Search Results
Found 30 results
510(k) Data Aggregation
(245 days)
Re: K233076
Trade/Device Name: Laser Speckle Imaging System (RFLSI CZW) Regulation Number: 21 CFR 870.2120
Extravascular blood flow probe Classification Name: Probe, Blood-Flow, Extravascular Regulation Number: 870.2120
Full-Field
Laser Perfusion
Imager | N/A |
| Regulation | 21 CFR Part 870.2120
| 21 CFR Part 870.2120 | Same |
| Product Code
The Laser Speckle Imaging System (RFLSI CZW) is intended for blood flow measurements in the micro-circulation. This device is intended for clinical research use.
The Laser Speckle Imaging System (RFLSI CZW) is intended for blood flow measurements in the micro-circulation. This device is intended for clinical research use. It is a measurement tool based on the laser speckle contrast analysis technology and provides real-time blood perfusion information of tissue and organs in a visual and quantitative way. The device is non-patient contacting and does not require the use of contrast agents.
The provided text describes the 510(k) premarket notification for the Laser Speckle Imaging System (RFLSI CZW) and its comparison to a predicate device. However, it does not contain detailed information about specific acceptance criteria and a study proving the device meets those criteria, particularly in the context of clinical performance or diagnostic accuracy. The document focuses on technological comparison, electrical safety, EMC, and laser safety testing.
Here's a breakdown of the information that is available and what is missing based on your request:
1. A table of acceptance criteria and the reported device performance
The document provides the following performance parameters for the subject device:
Feature | Acceptance Criteria (Subject Device) | Reported Device Performance (Subject Device) |
---|---|---|
Flux (Tissue perfusion) | Range: 0-5000 PU | Range: 0-5000 PU |
Resolution: 1 PU | Resolution: 1 PU | |
Accuracy: ± 10% | Accuracy: ± 10% | |
DC (Intensity) | Range: 0~255 AU | Range: 0~255 AU |
Accuracy: ± 1 AU | Accuracy: ± 1 AU | |
Resolution: 1 AU | Resolution: 1 AU |
Note: The document states that "Differences in device parameters do not raise new concerns regarding safety and effectiveness" and that "Verification and validation testing for the subject device demonstrate safety and effectiveness." It also mentions "Comparison tests to verify the substantial performance of the device and the predicate device were conducted... and the results conclude that the device shows comparable performance, safety, and effectiveness to the predicate device." However, specific numerical acceptance criteria for comparability in these comparison tests are not explicitly stated.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size:
- For the "laboratory testing model 'Flow Model' using a fluid simulator," the sample size is not specified.
- For the "volunteer test on the human body using post-occlusive reactive hyperemia method," the sample size is not specified.
- Data Provenance: Not specified. The document indicates the applicant and correspondent are in China, but it doesn't state where the testing data originated.
- Retrospective or Prospective: Not specified.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided in the document. The performance tests described (flow model and volunteer test) likely use objective physical measurements rather than expert assessment for ground truth, but the details are missing.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not provided.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This type of study is not mentioned in the document. The device is a Laser Speckle Imaging System for blood flow measurement, which typically outputs quantitative data directly, rather than images requiring human interpretation that AI might assist.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The performance testing described ("test with laboratory testing model 'Flow Model' using a fluid simulator" and "volunteer test on the human body") appears to evaluate the device's standalone performance in measuring blood flow. The document states that the testing demonstrated that "the device performs its intended purpose, and it meets the specified requirements and standards" and showed "comparable performance, safety, and effectiveness to the predicate device."
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- For the "Flow Model" testing, the ground truth would likely be the known flow rates or parameters of the fluid simulator.
- For the "volunteer test on the human body using post-occlusive reactive hyperemia method," the ground truth would be established by the physiological changes induced by the post-occlusive reactive hyperemia method, which creates a temporary blood flow cessation followed by a reactive increase. The device's measurements would be compared against the expected physiological response. The precise method of establishing 'ground truth' for comparison during these physiological tests is not detailed, but it's not expert consensus, pathology, or outcomes data in the traditional sense for diagnostic accuracy.
8. The sample size for the training set
This information is not applicable/not provided. The device described is a measurement system and not an AI/ML device that typically requires a large training set in the way a diagnostic imaging algorithm might. The document does not mention any machine learning components that would necessitate training data.
9. How the ground truth for the training set was established
This information is not applicable/not provided for the reasons stated in point 8.
Ask a specific question about this device
(125 days)
|
| Regulation Number | 21 CFR 876.1500 | 21 CFR 870.2120
The ActivSight Intraoperative Imaging System (ActivSight) is intended to provide real-time endoscopic fluorescence and near infrared imaging. ActivSight enables surgeons to visually assess vessels, blood flow, and related tissue perfusion using fluorescence and near infrared imaging, and at least one of the major bile ducts (cystic duct, common bile duct, or common hepatic duct) using fluorescence, all during minimally invasive surgery.
Fluorescence imaging of biliary ducts with ActivSight is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for stand-alone use for biliary duct visualization.
The ActivSight Intraoperative Imaging System (ActivSight) is an accessory to existing commercial surgical laparoscope systems, including cameras and video processor units. ActivSight provides real-time endoscopic fluorescence and near-infrared imaging. These imaging features allow surgeons to visually assess vessels, blood flow, and tissue perfusion (using fluorescence and near-infrared imaging), and to visually assess at least one of the major bile ducts (cystic duct, common bile duct, or common hepatic duct) using fluorescence. Fluorescence imaging is enabled through use of any commercially available Indocyanine Green (ICG). These visualization features are available for surgeons to use during minimally invasive surgery. ActivSight is intended to be used in a surgical environment.
ActivSight consists of the following reusable components:
- ActivSight Imaging Module, consisting of optics and sensing electronics. The imaging module attaches physically between the third-party laparoscope and the third-party imaging system camera. Reprocessing of this component requires cleaning and disinfection between uses.
- ActivSight Light Engine, equipment consisting of system electronics that provide laser fluorescence and near-infrared illumination, processing of video input from both the third-party Camera Control Unit (CCI) and the ActivSight lmaging Module, and outputs video to the surgical monitor. This component does not require reprocessing between uses.
- ActivSight Light Cable, component consisting of a Y-shaped bifurcated light a cable that connects to the third party white-light source, the ActivSight Light Engine, and the third party laparoscope light post (providing white-light illumination of the surgical site through the laparoscope). Reprocessing of this component requires cleaning and steam sterilization.
