Search Results
Found 67 results
510(k) Data Aggregation
(163 days)
38117
Re: K250037
Trade/Device Name: Incompass Total Ankle System
Regulation Number: 21 CFR 888.3110
Ankle System
Common Name: Total Ankle Prosthesis
Classification Name and Reference: 21 CFR 888.3110
Intended Use: The Incompass Total Ankle System is intended to give a patient limited mobility by reducing pain, restoring alignment and replacing the flexion and extension movement in the ankle joint.
Indications for Use: The Incompass Total Ankle is indicated for patients with ankle joints damaged by severe rheumatoid, post-traumatic, or degenerative arthritis. The Incompass Total Ankle is additionally indicated for patients with a failed previous ankle surgery.
Incompass™ Total Ankle System is a total ankle replacement system consisting of implants and instruments.
Incompass™ Total Ankle System will harmonize the Infinity™ and Inbone™ Total Ankle Systems into one comprehensive system consisting of modular stemmed and low-profile pegged tibial trays, tibial inserts, and talar domes. The harmonized system is facilitated by sharing the same tibial resection geometry between the pegged and the stemmed tibial tray designs.
The provided text is a 510(k) clearance letter for the Incompass Total Ankle System. It outlines the device's indications for use, its classification, and declares its substantial equivalence to predicate devices based on non-clinical evidence.
However, the clearance letter explicitly states: "Clinical testing was not necessary for the determination of substantial equivalence."
This means that the information requested regarding acceptance criteria, study details, sample sizes, expert involvement, and ground truth establishment (which are typical for studies proving a device meets acceptance criteria, especially for AI/ML-based devices or those requiring clinical trials) is not included in this document. The clearance was based on non-clinical performance bench testing to demonstrate substantial equivalence in wear, articular stability, lock detail, manufacturing processes, static strength, fatigue strength, fretting, corrosion, and MRI safety, rather than clinical efficacy studies.
Therefore, I cannot provide the requested information from this document. If this were a submission for a device that required clinical performance data to demonstrate safety and effectiveness (e.g., an AI/ML device), this section would typically contain summaries of those studies.
Based on the provided document, here's what can be inferred or explicitly stated regarding the acceptance criteria and study proving adherence to them:
The device in question is a Total Ankle System, a type of orthopedic implant, not an AI/ML-based device or one that underwent a clinical efficacy study for its 510(k) clearance.
The "study" that proves the device meets acceptance criteria, in this context, refers to non-clinical performance bench testing.
1. A table of acceptance criteria and the reported device performance:
Since no clinical efficacy study was conducted for this 510(k) clearance, there's no data to populate such a table detailing clinical performance metrics (like sensitivity, specificity, or human reader improvement).
The acceptance criteria for this device were related to substantial equivalence through bench testing. The document states:
Acceptance Criteria (Implied for Substantial Equivalence) | Reported Device Performance (from Non-Clinical Bench Testing) |
---|---|
Demonstrate wear performance comparable to predicate. | Testing performed to demonstrate substantial equivalence in wear. |
Demonstrate articular stability comparable to predicate. | Testing performed to demonstrate substantial equivalence in articular stability. |
Demonstrate lock detail comparable to predicate. | Testing performed to demonstrate substantial equivalence in lock detail. |
Adherence to manufacturing processes. | Testing performed to demonstrate substantial equivalence in manufacturing processes. |
Demonstrate static strength comparable to predicate. | Testing performed to demonstrate substantial equivalence in static strength. |
Demonstrate fatigue strength comparable to predicate. | Testing performed to demonstrate substantial equivalence in fatigue strength. |
Demonstrate fretting performance comparable to predicate. | Testing performed to demonstrate substantial equivalence in fretting. |
Demonstrate corrosion resistance comparable to predicate. | Testing performed to demonstrate substantial equivalence in corrosion. |
Demonstrate MRI safety. | Testing performed to demonstrate substantial equivalence in MRI safety. |
Overall, no new questions of safety or effectiveness raised compared to predicates. | "The differences in design specifications do not raise any new questions of safety and effectiveness over the predicate, which is demonstrated in the performance testing and process validation." |
2. Sample sized used for the test set and the data provenance:
- Sample Size: Not specified in the document, as it refers to engineering and materials testing (bench testing) rather than a dataset of patient cases. For bench testing, "samples" would refer to the number of devices or components tested.
- Data Provenance: Not applicable in the context of clinical patient data (e.g., country of origin, retrospective/prospective), as the evidence was non-clinical bench testing. The testing would have been conducted in a laboratory setting.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. Ground truth as typically defined for AI/ML or clinical studies (e.g., expert consensus on medical images) was not established. The "ground truth" for bench testing is determined by engineering standards, material science principles, and established testing methodologies.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. This is relevant for studies involving human reviewers or AI output, not for materials and mechanical bench testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This type of study is specifically for AI/ML-assisted diagnostic devices. The Incompass Total Ankle System is a physical medical implant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No. This is relevant for AI/ML algorithms.
