(26 days)
Cary is indicated for removal of fluids from the upper airway. The device creates a negative pressure (vacuum) that draws fluids through disposable suction tube that is connected to a disposable collection canister. The fluids are captured in the collection canister for proper disposal. Cary is for use only on the order of a physician or other licensed practitioner (e.g. Emergency Medical Technician in field use).
[Cary] is a handheld, electrically powered medical suction device used to clear the upper airway during emergency care. It consists of a reusable drive unit and disposable collection container (pumpister) and tubing. The drive unit powers a piston pump, creating negative pressure (vacuum) in the disposable collection container. The vacuum draws fluids and other matter through a disposable tubing connected to the container, where they are collected for disposal. The size and shape of the device enable use by one hand, and a rechargeable lithium-ion battery allows field and transport use of the device.
The provided FDA 510(k) clearance letter and summary for the "Cary" device ([K251111](https://510k.innolitics.com/search/K251111)) primarily discuss non-clinical performance testing to demonstrate substantial equivalence to a predicate device. This document does not describe a clinical study (e.g., MRMC or standalone human-in-the-loop study) involving human readers or expert panels for establishing ground truth, as would typically be seen for AI/ML-driven diagnostic devices. Instead, the testing focuses on the physical and functional performance of the suction pump itself against recognized consensus standards.
Therefore, many of the requested points regarding clinical study design, expert panels, ground truth, and MRMC studies are not applicable based on the provided text.
Here's an analysis based on the information provided:
1. Table of Acceptance Criteria and Reported Device Performance
The document describes several non-clinical tests with implied or explicit acceptance criteria based on industry standards and comparison to the predicate device.
| Test Category | Acceptance Criteria (Implied/Explicit from Standards/Predicate) | Reported Device Performance |
|---|---|---|
| Operating Position | Ability to operate in any spatial orientation. | Confirmed. |
| Protection Devices | Compliance with overfill protection and reverse flow protection. | Results within acceptance criteria. |
| Noise Level | Unspecified standard, implied safe/acceptable level. | 70.5 dB. "Within the acceptance criteria apart from 0.5dB excess when the suction tip was blocked. This is considered as an acceptable deviation from the standard requirement as it does not decrease the safety or effectiveness of the subject device when compared to the predicate device." |
| Air Leakage | As per standard requirements for general use suction equipment. | Results within acceptance criteria. |
| Vacuum Levels & Free Air Flow | Within acceptance criteria for High vacuum/high flow and Low vacuum/low flow (similar to predicate: High vacuum <600 mmHg vs <550 mmHg; Low vacuum ≤100 mmHg vs 50 mmHg; High flow ≥ 20 l/min vs 30 l/m). | Results within acceptance criteria. Suction performance "similar to LCSU 4". |
| Pharyngeal Suction Equipment | Ability to evacuate ≥ 200 ml of simulated vomitus in ≤ 10 s. | Results within acceptance criteria. |
| Reprocessing Validation | Appropriateness of manual reprocessing process; user ability to carry out instructions. | Outcome within acceptance criteria. |
| Biocompatibility | Compliance with ISO 10993 standards for cytotoxicity, sensitization, irritation, acute systemic toxicity, and material mediated pyrogenicity. | Confirmed for direct-contact materials. |
| Software | Compliance with IEC 62304 and FDA guidance for software functions, classified as Class B. | Verification and validation conducted, documentation provided as recommended. (Specific performance metrics are not applicable here as it's about compliance, not a clinical outcome). |
| EMC and Electrical Safety | Compliance with IEC 60601-1-2 and IEC 60601-1 series. | Confirmed. |
2. Sample Size Used for the Test Set and Data Provenance
The document describes non-clinical, laboratory-based testing of the device's physical and functional performance, not a clinical study involving a dataset of patient cases. Therefore, "sample size for the test set" and "data provenance" in the context of patient data are not applicable. The tests were performed on the device prototypes/production samples in controlled laboratory environments.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. Ground truth, in the context of device performance testing for a suction pump, refers to established engineering specifications, physical laws, and recognized standards (e.g., ISO 10079-1). There are no "experts" in the sense of medical professionals interpreting patient data to establish ground truth for this type of device.