- ActivSight Sterilization Tray, stainless-steel sterilization tray designed to a properly secure the ActivSight Light Cable for disinfection and sterilization. Reprocessing of this component is accomplished in its use- cleaning and steam sterilization.
ActivSight consists of the following disposable components:
- ActivSight Sterile Drape, a sterilized plastic drape that is provided for each use = to provide a sterile barrier between the imaging module and the patient.
- -ActivSight Calibration Target, a sterilized paper imaging target containing a checkerboard pattern for calibration of the ActivSight imaging module prior to each use of the device.
The provided text describes the ActivSight Intraoperative Imaging System, its indications for use, and a comparison to predicate devices, but it does not contain explicit acceptance criteria or a detailed study proving the device meets those criteria, as typically found in clinical trial reports or validation studies.
The document primarily focuses on regulatory approval (510(k) submission) by demonstrating substantial equivalence to a legally marketed predicate device. While "Performance Data" is mentioned, it refers to compliance with general safety and performance standards (e.g., IEC standards for electrical safety, software lifecycle) and an animal study for comparative visualization, not a clinical study to establish quantitative performance against defined acceptance criteria.
Therefore, much of the requested information cannot be extracted directly from this document. However, I can provide the available information based on the text:
1. Table of Acceptance Criteria and Reported Device Performance
This information is not explicitly stated in the provided document. The document focuses on demonstrating compliance with general safety and performance standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 62366-1, ANSI AAMI IEC 62304, IEC 60825-1, IEC 60601-2-18) and a "design validation in an animal study." However, specific quantitative acceptance criteria for parameters like sensitivity, specificity, accuracy, or other clinical efficacy metrics are not provided, nor are the detailed results of such performance against these criteria.
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: Not specified for any clinical performance or validation study against specific acceptance criteria. The document mentions an "animal study" for design validation, but details like the number of animals or specific data points are omitted.
- Data Provenance: The "design validation" was conducted as an "animal study." No information on a human test set or its provenance (e.g., country of origin, retrospective/prospective) is provided.
3. Number of Experts Used to Establish Ground Truth and Qualifications
- Number of Experts: Not specified.
- Qualifications of Experts: Not specified.
4. Adjudication Method for the Test Set
- Adjudication Method: Not specified.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- MRMC Study Done: No, an MRMC comparative effectiveness study is not mentioned in the provided text. The document refers to an "animal study" for comparative visualization against a predicate's ICG mode, but this is not a human MRMC study to assess reader improvement with AI assistance.
- Effect Size of Human Reader Improvement: Not applicable, as no MRMC study was mentioned.
6. Standalone (Algorithm Only) Performance Study
- Standalone Performance Study Done: The document describes the device (ActivSight) and its capabilities, including "real-time endoscopic fluorescence and near-infrared imaging" and enabling surgeons to "visually assess vessels, blood flow, and related tissue perfusion." The "animal study" compared ActivSight Perfusion (speckle laser) and ActivSight ICG (fluorescence) modes against the predicate's ICG mode in terms of "comparatively visualizing blood flow and perfusion." This suggests evaluating the device's imaging capabilities (algorithm included), but it is not framed as a formal "standalone performance study" with defined metrics like sensitivity/specificity for disease detection, or an "algorithm only" evaluation completely separated from the hardware and human interpretation, as typically meant by standalone performance in AI/ML medical devices. The primary focus is on the device as a whole system providing real-time visualization.
7. Type of Ground Truth Used
- Ground Truth Type: For the "animal study," the ground truth for "comparatively visualizing blood flow and perfusion" is implied to be direct visual assessment by the comparing technologies (ActivSight's modes vs. predicate's ICG mode). The document does not specify an independent, objective ground truth established by pathology, outcomes data, or expert consensus beyond the visual comparison of the imaging techniques themselves.
8. Sample Size for the Training Set
- Training Set Sample Size: Not specified. The document does not mention details about algorithm training or associated datasets.
9. How the Ground Truth for the Training Set Was Established
- Ground Truth Establishment for Training Set: Not specified, as training set details are not provided.
Ask a specific question about this device
(262 days)
: K163339
Trade Device Name:SpectralMD DeepView Wound Imaging System 2.0 Regulation Number: 21 CFR 870.2120
Classification
Name: | Extravascular Blood Flow Probe |
| Reference: | 21 CFR §870.2120
Imager Viewing Software for desktop computer (Accessory) |
| Regulation | 21 CFR Part 870.2120
| 21 CFR Part 870. 2120 | 21 CFR Part 870.2120
The SpectralMD™ DeepView™ Wound Imaging System 2.0 is an optical imaging device intended for studies of blood flow in the microcirculation. The DeepView system is suitable for a wide variety of clinical applications including plastic surgery, diabetes, dermatology, vascular surgery, wound healing, neurology, neurosurgery and anesthetics. In particular, it can be used for measuring perfusion of healthy and injured skin including burn wounds, skin flaps (plastic and reconstructive surgery), chronic wounds, decubitus ulcers and diabetic ulcers.
The DeepView System 2.0 is a prescription device that utilizes the principles of non-contact photoplethysmography (PPG) to capture images of tissue blood perfusion. This is accomplished by measuring the optical properties of tissues and blood as they vary in response to changing hemodynamic conditions. The device's software combines real-time digital analysis based on the interaction of light with vascular tissues below the skin's surface to produce 2-D color images on a touch-screen display depicting relative blood perfusion. The DeepView System consists of a Camera Head with LED optics, an Articulating Arm for Camera Head positioning, a Touch-Screen Display for image viewing, and for accessing and interacting with the Graphical User Interface (GUI). All components are integrated on a Mobile Cart that houses the hardware/software, uninterruptable power supply (UPS), and allows for transport between use environments. The DeepView System 2.0 is AC powered with a backup UPS, and is for use in healthcare/hospital facilities.
The provided text describes the SpectralMD DeepView Wound Imaging System 2.0 and its 510(k) submission for substantial equivalence to predicates.
Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided document:
Crucially, the document states: "No clinical performance data were needed to support substantial equivalence." This means that the device's acceptance was not based on human reader studies, AI assistance, or the analysis of detailed clinical performance metrics like sensitivity, specificity, or AUC for diagnostic purposes. Instead, the focus was on demonstrating technological equivalence and safety.
