K Number
K252169

Validate with FDA (Live)

Device Name
Access BNP II
Date Cleared
2026-03-24

(257 days)

Product Code
Regulation Number
862.1117
Age Range
22 - 120
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Access BNP II test is intended for use with the Beckman Coulter Access Family of Immunoassay Systems for the In Vitro quantitative measurement of B-type natriuretic peptide (BNP) in plasma specimens using EDTA as the anticoagulant. The test is intended to be used for the following indications:

  • as an aid in the diagnosis of heart failure (HF) in patients presenting to the emergency department (ED) with clinical suspicion of new onset, acutely decompensated, or exacerbated heart failure
  • for the risk stratification of patients with acute coronary syndromes
  • for the risk stratification of patients with heart failure
Device Description

The Access BNP II assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of B-type natriuretic peptide (BNP) levels in human EDTA plasma using the Access Immunoassay Systems.

The Access BNP Calibrators are intended to calibrate the Access BNP and Access BNP II tests for the quantitative determination of BNP levels in human EDTA plasma using the Beckman Coulter Access Family of Immunoassay Systems.

The Access BNP II test is a two-site immunoenzymatic ("sandwich") assay. A sample is added to a reaction vessel with mouse monoclonal anti-human BNP antibody-alkaline phosphatase conjugate and paramagnetic particles coated with mouse Omniclonal anti-human BNP antibody. BNP in human plasma binds to the immobilized anti-BNP on the solid phase, while the mouse anti-BNP conjugate reacts specifically with bound BNP.

After incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is inversely proportional to the concentration of analyte in the sample. Analyte concentration is automatically determined from a stored calibration.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - Access BNP II

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.04

March 24, 2026

Beckman Coulter, Inc.
Neha Desai
Staff Quality and Regulatory Affairs
1000 Lake Hazeltine Dr.
Chaska, MN 55318

Re: K252169
Trade/Device Name: Access BNP II
Regulation Number: 21 CFR 862.1117
Regulation Name: B-type natriuretic peptide test system
Regulatory Class: Class II
Product Code: NBC
Dated: March 2, 2026
Received: March 3, 2026

Dear Neha Desai:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K252169 - Neha Desai Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K252169 - Neha Desai Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

PAULA V. CAPOSINO -S

Paula Caposino, Ph.D.
Deputy Director
Division of Chemistry and
Toxicology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.

510(k) Number (if known): K252169

Device Name: Access BNP II

Indications for Use (Describe):

The Access BNP II test is intended for use with the Beckman Coulter Access Family of Immunoassay Systems for the In Vitro quantitative measurement of B-type natriuretic peptide (BNP) in plasma specimens using EDTA as the anticoagulant. The test is intended to be used for the following indications:

  • as an aid in the diagnosis of heart failure (HF) in patients presenting to the emergency department (ED) with clinical suspicion of new onset, acutely decompensated, or exacerbated heart failure
  • for the risk stratification of patients with acute coronary syndromes
  • for the risk stratification of patients with heart failure

Type of Use (Select one or both, as applicable):

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

Access BNP II 1 of 12
Confidential - Company Proprietary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

510(k) Number: K252169

Submitted By:
Beckman Coulter, Inc.
1000 Lake Hazeltine Drive
Chaska, MN 55318

Primary Contact:
Neha Desai
Staff Quality and Regulatory Affairs
Phone: (612) 244-9788
Email: nhdesai@beckman.com

Alternate Contact:
Kate Oelberg
Senior Staff Quality and Regulatory Affairs
Phone: (612) 431-7315
Email: kmoelberg@beckman.com

Device Name

Common Name: Access BNP
Trade Name: Access BNP II
Classification Name: B-Type Natriuretic Peptide Test System
Classification Regulation: [21 CFR 862.1117]

Predicate Device

Device Name: BNP Test
510(k) Numbers: K052789

Device Description

The Access BNP II assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of B-type natriuretic peptide (BNP) levels in human EDTA plasma using the Access Immunoassay Systems.