4. Adjudication Method for the Test Set
Not applicable. As this is non-clinical performance testing against engineering standards, an adjudication method for conflicting interpretations of medical data is not relevant.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. The document explicitly describes non-clinical testing and comparison to a predicate device based on device performance metrics (e.g., vacuum levels, flow rates, noise), not on human reader performance with or without AI assistance. This device is a powered suction pump, not an AI diagnostic/imaging device that would typically undergo MRMC studies.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
Not applicable. The device is a mechanical suction pump, not an AI algorithm. Its performance is inherent to its physical design and operation, not an output from an algorithm in the typical AI sense.
7. The Type of Ground Truth Used
The ground truth used for these non-clinical tests is based on:
- Recognized Consensus Standards: e.g., ISO 10079-1, ISO 10079-4, IEC 62304, ISO 10993, IEC 60601 series.
- Engineering Specifications: Explicit performance requirements (e.g., ≥ 20 l/min flow rate, < 1 kPa air leakage).
- Comparison to Predicate Device Performance: Establishing equivalence or superiority in key performance areas.
8. The Sample Size for the Training Set
Not applicable. There is no "training set" as this is not an AI/ML device that requires machine learning training.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for an AI/ML algorithm.
FDA 510(k) Clearance Letter - K251111
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.07.05
May 7, 2025
Excitus AS
℅ Dave Yungvirt
CEO
Third Party Review Group, LLC
25 Independence Blvd
Warren, New Jersey 07059
Re: K251111
Trade/Device Name: Cary
Regulation Number: 21 CFR 878.4780
Regulation Name: Powered Suction Pump
Regulatory Class: Class II
Product Code: JCX
Dated: April 10, 2025
Received: April 11, 2025
Dear Dave Yungvirt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2
K251111 - Dave Yungvirt Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
K251111 - Dave Yungvirt Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James H. Jang -S [Digitally signed by James H. Jang -S Date: 2025.05.07 13:27:17 -04'00']
For
Long Chen, Ph.D.
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K251111
Device Name
Cary
Indications for Use (Describe)
Cary is indicated for removal of fluids from the upper airway. The device creates a negative pressure (vacuum) that draws fluids through disposable suction tube that is connected to a disposable collection canister. The fluids are captured in the collection canister for proper disposal. Cary is for use only on the order of a physician or other licensed practitioner (e.g. Emergency Medical Technician in field use).
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF
Page 5
510(k) Summary K251111
7 May 2025
Submitter
Applicant Name: Excitus AS
Applicant Address: Hillevågsveien 101, 4106 Stavanger, Norway
Applicant Phone: +47 971 57 784
Applicant contact: Øystein Refseth, CEO, or@excitus.no
Correspondent Name: Innokas Medical Oy
Correspondent Address: Metsänneidonkuja 8 Espoo 02130 Finland
Correspondent Phone: +358 40 5856625
Correspondent Contact: Minna Eskola, QA/RA Specialist, minna.eskola@innokasmedical.fi
Device Classification
Trade Name: Cary
Common Name: Powered suction pump
Classification Name: Apparatus, Suction, Ward Use, Portable, Ac-Powered
Regulation Number: 878.4780
Product Code: JCX
Legally Marketed Predicate Device
Predicate: K120065
Predicate Trade Name: Laerdal Compact Suction Unit 4 (LCSU 4)
Product Code: JCX
Device Description Summary
[Cary] is a handheld, electrically powered medical suction device used to clear the upper airway during emergency care. It consists of a reusable drive unit and disposable collection container (pumpister) and tubing. The drive unit powers a piston pump, creating negative pressure (vacuum) in the disposable collection container. The vacuum draws fluids and other matter through a disposable tubing connected to the container, where they are collected for disposal. The size and shape of the device enable use by one hand, and a rechargeable lithium-ion battery allows field and transport use of the device.
