510(k) Data Aggregation
(257 days)
The Access BNP II test is intended for use with the Beckman Coulter Access Family of Immunoassay Systems for the In Vitro quantitative measurement of B-type natriuretic peptide (BNP) in plasma specimens using EDTA as the anticoagulant. The test is intended to be used for the following indications:
- as an aid in the diagnosis of heart failure (HF) in patients presenting to the emergency department (ED) with clinical suspicion of new onset, acutely decompensated, or exacerbated heart failure
- for the risk stratification of patients with acute coronary syndromes
- for the risk stratification of patients with heart failure
The Access BNP II assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of B-type natriuretic peptide (BNP) levels in human EDTA plasma using the Access Immunoassay Systems.
The Access BNP Calibrators are intended to calibrate the Access BNP and Access BNP II tests for the quantitative determination of BNP levels in human EDTA plasma using the Beckman Coulter Access Family of Immunoassay Systems.
The Access BNP II test is a two-site immunoenzymatic ("sandwich") assay. A sample is added to a reaction vessel with mouse monoclonal anti-human BNP antibody-alkaline phosphatase conjugate and paramagnetic particles coated with mouse Omniclonal anti-human BNP antibody. BNP in human plasma binds to the immobilized anti-BNP on the solid phase, while the mouse anti-BNP conjugate reacts specifically with bound BNP.
After incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is inversely proportional to the concentration of analyte in the sample. Analyte concentration is automatically determined from a stored calibration.
N/A
Ask a specific question about this device
(86 days)
The Access anti-HBc IgM assay is a paramagnetic particle, chemiluminescent immunoassay for the in vitro qualitative detection of IgM antibodies to hepatitis B virus core antigen (anti-HBc IgM) in human pediatric (3 through 21 years) and adult serum and serum separator tubes or plasma [lithium heparin, lithium heparin separator tubes, dipotassium (K2) EDTA, tripotassium (K3) EDTA, sodium citrate, acid citrate dextrose (ACD), and citrate phosphate dextrose (CPD)] using the DxI 9000 Access Immunoassay Analyzer.
The Access anti-HBc IgM assay results may be used as an aid in the laboratory diagnosis of acute or recent hepatitis B virus (HBV) infection in individuals with signs and symptoms of hepatitis, when used in conjunction with other serological and clinical information.
The Access anti-HBc IgM assay is for use on the DxI 9000 Access Immunoassay Analyzer only.
This assay is not intended for the screening of blood, plasma, and cell or tissue donors.
The Access anti-HBc IgM assay is a two-step enzyme immunoassay. Paramagnetic particles coated with anti-human IgM monoclonal antibody and prediluted sample are added to a reaction vessel. After incubation, material bound to the solid phase is held in a magnetic field while unbound materials are washed away. HBc antigen complexed to anti-HBc monoclonal antibody alkaline phosphatase conjugate is added and the conjugate binds to the IgM antibodies captured on the particles. A second separation and wash step remove unbound conjugate.
A chemiluminescent substrate is then added to the vessel and light generated by the reaction is measured with a luminometer. The light production is compared to the cutoff value defined during calibration of the instrument. The qualitative assessment is automatically determined from a stored calibration.
The Access anti-HBc IgM Calibrator is used to establish calibration (determine the cutoff value) for the Access anti-HBc IgM assay. By comparing the light intensity generated by a sample to the cutoff value, the presence or absence of IgM antibodies to hepatitis B virus core antigen (anti-HBc IgM) in the sample is determined.
N/A
Ask a specific question about this device
(90 days)
Not Found
Not Found
N/A
Ask a specific question about this device
(214 days)
The Access anti-HAV IgM assay is a paramagnetic particle, chemiluminescent immunoassay for the in vitro qualitative detection of IgM antibodies to hepatitis A virus (anti-HAV IgM) in human pediatric (2 through 21 years) and adult serum and serum separator tubes or plasma [lithium heparin, lithium heparin separator tubes, dipotassium (K2) EDTA, and tripotassium (K3) EDTA] using the DxI 9000 Access Immunoassay Analyzer. The Access anti-HAV IgM assay results may be used as an aid in the laboratory diagnosis of acute or recent hepatitis A virus (HAV) infection in individuals with signs and symptoms of hepatitis A virus, when used in conjunction with other serological and clinical information.
