K Number
K250757

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-05-29

(78 days)

Regulation Number
870.1025
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Radius VSM and accessories are intended to be used as both a wearable multi-parameter patient monitor and an accessory to a multi-parameter patient monitor that is intended for multi-parameter physiological patient monitoring in hospital and healthcare facilities.

The Radius VSM and accessories are indicated for the monitoring of hemodynamic (including ECG, arrhythmia detection, non-invasive blood pressure, SpO2, Pulse Rate, PVi, heart rate, and temperature), and respiratory (e.g., impedance, acoustic, and pleth-based respiration rate) physiological parameters along with the orientation and activity of adults.

The Radius VSM and accessories are indicated for the non-invasive continuous monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate (PR) of well or poorly perfused adults during both no motion and motion conditions.

The Radius VSM and accessories are indicated for continuous monitoring of skin temperature of adults.

The Radius VSM and accessories are indicated for monitoring of the orientation and activity of patients including those susceptible to pressure ulcers.

The Radius VSM and accessories are indicated for the continuous non-invasive monitoring of PVi as a measure of relative variability of the photoplethysmograph (pleth) of adults during no motion conditions. PVi may be used as a noninvasive dynamic indicator of fluid responsiveness in select populations of mechanically ventilated adult patients. Accuracy of PVi in predicting fluid responsiveness is variable and influenced by numerous patient, procedure and device related factors. PVi measures the variation in the plethysmography amplitude but does not provide measurements of stroke volume or cardiac output. Fluid management decisions should be based on a complete assessment of the patient's condition and should not be based solely on PVi.

Devices with Masimo technology are only indicated for use with Masimo accessories.

Radius VSM Accessories:

Radius VSM ECG Electrodes are disposable, single-patient use ECG electrodes intended to acquire ECG signals from the surface of the body. They are indicated for use on adults for up to 3 days of skin surface contact.

Radius VSM Blood Pressure Cuffs are accessories intended to be used with a noninvasive blood pressure measurement system to measure blood pressure. They are indicated for use on adults during no motion conditions.

Device Description

The Radius VSM and accessories are an FDA cleared (K223498), wearable, battery-operated, multi-modular patient monitoring platform that allows for the ability to scale and tailor the use of different monitoring technologies based upon the hospital and clinician's assessment of what technologies are appropriate.

As part of this submission, a MAP feature is being added to the Radius VSM. The feature is a software feature that uses the previously cleared systolic and diastolic measurement capabilities to automate the calculation of MAP using the following formula: MAP = 1/3* Systolic + 2/3*Diastolic.

The MAP is calculated by the Radius VSM NIBP Module and displayed on the Radius VSM Wearable Monitor. There were no other features added as part of this submission.

AI/ML Overview

The provided 510(k) clearance letter and summary discuss the addition of a Mean Arterial Pressure (MAP) feature to the previously cleared Radius VSM and Accessories device. The primary focus of the performance data section is on validating this new MAP feature.

Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided document:

Acceptance Criteria and Reported Device Performance

The document states that the acceptance criterion for Blood Pressure (including MAP) is:

"Meets ISO 81060-2 (Mean difference of ≤5 mmHg with a standard deviation of ≤8 mmHg)"

The document directly states that the results of the clinical testing supported the clinical performance of the MAP in accordance with ISO 81060-2. While specific numerical results (e.g., the exact mean difference and standard deviation achieved) are not explicitly provided in the summary table, the clearance implies that these metrics fell within the specified ISO 81060-2 limits for the MAP feature.

Table 1: Acceptance Criteria and Reported Device Performance for MAP Feature (as inferred from the document)

FeatureAcceptance CriteriaReported Device Performance
Mean Arterial Pressure (MAP)Meets ISO 81060-2: Mean difference of ≤5 mmHg with a standard deviation of ≤8 mmHgPerformance met ISO 81060-2 (i.e., mean difference and standard deviation were within the specified limits).