Page 6
Intended Use / Indications for Use
Indications for use: Cary is indicated for removal of fluids from the upper airway. The device creates a negative pressure (vacuum) that draws fluids through disposable suction tube that is connected to a disposable collection canister. The fluids are captured in the collection canister for proper disposal. Cary is for use only on the order of a physician or other licensed practitioner (e.g. Emergency Medical Technician in field use).
Device operation: Cary is a non-sterile, portable, handheld, battery-powered medical suction device designed for field and transport use and for evacuation of fluids.
Intended patient population: adult patients
Intended users: a physician or other licensed practitioner (e.g. Emergency Medical Technician in field use)
Indications for Use Comparison
Subject device Cary is indicated for removal of fluids from the upper airway and the predicate device is to be used to remove fluids from the airway or respiratory support system. This is similar between the subject device and predicate device, with the suction tip length of subject device Cary ensuring use in the upper airway.
Subject device Cary creates a negative pressure (vacuum) that draws fluids through disposable suction tube that is connected to a disposable collection canister. The fluids are captured in the collection canister for proper disposal. This is the same between the subject device and predicate device.
Subject device Cary is for use only on the order of a physician or other licensed practitioner (e.g. EMT-field use). This is the same between the subject device and predicate device.
The indications for use of the subject device and predicate device are essentially the same.
The restriction to use subject device Cary in the upper airway does not affect the safety and effectiveness of the subject device when compared to the predicate device as the design of the subject device (suction tip) ensures use as indicated and efficacy of performance (clearance of airway to allow respiration or ventilation).
Comparison of patient population
Subject device Cary and the predicate device are intended to be used for adult patients. This is the same between the subject device and predicate device.
Page 7
Technological Comparison
The technological characteristics of the Excitus Cary are functionally equivalent to the predicate device, the Laerdal Compact Suction Unit 4 (LCSU 4).
| Device & Predicate Device(s): | Subject Device Cary model IM006764 | Predicate Device Laerdal Compact Suction Unit 4 (LCSU 4) K120065 | Significant Difference |
|---|---|---|---|
| General Device Characteristics | |||
| Operation | Continuous | 30 min ON, 30 min OFF (intermittent) | Different |
| Air leakage | < 1 kPa | Not specified | -- |
| High vacuum | <600 mmHg | <550 mmHG | Similar |
| Low vacuum | ≤100 | 50 mmHg | Similar |
| High flow rate | ≥ 20 l/min | 30 l/m | Similar |
| Low flow rate | < 20 l/min | Not specified | -- |
| Vacuum indicator accuracy | No indicator | 5% of full scale | Different |
| Noise level | 70.5 dB | Not specified | -- |
| Collection capacity | 300 ml | 300 ml | Same |
| Suction tubing diameter | 8 mm ID | 6.6 mm ID | Similar |
| Power Supply | 14.4V, 2900mAh Li-Ion internal Battery. Cary cannot be used during charging cycle | NiMH 12V, 1.6 Ah internal Battery LCSU 4 can be operated during charging cycle | Similar / Different |
| Battery runtime | 3 hours | 45 minutes | Different |
| Ingress protection classification | IP54 | IP33 | Different |
While there are some differences in the materials that comprise the drive unit (main unit), pumpister (container), and suction tip (suction tubing) of the Cary, the strength and durability of the materials, and suitability for the intended use are comparable to those of the predicate device. Material of the subject device in direct contact with the human body was assessed for biocompatibility in accordance with ISO10993-1, 5, 10, 11, 18, 23: Biological Evaluation of Medical Devices.
Suction performance of Cary is similar to LCSU 4 when it comes to free air flow and vacuum levels. Cary can be used in any position while holding in hand whereas LCSU 4 shall be placed upright. Collection capacity in the devices is the same.
The weight of Cary is less than that of LCSU 4 and dimensions of Cary more compact than the size of LCSU 4. User interface for vacuum level adjustment is implemented with a toggle switch in Cary and with a vacuum regulator in the predicate device LCSU 4.