7. The type of ground truth used:
- The "ground truth" for this clearance was established through engineering standards, material specifications, and validated bench testing protocols comparing the device's physical and mechanical properties against those of legally marketed predicate devices. It is not based on expert consensus, pathology, or outcomes data in the clinical sense.
8. The sample size for the training set:
- Not applicable. This device is not an AI/ML algorithm that requires a training set.
9. How the ground truth for the training set was established:
- Not applicable. This device is not an AI/ML algorithm.
Ask a specific question about this device
(58 days)
: K250641**
Trade/Device Name: APEX 3D Total Ankle Replacement System
Regulation Number: 21 CFR 888.3110
System
Device Class and Common Name: Class II, Total Ankle Prosthesis
Classification: 21 CFR 888.3110
The APEX 3D™ Total Ankle Replacement System is indicated as a total ankle replacement in primary surgery for patients with ankle joints damaged by severe rheumatoid, post traumatic, or degenerative arthritis. Revision surgery for these patients is also indicated for patients with sufficient bone stock present. In the United States, components are intended for cemented use only.
The APEX 3D Total Ankle Replacement System has been cleared under K192994. The purpose of this submission is to introduce a new diffusion bonded talar dome to the APEX 3D Total Ankle Replacement System. The APEX 3D Total Ankle Replacement System is a cemented, fixed-bearing device comprised of a tibial component, a talar component, and a UHMWPE component used for ankle joint replacement. Based on patient anatomy, a number of component sizes and design configurations can be selected for best fit.
The provided document is an FDA 510(k) clearance letter for the APEX 3D Total Ankle Replacement System. This document does not describe AI/ML-driven medical devices or related acceptance criteria studies.
The 510(k) summary explains that the submission is to introduce a new diffusion-bonded talar dome to an already cleared total ankle replacement system. The performance testing described is focused on the physical and mechanical properties of the implant materials and manufacturing processes, such as:
- Porous Structure Characterization
- Diffusion Bonded Tensile Strength
- Diffusion Bonded Static Shear Strength
- Diffusion Bonded Fatigue Shear Strength
- Diffusion Bonded Abrasion Resistance
- Metallurgical Analysis
- Accelerated Corrosion Soak
The document explicitly states: "Clinical data are not needed to support the safety and effectiveness of the subject device."
Therefore, I cannot provide the information requested in your prompt regarding acceptance criteria and studies for an AI/ML-driven device, as this document pertains to a mechanical orthopedic implant, not an AI/ML product.
To answer your request, I would need a different source document that details the performance study of an AI/ML-driven medical device.
Ask a specific question about this device
(111 days)
The iTotal® Identity™ Cruciate Retaining 3DP Porous Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.
The Indications for Use include:
· Painful joint disease due to osteoarthritis, traumatic arthritis or osteonecrosis of the knee.
· Post traumatic loss of joint function.
· Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
· Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants.
· Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans
This implant intended for uncemented use, but porous implants may be used with cement if desired by the surgeon.
The CS (cruciate sacrificing) insert option should be utilized when additional anterior-posterior constraint is desired.
The iTotal® Identity™ Cruciate Retaining 3DP Porous Knee Replacement System is a tricompartmental semi-constrained knee prosthesis composed of three components: a Femoral Component, a Tibial Component, and a Patellar Component. The product design incorporates a bone preserving approach for the treatment of severe pain and/or disability of a knee damaged by osteoarthritis or trauma. The joint restoring design provides for more natural kinematics by maintaining the patient specific femoral sagittal curves, preserving the patient specific femoral offset, preserving the medial and lateral joint lines and having a patient specific fit. It is intended for use in those patients whose condition cannot be appropriately or effectively addressed using a device that treats only one or two compartments of the knee (i.e. a unicompartmental, bicompartmental, or patellofemoral prosthesis).
Using patient imaging (CT scans), a patient-specific implant is designed that best meets the geometric and anatomic requirements of the specific patient. The treatment allows for the placement of an uncemented metallic device designed from the patient's natural bone geometry. The femoral component is additively manufactured from a cobalt chromium molybdenum ("CoCrMo") alloy. The tibial tray is additively manufactured from titanium ("Ti-6Al-4V") alloy. The tibial inserts are manufactured and offered in either ultra-high molecular weight polyethylene (iPoly®) or highly crosslinked ultra-high molecular weight Vitamin-E enriched polyethylene (iPoly® XE). The patellar component is manufactured and offered in ultra-high molecular weight polyethylene (iPoly®) with a porous Ti6Al4V metal backing. The femoral, tibial, and patellar implants are additively manufactured using proprietary TIDAL Technology™ which allows biological fixation without the need for bone cement. The iTotal® ldentity™ CR 3DP Porous KRS is designed for press-fit use without cement, but may be used with a cemented technique if desired.