The Access BNP Calibrators are intended to calibrate the Access BNP and Access BNP II tests for the quantitative determination of BNP levels in human EDTA plasma using the Beckman Coulter Access Family of Immunoassay Systems.

The Access BNP II test is a two-site immunoenzymatic ("sandwich") assay. A sample is added to a reaction vessel with mouse monoclonal anti-human BNP antibody-alkaline phosphatase conjugate and paramagnetic particles coated with mouse Omniclonal anti-human BNP antibody. BNP in human

Page 6

Access BNP II 2 of 12
Confidential - Company Proprietary

plasma binds to the immobilized anti-BNP on the solid phase, while the mouse anti-BNP conjugate reacts specifically with bound BNP.

After incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is inversely proportional to the concentration of analyte in the sample. Analyte concentration is automatically determined from a stored calibration.

Intended Use

The Access BNP II test is intended for use with the Beckman Coulter Access Family of Immunoassay Systems for the In Vitro quantitative measurement of B-type natriuretic peptide (BNP) in plasma specimens using EDTA as the anticoagulant. The test is intended to be used for the following indications:

  • as an aid in the diagnosis of heart failure (HF) in patients presenting to the emergency department (ED) with clinical suspicion of new onset, acutely decompensated, or exacerbated heart failure
  • for the risk stratification of patients with acute coronary syndromes
  • for the risk stratification of patients with heart failure

Comparison of Technological Characteristics to the Predicate

ParameterAccess BNP Assay on Access 2 Immunoassay System (Predicate)Access BNP II Assay on DxI 9000 Access Immunoassay System
Intended useThe Triage BNP test is intended for use with Beckman Coulter Immunoassay Systems (Access, Access 2, Synchron LXi 725 and UniCel Dxl 800) for the In Vitro quantitative measurement of B-type natriuretic peptide (BNP) in plasma specimens using EDTA as the anticoagulant. The test is intended to be used as an aid in the diagnosis and assessment of severity of congestive heart failure (also referred to as heart failure). The test also is used for the risk stratification of patients with acute coronary syndromes and for the risk stratification of patients with heart failure.The Access BNP test is intended for use with the Beckman Coulter Access Family of Immunoassay Systems for the In Vitro quantitative measurement of B-type natriuretic peptide (BNP) in plasma specimens using EDTA as the anticoagulant. The test is intended to be used for the following indications: • as an aid in the diagnosis of heart failure (HF) in patients presenting to the emergency department (ED) with clinical suspicion of new onset, acutely decompensated, or exacerbated heart failure • for the risk stratification of patients with acute coronary syndromes • for the risk stratification of patients with heart failure
TechnologyTwo-site immunoenzymatic assaySame
FormatChemiluminescentSame

Page 7

Access BNP II 3 of 12
Confidential - Company Proprietary

ParameterAccess BNP Assay on Access 2 Immunoassay System (Predicate)Access BNP II Assay on DxI 9000 Access Immunoassay System
CalibrationUtilizes a stored calibration curveSame
Sample TypeEDTA PlasmaSame
Measuring Range1-5,000 pg/ml5-5,000 pg/ml
Sample Volume55 μL13 μL
InstrumentAccess Immunoassay systemDxI 9000 Access Immunoassay Analyzer
SubstrateAccess SubstrateLumi-Phos Pro Substrate

Standard/Guidance Document Referenced (if applicable):

