Search Filters

Search Results

Found 25 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ZeniCore™ Automatic Biopsy Instrument and optional Co-Axial Introducer Needle is a single-use device indicated for obtaining core samples of soft tissue masses via percutaneous needle biopsy. It is not intended for use in bone.

    Device Description

    The ZeniCore™ Automatic Biopsy Instrument and optional, coaxial Introducer needle is a single-use medical device intended for obtaining core soft tissue samples via percutaneous needle biopsy. It is not intended for use in bone lesions. The ZeniCore™ Automatic Biopsy Instrument is a sterile, single patient use device comprised of a spring-loaded biopsy needle fitted into a plastic handle permitting single handed specimen collection than can be used with an optional corresponding coaxial Introducer Needle. The position of the needle may be visualized by imaging techniques such as X-ray, fluoroscopy, and ultrasound. The needle has an echogenic tip with numerically ordered centimeter markings to facilitate precise depth placement. The throw length of the needle when collecting a specimen is 25mm. There are five available needle gauge (G) sizes (12G, 14G, 16G, 18G and 20G) which are identified by a color-coded actuator button. The device is sold either standalone, or as a kit, with an optional, corresponding coaxial needle (cleared under K980004).

    The optional Co-Axial Introducer Needle system includes an outer cannula and an inner stylet which terminate in a trocar tip and serves to support acquisition of tissue samples. Each Co-Axial Introducer Needle has an adjustable spring clip that serves as a locking mechanism to secure the position of the outer co-axial needle relative to the skin which ensures the biopsy needle does not accidentally slide forward or backward during the procedure, thereby maintaining the accuracy and safety of the biopsy. The Luer cap of the stylet is color coded to match the corresponding biopsy instrument it is sized to fit.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K260028

    Validate with FDA (Live)

    Date Cleared
    2026-02-05

    (31 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CLEANER™ Vac Thrombectomy System is indicated for the removal of fresh, soft thrombi and emboli from the vessels of the peripheral venous vasculature, and for the infusion of physician-specified fluids, including thrombolytics. The CLEANER™ Vac Thrombectomy System is not intended for use in the pulmonary vasculature for treating pulmonary embolism.

    The CLEANER™ Vac Aspiration Canister is indicated as a vacuum source for the CLEANER™ Vac Thrombectomy System.

    Device Description

    The CLEANER™ Vac Thrombectomy System is a single use device intended for the removal of fresh, soft emboli and thrombi and for the infusion of physician selected fluids through the side-port of the aspiration catheter.

    The disposable system consists of:

    • The CLEANER™ Vac Aspiration Catheter with Handpiece, and
    • The CLEANER™ Vac Aspiration Canister (external vacuum reservoir with pump).

    The Aspiration Catheter with handpiece includes a Dilator and may be placed over-the-wire to navigate the device to the target site. Once in the target site, to complete the system, the Aspiration Canister is connected to the handpiece. The dilator is removed, and the device is activated by the user to aspirate soft emboli and thrombi. The clot is aspirated from the distal portion of the device through the handpiece and then collected in the aspiration canister reservoir.

    Additionally, the device may be used for infusion of thrombolytics and/or contrast media. Once thrombus resolution is achieved, the device is removed from the patient and discarded.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251318

    Validate with FDA (Live)

    Date Cleared
    2026-01-20

    (266 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VariFuse Adjustable Infusion Catheter is indicated for the administration of fluids, including thrombolytic agents and contrast media, into the peripheral and pulmonary artery vasculature.

    Device Description

    The VariFuse™ Adjustable disposable Infusion Catheter System consists of:

    • An adjustable infusion catheter is comprised of the outer and inner catheters pre-assembled coaxially and inseparable, set at the maximum infusion length (i.e. 50 cm) with a Touhy Borst Adapter for shipping.
      • The inner infusion catheter is designed with infusion holes spaced unevenly along the infusion length of 50 cm, a valve at the distal tip, and a radiopaque marker at the distal tip.
      • The outer infusion catheter is designed to set the infusion length by covering the infusion holes and contains a radiopaque marker at the distal tip.
    • A 1cc Luer Lock Syringe is included in the packaging for flushing
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K253741

    Validate with FDA (Live)

    Date Cleared
    2025-12-15

    (21 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The V•Stick™ Vascular Access Set is intended for use in the introduction and placement of guidewires and/or catheters.

    Device Description

    The V•Stick™ Vascular Access Set consists of a 4F or 5F coaxial introducer set (with a 3F standard or 3F stiff dilator), a 21ga entry needle (with an echogenic or non-echogenic tip), and a 0.018" x 40cm Nitinol core and with Stainless Steel coil vascular access guidewire with an articulatable tip.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K242612

    Validate with FDA (Live)

    Date Cleared
    2024-09-10

    (7 days)

    Product Code
    Regulation Number
    870.3375
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Option™ELITE Filter is indicated for the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the inferior vena cava (IVC) in the following conditions:

    · Pulmonary thromboembolism when anticoagulants are contraindicated

    • Failure of anticoagulant therapy in thromboembolic disease

    · Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced

    · Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated

    The Option™ELITE Filter may be removed according to the instructions supplied in the Section IX, entitled "Optional Procedure for Filter Retrieval" in patients who no longer require a filter can only be performed by the jugular approach.

    The Option™ELITE Vena Cava Filter 100cm System is indicated for the prevention of recurrent pulmonary embolism (PE) via placement in the vena cava in the following conditions:

    · Pulmonary thromboembolism when anticoagulants are contraindicated

    · Failure of anticoagulant therapy in thromboembolic disease

    · Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced

    · Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated

    The Option™ELITE Filter may be removed according to the instructions supplied in the Section IX, entional Procedure for Filter Retrieval" in patients who no longer require a filter in patients who no longer require a filter. Retrieval of the filter can only be performed by the jugular approach.

