K Number
K244061

Validate with FDA (Live)

Device Name
X-Wire Guidewire
Date Cleared
2025-08-26

(238 days)

Product Code
Regulation Number
870.1330
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The X-Wire Guidewire is indicated for general intravascular use within the peripheral and neuro vasculature to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

Device Description

The Imperative Care's X-Wire Guidewires are guidewires with shapeable tips to aid in accessing the peripheral and neuro vasculature. The guidewires are available in 200cm - 300cm lengths, standard (S) and support (T) stiffness profiles, and 0.014", 0.018" and 0.024" diameters. The distal portion of the guidewire tip includes a radiopaque marker to facilitate fluoroscopic visualization. A hydrophilic coating on the distal segment and PTFE coating on the proximal segment serve to reduce friction during manipulation in vessels. The X-Wire Guidewire is supplied with a shaping mandrel, introducer, and torque device.

AI/ML Overview

This document is a 510(k) clearance letter for a medical device and therefore does not contain acceptance criteria and study data in the format typically found in clinical efficacy studies for AI/software devices. The acceptance criteria and "study" described herein relate to the safety and performance of the physical guidewire device (X-Wire Guidewire) based on bench testing, biocompatibility testing, sterilization validation, and shelf-life testing, rather than a clinical study evaluating the performance of an AI algorithm.

Here's a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document summarizes the performance testing done in Table 2: Summary of Bench Tests and Performance Specifications and Table 3: Biocompatibility Test Summary – X-Wire Guidewire.

Table: Acceptance Criteria and Reported Device Performance (Synthesized from document)

Test AttributeAcceptance Criteria (Specification)Reported Device Performance (Results)
Dimensional VerificationAll defined guidewire dimensions are within the specified tolerances.Pass
Visual InspectionThe guidewire shall be free of visual defects when removed from packaging.Pass
TorqueabilityThe guidewire shall transmit rotation from the proximal end to the distal tip to allow users to select branches of the vasculature and reach the target location.Pass
Torque StrengthDevice shall not fail under expected torsional input when distal end is unable to move/rotate.Pass
Tip FlexibilityTip of guidewire shall not cause vessel damage; defined by ability to buckle to reduce contact pressure and prevent perforation.Pass
Tensile Strength and Tip PullGuidewire shall withstand tensile forces expected in clinical use without breaking.Pass
Kink ResistanceGuidewire shall be able to traverse through clinically relevant bends without kinking.Pass
Coating IntegrityCoating shall remain intact during clinical use.Pass
Coating LubricityCoating shall be lubricious to reduce frictional forces for navigation.Pass
Particulate EvaluationGuidewire shall not generate particles at a level greater than the range generated by the predicate.Pass
Corrosion ResistanceGuidewire shall not corrode from time of manufacture through its shelf life.Pass
RadiopacityOpacity to x-ray shall allow physicians to visualize guidewire under fluoroscopy.Pass
Simulated UseGuidewire must be able to reach anatomical locations and deliver catheters and other interventional devices used in common neurovascular procedures.Pass
CytotoxicityTest article extracts must yield grade 2 or lower.Pass, Noncytotoxic
SensitizationOverall pattern, intensity, duration, character of reactions compared to control conditions.Pass, Nonsensitizer
Irritation or Intracutaneous ReactivityDifference between test extract mean score and corresponding control mean score ≤1.Pass, Nonirritating
Acute Systemic ToxicityNone of the animals treated with test article extracts must show significantly greater biological reactivity than control-treated animals.Pass, Non-toxic (acute systemic)
Material Mediated PyrogenicityTest article extract must yield <0.5°C rise of individual animal baseline temperature.Pass, Nonpyrogenic
Hemocompatibility – HemolysisTest article extract must result in hemolytic index of <2%.Pass, Nonhemolytic for both Direct and Extract Methods
Hemocompatibility - SC5b-9 Complement ActivationSC5b-9 concentration of the test article is similar to both the activated NHS and negative controls.Pass, Hemocompatible
Hemocompatibility – In vivo thromboresistanceTest article performs similarly to the predicate.Pass, Hemocompatible
Sterilization Assurance Level1 x 10^-6 (implicit)Achieved (via validated EO process)
Shelf Life (Accelerated Aging)Device, accessory, and packaging performance maintained for stated shelf life.Supports 8-month shelf life.