  • CLSI EP05-A3: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline – Third Edition
  • CLSI EP06-2nd Edition: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
  • CLSI EP07- 3rd Edition: Interference Testing in Clinical Chemistry
  • CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition
  • CLSI EP09c 3rd Edition: Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Third Edition
  • CLSI EP24-A2: Assessment of the Diagnostic Accuracy of Laboratory Tests Using Receiver Operating Characteristic Curves; Draft Guideline-Second Edition
  • CLSI EP12-A2: User Protocol for Evaluation of Qualitative Test Performance
  • CLSI EP25-Ed2: Evaluation of Stability of In Vitro Medical Laboratory Test Reagents - 2nd Edition
  • CLSI EP28-A3c: Defining Establishing and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition
  • ISO 20916: In vitro diagnostic medical devices- Clinical Performance studies using specimens from human subjects- Good Study Practice

Page 8

Access BNP II 4 of 12
Confidential - Company Proprietary

Summary of Non-Clinical Studies

Method Comparison:

A study based on CLSI EP09c, 3rd Edition using Weighted Deming regression and Pearson's correlation compared Access BNP II on the DxI 9000 Access Immunoassay Analyzer with Access BNP on the Access 2 Immunoassay System.

NConcentration Range (pg/mL)*SlopeSlope 95% CIInterceptIntercept 95% CICorrelation Coefficient R
1447.5 – 46411.061.02 – 1.10-6.8-9.5 – -4.11.00

*Range is Access 2 values

Imprecision:

The assay was designed to have within-laboratory imprecision as listed below:

  • SD ≤ 2.4 pg/mL for values < 30 pg/mL
  • CV ≤ 8 % for values ≥ 30 pg/mL

A study based on CLSI EP05-A3 performed on the DxI 9000 Access Immunoassay Analyzer tested multiple samples in duplicate in 2 runs per day for a minimum of 20 days.

SampleNMean pg/mLRepeatability (Within-run)Between-runBetween-dayWithin-Laboratory (Total)
SD pg/mL%CVSD pg/mL%CVSD pg/mL%CVSD pg/mL%CV
SAMPLE 1805.10.5100.000.470.612
SAMPLE 2801218.770.002.329.07
SAMPLE 38022611.458.445.0215.07
SAMPLE 48034912.748.6212.5419.76
SAMPLE 580103045.1436.1429.9365.06
SAMPLE 680141438.2318.4136.5355.94
SAMPLE 780385494.4278.32103.53160.54

Linearity: A study based on CLSI EP06-Ed2 performed on the DxI 9000 Access Immunoassay Analyzer determined the assay demonstrated linearity across the measuring interval.

Hook Effect: A study was performed to verify the hook effect for the Access BNP II assay immunoassay on the DxI 9000 Access Immunoassay Analyzer. Access BNP II assay has no high dose hook effect observed up to 150,000 pg/mL.

Page 9

Access BNP II 5 of 12
Confidential - Company Proprietary

Interfering Substances

EDTA plasma samples that contained BNP concentrations of approximately 100 and 500 pg/mL were spiked with the substances listed in the following table. The spiked samples were run on the DxI 9000 Access Immunoassay Analyzer. The values were calculated based upon CLSI EP07-A3 guidelines. The interference was determined by testing controls (with no interfering substance added) and matched test samples (with interfering substance added). None of the compounds tested were found to cause significant interference (as defined by a shift in dose > 10%) using the test concentrations provided in the following table.

SubstanceHighest Concentration AddedSubstanceHighest Concentration Added
Acetaminophen20 mg/dLHydrochlorothiazide20 μg/mL
Allopurinol40 mg/dLIbuprofen40 mg/dL
Ambroxol40 mg/dLIndomethacin36 μg/mL
Amiodarone4.2 mg/dLIsosorbide dinitrate0.593 mg/dL
Ampicillin5 mg/dLLisinopril16 μg/mL
Ascorbic acid3 mg/dLLovastatin0.021 mg/dL
Aspirin50 mg/dLMethyldopa2.5 mg/dL
Atenolol1 mg/dLNicotine1.6 μg/mL
Conjugated Bilirubin20 mg/dLNifedipine6 mg/dL
Unconjugated Bilirubin20 mg/dLNitrofurantoin6.4 mg/dL
Biotin3,510 ng/mLNystatin0.7 mg/dL
Caffeine10 mg/dLOxazepam12 μg/mL
Captopril1.25 mg/dLOxytetracycline0.5 mg/dL
Chloramphenicol7.8 mg/dLPhenobarbital69 mg/dL
Cinnarizine40 mg/dLPhenytoin10 mg/dL
Clopidogrel bisulfate30 μg/mLProbenecid600 μg/mL
Cyclosporine40 μg/mLProcainamide4.8 mg/dL
Diclofenac2 mg/dLPropranolol0.5 mg/dL
Digoxin0.02 mg/dLProtein (Human Serum Albumin)6000 mg/dL
Dipyridamole30 μg/mLQuinidine Sulfate5 mg/dL