This assay is not intended for use for screening donors of blood or blood products or human cells, tissues, or cellular or tissue-based products (HCT/Ps).
The Access anti-HAV IgM assay requires Access anti-HAV IgM (reagent packs), Access anti-HAV IgM Calibrator (C1), and Access anti-HAV IgM QC (QC1-QC2). The Access anti-HAV IgM assay is a two-step sandwich immunoassay. Paramagnetic particles coated with anti-human IgM monoclonal antibody and prediluted sample are added to a reaction vessel. After incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. HAV antigen and anti-HAV monoclonal antibody alkaline phosphatase conjugate are added. HAV antigen complexed to the conjugate binds to the IgM antibodies captured on the particles. A second separation and wash step removes unbound conjugate.
A chemiluminescent substrate is then added to the vessel and light generated by the reaction is measured with a luminometer. The light production is compared to the cut-off value defined during calibration of the instrument. The qualitative assessment is automatically determined from a stored calibration.
Quality control (QC) materials simulate the characteristics of patient samples and are essential for monitoring the system performance of the Access anti-HAV IgM immunoassay. In addition, they are an integral part of good laboratory practices. When performing assays with Access reagents for anti-HAV IgM, include quality control materials to validate the integrity of the assay. The assayed values should fall within the acceptable range if the test system is working properly.
The Access anti-HAV IgM reagents are provided in liquid ready-to-use format designed for optimal performance on the Beckman Coulter DxI 9000 Access Immunoassay Analyzer only. Each reagent kit contains two reagent packs. The Access anti-HAV IgM Calibrator kit contains one vial, and the Access anti-HAV IgM QC kit contains three vials each of anti-HAV IgM positive control and anti-HAV IgM negative control. Other items needed to run the assay include Lumi-Phos PRO (chemiluminescent substrate) and UniCel DxI Wash Buffer II.
N/A
Ask a specific question about this device
(263 days)
The Access Rubella IgG assay is a paramagnetic-particle, chemiluminescent immunoassay for the qualitative and quantitative determination of IgG antibodies to the rubella virus in human serum using the Access Immunoassay Systems. The Access Rubella IgG assay aids in the diagnosis of rubella infection and the determination of immunity.
The Access Rubella IgG assay is a paramagnetic-particle, chemiluminescent immunoassay for the qualitative and quantitative detection of IgG antibodies to the rubella virus in human serum using the Access Immunoassay Systems.
The Access Rubella IgG assay consists of the reagent pack, calibrators, and quality controls (QCs), packaged separately. Other items needed to run the assay include substrate and wash buffer.
N/A
Ask a specific question about this device
(26 days)
The iQ200 System is an in-vitro diagnostic device used to automate the complete urinalysis profile, including urine test strip chemistry panel, and microscopic sediment analysis. Optionally, the iQ200 Analyzer can be used as a stand-alone unit, or the results from the iQ200 analyzer can be combined with other urine chemistry results received from an LIS. It produces quantitative or qualitative counts of all formed sediment elements present in urine, including cells, casts, crystals, and organisms. A competent human operator can set criteria for auto-reporting and flagging specimens for review. All instrument analyte image decisions may be reviewed and overridden by a trained technologist.
The iQ200 Series Automated Urine Microscopy system utilizes a specimen sandwiched between lamina layers presented to a microscope with a CCD video camera. This ensures the specimen is precisely within the microscope's focus and field of view. The system automates sample handling and analyte classification for improved data reporting and management. Specimens are aspirated by an autosampler, and individual particle images are isolated in each frame. The Auto-Particle Recognition (APR) software classifies images into 12 categories, and more, with 27 additional sub-classifications available. Particle concentration is determined by the number of images and the analyzed volume. Results are checked against user-defined criteria and sent for operator review or directly uploaded to the LIS. Specimen results can be edited, imported, and exported.
