (RN) / Bone Level Ø 4.1 mm, Regular CrossFit® (RC) / Bone Level Ø 4.8 mm, Regular CrossFit® (RC).

The SFI-Anchor® CD 20 is intended as an additional retaining element on dental bars. The SFI-Anchor® D20 are straight abutments with gingival heights ranging from 1mm to 5 mm. The SFI-Anchor® CD 20 is a straight bar attachment. The SFI-Anchor® Housing and the SFI-Anchor® Retention Inserts can correct for up to 10° of divergence (20° of divergence between two implants).

The SFI-Anchor® is offered in a diameter 4.8 mm. The SFI-Anchor® D20 is connected to the corresponding implants via a mating screw thread and interface geometry. The SFI-Anchor® CD20 is directly screwed on a dental bar construct. The SFI-Anchor® Housing is cemented in the dental prostheses. The SFI-Anchor® Housing, along with an SFI-Anchor® Retention Insert, allow connection of the prosthesis to the abutment / bar attachment by retention-grip.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a logo with the text "CENDRES+" on the top line and "MÉTAUX" on the bottom line. To the left of the text is a stylized letter "C" that is thicker than the text. The text is in a sans-serif font and is black. The logo is simple and modern.

510(k) Summary: SFI-Anchor® D20, SFI-Anchor® CD20 510(k) no K130618

The abutments / bar attachment are machined from Titanium Alloy conforming to ASTM F-136. Other SFI-Anchor® key functional parts are made out of Elitor® (conforming to standard ISO 22674) and of Pekkton® conforming to standard ASTM D3418, ASTM D638 and ASTM D790.

Indications for Use:

The SFI-Anchor® D20 is intended for the fixation of (dental) prostheses to corresponding dental implants.

The compatible implant systems are specified in Table 1 below.

Implant Manufacturer: Institut Straumann
Implant System: ITI Dental Implant System® respectively Straumann® Dental Implant System
Implant line name(s): see below
-Standard Ø 4.1 mm Regular Neck (RN)-Standard Ø 4.8 mm Regular Neck (RN)
-Standard Plus Ø 4.1 mm Regular Neck (RN)-Standard Plus Ø 4.8 mm Regular Neck (RN)
-Tapered Effect Ø 4.1 mm Regular Neck (RN)-Bone Level Ø 4.1 mm, Regular CrossFit® (RC)
-Bone Level Ø 4.8 mm, Regular CrossFit® (RC)

Table 1 Compatible Implant Systems

The SFI-Anchor® CD20 is intended as an additional retaining element on CAD / CAM milled dental bars.

Summary Technological Characteristics

The following Figure 2 describes similarities and differences between the subject devices and legally cleared predicates:

Candidates:Predicate DevicesComparison
SFI-Anchor®D20SFI-Anchor®CD20SFI-Bar®LocatorImplantAnchorP.004RC/NC Barand BridgeAbutmentsLine
Platformcompatibility(implant)Refer toTable 1N/ARegular Neck(RN)(and others)---RegularCrossFit®(RC) (andothers)Sameimplantplatformsthan SFI-Bar® andP.004.
Basic Designof theAbutment /ElementStraight,gingivalheightsranging from1 mm to 5mm.StraightStraight,gingivalheightsranging from1 to 6.Straight,gingivalheightsranging from1 to 5.Straight (andothers)ComparabletoPredicates
RetentiveforceUp to 17.5 NUp to 17.5 NUp to 14 NUp to 5.0pounds(22.25 N)Comparableto SFI-Bar® andLocator.
DivergenceCompensationby Housingand RetentioninsertUp to 10° perimplant(20° between2 implants)Up to 10°per implant(20°between 2implants)

Figure 2 Comparison with Predicate Devices

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image contains a logo with the text "CENDRES+ METAUX" to the right of a stylized letter "C". The text is stacked, with "CENDRES+" on top and "METAUX" below. The letter "C" is bold and stylized, with a curved shape. The overall design is simple and modern.

510(k) Summary: SFI-Anchor® D20, SFI-Anchor® CD20 510(k) no K130618

The following materials with permanent patient contact are incorporated into key functional elements: Titanium; Elitor®; Pekkton® (blue, red, yellow, green). Similar material was used in the legally marketed predicate devices.

The proposed devices are similar in terms of design, angulations for use and incorporate the same technological characteristics as the predicate devices.

In order to assure safety of the SFI-Anchor® a failure mode, effect and criticality analysis has been performed.

Non-Clinical Performance Data

The specifications for a reliable connection between the dental implant and the SFI-Anchor® are developed in close cooperation with the dental implant manufacturer. To ensure compatibility today and in future there is a Quality Agreement between Cendres + Métaux SA and the Institut Straumann AG in place. This Quality Agreement handles among other things the Design Control process and Vigilance. Non-clinical test data was used to support the substantially equivalence claim. Clinical testing was not necessary. Non-clinical testing consisted of analysis of platforms to identify worst-case test samples. Retention force testing and simulated use testing were done. Furthermore, the evaluation was based on FDA guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments".

Conclusion as to Substantial Equivalence

Based on the comparison of the indications for use, the technological characteristics and the non-clinical testing it can be concluded that the SFI-Anchor® is substantially equivalent to the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's services. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged around the eagle in a circular fashion.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W066-G609 Silver Spring, MD 20993-0002

March 3, 2014

Cendres+Metaux SA Ms. Tanja Bongni Regulatory Affairs Manager Rue de Boujean 122 2501 Biel/Bienne SWITZERLAND

Re: K130618

Trade/Device Name: SFI-Anchor® D20, SFI-Anchor® CD20 S Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: January 29, 2014 Received: January 30, 2014

Dear Ms. Bongni:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.goy/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Image /page/4/Picture/7 description: The image shows the text "Erin" followed by a complex, maze-like design that incorporates the letters "F", "D", "K", "A", and "h". The text "-S" is located to the right of the maze-like design. The overall image appears to be a logo or a stylized wordmark.

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):

K130618

Device Names:

SFI-Anchor® D20, SFI-Anchor® CD 20

Indications for Use:

The SFI-Anchor® D20 is intended for the fixation of (dental) prostheses to corresponding dental implants.

The compatible implant systems are specified in Table 1 below.

Implant Manufacturer: Institut Straumann
Implant System: ITI Dental Implant System® respectively Straumann® Dental Implant System
Implant line name(s): see below
-Standard Ø 4.1 mm Regular Neck (RN)-Standard Plus Ø 4.1 mm Regular Neck (RN)-Tapered Effect Ø 4.1 mm Regular Neck (RN)-Bone Level Ø 4.8 mm, Regular CrossFit® (RC)-Standard Ø 4.8 mm Regular Neck (RN)-Standard Plus Ø 4.8 mm Regular Neck (RN)-Bone Level Ø 4.1 mm, Regular CrossFit® (RC)

Table 1 Compatible Implant Systems

The SFI-Anchor® CD20 is intended as an additional retaining element on CAD / CAM milled dental bars.

Prescription Use X (Part 21 CFR 801 Subpart D) AND / OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mary S. Runner -S
SR DOSNA 2014.02.28
11:09:55:05'00'

Page 1 of 1

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)