    Device Description

    The Option™ELITE Vena Cava Filter is designed for the prevention of recurrent pulmonary embolism via percutaneous delivery in the inferior vena cava (IVC).

    The Option™ELITE Vena Cava Filter 100cm System is designed for IVC filter insertion, delivery, deployment and placement via the popliteal and antecubital approach.

    The self-centering Option™ELITE Filter is laser cut from nickel - titanium alloy (Nitinol) tubing. The Option™ELITE Filter consists of shape memory Nitinol struts emanating from a central location and is designed for optimal clot capture. Retention anchors (retention hooks) are located at the caudal portion of the filter. These anchors are intended for filter fixation to the vessel wall. The Option™ELITE Filter is intended to be used in caval diameters up to 30mm. A retrieval hook is centrally located at the cranial extremity.

    The constrained Option™ELITE Filter is flexible and expands to the internal diameter of the IVC upon deployment. The Option™ELITE Filter imparts an outward radial force on the luminal surface of the vena cava to ensure proper positioning and stability. The Option™ ELITE Filter is designed to prevent pulmonary embolism while maintaining caval patency through central filtration.

    The introduction kit is comprised of a filter housed in a filter cartridge. Catheter Sheath Introducer (5F ID). Angiographic Vessel Dilator with an open end, and a Pusher with deployment marker.

    The Angiographic Vessel Dilator has side holes and 2 radiopaque markers, separated by 32mm (between the marker bands), that provide linear measurement of the inferior vena cava and assists in angiographic visualization when radiopaque contrast is delivered. The pusher advances the filter through the Catheter Sheath Introducer up to the deployment marker, and is then used to fix the filter in place during uncovering. The location of the distal end of the Catheter Sheath Introducer can be controlled by rotating the entire device to position the Catheter Sheath Introducer in the center of the vena cava. The Filter Cartridge houses the Option™ELITE Filter. The body of the Cartridge has text and colored arrows printed on it that identify assembly orientation, femoral is printed in green and jugular is printed in blue. The arrow of the desired access site will point into the Catheter Sheath Introducer hub. The Angiographic Vessel Dilator is designed to provide angiographic visualization and linear measurement of the vasculature when used in conjunction with the delivery of radiopaque contrast media to the vena cava.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the Option™ELITE Vena Cava Filter System and Option™ELITE Vena Cava Filter 100cm System. The purpose of the submission is to incorporate results from the PRESERVE study summary data into the Instructions for Use (IFU).

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not include a specific table of acceptance criteria or reported device performance metrics in the traditional sense of a performance study. This submission is for incorporating existing clinical study results (PRESERVE study) into the device's labeling, rather than presenting new performance data for a modified device. The document explicitly states:

    "No new operating principles have been introduced with the subject device. The subject device operates using the identical fundamental scientific technology as the predicate device; therefore, no performance testing was necessary nor was any performed. The clinical testing was previously provided to FDA under the Investigational Device Exemption and the PRESERVE study."

    Because no new performance testing was conducted for this specific 510(k) submission, there are no new device performance metrics to report against an acceptance criterion in this document. The existing performance would be based on the prior clearance of the predicate devices and the PRESERVE study.

    2. Sample Size Used for the Test Set and Data Provenance

    The document refers to the "PRESERVE study." However, it does not provide details on the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective) for this study within the provided text. To obtain this information, one would need to consult the original PRESERVE study documentation.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The provided document does not contain information on the number of experts used to establish ground truth or their qualifications for the PRESERVE study's test set.

    4. Adjudication Method for the Test Set

    The provided document does not specify any adjudication method (e.g., 2+1, 3+1, none) for the test set of the PRESERVE study.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC comparative effectiveness study is typically used for AI-assisted diagnostic devices to assess the impact of AI on human reader performance. Since the device is a Vena Cava Filter System, it is a physical medical device, not a diagnostic imaging AI algorithm. Therefore, no MRMC comparative effectiveness study was done for this device.

    6. Standalone (Algorithm Only) Performance

    The device is a physical Vena Cava Filter System, not an algorithm or software-only device. Therefore, the concept of "standalone performance" for an algorithm does not apply to this device.

    7. Type of Ground Truth Used

    Given that this is a physical medical device (vena cava filter) and the PRESERVE study would be a clinical trial, the "ground truth" would likely be derived from patient outcomes, imaging data, and clinical assessments related to the filter's effectiveness in preventing pulmonary embolism and its safety profile. However, the document does not explicitly state the type of ground truth used for the PRESERVE study.

    8. Sample Size for the Training Set

    The concept of a "training set" typically applies to machine learning algorithms. Since the device is a physical medical device and not an AI/ML product, no training set was used in the context of algorithm development. Clinical trials, like PRESERVE, gather data for validation and effectiveness, not for training an algorithm.

    9. How Ground Truth for the Training Set Was Established

    As noted above, no training set was used in the context of an algorithm for this physical device. Therefore, how ground truth for a training set was established is not applicable.

    In summary, the provided FDA 510(k) clearance letter and summary primarily focus on demonstrating substantial equivalence to a predicate device and incorporating pre-existing clinical study (PRESERVE study) data into the device's labeling. It does not present new performance data or detail the methodology of the PRESERVE study in the way one might expect for a new device's initial clearance or an AI/ML product. To get specific details about the PRESERVE study (sample size, ground truth, expert involvement, etc.), one would need to refer to its full study report or publications.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241145

    Validate with FDA (Live)

    Date Cleared
    2024-08-01

    (98 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TLAB® Transvenous Liver Biopsy System is intended to be used for percutaneous transjugular and transfemoral venous liver access during diagnostic and interventional procedures.

