K Number
K140015

Validate with FDA (Live)

Date Cleared
2014-02-05

(34 days)

Product Code
Regulation Number
870.1340
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Dexterity™ Steerable Introducer is intended to be used for the percutaneous introduction of various cardiovascular catheters into the heart, including the left side of the heart through the inter-atrial septum.

Device Description

The Cardiosolutions Dexterity™ Steerable Introducer is provided as a 9Fr and 14Fr Introducer. The set also consists of the following components based on the model#:
Size and Working Length: 14FR 65 and 85 CM, Included Accessories: Dilator, Stylet, Tear-away introducer sheath
Size and Working Length: 14Fr 105 CM, Included Accessories: Dilator, NA, Tear-away introducer sheath
Size and Working Length: 9Fr 65CM, Included Accessories: NA, Curved Stylet, NA
The Dexterity steerable introducer is designed to provide flexible catheter positioning in the cardiac anatomy. The steerable introducer provides both proximal tip and distal tip steering and is fitted with a hemostasis valve to minimize blood loss during catheter introduction and/or exchange. A sideport with stopcock is provided for air or blood aspiration, fluid infusion, blood sampling, and pressure monitoring. The introducer sheath is reinforced Pebax and the distal tip has a radiopaque marker to improve fluoroscopic visualization. The device is provided in various diameters and working lengths. The device is provided sterile and is intended for single use only.
The purpose of this submission is to add an additional device working length, 105 CM, to the 14Fr Introducer product offering. In addition the intended use statement is modified to include the term percutaneous for clarification only. This change is not a change to the intended use of the device.

AI/ML Overview

This document describes the Cardiosolutions Dexterity™ Steerable Introducer, a medical device designed for the introduction of cardiovascular catheters into the heart. The information provided outlines the acceptance criteria and the summary of testing performed to demonstrate substantial equivalence to predicate devices, rather than a standalone clinical study on human subjects with ground truth established by experts.

Here's an analysis of the provided text based on your request:

1. Table of Acceptance Criteria and Reported Device Performance:

The document lists performance tests conducted, implying these are the acceptance criteria, and states that "All test results demonstrate that the properties and performance of the device are substantially equivalent to the predicate device and suitable for its intended use." Specific quantitative performance values are not provided in this summary.

Acceptance Criterion (Test Name)Reported Device Performance
Visual Surface InspectionMet (implied by "All test results demonstrate...")
Dimensional VerificationMet (implied by "All test results demonstrate...")
Corrosion ResistanceMet (implied by "All test results demonstrate...")
Tensile Break ForceMet (implied by "All test results demonstrate...")
Tip Separation ForceMet (implied by "All test results demonstrate...")
Freedom from Leakage Under PressureMet (implied by "All test results demonstrate...")
Air Leakage During AspirationMet (implied by "All test results demonstrate...")
Luer Hub ComplianceMet (implied by "All test results demonstrate...")
Flexural Fatigue ToleranceMet (implied by "All test results demonstrate...")
In Vitro Simulated Use StudiesMet (implied by "All test results demonstrate...")
Functional Performance TestingMet (implied by "All test results demonstrate...")
RadiopacityMet (implied by "All test results demonstrate...")
Shelf LifeMet (implied by "All test results demonstrate...")

2. Sample Size Used for the Test Set and Data Provenance:

The document describes design verification testing on the modified device, specifically a 14Fr 105 CM Dexterity™ Steerable Introducer. It does not specify the sample size for each test. The testing is described as occurring in a laboratory setting (in vitro simulated use), and therefore the data provenance is laboratory testing, not human subject data from a specific country or retrospective/prospective study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

Not applicable. This device is a steerable introducer catheter, and the evaluation is based on engineering and performance specifications, not on expert interpretation of diagnostic images or clinical outcomes. The "ground truth" for these tests would be the established engineering specifications and standards (e.g., ISO 10555).

4. Adjudication Method for the Test Set:

Not applicable. The tests are objective engineering and performance tests, not subjective assessments requiring adjudication by multiple experts.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done:

No, an MRMC comparative effectiveness study was not done. The document focuses on demonstrating substantial equivalence to existing predicate devices through engineering and performance testing, not on assessing the improvement of human readers with or without AI assistance. The device itself is an introducer, not an AI-powered diagnostic tool.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This device is a physical medical instrument (catheter introducer), not an algorithm or AI system.

7. The Type of Ground Truth Used:

For the design verification tests, the ground truth is based on engineering specifications, international standards (ISO 10555), and FDA guidance documents for intravascular catheters.

8. The Sample Size for the Training Set:

Not applicable. This is not an AI/ML device that requires a training set. The device is evaluated based on its physical and mechanical properties.

9. How the Ground Truth for the Training Set was Established:

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

.