The iTotal® Identity™ CR 3DP Porous KRS is supplied with disposable, patient-specific instrumentation (ilig®) designed for use with the system. These patient-specific guides are pre-navigated to fit the contours of the patient's femoral and tibial anatomies and to facilitate a simpler surgical technique. Each set of instruments is designed for one-time use, specifically for one patient. The iJig® instrument set is manufactured from biocompatible nylon material and supplied sterile along with the implants.
The provided text describes a medical device, the "iTotal® Identity™ Cruciate Retaining 3DP Porous Knee Replacement System," and its submission for FDA 510(k) clearance. However, it explicitly states that "No clinical testing was warranted for the subject device."
This means that the information you requested regarding acceptance criteria and the study that proves the device meets those criteria cannot be found in this document because a clinical study was not conducted or submitted for this particular clearance.
Therefore, I cannot provide the following information:
- A table of acceptance criteria and the reported device performance: No clinical performance data is reported.
- Sample size used for the test set and the data provenance: No clinical test set.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: No clinical test set.
- Adjudication method for the test set: No clinical test set.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size: Not applicable as no clinical study was done.
- If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable as this is a physical medical implant, not a software algorithm.
- The type of ground truth used: No clinical ground truth established.
- The sample size for the training set: No clinical training set.
- How the ground truth for the training set was established: No clinical training set.
The document focuses on demonstrating substantial equivalence to previously cleared predicate devices through non-clinical performance evaluation, which includes various biomechanical and material tests. These non-clinical tests are listed, but their specific acceptance criteria and detailed results are not provided in this summary.
Non-Clinical Performance Evaluation (as listed in the document):
- Patella static tensile test
- Patella static shear test
- Patella shear fatigue test
- Tibial tray fatigue test
- Femoral fatigue test
- Tibial micromotion
- Tibiofemoral constraint test
- Tibiofemoral contact area, stress and wear test
- Tibial interlock test
- Patellofemoral subluxation, contact area and stress test
- Particle characterization
- Porous coating testing per FDA guidance on modified metallic surfaces
- Biocompatibility assessment
Conclusion from the document: "The results of the testing demonstrate that the subject device iTotal® Identity™ CR 3DP Porous KRS is as safe and effective and is substantially equivalent to the cleared predicates... and the differences between the subject and predicates do not raise different questions of safety or effectiveness."
Ask a specific question about this device
(286 days)
Re: K240475
Trade/Device Name: Hintermann Series H2 Total Ankle System Regulation Number: 21 CFR 888.3110
Polyethylene Inlay, Talar Component |
| Classification: | 21 CFR 888.3110
The Hintermann Series H2 Total Ankle (Hintermann Series H2) system is designed to treat ankle arthritis through replacement of the ankle joint with a prosthesis, thereby reducing pain, restoring alignment, replacing flexion and extension movement in the ankle joint, and allowing for movement at the replaced joint.
The Hintermann Series H2 is indicated as a total ankle replacement in primary of ankle joints damaged bv:
-
Systemic arthritis of the ankle (e.g., rheumatoid arthritis, hemochromatosis)
-
Primary arthritis (e.g., degenerative disease)
-
Secondary arthritis (e.g., post-traumatic, avascular necrosis, provided enough of the talus is preserved to support the implant)
The Hintermann Series H2 is also indicated for patients with a failed previous ankle surgery and revision surgeries following failed total ankle replacement or non-union of the ankle arthrodesis, provided sufficient bone stock is present.
Note: In the United States, this device is intended for cemented use only.
The Hintermann Series H2 Total Ankle System is a total ankle prosthesis designed to replace the ankle joint. The Hintermann Series H2 Total Ankle System includes the following implantable components.
- H2 Tibial Component
- H2 Tibial Polyethylene (PE) Inlay
- Hintermann Series Talar Component
The tibial and talar components are coated with plasma spray titanium on the cement contacting surfaces. These components are used together to address the indications as outlined.
The provided text is a 510(k) summary for the Hintermann Series H2 Total Ankle System, a medical device. This document focuses on demonstrating the substantial equivalence of the new device to a legally marketed predicate device, rather than providing an acceptance criteria table and detailed study results typical of an AI/ML medical device submission.
As such, the information required to answer your specific questions about acceptance criteria and a study proving the device meets those criteria (especially in the context of AI/ML performance) is not present in the provided text.