N/A
Ask a specific question about this device
(267 days)
The Access anti-HAV assay is a paramagnetic particle, chemiluminescent immunoassay for the in vitro qualitative detection of total antibodies (anti-HAV IgG and IgM) to hepatitis A virus (HAV) in human pediatric (2 through 21 years) and adult serum and serum separator tubes or plasma [lithium heparin, lithium heparin separator tubes, sodium citrate, acid-citrate-dextrose (ACD), and citrate phosphate-dextrose (CPD)] using the Dxl 9000 Access Immunoassay Analyzer.
The Access anti-HAV assay is indicated as an aid in the diagnosis of current or past HAV infection in persons with risk factors and/or signs or symptoms of hepatitis A, when used in conjunction with other serological and clinical information. The assay may also be used in the identification of HAV susceptible individuals and to determine the presence of an antibody response to HAV in vaccine recipients.
This assay is not intended for use for screening donors of blood or blood products or human cells, tissues, or cellular or tissue-based products (HCT/Ps).
The Access anti-HAV assay requires Access anti-HAV (reagent packs), Access anti-HAV Calibrator (C1), and Access anti-HAV QC (QC1-QC2). The Access anti-HAV assay is a two-step competitive immunoassay. During incubation, the anti-HAV antibodies present in the patient sample bind to the coated antigen. After incubation in a reaction vessel, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. A monoclonal anti-HAV antibody alkaline phosphatase conjugate is added to the reaction vessel and the conjugate competes with the bound patient antibodies to affix the HAV antigen coated on the particles. After a second incubation and wash step, the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is compared to the cutoff value defined during calibration. The Qualitative assessment is automatically determined from a stored calibration.
Quality control (QC) materials simulate the characteristics of patient samples and are essential for monitoring the system performance of the Access anti-HAV immunoassay. In addition, they are an integral part of good laboratory practices. When performing assays with Access reagents for anti-HAV, include quality control materials to validate the integrity of the assay. The assayed values should fall within the acceptable range if the test system is working properly.
The Access anti-HAV reagents are provided in liquid ready-to-use format designed for optimal performance on the Beckman Coulter DxI 9000 Access Immunoassay Analyzer only. Each reagent kit contains two reagent packs. The Access anti-HAV Calibrator kit contains one vial, and the Access anti-HAV QC kit contains three vials, each of anti-HAV positive control and anti-HAV negative control. Other items needed to run the assay include Lumi-Phos PRO (chemiluminescent substrate) and UniCel DxI Wash Buffer II.
N/A
Ask a specific question about this device
(252 days)
The MicroScan Dried Gram-Negative MIC/Combo Panel is used to determine quantitative and qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-negative bacilli. After inoculation, panels are incubated for 16-20 hours at 35°C ± 1°C in a non-CO2 incubator, and read either visually or with MicroScan instrumentation, according to the Package Insert.
This particular submission is for the addition of the antimicrobial cefepime at concentrations of 0.12-64 µg/mL to the test panel. Testing is indicated for Enterobacterales, Pseudomonas aeruginosa and Aeromonas spp., as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC) webpage.
The MicroScan Dried Gram-Negative MIC/Combo Panels with Cefepime (CPE) (0.12-64µg/mL) has demonstrated acceptable performance with the following organisms:
Enterobacterales (Enterobacter spp., Escherichia coli, Klebsiella pneumoniae, Proteus mirabilis, Citrobacter koseri, (formerly Citrobacter diversus), Citrobacter freundii complex (Citrobacter freudnii, Citrobacter werkmanii and Citrobacter youngae), Klebsiella oxytoca, Morganella morganii, Proteus vulgaris, Providencia stuartii, Providencia rettgeri, Serratia marcescens)
Pseudomonas aeruginosa
Aeromonas spp.
MicroScan Dried Gram-Negative MIC/Combo Panels are designed for use in determining quantitative and qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-negative bacilli.