Study Details for MAP Feature Validation

  1. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: The document does not explicitly state the numerical sample size (number of subjects/patients) used for the clinical test set. It only mentions "clinical study data."
    • Data Provenance: The document does not specify the country of origin. It indicates it was a "clinical study" and implies it was prospective ("clinical testing is provided to support its performance" for the added feature).
  2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts:

    • Not applicable as the ground truth was established by an objective reference device, not human experts.
  3. Adjudication Method for the Test Set:

    • Not applicable, as the method for ground truth establishment was comparison to a reference device.
  4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No, an MRMC study was not done. The study was a comparison of the device's calculated MAP to invasively measured MAP from a reference device. This is a technical performance validation, not a study assessing human reader improvement with AI assistance.
  5. If a Standalone Performance (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, this was a standalone performance study. The Radius VSM automatically calculates the MAP based on the NIBP measurements (Systolic and Diastolic Pressure). The clinical testing validated the accuracy of this calculation against a reference standard, without human intervention in the MAP calculation or interpretation for the test itself.
  6. The Type of Ground Truth Used:

    • Reference Ground Truth: Invasively measured MAP values from a 510(k) cleared reference device (K171801). This reference device is identified as "IntelliVue Multi-Measurement Module X3." This constitutes a device-based reference standard or instrument-based ground truth.
  7. The Sample Size for the Training Set:

    • The document does not provide information about a training set since the MAP feature appears to be a direct calculation using a standard formula (MAP = 1/3* Systolic + 2/3*Diastolic) rather than a machine learning model that requires a training phase. While the device as a whole (Radius VSM) likely had training and validation phases for its other parameters, the specific "addition of a Mean Arterial Pressure (MAP) feature" is described as a software feature that "automates the calculation" using a known formula. Therefore, a separate training set for this specific MAP feature is unlikely to have been required or used in the conventional machine learning sense.
  8. How the Ground Truth for the Training Set was Established:

    • As inferred above, a specific training set and ground truth establishment for this isolated MAP calculation feature are not described, given its nature as a direct formulaic calculation.

Summary of Key Information:

The core of this submission revolves around adding a simple, formula-based calculation for MAP. The primary study presented is a clinical validation confirming that the device's computed MAP aligns with a known industry standard (ISO 81060-2) when compared against an invasive reference device. This is a technical performance validation rather than a complex AI-driven diagnostic study.

U.S. Food & Drug Administration 510(k) Clearance Letter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

May 29, 2025

Masimo Corporation
Jazmin Laidet
Regulatory Affairs Specialist III
52 Discovery
Irvine, California 92618

Re: K250757
Trade/Device Name: Radius VSM and Accessories
Regulation Number: 21 CFR 870.1025
Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm)
Regulatory Class: Class II
Product Code: MHX, BZQ, DQA, DPS, DPZ, DRT, DSJ, DXN, FLL, KMI, DSI, DQK, DXQ
Dated: March 12, 2025
Received: March 12, 2025

Dear Jazmin Laidet:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K250757 - Jazmin Laidet
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

Page 3

K250757 - Jazmin Laidet
Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen C. Browning -S

LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K250757

Please provide the device trade name(s).
Radius VSM and Accessories

Please provide your Indications for Use below.

Radius VSM:

The Radius VSM and accessories are intended to be used as both a wearable multi-parameter patient monitor and an accessory to a multi-parameter patient monitor that is intended for multi-parameter physiological patient monitoring in hospital and healthcare facilities.

The Radius VSM and accessories are indicated for the monitoring of hemodynamic (including ECG, arrhythmia detection, non-invasive blood pressure, SpO2, Pulse Rate, PVi, heart rate, and temperature), and respiratory (e.g., impedance, acoustic, and pleth-based respiration rate) physiological parameters along with the orientation and activity of adults.

The Radius VSM and accessories are indicated for the non-invasive continuous monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate (PR) of well or poorly perfused adults during both no motion and motion conditions.

The Radius VSM and accessories are indicated for continuous monitoring of skin temperature of adults.

The Radius VSM and accessories are indicated for monitoring of the orientation and activity of patients including those susceptible to pressure ulcers.

The Radius VSM and accessories are indicated for the continuous non-invasive monitoring of PVi as a measure of relative variability of the photoplethysmograph (pleth) of adults during no motion conditions. PVi may be used as a noninvasive dynamic indicator of fluid responsiveness in select populations of mechanically ventilated adult patients. Accuracy of PVi in predicting fluid responsiveness is variable and influenced by numerous patient, procedure and device related factors. PVi measures the variation in the plethysmography amplitude but does not provide measurements of stroke volume or cardiac output. Fluid management decisions should be based on a complete assessment of the patient's condition and should not be based solely on PVi.