The document describes non-clinical testing/performance data for the ankle system, which refers to mechanical and material tests to ensure the physical properties and safety of the implant, not the performance of an AI/ML algorithm.
Here's why the document doesn't contain the requested information:
- Device Type: The Hintermann Series H2 Total Ankle System is a physical orthopedic implant (an ankle joint prosthesis), not an AI/ML-driven diagnostic or therapeutic device.
- Regulatory Pathway: A 510(k) submission for a physical device like this primarily focuses on demonstrating substantial equivalence to a predicate device through engineering analysis, material characterization, and mechanical testing (e.g., wear, fatigue, range of motion). It does not involve AI/ML performance metrics, ground truth establishment, or human reader studies.
Therefore, for each of your points, the answer based solely on the provided text would be:
- A table of acceptance criteria and the reported device performance: Not applicable. The document lists mechanical tests (e.g., Range of Motion per ASTM F2665-09, Locking Mechanism Testing per ASTM F1814) that were performed, but it does not present a table of specific quantitative acceptance criteria or the numerical results of these tests. It simply states that testing was done to support substantial equivalence.
- Sample sizes used for the test set and the data provenance: Not applicable in the context of AI/ML. For mechanical testing, sample sizes would refer to the number of prostheses tested, but this information is not detailed in the summary.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No human experts are establishing ground truth for an AI/ML algorithm in this context.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable. The "ground truth" for a mechanical device would be established by standardized testing protocols (e.g., ASTM standards) measuring physical parameters, not by expert consensus on clinical cases.
- The sample size for the training set: Not applicable.
- How the ground truth for the training set was established: Not applicable.
In summary, the provided FDA 510(k) document is for a physical medical device (an ankle prosthesis) and not an AI/ML-driven device. As such, it does not contain the information related to acceptance criteria, test sets, ground truth, or human reader studies typically associated with AI/ML medical device performance evaluation.
Ask a specific question about this device
(115 days)
38117
Re: K241999
Trade/Device Name: Prophecy Surgical Planning System Regulation Number: 21 CFR 888.3110
|
| Classification Name and Reference: | 21 CFR 888.3110
The Prophecy® Surgical Planning System is intended to be used as patient-specific surgical instrumentation to assist in the positioning of total ankle replacement components intraoperatively, in guiding the marking of bone before cutting, and in the pre-surgical planning of the ankle and surrounding anatomy to support the total ankle implant. The Prophecy® Surgical Planning Guides and Reports are intended for use with the Inbone®. Infinity® and Invision® Total Ankle Systems and their cleared indications for use, provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans (e.g. CT scans and X-rays). The Prophecy® Surgical Planning guides are intended for single use only.
The Prophecy® Surgical Planning System is composed of three components:
- Prophecy® patient-specific guides
- Prophecy® 3D Planner
- Prophecy® Preoperative report
The Prophecy® Surgical Planning System is compatible with the Inbone®, Infinity®, and Invision® Total Ankle Systems.
The provided text describes a 510(k) premarket notification for the Prophecy® Surgical Planning System. The submission focuses on modifications to the Prophecy® 3D Planner software. The information available details non-clinical testing performed to demonstrate substantial equivalence, but it does not provide specific acceptance criteria or an explicit study proving performance against those criteria in a format you requested for a device that involves performance against defined metrics (e.g., accuracy, sensitivity, specificity).
Here's a breakdown of the available information based on your questions:
1. Table of Acceptance Criteria and Reported Device Performance
The submission states, "All test results met the acceptance criteria, demonstrating the subject device performs as intended and is substantially equivalent to the predicate device." However, the text does not explicitly list the specific acceptance criteria or the reported device performance metrics in a quantitative manner.
2. Sample size used for the test set and the data provenance
The document mentions "Software verification testing," "Software validation," "Usability test," and "Cybersecurity testing." These are general descriptions of testing types. No specific sample sizes for test sets or data provenance (e.g., country of origin, retrospective/prospective) are provided in the given text.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The document states, "Clinical testing was not necessary for the determination of substantial equivalence." This implies that the validation did not rely on expert-established ground truth in a clinical context for performance metrics as you typically find for AI diagnostic devices. For the usability testing, the number and qualifications of intended users (surgeons or engineers) involved in establishing ground truth are not specified.