The principle of MicroScan panels with antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test that have been diluted in broth and dehydrated. Various antimicrobial agents are diluted in broth to concentrations bridging the range of clinical interest. Panels are rehydrated with water after inoculation with a standardized suspension of the organism. After incubation in a non-CO2 incubator for 16-20 hours, the minimum inhibitory concentration (MIC) for the test organism is read by determining the lowest antimicrobial concentration showing inhibition of growth.
The product is single-use and intended for laboratory professional use.
Device Performance Acceptance Criteria and Study Details for MicroScan Dried Gram-Negative MIC/Combo Panels with Cefepime
Based on the provided FDA 510(k) Clearance Letter, the device in question is the MicroScan Dried Gram-Negative MIC/Combo Panels with Cefepime (CPE) (0.12-64 µg/mL), which is an Antimicrobial Susceptibility Test (AST) System. The study described focuses on demonstrating the substantial equivalence of this new configuration (with Cefepime) to a predicate device.
Given the nature of the device (an AST System), the "acceptance criteria" are typically related to the accuracy of determining Minimum Inhibitory Concentration (MIC) and the resulting categorical agreement (Susceptible, Intermediate, Resistant) compared to a reference method. The "study that proves the device meets the acceptance criteria" refers to the performance evaluation conducted for the 510(k) submission.
1. Table of Acceptance Criteria and Reported Device Performance
For AST systems, the key performance metrics are Essential Agreement (EA) and Categorical Agreement (CA) when compared to a CLSI (Clinical and Laboratory Standards Institute) frozen reference panel. The FDA document "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", dated August 28, 2009, likely outlines the specific acceptance criteria thresholds for EA and CA. While the exact numerical acceptance criteria are not explicitly stated in the provided text, the performance "demonstrated acceptable performance" implies meeting these pre-defined thresholds.
| Performance Metric | Organism Group (Inoculation/Read Method) | Reported Device Performance (Essential Agreement) | Reported Device Performance (Categorical Agreement) | Acceptance Criteria (Implied / Based on FDA Guidance for AST) |
|---|---|---|---|---|
| Essential Agreement (EA) | Aeromonas spp. (Prompt Inoculation/WalkAway Instrument) | 93.5% | N/A | Typically ≥ 90% (Guidance based, not explicitly stated as a number) |
| Categorical Agreement (CA) | Aeromonas spp. (Prompt Inoculation/WalkAway Instrument) | N/A | 90.3% | Typically ≥ 90% (Guidance based, not explicitly stated as a number) |
| Essential Agreement (EA) | Pseudomonas aeruginosa (Prompt Inoculation/WalkAway Instrument) | 95.7% | N/A | Typically ≥ 90% (Guidance based, not explicitly stated as a number) |
| Categorical Agreement (CA) | Pseudomonas aeruginosa (Prompt Inoculation/WalkAway Instrument) | N/A | 91.4% | Typically ≥ 90% (Guidance based, not explicitly stated as a number) |
| Essential Agreement (EA) | Enterobacterales (Turbidity Method/WalkAway Instrument) | 94.7% | N/A | Typically ≥ 90% (Guidance based, not explicitly stated as a number) |
| Categorical Agreement (CA) | Enterobacterales (Turbidity Method/WalkAway Instrument) | N/A | 96.3% | Typically ≥ 90% (Guidance based, not explicitly stated as a number) |
| Essential Agreement (EA) | Aeromonas spp. (Turbidity Inoculation/autoSCAN-4 and Manual Reads) | 100.0% | N/A | Typically ≥ 90% (Guidance based, not explicitly stated as a number) |
| Essential Agreement of Evaluable Isolates | Aeromonas spp. (Turbidity Inoculation/autoSCAN-4 and Manual Reads) | 100.0% | N/A | N/A (Supplementary metric) |
| Categorical Agreement (CA) | Aeromonas spp. (Turbidity Inoculation/autoSCAN-4 and Manual Reads) | N/A | 87.1% | Typically ≥ 90% (Guidance based, not explicitly stated as a number) |
| Categorical Agreement (CA) | Aeromonas spp. (Turbidity Inoculation/WalkAway Read Method) | N/A | Below 90% | Typically ≥ 90% (Guidance based, not explicitly stated as a number) |
| Inoculum and Instrument Reproducibility | Cefepime (Turbidity/Prompt, autoSCAN-4/WalkAway) | Acceptable Reproducibility and Precision | N/A | (Implied acceptable performance) |
| Quality Control Testing | Cefepime | Acceptable Results | N/A | (Implied acceptable performance) |
Important Note: The document highlights some instances where the performance was "outside of essential agreement" for Enterobacterales with Prompt inoculation and "below 90%" for Aeromonas spp. with turbidity inoculation and WalkAway read method. These discrepancies are "mitigated with a limitation" in the product labeling, suggesting that while initial performance in those specific conditions did not meet implicit criteria, the overall robust performance with other methods/organisms, coupled with labeling limitations, made the device acceptable for clearance.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not explicitly provide a total number for the test set sample size (e.g., number of isolates tested). It refers to "contemporary and stock Efficacy isolates and stock Challenge strains" used for external evaluations.
