Search Filters

Search Results

Found 9 results

510(k) Data Aggregation

    K Number
    K221722
    Date Cleared
    2022-11-15

    (154 days)

    Product Code
    Regulation Number
    868.5830
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    |
    | Regulation Number | 21 CFR 864.9245

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Haemonetics® Cell Saver® Elite®+ Autotransfusion System and its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping the processed red blood cells to a product bag. The intended use of the Sequestration Protocol is to collect an autologous, preoperative, plasma product for reinfusion to the same patient within the recommended time of the American Association of Blood Banks (AABB), 9th Edition.

    Device Description

    The subject of this Traditional 510(k) is the Haemonetics Cell Saver Elite/Elite+ Autotransfusion System 7.3 (AQ) software update which allows users the ability to manually control the cell salvage procedure through manual mode, quick transfer and decreased minimum wash volume. The Cell Saver Elite/Elite+ System consists of a single use disposable set and reusable equipment. One disposable set is used throughout an individual patient's surgical procedure and then discarded. The Cell Saver Elite/Elite+ System utilizes a unique bowl processing kit, but is compatible with Haemonetics standard reservoirs and A&A lines. The collected blood is processed through a centrifugal separation chamber (bowl) where RBCs are concentrated and then washed, removing unwanted substances such as hemolized cells, anticoagulant and irrigating fluids. The washed RBC product is available for return via a product bag to the patient. The Elite+ System is designed to perform plasma sequestration using the autotransfusion disposable in conjunction with an ancillary sequestration set prior to performing autotransfusion.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Haemonetics Cell Saver Elite/Elite+ Autotransfusion System (CSE-E-US/CSE-EW-US) software update. It describes the device, its intended use, and the non-clinical testing performed to demonstrate substantial equivalence to a predicate device. However, this document does not contain information about the acceptance criteria and study design for proving the device meets those criteria from an AI/ML perspective.

    The changes in this 510(k) are related to a software update (version 7.3 AQ) for an autotransfusion system, specifically adding "manual mode, quick transfer and decreased minimum wash volume" features. This device processes blood (concentrating and washing red blood cells) rather than interpreting medical images or data using AI/ML algorithms.

    Therefore, many of the requested items (sample size for test/training sets, data provenance, number/qualifications of experts, adjudication methods, MRMC studies, standalone AI performance, type of ground truth for AI, how ground truth for training set was established) are not applicable to this type of device and submission.

    The document focuses on:

    • Software Verification: To verify the new software revision.
    • Functional Testing: To validate washout performance (a physical function of the blood processing).
    • Usability Testing: To validate operational needs and usability.

    These tests are standard for a medical device software update and functional changes, but they do not involve AI/ML performance evaluation as typically understood in the context of diagnostic or prognostic AI systems that require ground truth, expert readers, and rigorous statistical analysis of AI model performance.

    In summary, this document is for a medical device software update, not an AI/ML device. Therefore, it does not provide the information requested about AI/ML acceptance criteria and study paradigms.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200530
    Manufacturer
    Date Cleared
    2020-09-11

    (193 days)

    Product Code
    Regulation Number
    864.9245
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    a regulation by the Center for Devices and Radiological Health due to pre-amendment status.

    21 CFR 864.9245

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Depending on the AMICUS Separator System disposable used in the therapeutic apheresis procedure, the AMICUS Separator System has been cleared for the following:

    The AMICUS Separator System is an automated blood cell separator indicated to perform therapeutic plasma exchange (TPE). (K111702)

    The AMICUS Exchange Kit is indicated for use in therapeutic plasma exchange (TPE). The kit is for use with the AMICUS separator. (K111702)

    The AMICUS Separator System is an automated blood component separator indicated to perform red blood cell exchange (RBCX), including exchange and depletion/exchange procedures, for the transfusion management of sickle cell disease in adults and children. (K180615)

    The AMICUS Exchange Kit – Therapeutics is indicated for use in therapeutic plasma exchange (TPE) and red blood cell exchange (RBCX). The kit is for use with the AMICUS separator. (K111702, K180615)

    The waste transfer set is indicated for use in red blood cell exchange (RBCX). The set is for use with the AMICUS separator. (K180615)

    The Blood Component Filter Set with Vented Spike and Luer Adapter is indicated for the administration of blood and blood components during a therapeutic plasma exchange (TPE) or red blood cell exchange (RBCX) therapeutic apheresis procedure. The set is for use with the AMICUS separator. (K111702, K180615)

    The AMICUS Separator System is an automated blood cell separator indicated for the collection of blood components and mononuclear cells.

    Depending on the AMICUS Separator System apheresis kit used in the collection of products, the AMICUS Separator System has been cleared to collect:

    • Platelets Pheresis, Leukocytes Reduced (single, double, or triple units)(BK960005, BK990009)
    • Platelets Pheresis, Leukocytes Reduced, Platelet Additive Solution (InterSol)(single, double or triple units) (BK090065)
    • Red Blood Cells, Leukocytes Reduced (by apheresis) (BK000039)
    • Mononuclear Cells (BK000047)
    • Plasma (BK960005, BK120041)
    • Fresh Frozen Plasma
    • Must be prepared and placed in a freezer at -18° C or colder within eight hours after phlebotomy.
    • Plasma Frozen Within 24 Hours After Phlebotomy (PF24)
    • Must be stored at 1-6°C within eight hours after phlebotomy and placed in a freezer at -18° C or colder within 24 hours after phlebotomy.
    • Indicated for replacement of non-labile clotting factors. This product is not equivalent to Fresh Frozen Plasma.
    • Plasma Frozen Within 24 Hours After Phlebotomy (PF24) Held at Room Temperature Up to 24 Hours After Phlebotomy (PF24RT24)
    • Can be stored at room temperature for up to 24 hours after phlebotomy. Product must be placed in a freezer at -18° C or colder within 24 hours after phlebotomy.
    • Indicated for replacement of non-labile clotting factors. This product is not equivalent to Fresh Frozen Plasma.
    • Source Plasma

    The device is designed to collect products while maintaining an extracorporeal volume at or below 10.5ml/kg and a donor post-platelet count greater than or equal to 100,000 platelets/microliter.

    Platelet Pheresis (single, double, or triple units) may be manufactured from products that do not meet leukocyte reduction product standards. This does not apply to Platelet Additive Solution (InterSol) (single, double or triple units).

    The AMICUS platelet storage container is cleared to store Platelets Pheresis, Leukocytes Reduced in 100% plasma for up to seven days. Additionally, for platelet units stored past five days and through seven days, every product must be tested with a bacterial device cleared by FDA and labeled as a "safety measure."

    Device Description

    The AMICUS Separator System is comprised of the AMICUS separator instrument and a disposable apheresis kit specific to the procedure being performed. The instrument is a continuous-flow, centrifugal device that draws whole blood from a donor/patient, separates the blood into its components, collects one or more of the blood components, and returns the remainder of the blood components to the donor/patient. The instrument operates using pumps, clamps and valves that move donor/patient blood through a singleuse, sterile fluid path disposable kit. The cells are centrifugally separated within the kit by density differences.