Therefore, many of the typical criteria for evaluating AI/ML-based medical devices (especially those involving diagnostic performance improvement or standalone AI performance) are not applicable in this context, as the device is an imaging system measuring blood flow, not a diagnostic AI.
However, I can extract information related to the device's functional performance acceptance criteria and the engineering studies performed.
Acceptance Criteria and Reported Device Performance
The "acceptance criteria" here are framed around demonstrating equivalence to the predicate device in terms of fundamental functional performance related to blood flow detection and safety.
Acceptance Criterion (Implicit/Explicit) | Reported Device Performance (as stated in Summary of Testing) |
---|---|
Equivalent Detection of Pulsatile Fluid Flow | The DeepView 2.0 is "capable of detecting the pulsatile component of fluid flow in an equivalent manner to that of the primary predicate." Specific results: |
- Ability to identify the 2% alternating change (AC) modulation consistent with tissue-volume change from blood flow.
- Ability to identify AC modulations at various frequencies within human heart rate frequencies.
- Capability to detect fluid flow beneath the surface of an optically dense medium. These results "demonstrate that the DeepView System 2.0 performs in an equivalent manner to the original DeepView System (primary predicate)." |
| Equivalent Frequency Detection of Pulsatile Flow | (See above) Demonstrated ability to identify AC modulations at various frequencies within human heart rate range. |
| Equivalent Capability under Simulated Physiological Conditions | (See above) Demonstrated ability to detect fluid flow beneath the surface of an optically dense medium. |
| Electrical Safety | Met AAMI/ANSI ES60601-1:2005/(R) 2012 & A1:2012; IEC 60601-1:2005 (Third Edition) + CORR. 1:2006 + CORR. 2:2007. |
| Electromagnetic Compatibility (EMC) | Met IEC 60601-1-2 (2007)/(R) 2012. |
| Human Factors and Usability Engineering Compliance & Validation | Designed "in accordance with FDA guidance on human factors and usability engineering" and "subjected to usability testing validation." |
Study Details (Based on the provided text)
Given the stated "No clinical performance data were needed," the "study" is primarily a series of engineering and bench testing studies to demonstrate functional and safety equivalence, not a traditional clinical trial.
-
Sample size used for the test set and the data provenance:
- Test Set Sample Size: Not specified in terms of number of "cases" or "patients" as this was not a clinical performance study. The tests likely involved physical phantoms or controlled bench setups to simulate blood flow conditions.
- Data Provenance: Not applicable in the sense of patient data from specific countries. This was likely laboratory/bench testing.
- Retrospective/Prospective: Not applicable in a clinical sense. These were controlled engineering experiments.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. The ground truth for these engineering tests would be derived from the controlled experimental setup (e.g., known fluid flow rates, known material properties of optically dense media, precise electrical and EMC standards). No human "experts" establishing "ground truth" in terms of clinical interpretation were needed for these specific tests.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. This type of adjudication is for clinical ground truth establishment, which was not performed here. The "adjudication" would be based on instrument readings and engineering standards.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC study was done. The document explicitly states: "No clinical performance data were needed to support substantial equivalence." The device is an imaging system, and its software provides "specific wound modules for facilitating patient/wound documentation," but it's not described as having an AI component that assists human interpretation for a diagnostic outcome that would be subject to an MRMC study.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No standalone AI performance study was done in the diagnostic sense. The "algorithm" here processes light interaction with tissue to produce 2D perfusion images. The performance assessed was its ability to detect pulsatile flow equivalently to the predicate device, not its ability to make a diagnostic determination on its own.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- The "ground truth" for the functional tests was engineered and controlled experimental conditions (e.g., known AC modulation percentages, known frequency inputs, known optical properties of test media). For safety tests, it was adherence to recognized international electrical safety and EMC standards.
-
The sample size for the training set:
- Not applicable. The document describes a traditional medical device submission based on predicate equivalence, not an AI/ML device that requires a separate training set for algorithm development. While there's a "Proprietary Software Algorithm" for data analysis, there's no mention of a machine learning training phase or associated dataset.
-
How the ground truth for the training set was established:
- Not applicable. As no training set was described for an AI/ML model, no ground truth establishment for such a set is relevant here.
In summary, the DeepView System 2.0's acceptance was based on demonstrating technical and functional equivalence to its predicate device through bench and engineering testing, alongside adherence to safety and usability standards. It was not cleared as an AI-enabled diagnostic device requiring clinical performance studies with human readers or standalone AI performance metrics.
Ask a specific question about this device
(232 days)
Infrared Laser Doppler Imager, moorLDI2-VR Visible Red Laser Doppler Imager Regulation Number: 21 CFR 870.2120
| |
| Classification Name: | Extravascular blood flow probe
(21 CFR 870.2120
|
| Classification Name: | Extravascular blood flow probe
(21 CFR 870.2120
The moorLDI2 Laser Doppler Imager is intended for blood flow measurements in the microcirculation. It is intended to be used for clinical research applications and pre-clinical research applications.
The moorLDI2 is a device to perform non-invasive blood flow measurements in the microcirculation, for example skin, using the established laser Doppler technique to quantify movement of blood cells beneath the skin surface. The system scans a very low power laser beam across the tissue. Moving blood in the microvasculature causes a Doppler frequency shift of the laser light, which is photo detected and processed to generate a colour coded blood flow map, line by line. An in built CCD camera records a colour photograph to aid visualisation of the scan site.
The Moor Instruments Ltd moorLDI2 Laser Doppler Imager is a device intended for blood flow measurements in the microcirculation for clinical and pre-clinical research applications. The 510(k) summary provides information regarding its performance and comparison to a predicate device (moorLDI2-IR laser Doppler imager, K032841).
1. Table of Acceptance Criteria and Reported Device Performance:
The document doesn't explicitly state "acceptance criteria" in a pass/fail sense for a clinical study comparing the new device against specific benchmarks. Instead, it focuses on demonstrating substantial equivalence to the predicate device. The performance characteristics of the device itself are given.