- Data Provenance: The document does not specify the country of origin of the data. It mentions "external evaluations," which generally implies testing conducted at clinical sites or contract research organizations. The study appears to be retrospective in the sense that it uses "stock Efficacy isolates and stock Challenge strains" which are pre-existing collections of bacterial isolates. It also mentions "contemporary" isolates, suggesting some recent collection. It implies a laboratory-based performance study rather than a clinical trial with patient outcomes.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This type of device (AST System) does not typically rely on human expert interpretation for establishing the "ground truth" of the test set. The ground truth for antimicrobial susceptibility testing is established by a reference method, which for this device is stated as a "CLSI frozen Reference Panel."
Therefore:
- Number of Experts: Not applicable in the context of creating the ground truth for AST.
- Qualifications of Experts: Not applicable.
4. Adjudication Method for the Test Set
As the ground truth is established by a reference method (CLSI frozen Reference Panel), there is no human adjudication method like 2+1 or 3+1 typically used for image-based diagnostics. The device's results are directly compared to the quantitatively or qualitatively determined results from the CLSI reference method.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
There is no indication that an MRMC comparative effectiveness study was performed. This type of study is not relevant for this device, which is an automated or manually read laboratory diagnostic for antimicrobial susceptibility, not an AI-assisted diagnostic tool that aids human readers in interpretation. The device itself performs the susceptibility test.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the performance data presented is effectively standalone performance of the device (MicroScan Dried Gram-Negative MIC/Combo Panels with Cefepime). The device "read either visually or with MicroScan instrumentation" and its performance (Essential Agreement, Categorical Agreement) is directly compared to the reference standard. The "human-in-the-loop" would be the laboratory professional reading the results, and the study evaluates the accuracy of the device itself in producing those results. Where visual reads are mentioned, it's about the device's ability to produce clear inhibition patterns for visual interpretation, not a human independently interpreting raw data without the device.
7. The Type of Ground Truth Used
The ground truth used was established by a CLSI frozen Reference Panel. This is a recognized and standardized method for determining antimicrobial susceptibility, often involving broth microdilution or agar dilution methods where organisms are tested against known concentrations of antimicrobials. It is a highly controlled and quantitative method to determine the true MIC value against which the device's performance is compared.
8. The Sample Size for the Training Set
The document does not mention a training set or any details about its sample size. This is consistent with the nature of the device. AST systems are generally rule-based or empirically derived systems based on established microbiological principles, rather than machine learning models that require distinct training sets. The development of such panels involves extensive empirical testing during the R&D phase to ensure the correct concentrations and formulations, but this isn't typically referred to as a "training set" in the context of an AI/ML model.
9. How the Ground Truth for the Training Set was Established
As no training set (in the AI/ML sense) is indicated, this point is not applicable.
Ask a specific question about this device
(219 days)
The MicroScan Dried Gram-Positive MIC/Combo Panel is used to determine quantitative and qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic gram-positive cocci and Listeria monocytogenes. After inoculation, panels are incubated for 16-20 hours at 35°C ± 1°C in a non-CO2 incubator, and read either visually or with MicroScan instrumentation, according to the Package Insert.