4. Adjudication method for the test set
Since specific acceptance criteria involving expert-reviewed ground truth are not detailed, no adjudication method is mentioned or implied in the provided text.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
The document explicitly states, "Clinical testing was not necessary for the determination of substantial equivalence." Therefore, no MRMC comparative effectiveness study was performed or is referenced.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The Prophecy® Surgical Planning System is described as a tool that "assists in the positioning of total ankle replacement components intraoperatively, in guiding the marking of bone before cutting, and in the pre-surgical planning," and involves "user interface software intended to be used by orthopedic surgeons" and "trained employees/engineers." This suggests a human-in-the-loop process. While software verification and validation would test the algorithm's performance, the context implies it's not a standalone diagnostic AI, but rather a planning and guidance tool. The text does not explicitly detail a standalone algorithm-only performance study in the manner of a diagnostic AI.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Given that clinical testing was not performed for substantial equivalence, the "ground truth" for the software's performance would likely be based on engineering specifications, design requirements, and possibly simulated or phantom data for accuracy of measurements and guide design. For usability, it would relate to user feedback and completion of tasks as per the system's design. The document mentions "software verification testing to ensure all design outputs meet all specified requirements" and "software validation to ensure software specifications conform to user needs and intended uses," which aligns with this interpretation.
8. The sample size for the training set
The document refers to a "Prophecy® 3D Planner" software update. It does not mention any machine learning or AI model training. Therefore, a training set size is not applicable or provided in this context.
9. How the ground truth for the training set was established
As there's no mention of a traditional machine learning training set, this question is not applicable. The "ground truth" for the software's development (not training) would have been established through engineering design processes and requirements.
Ask a specific question about this device
(25 days)
Colorado 80127
Re: K243173
Trade/Device Name: Salto Talaris Ankle PSI System Regulation Number: 21 CFR 888.3110
cemented prosthesis |
| Classification: | Class II, 21 CFR 888.3110
The Salto Talaris Ankle PSI System is intended to be used as patient specific surgical planning and instrumentation to assist in the positioning of total ankle replacement components intraoperatively, and in guiding bone cutting. The Salto Talaris Ankle PSI System is intended for use with Smith + Nephew's Salto Talaris Total Ankle System and its cleared indications for use, provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging CT scans.
3D Systems' Salto Talaris Ankle PSI System consists of patient specific outputs including surgical guides, anatomical models, and case reports. The Salto Talaris Ankle PSI System guides are made from biocompatible nylon and surgical grade stainless steel and are designed to fit the contours of the patient's distal tibial and proximal talar anatomy. The surgical guides in combination with the Smith+Nephew Salto Talaris Total Ankle System instruments, facilitate the positioning of Salto Talaris Total Ankle Prostheses.
The provided document describes the FDA clearance (K243173) for the Salto Talaris Ankle PSI System. However, it does not contain a detailed study with acceptance criteria and reported device performance in the format requested.
The document states:
- "Non-clinical performance testing on the Salto Talaris Ankle PSI System included cadaveric comparison testing in order to compare Implant Alignment Accuracy and Guide Usability between the subject device and reference device."
- "Accuracy and functionality were shown to be similar to that of the standard instrumentation used for the Salto Talaris Total Ankle System."
This indicates that a comparison study was performed, but the specific acceptance criteria, reported performance values, sample size, ground truth establishment, or expert details are not provided in this 510(k) summary. The summary concludes that the device is substantially equivalent and "performs as well as the predicate device" based on this testing, but the numerical data from the study is not included.
Therefore, I cannot populate the table or answer the specific questions about the study design with the information available in the provided text. The document focuses on regulatory clearance based on substantial equivalence, rather than a detailed presentation of performance study results against predefined acceptance criteria.
Ask a specific question about this device
(25 days)
Colorado 80127
Re: K243174
Trade/Device Name: Cadence Ankle PSI System Regulation Number: 21 CFR 888.3110
cemented prosthesis |
| Classification: | Class II, 21 CFR 888.3110
The Cadence Ankle PSI System is intended to be used as patient specific surgical planning and instrumentation to assist in the positioning of total ankle replacement components intraoperatively, and in guiding bone cutting. The Cadence Ankle PSI System is intended for use with Smith + Nephew's Cadence Total Ankle System and its cleared indications for use, provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging CT scans.
3D Systems' Cadence Ankle PSI System consists of patient specific outputs including surgical guides, anatomical models, and case reports. The Cadence Ankle PSI System guides are made from biocompatible nylon and surgical grade stainless steel and are designed to fit the contours of the patient's distal tibial and proximal talar anatomy. The surgical guides in combination with the Smith+Nephew Cadence Total Ankle System instruments, facilitate the positioning of Cadence Total Ankle Prostheses.
The provided text does not contain detailed information about the acceptance criteria or a specific study that proves the device meets those criteria in a format that would allow for a precise population of the requested table and points. The document is a 510(k) summary for a medical device (Cadence Ankle PSI System), which focuses on demonstrating substantial equivalence to a predicate device rather than presenting extensive performance study results like those typically found for AI/ML-based diagnostic devices.