Devices with Masimo technology are only indicated for use with Masimo accessories.

Radius VSM Accessories:

  • Radius VSM ECG Electrodes are disposable, single-patient use ECG electrodes intended to acquire ECG signals from the surface of the body. They are indicated for use on adults for up to 3 days of skin surface contact.

  • Radius VSM Blood Pressure Cuffs are accessories intended to be used with a noninvasive blood pressure measurement system to measure blood pressure. They are indicated for use on adults during no motion conditions.

Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

Radius VSM and Accessories
Page 1 of 1
Page 8 of 34

Page 5

510(k) Summary K250757

Submitter and Address of Manufacturing Facility:
Masimo Corporation
52 Discovery
Irvine, CA 92618
Phone: (949) 297-7000

Date: May 20, 2025
Contact: Jazmin Laidet
Regulatory Affairs Specialist III
Masimo Corporation
Phone: (949) 577-2849

Trade Name: Radius VSM and Accessories
Common Name: Patient Monitor (with Arrhythmia Detection)
Classification Regulation/Product Code/Panel: 21 CFR 870.1025 / MHX / Cardiovascular
Regulatory Class: Class II

Additional Product Code:

  • 21 CFR 868.2375/BZQ
  • 21 CFR 870.2700/DQA
  • 21 CFR 870.2340/DPS
  • 21 CFR 870.2710/DPZ
  • 21 CFR 870.2300/DRT
  • 21 CFR 870.1100/DSJ
  • 21 CFR 870.1130/DXN
  • 21 CFR 880.2910/FLL
  • 21 CFR 880.2400/KMI
  • 21 CFR 870.1025/DSI
  • 21 CFR 870.1425/DQK
  • 21 CFR 870.1120/DXQ

Establishment Registration Number: 3011353843
Reason for Premarket Notification: Addition of Mean Arterial Pressure (MAP) feature
Primary Predicate: K223489 Radius VSM and Accessories
Reference Device: K171801 IntelliVue Multi-Measurement Module X3

1. Device Description

The Radius VSM and accessories are an FDA cleared (K223498), wearable, battery-operated, multi-modular patient monitoring platform that allows for the ability to scale and tailor the use of different monitoring technologies based upon the hospital and clinician's assessment of what technologies are appropriate.

Page 1 of 13

Page 6

510(k) Summary K250757

As part of this submission, a MAP feature is being added to the Radius VSM. The feature is a software feature that uses the previously cleared systolic and diastolic measurement capabilities to automate the calculation of MAP using the following formula:

MAP = 1/3 Systolic + 2/3Diastolic**

The MAP is calculated by the Radius VSM NIBP Module and displayed on the Radius VSM Wearable Monitor. There were no other features added as part of this submission.

There were no changes made to the Radius VSM Wearable Monitor specifications from the previous clearance under K223498. The specifications for the Radius VSM Wearable Module are provided in Table 1-1 below:

Table 1-1: Radius VSM Wearable Monitor Specifications

FeatureSpecifications
Display
Display TypeTouchscreen
Alarms
Type of AlarmsVisual/Audible Alarms
Technological Characteristics
Measured ParametersSpO2, PR, RRp, RRa
Calculated or Derived ParametersPVi, Pi
Performance Specifications
SpO2, No Motion70-100%, 1.5% Arms, Adults
SpO2, Motion70-100%, 1.5% Arms, Adults
SpO2, Low perfusion70-100%, 2% Arms, Adults
PR, No Motion25-240 bpm, 3 bpm, Adults
PR, Motion25-240 bpm, 5 bpm, Adults
PR, Low Perfusion25-240 bpm, 5 bpm, Adults
RRp, No Motion4-70, 3 rpm Arms, 1 rpm mean error, Adults
RRa4-70 rpm, 1 rpm Arms, Adults
Interfaces
PhysicalModules, Charging Adapter
WirelessBluetooth LE, Wi-Fi
Electrical
Internally PoweredRechargeable, Lithium Ion Battery
Mechanical Specifications
Dimensions10.9 cm x 5.8 cm x 2.1 cm (4.28" x 2.28" x 0.83")
Weight122 g (0.27 lbs.)
Ingress ProtectionIPX4
Environmental Specifications
Operating Temperature0°C to 40°C (32°F to 104°F)