    Device Description

    The TLAB® Transvenous Liver Biopsy System is a single use, disposable, sterile device. The TLAB® Transvenous Liver Biopsy System consists of the following components: 18-Gauge (18Ga) Flexcore Biopsy Needle, 7 French (7Fr) Introducer Sheath with a curved metal stiffener, 5 French (5Fr) Straight Catheter, 5 French (5Fr) Curved Catheter, Tissue Removal Swabs, Bending Tool (used for transfemoral access).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the TLAB® Transvenous Liver Biopsy System, based on the provided text:

    This document describes a medical device (TLAB® Transvenous Liver Biopsy System) not an AI/ML device, therefore, sections related to AI/ML specific criteria (such as MRMC studies, standalone algorithm performance, training set details) are not applicable.


    Acceptance Criteria and Reported Device Performance

    Device Trade Name: TLAB® Transvenous Liver Biopsy System (TF-18C)
    Regulation Number: 21 CFR 870.1340 - Catheter Introducer
    Regulatory Class: Class II
    Product Code: DYB

    The device demonstrated substantial equivalence to predicate devices (TLAB® Transjugular Liver Access Set K022634 and Traveler Portal Vein Access Set K213638) based on a comparison of technological characteristics, indications for use, materials, packaging, principle of operation, design features, sterilization process, and in-vitro performance testing.

    Acceptance Criteria CategorySpecific Criteria / Test PerformedReported Device Performance / Outcome
    Performance TestingDimensionalMet acceptance criteria
    VisualMet acceptance criteria
    RadiopacityMet acceptance criteria
    Corrosion ResistanceMet acceptance criteria
    Simulated UseMet acceptance criteria
    Tensile StrengthMet acceptance criteria
    Torque StrengthMet acceptance criteria
    Liquid LeakageMet acceptance criteria
    Luer Functional Testing (ISO 80369-7)Met acceptance criteria
    Catheter Functional Testing (ISO 10555-1)Met acceptance criteria
    Resistance to Breakage (ISO 9626)Met acceptance criteria
    Shipping Qualification TestingMet acceptance criteria
    Design ValidationMet acceptance criteria
    Summative UsabilityMet acceptance criteria
    Biocompatibility (ISO 10993-1:2020)Cytotoxicity (ISO 10993-5)Previously performed, applicable to new device; met criteria
    Sensitization (ISO 10993-23)Previously performed, applicable to new device; met criteria
    Irritation or Intracutaneous Reactivity (ISO 10993-10)Previously performed, applicable to new device; met criteria
    Material Mediated Pyrogen (ISO 10993-11)Previously performed, applicable to new device; met criteria
    Acute Systemic Toxicity (ISO 10993-11)Previously performed, applicable to new device; met criteria
    Hemocompatibility (ISO 10993-4): ASTM Hemolysis Assay, Direct and Extract Methods (ISO)Previously performed, applicable to new device; met criteria
    Hemocompatibility (ISO 10993-4): Complement Activation Assay, SC5b-9 Method with Comparison Article (ISO)Performed anew; met criteria
    Hemocompatibility (ISO 10993-4): Partial Thromboplastin Time (PTT) Assay with Comparison Article (ISO)Performed anew; met criteria
    Hemocompatibility (ISO 10993-4): Heparinized Platelet and Leukocyte Count Assay with Comparison Article (ISO)Performed anew; met criteria
    SterilizationMinimum SAL 10-6, EtOMet acceptance criteria
    Shelf-Life3 yearsMet acceptance criteria (based on predicate)

    Study Details

    1. Sample size used for the test set and the data provenance:

      • The document primarily describes bench-top testing (in-vitro performance) and biocompatibility testing. Specific sample sizes for each non-clinical test are not explicitly stated in the provided text, but it mentions that "A series of testing was conducted".
      • Data Provenance: All studies were performed following approved protocols under Good Laboratory Practices (GLP) in compliance with FDA GLP, 21 CFR Part 58. This indicates the testing was conducted in a controlled laboratory environment. No specific country of origin or retrospective/prospective nature for these non-clinical tests is given beyond GLP compliance.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This question is not applicable as the described studies are non-clinical, bench-top performance and biocompatibility tests. They do not involve human interpretation or subjective clinical "ground truth" that would require expert consensus. The acceptance criteria themselves serve as the 'ground truth' for these engineering and materials tests.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This question is not applicable as the described studies are non-clinical, bench-top performance and biocompatibility tests. Adjudication methods are typically used in clinical studies or studies involving subjective human interpretation (e.g., image reading) where disagreement among experts needs resolution.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • This question is not applicable as this document describes a conventional medical device (a biopsy system), not an AI/ML powered device. Therefore, no MRMC comparative effectiveness study was performed in the context of AI assistance.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • This question is not applicable as this document describes a conventional medical device (a biopsy system), not an AI/ML powered device. There is no algorithm to test in a standalone manner.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For Performance Testing: The "ground truth" is defined by the acceptance criteria established in accordance with protocols based on guidance and industry standards. These include established engineering and material science standards (e.g., ISO for luer, catheter, and breakage testing) that define acceptable performance.
      • For Biocompatibility Testing: The "ground truth" is defined by the internationally recognized standards for biocompatibility (ISO 10993 series) and FDA Guidance, which specify test methods and acceptable limits for biological responses.
    7. The sample size for the training set:

      • This question is not applicable as this document describes a conventional medical device (a biopsy system), not an AI/ML powered device. There is no machine learning model that requires a training set.
    8. How the ground truth for the training set was established:

      • This question is not applicable as this document describes a conventional medical device (a biopsy system), not an AI/ML powered device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K233909

    Validate with FDA (Live)

    Date Cleared
    2024-03-28

    (107 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cleaner™ Plus 18F Thrombectomy System is indicated for mechanical de-clotting, aspiration, and controlled and selective infusion of physician-specified fluids, including thrombolytics, in the peripheral venous vasculature.