Parameter | Acceptance Criteria (Equivalent to Predicate) | Reported Device Performance (moorLDI2) |
---|---|---|
Flux (Tissue perfusion) | Range: 0-5000 PU | Range: 0-5000 PU |
Accuracy: ±10% relative to Moor Instruments 'standard' moorLDI; ±3% of measurement from temperature controlled (22 ±1°C) motility standard | Accuracy: ±10% relative to Moor Instruments 'standard' moorLDI; ±3% of measurement from temperature controlled (22 ±1°C) motility standard | |
Conc (Concentration of blood flow) | Range: 0-5000 AU | Range: 0-5000 AU |
Accuracy: ±10%; ±3% of measurement value | Accuracy: ±10%; ±3% of measurement value | |
DC (Intensity) | Range: 0-5000 AU | Range: 0-5000 AU |
Accuracy: ±10%; ±3% of measurement value | Accuracy: ±10%; ±3% of measurement value | |
Maximum image resolution | 256 x 256 pixels | 256 x 256 pixels |
Scan speed | 4ms/pixel, 10ms/pixel, 50ms/pixel | 4ms/pixel, 10ms/pixel, 50ms/pixel |
Operating principle | Same as predicate device | Same as predicate device |
Image resolution options | Same as predicate device | Same as predicate device |
Laser classification | Same as predicate device (Class 3R) | Class 3R (per IEC 60825-1:2007) |
Scan head external dimensions | Same as predicate device | Same as predicate device |
2. Sample Size Used for the Test Set and Data Provenance:
The document states: "The moorLDI2 laser Doppler imager was tested in direct comparison to the predicate device using laboratory models and skin blood flow measurements on volunteers."
- Sample Size: Not explicitly provided. The number of laboratory models or volunteers used is not specified.
- Data Provenance: The study involved "laboratory models" and "skin blood flow measurements on volunteers." The country of origin for the data is implicitly the United Kingdom, where Moor Instruments Ltd is based. The data would be considered prospective as it involves new testing for the purpose of this submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:
This information is not provided in the document. The type of testing performed (laboratory models, skin blood flow measurements) suggests objective measurements rather than subjective expert consensus for ground truth establishment.
4. Adjudication Method for the Test Set:
This information is not applicable as the document describes direct comparison testing using objective measurements of performance characteristics rather than an expert-adjudicated test set in the traditional sense.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
An MRMC study was not conducted or described. The device is a measurement instrument, and the study focused on technical performance comparison rather than human reader interpretation with and without AI assistance.
6. Standalone Performance:
Yes, a standalone performance evaluation was done. The document provides detailed specifications and performance parameters like "Accuracy" for Flux, Concentration, and DC (Intensity), stating these accuracies relative to a standard. The study also concludes that the device performs "as well as or better than the predicate device," which implies standalone performance evaluation of the new device and its comparison to the predicate.
7. Type of Ground Truth Used:
The ground truth appears to be based on objective physical measurements from "laboratory models" and "skin blood flow measurements on volunteers." For parameters like Flux, Conc, and DC, accuracy is given relative to "Moor Instruments 'standard' moorLDI" and a "temperature controlled (22 ±1°C) motility standard." This suggests a reliance on established measurement standards and controlled experimental conditions to define the true values.
8. Sample Size for the Training Set:
This information is not provided and is likely not applicable. The device is not described as an AI/ML algorithm that requires a training set in the conventional sense. It's a measurement instrument, and its development would involve engineering design and calibration rather than AI model training.
9. How the Ground Truth for the Training Set Was Established:
This information is not provided and is not applicable, as there is no mention of an AI/ML component requiring a training set with established ground truth.
Ask a specific question about this device
(105 days)
GB
Re: K142932
Trade/Device Name: Deltex Medical KDP72 Doppler Probe Regulation Number: 21 CFR 870.2120
Extravascular blood flow probe
The FDA has classified: "Extravascular blood flow probe" in 21 CFR 870.2120
The probe is for use with the Deltex Medical CardioQ-EDM and CardioQ-EDM+ for Monitoring of cardiac output and fluid status. The probe is only approved for oral placement into the esophagus of a single anesthetized patient 15 years of age or younger, 50cm (20") to 170cm (67") in height.
The Deltex Medical Ltd KDP72 Pediatric Doppler Probe is an oral extravascular blood flow probe designed to work with the CardioQ-EDM and CardioQ-EDM+ Systems (K111542 and K132139 respectively). It consists of a shaft, which is a spring reinforced silicone tube, with an electrical connector on the machine end and an ultrasonic transmitting and receiving tip on the patient end. The tip is fully covered and sealed to the shaft with a silicone rubber boot and by wires running through the shaft to the connector. Visual product identification is provided at the machine end and the device is provided single packed sterile for single patient use.
The provided document is a 510(k) summary for the Deltex Medical KDP72 Doppler Probe, asserting its substantial equivalence to previously cleared devices. It describes the device, its intended use, and performance data related to its design and safety standards, but does not contain information about a study proving the device meets specific acceptance criteria related to its clinical performance as a diagnostic tool for cardiac output and fluid status.
The "Performance Data" section briefly mentions:
- "The performance data recommended in 'Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers,' issued on September 9, 2008, has been included." This suggests that general requirements for diagnostic ultrasound devices were addressed, likely related to acoustic output and image quality, not the accuracy of cardiac output measurements.
- "Additionally a flexibility test has been conducted on the subject and predicate devices which demonstrates the comparative flexibility." This is a mechanical performance test, not a clinical diagnostic performance study.
Therefore, many of the requested categories for acceptance criteria and a study to prove they are met cannot be extracted from this document, as the document focuses on demonstrating substantial equivalence based on design, materials, sterilization, biocompatibility, and electrical safety standards rather than clinical diagnostic accuracy.
However, based on the provided text, here’s what can be extracted:
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria Category | Acceptance Criteria | Reported Device Performance (KDP72 Doppler Probe) |
---|---|---|
Sterilization | Sterilized in accordance with ISO 11135-1 (version of standard current at time of submission). | Meets ISO 11135-1 (version current at submission). |
Shelf Life | Meets ISO 11607-1 (version of standard current at time of submission). | Meets ISO 11607-1 (version current at submission). |
Biocompatibility | Tested in accordance with ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Irritation and Skin Sensitization), ISO 10993-7 (Ethylene Oxide Sterilization Residuals) (version of standard current at time of submission). Note: ISO 10993-7 is about ETO residuals, not a direct biocompatibility test type, but often included for ETO sterilized devices. | Meets ISO 10993-5, ISO 10993-10, ISO 10993-7 (versions current at submission). |
EMC and Electrical Safety | Meets IEC 60601-1 (Medical Electrical Equipment - General requirements for basic safety and essential performance), IEC 60601-1-2 (Electromagnetic disturbances - Requirements and tests), IEC 60601-2-37 (Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment) (version of standard current at time of submission). | Meets IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37 (versions current at submission). |
Packaging | Meets ISO 11607-1 (Packaging for terminally sterilized medical devices - Requirements for materials, sterile barrier systems and packaging systems), ISO 11607-2 (Packaging for terminally sterilized medical devices - Validation requirements for forming, sealing and assembly processes) (version of standard current at time of submission). | Meets ISO 11607-1, ISO 11607-2 (versions current at submission). |
Flexibility | Demonstrates comparative flexibility to predicate devices. (No specific quantitative acceptance criterion stated). | Flexibility test conducted; demonstrates comparative flexibility. |
Diagnostic Ultrasound Performance Data | Based on "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers", issued on September 9, 2008. (Likely relates to acoustic output, measurement accuracy of general ultrasound parameters, not necessarily cardiac output accuracy specifically). | "Included" (no specific results provided). |
2. Sample size used for the test set and the data provenance:
Not specified in the document for any clinical performance or diagnostic accuracy study. The document mentions a flexibility test, but details on sample size or provenance are not provided.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable/not specified. The document does not describe a study involving expert-established ground truth for clinical diagnostic performance.