    The operator is responsible for connecting and monitoring the donor/patient and operating and monitoring the AMICUS separator during the procedure. The operator controls the separator through a touch screen. When necessary, the operator is warned of problems with messages on the screen and corresponding audible alerts.

    Once the cell separation is complete, the operator disconnects the donor/patient, dismantles the kit, and disposes of the kit in a safe manner. The kit is packaged in a recyclable plastic tray.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for the Fresenius Kabi AMICUS Separator System. This document is a regulatory approval letter and a summary of the device and its modifications, not a study report. As such, it does not contain the detailed performance study data, acceptance criteria, ground truth establishment methods, or sample sizes typically found in a clinical study report or a pre-market approval application for a novel device or a significant new indication.

    The document states that the primary changes to the AMICUS device are:

    • Replacement of obsolete controller circuit boards and associated redesign of other internal boards.
    • Modification to the rear door design to accommodate new USB ports, Ethernet port, serial port, and Wi-Fi antenna.
    • An update to the 6.0 software, including a change in the operating system from pSOS to QNX, to run on the proposed device with the new hardware.
    • Qualification of a new off-the-shelf USB barcode scanner.

    It explicitly states: "The software changes do not involve changes to the core blood component collection or therapeutic procedure functionality." and "The AMICUS disposable apheresis kits remain the same as the currently cleared kits, including design, materials and manufacturing methods."

    The conclusion section states: "Software and systems verification testing of impacted requirements were performed with the replacement hardware components and updated 6.0 software. Regression testing was performed including paired testing with the predicate device on key product quality and performance characteristics. Results from the software and systems verification testing indicate that the subject device is as safe and performs as well as the predicate device."

    Therefore, based on the provided text, I cannot extract the specific information requested in the prompt regarding acceptance criteria tables, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, or detailed ground truth methodologies, because these types of studies were not conducted or reported in this 510(k) summary for a device modification, where the manufacturer is claiming substantial equivalence to a previously cleared predicate device due to hardware and software updates that do not alter core functionality. The "performance data" section primarily refers to verification testing (software and systems verification, regression testing compared to predicate, and electrical safety/EMC testing) rather than clinical performance studies with human subjects or expert readers that would typically involve the criteria you've listed.

    In summary, the document describes a device modification rather than a new device or significant new indication. The approval relies on demonstrating substantial equivalence to existing, cleared devices, primarily through engineering verification and validation testing, and not necessarily new extensive human-based performance studies as would be the case for an AI/ML algorithm with diagnostic capabilities.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192150
    Manufacturer
    Date Cleared
    2019-11-13

    (96 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    .

    21 CFR 864.9245 Automated Blood Cell Separator

    Automated blood cell separators which are based

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Depending on the AMICUS Separator System disposable used in the therapeutic apheresis procedure, the AMICUS Separator System has been cleared for the following:

    The AMICUS Separator System is an automated blood cell separator indicated to perform therapeutic plasma exchange (TPE). (K111702)

    The AMICUS Exchange Kit is indicated for use in therapeutic plasma exchange (TPE). The kit is for use with the AMICUS separator. (K111702)

    The AMICUS Separator System is an automated blood component separator indicated to perform red blood cell exchange (RBCX), including exchange and depletion/exchange procedures, for the transfusion management of sickle cell disease in adults and children. (K180615)

    The AMICUS Exchange Kit – Therapeutics is indicated for use in therapeutic plasma exchange (TPE) and red blood cell exchange (RBCX). The kit is for use with the AMICUS separator. (K111702, K180615)

    The waste transfer set is indicated for use in red blood cell exchange (RBCX). The set is for use with the AMICUS separator. (K180615)

    The Blood Component Filter Set with Vented Spike and Luer Adapter is indicated for the administration of blood and blood components during a therapeutic plasma exchange (TPE) or red blood cell exchange (RBCX) therapeutic apheresis procedure. The set is for use with the AMICUS separator. (K111702, K180615)

    The AMICUS Separator System is an automated blood cell separator indicated for the collection of blood components and mononuclear cells.

    Depending on the AMICUS Separator System apheresis kit used in the collection of products, the AMICUS Separator System has been cleared to collect:

    • · Platelets Pheresis, Leukocytes Reduced (single, double, or triple units) (BK960005) (BK990009)
    • · Platelets Pheresis, Leukocytes Reduced, Platelet Additive Solution (InterSol) (single, double or triple units) (BK090065)
    • · Red Blood Cells, Leukocytes Reduced (by apheresis) (BK000039)
    • Mononuclear Cells (BK000047)
    • · Plasma (BK960005, BK120041)
      • o Fresh Frozen Plasma

    · Must be prepared and placed in a freezer at -18° C or colder within 8 hours after phlebotomy.

    • o Plasma Frozen Within 24 Hours After Phlebotomy (PF24)
      • Must be stored at 1-6℃ within 8 hours after phlebotomy and placed in a freezer at -18° C or colder within 24 hours after phlebotomy.

    · Indicated for replacement of non-labile clotting factors. This product is not equivalent to Fresh Frozen Plasma.

    o Plasma Frozen Within 24 Hours After Phlebotomy (PF24) Held at Room Temperature Up to 24 Hours After Phlebotomy (PF24RT24)

    • · Can be stored at room temperature for up to 24 hours after phlebotomy. Product must be placed in a freezer at -18° C or colder within 24 hours after phlebotomy.

    • · Indicated for replacement of non-labile clotting factors. This product is not equivalent to Fresh Frozen Plasma.
      o Source Plasma

    The device is designed to collect products while maintaining an extracorporeal volume at or below 10.5ml/kg and a donor post platelet count greater than or equal to 100,000 platelets/microliter. (BK080018)

    Platelet Pheresis (single, double, or triple units) may be manufactured from products that do not meet leukocyte reduction product standards. This does not apply to Platelet Additive Solution (InterSol) (single, double, or triple units). (BK090065)

    The AMICUS platelet storage container is cleared to store Platelets Pheresis, Leukocytes Reduced in 100% plasma for up to 7 days. Additionally, for platelet units stored past 5 days and through 7 days, every product must be tested with a bacterial device cleared by FDA and labeled as a "safety measure." (BK040059, BK150242)

    Device Description

    The AMICUS Separator System is comprised of the AMICUS separator instrument and a disposable apheresis kit specific to the procedure being performed. The instrument is a continuous-flow, centrifugal device that draws whole blood from a donor/patient, separates the blood into its components, collects one or more of the blood components, and returns the remainder of the blood components to the donor/patient. The instrument operates using pumps, clamps and valves that move donor/patient blood through a single-use, sterile fluid path disposable kit. The cells are centrifugally separated within the kit by density differences.

    The operator is responsible for connecting and monitoring the donor/patient and operating and monitoring the AMICUS separator during the procedure. The operator controls the separator through a touch screen. When necessary, the operator is warned of problems with messages on the screen and corresponding audible alarms.