    Device Description

    The Cleaner Plus™ 18F Thrombectomy System is a single use device used to provide thrombectomy in the peripheral venous vasculature. The device provides additional features, such as aspiration and over-the-wire device placement.

    The disposable system consists of: (1) the Aspiration Catheter & Dilator, (2) the Handpiece that includes an aspiration control and an integrated Maceration Wire, and a Peel-Away Introducer and (3) the external vacuum reservoir with pump known as the Aspiration Canister.

    The system is inserted percutaneously into the vessel using an introduction catheter, the system Aspiration Catheter and Dilator may be placed over-the-wire to the site of thrombus. The device macerates intra-lumen and wall adherent thrombus with a maceration wire. The macerated thrombus is removed from the vessel using an aspiration system. The aspiration of the clot can be performed simultaneous or independently.

    Additionally, the device may be used for infusion of thrombolytics and/or contrast media. Once thrombus resolution is achieved, the device is removed from the patient and discarded.

    AI/ML Overview

    The provided text describes information about the Cleaner™ Plus 18F Thrombectomy System, a medical device, and the basis for its FDA clearance through a 510(k) premarket notification (K233909). The information primarily focuses on establishing substantial equivalence to a predicate device (K211798), rather than detailing a study that proves the device meets specific acceptance criteria in the context of an AI/ML medical device.

    Therefore, many of the requested items related to AI/ML device study design (e.g., sample size for test sets, expert ground truth establishment, MRMC studies, standalone algorithm performance, training set details) are not applicable or cannot be extracted from this document, as it concerns a mechanical thrombectomy system, not an AI/ML-driven device.

    However, I can provide the available information regarding the device's performance assessment and the basis for its clearance.


    Acceptance Criteria and Device Performance (Non-AI/ML Medical Device)

    The document does not present a table of numerical "acceptance criteria" in the same way one might find for an AI/ML model's performance metrics (e.g., AUC, sensitivity, specificity thresholds). Instead, the acceptance criteria for this medical device are qualitative, based on established industry standards, guidance, and the successful outcome of various performance and safety tests demonstrating substantial equivalence to a predicate device.

    The "reported device performance" refers to the successful completion and meeting of predefined parameters for each test, indicating the device performs as intended and is as safe and effective as the predicate.

    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a mechanical device, not an AI/ML system, the "acceptance criteria" are the successful completion of various tests and the demonstration of substantial equivalence. The "reported device performance" refers to the outcomes of these tests.

    Test Category / CharacteristicAcceptance Criteria (Qualitative)Reported Device Performance (Outcome)
    Substantial EquivalenceDemonstrates comparable indications for use, mechanism of action, materials, technological characteristics, principle of operation, design features, and sterilization process to the predicate device, introducing no additional clinical risk.Established based on in vitro performance testing, similarities in indications for use, materials, technological characteristics, principle of operation, design features, and sterilization process with the predicate device. Confirmed by in-vitro and in-vivo tests. Conclusion: Cleaner Plus 18F Thrombectomy System is substantially equivalent to the predicate device and introduces no additional clinical risk.
    In-Vitro Performance TestingMeet acceptance criteria based on requirements outlined in guidance and industry standards for each specific test.Cleaner Plus 18F Thrombectomy System Performance: Visual, Dimensional, Simulated Use, Leak Testing.Handpiece Performance: Aspiration and Maceration Performance, Tensile and Torque, Kink Testing, Catheter Torque.Canister Performance: Simulated Use Maceration and Aspiration Performance.Maceration Wire Performance: Bond Strength, Resistance to Corrosion, Shape Retention, Flexing and Fracture, Kink Radius, Fatigue, Endurance, Torque, Tensile, Dimensional, Visual.Handpiece and Catheter Performance: Peel-Away Introducer Removal Force, Maceration Wire RPM, Wire to Coupler Tensile, Peel-Away Introducer Functionality, Tip Collapse.Other Tests: Particulates, Canister Performance (Visual, Functional, Tensile - joint strength), Luer Compliance, Pump Performance, IEC Compliance, Software Testing, Shipping Qualification. All tests met acceptance criteria.
    BiocompatibilityCompliant with ISO 10993-1:2020 for an external communicating, circulating blood device with limited duration (<24hrs).A Biological Risk Assessment was performed. Biocompatibility testing was appropriately leveraged from the predicate device (K211798) and a comparator device (K232679), performed following approved protocols under GLP and in compliance with FDA GLP, 21 CFR Part 58.
    Animal Studies (In-vivo Safety & Performance)Demonstrate substantially equivalent safety and performance outcomes, with no new questions of safety or effectiveness. Show no histological signs of thromboembolism, ruptures, hemorrhage, perforation, or dissection in treated vessels.First animal study: Assessed safety and performance on vascular endothelium (vessel patency, clot burden, pulmonary embolism). No histological signs of thromboembolism, ruptures, hemorrhage, perforation, or dissection noted. All endpoints met.Second animal study: Assessed safety and performance of the aspiration aspect (vessel injury, patency, gross thrombogenicity). No histological signs of thromboembolism, ruptures, hemorrhage, perforation, or dissection noted. All endpoints met.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated numerically as a "test set" in the context of an AI/ML model for classification or prediction. The "test set" here refers to the actual physical devices (or components) used for the various in-vitro and in-vivo performance and safety tests. The specific number of devices/samples tested for each individual performance test (e.g., tensile strength, kink testing) is not provided in this summary.
    • Data Provenance:
      • In-vitro Performance Testing: Conducted by the manufacturer (Argon Medical Devices, Inc.)
      • Biocompatibility Testing: Leveraging data from predicate device (K211798) and a comparator device (K232679). These studies were performed under Good Laboratory Practices (GLP) in compliance with FDA GLP, 21 CFR Part 58.
      • Animal Studies: Conducted as part of the submission to demonstrate safety and performance. (Specific country/institution not mentioned, implied to be part of the submission package leveraged by the manufacturer).
      • Retrospective or Prospective: All tests described are prospective, conducted specifically for this submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not applicable for this type of device. The "ground truth" for a mechanical device like this is established through engineering and scientific measurement, and observable biological/physical outcomes, rather than expert interpretation of images or clinical data. Device performance is measured against established physical and safety parameters.