Important Note: This document describes a physical medical device (guidewire), not a software/AI device. Therefore, many of the subsequent questions related to AI studies (test sets, ground truth, experts, MRMC studies, standalone performance, training sets) are not applicable to this 510(k) submission.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: Not explicitly stated as a number of "cases" or "patients" in the typical sense of a clinical study. The testing here refers to physical guidewire samples subjected to bench tests and biological evaluations. The specific number of devices tested for each attribute (e.g., number of guidewires for torqueability, number of animal subjects for biocompatibility) is not detailed in this summary, but would be found in the full test reports.
  • Data Provenance: N/A for clinical data. The data is generated from laboratory (bench and biocompatibility) testing of the physical device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Number of Experts: Not applicable. Ground truth for the physical device performance (e.g., whether a guidewire kinks or corrodes) is established by objective measurements and standardized test protocols, not by expert consensus in a clinical diagnostic sense. However, for the biocompatibility testing, an "expert evaluation" is mentioned as confirming that the device met biological safety requirements. The number and qualifications of this expert(s) are not specified.

4. Adjudication Method (for the test set)

  • Adjudication Method: Not applicable in the context of an AI study. For the bench and biocompatibility tests, "adjudication" would refer to the interpretation of raw test data against predefined acceptance criteria by qualified laboratory personnel, which is inherent in the "Pass" results reported.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

  • MRMC Study: No. This is a physical guidewire device, not an AI or software device that would typically involve human reader performance evaluation.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

  • Standalone Performance: Not applicable. This is not an algorithm or software device.

7. The Type of Ground Truth Used

  • Ground Truth Type:
    • Bench Testing: Objective physical measurements, engineering specifications, and performance defined by established standards (e.g., ability to transmit torque, resistance to kinking, coating integrity). Performance is compared against similar predicate devices and general industry standards.
    • Biocompatibility Testing: Established by standardized biological assays (e.g., ISO 10993 series) that measure cellular response, material-mediated effects, and systemic reactions in vitro/in vivo, compared to predefined safety thresholds.

8. The Sample Size for the Training Set

  • Training Set Sample Size: Not applicable. This is a physical device, not an AI or machine learning model that requires a training set.

9. How the Ground Truth for the Training Set was Established

  • Training Set Ground Truth Establishment: Not applicable, as there is no training set for an AI model.

FDA 510(k) Clearance Letter - X-Wire Guidewire

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

August 26, 2025

Imperative Care, Inc.
Shivani Patel
Principal Regulatory Affairs Specialist
1359 Dell Avenue
Campbell, California 95008

Re: K244061
Trade/Device Name: X-Wire Guidewire
Regulation Number: 21 CFR 870.1330
Regulation Name: Catheter Guide Wire
Regulatory Class: Class II
Product Code: MOF, DQX
Dated: July 23, 2025
Received: July 24, 2025

Dear Shivani Patel:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K244061 - Shivani Patel Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K244061 - Shivani Patel Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Naira Muradyan -S

Naira Muradyan, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K244061

Device Name: X-Wire Guidewire

Indications for Use (Describe):
The X-Wire Guidewire is indicated for general intravascular use within the peripheral and neuro vasculature to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary K244061

A. Submitter Information:

Submitter's Name: Imperative Care, Inc.
Address: 1359 Dell Avenue
Campbell, CA 95008
Contact Person: Shivani Patel
Telephone: 510-468-8862
Email: spatel@imperativecare.com
Date of Preparation: August 21, 2025

B. Subject Device:

Proprietary Name: X-Wire Guidewire
Common/Usual Name: Guidewire
Classification Name: Catheter guide wire
Product Codes: MOF, DQX
Regulation: 21 CFR 870.1330
Manufacturer: Imperative Care, Inc.

C. Predicate Devices:

Proprietary Names: Aristotle 18 Guidewire; Aristotle 24 Guidewire (K231954)
Aristotle 14 Guidewire (K220398)
Common/Usual Name: Guidewire
Classification Name: Catheter guide wire
Product Codes: MOF, DQX
Regulation: 21 CFR. 870.1330
Manufacturer: Scientia Vascular, Inc.

D. Device Description:

The Imperative Care's X-Wire Guidewires are guidewires with shapeable tips to aid in accessing the peripheral and neuro vasculature. The guidewires are available in 200cm - 300cm lengths, standard (S) and support (T) stiffness profiles, and 0.014", 0.018" and 0.024" diameters. The distal portion of the guidewire tip includes a radiopaque marker to facilitate fluoroscopic visualization. A hydrophilic coating on the distal segment and PTFE coating on the proximal segment serve to reduce friction during manipulation in vessels. The X-Wire Guidewire is supplied with a shaping mandrel, introducer, and torque device.

Page 6

E. Indications for Use:

The X-Wire Guidewire is indicated for general intravascular use within the peripheral and neuro vasculature to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

F. Comparison with the Predicate Devices:

The predicate devices are the Aristotle 18, 24, and 14 Guidewires cleared under K231954 and K220398. The predicate and subject devices share the same intended use and basic technological characteristics as shown in Table 1, below.