Page 10

Access BNP II 6 of 12
Confidential - Company Proprietary

Cross Reactivity

A study evaluated the potential for cross-reactivity of substances that are similar in structure to BNP. EDTA plasma samples that contained BNP concentrations of approximately 100 and 500 pg/mL were spiked with concentrations of the substances listed in the following table. The spiked samples were run on the DxI 9000 Access Immunoassay Analyzer. The values were calculated based upon CLSI EP07-A3, guidelines. Cross-reactivity was observed when the listed substances were tested at the indicated concentrations. The cross reactivity was determined by testing controls (with no substance added) and matched test samples (with substance added). No significant cross-reactivity (as defined by a shift in dose >10%) was observed when the listed substances were tested at the indicated concentrations.

SubstanceHigh Concentration
Adrenomedullin1000 pg/mL
Aldosterone600 pg/mL
α Atrial Natriuretic polypeptide 1-281000 pg/mL
Angiotensin I600 pg/mL
Angiotensin II600 pg/mL
Angiotensin III1000 pg/mL
Arg Vasopressin1000 pg/mL
C type Natriuretic Peptide 531000 pg/mL
Endothelin I20 pg/mL
Prepro ANF 104-1231000 pg/mL
Prepro ANF 26-551000 pg/mL
Prepro ANF 56-921000 pg/mL
Prepro BNP 1-211000 pg/mL
Prepro BNP 22-461000 pg/mL
Renin50 ng/mL
Urodilatin 95-1261000 pg/mL

Page 11

Access BNP II 7 of 12
Confidential - Company Proprietary

Detection Capability:

Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ) studies were conducted on the DxI 9000 Access Immunoassay Analyzer following CLSI guideline EP17-A2. The LoB study included multiple reagent lots and 3 instruments over a minimum of 3 days. The LoD and LoQ studies included multiple reagent lots and 3 instruments over a minimum of 5 days.

Maximum Observed Result pg/mLDesign Criteria pg/mL
Limit of Blank (LoB)0.3≤ 1
Limit of Detection (LoD)1≤ 3
Limit of Quantitation (LoQ)2≤ 5

Page 12

Access BNP II 8 of 12
Confidential - Company Proprietary

CLINICAL PERFORMANCE EVALUATION

A multicenter prospective study was conducted in the intended use population to validate the recommended cutoff. BNP concentrations were determined in samples from 1323 adult patients ≥22 years old presenting to the Emergency Department with clinical suspicion of new onset heart failure or worsening symptoms suggestive of decompensated or exacerbated heart failure. Heart failure diagnosis collected by the clinical site was used to evaluate clinical performance of the Access BNP II assay. The HF incidence was 34% (449/1323). Descriptive statistics for the enrolled population without and with a HF diagnosis are presented in Tables 1.0 and 2.0. Clinical data demonstrated the recommended age-independent cutoff of 100 pg/mL to be an appropriate decision threshold.

Note: The Access BNP II assay is not intended to be used in isolation; results should be interpreted in conjunction with other diagnostic tests and clinical information. BNP results less than or equal to 100 pg/mL are representative of normal values in patients without HF. BNP results greater than 100 pg/mL are considered abnormal and suggestive of patients with HF.