    Once the cell separation is complete, the operator disconnects the donor/patient, dismantles the kit, and disposes of the kit in a safe manner. The kit is packaged in a recyclable plastic tray.

    AI/ML Overview

    The document provided describes a 510(k) premarket notification for the AMICUS Separator System, focusing on a software update (version 6.0). A 510(k) submission generally claims substantial equivalence to a predicate device, rather than providing extensive de novo clinical studies to establish new acceptance criteria and prove performance against those criteria in a traditional sense.

    Therefore, the information requested, which typically relates to clinical studies establishing performance against defined acceptance criteria for a novel device, is not directly available in this type of FDA submission.

    However, I can extract the information provided about the verification and validation activities for the software update and explain why some of the requested information is not applicable to a 510(k) for a software modification.

    Here's a breakdown based on the provided text:

    1. Table of acceptance criteria and the reported device performance:

    The document states that "Software and systems verifications were performed in support of this submission. The results of the testing were acceptable." However, it does not explicitly list specific acceptance criteria (e.g., in terms of sensitivity, specificity, accuracy for a diagnostic device, or specific performance metrics for a therapeutic device's outcome) and then report quantified performance against those criteria. This type of detailed breakdown is not typically required or provided in a 510(k) for a software enhancement to an already cleared device. The "acceptance criteria" here would be that the software performs its intended functions correctly and safely, as per the design specifications and referenced standards.

    The performance is implicitly reported as "acceptable" based on the verification and validation, implying that the software modification did not negatively impact the previously established performance of the AMICUS Separator System, and likely improved the listed enhancements.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    The document does not specify a "test set" in the context of clinical data for performance assessment. The "testing" mentioned refers to software and system verification, which would typically involve testing new functionalities, ensuring existing functionalities are preserved, and assessing system stability. The number of test cases run, or the number of simulated or actual apheresis procedures performed during verification, is not disclosed. Data provenance in this context would refer to the details of the testing environment rather than patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable. For a software update to an existing apheresis system, "ground truth" as established by medical experts (e.g., radiologists, pathologists) is not relevant in the same way it would be for a diagnostic AI device. Software verification focuses on functional correctness, safety, and adherence to design specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable for software verification. Adjudication methods are typically used in clinical studies where multiple experts assess cases and their opinions need to be reconciled to establish a ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is an automated blood cell separator, not a diagnostic AI device that assists human readers. No MRMC study would be performed for this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device (AMICUS Separator System) is inherently a "human-in-the-loop" system, as an operator connects and monitors the donor/patient and operates the separator. The software is part of this system. Therefore, the concept of "standalone" algorithm performance as typically applied to AI for diagnosis is not directly relevant here. The software verification assesses the behavior and functionality of the software within the context of the overall system and its intended use.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    Not applicable. The "ground truth" for software verification typically involves the design specifications and expected behavior of the software functions. For example, if a new feature is added to calculate a specific parameter, the "ground truth" is that the calculation is correct according to the mathematical formula and clinical parameters.

    8. The sample size for the training set:

    Not applicable. This is a software update for a control system, not a machine learning model that requires a training set of data.

    9. How the ground truth for the training set was established:

    Not applicable. See point 8.


    Summary of what is known from the document regarding meeting acceptance criteria:

    The document emphasizes that "Software and systems verifications were performed in support of this submission. The results of the testing were acceptable."

    The study (or rather, the verification and validation activities) that proves the device meets (implicitly defined) acceptance criteria is described as compliant with recognized standards and guidance documents:

    • IEC 62304 Ed. 1.0, "Medical device software Software life cycle processes": This standard outlines requirements for the software development life cycle of medical device software, ensuring safety and quality.
    • ISO 14971:2012 Medical Devices – Applications of Risk Management to Medical Devices: This standard specifies a process for a manufacturer to identify the hazards associated with medical devices, including in vitro diagnostic medical devices, to estimate and evaluate the associated risks, to control these risks, and to monitor the effectiveness of the controls.
    • FDA/CDRH "General Principles of Software Validation; Final Guidance for Industry and FDA Staff." Issued January 11, 2002: This guidance outlines the FDA's current thinking on general principles of software validation for medical devices.
    • ES 60601-1:2005/(R)2012 and A1:2012, C1: 2009/(R)2012 and A2:2010/(R)2012. Medical electrical equipment - Part 1: General requirements for basic safety and essential performance: This standard addresses the basic safety and essential performance requirements for medical electrical equipment.

    The conclusion states: "Based on the verification activities performed, the AMICUS Separator System modified with software 6.0 provides a device system that is substantially equivalent to the currently marketed AMICUS Separator System." This indicates that the software enhancements were verified to perform as intended and not introduce new safety or performance concerns, thus maintaining substantial equivalence to the predicate device.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    510k Summary Text (Full-text Search) :

    .

    21 CFR 864.9245 Automated Blood Cell Separator

    Automated blood cell separators which are based

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Depending on the AMICUS Separator System disposable used in the therapeutic apheresis procedure, the AMICUS Separator System has been cleared for the following:

    The AMICUS Separator System is an automated blood cell separator indicated to perform Therapeutic Plasma Exchange (TPE). (K111702)

    The AMICUS Exchange Kit is indicated for use in Therapeutic Plasma Exchange (TPE). The kit is for use with the AMICUS separator. (K111702)

    The AMICUS Separator System is an automated blood component separator indicated to perform Red Blood Cell Exchange (RBCX), including Exchange and Depletion/Exchange procedures, for the transfusion management of Sickle Cell Disease in adults and children. (K180615)

    The AMICUS Exchange Kit - Therapeutics is indicated for use in Therapeutic Plasma Exchange (TPE) and Red Blood Cell Exchange (RBCX). The kit is for use with the AMICUS separator. (K111702, K180615)

    The Waste Transfer Set is indicated for use in Red Blood Cell Exchange (RBCX). The set is for use with the AMICUS separator. (K180615)

    The Blood Component Filter Set with Vented Spike and Luer Adapter is indicated for the administration of blood and blood components during a Therapeutic Plasma Exchange (TPE) or Red Blood Cell Exchange (RBCX) therapeutic apheresis procedure. The set is for use with the AMICUS separator. (K111702, K180615)

    Device Description

    The AMICUS Separator System is comprised of the AMICUS separator instrument and a disposable apheresis kit specific to the procedure being performed. The instrument is a continuous-flow, centrifugal device that draws whole blood from a donor/patient, separates the blood into its components, collects one or more of the blood components, and returns the remainder of the blood components to the donor/patient. The instrument operates using pumps, clamps and valves that move donor/patient blood through a single-use, sterile fluid path disposable kit. The cells are centrifugally separated within the kit by density differences.

    The operator is responsible for connecting and monitoring the donor/patient and operating and monitoring the AMICUS separator during the procedure. The operator through a touch screen. When necessary, the operator is warned of problems with messages on the screen and corresponding audible alarms.