4. Adjudication method for the test set:
Not applicable/not specified.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This device is a probe for a cardiac output monitor, not an AI-assisted diagnostic imaging system that involves human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not specified. The document refers to "performance data recommended in 'Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers'" but does not detail what kind of study was performed or if it was standalone. The device is a probe to be used with a monitoring system, implying a system performance, not a standalone algorithm.
7. The type of ground truth used:
Not specified for diagnostic accuracy. For the engineering criteria (sterilization, biocompatibility, etc.), the ground truth is established by adherence to the respective international standards (e.g., ISO, IEC).
8. The sample size for the training set:
Not applicable/not specified. No information on an algorithm training set is provided; this is a medical device (probe) for measurement, not an AI/ML diagnostic algorithm.
9. How the ground truth for the training set was established:
Not applicable/not specified.
Ask a specific question about this device
(111 days)
Blood-Flow, Extravascular
Class II, 21 CFR 870.2120
DPT
Name of Predicate Device:
Bio-Probe Blood
Bio-probe Blood-Flow Transducer, models TX50 (adult) and TX50P (pediatric)
Regulation Number: 21 CFR 870.2120
The Bio-Probe blood flow monitoring system is to be used with an appropriate model Bio-Console extracorporeal blood pumping console to measure directly the blood flow in the extracorporeal perfusion circuit.
The Bio-Probe 10 Blood Flow Monitoring System consists of a flow transducer and a sterile, single-use insert. The flow transducer consists of a flow-meter, cable and connector. The TX50 (adult) and TX50P (pediatric) transducer models are reusable. The Bio-Probe blood flow monitoring system can be used to measure the patient blood flow during the extracorporeal procedure.
The provided document is a 510(k) summary for the Bio-Probe Blood Flow Transducer. This particular submission concerns the addition of a contraindication statement to the device.
Therefore, the document explicitly states that "Testing was not required for addition of a contraindication statement. Addition of the contraindication statement does not change the indications for use, technology and performance specifications of this device." This means there is no study described in this document proving the device meets acceptance criteria, as the submission is not about performance.
Here's the breakdown based on your request, highlighting the lack of relevant information for a performance study in this specific submission:
1. A table of acceptance criteria and the reported device performance
- Acceptance Criteria: Not applicable/Provided.
- Reported Device Performance: Not applicable/Provided.
Explanation: This 510(k) is solely for adding a contraindication statement, not for demonstrating new performance or design changes.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size (Test Set): Not applicable.
- Data Provenance: Not applicable.
Explanation: No testing was performed for this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Number of Experts: Not applicable.
- Qualifications of Experts: Not applicable.
Explanation: No ground truth establishment was needed as no performance testing was conducted.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Adjudication Method: Not applicable.
Explanation: No adjudication was needed as no performance testing was conducted.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Study: No.
- Effect Size: Not applicable.
Explanation: This device is a blood flow transducer, not an AI-assisted diagnostic or interpretation device that would involve human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Standalone Performance: Not applicable.
Explanation: This device is a physical transducer for measuring blood flow, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Type of Ground Truth: Not applicable.
Explanation: No ground truth was established as no performance testing was conducted.
8. The sample size for the training set
- Training Set Sample Size: Not applicable.
Explanation: This submission does not involve an algorithm or machine learning that would require a training set.
9. How the ground truth for the training set was established
- Ground Truth Establishment (Training Set): Not applicable.
Explanation: No training set or ground truth for it was established.
In summary, this 510(k) submission is a regulatory update for an existing device (Bio-Probe Blood Flow Transducer, predicate device K070286) to add a contraindication statement. It explicitly states that no testing was required or performed because the change does not impact the device's indications for use, technology, or performance specifications. Therefore, information regarding acceptance criteria and performance study details is not present in this document.
Ask a specific question about this device
(201 days)
| |
| Classification Name: | Extravascular blood flow probe, DPT, 21 CFR 870.2120
2014
Re: K132163
Trade/Device Name: MoorLDLS-BI laser-doppler burns imager Regulation Number: 21 CFR 870.2120
The moorLDLS-BI laser Doppler burns imager assesses the blood flow in burn wounds of the skin, when cleaned of surface debris, to aid in the clinician's assessment of burn wound healing potential. It is intended to be used as an aid to burn wound management for patients with Total Body Surface Area burn of up to 30%.
The device is intended to be used as an aid to burn wound assessment, and not as a stand-alone prediction device.
The moor DLS-BI laser Doppler burns imager is an imaging device to aid the clinician to judge the healing potential of bums and the need for surgery.
It uses the laser Doppler imaging technique to quantify the blood flow in an area of skin damaged by a burn. The device uses a line of laser light projected onto the tissue and a linear detector arrav that sample from the line as it is swept across the tissue to rapidly build up a colour coded image of blood flow in the burn area and the surrounding normal skin for healing potential prediction. In addition, a CCD camera is integrated into the scanner unit for recording a colour photograph at the time of scanning, corresponding closely with the blood flow image in size and aspect.
The moorLDLS-BI laser Doppler burns imager is a substantial equivalent to the predicate device moorLDI2-B1 and assesses burn wound healing potential.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the clinical investigation's objectives and the comparative analysis against the predicate device. The primary performance metrics are accuracy in predicting healing potential and agreement with the predicate device.