Page 2 of 13

Page 7

510(k) Summary K250757

Table 1-1: Radius VSM Wearable Monitor Specifications (continued)

FeatureSpecifications
Storage Temperature-20°C to 60°C (-4°F to 140°F)
Operating/Storage/Transport Humidity10% to 95%, non-condensing
Operating Atmospheric Pressure540 mbar to 1060 mbar
Classification per IEC 60601-1
Electrical SafetyIEC 60601-1
EMCIEC 60601-1-2
Electrical Isolation TypeClass II (Internally Powered)
Applied Part TypeBF Applied Part
Ingress ProtectionIP22
Mode of OperationContinuous

There were no changes made to the Radius VSM ECG Module specifications from the previous clearance under K223498. The specifications for the Radius VSM ECG Module are provided in Table 1-2 below:

Table 1-2: Radius VSM ECG Module Specifications

FeatureSpecifications
Technical Characteristics
Measured ParametersECG waveform, HR, RRe, Temperature, Time in Position, Patient Incline Angle
Calculated or Derived ParametersArrhythmia Detection, Fall Detection
Performance Specifications
ECG, Monitoring Bandwidth0.67 Hz to 40 Hz
ECG, Diagnostic Bandwidth0.05 Hz to 150 Hz
HR (15bpm to 300 bpm)1% or ± 2 bpm (whichever is greater)
RRe (4-120 rpm)≤1 rpm mean error
Patient Incline Angle (-180° to 180°C)±1°
Laboratory Accuracy (77°F to 109.4°F (25°C to 43°C))± 0.3°C (±.54°F)
Lethal Arrhythmias DetectedAsystole, Ventricular Fibrillation/Ventricular Tachycardia, Ventricular Tachycardia (greater than 30s)
Non-Lethal Arrhythmias DetectedAtrial Fibrillation greater than 30 seconds, Ventricular Tachycardia (less than 30s)
Interface
PhysicalModule connector
Electrical
DC PoweredRadius VSM Module
Mechanical
Dimensions4.7 cm x 4.06 cm (1.85" x 1.60")
Weight20 g (0.04 lbs.)
Ingress ProtectionIPX4

Page 3 of 13

Page 8

510(k) Summary K250757

Table 1-2: Radius VSM ECG Module Specifications (continued)

FeatureSpecifications
Environmental
Operating Temperature0°C to 40°C (32°F to 104°F)
Storage/Transport Temperature-20°C to 60°C (-4°F to 140°F)
Operating/Storage/Transport Humidity10% to 95%, non-condensing
Operating Atmospheric Pressure540 mbar to 1060 mbar (540 hPa to 1060 hPa)

Apart from the added MAP feature, there were no changes made to the specifications from the previous clearance under K223498. The specifications for the Radius VSM NIBP Module are provided in Table 1-3 below.

Table 1-3: Radius VSM NIBP Module Specifications

FeatureSpecifications
Technical Characteristics
Measured ParametersSystolic, Diastolic, Pulse Rate
Calculated or Derived ParametersMean Arterial Pressure (MAP)
Performance Specifications
Pressure Transducer (Between 0 mmHg and 300 mmHg)±3 mmHg
Blood PressureMeets ISO 81060-2 (Mean difference of ≤5 mmHg with a standard deviation of ≤8 mmHg)
Interface
PhysicalModule connector
Electrical
DC PoweredRadius VSM Module
Mechanical
Dimensions9.3 cm x 5.5 cm x 2.9 cm (3.66" x 2.17" x 0.86")
Weight111 g (0.24 lbs.)
Ingress ProtectionIPX2
Environmental
Operating Temperature0°C to 40°C (32°F to 104°F)
Storage/Transport Temperature-20°C to 60°C (-4°F to 140°F)
Operating/Storage/Transport Humidity10% to 95%, non-condensing
Operating Atmospheric Pressure540 mbar to 1060 mbar (540 hPa to 1060 hPa)

2. Intended Use/ Indications for Use

There are no changes to the intended use for the subject device, Radius VSM, from the previous clearance under K223498. The intended use statement is provided below for ease of reference:

Radius VSM:

The Radius VSM and accessories are intended to be used as both a wearable multi-parameter patient monitor and an accessory to a multi-parameter patient monitor that is intended for multi-parameter physiological patient monitoring in hospital and healthcare facilities.