    Once the cell separation is complete, the operator disconnects the donor/patient, dismantles the kit, and disposes of the kit in a safe manner. The kit is packaged in a recyclable plastic tray.

    AI/ML Overview

    Acceptance Criteria and Device Performance for AMICUS Separator System (K180615)

    The AMICUS Separator System, specifically for Red Blood Cell Exchange (RBCX) procedures, underwent clinical studies to demonstrate its safety and effectiveness.

    1. Table of Acceptance Criteria and Reported Device Performance

    Study ParameterAcceptance Criteria (Target Range for 95% CI)Reported Device Performance (Mean ± SD, 95% CI)Study
    AMIC-003-CMD (RBC Exchange and Depletion/Exchange)
    Actual to Target (A:T) FCR ratio0.75 to 1.250.978 ± 0.1933 (0.927 to 1.028)AMIC-003-CMD
    Target End Hct AccuracyNot explicitly stated an acceptance range, but reported1.19 (0.817) %AMIC-003-CMD
    AMIC-004-CMD (RBC Depletion)
    Actual to Target (A:T) End Hct ratio0.85 to 1.151.0 ± 0.05 (0.95 to 0.98)AMIC-004-CMD

    2. Sample Size Used for the Test Set and Data Provenance

    • AMIC-003-CMD:
      • Sample Size: 59 evaluable procedures.
      • Data Provenance: Prospective, multi-site investigational study. The country of origin is not explicitly stated, but it is implied to be within the scope of where FDA clinical trials are conducted (e.g., USA).
    • AMIC-004-CMD:
      • Sample Size: 36 evaluable procedures.
      • Data Provenance: Prospective, single-site investigational study. The country of origin is not explicitly stated, but it is implied to be within the scope of where FDA clinical trials are conducted (e.g., USA).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The provided document describes clinical studies that directly measured device performance (e.g., FCR, End Hct) using laboratory analyses of patient blood samples. It does not mention the use of "experts" to establish a separate "ground truth" or reference standard for the device's analytical performance, as would be common in diagnostic imaging or clinical decision support AI. The ground truth for this device's performance metrics (FCR, hematocrit) would have been established through standard clinical laboratory procedures performed on patient samples, likely adhering to existing clinical guidelines.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method in the context of expert review for establishing ground truth, as the studies are focused on direct clinical performance measurements (e.g., FCR, hematocrit) from patient data. Clinical events (adverse events) would typically be reviewed by an independent safety committee or investigators.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The AMICUS Separator System is an automated blood processing device, and its performance is evaluated based on its technical output (e.g., FCR, hematocrit) in patients, not on human interpretation of data. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply to this device.

    6. Standalone Performance Study

    Yes, standalone performance studies were done. The clinical studies (AMIC-003-CMD and AMIC-004-CMD) evaluated the performance of the AMICUS Separator System as a standalone device in performing RBCX procedures. The results reported (e.g., A:T FCR ratio, A:T End Hct ratio) are direct measures of the algorithm's and system's performance.

    7. Type of Ground Truth Used

    The ground truth used was outcomes data and patient laboratory measurements.

    • For AMIC-003-CMD, the primary endpoint focused on the Actual FCR (Fraction of Cells Remaining) as measured by the subject's post-procedure Hb S, compared to the Target FCR. This is a direct laboratory measurement of the effectiveness of the exchange.
    • For AMIC-004-CMD, the primary objective was the Actual End Hct (hematocrit) post-procedure compared to the Target End Hct. This is also a direct laboratory measurement indicating the effectiveness of the depletion.
    • Additionally, safety was assessed by observing adverse events and changes in subject cellular losses (WBC and platelet counts), which also constitute clinical outcomes data.

    8. Sample Size for the Training Set

    The document does not explicitly mention a separate "training set" with a specified sample size for algorithm development. It refers to "internal in vitro studies using bagged blood" conducted during product development to assess performance and safety. These in vitro studies, along with system verification and validation activities, would have implicitly served some function akin to algorithm training and refinement, but a distinct "training set" with specific sample sizes from patients is not detailed as per AI/ML terminology. The clinical studies (AMIC-003-CMD and AMIC-004-CMD) are evaluation studies, acting as the test sets for the finalized device software.

    9. How the Ground Truth for the Training Set was Established

    As noted above, a distinct "training set" in the AI/ML sense is not explicitly described. However, the ground truth for the in vitro studies during development would have been established through:

    • Direct laboratory measurements: For parameters like FCR and hematocrit using established analytical methods from the bagged blood samples.
    • Comparison to predicate devices: The document states that these in vitro studies "concurrently collected data on the COBE Spectra Apheresis System for comparison. Overall study results showed that both devices are capable of meeting performance targets." This indicates that the performance of the predicate device served as a benchmark for establishing expected performance characteristics during the development phase.
    Ask a Question

    Ask a specific question about this device

    K Number
    K162423
    Date Cleared
    2017-01-03

    (126 days)

    Product Code
    Regulation Number
    868.5830
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    |
    | Regulation Number | 21 CFR 864.9245

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Haemonetics Cell Saver® Elite®+ Autotransfusion System and its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for gravity reinfusion into the patient or to the arterial line of an extracorporeal circuit for reinfusion into the patient.

    The intended use of the Sequestration Protocol is to collect an autologous, preoperative, platelet rich plasma product for reinfusion to the same patient within 6 hours of collection.

    The Cell Saver Elite/Elite+ System is intended to be used by trained physicians, operating room nurses or floor nurses, anesthesia technicians and autotransfusion service providers to provide intra-operative and post-operative blood salvage for surgical procedures with medium to high blood loss including, but not limited to CABG, AAA, joint replacement, spinal, trauma and transplant surgeries.

    Device Description

    The Cell Saver Elite/Elite+ System is intended to be used by trained physicians, operating room nurses or floor nurses, anesthesia technicians and autotransfusion service providers to provide intra-operative and post-operative blood salvage for surgical procedures with medium to high blood loss including, but not limited to CABG, AAA, joint replacement, spinal, trauma and transplant surgeries.

    The Cell Saver Elite/Elite+ System consists of a single use disposable set and reusable equipment. One disposable set is used throughout an individual patient's surgical procedure and then discarded. The Cell Saver Elite/Elite+ System utilizes a unique bowl processing kit, but is compatible with Haemonetics standard reservoirs and A&A lines.

    The collected blood is processed through a centrifugal separation chamber (bowl) where RBCs are concentrated and then washed, removing unwanted substances such as hemolized cells, anticoagulant and irrigating fluids. The washed RBC product is available for return via a product bag to the patient.

    The Elite/Elite+ System is designed to perform plasma sequestration using the autotransfusion disposable in conjunction with an ancillary sequestration set prior to performing autotransfusion.

    The subject of this Special 510(k) is the Haemonetics Cell Saver Elite/Elite+ Autotransfusion System software and hardware to enable use of wired and wireless connectivity.