Acceptance Criteria / Performance Metric | Reported Device Performance (moorLDLS-BI) | Notes |
---|---|---|
Overall accuracy in predicting burn wound healing potential (compared to healing records) | 94.2% | This is a standalone performance metric for the moorLDLS-BI. |
Agreement with predicate device (moorLDI2-BI) for HP14 (healing in 21 (healing in >21 days) | 98.5% | Demonstrates close correlation and substantial equivalence for predicting long healing times. |
Non-inferiority to the predicate device (moorLDI2-BI) | Unlikely to perform more than 1.7% worse. (In fact, performed slightly better) | This indicates that the moorLDLS-BI is at least as effective as the predicate device, fulfilling a key aspect of substantial equivalence. |
Electrical safety, laser radiation safety, electromagnetic compatibility, and programmable medical device standards conformity | Designed and tested for compliance with standards | Although specific compliance percentages or detailed results aren't provided, this statement confirms that regulatory safety criteria were met. |
Good correlation for tissue blood flow measurement (bench testing) | Good correlation between moorLDLS-BI and predicate device in flow model and normal skin tissue scan results | This bench testing supports the foundational equivalence of the physiological measurement principle between the two devices before clinical application. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: 596 burn cases for 204 burn patients.
- Data Provenance: Multi-center clinical investigation conducted in five burns centers across various countries: 2 in the UK, 1 in the USA, 1 in Belgium, and 1 in Australia. The study was prospective in nature, as it was a "Clinical Investigation" designed to assess the performance of a new device.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts
The document does not explicitly state the number of experts used to establish the ground truth or their specific qualifications (e.g., radiologist with X years of experience).
However, given the nature of "healing records" being the ground truth and the device's intended use as an aid to "clinician's assessment of burn wound healing potential" by "burn surgeons," it is highly probable that the ground truth was established by burn care clinicians/surgeons responsible for patient management and outcome tracking.
4. Adjudication Method for the Test Set
The document does not explicitly state the adjudication method used for establishing the ground truth from "healing records."
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
- No, a MRMC comparative effectiveness study was not explicitly done for human readers. The study focused on comparing the performance of the device (moorLDLS-BI) against the predicate device (moorLDI2-BI) and against healing records, not on comparing human reader performance with and without AI assistance.
- Effect size of human reader improvement: Not applicable, as this type of study was not conducted as described. The device is intended as an "aid to burn wound management," implying human oversight, but the study directly compares device performance, not human-AI synergy.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Yes, a standalone performance assessment was done. The "overall accuracy of 94.2% was found for moorLDLS-Bl when compared with healing records" represents the device's standalone performance in predicting healing potential.
- It's important to note that while the device has standalone accuracy, its intended use statement explicitly says, "It is intended to be used as an aid to burn wound assessment, and not as a stand-alone prediction device," indicating that clinical interpretation by a human is still required.
7. The Type of Ground Truth Used
- Outcomes Data (Healing Records): The primary ground truth for the overall accuracy of the moorLDLS-BI was based on "healing records." This refers to the actual observed healing time of the burn wounds, which is a direct patient outcome.
8. The Sample Size for the Training Set
The document does not explicitly state the sample size for a training set. The description mentions a "clinical investigation" and performance data, but does not distinguish between a training set and a testing set in the context of machine learning model development. This suggests the moorLDLS-BI, while a new device, might not be heavily reliant on a trainable "algorithm" in the modern AI sense, but rather on a refined physical measurement principle and associated processing algorithms for which the mentioned clinical investigation serves as a validation/test set.
9. How the Ground Truth for the Training Set Was Established
Since a distinct "training set" is not mentioned or detailed, the method for establishing its ground truth is also not described. If any internal development process involved a training phase, that information is not part of this 510(k) summary.
Ask a specific question about this device
(91 days)
Codes: | DPW, DPT, DSA |
| Regulation Numbers: | 21 CFR 870.2100, 870.2120
|
| Device Classification | 21 CFR 870.2100, 870.2120
| 21 CFR 870.2100, 870.2120
The CardioQ-EDM+ cardiac function and fluid status monitoring system is designed to provide clinicians with real-time information about a patient's left ventricular blood flow and key hemodynamic parameters. The CardioQ-EDM+ beat-to-beat data on cardiovascular status can be used by the managing clinician to evaluate and optimize hemodynamic performance in anesthetized, sedated or conscious patients in the operating room, intensive care unit, emergency room or ward.
The CardioQ-EDM+, is a medical instrument designed to monitor cardiac function and fluid status, providing clinicians with real-time information about a patient's left ventricular blood flow and key hemodynamic parameters in anesthetized, sedated or conscious patients in the operating room, intensive care unit, emergency room or ward. In addition, the CardioQ-EDM+ includes a function to calculate arterial blood pressure based parameters from an output slaved from a vital signs monitor. The CardioQ-EDM+ combines Doppler measurement of blood flow with Pulse Pressure Waveform Analysis (PPWA). In "Flow Monitoring Mode" the system employs esophageal Doppler monitoring (EDM) techniques using 4 MHz continuous wave ultrasound to monitor and quantify the blood flow in the descending thoracic aorta. displaying this data as a maximum velocity curve, a velocity spectrum and derived measurements. Real-time information about cardiac function, in particular left ventricular flow, is displayed continuously. For the CardioQ-EDM+, the newly added "Pressure Monitoring Mode" the system slaves the arterial blood pressure signal supplied by the hospital patient monitoring system to provide systolic and diastolic pressures and derived parameters. The CardioQ-EDM+ uses these classical blood pressure measurements to calculate Stroke Volume (SV), Cardiac Output (CO), Stroke Volume Variation (SVV). Pulse Pressure Variation (PPV) and a small number of derived parameters. The pressure derived stroke volume is calibrated from the CardioQ-EDM+'s Doppler ultrasound measurement of stroke volume ensuring consistency and allowing frequent recalibration.
Here's a breakdown of the acceptance criteria and study information for the Deltex Medical CardioQ-EDM+ device, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The provided 510(k) summary does not contain a specific table detailing acceptance criteria for the CardioQ-EDM+ device's performance in terms of accuracy, precision, sensitivity, or specificity. Instead, the submission focuses on demonstrating substantial equivalence to a predicate device (CardioQ-EDM) by highlighting that the new device is a modification incorporating additional functionalities for blood pressure-based parameter calculation.