Table 1: Comparison of Subject and Predicate Devices

Predicate Aristotle 18 and 24 Guidewires (K231954)Predicate Aristotle 14 Guidewire (K220398)Subject Device X-Wire Guidewire (K244061)
Regulation Number21 CFR 870.1330SameSame
Regulation NameCatheter Guide WireSameSame
Regulatory ClassClass IISameSame
Product CodeMOF, DQXSameSame
ManufacturerScientia Vascular, Inc.Scientia Vascular, Inc.Imperative Care., Inc
Anatomical LocationPeripheral and Neuro VasculatureSameSame
Indications for UseThe Aristotle 18 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.The Aristotle 24 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for useThe Aristotle 14 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.The X-Wire Guidewire is indicated for general intravascular use within the peripheral and neuro vasculature to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.

Page 7

Predicate Aristotle 18 and 24 Guidewires (K231954)Predicate Aristotle 14 Guidewire (K220398)Subject Device X-Wire Guidewire (K244061)
in the coronary vasculature.
Condition SuppliedSterile, single-useSameSame
Accessory DevicesShaping Mandrel, Introducer, Torque DeviceSameSame
Wire Outer Diameter0.018" (0.46mm)0.024" (0.61mm)0.014" (0.36mm)0.014" (0.36mm)0.018" (0.46mm)0.024" (0.61mm)
Device Length200 cm, 300 cm200cm, 300cmX-Wire 14: Same as Aristotle 14X-Wire 18: 200cm, 260cm, 300cmX-Wire 24: Same as Aristotle 24
Tip Length (Hypotube)35cm35cmX-Wire 14S: 29cmX-Wire 14T: 19cmX-Wire 18S: 35cmX-Wire 18T: 25cmX-Wire 24S: 43.5cmX-Wire 24T: 50cm
Tip Type and ShapeCore-to-tip ShapeableSameSame
Flexible Tip Material (Distal Tip)NitinolSameSame
Core Wire MaterialStainless SteelSameSame
Stiffness ProfileSupport, Standard, SoftSoft, StandardSupport, Standard
CoatingsDistal End: HydrophilicProximal End: PTFESameSame
Hydrophilic Coating Length46cm46cmX-Wire 14S: 41cmX-Wire 14T: 27.5cmX-Wire 18S: 50cmX-Wire 18T: 41cmX-Wire 24S: 60cmX-Wire 24T: 60cm

Page 8

Predicate Aristotle 18 and 24 Guidewires (K231954)Predicate Aristotle 14 Guidewire (K220398)Subject Device X-Wire Guidewire (K244061)
Radiopaque Marker1 radiopaque marker at distal tipSameSame
Radiopaque Length10cm10cm85mm (8.5cm)
BushingAristotle 24: Nitinol BushingAristotle 18: NoneNoneNone
Centering CoilAristotle 18: 1 centering coilAristotle 24: 2 centering coils1 centering coilX-Wire 14S:1 centering coilX-Wire 14T: 1 centering coilX-Wire 18S: 2 centering coilsX-Wire 18T: 1 centering coilX-Wire 24S: 3 centering coilsX-Wire 24T: 2 centering coils
Sterilization MethodEOSameSame
Sterility Assurance LevelSAL of 10⁻⁶SameSame
Shelf Life36 months4 months8 months

G. Performance Data - Bench:

Bench testing was completed to evaluate the differences between the subject devices and predicate devices and to ensure the subject device meets the specifications and performs as intended. The test methods were based primarily on the FDA Guidance, "Coronary, Peripheral, and Neurovascular Guidewires - Performance Tests and Recommended Labeling, Guidance for Industry and Food and Drug Administration Staff," issued in October 2019. A summary of the evaluated performance specifications is presented in Table 2.

The test results demonstrated that any differences between the subject devices and predicate devices do not significantly impact performance parameters that would negatively affect the safety or effectiveness of the subject devices.

Page 9

Table 2: Summary of Bench Tests and Performance Specifications

Test AttributeSpecificationResults
Dimensional VerificationAll defined guidewire dimensions are within the specified tolerances.Pass
Visual InspectionThe guidewire shall be free of visual defects when removed from packaging.Pass
TorqueabilityThe guidewire shall transmit rotation from the proximal end to the distal tip to allow users to select branches of the vasculature and reach the target location.Pass
Torque StrengthIn the event where the distal end of the guidewire is unable to move/rotate during the procedure, the device shall not fail under expected torsional input.Pass
Tip FlexibilityThe tip of the guidewire shall not cause vessel damage. Tip flexibility is defined by the ability of the tip to buckle when contacting vessel walls. In doing so it reduces the contact pressure to prevent vessel perforation.Pass
Tensile Strength and Tip PullThe guidewire shall withstand tensile forces expected in clinical use without breaking.Pass
Kink ResistanceThe guidewire shall be able to traverse through clinically relevant bends without kinking.Pass
Coating IntegrityThe coating shall remain intact during clinical use.Pass
Coating LubricityThe coating shall be lubricious to reduce frictional forces to allow for navigation through tortuous anatomy.Pass
Particulate EvaluationThe guidewire shall not generate particles at a level that is greater than the range generated by the predicate.Pass
Corrosion ResistanceThe guidewire shall not corrode from the time of manufacture through its shelf life.Pass
RadiopacityOpacity to x-ray shall allow physicians to visualize guidewire under fluoroscopy.Pass
Simulated UseThe guidewire must be able to reach anatomical locations and deliver catheters and other interventional devices used in common neurovascular procedures.Pass