    AI/ML Overview

    This document describes the regulatory approval (K162423) for the Haemonetics Cell Saver Elite/Elite+ Autotransfusion System with added wired and wireless connectivity features. The approval is based on demonstrating substantial equivalence to a predicate device (K160197).

    Here's an analysis of the provided information concerning acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission primarily focuses on the safety and performance of the added connectivity features and modifications to the user interface hardware. The acceptance criteria essentially revolve around demonstrating that these changes do not compromise the existing performance requirements of the autotransfusion system and comply with relevant standards.

    Acceptance Criteria (What was measured/verified)Reported Device Performance (Result)
    Electromagnetic Compatibility (EMC) compliance (per IEC 60601-1-2)Pass
    Electrical Safety compliance (per IEC 60601-1)Pass
    Wireless Coexistence of the Cell Saver Elite/Elite+ with potential interference appliancesPass
    Software Validation of Version AN of the CS Elite/Elite+ Software to ensure correct functionality and communication capabilitiesPass
    Maintenance of previous performance requirements despite UI hardware and software changes (implied)Subject device met all performance requirements, demonstrating substantial equivalence to the predicate.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the "sample size" in terms of number of devices or number of test cases run for each test. The non-clinical testing summary simply lists the tests conducted.

    The data provenance is from non-clinical testing performed by the manufacturer, Haemonetics Corporation, as part of their 510(k) submission. This is internal testing, not patient data, and is thus prospective in the sense that it was performed specifically for this submission. The country of origin for the data generation would be where Haemonetics conducted their internal R&D and testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable in the context of this 510(k) submission for device modifications. No "ground truth" based on expert medical opinion (like clinical diagnosis) was established or used for the technical performance tests (EMC, electrical safety, wireless coexistence, software validation). These tests rely on engineering standards and functional verification.

    4. Adjudication Method for the Test Set

    This is not applicable. The tests performed (EMC, electrical safety, wireless coexistence, software validation) are objective engineering and software verification tests, not subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This submission is for modifications to an autotransfusion system, a medical device that processes blood. It does not involve "human readers" or "AI assistance" in the context of interpreting medical images or making diagnostic decisions, which is typically where MRMC studies are conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is an autotransfusion system, not an algorithm performing a standalone diagnostic or interpretive function. The "software validation" was for the embedded software controlling the device's functionality and connectivity, which operates as part of the overall system with human operators.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical performance tests was based on engineering specifications, regulatory standards (e.g., IEC 60601-1-2, IEC 60601-1), and the defined functional requirements of the device. For example, for EMC, the ground truth is compliance with the specified limits in the standard. For wireless coexistence, the ground truth is the device operating without unacceptable interference. For software, the ground truth is the software performing as designed according to its requirements.

    8. The Sample Size for the Training Set

    This is not applicable. This device is not an AI/ML system that undergoes a "training phase" with a training set of data in the typical sense (e.g., for image recognition or predictive models). The software validation refers to the testing of the developed software against its requirements.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the reasons stated in point 8. The device's software is developed through standard software engineering practices and validated against pre-defined functional and performance requirements.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162462
    Date Cleared
    2016-11-23

    (82 days)

    Product Code
    Regulation Number
    864.9245
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    .

    21 CFR 864.9245 Automated Blood Cell Separator

    Automated blood cell separators which are based

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AMICUS Separator System is an automated blood cell separator indicated to perform Therapeutic Plasma Exchange (TPE).

    The AMICUS Separator System is an automated blood cell separator indicated for the collection of blood components and mononuclear cells.

    The device is designed to collect products while maintaining an extracorporeal volume at or below 10.5 mL/kg and a donor post platelet count greater than or equal to 100,000 platelets/microliter.

    Depending on the AMICUS Separator System apheresis kit used in the collection of products, the AMICUS Separator System has been cleared to collect:

    • · Platelets Pheresis, Leukocytes Reduced (single, double, or triple units)
    • · Platelets Pheresis, Leukocytes Reduced, Platelet Additive Solution (InterSol) (single, double or triple units)
    • · Red Blood Cells, Leukocytes Reduced (by apheresis)
    • Mononuclear Cells
    • Plasma
    • o Fresh Frozen Plasma
    • · Must be prepared and placed in a freezer at -18° C or colder within 8 hours after phlebotomy.
    • o Plasma Frozen Within 24 Hours After Phlebotomy (PF24)
    • · Must be stored at 1-6°C within 8 hours after phlebotomy and placed in a freezer at -18° C or colder within 24 hours after phlebotomy.
    • · Indicated for replacement of non-labile clotting factors. This product is not equivalent to Fresh Frozen Plasma.
    • o Plasma Frozen Within 24 Hours After Phlebotomy (PF24) Held at Room Temperature Up to 24 Hours After Phlebotomy (PF24RT24)
    • Can be stored at room temperature for up to 24 hours after phlebotomy. Product must be placed in a freezer at -18° C or colder within 24 hours after phlebotomy.
    • · Indicated for replacement of non-labile clotting factors. This product is not equivalent to Fresh Frozen Plasma.
    • o Source Plasma

    Platelet Pheresis (single, double, or triple units) may be manufactured from products that do not meet leukocyte reduction product standards. This does not apply to Platelet Additive Solution (InterSol) (single, double, or triple units).

    The AMICUS platelet storage container is cleared to store Platelets Pheresis, Leukocytes Reduced in 100% plasma for up to 7 days. Additionally, for platelet units stored past 5 days and through 7 days, every product must be tested with a bacterial device cleared by FDA and labeled as a "safety measure."

    NOTE - No changes to the AMICUS Separator System indications for use are requested in this 510(k) filing.

    Device Description

    The AMICUS Separator System is comprised of the AMICUS separator instrument and a disposable apheresis kit specific to the procedure being performed. The instrument is a continuous-flow, centrifugal device that draws whole blood from a donor/patient, separates the blood into its components, collects one or more of the blood components, and returns the remainder of the blood components to the donor/patient. The instrument operates using pumps, clamps and valves that move donor/patient blood through a single-use, sterile fluid path disposable kit. The cells are centrifugally separated within the kit by density differences.

    The operator is responsible for preparing and monitoring the donor/patient and operating and monitoring the AMICUS separator during the automatic blood collection cycle. The operator controls the separator through a touch screen. When necessary, the operator is warned of problems with messages on the screen and corresponding audible alarms.

    Once the cell separation is complete, the operator removes the needle(s) from the donor/patient, dismantles the kit, and disposes of the kit in a safe manner. The kit is packaged in a recyclable plastic tray.

    AI/ML Overview

    The Fresenius Kabi AMICUS Separator System, with software version 5.1, underwent performance testing to support its substantial equivalence claim. While the provided document states that "Software and systems verification were performed in support of this submission. The results of the testing were acceptable," it does not include a detailed table of acceptance criteria nor the reported device performance metrics from a formal study. It references that the changes were "procedure enhancements and correction of anomalies" and does not present data in the format of a clinical study with specific acceptance criteria and results.