The performance comparison is implicitly made through the "Comparison technological features" table, which shows that the CardioQ-EDM+ offers the same ranges for directly measured Doppler parameters as the predicate device. The new "Pressure Monitoring Mode" adds new calculated parameters with specified ranges, but these are not compared against a defined acceptance criterion from a performance study within this document.
Implicit Performance Claim (based on substantial equivalence):
The CardioQ-EDM+ is designed to provide real-time information about a patient's left ventricular blood flow and key hemodynamic parameters, similar to the predicate device. The new pressure-based calculations are presented as additions to the existing capabilities, relying on the established performance of the Doppler measurement and the slaved arterial blood pressure signal.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size:
- Clinical Testing: The document explicitly states: "No clinical testing was conducted in support of this 510(k)." Therefore, there is no clinical test set sample size.
- Bench Testing: The document mentions "Comparative bench testing," but does not provide details on the number of samples, cases, or specific data points used in this testing.
- Data Provenance: Not applicable for clinical data. For bench testing, the provenance is internal to Deltex Medical, as it was "Comparative bench testing... using a CardioQ-EDM Cardiac Output and Fluid Status Monitoring System (K111542)."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Since "No clinical testing was conducted in support of this 510(k)," there was no ground truth established by experts for a clinical test set.
4. Adjudication Method for the Test Set
Not applicable, as no clinical test set was used to establish performance against a ground truth.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Was it done? No. The document explicitly states: "No clinical testing was conducted in support of this 510(k)."
- Effect size of human readers improvement: Not applicable, as no such study was performed.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
While the device operates as a standalone monitoring system, the term "standalone study" typically refers to an evaluation of an AI algorithm's performance independent of human input or review. As this device is a medical instrument (hardware and integrated software) designed to measure physiological parameters rather than an AI-driven diagnostic or interpretative tool, this concept doesn't directly apply in the same way it would for, for instance, an AI-powered image analysis product.
The device's performance is demonstrated through its ability to measure and calculate parameters, and the validation of these calculations would be part of the bench testing. The 510(k) focuses on the "substantial equivalence" of the device, particularly the new features, to the predicate. The "Bench Testing" section confirms that the new device's performance is substantially equivalent to the predicate, implying that the algorithm's calculations were validated against expected outputs or the predicate's known outputs, but specific details of this validation are not provided.
7. Type of Ground Truth Used
- Clinical: Not applicable, as no clinical testing was performed.
- Bench Testing: The implicit ground truth for the bench testing would have been reference measurements or known values used to compare the CardioQ-EDM+ against the predicate CardioQ-EDM. The document states: "Comparative bench testing... supports the conclusion that the two systems have substantially equivalent performance." This implies that the CardioQ-EDM+'s outputs for new and existing parameters were compared against established and accepted values (e.g., from the predicate device or simulated physiological signals) during bench testing.
8. Sample Size for the Training Set
This device does not appear to use machine learning or AI in a way that requires a "training set" in the conventional sense (e.g., for image recognition or predictive modeling from a large dataset). Its algorithms are based on established physiological principles and signal processing for Doppler measurements and arterial pressure analysis. Therefore, there's no reported training set sample size.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no apparent training set for machine learning. The algorithms are likely deterministic and based on validated mathematical models and physiological relationships.
Ask a specific question about this device
(108 days)
| |
| Classification Name: | DPT 870.2120
Device Name: DeepView Digital Video Physiological Portable Imaging System Regulation Number: 21 CFR 870.2120
The Spectral MD DeepView system is intended for studies of blood flow in the microcirculation. The DeepView system is suitable for a wide variety of clinical applications including plastic surgery, diabetes, dermatology, vascular surgery, wound healing, neurology, physiology, neurosurgery and anesthetics.
Deep View system-based technology combines real-time digital analysis of optical signatures, thereby sensitizing an imager to photon-tissue interactions deep below the skin's surface. These image signatures are unique to the body and relate directly to a person's dynamic nature - both in terms of the quantity and quality of important physiological properties. This technology is non-invasive and uses no harmful radiation such as X-rays and allows clinical investigators to look deeper into the body, delivering images of blood flow under the skin's surface without ever touching the patient.
The DeepView system is composed of a mobile cart with uninterruptible power supply, a laptop computer with remote multimedia keyboard, an LCD screen mounted on a bracket that allows for side-to-side panning, a mechanical arm, a CMOS camera with DSP electronics, and disposable LED cartridges with an associated LED driver control board.
The provided document describes the DeepView Digital Video Physiological Portable Imaging System, a device intended for studies of blood flow in the microcirculation. It focuses on demonstrating substantial equivalence to predicate devices rather than establishing novel acceptance criteria and proving performance against them in a de novo study.
Here's an analysis based on the available information:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly define acceptance criteria in terms of specific performance metrics (e.g., sensitivity, specificity, accuracy, or quantitative blood flow measurements) that the DeepView device needed to meet. Instead, the study's goal was to demonstrate substantial equivalence to existing predicate devices.
Acceptance Criterion (Implicit) | Reported Device Performance |
---|---|
Ability to detect blood flow optically | DeepView uses optical methods to detect blood flow and pulse pressure. |
Ability to detect pulse frequency | DeepView was tested alongside predicate devices to show substantial equivalence in detecting pulse frequency. |
Ability to produce flow images | DeepView displays 2D color images demonstrating relative blood flow, similar to moorLDI and moorLDI2-IR. |
No new issues of safety and efficacy compared to predicate devices | Comparison testing conducted demonstrates that the DeepView is substantially equivalent and does not introduce any new issues of safety and efficacy. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not specified. The document mentions "comparison testing" which included "camera distance testing and tissue phantom testing." This implies the use of a controlled test environment (tissue phantoms) rather than human subjects. No specific number of phantoms or test cases is provided.
- Data Provenance: The testing appears to be conducted in a laboratory setting using "tissue phantoms." There is no mention of human data, country of origin, or whether it was retrospective or prospective.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable in the context of this submission. The ground truth for the comparison testing seems to be the performance of the predicate devices themselves, as the DeepView's output was compared to theirs. There's no indication of independent expert review to establish a separate "ground truth" for the test set.