Page 4 of 13

Page 9

510(k) Summary K250757

The Radius VSM and accessories are indicated for the monitoring of hemodynamic (including ECG, arrhythmia detection, non-invasive blood pressure, SpO2, Pulse Rate, PVi, heart rate, and temperature), and respiratory (e.g., impedance, acoustic, and pleth-based respiration rate) physiological parameters along with the orientation and activity of adults.

The Radius VSM and accessories are indicated for the non-invasive continuous monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate (PR) of well or poorly perfused adults during both no motion and motion conditions.

The Radius VSM and accessories are indicated for continuous monitoring of skin temperature of adults.

The Radius VSM and accessories are indicated for monitoring of the orientation and activity of patients including those susceptible to pressure ulcers.

The Radius VSM and accessories are indicated for the continuous non-invasive monitoring of PVi as a measure of relative variability of the photoplethysmograph (pleth) of adults during no motion conditions. PVi may be used as a noninvasive dynamic indicator of fluid responsiveness in select populations of mechanically ventilated adult patients. Accuracy of PVi in predicting fluid responsiveness is variable and influenced by numerous patient, procedure and device related factors. PVi measures the variation in the plethysmography amplitude but does not provide measurements of stroke volume or cardiac output. Fluid management decisions should be based on a complete assessment of the patient's condition and should not be based solely on PVi.

Devices with Masimo technology are only indicated for use with Masimo accessories.

Radius VSM Accessories:

Radius VSM ECG Electrodes are disposable, single-patient use ECG electrodes intended to acquire ECG signals from the surface of the body. They are indicated for use on adults for up to 3 days of skin surface contact.

Radius VSM Blood Pressure Cuffs are accessories intended to be used with a noninvasive blood pressure measurement system to measure blood pressure. They are indicated for use on adults during no motion conditions.

3. Technological Characteristics

Principle of Operation

As part of this submission, Masimo is requesting the addition of a mean arterial pressure (MAP) calculation feature. The MAP feature relies on the calculation of the mean arterial pressure using the established weighted average of the systolic and diastolic blood pressure values. The Radius VSM automates the calculation to output an MAP value.

Mechanism of Action for Achieving the Intended Effect

Page 5 of 13

Page 10

510(k) Summary K250757

There were no changes to the mechanism of action of the Radius VSM and Accessories in the addition of the MAP feature from the previous clearance under K223498.

The MAP value from the Radius VSM is provided along with the previously cleared systolic and diastolic values without any additional user actions. To obtain the MAP value and the other NIBP parameter data, the Radius VSM NIBP module is attached to the Radius VSM NIBP cuff which is applied to the patient. The Radius VSM NIBP module determines the Systolic Pressure, Diastolic Pressure, Pulse Rate, and MAP values that are communicated to the Radius VSM Wearable Monitor for display and wireless communication.

4. Discussion of Similarities and Differences Between Primary Predicate and Subject Device

Similarities and Differences between Primary Predicate and Subject Device

The subject device, Radius VSM and Accessories, and the predicate device, Radius VSM and Accessories (K223498) have the following key similarities:

  • Both devices have the same intended use.
  • Both devices have the same principle of operation and mechanism of action.
  • Both devices are indicated for the same patient population.

The subject device, Radius VSM and Accessories, and the predicate device, Radius VSM and Accessories (K223498), have the following key difference:

  • The subject device includes the Mean Arterial Pressure (MAP) feature.

The purpose of this submission is the addition of a Mean Arterial Pressure (MAP) feature to the predicate device, Radius VSM (K223498). The subject and the predicate device have the same intended use. To support the addition of MAP feature does not raise different questions of safety and effectiveness, clinical testing is provided to support its performance. The testing supports the substantial equivalence of the subject device.

Refer to Table 4-1 below for the detailed comparison between the subject and predicate devices.