However, I can extract the available information regarding non-clinical performance testing:
Non-clinical Performance Testing:
The relevant section states: "Non-clinical performance testing on the Cadence Ankle PSI System included cadaveric comparison testing in order to compare Implant Alignment Accuracy and Guide Usability between the subject device and reference device. Accuracy and functionality were shown to be similar to that of the standard instrumentation used for the Cadence Total Ankle System."
This indicates that the study focused on Implant Alignment Accuracy and Guide Usability. The acceptance criteria are implied to be "similar to that of the standard instrumentation used for the Cadence Total Ankle System," which served as the reference device.
Based on the provided text, the following information is not available:
- A formal table of acceptance criteria with reported device performance metrics (e.g., specific thresholds for accuracy or usability scores).
- Detailed sample size for the test set (only "cadaveric comparison testing" is mentioned).
- Data provenance (e.g., country of origin, retrospective/prospective).
- Number of experts, their qualifications, or adjudication methods for ground truth.
- Whether a multi-reader multi-case (MRMC) comparative effectiveness study was done.
- If a standalone (algorithm only) performance was measured (this device is a surgical guide system, not an AI diagnostic algorithm in the typical sense).
- The specific type of ground truth used beyond "comparison to the standard instrumentation."
- Sample size for the training set.
- How the ground truth for the training set was established.
In summary, the document states performance testing focused on implant alignment accuracy and guide usability through cadaveric comparison, indicating similarity to standard instrumentation as the performance benchmark. However, granular details about specific acceptance criteria metrics, study design, and ground truth establishment are not provided.
Ask a specific question about this device
(28 days)
Carolina 27709
Re: K242868
Trade/Device Name: Kinos Total Ankle System Regulation Number: 21 CFR 888.3110
| Ankle Joint Metal/Polymer Semi-Constrained Cemented Prosthesis |
| Regulation Number: | 21 CFR 888.3110
The Kinos Total Ankle System is indicated for patients with ankle joints damaged by severe rheumatoid, post-traumatic, or degenerative arthritis. The Kinos Total Ankle System is additionally indicated for patients with failed previous ankle surgery. The Kinos Total Ankle System is intended for cement use only.
The subject Kinos Total Ankle System consists of implant and instrument components designed to preserve motion in patients with ankle arthritis or previously failed ankle surgery. The device is a fixed bearing and semi-constrained implant construct, intended for replacement of the articulating surfaces of the ankle that have been affected by a disease state or injury. The subject device provides limited mobility to a patient by restoring alignment of the articulating surfaces, reducing pain, and providing flexion- extension movement within the ankle joint. The tibial and talar implants are intended to be used with bone cement. The subject Kinos Total Ankle System consists of three implant components - a tibial implant affixed to the tibia, a bearing implant assembled to the tibial implant and a talar implant affixed to the talus. The tibial implant and bearing implant, assembled intraoperatively, function together and articulate with the talar implant to create a fixed bearing prosthesis, replacing the articulating surface of the tibiotalar joint. The subject line extension includes tibial segments which are assembled intraoperatively to the tibial base plate to enhance tibial fixation height in the tibia.
This document is a 510(k) Summary for the Kinos Total Ankle System. It describes the device, its intended use, and its substantial equivalence to predicate devices. However, it does not contain any information about acceptance criteria or a study proving the device meets those criteria, particularly not in the context of an AI/ML-driven device.
The Kinos Total Ankle System is a physical medical device (an ankle prosthesis), and the submission is for an expansion of indications for use. The performance testing section specifically states: "No performance testing was performed to establish substantial equivalence of the subject device to the predicates." Instead, a "dimensional analysis comparison of device components and a surgical technique comparison" was conducted.
Therefore, I cannot provide the requested information, as the provided text relates to a traditional medical device and its regulatory submission, not a study validating an AI/ML device's performance against specific acceptance criteria.
Ask a specific question about this device
(127 days)
Colorado 80127
Re: K241148
Trade/Device Name: Salto Talaris Ankle PSI System Regulation Number: 21 CFR 888.3110
cemented prosthesis |
| Classification: | Class II, 21 CFR 888.3110
The Salto Talaris PSI System is intended to be used as patient specific surgical planning and instrumentation to assist in the positioning of total ankle replacement components intraoperatively, and in guiding bone cutting. The Salto Talaris PSI System is intended for use with Smith + Nephew's Salto Talaris Total Ankle System and its cleared indications for use.
3D Systems' Salto Talaris Ankle PSI System consists of patient specific outputs including surgical guides, anatomical models, and case reports. The Salto Talaris Ankle PSI System guides are made from biocompatible nylon and surgical grade stainless steel and are designed to fit the contours of the patient's distal tibial and proximal talar anatomy. The surgical guides in combination with the Smith+Nephew Salto Talaris Total Ankle System instruments, facilitate the positioning of Salto Talaris Total Ankle Prostheses.