Table 1.0: Descriptive Statistics for Non-HF Population by Age and Sex

StatisticsAll AgesAge < 45Age 45-54Age 55-64Age 65-74Age 75 +
All Subjects
N874141145232203153
Mean (pg/mL)16297100161179262
SD (pg/mL)323.3236.4186.5363.5274.3444.0
Median (pg/mL)5220254075126
IQR: 25th – 75th percentile (pg/mL)17 - 15910 - 5111 - 9114 - 13531 - 21959 - 304
Min (pg/mL)< 5< 5< 5< 558
Max (pg/mL)429413631100287918354294
Male
N469718113210778
Mean (pg/mL)16810088201157270
SD (pg/mL)361.8273.7194.8449.7205.7511.7
Median (pg/mL)4613213969126
IQR: 25th – 75th percentile (pg/mL)15 - 1555 - 299 - 6314 - 17035 - 20167 - 304
Min (pg/mL)< 5< 5< 5< 568
Max (pg/mL)429413631100287910954294
Female
N40570641009675
Mean (pg/mL)15695116108203253

Page 13

Access BNP II 9 of 12
Confidential - Company Proprietary

| SD (pg/mL) | 272.4 | 193.4 | 175.6 | 188.7 | 334.0 | 363.9 |
| Median (pg/mL) | 56 | 25 | 43 | 43 | 86 | 128 |
| IQR: 25th – 75th percentile (pg/mL) | 20 - 176 | 13 - 60 | 12 - 134 | 16 - 97 | 29 - 233 | 56 - 318 |
| Min (pg/mL) | < 5 | < 5 | < 5 | < 5 | 5 | 12 |
| Max (pg/mL) | 2317 | 932 | 840 | 1182 | 1835 | 2317 |

Table 2.0: Descriptive Statistics for HF Population by Age and Sex

StatisticsAll AgesAge < 45Age 45-54Age 55-64Age 65-74Age 75 +
All Subjects
N4454072109114110
Mean (pg/mL)10461163101011231096899
SD (pg/mL)1080.01049.31042.81217.61130.5904.9
Median (pg/mL)729919660800831626
IQR: 25th – 75th percentile (pg/mL)318 - 1419336 - 1578304 - 1363354 - 1450326 - 1492318 - 1105
Min (pg/mL)105420103013
Max (pg/mL)> 50004457> 5000> 5000> 50004618
Male*
N2782548776959
Mean (pg/mL)11491393111311451248964
SD (pg/mL)1118.31119.71107.51046.11292.9996.6
Median (pg/mL)819982772857864630
IQR: 25th – 75th percentile (pg/mL)387 - 1515718 - 1596374 - 1426394 - 1450360 - 1719360 - 1078
Min (pg/mL)109432103013
Max (pg/mL)> 50004457> 5000> 5000> 50004618
Female
N1671524324551
Mean (pg/mL)8757808031070863824
SD (pg/mL)992.7816.4885.11574.1779.2788.8
Median (pg/mL)615356390659680580
IQR: 25th – 75th percentile (pg/mL)253 - 1151191 - 1440265 - 1149152 - 1315276 - 1237279 - 1118
Min (pg/mL)205420564748
Max (pg/mL)> 500027783325> 500034713858

*Four (4) male subjects with adjudicated diagnosis of Acute HF are excluded due to undetermined Access BNP II numerical values for descriptive statistics reporting.

Page 14

Access BNP II 10 of 12
Confidential - Company Proprietary

The clinical sensitivity and specificity of the Access BNP II test using a cutoff of 100 pg/mL and the clinical site diagnosis across age groups and by sex is described in Table 3.0.