    Therefore, the following information is extracted and synthesized based on the available text, with caveats for missing details.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a table with specific acceptance criteria or quantitative performance results from a study. It generally states that the "results of the testing were acceptable." The device's performance claims are tied to its indications for use, such as maintaining an extracorporeal volume at or below 10.5 mL/kg and a donor post platelet count greater than or equal to 100,000 platelets/microliter. However, no specific study data is presented to demonstrate meeting these criteria with software 5.1.

    Acceptance Criteria Category (Derived from Indications for Use)Stated Performance (Based on "acceptable results" and existing cleared claims)
    Extracorporeal volumeMaintained at or below 10.5 mL/kg
    Donor post platelet countMaintained greater than or equal to 100,000 platelets/microliter
    Collection of specified blood componentsAcceptable (implies meeting purity, quantity, and viability standards)
    Procedure enhancements and anomaly correctionsAcceptable (verified in software and systems verification)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample size for any test sets or the provenance of data (e.g., country of origin, retrospective or prospective) for the software and systems verification.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    The document does not mention the use of experts to establish ground truth or their qualifications for the testing of software version 5.1. The "software and systems verification" typically involves engineering and quality assurance personnel, not necessarily medical experts establishing ground truth in a clinical context.

    4. Adjudication Method for the Test Set

    No information is provided regarding any adjudication method, as the testing described appears to be software and system verification, not a clinical study requiring expert consensus or adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC study was not done or at least not reported in this document. The device is an automated blood cell separator, not an imaging or diagnostic AI requiring human-in-the-loop performance evaluation.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    The "Software and systems verification" serves as a standalone performance evaluation of the modified software. While the document does not label it as such, the nature of the device (automated blood cell separator) means its primary function is standalone operation without continuous human interpretation similar to an AI diagnostic tool. No specific performance metrics for this standalone evaluation are provided beyond "acceptable."

    7. Type of Ground Truth Used

    The concept of "ground truth" as typically used in AI/diagnostic studies (e.g., pathology, outcomes data, expert consensus) is not directly applicable here. For software and systems verification, the "ground truth" would be the pre-defined functional specifications and performance requirements for the software and the overall system. These would be established through engineering design, regulatory standards, and previous device clearances.

    8. Sample Size for the Training Set

    The document does not mention a training set. This is a software and system update for an existing medical device, not a machine learning model developed with training data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no mention of a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160197
    Date Cleared
    2016-05-24

    (118 days)

    Product Code
    Regulation Number
    868.5830
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    |
    | Regulation Number | 21 CFR 864.9245

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Haemonetics Cell Saver® Elite® Autotransfusion System and its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for gravity reinfusion into the patient or to the arterial line of an extracorporeal circuit for reinfusion into the patient.

    The intended use of the Sequestration Protocol is to collect an autologous, preoperative, platelet rich plasma product for reinfusion to the same patient within 6 hours of collection.

    The Cell Saver Elite System is intended to be used by trained physicians, operating room nurses or floor nurses, anesthesia technicians and autotransfusion service providers to provide intraoperative and post-operative blood salvage for surgical procedures with medium to high blood loss including, but not limited to CABG, AAA, joint replacement, spinal, trauma and transplant surgeries.

    Device Description

    The subject of this Special 510(k) is the Haemonetics Cell Saver Elite Autotransfusion System fat washing protocol and modified 70mL bowl algorithm.

    The Cell Saver Elite System consists of a single use disposable set and reusable equipment. One disposable set is used throughout an individual patient's surgical procedure and then discarded. The Cell Saver Elite System utilizes a unique bowl processing kit, but is compatible with Haemonetics standard reservoirs and A&A lines.

    The collected blood is processed through a centrifugal separation chamber (bowl) where RBCs are concentrated and then washed, removing unwanted substances such as hemolized cells, anticoagulant and irrigating fluids. The washed RBC product is available for return via a product bag to the patient.

    The Elite System is designed to perform plasma sequestration using the autotransfusion disposable in conjunction with an ancillary sequestration set prior to performing autotransfusion.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Haemonetics Cell Saver Elite Autotransfusion System, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance (for Fat Washing Protocol):

    Acceptance CriteriaReported Device Performance (Reference Device - Sorin Xtra)Reported Device Performance (Subject Device - Haemonetics Cell Saver Elite)
    HCT%≈ 50% (***)≥ 40%
    RBC RecoveryN/A (implied good hematocrit)≥ 80%
    Plasma HgB Washout>95% (***)≥ 95%
    Heparin Washout>95% (***)≥ 95%
    Albumin Washout>95% (***)≥ 95%
    Fat Removal>99% (***)≥ 99% (*)

    Note: (**) refers to data from the reference device Sorin Xtra, cited as "Fat removal during cell salvage - An optimized program in the XTRA® autotransfusion device" by Timo Seyfried, MD, Michael Gruber, MD; Lilith Haas; Emil Hansen, PhD, MD 13th ECOPEAT Vienna - Austria 2013.*
    Note: () for the subject device indicates "depending on bowl size used".*

    2. Sample Size and Data Provenance for the Test Set:

    • Sample Size: The document does not explicitly state the sample size used for the performance tests (functional testing and software validation). It lists several test reports by number (e.g., TR-SOF-100562, TR-OTH-100649), but these reports themselves are not included in the provided text.
    • Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective. It only states that "non-clinical performance testing was submitted."

    3. Number of Experts and Qualifications for Ground Truth:

    • This information is not provided in the document. The testing described appears to be primarily technical and functional validation against predefined performance metrics for a medical device rather than studies requiring expert human interpretation of medical images or patient data.

    4. Adjudication Method for the Test Set:

    • This information is not applicable and not provided. The testing described is against established performance requirements and is not a clinical study requiring adjudication of expert interpretations.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study is mentioned. The submission is for a modification to an existing autotransfusion system, focused on technical performance improvements rather than reader interpretation.

    6. Standalone Performance Study (Algorithm Only):

    • The document implies standalone performance studies were conducted for the device's functionality. Table 1 "Summary of Performance Studies" lists "Software Validation" and "Functional Testing" with corresponding report numbers and "Test Intent" that demonstrate the device (or its software components and new features like fat washing protocol and modified 70mL bowl algorithm) met performance requirements. The results are simply "Passed," indicating the algorithm's performance in achieving the specified criteria.

    7. Type of Ground Truth Used:

    • The ground truth used for the performance studies appears to be based on objective, quantitative measurements related to blood processing parameters. For example:
      • Hematocrit (HCT%)
      • Red Blood Cell (RBC) Recovery
      • Plasma Hemoglobin (HgB) Washout
      • Heparin Washout
      • Albumin Washout
      • Fat Removal
    • These are physical and chemical properties of blood that can be measured directly by laboratory methods, establishing a clear objective ground truth for the device's performance in processing blood.

    8. Sample Size for the Training Set:

    • The document does not provide details about a "training set" as this device is not an AI/ML algorithm that typically requires a large training dataset in the same way. The mentioned "Software Validation" and "Functional Testing" refer to verification and validation activities for the device's software and hardware, where "training" in the context of machine learning is not directly applicable.