Page 10

H. Biocompatibility Testing:

Biocompatibility testing was performed on a representative final, finished guidewire to evaluate all direct patient contacting components of the device. This testing was conducted to ensure that the components and raw materials used in the subject X-Wire Guidewires and the manufacturing processes and sterilization processes result in a biocompatible product. Test results and an expert evaluation confirmed that the subject X-Wire Guidewires met biological safety requirements per ISO 10993-1 for externally communicating medical devices with direct circulating blood contact for ≤24 hours.

Table 3: Biocompatibility Test Summary – X-Wire Guidewire

TestTest MethodExtraction Methods/ConditionsAcceptance CriteriaResults
CytotoxicityISO 10993-5ISO 10993-12Test device extracted with MEM with 5% serum at 37°C for 72 hours.Test article extracts must yield grade 2 or lower.Pass, Noncytotoxic
SensitizationISO 10993-10ISO 10993-12Test device extracted with both SC and SO at 50°C for 72 hours.Overall pattern, intensity, duration, character of reactions compared to control conditions.Pass, Nonsensitizer
Irritation or Intracutaneous ReactivityISO 10993-23ISO 10993-12Test device extracted with both SC and SO at 50°C for 3 hours.Difference between test extract mean score and corresponding control mean score ≤1.Pass, Nonirritating
Acute Systemic ToxicityISO 10993-11ISO 10993-12Test device extracted with both SC and SO at 50°C for 72 hours.None of the animals treated with the test article extracts must show a significantly greater biological reactivity than animals treated with the control article.Pass, Non-toxic (acute systemic)
Material Mediated PyrogenicityISO 10993-11USP <151>ISO 10993-12Test device extracted with SNPS at 50°C for 72 hours.Test article extract must yield <0.5°C rise ofPass, Nonpyrogenic

Page 11

TestTest MethodExtraction Methods/ConditionsAcceptance CriteriaResults
individual animal baseline temperature.
Hemocompatibility – Hemolysis (Direct and Indirect Contact)ISO 10993-4ASTM F756ISO 10993-12Direct Contact: Test device incubated directly in diluted blood at 37°C for 72 hours.Indirect Contact: Test device extracted in CMF-PBS at 50°C for 72 hours and then extracts incubated with diluted blood at 37⁰C for 3 hours.Test article extract must result in hemolytic index of <2%.Pass, Nonhemolytic for both Direct and Extract Methods
Hemocompatibility - SC5b-9 Complement ActivationISO 10993-4Test device exposed to Normal Human Serum at 37°C for 60 minutes.SC5b-9 concentration of the test article is similar to both the activated NHS and negative controls.Pass, Hemocompatible
Hemocompatibility – In vivo thromboresistanceISO 10993-4N/A, nonanticoagulated venous implant study.Test article performs similarly to the predicate.Pass, Hemocompatible

I. Sterilization:

The X-Wire Guidewires are sterilized using a validated EO process with a sterility assurance level of 1x10⁻⁶ validated per the overkill method in accordance with ISO 11135, "Sterilization of Health-Care Products - Ethylene Oxide - Requirements for the Development, Validation and Routine Control of a Sterilization Process for Medical Devices".

J. Shelf Life and Packaging:

Accelerated aging testing based on ASTM F1980 was conducted to verify device, accessory, and packaging performance. An accelerated aging equivalent of 8 months was used to support 8-month shelf life. Device performance was verified by functional and performance testing.

Page 12

Packaging and sterile barrier integrity through transportation challenge has been verified. Aging testing has also been performed that supports the sterile barrier integrity following 8 months of accelerated aging.

K. Animal Study:

Animal study was not deemed necessary to demonstrate substantial equivalence.

L. Clinical Study:

Clinical study was not deemed necessary to demonstrate substantial equivalence.

M. Conclusions:

The X-Wire Guidewire has the same intended use and similar technological characteristics as the predicate devices. The performance testing was conducted to ensure the subject device performs as intended, meets all design specifications, and performs consistently throughout its labeled shelf life. The differences between subject and predicate devices do not raise any new or different questions of safety and effectiveness. Therefore, the X-Wire Guidewire is substantially equivalent to the predicate devices.

§ 870.1330 Catheter guide wire.

(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.