Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K140015

    Validate with FDA (Live)

    Date Cleared
    2014-02-05

    (34 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dexterity™ Steerable Introducer is intended to be used for the percutaneous introduction of various cardiovascular catheters into the heart, including the left side of the heart through the inter-atrial septum.

    Device Description

    The Cardiosolutions Dexterity™ Steerable Introducer is provided as a 9Fr and 14Fr Introducer. The set also consists of the following components based on the model#:
    Size and Working Length: 14FR 65 and 85 CM, Included Accessories: Dilator, Stylet, Tear-away introducer sheath
    Size and Working Length: 14Fr 105 CM, Included Accessories: Dilator, NA, Tear-away introducer sheath
    Size and Working Length: 9Fr 65CM, Included Accessories: NA, Curved Stylet, NA
    The Dexterity steerable introducer is designed to provide flexible catheter positioning in the cardiac anatomy. The steerable introducer provides both proximal tip and distal tip steering and is fitted with a hemostasis valve to minimize blood loss during catheter introduction and/or exchange. A sideport with stopcock is provided for air or blood aspiration, fluid infusion, blood sampling, and pressure monitoring. The introducer sheath is reinforced Pebax and the distal tip has a radiopaque marker to improve fluoroscopic visualization. The device is provided in various diameters and working lengths. The device is provided sterile and is intended for single use only.
    The purpose of this submission is to add an additional device working length, 105 CM, to the 14Fr Introducer product offering. In addition the intended use statement is modified to include the term percutaneous for clarification only. This change is not a change to the intended use of the device.

    AI/ML Overview

    This document describes the Cardiosolutions Dexterity™ Steerable Introducer, a medical device designed for the introduction of cardiovascular catheters into the heart. The information provided outlines the acceptance criteria and the summary of testing performed to demonstrate substantial equivalence to predicate devices, rather than a standalone clinical study on human subjects with ground truth established by experts.

    Here's an analysis of the provided text based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document lists performance tests conducted, implying these are the acceptance criteria, and states that "All test results demonstrate that the properties and performance of the device are substantially equivalent to the predicate device and suitable for its intended use." Specific quantitative performance values are not provided in this summary.

    Acceptance Criterion (Test Name)Reported Device Performance
    Visual Surface InspectionMet (implied by "All test results demonstrate...")
    Dimensional VerificationMet (implied by "All test results demonstrate...")
    Corrosion ResistanceMet (implied by "All test results demonstrate...")
    Tensile Break ForceMet (implied by "All test results demonstrate...")
    Tip Separation ForceMet (implied by "All test results demonstrate...")
    Freedom from Leakage Under PressureMet (implied by "All test results demonstrate...")
    Air Leakage During AspirationMet (implied by "All test results demonstrate...")
    Luer Hub ComplianceMet (implied by "All test results demonstrate...")
    Flexural Fatigue ToleranceMet (implied by "All test results demonstrate...")
    In Vitro Simulated Use StudiesMet (implied by "All test results demonstrate...")
    Functional Performance TestingMet (implied by "All test results demonstrate...")
    RadiopacityMet (implied by "All test results demonstrate...")
    Shelf LifeMet (implied by "All test results demonstrate...")

    2. Sample Size Used for the Test Set and Data Provenance:

    The document describes design verification testing on the modified device, specifically a 14Fr 105 CM Dexterity™ Steerable Introducer. It does not specify the sample size for each test. The testing is described as occurring in a laboratory setting (in vitro simulated use), and therefore the data provenance is laboratory testing, not human subject data from a specific country or retrospective/prospective study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    Not applicable. This device is a steerable introducer catheter, and the evaluation is based on engineering and performance specifications, not on expert interpretation of diagnostic images or clinical outcomes. The "ground truth" for these tests would be the established engineering specifications and standards (e.g., ISO 10555).