    4. Adjudication Method for the Test Set

    • Not applicable. Adjudication methods (e.g., 2+1, 3+1) are common in scenarios where human interpretation of complex data (like medical images) is being used to establish ground truth or evaluate diagnostic performance. For a mechanical device, performance is evaluated against objective, measurable criteria and laboratory test results.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    • No. This type of study is relevant for AI/ML devices where reader performance (e.g., radiologists, pathologists) is being evaluated with and without AI assistance. This document describes a mechanical device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not applicable. This relates to AI/ML algorithm performance.

    7. The Type of Ground Truth Used

    • Objective Engineering & Biological Measurements: The ground truth for this device is based on:
      • Physical/Dimensional conformity: Verified dimensions and visual appearance.
      • Functional performance: Successful operation during simulated use, aspiration and maceration efficiency, tensile strength, torque, kink resistance, etc.
      • Biocompatibility: Established through ISO standards and GLP-compliant testing results (e.g., absence of cytotoxicity, sensitization, irritation).
      • Biological safety and effectiveness (animal studies): Observable clinical and histological outcomes in vivo (e.g., vessel patency, absence of thromboembolism, rupture, hemorrhage).
      • Compliance with industry standards: Meeting parameters set forth in standards like IEC 60601.

    8. The Sample Size for the Training Set

    • Not applicable. This document describes a mechanical device, not an AI/ML model that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there is no AI/ML training set, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K232679

    Validate with FDA (Live)

    Date Cleared
    2024-01-19

    (140 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cleaner™ Pro Thrombectomy System is indicated for the removal of fresh, soft thrombi and emboli from the vessels of the peripheral venous vasculature, and for the infusion of physician-specified fluids, including thrombolytics. The Cleaner™ Pro Thrombectomy System is not intended for use in the pulmonary vasculature for treating of pulmonary embolism.

    Device Description

    The Cleaner Pro Thrombectomy System is a single use device in the removal of fresh, soft emboli and thrombi and for the infusion of physician-selected fluids through the side-port of the aspiration catheter.

    The disposable system consists of: (1) the Aspiration Catheter with Dilator, (2) the Handpiece with Flushing Adapter, and (3) the external vacuum reservoir with pump known as the Aspiration Canister.

    The Aspiration Catheter with Dilator may be placed over-the-wire to navigate the device to the target site. Once in the target site, to complete the system, the Aspiration Canister is connected to the handpiece. The dilator is removed, and the device is activated by the user to aspirate soft emboli and thrombi. The clot is aspirated from the distal portion of the device through the handpiece and then collected in the aspiration canister reservoir.

    Additionally, the device may be used for infusion of thrombolytics and/or contrast media. Once thrombus resolution is achieved, the device is removed from the patient and discarded.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Cleaner™ Pro Thrombectomy System. This document focuses on demonstrating substantial equivalence to previously cleared predicate devices, rather than proving the device meets specific acceptance criteria based on clinical outcomes or standalone AI algorithm performance.

    Therefore, many of the requested details about acceptance criteria, test sets, ground truth establishment, expert adjudication, MRMC studies, and training sets are not present in this type of regulatory submission. This document describes the bench-top and animal performance testing to support the claim of substantial equivalence for a medical device, which is different from studies demonstrating efficacy of an AI algorithm or performance directly tied to a specific set of clinical acceptance criteria.

    However, I can extract the information that is available:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not present a table of specific acceptance criteria with quantitative performance metrics for a clinical outcome study. Instead, it lists performance testing conducted to demonstrate substantial equivalence to the predicate devices. The acceptance criteria for these tests are generally that the device performs comparably to the predicate or meets established engineering/design specifications for safety and functionality.