Page 6 of 13

Page 11

510(k) Summary K250757

Table 4-1 Comparison between Subject and Predicate Devices

FeatureSubject Predicate Radius VSMPredicate Device Radius VSM (K223498)Comparison to Predicate
Intended UseThe Radius VSM and accessories are intended to be used as both a wearable multi-parameter patient monitor and an accessory to a multi-parameter patient monitor that is intended for multi-parameter physiological patient monitoring in hospital and healthcare facilities.The Radius VSM and accessories are intended to be used as both a wearable multi-parameter patient monitor and an accessory to a multi-parameter patient monitor that is intended for multi-parameter physiological patient monitoring in hospital and healthcare facilities.Same.
Indications for UseThe Radius VSM and accessories are indicated for the monitoring of hemodynamic (including ECG, arrhythmia detection, non-invasive blood pressure, SpO2, Pulse Rate, PVi, heart rate, and temperature), and respiratory (e.g., impedance, acoustic, and pleth-based respiration rate) physiological parameters along with the orientation and activity of adults.The Radius VSM and accessories are indicated for the non-invasive continuous monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate (PR) of well or poorly perfused adults during both no motion and motion conditions.The Radius VSM and accessories are indicated for continuous monitoring of skin temperature of adults.The Radius VSM and accessories are indicated for monitoring of the orientation and activity of patients including those susceptible to pressure ulcers.The Radius VSM and accessories are indicated for the continuous non-invasive monitoring of PVi as a measure of relative variability of the photoplethysmograph (pleth) of adults during no motion conditions. PVi may be used as a noninvasive dynamic indicator of fluid responsiveness in select populations of mechanically ventilated adult patients. Accuracy of PVi in predicting fluid responsiveness is variable and influenced by numerous patient, procedure and device related factors. PVi measures the variation in the plethysmography amplitude but does not provide measurements of stroke volume or cardiac output. Fluid management decisions should be based on a complete assessment of the patient's condition and should not be based solely on PVi.The Radius VSM and accessories are indicated for the monitoring of hemodynamic (including ECG, arrhythmia detection, non-invasive blood pressure, SpO2, Pulse Rate, PVi, heart rate, and temperature), and respiratory (e.g., impedance, acoustic, and pleth-based respiration rate) physiological parameters along with the orientation and activity of adults.The Radius VSM and accessories are indicated for the non-invasive continuous monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate (PR) of well or poorly perfused adults during both no motion and motion conditions.The Radius VSM and accessories are indicated for continuous monitoring of skin temperature of adults.The Radius VSM and accessories are indicated for monitoring of the orientation and activity of patients including those susceptible to pressure ulcers.The Radius VSM and accessories are indicated for the continuous non-invasive monitoring of PVi as a measure of relative variability of the photoplethysmograph (pleth) of adults during no motion conditions. PVi may be used as a noninvasive dynamic indicator of fluid responsiveness in select populations of mechanically ventilated adult patients. Accuracy of PVi in predicting fluid responsiveness is variable and influenced by numerous patient, procedure and device related factors. PVi measures the variation in the plethysmography amplitude but does not provide measurements of stroke volume or cardiac output. Fluid management decisions should be based on a complete assessment of the patient's condition and should not be based solely on PVi.Same.

Page 7 of 13

Page 12

510(k) Summary K250757

Table 4-1 Comparison between Subject and Predicate Devices (continued)