    4. Adjudication Method for the Test Set:

    Not applicable. The tests are objective engineering and performance tests, not subjective assessments requiring adjudication by multiple experts.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done:

    No, an MRMC comparative effectiveness study was not done. The document focuses on demonstrating substantial equivalence to existing predicate devices through engineering and performance testing, not on assessing the improvement of human readers with or without AI assistance. The device itself is an introducer, not an AI-powered diagnostic tool.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a physical medical instrument (catheter introducer), not an algorithm or AI system.

    7. The Type of Ground Truth Used:

    For the design verification tests, the ground truth is based on engineering specifications, international standards (ISO 10555), and FDA guidance documents for intravascular catheters.

    8. The Sample Size for the Training Set:

    Not applicable. This is not an AI/ML device that requires a training set. The device is evaluated based on its physical and mechanical properties.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131332

    Validate with FDA (Live)

    Date Cleared
    2013-07-19

    (71 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Percu-Pro™ Steerable Introducer is intended to be used for the introduction of various cardiovascular catheters into the heart, including the left side of the heart through the inter-atrial septum.

    Device Description

    The Cardiosolutions Percu-Pro™ Steerable Introducer is provided as a 9Fr and 14Fr Introducer. The set also consists of a dilator (14Fr only), stylet, and tear-away introducer sheath. The steerable introducer is designed to provide flexible catheter positioning in the cardiac anatomy. The steerable introducer provides both proximal tip and distal tip steering and is fitted with a hemostasis valve to minimize blood loss during catheter introduction and/or exchange. A sideport with stopcock is provided for air or blood aspiration, fluid infusion, blood sampling, and pressure monitoring. The introducer sheath is reinforced Pebax and the distal tip has a radiopaque marker to improve fluoroscopic visualization. The device is provided in 65 cm working lengths. The device is provided sterile and is intended for single use only. The purpose of this submission is to add an additional device diameter; 9Fr has 65 cm working lengths.

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device, not a study evaluating an AI algorithm. Therefore, many of the requested categories are not applicable.

    Here's an analysis based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test CategoryAcceptance Criteria (Implied)Reported Device Performance
    Mechanical TestingConformity to ISO 10555 "Sterile, single-use intravascular catheters Part 1: General requirements" (as amended, 1999, 2004) and FDA Guidance on Premarket Notification 510(k) Submission for Short Term and Long Term Intravascular Catheters. This implies that the device should meet specified standards for: - Visual Surface Inspection - Dimensional Verification - Corrosion Resistance - Tensile Break Force - Tip Separation Force - Freedom from Leakage Under Pressure - Air Leakage During Aspiration - Luer Hub Compliance - Flexural Fatigue Tolerance - In Vitro Simulated Use Studies - Functional Performance Testing - Radiopacity"All test results demonstrate that the properties and performance of the device are suitable for its intended use." (Specific quantitative results are not provided in this summary.)
    Material/BiocompatibilityNew device materials should be biocompatible or identical to predicate device materials."The materials used in the 9Fr Percu-Pro™ Steerable Introducer are identical to the 14Fr Percu-Pro™ Steerable Introducer predicate device. No additional biocompatibility is required."
    Shelf LifeThe device should maintain its properties and performance over its intended shelf life.Shelf life testing was completed (specific results not provided).
    Substantial EquivalenceThe device must be substantially equivalent to the predicate device in terms of intended use, design, materials, technology, and function, with no new issues of safety or effectiveness."The 9Fr Percu-Pro™ Steerable Introducer is substantially equivalent to the predicate devices in terms of intended use, design, materials, technology, and function. There are no differences between devices which would raise new issues of safety or effectiveness."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not provide specific sample sizes for each mechanical and functional test. It generally states that "Design verification testing performed on the 9Fr device consisted of mechanical testing conducted in accordance with..."

    • Sample Size: Not explicitly stated as a number of devices/units. The testing generally applies to the 9Fr device.
    • Data Provenance: Not applicable in the context of clinical data for AI. These are engineering/device performance tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This is a device performance evaluation, not an AI study requiring expert ground truth for classification.