    Test CategorySpecific Tests Performed (examples)Acceptance Criteria (Implied)Reported Performance (Implied)
    In-Vitro (Bench-top) Performance TestingCatheter - Dimensional, Visual, Radiopacity, Hemostasis Valve Leak, System Leak, Burst, Kink Radius, Tensile Break, Torsional Break, Aspiration Tip Collapse. Dilator - Dimensional, Leak, Burst, Functional, Visual, Tensile Break, Radiopacity. Aspiration Canister - Visual, Functional, Tubing Tensile Break, Weld Strength. Handpiece - Functional, Simulate Use. System - Functional, Vacuum Decay, Noise Level, Integrity, Visual, Simulated Use. Shipping Qualification, IEC-60601 Compliance, Software Validation.The device must meet predefined engineering specifications and perform comparably to predicate devices for these physical and functional characteristics, ensuring safety and intended function."The below were shown to meet the acceptance criteria that were determined to demonstrate substantial equivalence." (General statement that tests were successful)
    Biocompatibility TestingCytotoxicity (ISO 10993-5), Sensitization (ISO 10993-10), Irritation or Intracutaneous Reactivity (ISO 10993-10), Material Mediated Pyrogen (ISO 10993-11), Acute Systemic Toxicity (ISO 10993-11), Hemocompatibility (ISO10993-4) - In-vitro Blood Assay Complement Activation, SC5b-9 Heparinized Platelet and Leucocyte Counts, Partial Thromboplastin Time (PTT), ASTM Hemolysis Assay (Direct and Extract Methods).Compliance with ISO 10993 standards for the specified biological endpoints, ensuring the device is biocompatible for its intended use."Biocompatibility is established for the Cleaner Pro Thrombectomy System according to ISO 10993-1:2020 as an external communicating, circulating blood with limited duration <24hrs. All studies were performed following the approved protocol under Good Laboratory Practices (GLP) in compliance to FDA GLP, 21 CFR Part 58." (Overall claim of meeting biocompatibility requirements)
    Animal StudyEvaluation for vessel injury, patency, and gross thrombogenicity in the target vasculature.Demonstrating substantially equivalent safety and performance, specifically no vessel injury (ruptures, hemorrhage, perforation, dissection) or histological signs of thromboembolism."There were no histological signs of thromboembolism noted in the treated vessels, there were no ruptures, hemorrhage, perforation, or dissection noted histologically in the treated veins. All the end points of this study were met."

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set (Animal Study): The document states "An animal study was performed..." but does not specify the sample size (number of animals or procedures).
    • Data Provenance: The animal study data provenance is "animal study," implying a prospective, controlled study within a laboratory setting. No country of origin for the animals or study location is explicitly stated, but it would typically be conducted under US regulations (GLP).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This document does not describe the use of human experts to establish ground truth for a test set in the context of an AI algorithm or image analysis. The animal study's "ground truth" would be established by veterinary pathologists or other trained personnel performing histological analysis and gross observation, but the number and specific qualifications are not mentioned.

    4. Adjudication Method for the Test Set

    Not applicable, as the document doesn't describe ground truth establishment by multiple human readers for a test set in the context of an AI/image analysis setting. The animal study results would likely be adjudicated if multiple observers interpreted the findings, but no such method is detailed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not conducted or described. This type of study is typically done to assess the impact of an AI system on human reader performance, which is not the focus of this 510(k) submission for a mechanical thrombectomy device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Not applicable. This device is a mechanical thrombectomy system, not an AI algorithm. Its performance is inherent in its physical and mechanical operation.

    7. The Type of Ground Truth Used

    For the animal study, the ground truth was based on post-procedural histological analysis of treated vessels and gross observation for signs of injury or thrombosis. For bench-top testing, the ground truth is against engineering specifications and physical measurements.

    8. The Sample Size for the Training Set

    Not applicable. This is a mechanical device, not an AI algorithm, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As above, there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233432

    Validate with FDA (Live)

    Date Cleared
    2023-12-08

    (57 days)

    Product Code
    Regulation Number
    870.1340
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 10F Sheath and Dilator Set is indicated for introduction of therapeutic devices into the vasculature, excluding coronary and neuro vasculature.

    Device Description

    The 10F Sheath and Dilator Set is designed for single use and intended to introduce therapeutic or diagnostic devices into the vasculature, excluding coronary and neuro vasculature. The 10F Sheath and Dilator Set consists of a 10F reinforced Introducer Sheath and a matching 10F radiopaque Dilator.

    The 10F Introducer sheath is a coil reinforced sheath with a Platinum Iridium radiopaque marker band. The sheath has a hemostasis valve and a side port with stopcock. The 10F dilator is made of radiopaque, non-reinforced material with an atraumatic tip to minimize blood loss. The device does not include any coating.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for the "10F Sheath and Dilator Set" and establishes its substantial equivalence to a predicate device based on in vitro performance testing and technological characteristics. It does not include information about an AI-powered device or a study proving that an AI device meets acceptance criteria. Thus, I cannot populate the requested table or answer the questions related to AI device performance.

    However, I can extract the acceptance criteria and the study performed for the 10F Sheath and Dilator Set (a non-AI medical device).

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Radiopacity(Implicit performance standard)Met acceptance criteria demonstrating substantial equivalence.
    Visual(Implicit performance standard)Met acceptance criteria demonstrating substantial equivalence.
    Dimensional(Implicit performance standard)Met acceptance criteria demonstrating substantial equivalence.
    Simulative Use(Implicit performance standard)Met acceptance criteria demonstrating substantial equivalence.
    Leakage (Sheath and Dilator)(Implicit performance standard)Met acceptance criteria demonstrating substantial equivalence.
    Kink Test (System)(Implicit performance standard)Met acceptance criteria demonstrating substantial equivalence.
    Tensile (Sheath and Dilator)(Implicit performance standard)Met acceptance criteria demonstrating substantial equivalence.
    Luer testing(Implicit performance standard)Met acceptance criteria demonstrating substantial equivalence.
    Shipping Qualification(Implicit performance standard)Met acceptance criteria demonstrating substantial equivalence.
    BiocompatibilityAccording to ISO 10993-1:2020 for an external communicating, circulating blood device with limited duration <24hrs.
        - Cytotoxicity(Implicit standard per ISO 10993-5)Performed following approved protocols under GLP, in compliance with FDA GLP, 21 CFR Part 58.
        - Sensitization(Implicit standard per ISO 10993-10)Performed following approved protocols under GLP, in compliance with FDA GLP, 21 CFR Part 58.
        - Irritation or Intracutaneous Reactivity(Implicit standard per ISO 10993-23)Performed following approved protocols under GLP, in compliance with FDA GLP, 21 CFR Part 58.
        - Material Mediated Pyrogenicity(Implicit standard per ISO 10993-11)Performed following approved protocols under GLP, in compliance with FDA GLP, 21 CFR Part 58.
        - Acute Systemic Toxicity(Implicit standard per ISO 10993-11)Performed following approved protocols under GLP, in compliance with FDA GLP, 21 CFR Part 58.
        - Hemocompatibility (In-vitro Blood Loop Assay)(Implicit standard per ISO 10993-4)Performed with comparison article following approved protocols under GLP, in compliance with FDA GLP, 21 CFR Part 58.
        - Hemocompatibility (Complement Activation Assay, SC5b-9 Method)(Implicit standard per ISO 10993-4)Performed with comparison article following approved protocols under GLP, in compliance with FDA GLP, 21 CFR Part 58.
        - Hemocompatibility (Heparinized Platelet and Leukocyte Count Assay)(Implicit standard per ISO 10993-4)Performed with comparison article following approved protocols under GLP, in compliance with FDA GLP, 21 CFR Part 58.
        - Hemocompatibility (Partial Thromboplastin Time (PTT) Assay)(Implicit standard per ISO 10993-4)Performed with comparison article following approved protocols under GLP, in compliance with FDA GLP, 21 CFR Part 58.
        - Hemocompatibility (ASTM Hemolysis Assay, Direct and Extract Methods)(Implicit standard per ISO 10993-4)Performed following approved protocols under GLP, in compliance with FDA GLP, 21 CFR Part 58.