FeatureSubject Predicate Radius VSMPredicate Device Radius VSM (K223498)Comparison to Predicate
Devices with Masimo technology are only indicated for use with Masimo accessories.Radius VSM ECG Electrodes are disposable, single-patient ECG electrodes intended to acquire ECG signals from the surface of the body. They are indicated for use on adults for up to 3 days of skin surface contact.Radius VSM Blood Pressure Cuffs are accessories intended to be use with a noninvasive blood pressure measurement system to measure blood pressure. They are indicated for use on adults during no motion conditions.Devices with Masimo technology are only indicated for use with Masimo accessories.Radius VSM ECG Electrodes are disposable, single-patient ECG electrodes intended to acquire ECG signals from the surface of the body. They are indicated for use on adults for up to 3 days of skin surface contact.Radius VSM Blood Pressure Cuffs are accessories intended to be use with a noninvasive blood pressure measurement system to measure blood pressure. They are indicated for use on adults during no motion conditions.
Primary Classification Regulation/Product Code21 CFR 870.1025/ MHX21 CFR 870.1025/ MHXSame.
Additional Classification Regulation/Product Code(s)21 CFR 868.2375/ BZQ21 CFR 870.2700/ DQA21 CFR 870.2340/ DPS21 CFR 870.2710/ DPZ21 CFR 870.2300/ DRT21 CFR 870.1100/ DSJ21 CFR 870.1130/ DXN21 CFR 880.2910/ FLL21 CFR 870.1025/ DSI21 CFR 870.1425/ DQK21 CFR 880.2400/ KMI21 CFR 870.1120/ DXQ21 CFR 868.2375/ BZQ21 CFR 870.2700/ DQA21 CFR 870.2340/ DPS21 CFR 870.2710/ DPZ21 CFR 870.2300/ DRT21 CFR 870.1100/ DSJ21 CFR 870.1130/ DXN21 CFR 880.2910/ FLL21 CFR 870.1025/ DSI21 CFR 870.1425/ DQK21 CFR 880.2400/ KMI21 CFR 870.1120/ DXQSame.
Pulse Oximetry Monitored Parameters/FeaturesSpO2, PR, PVi, RRpSpO2, PR, PVi, RRpSame.
ECG Monitored Parameters/FeaturesECG waveform, Heart Rate, Impedance Respiration Rate, Arrhythmia Detection.ECG waveform, Heart Rate, Impedance Respiration Rate, Arrhythmia Detection.Same.

Page 8 of 13

Page 13

510(k) Summary K250757

Table 4-1 Comparison between Subject and Predicate Devices (continued)

FeatureSubject Predicate Radius VSMPredicate Device Radius VSM (K223498)Comparison to Predicate
NiBP Monitored Parameters/FeaturesSystolic Pressure, Diastolic Pressure and Pulse Rate, Mean Arterial Pressure (MAP)Systolic Pressure, Diastolic Pressure and Pulse RateDifferent. The subject device supports a Mean Arterial Pressure (MAP) feature.Clinical performance validation is provided to support the feature does not raise different questions of safety and effectiveness.
Thermometer Monitored Parameters/FeaturesTemperatureTemperatureSame.
Acoustic Respiration Rate Monitored Parameters/FeaturesRRaRRaSame.
Position/Orientation Monitored Parameters/FeaturesTime in Position, Patient Incline AngleTime in Position, Patient Incline AngleSame.
Aggregate Respiration RateYesYesSame.
Indicated populationAdultAdultSame.
Type of displayLCD, TouchscreenLCD, TouchscreenSame.
Battery Internal PowerRechargeable, Lithium ionRechargeable, Lithium ionSame.
Alarms NotificationsAudible and visualAudible and visualSame.
Physical Characteristics
Dimensions10.9 cm x 5.8 cm x 2.1 cm10.9 cm x 5.8 cm x 2.1 cmSame.
Weight0.27 lbs0.27 lbsSame.
Performance Specifications
SpO2, No Motion Accuracy1.5% Arms, 70-100%1.5% Arms, 70-100%Same.
SpO2, Motion Accuracy1.5% Arms, 70-100%1.5% Arms, 70-100%Same.
SpO2, Low perfusion2% Arms, 70-100%2% Arms, 70-100%Same.
PR, No motion3 bpm, 25-240 bpm3 bpm, 25-240 bpmSame.
PR, Motion5 bpm, 25-240 bpm5 bpm, 25-240 bpmSame.

Page 9 of 13

Page 14

510(k) Summary K250757

Table 4-1 Comparison between Subject and Predicate Devices (continued)