This particular submission is for the addition of the antimicrobial daptomycin at concentrations of 0.06-32 µg/mL to the test panel. Testing is indicated for Enterococcus faecium, Enterococcus spp. other than E. faecium, and Staphylococcus spp., as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC) webpage.
The MicroScan Dried Gram-Positive MIC/Combo Panels with Daptomycin (DAP) (0.06-32 µg/mL) has demonstrated acceptable performance with the following organisms:
- Enterococcus faecium
- Enterococcus spp. other than E. faecium (Enterococcus faecalis, Enterococcus avium, Enterococcus raffinosus, Enterococcus casseliflavus and Enterococcus durans)
- Staphylococcus spp. (Staphylococcus aureus, Staphylococcus epidermidis, Staphylococcus capitis, Staphylococcus haemolyticus, Staphylococcus lugdunensis, Staphylococcus hominis, Staphylococcus warneri, Staphylococcus simulans, Staphylococcus saprophyticus, Staphylococcus intermedius, and Staphylococcus sciuri)
MicroScan Dried Gram-Positive MIC/Combo Panels are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-positive bacteria.
The principle of MicroScan panels with antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test that have been diluted in broth and dehydrated. Various antimicrobial agents are diluted in broth to concentrations bridging the range of clinical interest. Panels are rehydrated with water after inoculation with a standardized suspension of the organism. After incubation in a non-CO2 incubator for 16-20 hours, the minimum inhibitory concentration (MIC) for the test organism is read by determining the lowest antimicrobial concentration showing inhibition of growth.
This product is single-use and intended for laboratory professional use.
The provided text specifies the performance validation of the MicroScan Dried Gram-Positive MIC/Combo Panels with Daptomycin (DAP). Here's a breakdown of the acceptance criteria and study information:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the reported performance metrics against a CLSI (Clinical and Laboratory Standards Institute) frozen Reference Panel. The primary metrics are Essential Agreement (EA) and Categorical Agreement (CA).
| Organism Group | Performance Metric | Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|---|---|
| Staphylococcus spp. | Essential Agreement (EA) | Acceptable performance | 94.3% |
| Categorical Agreement (CA) | Acceptable performance | 99.5% | |
| Enterococcus faecium | Essential Agreement (EA) | Acceptable performance | 90.8% |
| Categorical Agreement (CA) | Acceptable performance | 92.0% | |
| Enterococcus species other than E. faecium | Essential Agreement (EA) | Acceptable performance | 100.0% |
| Categorical Agreement (CA) | Acceptable performance | 94.1% |
Note: The document states "acceptable performance" without defining specific numerical thresholds for EA and CA as explicit "acceptance criteria." However, the reported values are presented as meeting this "acceptable performance." In antimicrobial susceptibility testing, typical FDA guidance for acceptable EA is generally $\geq$90% and for CA is general $\geq$90% to 95%, depending on the organism/drug combination and resistance rates.
2. Sample Size Used for the Test Set and Data Provenance
The document mentions "external evaluations" conducted with:
- Fresh and stock Efficacy isolates
- Stock Challenge strains
It does not explicitly state the numerical sample size (number of isolates/strains) used for the test set.
The data provenance is not explicitly stated regarding country of origin or whether it was retrospective or prospective. Given the mention of "external evaluations," it implies that the testing was performed, but the location and study design (retrospective/prospective) are not detailed. Standard AST studies typically involve prospective collection of clinical isolates and/or the use of well-characterized reference strains.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The ground truth for the test set was established by comparison with a CLSI frozen Reference Panel. This implies that the reference standard, and thus the "ground truth," is determined by a highly standardized and validated laboratory method (broth microdilution or similar CLSI-recognized standard).
The document does not mention human experts being used directly to establish the ground truth for individual cases in the test set. Instead, the CLSI frozen Reference Panel itself serves as the gold standard.