    Study Information for 10F Sheath and Dilator Set:

    Since this document pertains to a traditional medical device (sheath and dilator set) and not an AI/software device, the following points regarding AI studies are not applicable.

    1. Sample size used for the test set and the data provenance: Not applicable to this type of device. The tests mentioned are bench-top (in vitro) and biocompatibility tests on device components. No "test set" in the context of diagnostic performance, images, or patient data is described.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for diagnostic purposes is not involved.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for the device's performance is established by meeting predefined engineering specifications, industry standards (e.g., ISO 10993), and regulatory requirements for physical and biological properties.
    7. The sample size for the training set: Not applicable. This device does not have a "training set" in the context of machine learning.
    8. How the ground truth for the training set was established: Not applicable.

    Summary of the study conducted for the 10F Sheath and Dilator Set:

    The study involved a series of non-clinical (bench-top) performance tests and biocompatibility tests. The non-clinical tests included:

    • Radiopacity
    • Visual inspection
    • Dimensional analysis
    • Simulative Use
    • Leakage (Sheath and Dilator)
    • Kink Test (System)
    • Tensile (Sheath and Dilator)
    • Luer testing
    • Shipping Qualification

    The biocompatibility testing was conducted according to ISO 10993-1:2020 for an external communicating, circulating blood device with limited duration (<24 hours). These tests included Cytotoxicity, Sensitization, Irritation or Intracutaneous Reactivity, Material Mediated Pyrogenicity, Acute Systemic Toxicity, and Hemocompatibility assays (In-vitro Blood Loop, Complement Activation, Heparinized Platelet and Leukocyte Count, Partial Thromboplastin Time (PTT), and ASTM Hemolysis). All biocompatibility studies were performed under Good Laboratory Practices (GLP) in compliance with FDA GLP, 21 CFR Part 58.

    The results of these tests demonstrated that the Argon Medical Devices, Inc. 10F Sheath and Dilator Set met the acceptance criteria and was substantially equivalent to the predicate device (Flexor® Check-Flo® Introducer Set) in terms of safety and performance for its intended use.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230669

    Validate with FDA (Live)

    Date Cleared
    2023-11-30

    (265 days)

    Product Code
    Regulation Number
    880.5970
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The L-Cath Single and Dual Lumen Catheters and the L-Cath Midline Catheters are indicated for short or long term peripheral access to the central venous system for the administrations, and nutrients; the sampling of blood; and monitoring blood pressure and temperature intravenously.

    Device Description

    The L-Cath™ Single and Dual Lumen and L-Cath™ Midline Catheters are sterile, single use devices, sold as standalone devices which includes the following accessories:
    • Trim Tool
    • Tape measure
    • Stylet (include in the final catheter assembly as applicable)
    • Polyurethane catheter
    These catheters are also sold as a component in a Basic Kit (catheter and introducer packaged together) which includes the following:

    • BD Introsyte Autoguard Introducer or Splittable Introducer ●
    • Stylet (include in the final catheter assembly as applicable)
    • Trim Tool
    • Tourniquet
    • Tape measure
    AI/ML Overview

    The provided text describes a 510(k) premarket notification for Argon Medical Devices' L-Cath™ Single and Dual Lumen Catheters and L-Cath™ Midline Catheters. This document focuses on demonstrating substantial equivalence to an existing predicate device (BD L-Cath™ Midline Catheter, BD Cath™ Single Lumen, BD Cath™ Dual Lumen Peripherally Inserted Central Catheters PICC, K091670), rather than proving the device meets clinical acceptance criteria based on studies involving AI or human readers.

    Therefore, many of the requested criteria in your prompt (e.g., acceptance criteria for device performance in terms of AI metrics, sample sizes for test/training sets in an AI study, number of experts, adjudication methods, MRMC studies, standalone AI performance, ground truth establishment) are not applicable to this type of regulatory submission. This document details the engineering and biocompatibility aspects of the device to show it functions safely and effectively similarly to devices already on the market.

    However, I can extract the relevant information from the provided text regarding the performance testing conducted to support the substantial equivalence claim.