FeatureSubject Predicate Radius VSMPredicate Device Radius VSM (K223498)Comparison to Predicate
PR, Low Perfusion5 bpm, 25-240 bpm5 bpm, 25-240 bpmSame.
RRa1 rpm Arms, 4-70 rpm1 rpm Arms, 4-70 rpmSame.
RRp, No Motion3 rpm Arms, 1 mean error, 4-70 rpm3 rpm Arms, 1 mean error, 4-70 rpmSame.
Heart Rate≤1% or ≤2 bpm (whichever is greater)≤1% or ≤2 bpm (whichever is greater)Same.
Impedance Respiration Rate≤1 rpm Arms, 4-120 rpm≤1 rpm Arms, 4-120 rpmSame.
Pressure Transducer (Between 0 mmHg and 300 mmHg)±3 mmHg±3 mmHgSame.
NIBP Performance StandardsANSI/AAMI SP10 and ISO 81060-2ANSI/AAMI SP10 and ISO 81060-2Same.
Temperature±0.3°C,77°F to 109.4°F (25°C to 43°C)±0.3°C,77°F to 109.4°F (25°C to 43°C)Same.
Patient Incline Angle-180° to 180°-180° to 180°Same.
Environmental Specifications
Operating conditions
Temperature0º C to 40 º C0º C to 40 º CSame.
Humidity10% to 95%, non-condensing10% to 95%, non-condensingSame.
Storage conditions
Temperature-20°C to 60°C-20°C to 60°CSame.
Humidity10% to 95%, non-condensing10% to 95%, non-condensingSame.
ECG General
Lead Configurations Supported33Same.
Pacemaker Pulse Rejection Capability± 2 mV to ± 700 mV± 2 mV to ± 700 mVSame.
ECG Monitoring Bandwidth0.67 Hz to 40 Hz0.67 Hz to 40 HzSame.
ECG Diagnostic Bandwidth0.05 Hz to 150 Hz0.05 Hz to 150 HzSame.
Supported Arrhythmia Classificationslethal and non-arrhythmialethal and non-arrhythmiaSame.

Page 10 of 13

Page 15

510(k) Summary K250757

Table 4-1 Comparison between Subject and Predicate Devices (continued)

FeatureSubject Predicate Radius VSMPredicate Device Radius VSM (K223498)Comparison to Predicate
Thermometer Validation MethodISO 80601-2-56ISO 80601-2-56Same.
Thermometer Application Site(s)ChestChestSame.
Temperature Measurement ModeContinuousContinuousSame.
NIBP Measurement MethodOscillometricOscillometricSame.
Sensor Technology3-axis accelerometer3-axis gyroscope3-axis accelerometer3-axis gyroscopeSame.
Mode of Operation per IEC 60601-1
Mode of OperationContinuousContinuousSame.

Page 11 of 13

Page 16

510(k) Summary K250757

5. Performance Data

As part of this submission, clinical and non-clinical testing was included to support the addition of the Mean Arterial Pressure (MAP) Feature.

Performance Bench Testing

Bench testing was conducted to support the performance of the Mean Arterial Pressure (MAP) using a simulator.

Biocompatibility Testing

As there were no changes made to the patient contacting materials of the subject device from the previous clearance under K223498, no new biocompatibility testing was included in this submission.

Electromagnetic Compatibility, Electrical Safety, Environmental, Mechanical and Cleaning

As there were no hardware changes made to the subject device from the previous clearance under K223498, no new electrical safety, environmental, mechanical, and cleaning testing was included as part of this submission.

Software Verification and Validation Testing

Software verification and validation testing were conducted and provided as part of this submission as recommended by the FDA Guidance, "Content of Premarket Submissions for Software Device Software Functions", June 2023 to support the software change to add the Mean Arterial Pressure (MAP) feature.

Wireless Testing

As there were no changes to the wireless capabilities from the previous clearance under K223498, no new wireless testing was included as part of this submission.

Cybersecurity Testing

As there were no changes made to the subject device that would affect its cybersecurity from the previous clearance under K223498, no new cybersecurity testing was included as part of this submission.

Human Factors and Usability Testing

Previous human factors testing of the predicate device were found to support the acceptable human factors and usability risk of the subject device.

Clinical Testing

Clinical study data was included as part of this submission to support the addition of the calculation of the mean arterial pressure (MAP) feature by Radius VSM. The clinical study used invasively measured MAP

Page 12 of 13

Page 17

510(k) Summary K250757

values from a 510(k) cleared reference device (K171801) as a comparator in accordance with the FDA recognized standard ISO 81060-2.

The results of the testing supported the clinical performance of the MAP in accordance with ISO 81060-2.

6. Conclusion

The data provided as part of this submission was found to support the substantial equivalent of the subject device.

Page 13 of 13

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.