    9. How Ground Truth for the Training Set Was Established:

    • As mentioned above, the concept of a "training set" and its associated ground truth establishment is not relevant to this type of device submission based on the provided information. The validation focuses on ensuring the device meets pre-defined performance specifications for blood processing.
    Ask a Question

    Ask a specific question about this device

    K Number
    K141019
    Manufacturer
    Date Cleared
    2014-06-10

    (50 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    a regulation by the Center for Devices and Radiological Health due to pre-amendment status.

    21 CFR 864.9245

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AMICUS Separator System is an automated blood cell separator intended for use in therapeutic apheresis applications and may be used to perform Therapeutic Plasma Exchange (TPE).

    The AMICUS Separator System is an automated blood cell separator intended for use in the collection of blood components and mononuclear cells.

    The AMICUS Separator System is an automated blood cell separator indicated to perform Therapeutic Plasma Exchange (TPE).

    The AMICUS Separator System is an automated blood cell separator indicated for the collection of blood components and mononuclear cells.

    The device is designed to collect products while maintaining an extracorporeal volume at or below 10.5 ml/kg and a donor post platelet count greater than or equal to 100,000 platelets/microliter.

    Depending on the AMICUS Separator System apheresis kit used in the collection of products, the AMICUS Separator System has been cleared to collect:

    • Platelets Pheresis, Leukocytes Reduced (single, double, or triple units) .
    • Platelets Pheresis, Leukocytes Reduced, Platelet Additive Solution (InterSol) (single, double . or triple units)
    • Red Blood Cells, Leukocytes Reduced (by apheresis) .
    • . Mononuclear Cells
    • Plasma .
      • o Fresh Frozen Plasma
        • . Must be prepared and placed in a freezer at -18° C or colder within 8 hours after phlebotomy.
      • o Plasma Frozen Within 24 Hours After Phlebotomy (PF24)
        • Must be stored at 1-6°C within 8 hours after phlebotomy and placed . in a freezer at -18° C or colder within 24 hours after phlebotomy.
        • Indicated for replacement of non-labile clotting factors. This product . is not equivalent to Fresh Frozen Plasma.
      • o Plasma Frozen Within 24 Hours After Phlebotomy (PF24) Held at Room Temperature Up to 24 Hours After Phlebotomy (PF24RT24)
    • . Can be stored at room temperature for up to 24 hours after phlebotomy. Product must be placed in a freezer at -18° C or colder within 24 hours after phlebotomy.
    • Indicated for replacement of non-labile clotting factors. This product . is not equivalent to Fresh Frozen Plasma.
    • o Source Plasma

    Platelet Pheresis (single, double, or triple units) may be manufactured from products that do not meet leukocyte reduction product standards. This does not apply to Platelet Pheresis, Platelet Additive Solution (InterSol) (single, double, or triple units).

    Device Description

    The AMICUS Separator System is comprised of the AMICUS separator instrument and a disposable apheresis kit specific to the procedure being performed. The instrument is a continuous-flow, centrifugal device that draws whole blood from a donor/patient, separates the blood into its components, collects one or more of the blood components, and returns the remainder of the blood components to the donor/patient. The instrument operates using pumps, clamps and valves that move donor/patient blood through a single-use, sterile fluid path disposable kit. The cells are centrifugally separated within the kit by density differences.

    The operator is responsible for preparing and monitoring the donor/patient and operating and monitoring the AMICUS separator during the automatic blood collection cycle. The operator controls the separator through a touch screen. When necessary, the operator is warned of problems with messages on the screen and corresponding audible alarms.

    Once the cell separation is complete, the operator removes the needle(s) from the donor/patient, dismantles the kit, and disposes of the kit in a safe manner. The kit is packaged in a recyclable plastic tray.

    AI/ML Overview

    This document describes a 510(k) premarket notification for a software update (version 4.5) to the AMICUS Separator System, an automated blood cell separator. The purpose of this submission is to demonstrate substantial equivalence to the currently marketed AMICUS Separator System.

    Here's an analysis of the provided information, focusing on acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state numerical acceptance criteria in a dedicated table format with corresponding performance results. Instead, it broadly mentions that "Software verification, systems verification and systems validation were performed in support of this submission. The results of the testing were acceptable."

    The core acceptance criterion implicitly stated and tested is substantial equivalence to the predicate device, especially regarding the described enhancements in software version 4.5. These enhancements primarily focus on:

    • Automated custom prime in Mononuclear (MNC) and Therapeutic Plasma Exchange (TPE) procedures to maintain isovolemia.
    • Functionality to pump saline during TPE procedures.
    • Additional input parameters and output values (e.g., Product Volume and ACD in Product for MNC, in line with FACT requirements).
    • Additional instruction screens and other minor enhancements.

    The "reported device performance" is summarized as:

    Acceptance Criteria (Implied)Reported Device Performance
    Functional equivalence of software version 4.5 features: - Automated custom prime for isovolemia management (MNC, TPE)All new functionalities of software 4.5 perform as intended and designed.
    - Saline pumping functionality during TPE
    - Accuracy/Correctness of new input parameters and output values (e.g., FACT-aligned MNC data)
    - Improvements in operator instructions/user interface
    Maintenance of existing safety and efficacy of the AMICUS Separator System: (e.g., extracorporeal volume, post platelet count)Original safety and efficacy profiles of the AMICUS Separator System are maintained with software 4.5.
    Technological Characteristics: Centrifuge system, fluid control system, safety management, anticoagulant management, physical design, apheresis kits, data management capabilities.Technological characteristics remain the same as the predicate AMICUS device.

    The conclusion explicitly states: "Based on the validation and verification activities performed, the AMICUS Separator System modified with software 4.5 provides a device system that is substantially equivalent to the currently marketed AMICUS Separator System." This implies all these implicit criteria were met.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document provides very limited detail on the specifics of the test set, sample size, or data provenance. It only states: "Software verification, systems verification and systems validation were performed in support of this submission."

    • Sample Size for Test Set: Not specified.
    • Data Provenance: Not specified (e.g., country of origin, retrospective/prospective). Given this is a software update for an existing device, it's likely that internal testing, possibly using simulated data or real-world operational data from in-house labs or clinical sites, was conducted.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable in the context of this submission. The device is an automated blood cell separator, and the submission concerns a software update. "Ground truth" in this context would typically refer to the correct functioning of the software according to specifications, and the accuracy of its calculations and controls. This would be established through engineering and software testing protocols, not by expert consensus on clinical findings.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This type of adjudication method is used in studies where human interpretation (e.g., image reading) requires consensus, which is not the nature of the validation described. Software and system validation typically involve comparing system outputs against predefined specifications and expected results, not human adjudication of subjective interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The AMICUS Separator System is an automated blood processing device, not an AI-assisted diagnostic or interpretive tool for human readers. Therefore, an MRMC study related to human reading performance with or without AI assistance is not relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, implicitly. The "Software verification, systems verification and systems validation" would involve extensive testing of the algorithm (software) in various scenarios, both in isolation ("standalone") and within the full system, to ensure it performs according to its design specifications. While a specific "standalone" study isn't detailed, the nature of software and system validation inherently includes rigorous testing of the automated functions without direct human intervention impacting the internal logic or calculations. The human "operator" monitors the system, but the core function of separation and collection is automated by the device's algorithms.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the software verification and validation, the "ground truth" would be the pre-defined design specifications and expected outputs of the software and system. For example:

    • For the custom prime function: The ground truth would be that the system accurately calculates and delivers the prescribed priming fluid volume while maintaining isovolemia as specified in the design requirements.
    • For saline pumping: The ground truth would be precise delivery of saline according to operator input and system control.
    • For new output parameters (e.g., MNC Product Volume, ACD in Product): The ground truth would be the accurate calculation and display of these values based on the internal measurements and algorithms, consistent with FACT requirements where applicable.