Table 3.0: Diagnostic Accuracy of the Established Access BNP II Cutoff (100 pg/mL) by Age and Sex

StatisticsAll AgesAge < 50Age 50-75Age > 75
All Subjects
Sensitivity Estimate93.1% (418/449)93.7% (74/79)91.8% (246/268)96.1% (98/102)
95% Confidence Interval(90.4-95.1%)(86.0-97.3%)(87.9-94.5%)(90.3-98.5%)
Specificity Estimate65.6% (573/874)82.0% (168/205)66.4% (352/530)38.1% (53/139)
95% Confidence Interval(62.3-68.6%)(76.1-86.6%)(62.3-70.3%)(30.5-46.4%)
NPV Estimate94.9% (573/604)97.1% (168/173)94.1% (352/374)93.0% (53/57)
95% Confidence Interval(92.8-96.4%)(93.4-98.8%)(91.3-96.1%)(83.3-97.2%)
PPV Estimate58.1% (418/719)66.7% (74/111)58.0% (246/424)53.3% (98/184)
95% Confidence Interval(54.5-61.7%)(57.5-74.7%)(53.3-62.6%)(46.1-60.3%)
Male
Sensitivity Estimate94.3% (266/282)98.0% (49/50)93.2% (165/177)94.5% (52/55)
95% Confidence Interval(91.0-96.5%)(89.5-99.6%)(88.5-96.1%)(85.1-98.1%)
Specificity Estimate65.7% (308/469)85.6% (89/104)65.8% (194/295)35.7% (25/70)
95% Confidence Interval(61.3-69.8%)(77.6-91.1%)(60.2-70.9%)(25.5-47.4%)
NPV Estimate95.1% (308/324)98.9% (89/90)94.2% (194/206)89.3% (25/28)
95% Confidence Interval(92.1-96.9%)(94.0-99.8%)(90.1-96.6%)(72.8-96.3%)
PPV Estimate62.3% (266/427)76.6% (49/64)62.0% (165/266)53.6% (52/97)
95% Confidence Interval(57.6-66.8%)(64.9-85.3%)(56.1-67.7%)(43.7-63.2%)
Female
Sensitivity Estimate91.0% (152/167)86.2% (25/29)89.0% (81/91)97.9% (46/47)

Page 15

Access BNP II 11 of 12
Confidential - Company Proprietary

StatisticsAll AgesAge < 50Age 50-75Age > 75
95% Confidence Interval(85.7-94.5%)(69.4-94.5%)(80.9-93.9%)(88.9-99.6%)
Specificity Estimate65.4% (265/405)78.2% (79/101)67.2% (158/235)40.6% (28/69)
95% Confidence Interval(60.7-69.9%)(69.2-85.2%)(61.0-72.9%)(29.8-52.4%)
NPV Estimate94.6% (265/280)95.2% (79/83)94.0% (158/168)96.6% (28/29)
95% Confidence Interval(91.4-96.7%)(88.3-98.1%)(89.4-96.7%)(82.8-99.4%)
PPV Estimate52.1% (152/292)53.2% (25/47)51.3% (81/158)52.9% (46/87)
95% Confidence Interval(46.3-57.7%)(39.2-66.7%)(43.5-58.9%)(42.5-63.0%)

Page 16

Access BNP II 12 of 12
Confidential - Company Proprietary

Substantial Equivalence Comparison Conclusion

Beckman Coulter's Access BNP II Assay on the DxI 9000 Access Immunoassay Analyzer is substantially equivalent to the Access BNP Assay on the Access 2 Immunoassay System as demonstrated through the information and data provided in this submission. The performance testing presented in this submission provides evidence that the device is safe and effective in its intended use.

§ 862.1117 B-type natriuretic peptide test system.

(a)
Identification. The B-type natriuretic peptide (BNP) test system is an in vitro diagnostic device intended to measure BNP in whole blood and plasma. Measurements of BNP are used as an aid in the diagnosis of patients with congestive heart failure.(b)
Classification. Class II (special controls). The special control is “Class II Special Control Guidance Document for B-Type Natriuretic Peptide Premarket Notifications; Final Guidance for Industry and FDA Reviewers.”