    Here's a breakdown of the requested information based on the provided text, highlighting what is applicable and what is not:


    Acceptance Criteria and Study to Prove Device Meets Acceptance Criteria

    This submission is a 510(k) for substantial equivalence, not a clinical study proving performance against AI-related metrics. The "acceptance criteria" here refer to engineering, biocompatibility, and sterilization performance standards, demonstrating the modified device is as safe and effective as its predicate.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states: "The subject device met all the predetermined acceptance criteria derived from the above listed reference standards and internal test protocols and demonstrated substantially equivalent performance compared to the predicate device." While specific numerical acceptance criteria (e.g., "pressure burst must exceed X psi") are not explicitly detailed in a table, the document lists the types of performance tests conducted and confirms successful completion.

    Acceptance Criteria Category (Implied)Performance Test Conducted (Reported)Device Performance (Reported)
    Mechanical/Physical PerformanceCatheter – Joint Leak TestMet acceptance criteria (Substantially equivalent to predicate)
    Catheter – DimensionalMet acceptance criteria (Substantially equivalent to predicate)
    Catheter – Pressure BurstMet acceptance criteria (Substantially equivalent to predicate)
    Catheter – DurometerMet acceptance criteria (Substantially equivalent to predicate)
    Catheter – RadiopacityMet acceptance criteria (Substantially equivalent to predicate)
    Stylet – Retaining forceMet acceptance criteria (Substantially equivalent to predicate)
    Luer Functional Mechanical Bonds – Pull forceMet acceptance criteria (Substantially equivalent to predicate)
    ParticulatesMet acceptance criteria (Substantially equivalent to predicate)
    Material DurabilityMet acceptance criteria (Substantially equivalent to predicate)
    Supplemental Testing for Design Change (Luer Lock material change)
    Luer Integrity (Functional) Report 2023-042-RPTMet acceptance criteria
    ISO 80369-7 ComplianceMet acceptance criteria
    Luer Integrity (Functional); Functional Leak Test Report 2023-134-RPTMet acceptance criteria
    BiocompatibilityPyrogenicity (ISO 10993-11)Met acceptance criteria (Substantially equivalent to predicate)
    Cytotoxicity (ISO 10993-5)Met acceptance criteria (Substantially equivalent to predicate)
    Hemocompatibility (ISO 10993-4)Met acceptance criteria (Substantially equivalent to predicate)
    Irritation/Sensitization (ISO 10993-10)Met acceptance criteria (Substantially equivalent to predicate)
    Systemic Toxicity (ISO 10993-11)Met acceptance criteria (Substantially equivalent to predicate)
    Ethylene oxide Residuals (ISO 10993-7)Met acceptance criteria (Substantially equivalent to predicate)
    Local effects after implantation (ISO 10993-7)Met acceptance criteria (Substantially equivalent to predicate)
    Supplemental Testing for Material Change (Luer lock material)
    Systemic Toxicity (10993-17) Report 2022-026-RPTMet acceptance criteria
    Genotoxicity (10993-3)Met acceptance criteria
    Sterilization100% EtOMet acceptance criteria (SAL 10-6)
    ISO 11135:2014 & AAMI TIR28:2016 complianceMet acceptance criteria

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not explicitly stated in terms of number of units tested for each criterion, but implied to be sufficient for demonstrating compliance with ISO standards and internal protocols.
    • Data Provenance: The tests were conducted internally by Argon Medical Devices, as indicated by the company providing the summary. The regulatory context implies these are laboratory tests, not clinical data from patients. They are likely prospective bench and lab tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. This is a medical device 510(k) submission based on engineering and biocompatibility testing, not an AI performance study requiring human expert ground truth for image or diagnostic interpretation. The "ground truth" here is established by validated test methods and reference standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. See point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is not an AI-assisted diagnostic device. Clinical testing (human studies) was explicitly stated as "not required for the determination of substantial equivalence."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. See point 5.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Engineering and Biocompatibility Test Standards: The "ground truth" for the device's performance is established by recognized international standards (e.g., ISO 10993 series for biocompatibility, ISO 80369-7 for luer connectors, ISO 11135 for sterilization) and internal validated test protocols designed to assess the device's physical properties and safety characteristics.

    8. The sample size for the training set

    • Not Applicable. This document describes a medical device, not an AI algorithm. There is no "training set."

    9. How the ground truth for the training set was established

    • Not Applicable. See point 8.

    Summary of the Study Proving Acceptance:

    The study presented in this 510(k) summary is a series of bench testing, laboratory analyses (biocompatibility), and sterilization validations designed to demonstrate that the modified L-Cath™ catheters are substantially equivalent to the previously cleared predicate devices. The primary change was the material used in the Luer Lock assembly.

    The "study" involved:

    • Performance Testing: Assessing mechanical characteristics like leak integrity, dimensions, pressure burst, durometer, radiopacity, retaining force, pull force, particulates, and material durability. These tests were stated to be the "SAME as K091670" (the predicate's testing) with additional supplemental testing specific to the Luer Lock material change (Luer Integrity, ISO 80369-7 Compliance).
    • Biological Comparison (Biocompatibility Testing): Evaluation of potential biological risks including pyrogenicity, cytotoxicity, hemocompatibility, irritation/sensitization, systemic toxicity, ethylene oxide residuals, and local effects after implantation. Supplemental testing (Systemic Toxicity, Genotoxicity) was performed due to the Luer Lock material change.
    • Sterilization Validation: Confirmation that the device can be consistently sterilized to meet a sterility assurance level (SAL) of 10-6, in accordance with ISO 11135:2014 and AAMI TIR28:2016.

    The conclusion is that the modified device "met all the predetermined acceptance criteria derived from the above listed reference standards and internal test protocols and demonstrated substantially equivalent performance compared to the predicate device." This means the tests confirmed the new device performs within acceptable limits, similar to the predicate, and does not raise new questions of safety or effectiveness.

    Ask a Question

    Ask a specific question about this device