    This "ground truth" is established by the engineering and scientific principles governing the device's operation and the regulatory/clinical standards it must meet (e.g., FACT requirements mentioned for MNC output values).

    8. The sample size for the training set

    Not applicable. This submission is for a software update to an automated device, not a machine learning or AI-driven system that would typically require a "training set" in the context of learning algorithms. The software's logic is explicitly programmed based on engineering principles and clinical requirements, not "trained" on a dataset.

    9. How the ground truth for the training set was established

    Not applicable, as there is no "training set" in the machine learning sense. The "ground truth" (i.e., correct behavior and outputs) for the software was established through engineering design, scientific principles, and adherence to specific performance requirements and regulatory standards for blood processing devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K111702
    Manufacturer
    Date Cleared
    2012-03-22

    (279 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Principle) Automated Separator, Blood Cell and Plasma, Therapeutic*

    Classification Name:

    21 CFR 864.9245

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AMICUS Separator System is an automated blood cell separator indicated for the collection of blood components and mononuclear cells.

    The AMICUS Separator System is an automated blood cell separator indicated to perform Therapeutic Plasma Exchange (TPE).

    The Blood Component Filter Set with Vented Spike and Luer Adapter is indicated for the administration of blood and blood components during a therapeutic plasma exchange procedure. For use with the AMICUS Separator System.

    Device Description

    The AMICUS separator is a continuous-flow, centrifugal device that separates whole blood into its components. Blood components are collected using sterile fluid path, single-use, apheresis kits. The cells are centrifugally separated within the kit by density differences.

    The Therapeutic Plasma Exchange (TPE) procedure is an apheresis collection procedure coupled with an infusion procedure. The collected plasma, which contains most of the anticoagulant, is pumped to a waste container(s). The patient's WBCs, the majority of platelets and a small amount of plasma and anticoagulant are mixed with a physician-prescribed replacement fluid, such as FFP, human albumin and/or crystalloid solutions, and returned to the patient. This process continues until the target plasma volume has been removed and replaced per physician order.

    The operator is responsible for preparing and monitoring the patient as well as operating and monitoring the AMICUS separator during the TPE procedure. The operator controls the separator through a touch screen. When necessary, the operator is alerted to problems, or given notes via messages displayed on the screen with corresponding audible alarms. Once complete, the operator disconnects the donor/patient, removes the kit, and disposes of the kit per institutional SOPs. The kit is packaged in a recyclable plastic tray.

    The Blood Component Filter Set with Vented Spike and Luer Adapter is an optional accessory intended to be used in conjunction with the AMICUS Separator System in Therapeutic Plasma Exchange (TPE) procedures to administer physician prescribed replacement fluids during the TPE procedure, when blood component filtration is needed (e.g. when using fresh frozen plasma). The filter set has a common blood transfusion set design and is comprised of a blood component filter, tubing, vented spike and a luer lock/cap. The filter set is a single use, sterile fluid path intravascular administration set. The filter set is connected to the TPE kit via the male luer connector and the vented spike is used to access the prescribed replacement fluid container. The flow rate through the filter set is controlled by the AMCUS Separator System.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    AMICUS® Separator System - Use in Therapeutic Plasma Exchange

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Primary Study Objective)Reported Device Performance
    Efficiency of plasma removal in the AMICUS separator (Test group) statistically non-inferior to the Control group at a margin of 15%.Mean efficiency of plasma removal for the test procedures was 81.9% (range 68% to 96%).
    This was statistically non-inferior to the control group and, in fact, statistically superior to the control group.
    (Implicit) Accurate maintenance of fluid balance.Accurately maintaining fluid balance.
    (Implicit) Assessment of platelet loss in waste plasma within the range of the control device.Only low levels of platelets in the waste plasma, within the range of the control device.
    (Implicit) Plasma hemoglobin in waste plasma low for AMICUS device and similar to control procedures.Plasma hemoglobin in the waste plasma was low for the AMICUS device and similar to that of the control procedures.

    2. Sample Size and Data Provenance

    • Test Set Sample Size: Not explicitly stated as a number of patients or procedures. The text refers to "test procedures" and "control procedures" as groups within a clinical evaluation. It doesn't break down the exact number of participants in each.
    • Data Provenance: The study was a "clinical evaluation," suggesting a prospective study. The country of origin of the data is not specified in the provided text.

    3. Number of Experts and Qualifications for Ground Truth

    This information is not provided in the text. The study evaluates the device's technical performance (plasma removal efficiency, fluid balance, etc.) rather than diagnostic accuracy requiring expert consensus on images or interpretations.

    4. Adjudication Method for the Test Set

    This information is not applicable as the study focused on objective performance metrics of a medical device, not on diagnostic interpretations requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A MRMC comparative effectiveness study was not conducted. This study evaluates the performance of a medical device (AMICUS Separator System) in performing a therapeutic procedure (Therapeutic Plasma Exchange), not the interpretive skills of human readers.

    6. Standalone Performance

    Yes, a standalone performance study (clinical evaluation) was done. The study specifically compared the AMICUS Separator System's performance (Test group) against a "Control group" (which, based on context, appears to be the predicate device, COBE Spectra). The reported metrics (plasma removal efficiency, fluid balance, platelet loss, plasma hemoglobin) directly reflect the algorithm's (device's) performance.

    7. Type of Ground Truth Used

    The "ground truth" for this study was objective clinical performance metrics measured during the Therapeutic Plasma Exchange procedures. These include:

    • Efficiency of plasma removal (quantified percentage).
    • Fluid balance (accuracy of maintenance).
    • Levels of platelet loss in waste plasma.
    • Levels of plasma hemoglobin in waste plasma.

    8. Sample Size for the Training Set

    This information is not applicable/not provided. The AMICUS Separator System is a medical device, not an AI/ML algorithm that is "trained" on a dataset in the typical sense. Its operational parameters and software are developed through engineering and verification/validation processes, not machine learning training.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable/not provided for the same reason as point 8. The device's operation is based on established engineering principles and extensive system validation and verification activities, rather than a machine learning training paradigm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1