The provided text is a 510(k) summary for the Salto Talaris Ankle PSI System. It describes the device, its intended use, and a summary of non-clinical testing. However, it does not contain the detailed information necessary to fully answer all aspects of your request, particularly regarding specific acceptance criteria, expert qualifications, adjudication methods, details of ground truth establishment, or specific sample sizes for training and testing sets in the context of an AI-based system.
The Salto Talaris Ankle PSI System is described as patient-specific surgical planning and instrumentation to assist in positioning total ankle replacement components and guiding bone cutting. This device primarily involves physical surgical guides and anatomical models derived from CT data, rather than a standalone AI algorithm for diagnosis or image analysis. Therefore, some of the requested information, such as an MRMC comparative effectiveness study with AI assistance or standalone algorithm performance, is not directly applicable to this type of device as typically described for AI/ML-based diagnostic software.
Here's a breakdown of the available information based on your request:
1. Table of acceptance criteria and the reported device performance
Acceptance Criteria | Reported Device Performance |
---|---|
Mechanical Integrity (post-processing) | Met all acceptance criteria. |
Debris Generation | Met all acceptance criteria. |
Inter-Designer Variability analysis | Met all acceptance criteria. |
Implant Alignment Accuracy (cadaveric comparison) | Accuracy shown to be similar to standard instrumentation for the Salto Talaris Total Ankle System. |
Guide Usability (cadaveric comparison) | Functionality shown to be similar to standard instrumentation for the Salto Talaris Total Ankle System. |
The document states that the Salto Talaris Ankle PSI System met all acceptance criteria for mechanical integrity, debris generation, and inter-designer variability analysis. For implant alignment accuracy and guide usability, cadaveric comparison testing showed similarity to standard instrumentation. The specific quantitative acceptance criteria values (e.g., specific thresholds for mechanical integrity or debris generation, or a numerical range for alignment accuracy) are not provided in this summary.
2. Sample size used for the test set and the data provenance
The document mentions "Non-clinical cadaveric comparison testing." However, it does not specify the sample size (number of cadavers or cases) used for this test set, nor does it provide information about the data provenance (e.g., country of origin, retrospective or prospective).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The document does not provide this information. Since the device involves surgical guides and bone cutting, "ground truth" in this context would likely relate to anatomical measurements or surgical outcomes, potentially assessed by orthopedic surgeons. However, no details on experts or their qualifications are given.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
The document does not specify any adjudication method.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC comparative effectiveness study of the type typically performed for AI/ML diagnostic software (where human readers evaluate cases with and without AI assistance) was not mentioned and is unlikely to be applicable based on the device description. This device provides physical guides for surgery, not AI-based image analysis for diagnosis. The non-clinical cadaveric testing compared the device to standard instrumentation, not to human readers using or not using AI.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
The device is described as "patient specific surgical planning and instrumentation to assist in the positioning... and in guiding bone cutting." This inherently implies a "human-in-the-loop" scenario (a surgeon using the guides). The "non-clinical cadaveric comparison testing" assessed the performance of the device in use, which is a form of standalone performance for the instrumentation itself but not in the context of an AI algorithm without human interaction for diagnosis or interpretation. The document does not describe a standalone algorithm performance test in the way it would be applied to AI/ML diagnostic software.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the cadaveric comparison testing, the "ground truth" implicitly relates to implant alignment accuracy and guide usability, compared against the results achieved using standard (non-PSI) instrumentation. The summary doesn't explicitly state how this ground truth was definitively established (e.g., by highly accurate post-operative CT measurements validated by multiple experts), but it refers to comparison rather than an absolute ground truth method.
8. The sample size for the training set
The document describes the device as being "designed with CT-based methods to produce patient-specific instrumentation." This suggests a design process based on anatomical data, but there is no mention of a "training set" in the context of an AI/ML algorithm. The device is a custom-manufactured surgical guide, not a learned AI model.
9. How the ground truth for the training set was established
Since there's no mention of a "training set" for an AI/ML algorithm, this question is not applicable in the context of the provided document. The "design" information would likely come from anatomical studies, engineering specifications, and clinical experience with total ankle arthroplasty, rather than a formal "ground truth" establishment for an AI training set.
Ask a specific question about this device
(112 days)
Colorado 80127
Re: K241326
Trade/Device Name: Cadence Ankle PSI System Regulation Number: 21 CFR 888.3110
cemented prosthesis |
| Classification: | Class II, 21 CFR 888.3110
The Cadence Ankle PSI System is intended to be used as patient specific surgical planning and instrumentation to assist in the positioning of total ankle replacement components intraoperatively, and in guiding bone cutting. The Cadence Ankle PSI System is intended for use with Smith + Nephew's Cadence Total Ankle System and its cleared indications for use.