4. Adjudication Method for the Test Set
No human adjudication method (e.g., 2+1, 3+1) is mentioned or implied, as the ground truth is established by the CLSI frozen Reference Panel, not by human interpretation or consensus.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
This is not applicable to this device. This device is an Antimicrobial Susceptibility Test (AST) panel, which determines the Minimum Inhibitory Concentration (MIC) of an antimicrobial agent against bacteria. It does not involve human readers interpreting images, and therefore, an MRMC study is not relevant. The device measures a quantitative result (MIC) which is then interpreted categorically (susceptible, intermediate, resistant) based on established breakpoints.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, this study represents a standalone performance evaluation of the device. The "MicroScan Dried Gram-Positive MIC/Combo Panels with Daptomycin" is a diagnostic test system. Its performance is evaluated against a recognized reference method (CLSI frozen Reference Panel) to determine its accuracy in reporting MIC values and categorical interpretations. While human operators inoculate and read the panels (either visually or with MicroScan instrumentation), the "performance" being assessed here is the device's ability to produce accurate results compared to the reference standard, not an AI algorithm's ability to interpret data without human input.
7. The type of Ground Truth Used
The ground truth used was comparison to a CLSI frozen Reference Panel. This is a laboratory-based gold standard for antimicrobial susceptibility testing, representing the accepted accurate measurement of MIC. It is a highly controlled and standardized method.
8. The Sample Size for the Training Set
The document does not mention a "training set" in the context of machine learning or AI. This is an AST panel, not an AI/ML diagnostic algorithm that requires a training set. The performance validation is based on a comparison to a reference standard using a test set of bacterial isolates/strains.
9. How the Ground Truth for the Training Set Was Established
As there is no concept of a training set for this type of device (AST panel based on traditional microbiological principles), this question is not applicable.
Ask a specific question about this device
(266 days)
Access hsTnI is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cardiac troponin I (cTnI) levels in human serum and plasma using the Unicel DxI Immunoassay Systems to aid in the diagnosis of myocardial infarction (MI).
The Access hsTnI is a two–site immunoenzymatic ("sandwich") assay. Monoclonal anti–cTnI antibody conjugated to alkaline phosphatase is added to a reaction vessel along with a surfactant–containing buffer and sample. After a short incubation, paramagnetic particles coated with monoclonal anti–cTnI antibody are added. The human cTnI binds to the anti–cTnI antibody on the solid phase, while the anti–cTnI antibody–alkaline phosphatase conjugate reacts with different antigenic sites on the cTnI molecules. After incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is directly proportional to the concentration of analyte in the sample. Analyte concentration is automatically determined from a stored calibration.
Here's a breakdown of the acceptance criteria and study information for the Access hsTnI device, based on the provided FDA 510(k) clearance letter:
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Acceptance Criteria | Reported Device Performance (Access hsTnI Candidate on UniCel DxI 800) |
|---|---|---|
| Method Comparison | Slope of Passing-Bablok linear regression model: 1.00 ± 0.10 | Met acceptance criteria (exact slope not provided, but stated that the study "met the acceptance criteria of slope 1.00 ± 0.10"). Bias data supported that reference intervals have not changed appreciably from the commercialized product. |
| Imprecision | Within-laboratory (total) CV: ≤ 10% for concentrations ≥ 11.5 pg/mLWithin-laboratory (total) SD: ≤ 1.15 pg/mL for concentrations < 11.5 pg/mL | Within-laboratory (total) % CV ranged from 3% to 4% for Access hsTnI concentrations ≥ 11.5 pg/mL. The within-laboratory (total) SD was 0.48 pg/mL for Access hsTnI concentration < 11.5 pg/mL. |
| Linearity | For each sample concentration range, the higher order (2nd or 3rd) term of the polynomial fit should be non-significant (p > 0.05). If significant, the fit of the polynomial regression demonstrating significance should have ≤ 10% bias across the analytical measuring range. | The analysis of the data found that across the UniCel DxI 800 instruments, and for each sample concentration range, the higher order (2nd or 3rd) term of the polynomial fit is non-significant (p > 0.05), and if significant, the fit of the polynomial regression demonstrating significance had ≤ 10% bias across the analytical measuring range. |