    4. Adjudication Method for the Test Set

    Not applicable.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, this is not an MRMC study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a medical instrument, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" here is adherence to established engineering standards (ISO 10555) and FDA guidance for the performance of intravascular catheters. For material equivalence, the ground truth is the chemical composition of the predicate device.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This is not an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K094048

    Validate with FDA (Live)

    Date Cleared
    2011-07-14

    (560 days)

    Product Code
    Regulation Number
    870.1240
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Percu-Pro™ Messenger Balloon Catheter is intended to facilitate the placement of interventional devices during diagnostic and interventional cardiovascular procedures. The device is not intended for coronary or neurovascular use.

    Device Description

    The Cardiosolutions Percu-Pro™ Messenger Balloon Catheter is a 5Fr catheter with a bifurcated proximal hub. The catheter shaft and connection ends are made from Pebax and the distal tip balloon is made from latex.

    AI/ML Overview

    This document does not contain information about a study proving that a device meets acceptance criteria, an AI/ML device, or a standalone algorithm.

    The document is a 510(k) summary for the Percu-Pro™ Messenger Balloon Catheter, which is a physical medical device (a catheter). The summary details its design, intended use, and comparison to predicate devices to establish substantial equivalence, rather than providing performance metrics like accuracy, sensitivity, or specificity that would be typical for an AI system.

    Therefore, I cannot provide the requested information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K094062

    Validate with FDA (Live)

    Date Cleared
    2010-09-27

    (270 days)

    Product Code
    Regulation Number
    870.1330
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Percu-Pro™ Guidewire is intended to facilitate the placement of devices during diagnostic and interventional vascular procedures in the peripheral vasculature. The guidewire is not intended for use in coronary arteries or for neurovascular use.

    Device Description

    The Percu-Pro™ GuideWire is constructed of a stainless steel core wire The proximal end of the core and is finished with a PTFE jacket. The coil covered distal end of the core is tapered to a ribbon tip to increase flexibility and is finished with a silver brazed ball tip. The guidewire is provided sterile and is a single use device.

    AI/ML Overview

    The provided 510(k) summary for the Percu-Pro™ GuideWire (K094062) describes pre-clinical safety and performance testing but does not include acceptance criteria or a study with specific performance metrics for the device itself against established criteria. Instead, it states that "No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for these devices."

    The submission focuses on demonstrating substantial equivalence to predicate devices (TechDevice Guidewire K053251 and TFX Guidewire K963320) based on "intended use, design, materials, technology, and performance," asserting that "There are no differences between devices which would raise new issues of safety or effectiveness."

    Therefore, I cannot fulfill the request to provide a table of acceptance criteria and reported device performance, or details about a study proving the device meets those criteria, as such information is not present in the provided document.

    However, I can extract information about the pre-clinical testing conducted, which supports the safety and performance claims for substantial equivalence:

    Pre-Clinical Safety and Performance Testing Information (from the 510(k) Summary):

    • Type of Testing: Biocompatibility and hemocompatibility, tensile strength, torque strength, resistance to fracture, torqueability, and tip flexibility.
    • Methodology: Conducted "in accordance with protocols based on the requirements of industry standards and guidance documents."

    Missing Information (Not available in the provided document):

    1. Table of Acceptance Criteria and Reported Device Performance: Not specified.
    2. Sample Size for Test Set and Data Provenance: Not specified for any performance testing.
    3. Number of Experts and Qualifications for Ground Truth: Not applicable, as performance testing against specific criteria is not detailed.
    4. Adjudication Method: Not applicable.
    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: Not applicable (this is a medical device, not an AI diagnostic tool).
    6. Standalone Performance Study (Algorithm only): Not applicable (this is a medical device, not an AI algorithm).
    7. Type of Ground Truth Used: Not applicable, as performance testing against specific criteria is not detailed.
    8. Sample Size for Training Set: Not applicable (this is a medical device, not an AI algorithm).
    9. How Ground Truth for Training Set was Established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1