4. Adjudication Method for the Test Set
Not applicable. There is no mention of human-in-the-loop assessment or expert adjudication for the "comparison testing." The comparison was against the output of the predicate devices.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, an MRMC comparative effectiveness study involving human readers and AI assistance was not conducted or reported. This submission focuses on the standalone device's equivalence to existing technology, not on improving human reader performance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done
Yes, the testing described appears to be a standalone performance evaluation of the DeepView device against the predicate devices. The "comparison testing" suggests the DeepView's output was directly compared to the outputs of the moorLDI, moorLDI2-IR, and Avant 9600. The device's ability to "detect blood flow and pulse pressure" and produce "2D color images of relative perfusion distribution" constitutes its standalone performance.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the comparison testing was effectively the performance of the predicate devices. The DeepView was evaluated for its ability to produce similar results (detect blood flow, pulse frequency, and produce flow images) to the already legally marketed and accepted predicate devices (moorLDI, moorLDI2-IR, and Avant 9600). For the tissue phantom testing, the "ground truth" would implicitly be the known properties of the phantoms and the expected measurements, as validated by the predicate devices.
8. The Sample Size for the Training Set
Not applicable. As this is a 510(k) premarket notification for a device using established optical principles, there is no mention of an "AI algorithm" requiring a training set in the contemporary sense. The device's operation is based on "real-time digital analysis of optical signatures" and "non contact Photoplethysmography (PPG)," which points to signal processing rather than machine learning models that require training data.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no mention of a training set or an AI algorithm that would require one.
Ask a specific question about this device
(115 days)
| Class II |
| Classification name: | 21 CFR 870.2120
Switzerland 1015
Re: K123216
Trade/Device Name: EasyLDI Microcirculation Camera Regulation Number: 21 CFR 870.2120
The Aïmago EasyLDI Microcirculation Camera is intended for blood flow measurements in the microcirculation. In particular, it can be used for measuring perfusion of healthy and injured skin including burn wounds, skin flaps (plastic and reconstructive surgery) and hand surgery.
EasyLDI Studio is intended to be used as offline viewer application for snapshots, videos and references recorded with the Aïmago EasyLDI Microcirculation Camera.
The Aïmago EasyLDI microcirculation camera is a device for imaging blood flow in the microcirculation. It is a medical diagnostic imaging device which serves to visualize the perfusion of cutaneous microcirculation in the form of arbitrary units in real-time. The EasyLDI uses the established laser Doppler technique performing a 2-dimensional area scan to build up a color coded image of the blood flow in the tissue. In the form of arbitrary units, this image allows the surgeon to quantify movement of blood cells beneath the skin surface.
The software changes implemented in the Aïmago EasyLDI microcirculation camera V2.X allow the user different modes of displaying the information on the built-in screen, thus facilitating the assessment of the microcirculation patterns for the specified applications.
EasyLD! Studio is a standalone software which runs on Windows systems. It is an optional accessory to the Aïmago EasyLDI microcirculation camera. It can be used to view LDI items (i.e. LDI snapshots, videos or references previously recorded with the Aïmago EasyLDI) on a commercially available desktop computer.
The provided text describes a 510(k) summary for the Aïmago EasyLDI Microcirculation Camera and its accessory, EasyLDI Studio. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study with specific acceptance criteria and detailed device performance metrics in the way a clinical trial would.
Therefore, many of the requested details about acceptance criteria, sample sizes, expert involvement, and ground truth establishment are not present in this type of regulatory submission. This submission primarily relies on comparing technological characteristics and demonstrating safety and effectiveness based on in-house and contract laboratory testing for regulatory compliance (e.g., FCC, IEC standards).
Here's a breakdown of what can be extracted and what is not available based on the provided text:
Acceptance Criteria and Device Performance
The document does not specify performance-based acceptance criteria in the typical sense of a clinical or functional study (e.g., sensitivity, specificity, accuracy for a particular clinical outcome). Instead, "acceptance criteria" here refer to regulatory compliance and equivalence:
Acceptance Criteria (Regulatory/Equivalence) | Reported Device Performance (Compliance) |
---|---|
Substantial Equivalence to Predicate Device (K121449) | Cleared as substantially equivalent. Changes do not adversely affect safety and effectiveness. |
FCC Rules for Digital Devices (Subpart B of Part 15 for Class A) | Fulfills the requirements. |
ESD safety (IEC 60601-1-2) | Fulfills the requirements. |
Electromagnetic immunity (IEC 60601-1-2) | Fulfills the requirements. |
Electrical safety requirements (IEC 60601-1) | Fulfills all requirements. |
Ability to display information in different modes (V2.X software) | Allows user different modes of displaying information, facilitating assessment of microcirculation. |
EasyLDI Studio: Ability to view LDI items (snapshots, videos, references) | Intended to be used as offline viewer application for snapshots, videos, and references recorded with the Aïmago EasyLDI Microcirculation Camera. |
Study Details
2. Sample size used for the test set and the data provenance
- Sample Size: Not specified for any performance testing related to clinical application. The testing mentioned refers to regulatory compliance tests (e.g., FCC, IEC), for which sample sizes for device hardware/software are typically small and not relevant to clinical data.
- Data Provenance: Not applicable in the context of clinical data. The testing mentioned ("performed in-house as well as at contract laboratories") is for regulatory compliance, not clinical data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. No ground truth for a test set in a clinical context is mentioned. Clinical assessments are explicitly stated to be "aid to healthcare professionals," and the device "do not provide specific clinical assessments such as burn depth assessments or potential healing times."
4. Adjudication method for the test set
- Not applicable. No clinical test set or adjudication process is described.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, an MRMC comparative effectiveness study was not done or reported. This device precedes the widespread use of clinical AI assistance in this context. The software changes are described as facilitating assessment, but no studies on reader improvement are mentioned.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- The device's core function is to visualize and measure blood flow, which is then interpreted by a human professional. The "new software applications are intended only as an aid to healthcare professionals in their clinical assessments." This implies a human-in-the-loop scenario. No standalone algorithm performance without human interpretation is described for clinical outcomes. The device itself is a standalone imaging device, but its utility for "assessment" relies on the human user.
7. The type of ground truth used
- Not applicable for clinical efficacy. The "ground truth" for the regulatory compliance testing would be established by the standards themselves (e.g., an ESD test either passes or fails according to IEC 60601-1-2 criteria). The device's output is "arbitrary units" of blood flow, which doesn't directly map to a "ground truth" in the sense of a definitive diagnosis or pathology.
8. The sample size for the training set
- Not applicable. This device is cleared based on predicate equivalence and compliance with engineering standards, not through machine learning or AI model development that would require a training set.
9. How the ground truth for the training set was established
- Not applicable, as no training set for a machine learning model is mentioned.
Ask a specific question about this device
Page 1 of 3