ﻬﺎ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟ

007-1

K140015
FEB - 5 2014

SECTION 7.0

·

SPECIAL 510(K) SUMMARY

:

Premarket Notification for the Cardiosolutions Dexterity™ Steerable Introducer

{1}------------------------------------------------

510(k) Summary

Special 510(k)

As required by section 807.92(c)

CompanyNameCardiosolutions Inc.
Address375 West St.West Bridgewater MA 02379Phone: 781-344-0801Fax: 781-344-0803
Contact PersonChristine Santagate
Date PreparedDecember 17, 2013
Trade NameDexterity™ Steerable Introducer
Common NameSteerable Introducer
ClassificationNameCatheter Introducer
Product CodeDYB, DRA
Regulation #21 CFR 870.1340
Class2
PanelCardiovascular
PredicateDevicesPercu-Pro™ Steerable Introducer K120086, 131332
DeviceDescriptionThe Cardiosolutions Dexterity™ Steerable Introducer is provided as a 9Fr and 14FrIntroducer. The set also consists of the following components based on the model#:
Size and WorkingLengthIncluded Accessories
14FR 65 and 85CMDilatorStyletTear-awayintroducer sheath
14Fr 105 CMDilatorNATear-awayintroducer sheath
9Fr 65CMNACurved StyletNA
The Dexterity steerable introducer is designed to provide flexible catheter positioning inthe cardiac anatomy. The steerable introducer provides both proximal tip and distal tipsteering and is fitted with a hemostasis valve to minimize blood loss during catheterintroduction and/or exchange. A sideport with stopcock is provided for air or bloodaspiration, fluid infusion, blood sampling, and pressure monitoring. The introducersheath is reinforced Pebax and the distal tip has a radiopaque marker to improvefluoroscopic visualization. The device is provided in various diameters and workinglengths. The device is provided sterile and is intended for single use only.
The purpose of this submission is to add an additional device working length, 105 CM,to the 14Fr Introducer product offering. In addition the intended use statement ismodified to include the term percutaneous for clarification only. This change is not achange to the intended use of the device.
Intended UseThe Dexterity™ Steerable Introducer is intended to be used for the percutaneousintroduction of various cardiovascular catheters into the heart, including the left side ofthe heart through the inter-atrial septum.
Safety andPerformanceTestingNo performance standards have been established under Section 514 of the Food, Drugand Cosmetic Act for these devices.
The materials used in the 14 Fr 105 CM Dexterity™ Steerable Introducer are identical tothe Dexterity™ Steerable Introducer predicate device. No additional biocompatibility isrequired.
Design verification testing performed on the modified device consisted of mechanicaltesting conducted in accordance with the ISO 10555 Sterile, single-use intravascularcatheters Part 1: General requirements (as amended, 1999, 2004) and in consideration ofFDA Guidance on Premarket Notification 510(k) Submission for Short Term and LongTerm Intravascular Catheters. The following tests were completed:
oVisual Surface Inspection
oDimensional Verification
oCorrosion Resistance
oTensile Break Force
oTip Separation Force
oFreedom from Leakage Under Pressure
oAir Leakage During Aspiration
oLuer Hub Compliance
oFlexural Fatigue Tolerance
oIn Vitro Simulated Use Studies
oFunctional Performance Testing
oRadiopacity

Premarket Notification for the Cardiosolutions Dexterity™ Steerable Introducer

{2}------------------------------------------------

ﺮ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤ

.

Premarket Notification for the Cardiosolutions Dexterity™ Steerable Introducer


{3}------------------------------------------------

. .

oShelf Life
All test results demonstrate that the properties and performance of the device aresubstantially equivalent to the predicate device and suitable for its intended use.
SubstantialEquivalenceThe 14Fr 105 CM Dexterity™ Steerable Introducer is substantially equivalent to thepredicate devices in terms of intended use, design, materials, technology, and function.There are no differences between devices which would raise new issues of safety oreffectiveness.

Premarket Notification for the Cardiosolutions Dexterity™ Steerable Introducer

.

.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 5, 2014

Cardiosolutions Inc. Christine Santagate Manager of Regulatory Consulting Services 375 West St. West Bridgewater, MA 02379

Re: K140015

Trade/Device Name: Cardiosolutions Dexterity Steerable Introducer Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter, Introducer Regulatory Class: Class II Product Code: DYB Dated: January 7, 2014 Received: January 8, 2014

Dear Ms. Santagate,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{5}------------------------------------------------

Page 2 - Christine Santagate

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

M. Ar. Heele

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use Statement

K140015 510(k) Number:

Device Name: Dexterity™ Steerable Introducer

Indications for Use:

The Dexterity™ Steerable Introducer is intended to be used for the percutaneous introduction of various cardiovascular catheters into the heart, including the left side of the heart through the inter-atrial septum.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

M. L. Hillman

§ 870.1340 Catheter introducer.

(a)
Identification. A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.(b)
Classification. Class II (performance standards).