| LoB/LoD | Not explicitly stated as an "acceptance criteria" but limits are reported for the predicate. | LoB estimate of the Access hsTnI is 1.5 (serum and plasma).LoD estimate of the Access hsTnI is 1.8 (serum and plasma). |
| LoQ | Not explicitly stated as an "acceptance criteria" but limits are reported for the predicate. | The LoQ for Access hsTnI at ≤20% within-lab CV was determined to be 1.3 pg/mL (serum) and 1.2 pg/mL (plasma). |
| Carryover | Not explicitly stated as an "acceptance criteria" for numeric limits, but the sponsor performed studies and included a limitation in labeling acknowledging observed carryover. | When a sample with cTnI > 150,000 pg/mL (ng/L) was tested, intra-assay carryover was observed if an Access hsTnI was tested after a high cTnI sample. Estimated carryover was 3-5 pg/mL (ng/L) from a high sample at 270,000 pg/mL (ng/L) and 5-8 pg/mL (ng/L) from a high sample at 500,000 pg/mL (ng/L). Limitation statements related to carryover are to be added. |
| Analytical Measuring Range | 2.3 pg/mL to 27,027 pg/mL (Predicate) | Similar (Candidate) |
Study Details
The provided document describes a study primarily focused on demonstrating the substantial equivalence of the Access hsTnI assay when run on the UniCel DxI 800 Immunoassay System compared to its predicate device (Access hsTnI on the Access 2 Immunoassay System). This is achieved through performance testing of various analytical aspects.
-
Sample Size Used for the Test Set and Data Provenance:
- Method Comparison: 239 samples (119 Lithium Heparin Plasma and 120 Serum).
- Data Provenance: Not specified (e.g., country of origin). The document indicates it's a retrospective comparison between the "IVD Access hsTnI (Current Assay Protocol File (APF))" and the "proposed Access hsTnI (Proposed APF)" on the UniCel DxI 800 instruments, implying existing samples or previously collected data.
-
Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- Not Applicable. This device is an in-vitro diagnostic (IVD) immunoassay for quantitative determination of cTnI levels. The "ground truth" for the test set in this context refers to the measured cTnI values, which are inherently quantitative and determined by the predicate device's method and the proposed device's method, not by expert consensus or interpretation of images/clinical findings.
-
Adjudication Method for the Test Set:
- Not Applicable. As noted above, this is a quantitative analytical method comparison, not a diagnostic interpretation or clinical outcome study that would require adjudication.
-
If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:
- Not Applicable. This is an in-vitro diagnostic device, not an AI-based image interpretation or diagnostic aid system involving human readers.
-
If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
- Yes, implicitly. The studies described (Method Comparison, Imprecision, Linearity, LoB/LoD, LoQ, Carryover) evaluate the performance of the analytical instrument and assay without human intervention in the measurement process. The device itself is an automated immunoassay system that produces quantitative results.
-
The Type of Ground Truth Used:
- The "ground truth" in this context refers to the quantitative measurements of cTnI levels themselves. For the method comparison, the predicate device (Access hsTnI on the Access 2 Immunoassay System, or the "Current Assay Protocol File (APF)" on the DxI 800) essentially serves as the reference for comparison against the "Proposed APF" on the UniCel DxI 800. Therefore, it's a comparison against an established, legally marketed reference measurement method.
-
The Sample Size for the Training Set:
- Not specified. This documentation primarily focuses on the validation of the device's analytical performance. While there would have been internal development and optimization (which could be considered "training"), the document does not distinguish a formal "training set" (as might be seen with AI/ML models) from "internal validation" data. The tested datasets described are for analytical validation.
-
How the Ground Truth for the Training Set Was Established:
- Not specified / Not applicable in the traditional sense. As an IVD assay, the "ground truth" for developing such a test is the accurate quantitative measurement of the analyte (cTnI) in biological samples, requiring highly controlled reference methods and materials. The document indicates the device's principle is a "two-site immunoenzymatic ('sandwich') assay," which is a well-established biochemical technique. The development process would involve extensive characterization against reference standards and known concentrations, rather than establishing ground truth through, for example, expert labeling of clinical data.