Page 8
Medical electrical equipment classification of the subject device is comparable to that of the predicate device. Unlike the predicate, subject device can be charged safely from sockets that are and those that are not protective earthed.
Cary uses rechargeable lithium-ion (li-ion) battery as energy source, whereas LCSU 4 includes a nickel–metal hydride (NiMH) battery. Battery runtime of Cary is longer than that of LCSU 4.
Excitus has submitted performance testing for the above referenced design differences with this 510(k) that demonstrate that the Cary is as safe and effective as the predicate device.
Non-Clinical Tests Summary and Conclusions
Recognized Consensus Standards
1-157 ISO 10079-1 Fourth edition 2022-03 Medical suction equipment - Part 1: Electrically powered suction equipment, and ISO 10079-4 First edition 2021-08 Medical suction equipment - Part 4: General requirements, and Guidance for Industry and FDA Reviewers/Staff Guidance Document for Powered Suction Pump 510(k)s have been adopted as special controls applicable to this device.
The following non-clinical testing are provided in support of the substantial equivalence determination.
Operating position
Ability of Cary to operate in any spatial orientation was tested and confirmed.
Protection devices
Overfill protection and protection against reverse flow were tested to ensure the compliance of Cary protection devices. The results of the testing were within the acceptance criteria.
Noise
The maximum A-weighted sound pressure level was tested for Cary. The result was within the acceptance criteria apart from 0,5dB excess when the suction tip was blocked. This is considered as an acceptable deviation from the standard requirement as it does not decrease the safety or effectiveness of the subject device when compared to the predicate device.
Air leakage
Leakage into the collection container assembly was verified for Cary as per standard requirement for general use suction equipment. The results of the testing were within the acceptance criteria.
Vacuum levels and free air flows
Vacuum levels and free air flows for categories High vacuum/high flow and Low vacuum/low flow were tested for Cary. The results of the testing were within the acceptance criteria.
Page 9
Pharyngeal suction equipment
Ability of Cary to evacuate ≥ 200 ml of simulated vomitus in not more than 10 s was tested. The result of the test was within the acceptance criteria.
Reprocessing Validation
An assessment was conducted for the appropriateness of the manual reprocessing process and for the ability of the intended user to carry out the reprocessing of Cary drive unit per provided instructions (human factors validation testing including reprocessing instructions). The outcome of the validation was within the acceptance criteria.
Sterilization and Shelf-Life
Cary is not provided as sterile. Shelf-life is applicable only to a filter in Cary pumpister. The filter is not sterile but its shelf-life, determined by the supplier, defines the expiration date of the pumpister which is included the pumpister label. Shelf-life testing is not required for usage of the filter in Cary pumpister.
Biocompatibility
Biocompatibility of direct-contact material in Cary was confirmed by assessing the cytotoxicity, sensitization, irritation, acute systemic toxicity, and material mediated pyrogenicity testing in accordance with ISO 10993-1,-5,-10,-11,-18,-23: Biological Evaluation of Medical Devices.
Software
Software verification and validation were conducted, and documentation provided as recommended by FDA's Guidance "Content of Premarket Submissions for Device Software Functions -Guidance for Industry and Food and Drug Administration Staff". Cary software is classified as CLASS B under the Software Safety Classification per IEC 62304:2006+AMD1:2015 Medical device software - Software life cycle processes, and the software documentation level is "Basic Documentation Level" based on the Guidance "Content of Premarket Submissions for Device Software Functions -Guidance for Industry and Food and Drug Administration Staff".
EMC and Electrical Safety
The electromagnetic compatibility (EMC) and electrical safety (ES) of Cary were confirmed by the following standards: IEC 60601-1-2:2014+AMD1:2020; IEC 60601-1:2005+AMD1:2012+AMD2:2020; ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021]
Page 10
Conclusion
The result from the comparison between Cary and the predicate device LCSU 4 and from the test reports is that the subject device Cary performs acceptably and does not raise any different questions regarding safety or effectiveness, thus supporting a determination of substantial equivalence.
§ 878.4780 Powered suction pump.
(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.