3D Systems' Cadence Ankle PSI System consists of patient specific outputs including surgical guides, anatomical models, and case reports. The Cadence Ankle PSI System guides are made from biocompatible nylon and surgical grade stainless steel and are designed to fit the contours of the patient's distal tibial and proximal talar anatomy. The surgical guides in combination with the Smith+Nephew Cadence Total Ankle System instruments, facilitate the positioning of Cadence Total Ankle Prostheses.
The manufacturer, 3D Systems, Inc., has introduced the Cadence Ankle PSI System, a device intended for patient-specific surgical planning and instrumentation to assist in total ankle replacement component positioning and bone cutting. This device is designed for use with Smith + Nephew's Cadence Total Ankle System.
The information provided by the FDA 510(k) summary for K241326 primarily focuses on non-clinical performance testing to demonstrate substantial equivalence to the predicate device, not on clinical performance with human readers or standalone AI performance. Therefore, many of the requested details regarding clinical study design (e.g., MRMC studies, human reader improvement, expert consensus for ground truth) are not applicable based on the provided document.
Here's an analysis of the acceptance criteria and the study conducted, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The 510(k) summary mentions "all acceptance criteria for all performance tests" were met, but it does not explicitly list the quantitative acceptance criteria. It broadly states the types of tests conducted and their qualitative outcomes.
Acceptance Criteria Category | Reported Device Performance |
---|---|
Non-clinical Performance Testing: Mechanical Integrity (post-processing) | Met all acceptance criteria. |
Non-clinical Performance Testing: Debris Generation | Met all acceptance criteria. |
Non-clinical Performance Testing: Intra- and Inter-Designer Variability analysis | Met all acceptance criteria. |
Non-clinical Cadaveric Comparison Testing: Implant Alignment Accuracy (vs. reference device) | Shown to be similar to that of the standard instrumentation currently used for the Cadence Total Ankle System. |
Non-clinical Cadaveric Comparison Testing: Guide Usability (vs. reference device) | Shown to be similar to that of the standard instrumentation currently used for the Cadence Total Ankle System. |
2. Sample Size Used for the Test Set and Data Provenance
The document indicates "Non-clinical cadaveric comparison testing", which implies a test set was used. However, the exact sample size (number of cadavers or anatomic specimens) is not specified.
The data provenance is cadaveric testing, which is a form of pre-clinical, laboratory-based testing, not human patient data (retrospective or prospective). The country of origin for the data is not specified, though 3D Systems, Inc. is based in the USA.
3. Number of Experts Used to Establish Ground Truth and Qualifications
This information is not applicable as the studies described are non-clinical, cadaveric, and mechanical/design verification tests. There is no mention of experts establishing a "ground truth" in the context of diagnostic or clinical interpretation. Performance was assessed mechanically or by comparison to standard instrumentation.
4. Adjudication Method for the Test Set
This is not applicable as the studies are non-clinical and do not involve human diagnostic interpretation requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. The document describes non-clinical and cadaveric testing, not studies involving human readers or clinical cases.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
The device is a "patient specific surgical planning and instrumentation" system. While it likely involves algorithms for design and manufacturing, the studies described are for the physical outputs (surgical guides, anatomical models) and their performance (mechanical integrity, debris, accuracy in cadavers), rather than a standalone AI algorithm's diagnostic performance. Therefore, a standalone AI performance study in the typical sense of evaluating an algorithm’s output without human-in-the-loop was not explicitly described for this submission. The focus is on the device's accuracy in physical guidance.
7. The Type of Ground Truth Used
For the non-clinical tests (mechanical integrity, debris generation, intra-/inter-designer variability), the "ground truth" would be established by engineering specifications, quality control standards, and measurement protocols.
For the cadaveric comparison testing, the "ground truth" for "Implant Alignment Accuracy" and "Guide Usability" would be based on direct measurements against design specifications and comparison to the performance of established standard instrumentation (the reference device, K151459 Cadence Total Ankle Replacement System). This is a technical, rather than a clinical, ground truth.
8. The Sample Size for the Training Set
The document does not describe the use of a "training set" in the context of machine learning or AI algorithm development. The device is a patient-specific instrument system, and its design is CADD-based. If there's an underlying AI component that uses a training set, this information is not provided in the summary.
9. How the Ground Truth for the Training Set Was Established
As no training set is described for an AI algorithm, this information is not applicable based on the provided document. If there are proprietary algorithms involved in the patient-specific design, the "training" data (if any) and its ground truth would be part of the manufacturer's internal development process, not typically disclosed in this level of detail in a 510(k) summary focused on the final product's performance.
Ask a specific question about this device
Page 1 of 7